The Advantage Multiple Drugs of Abuse Test is a qualitative, one step, immunochromatographic competitive assay used to screen human urine for the presence of one to five of the following Drugs of Abuse at the indicated cut off: THC 50ng/mL, Phencyclidine (PCP) 25ng/mL, Opiates 2000 ng/mL, Cocaine 300ng/mL, Amphetamine 1,000ng/mL and Methamphetamine 1,000ng/mL. The tests are formatted in multiple plastic test cards, which hold up to 5 drug test strips, and single plastic test cards which hold one drug test strip. The test is qualitative and provides only a preliminary analytical result, which must be confirmed by an alternate methodology preferably, GC/MS. The test is for professional use.
Device Story
Advantage Multiple Drugs of Abuse Test is an in vitro immunochromatographic competitive assay. Device utilizes plastic test cards containing 1 to 5 drug test strips. Input is human urine sample. Principle of operation is lateral flow immunoassay; presence of drugs at specified cut-offs produces qualitative visual result on test strip. Intended for professional use in clinical settings. Output is preliminary analytical result; requires confirmation by GC/MS or similar methodology. Benefits include rapid screening for common drugs of abuse to inform clinical decision-making.
Clinical Evidence
Bench testing only. Performance evaluated by comparing results to GC/MS and EMIT testing. Individual drug tests demonstrated 90-96% correlation with GC/MS and >99% correlation in drug-free urine compared to EMIT testing.
Technological Characteristics
Immunochromatographic competitive assay. Formatted as single or multiple (up to 5) drug test strips housed in plastic cards. Qualitative visual readout. No electronic components or software.
Indications for Use
Indicated for professional use to screen human urine for presence of THC, PCP, Opiates, Cocaine, Amphetamine, and Methamphetamine at specified cut-off concentrations. Provides preliminary analytical results requiring confirmation by alternate methodology (e.g., GC/MS).
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Reference Devices
GC/MS
EMIT testing
Related Devices
K021006 — STARTOX DRUG OF ABUSE SCREENING TEST · Starplex Scientific, Inc. · Jun 10, 2002
K063015 — AMEDITECH IMMUTEST MULTI-DRUG SCREEN BUP/PPX/COC150 · Ameditech, Inc. · Dec 20, 2006
K163704 — Healgen Multi-Drug Urine Test Cup; Healgen Multi-Drug Urine Test Dip Card · Healgen Scientific,, LLC · Aug 2, 2017
K061718 — INNOVACON SPECTRUM II TEST CARD · Innovacon, Inc. · Nov 13, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
## JAN 31 2003
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR Part 807.92.
The assigned 510(k) number is:_K033712-
1. Date of summary: October 30, 2002
- 2. Submitted by: Advantage Diagnostics Corporation 764 San Aleso Avenue Sunnyvale, CA 94085
Advantage Multiple Drugs of Abuse Test. 3. Device Name:
4. Device Classification: CFR 21 862.3870. 862.3250. 8610.3610. Class II. Panel 91 Toxicology
5. Device description: The Advantage Multiple Drugs of Abuse Test is an immunochromatographic based one step in vitro test.
6. Intended Use: The Advantage Multiple Drugs of Abuse Test is a qualitative, one step immunochromatographic competitive assay used to screen human urine for the presence of the six commonly abused drugs at cut off concentrations of 50 ng/mL THC. 300ng/mL benzoylecgonine/cocaine, 2,000ng/mL opiates, 1,000ng/mL methamphetamine 1,000ng/mL amphetamine and 25ng/mLPCP. The tests are formatted in a multiple plastic card, which holds up to 5 drug test strips, and a single plastic test card, which holds a single drug test strip. The test is qualitative and provides only a preliminary analytical result, which must be confirmed by an alternate methodology preferably. GC/MS.
7. Substantial Equivalence: The Advantage Multible Drugs of Abuse Test was found substantially equivalent to GC/MS and EMIT testing. The individual Advantage drug tests demonstrated correlations between 90-96% as compared to GC/MS and >99% in drug free urine as compared to EMIT testing.
## Conclusion:
The Advantage Multiple Drugs of Abuse Test is substantially equivalent in performance characteristics to GC/MS testing and EMIT testing.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Janis Freestone Director. Regulatory Affairs Advantage Diagnostics Corporation 764 San Aleso Avenue Sunnyvale, CA 94085
JAN 31 2003
k023712 Trade/Device Name: Advantage Multiple Drugs of Abuse Test Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Code: DKZ; LDJ; DIO; DJC; DJG; LCM Dated: January 7, 2003 Received: January 9, 2003
Dear Ms. Freestone:
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{2}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Dutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510k Number:
Device Name: Advantage Multiple Drugs of Abuse Test
Indications for Use:
The Advantage Multiple Drugs of Abuse Test is a qualitative, one step, immunochromatographic competitive assay used to screen human urine for the presence of one to five of the following Drugs of Abuse at the indicated cut off: THC 50ng/mL, Phencyclidine (PCP) 25ng/mL, Opiates 2000 ng/mL, Cocaine 300ng/mL, Amphetamine 1,000ng/mL and Methamphetamine 1,000ng/mL. The tests are formatted in multiple plastic test cards, which hold up to 5 drug test strips, and single plastic test cards which hold one drug test strip. The test is qualitative and provides only a preliminary analytical result, which must be confirmed by an alternate methodology preferably, GC/MS. The test is for professional use.
Sean Cooper
Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number .
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over the counter use
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.