GRIP LOCK ANCHOR, MODEL 411374
K013063 · Medtronic Neurological · GZB · Dec 11, 2001 · Neurology
Device Facts
| Record ID | K013063 |
| Device Name | GRIP LOCK ANCHOR, MODEL 411374 |
| Applicant | Medtronic Neurological |
| Product Code | GZB · Neurology |
| Decision Date | Dec 11, 2001 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 882.5880 |
| Device Class | Class 2 |
Intended Use
The Medtronic Grip-lock™ Anchor is intended for use to facilitate Medtronic neurological lead fixation.
Device Story
Implantable, sterile, nonpyrogenic anchor; facilitates fixation of Medtronic neurological leads. Device body constructed of polysulfone with nitinol band closing mechanism. Provided in default-closed configuration; physician squeezes polysulfone wings to open anchor, allowing sliding adjustment along lead body to desired position. Once positioned, physician sutures anchor to tissue via suture holes on wings. Used by physicians during lead implantation procedures. Provides secure lead fixation; prevents lead migration; improves clinical outcomes for patients undergoing neurological stimulation therapy.
Clinical Evidence
Bench testing only. Performance data included adjustability and placement testing, perpendicular pull force, suture hole strength, angular pull force, and lead flex life. Material analysis of nitinol performed per ISO 10993-1.
Technological Characteristics
Materials: Polysulfone body, nitinol band. Mechanism: Default-closed, squeeze-to-open wings. Dimensions: Compatible with Medtronic neurological leads. Biocompatibility: ISO 10993-1 compliant. Sterilization: Sterile, nonpyrogenic.
Indications for Use
Indicated for use to facilitate Medtronic neurological lead fixation in patients requiring neurological lead placement.
Regulatory Classification
Identification
An implanted spinal cord stimulator for pain relief is a device that is used to stimulate electrically a patient's spinal cord to relieve severe intractable pain. The stimulator consists of an implanted receiver with electrodes that are placed on the patient's spinal cord and an external transmitter for transmitting the stimulating pulses across the patient's skin to the implanted receiver.
Predicate Devices
- Medtronic Twist-Lock Anchor (K972906)
Reference Devices
- Medtronic silicone anchors (P970004)
Related Devices
- K150924 — A.I.M. Lead Loop Suturing Device (1), A.I.M. Tension Assist/Suture Cutter Device, A.I.M. Lead Loop Suturing Device (1) plus Tension Assist/Suture Cutter Device, A.I.M. Lead Loop Suturing Device (2), A.I.M. Lead Loop Suturing Device plus Tension Assist/Suture Cutter Device · Anchor Innovation Medical (A.I.M.) · Jul 27, 2015
- K140272 — SWIFT-LOCK ANCHOR · St Jude Medical · May 2, 2014
- K081208 — CINCH ANCHOR, MODEL 1194 · Advanced Neuromodulation Systems · Jul 18, 2008
- K092371 — SWIFT-LOCK ANCHOR, MODEL 1192 · Advanced Neuromodulation Systems · Dec 23, 2009
- K052418 — AXXESS BUTTERFLY ANCHOR · Advanced Neuromodulation Systems · Sep 26, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
## APPENDIX IV
DEC 1 1 2001
# SUMMARY OF SAFETY AND EFFECTIVENESS TRUTHFUL AND ACCURATE STATEMENT
KU13063
510(k) Notification Medtronic, Inc. Medtronic Functional Stimulation Medtronic Grip-lock™ anchor September 10, 2001
Confidential Appendix IV Page 1
{1}------------------------------------------------
## SUMMARY OF SAFETY AND EFFECTIVENESS
K013063
#### COMPANY AND CONTACT PERSON
DEC 1 1 2001
Medtronic Neurological (Sullivan Lake Facility) Columbia Heights, MN 55421 (763) 514-5068
Monaya Lee, Associate Product Regulations Manager
#### DEVICE NAME
Grip-lock™ Anchor
## NAME OF PREDICATE OR LEGALLY MARKETED DEVICE
Medtronic Twist-Lock Anchor (Part number 103963, or Model 3550TLA (kit) / 510(k): K972906); deemed substantially equivalent on February 6, 1998.
#### DESCRIPTION OF DEVICE
The Grip-lock™ Anchor is an implantable, sterile, nonpyrogenic anchor which has been The Grip Took - ranatible with existing Metronic Neurological leads. The design is based on existing anchors (Twist-Lock Anchor) currently marketed for use with Medtronic Neurological Leads.
The anchor is not pre-attached to the lead but is provided separate from the lead in a default-The anchor is not pro anazing the polysulfone wings together, the physician can slide the anchor onto the lead. Grooves in the polysulfone wings provide a location for the physician to grip the wings with rubber-shod forceps. The physician can readjust the position of the anchor by wings with rabber sulfone wings and sliding the anchor along the lead body to the desired position.
Once the anchor is in the desired position the physician can suture the anchor to tissue using the Onec the and of each polysulfone wing and the suture T attached to the anchor body.
The gripping mechanism of the Grip-lock™ Anchor is similar to that used by the Twist-Lock Anchor currently cleared for use with spinal cord stimulation (K972906).
Medtronic Grip-lock™ anchor September 10, 2001
Confidential Appendix IV Page 2
{2}------------------------------------------------
#### STATEMENT OF INTENDED USE
The Medtronic Grip-lock™ Anchor is intended for use to facilitate Medtronic neurological lead fixation.
