K Number
K023947

Validate with FDA (Live)

Device Name
GC OIA
Date Cleared
2003-04-17

(141 days)

Product Code
Regulation Number
866.3390
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Thermo BioStar® GC OIA assay is an Optical ImmunoAssay test for the rapid qualitative detection of gonococcal antigen (L7/L12 ribosomal protein) in female endocervical swab and male urine specimens. Urine specimens must be prepared using an accessory Urine Filtration Device (UFD) for concentration and extraction. This test is intended for in vitro diagnostic use as an aid in identifying the presence of Neisseria gonorrhoeae antigen. The assay is intended for use with symptomatic females and males, in populations at risk for sexually transmitted diseases.

Device Description

The GC OIA test involves the qualitative extraction and detection of an antigen unique to N. gonorrhoeae. The Optical ImmunoAssay technology enables the direct visual detection of a physical change in the optical thickness of molecular thin films. This change is a result of antigen-antibody binding on an optical surface (silicon wafer). When an extracted specimen is placed directly on the optical surface, the immobilized specific antibodies capture the antibody-enzyme conjugate is then added for form an immune complex "sandwich" of immobilized antibody-sample-antibody HRP on the surface. After washing, the substrate is added, increasing the thickness (mass enhancement) of the molecular thin film. This change in thickness alters the reflected light path and is visually perceived as a color change. Slight changes in optical thickness produce a distinct, visible color change. A positive result appears as a purple spot on the predominant gold background. When antigen is not present in the specimen, no binding takes place. Therefore, the optical thickness remains unchanged and the surface retains the original gold color indicating a negative result.

AI/ML Overview

Here's a summary of the acceptance criteria and study details for the GC OIA® device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

MetricAcceptance Criteria (Implied)Reported Device Performance (Female Endocervical Swabs)Reported Device Performance (Male Urine)
SensitivityNot explicitly stated70.7%93.2%
SpecificityNot explicitly stated99.4%97.5%
Reproducibility (Urine)Not explicitly statedNot applicable98.1% (95% CI = 93.8-99.3)
Reproducibility (Swab)Not explicitly stated88.9% (95% CI = 93.8-99.3)Not applicable
PPV (Positive Predictive Value)Not explicitly stated90.6%94.0%
NPV (Negative Predictive Value)Not explicitly stated97.5%97.2%

Note: The document does not explicitly state numerical acceptance criteria for sensitivity, specificity, or reproducibility. The reported performance is presented as the outcome of the clinical study, implying these values were deemed acceptable for substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: A total of 904 valid results were obtained from symptomatic patient specimens.
  • Data Provenance: The study was a "multicenter study" conducted at "four geographically diverse clinical sites." The data is prospective as it was collected from "symptomatic patient specimens" for the purpose of the study.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the given text. It mentions a "commercial culture media, with secondary confirmation testing by LCR (Ligase Chain Reaction)" as the comparison method for establishing ground truth, but does not detail the involvement or qualifications of human experts in this process.

4. Adjudication Method for the Test Set

This information is not provided in the given text. The ground truth was established by "commercial culture media, with secondary confirmation testing by LCR," but the specific adjudication method if there were discrepancies between these methods or if human experts were involved in adjudication is not described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study compared the device with a "historical reference method" (Neisseria gonorrhoeae culture) and used secondary confirmation by LCR, not human readers with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, a standalone study was done. The performance data (sensitivity, specificity, etc.) are reported for the GC OIA® assay itself, functioning without human interpretation affecting the result. The device is described as enabling "direct visual detection of a physical change...visually perceived as a color change," where a "positive result appears as a purple spot." This implies the human component is simply observing the color change, not interpreting complex images or performing diagnostic reasoning.

7. The Type of Ground Truth Used

The ground truth used was a combination of:

  • Culture: "Commercial culture media" was the primary comparison method.
  • Molecular Confirmation: "Secondary confirmation testing by LCR (Ligase Chain Reaction)" was used. This indicates a robust ground truth standard, likely aiming to resolve ambiguous culture results or provide higher certainty.

