K Number
K013247

Validate with FDA (Live)

Device Name
CANNABINOIDS
Date Cleared
2002-03-13

(166 days)

Product Code
Regulation Number
862.3870
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cannabinoids assay is used for the qualitative analysis of cannabinoids in human urine with a cutoff of 50 ng/mL for use in clinical laboratories. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoids use or overdose.

The Cannabinoids assay is calibrated with 11-nor-Δ'-THC-9-COOH and will detect a variety of -Δ'-THC metabolites.

The Cannabinoids assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Device Description

The Cannabinoids assay is an in vitro diagnostic assay for the qualitative analysis of cannabinoids in human urine. The assay is a homogeneous enzyme immunoassay with a 50 ng/mL cutoff. The assay is based on competition between drug in the specimen and drug labeled with the enzyme glucose-6phosphate dehydrogenase (G6PDH) for antibody binding sites. Enzyme activity decreases upon binding to the antibody, so the drug concentration in the specimen can be measured in terms of enzyme activity. Active enzyme converts NAD to NADH, resulting in an absorbance change that can be measured spectrophotometrically.

AI/ML Overview

This 510(k) summary describes a new Cannabinoids assay for qualitative analysis of cannabinoids in human urine, using a 50 ng/mL cutoff.

Here's an analysis of the provided information according to your criteria:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Qualitative analysis of cannabinoids in human urine with a 50 ng/mL cutoffThe Cannabinoids assay is an in vitro diagnostic assay for the qualitative analysis of cannabinoids in human urine with a 50 ng/mL cutoff.
Acceptable concordance with predicate device (Emit® II Cannabinoid assay on SYVA®-30R Analyzer)100% agreement between the new Cannabinoids assay and the Emit® II Cannabinoid assay.
Acceptable concordance with GC/MS (Gas Chromatography/Mass Spectrometry), generally considered the gold standard for confirmation.97% agreement with GC/MS.
Precision (within-run and total %CV) for various control levels should be low, indicating reliability and reproducibility.Total %CV for Verifier I: 1.19% Total %CV for Cutoff Calibrator: 1.13% Total %CV for Verifier II: 0.63% Total %CV for -25% Control of Cutoff Calibrator: 2.32% Total %CV for +25% Control of Cutoff Calibrator: 2.70%
Limit of detection (sensitivity)15 ng/mL

Study Details

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: The document does not explicitly state the total number of clinical specimens used for the concordance studies. It only mentions that "The clinical specimens tested ranged from 14.2 to 61.8 ng/mL."
    • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as the ground truth was established by laboratory methods (predicate device and GC/MS), not by human expert interpretation of images or other subjective data.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. Comparisons were made against a predicate device and a confirmatory analytical method (GC/MS).
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC study was not done. This device is an in vitro diagnostic assay, not an AI-powered diagnostic imaging tool that would typically involve human readers.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, this was a standalone performance study. The Cannabinoids assay (the "algorithm" equivalent in this context) was evaluated for its direct comparison against a predicate device and GC/MS without human intervention in the interpretation of the assay's output.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The primary ground truth used for performance evaluation was:
      • Predicate Device: Emit® II Cannabinoid assay on the SYVA®-30R Analyzer.
      • Confirmatory Method: Gas Chromatography/Mass Spectrometry (GC/MS), which is cited as the "preferred confirmatory method."
  7. The sample size for the training set:

    • Not applicable/Not mentioned. Immunoassays like this one do not typically involve "training sets" in the same way machine learning algorithms do. The assay is developed based on chemical principles and reagent optimization rather than data-driven training.
  8. How the ground truth for the training set was established:

    • Not applicable/Not mentioned, as there isn't a "training set" in the context of this immunoassay. The development and optimization of such assays would involve extensive analytical chemistry and biochemical validation rather than establishing ground truth on a data set in the AI sense.

{0}------------------------------------------------

MAR 1 3 2002

510(k) Summary

K013247

Submitter's Name/Address Abbott Laboratories 1920 Hurd Drive MS 1-8 Irving, Texas 75038

Contact Person Linda Morris Senior Regulatory Affairs Specialist Regulatory Affairs (972) 518-6711 Fax (972) 753-3367

November 21, 2001 Date of Preparation of this Summary: Device Trade or Proprietary Name: Cannabinoids Device Common/Usual Name or Classification Name: Cannabinoids Classification Number/Class: LDJ/Class II

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K013247.

