K Number
K012606

Validate with FDA (Live)

Date Cleared
2002-02-06

(177 days)

Product Code
Regulation Number
870.5800
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To reduce the incidence of deep vein thrombosis, reduce pain and swelling after injury or surgery and increase arterial blood flow.

Device Description

The ClearMedical/Kendall SCDs are an accessory device to a SCD System which is used for the prevention of deep vein thrombosis and pulmonary embolism. The SCD Sleeve is designed with inflatable bladders that fill with air to provide compression to the legs. A connector tubing system attaches the SCD to the Controller system. The SCD Sleeves are latex-free, made of a polyester-cotton fabric. They are designed with velcro hook straps, three bladders and molded polyethylene tubing with a polypropylene snap-lock connector. Kendall has recently changed the device from the color blue to white. ClearMedical will designate the white SCDs with a "w" after the part number (i.e. 5330W).

AI/ML Overview

This 510(k) summary describes a reprocessed medical device, the ClearMedical/Kendall Sequential Compression Device (SCD) Sleeves, which are intended to be substantially equivalent to the Kendall SCD Sleeves. The submission focuses on demonstrating that the reprocessed devices maintain the same safety and performance characteristics as the original, legally marketed predicate devices.

Here's an analysis of the provided information, focusing on your requested categories:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not explicitly present a table of numerical acceptance criteria with corresponding performance data in the typical format of a clinical study report. Instead, it relies on a qualitative assessment of "technological indicators of substantial equivalence" and bench testing results. The acceptance criterion is essentially "equality in safety and performance" compared to the predicate device.

Acceptance Criteria CategorySpecific Indicators/TestsReported Device Performance (Summary)
Overall EquivalenceIn all relevant safety and performance indicatorsClearMedical devices demonstrated equality in safety and performance.
Technological CharacteristicsMethods of infection controlClearMedical's methods meet or exceed CDC standards.
Fit/attachmentSame as predicate device.
Bladder functionSame as predicate device (functionally inflate with air to control circulation).
Velcro adhesionImplied as same/equivalent to predicate.
Connector functionImplied as same/equivalent to predicate.
Physical FormForm factorSame as predicate device.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated. The document refers to "bench tests" and "non-clinical performance data" but does not quantify the number of reprocessed sleeves or components tested.
  • Data Provenance: The data is generated from bench testing and "non-clinical performance data" conducted by ClearMedical, Inc. This is internal company data. The country of origin of the data is the USA (Bellevue, WA, where ClearMedical is located). The data is retrospective in the sense that it evaluates the reprocessed devices against the established characteristics of the predicate device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Not specified.
  • Qualifications of Experts: Not specified. The submission appears to rely on internal engineering and quality control expertise within ClearMedical for evaluating the bench test results against the predicate device's characteristics. There is no mention of independent clinical experts or a consensus panel.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable or specified. Given the nature of bench testing for substantial equivalence, there isn't a "ground truth" to be adjudicated by multiple readers in the traditional sense of a diagnostic medical device. The comparison is against engineering specifications and the known characteristics of the predicate device.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

  • MRMC Study: No, an MRMC comparative effectiveness study was not done. The device is a physical, accessory compression sleeve, not an imaging or diagnostic device that would typically involve human readers interpreting cases.
  • Effect Size: Not applicable.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Standalone Performance: Not applicable. This is a physical, non-electronic device. There is no algorithm involved. "Standalone" performance would refer to the device's physical function by itself, which is what the bench tests were designed to assess.

7. The Type of Ground Truth Used

  • Ground Truth: The "ground truth" for this substantial equivalence determination is the established characteristics and performance of the legally marketed predicate device (Kendall SCD Sleeves). This includes its technical labeling, promotional materials, and assumed baseline performance for safety and efficacy. The reprocessed device's performance benchmarks were derived from the predicate device's specifications and behavior.

8. The Sample Size for the Training Set

  • Training Set Sample Size: Not applicable. This device is not an AI/ML algorithm that requires a "training set." It is a reprocessed physical medical device.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: Not applicable, as there is no training set for an AI/ML algorithm.

{0}------------------------------------------------

K02606

Image /page/0/Picture/1 description: The image shows a logo with a water droplet inside a diamond shape. To the right of the logo, the text "FEB 6 2002" is printed in a bold, sans-serif font. The text likely represents a date, with "FEB" indicating February, "6" indicating the day, and "2002" indicating the year.

510(k) Summary

Image /page/0/Picture/3 description: The image shows the word "ClearMed" in a bold, sans-serif font. The letters are black against a white background. A small trademark symbol is located to the right of the word.