8. The Sample Size for the Training Set

The sample size for the training set is not explicitly stated in the provided text. The document describes clinical studies that established "performance characteristics," implying these involved a test set. There's no separate mention of a distinct training set.

9. How the Ground Truth for the Training Set Was Established

Since a separate training set is not explicitly mentioned, the method for establishing its ground truth is also not provided. If the reported clinical study data involved an internal training phase, the ground truth would presumably have been established similarly to the test set (culture with LCR confirmation).

{0}------------------------------------------------

510(k) SUMMARY (page 1 of 3) 8.0

APR 1 7 ZUU3

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K023947

A. Submitter's information required per [§807.92(a)(1)]:

  • SUBMITTER'S NAME: Thermo BioStar, Inc. .
  • ADDRESS: 331 South 104th Street, Louisville CO 800271 ●
  • TELEPHONE: (303) 530-3888 ext. 612
  • FAX: (303) 581-6405
  • CONTACT PERSON: John G. Adams .
  • DATE 510(k) SUMMARY PREPARED: April, 2003 ●

B. Device information required per [§807.92(a)(2)]:

  • TRADE OR PROPRIETARY NAME: GC OIA®
  • COMMON NAME: Neisseria gonorrhoeae antigen assay
  • CLASSIFICATION NAME: Antigen, Enzyme Linked Immunoabsorbent Assay, Neisseria gonorrhoeae ●
  • C. Identification of legally marketed device to which we are comparing performance.

Historical Reference Method

  • Neisseria gonorrhoeae culture .
    Device Technology:
Trade or Proprietary Name:Chlamydia OIA Assay
Regulatory Class:I
Manufacturer:Thermo BioStar
510(k) Number:K951010
  • D. Intended use of device [§807.92(a)(5)]:
    The Thermo BioStar® GC OIA assay is an Optical ImmunoAssay test for the rapid qualitative detection of gonococcal antigen (L7/L12 ribosomal protein) in female endocervical swab and male urine specimens. Urine specimens must be prepared using an accessory Urine Filtration Device (UFD) for concentration and extraction. This test is intended for in vitro diagnostic use as an aid in identifying the presence of Neisseria gonorrhoeae antigen. The assay is intended for use with symptomatic females and males, in populations at risk for sexually transmitted diseases.

  • E. Description of device [§807.92(a)(4)]:
    Principle of the Test:

The GC OIA test involves the qualitative extraction and detection of an antigen unique to N. gonorrhoeae. The Optical ImmunoAssay technology enables the direct visual detection of a physical change in the optical thickness of molecular thin films. This change is a result of antigen-antibody binding on an optical surface (silicon wafer). When an extracted specimen is placed directly on the optical surface, the immobilized specific antibodies capture the antibody-enzyme conjugate is then added for form an immune complex "sandwich" of immobilized antibody-sample-antibody HRP on the surface. After washing, the

{1}------------------------------------------------

8.0 510(k) SUMMARY (page 2 of 3)

substrate is added, increasing the thickness (mass enhancement) of the molecular thin film. This change in thickness alters the reflected light path and is visually perceived as a color change. Slight changes in optical thickness produce a distinct, visible color change. A positive result appears as a purple spot on the predominant gold background. When antigen is not present in the specimen, no binding takes place. Therefore, the optical thickness remains unchanged and the surface retains the original gold color indicating a negative result.

Device Comparison [§807.92(a)(6)]: F.

Comparison to Historical reference method

The GC OIA assay is similar to culture methods in that:

  • Both assays are used to detect and identify N. gonorrhoeae ; ●
  • . Both assays detect N. gonorrhoeae in endocervical swabs and male urine specimens.

The GC OIA assay differs from traditional culture methods in that:

  • GC OIA assay detects an antigen unique to the N. gonorrhoeae organism, while the traditional . culture methods detect the whole living organism;
  • The GC OIA assay can provide results in less than 30 minutes, in contrast to culture methods that ● can take up to 72 hours.