## STATEMENT OF INTENDED USE OF PREDICATE/MARKETED DEVICES
The Medtronic Twist-Lock Anchor, Part Number 103953 or Model 3550TLA, is indicated as an alternate anchor accessory for use with Spinal Cord Stimulation (SCS) Systems for the treatment of chronic pain of the trunk or limbs. The Twist-Lock Anchor is only used with coiled conductor Medtronic Neurological Spinal Cord Stimulation leads with an outside diameter of 1.27 mm. All other uses are contraindicated.
#### DETERMINATION OF SUBSTANTIAL EQUIVALENCE
This premarket notification is being submitted for a modified neurological lead anchor. Other anchors, previously cleared by FDA, and currently marketed include:
- Medtronic Twist-Lock Anchor (Part number 103963, or Model 3550TLA(kit) / . 510(k): K972906); was deemed substantially equivalent on February 6, 1998.
- Medtronic silicone anchors currently packaged in the lead kits approved for use . with sacral nerve stimulation systems (P970004, approved September 29, 1997).
These Medtronic anchors are devices which are currently in commercial distribution. The Twist-Lock Anchor was deemed substantially equivalent on February 6, 1998. The silicone anchors are currently supplied as components in the lead kits used for sacral nerve stimulation (P970004, approved September 29, 1997).
In determining substantial equivalence of the Medtronic Grip-lock™ Anchor, the decision-making process follows the 510(k) "Substantial Equivalence" flow diagram as follows:
The Medtronic Grip-lock" Anchor is being "compared to the following Marketed Device":
- Medtronic Twist-Lock Anchor (K972906) .
The Medtronic Grip-lock™ Anchor has the "same intended uses" as the:
- Medtronic Twist-Lock Anchor (K972906) .
The proposed indication statement for the Grip-lock™ Anchor is "for use to facilitate Medtronic neurological lead fixation" while the indication for the Twist-Lock Anchor (K972906) is "for use with spinal cord stimulation systems for chronic intractable pain of the trunk and limbs." The intended use of both devices is to facilitate neurological lead fixation
| 510(k) Notification | Medtronic Grip-lock™ Anchor | Confidential |
|----------------------|-----------------------------|--------------|
| Medtronic, Inc. | Updated December 7, 2001 | |
| Medtronic Functional | | Page 4 |
| Stimulation | | |
{3}------------------------------------------------
The Medtronic Grip-lock™ Anchor has "new technological characteristics (e.g., design, materials and manufacturing processes)" from the current Medtronic Twist-Lock Anchor. The new technological characteristics are listed below:
- Materials : Nitinol Band of the Grip-lock™ Anchor .
The Grip-lock Anchor is made primarily of polysulfone. A nitinol band encircles the anchor and serves as the closing mechanism of the anchor. Nitinol is commonly used is the medical device industry including use in cardiac stents.
- Closing mechanism of the Grip-lock " Anchor: Default closed, polysulfone wings squeezed . to open.
The Grip-lock" Anchor is provided in a "default-closed" configuration and is not preattached to the lead. When placing the anchor on the lead the physician can squeeze the polysulfone wings which open the anchor and allow the anchor to be moved along the length of the lead body. The Twist-Lock Anchor is also not pre-attached to the lead. When placing the anchor the physician twists the anchor to close it on the lead body.
These technological characteristics "could affect the safety and effectiveness of the device". However these "new technological characteristics do not raise new types of safety or effectiveness questions". In addition, "there are accepted scientific methods which exist for assessing effects of these new technological characteristics".
- Material analysis of nitinol was performed in accordance with ISO 10993-1, ● Biological Evaluation of Medical Devices - Part 1
"Performance data to assess the effects of these new technological characteristics" has been performed. The tests included;
- Adjustability and Placement on Lead ●
- . Pull Force-Perpendicular
- Suture Hole Strength and Angular Pull Force .
- . Flex Life of Lead at Anchor
These "performance data demonstrate" that the Medtronic Grip-lock™ Anchor is substantially equivalent to the currently marketed Medtronic Twist-Lock Anchor.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# DEC 1 1 2001
Ms. Monaya Krause Associate Product Regulation Manager Medtronic Functional Stimulation 800 5310 Avenue NE Minneapolis, Minnesota 55421-1200
Re: K013063
Trade/Device Name: Grip-lock™ Anchor Regulation Number: 21 CFR 882.5880 Regulation Name: Implanted Spinal Cord Stimulator for Pain Relief Regulatory Class: Class II Product Code: GZB Dated: September 10, 2001 Received: September 12, 2001
Dear Ms. Krause:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Ameridate overnments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if
{5}------------------------------------------------
#### Page 2 - Ms. Monaya Krause
applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely yours,
by Mark A. Mulhausen
elia M. Witten, Ph.D., M.D. Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Indications For Use C.
# INDICATIONS FOR USE
K013063 510(k) Number:
Grip-lock™ Anchor Device Name:
Indications for use:
The Medtronic Grip-lock™ Anchor is indicated for use to facilitate Medtronic neurological lead fixation.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR Over-The-Counter Use
Mark Millican
(Division Sign-Off)
(Division of General, Restorative Division ological Devices
510(k) Notification Medtronic, Inc. Medtronic Functional Stimulation
Confidential Medtronic Grip-lock " Anc. Ofk) Number . Updated December 7, 20
Page 1