ClearMedical, Inc. Submitter's name and Address: 1776 - 13610 Place NE Bellevue, WA 98005 Ph (425) 401-1414 Fax (425) 401-1515 3017110 FDA Registration Number: Richard Radford Contact Person: Director of Research and Product Development August 8, 2001 Date Summary Prepared:

Trade or Proprietary Name:

Common Name:

Classification Name:

Sleeve, Limb, Compressible (per 21 CFR section 870.5800)/JOW

Sequential Compression Sleeves

ClearMedical/Kendall Sequential Compression

Equivalent Device

The ClearMedical/Kendall reprocessed Sequential Compression Device (SCD) Sleeves are substantially equivalent to the Kendall SCD Sleeves, Parts 5329, 5329W. 5330, 5330W. 5345, 5345W. 5480, and 5480W. This determination has been reached based on an evaluation and analysis of the predicate device's technical and promotional labeling and specific bench testing. For all established indicators of substantial equivalence, the ClearMedical devices demonstrated equality in safety and performance.

Device (SCD) Sleeves

Device Description:

The ClearMedical/Kendall SCDs are an accessory device to a SCD System which is used for the prevention of deep vein thrombosis and pulmonary embolism. The SCD Sleeve is designed with inflatable bladders that fill with air to

3

{1}------------------------------------------------

510(k) Summary (Cont'd)

provide compression to the legs. A connector tubing system attaches the SCD to the Controller svstem.

The SCD Sleeves are latex-free, made of a polyester-cotton fabric. They are designed with velcro hook straps, three bladders and molded polyethylene tubing with a polypropylene snap-lock connector. Kendall has recently changed the device from the color blue to white. ClearMedical will designate the white SCDs with a "w" after the part number (i.e. 5330W).

Intended Use:

The ClearMedical/Kendall SCDs are intended as a single patient use accessory to a Kendall Controller System. The role of the SCD is to reduce the incidence of deep vein thrombosis, reduce pain and swelling after injurv and surgery, and increase arterial blood flow. The SCD Sleeves are used in conjunction with the Controller System for non-ambulatory patients in a hospital environment.

Technological Characteristics of ClearMedical and Predicate Devices:

The predicate device and the ClearMedical/Kendall SCDs contain bladders that inflate with air to control circulation of a limb. Attached to the Sleeve is a connector tubing system that connects to the Controller System. In form, the predicate device and the ClearMedical reprocessed SCD are the same.

Technological indicators of substantial equivalence were identified and included methods of infection control, fit/attachment, bladder function, velcro adhesion and connector function.

The predicate device is delivered to the customer labeled 'non-sterile' whereas the ClearMedical/Kendall SCDs are delivered to the customer labeled "High Level-Disinfected". ClearMedical's infection control methods meet or exceed the CDC standards ("Guideline for Handwashing and Hospital Environmental Control", 1985, "APIC Guideline for Selection and Use of Disinfectants") for this class of device.

Summary of ClearMedical/Kendall SCD Performance:

Based on an assessment of bench tests and non-clinical performance data, we believe that in all relevant safety and performance indicators the ClearMedical/ Kendall SCD Sleeves demonstrates substantial equivalence to the predicate devices, the Kendall SCD Sleeves.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 0 6 2002

Mr. Richard Radford Director of Research and Product Development ClearMedical, Inc. 1776 136th Place NE Belluvue, WA 98005-2328

Re: K012606

Trade Name: ClearMedical/Kendall Sequential Compression Device (SCD) Sleeves Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II (two) Product Code: JOW Dated: January 16, 2002 Received: January 17, 2002

Dear Mr. Radford:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your deyice can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Richard Radford

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r reaso be actived that I Dr usean that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of It in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) rms lotter will and will you to FDA finding of substantial equivalence of your device to a legally premation noticate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 11 you docure bpoints art 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Oiner general mironmation the mational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

cj

Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Statement of Indications for Use

510(K) NUMBER (IF KNOWN):K012606
DEVICE NAME:ClearMedical/Kendall Sequential Compression Devices (SCD)

Part Numbers: 5329, 5330, 5345, 5480, 5329W, 5330W, 5345W, 5480W

INDICATIONS FOR USE:

To reduce the incidence of deep vein thrombosis, reduce pain and swelling 10 rought or surgery and increase arterial blood flow.

(Please do not write below this line-continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE)

Division of Cardiovascular & Respiratory Devices
510(k) Number K012606

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter-Use _

§ 870.5800 Compressible limb sleeve.

(a)
Identification. A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.(b)
Classification. Class II (performance standards).