Comparison to existing device technology

The GC OIA assay is similar to the Chlamydia OIA assay in that:

  • Both assays have the same Optical ImmunoAssay (OIA) technology; ●
  • . Both assays detect antigen in endocervical swab specimens;
  • . The intended use for both assays is to evaluate symptomatic patients for a sexually transmitted disease infection;
  • Both assays are qualitative;
  • . Both assays utilize liquid reagents

The GC OIA assay differs with the Chlamydia assay in that:

  • GC OIA is testing for the presence of a different STD, Neisseria gonorrhoeae;
  • . GC OIA detects antigens in symptomatic male urine specimens, in addition to endocervical swabs

SUMMARY OF PERFORMANCE DATA:

CLINICAL STUDIES

Performance characteristics for the GC OIA assay were initially established in a multicenter study at four geographically diverse clinical sites.

  • G. Summary of clinical testing [§807.92(b)(2)]:

Reproducibility

Urine: Reproducibility testing was conducted at four clinical sites and two point of care settings including physician offices. Urine testing was performed with 2 mL samples on three days with nine blinded specimens each day. This testing resulted in an overall reproducibility of 98.1% (95%CI = 93.8-99.3).

Swab: Reproducibility testing was conducted at three physician office laboratories. Swab testing was performed on three days with nine blinded samples each day. The negative swabs contained no GC cells. Overall reproducibility of the testing using the swab panels was 88.9% at the three sites. (95% CI = 93.8-99.3)

{2}------------------------------------------------

8.0 510(k) SUMMARY (page 3 of 3)

Clinical Sensitivity and Specificity

A study comparing the GC OLA test to commercial culture media, with secondary confirmation testing by LCR (Ligase Chain Reaction). A total of 904 valid results were obtained from symptomatic patient specimens collected at four clinical trial locations. Sensitivity and specificity for symptomatic male unnes was 93.2 % and 97.5 % respectively. Sensitivity and specificity for symptomatic female endocervical swabs was 70.7% and 99.4% respectively. Overall PPV for males were 94.0 % and 97.2 % respectively; and for females were 90.6% and 97.5% respectively.

  • H. Conclusions from nonclinical / clinical testing [§807.92(b)(3)]:
    The results of the above described internal and external studies demonstrate that the GC OIA test is as safe and effective as the comparative device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES U.S.A." arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

APR 1 7 2003

Mr. John G. Adams Regulatory Affairs Manager Thermo BioStar, Inc. 331 South 104th Street Louisville, CO 80027

K023947 Re: Trade/Device Name: GC OIA® Test Kit Regulation Number: 21 CFR 866.3390 Regulation Name: Neisseria spp. Direct Serological Test Reagents Regulatory Class: Class II Product Code: LIR Dated: February 18, 2003 Received: February 19, 2003

Dear Mr. Adams:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE STATEMENT 12.0

INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K 023947

GC OIA® Device Name:

Indications For Use:

The Thermo BioStar® GC OIA assay is an Optical ImmunoAssay test for the rapid qualitative detection of gonococcal antigen (L7/L12 ribosomal protein) in female endocervical swab and male urine specimens. Urine specimens must be prepared using an accessory Urine Filtration Device (UFD) for concentration and extraction. This test is intended for in vitro diagnostic use as an aid in identifying the presence of Neisseria gonorrhoeae antigen. The assay is intended for use with symptomatic females and males, in populations at risk for sexually transmitted diseases.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use the text is the following:

Division of Clinical Laboratory Devices

510(k) Number K023947

§ 866.3390

Neisseria spp. direct serological test reagents.(a)
Identification. Neisseria spp. direct serological test reagents are devices that consist of antigens and antisera used in serological tests to identifyNeisseria spp. from cultured isolates. Additionally, some of these reagents consist ofNeisseria spp. antisera conjugated with a fluorescent dye (immunofluorescent reagents) which may be used to detect the presence ofNeisseria spp. directly from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genusNeisseria, such as epidemic cerebrospinal meningitis, meningococcal disease, and gonorrhea, and also provides epidemiological information on diseases caused by these microorganisms. The device does not include products for the detection of gonorrhea in humans by indirect methods, such as detection of antibodies or of oxidase produced by gonococcal organisms.(b)
Classification. Class II (performance standards).