Test Description:

The Cannabinoids assay is an in vitro diagnostic assay for the qualitative analysis of cannabinoids in human urine. The assay is a homogeneous enzyme immunoassay with a 50 ng/mL cutoff. The ussay is based on competition between drug in the specimen and drug labeled with the enzyme glucose-6phosphate dehydrogenase (G6PDH) for antibody binding sites. Enzyme activity decreases upon binding to the antibody, so the drug concentration in the specimen can be measured in terms of enzyme activily. Active enzyme converts NAD to NADH, resulting in an absorbance change that can be measured spectrophotometrically.

{1}------------------------------------------------

Substantial Equivalence:

The Cannabinoids assay is substantially equivalent to the Emit® II Cannabinoid assay (K904571) on the SYVA®-30R Analyzer.

Both assays vield similar Performance Characteristics.

Similarities:

  • Both assays are in vitro immunoassays. .
  • Both assays can be used for the qualitative analysis of cannabinoids. .
  • . Both assays vield similar results.
  • Both assays are based on the competition between drug in the specimen and drug labeled with the . enzyme glucose-6-phosphate dehydrogenase (G6PDH) for antibody binding sites.
  • . Both assays have the same assay range

Differences:

  • Cannabinoids is a qualitative assay. Emit II is a qualitative and semiquantitative assay. .

Intended Use:

The Cannabinoids assay is used for the qualitative analysis of cannabinoids in human urine with a cutoff of 50 ng/mL. For use in clinical laboratories.

The Cannabinoids assay is calibrated with 11-nor-Δ'-THC-9-COOH and will detect a variety of -0 -THC metabolites.

Performance Characteristics:

Comparative porformance studies were conducted using the AEROSET® System. The Cannabinoids assay method comparison yielded acceptable concordance with the Emit II Cannabinoid assay on the SYVA-30R Analyzer. The concordance table for the Cannabinoids assay shows 100% agreement. The Cannabinoids assay method comparison yielded acceptable concordance with GC/MS. The concordance table for the Cannabinoids assay shows 97% agreement with GC/MS. The clinical specimens tested ranged from 14.2 to 61.8 ng/mL. Precision studies were conducted using the

{2}------------------------------------------------

Cannabinoids assay. A within-run and total precision study was performed using five levels of control material. The total %CV for Verifier I is 1.19%. The total %CV for the Cutoff Calibrator is 1.13%. The total %CV for Verifier II is 0.63%. The total %CV for the - 25% Control of Cutoff Calibrator and the + 25% Control of Cutoff Calibrator samples are 2.32% and 2.70%, respectively. The Cannabinoids assuy cutoff is 50 ng/mL. The limit of detection (sensitivity) of the Cannabinoids assay is 15 ng/mL. These data demonstrate that the performance of the Cannabinoids assay is substantially equivalent to the performance of the Emit II Cannabinoid assay on the SYVA-30R Analyzer.

Conclusion:

The Cannabinoids assay is substantially equivalent to the Emit II Cannabinoids assay on the SYVA-30R Analyzer as demonstrated by results obtained in the studies.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Linda Morris Senior Regulatory Affairs Specialist Abbott Laboratories 1921 Hurd Dr. P.O. Box 152020 Irving. Texas 75015-2020

Re: K013247 Trade/Device Name: Cannabinoids Regulation Name: 21 CFR 862.3870 Regulatory Class: Class II Product Code: LDJ Dated: November 26, 2001 Received: November 28, 2001

Dear Ms. Morris:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

MAR 1 3 2002

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594- . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): K013247

Device Name: ____ Cannabinoids

Indications For Use:

The Cannabinoids assay is used for the qualitative analysis of cannabinoids in human urine with a cutoff of 50 ng/mL for use in clinical laboratories. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoids use or overdose.

The Cannabinoids assay is calibrated with 11-nor-A -THC-9-COOH and will detect a variety of -A -THC metabolites.

The Cannabinoids assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

le

vision Sinn- 399 vision of Children Caporatory Devices 510(k) Number

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use V (Per 21 CFR 801. 109)

OR

Over-The-Counter Use_

(Optional Format 1-2-96

§ 862.3870 Cannabinoid test system.

(a)
Identification. A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds includedelta -9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.(b)
Classification. Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).