K012308 · Kavo America · EFB · Jun 26, 2002 · Dental
Device Facts
Record ID
K012308
Device Name
KAVO QUATTROCARE
Applicant
Kavo America
Product Code
EFB · Dental
Decision Date
Jun 26, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Indications for Use
KaVo QUATTROcare is intended for internal cleaning, i.e., purging of old lubricant, for the maintenance of rotating dental and surgical instruments. NOTE: QUATTROcare should be used with only pre-cleaned dental handpieces and before they are sterilized. CAUTION: Federal (US) law restricts the use of this device to licensed professionals.
Device Story
KaVo QUATTROcare is an aerosol-based maintenance system for dental and surgical handpieces; utilizes pressurized cans containing hydrocarbon propellants, lubricants, and flavor/odor agents. Device performs internal cleaning by purging old lubricant from handpieces. Used in dental clinical settings by licensed professionals; applied to pre-cleaned instruments as a mandatory step prior to sterilization. System ensures proper instrument lubrication and debris removal, extending instrument life and maintaining hygiene standards.
Clinical Evidence
No clinical data. Substantial equivalence is based on performance data comparing formulations and intended uses to the predicate device.
Technological Characteristics
Aerosol-based maintenance system; consists of unit (2104 A) and pressurized cans (2106 A). Contains hydrocarbon propellants, lubricants, and flavor/odor ingredients. Manual application process for dental handpiece maintenance.
Indications for Use
Indicated for internal cleaning and maintenance of rotating dental and surgical instruments. For use by licensed professionals only on pre-cleaned instruments prior to sterilization.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
Predicate Devices
KaVo Spray
Related Devices
K071288 — MODIFICATION TO KAVO QUATTROCARE · Sybron Dental Specialties, Inc. · Jul 19, 2007
K143183 — Maxima XTEND Handpiece Maintenance System · Handpiece Headquarters · Feb 17, 2015
K070074 — LUBRINA, MODEL HIM-1 · J. Morita USA, Inc. · May 31, 2007
K113674 — SPRAY & CLEAN HANDPIECE CLEANER& LUBRICANT MAXIMA HANDPIECE CLEANER & LUBRICANT EZCARE HANDPIECE CLEANER & LUBRICANT · Handpiece Headquarters, Inc. - Hpr, Inc. · Apr 5, 2012
K081811 — CARE3 PLUS · Nakanishi, Inc. · Jan 21, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
# JUN 2 6 2002
K012308
## SECTION 9
### 510(k) SUMMARY
This 510(k) summary of safety and effectiveness for KaVo QUATTROcare (unit and pressurized cans) is submitted in accordance with the requirements of SMDA 1990 and follows Office of Device Evaluation guidance concerning the organization and content of a 510(k) summary.
- KaVo America Applicant:
- 340 East Main Street Address: Lake Zurich, IL 60045
- Manufacturer: KaVo Dental GmbH Bahnhofstr. 20 D-8847 Warthausen Biberach GERMANY
- Contact Person: Mr. John Westermeier
- Telephone: 847-550-6800 847-550-6825 (Fax) 800-323-8029
- February 2002 Preparation Date: (of the Summary)
KaVo QUATTROcare (Unit 2104 A; 6-pack of pressurized cans 2106 A) Device Name:
- Dental Handpiece Accessory (for maintenance; cleaning, and lubrication) Common Name:
Classification: Accessory to Dental Handpieces (Class I medical device: (21 CFR 872,4200),
> Product Code: EFB (Product Code for Dental Handpieces) Panel: 76
Predicate devices: KaVo Spray
Description: The KaVo QUATTROcare is an aerosol product with hydrocarbon propellants and lubricant and flavor/odor ingredients for use in the routine maintenance (cleaning and lubrication) of dental handpieces prior to their sterilization.
{1}------------------------------------------------
- Indications: KaVo QUATTROcare is intended for internal cleaning, i.e., purging of old lubricant, for the maintenance of rotating dental and surgical instruments .
NOTE: QUATTROcare should be used only with pre-cleaned dental handpieces and before they are sterilized.
CAUTION: Federal (US) law restricts the use of this device to licensed professionals.
- None required. The claim of substantial equivalence is based on Performance Data: comparisons of formulations and intended uses of the QUATTROcare and its claimed predicate.
- CONCLUSION: Based on the information in the notification KaVo America believes that QUATTROcare is substantially equivalent to the claimed predicate (i.e., KaVo Spray, a pre-amendments product).
Rev. April 2002
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, with flowing lines suggesting movement or progress.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUN 2 6 2002
Mr. John Westermeir Kavo America 340 East Main Street Lake Zurich, Illinois 60045
Re: K012308 Trade/Device Name: KaVo QUATTROcare
Regulation Number: 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EFB Dated: April 12, 2002 Received: April 17, 2002
Dear Mr. Westermeier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
Page 2 - Mr. Westermeier
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Klatruds
Timothy A. Ulatowski
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### SECTION 7
### INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K012308 -
Device Name: KaVo QUATTROcare
Indications for Use Statement:
KaVo QUATTROcare is intended for internal cleaning, i.e., purging of old lubricant, for the maintenance of rotating dental and surgical instruments.
NOTE: QUATTROcare should be used with only pre-cleaned dental handpieces and before they are sterilized.
KaVo proposes that KaVo QUATTROcare carry the following label:
CAUTION: Federal (US) law restricts the use of this device to licensed professionals.
Rev. April 2002
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
Prescription Use کر (Per 21 CFR 801.109) OR
Over-The Counter Use
Susan Pum
(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital, Infection Control, and General Hospi 510(k) Number
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.