(330 days)
Tyto Insights for Eardrum Bulging Detection is an over-the-counter web-based AI-enabled software analytics device intended to automatically detect the bulging of the eardrum in Otoscopic video recordings acquired by the compatible electronic otoscope, as identified in the product labeling, in patients aged 6 months and older. The device is not intended for diagnosis. A healthcare provider's advice is required to understand the meaning of the Tyto Insights for Eardrum Bulging Detection result. Healthcare providers should consider the device result in conjunction with otoscopic recordings and other relevant patient data. The device is not intended to detect other abnormal otoscopic findings.
Tyto Insights for Eardrum Bulging Detection is a web-based AI enabled software analytics device intended to automatically detect the bulging of the Eardrum in Otoscopic video recordings acquired by the compatible Tyto Otoscope in pediatrics and adults. The device is intended to process recordings only from the compatible Tyto Otoscope (Class I medical device, exempt from premarket notification procedures), manufactured by TytoCare Ltd. The acquisition of the video otoscopy data (recordings) is carried out by a professional user in a clinical environment or by a lay user in a non-clinical environment.
Bulging is a medical characteristic of the tympanic membrane that is commonly used for the diagnosis of Acute Otitis Media, but is also observed in ear trauma, barotrauma, bullous myringitis, etc. Therefore, the device can provide an aid to diagnosis when used in conjunction with otoscopic ear examinations and when considering all patient symptoms and medical history.
The Tyto Insights for Eardrum Bulging Detection system is a processing service and has an interface with a Telehealth system which is not part of this product. The Tyto Insights for Eardrum Bulging Detection device is responsible for the data transfer, store and read-only display of device output (otoscopic video analysis results).
Here's a breakdown of the acceptance criteria and the study that demonstrates the device's adherence to those criteria, based on the provided document:
Acceptance Criteria and Device Performance Study for Tyto Insights for Eardrum Bulging Detection
1. Table of Acceptance Criteria and Reported Device Performance
The primary acceptance criteria for the Tyto Insights for Eardrum Bulging Detection device were defined in a comparative effectiveness study, comparing the device's performance against a panel of non-ENT clinical readers. The criteria focused on demonstrating superiority in both sensitivity and specificity.
Acceptance Criteria for Primary Endpoint (Comparative Clinical Performance):
| Metric (Difference: Device vs. Readers) | Success Criteria (Lower Bound of 95% CI) | Reported Performance (Estimate) | Reported Performance (LCI) | Reported Performance (UCI) | Met? |
|---|---|---|---|---|---|
| Sensitivity Difference | > 0 | 0.1989 | 0.1691 | 0.2315 | Yes |
| Specificity Difference | > 0 | 0.2087 | 0.1756 | 0.2442 | Yes |
Standalone Performance (Against Expert GT)
While not explicitly stated as acceptance criteria in the same manner as the comparative study, the standalone performance against the expert ground truth provides the baseline for the device's accuracy.
| Measure | Reported Performance (Estimate) | LCI | UCI |
|---|---|---|---|
| Sensitivity | 0.9231 | 0.8902 | 0.9443 |
| Specificity | 0.9463 | 0.9227 | 0.9656 |
| PPV* | 0.6009 | 0.5097 | 0.7025 |
| NPV* | 0.9929 | 0.99 | 0.9948 |
Note: PPV and NPV are calculated based on Bayes theorem to account for eardrum prevalence of 8.06%.
Secondary Endpoint (Eardrum Quality Filter Performance):
No success criteria was applied to the secondary endpoint.
| Measure | Reported Performance (Estimate) | LCI | UCI |
|---|---|---|---|
| Sensitivity (Eardrum Quality Filter) | 0.9754 | 0.9633 | 0.9838 |
| Specificity (Eardrum Quality Filter) | 0.8997 | 0.8605 | 0.9252 |
2. Sample Size and Data Provenance for Test Set
- Sample Size for Test Set: 857 otoscopic recordings (857 patients).
- Data Provenance: Retrospective validation dataset sourced from real-world use of the Tyto Otoscope recordings. While specific countries are not mentioned for the test set, the training data originates from the TytoCare RWD (Real-World Data) Database, implying a similar origin of real-world telehealth use. The comparative human performance study specifically states the non-ENT clinicians were U.S.-licensed, suggesting the data may represent typical U.S. patient populations.
3. Number of Experts and Qualifications for Test Set Ground Truth
- Number of Experts: For the standalone performance (bench study) against expert ground truth, the document states the external ground truth was established by "a panel of ENTs." For the comparative clinical performance study (human reader comparison), the ground truth for the device's standalone performance within that study was also established by a panel of ENTs. The exact number of ENTs for the test set ground truth is not explicitly stated, but the training set ground truth process provides insight (see point 9).
- Qualifications of Experts: ENT (Ear Nose Throat) specialists (otolaryngologists) for establishing the primary ground truth. For the comparative study, the human reader panel consisted of two pediatricians and one general practitioner (U.S.-licensed), who were chosen to represent the intended clinical users of the device in U.S. primary care settings.
4. Adjudication Method for the Test Set
The document does not explicitly state the adjudication method (e.g., 2+1, 3+1) for establishing the ground truth of the test set. However, for the training set, it describes a process where 3 ENTs annotated data, and the "final ground truth is determined by majority vote among the three truthers," suggesting a 3+1 (majority vote) approach was likely used for expert ground truth establishment for the test set as well, or at least a consensus-based approach among a panel.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
-
Was it done? Yes, a comparative effectiveness study was performed, comparing the device's standalone performance against a panel of human readers (two pediatricians and one general practitioner). This implicitly functions similar to an "AI vs. human" comparative study, although the AI was standalone and not assistive.
-
Effect Size (Device vs. Human Readers):
- Sensitivity Improvement: The device had a sensitivity improvement of 0.1989 (LCI: 0.1691, UCI: 0.2315) compared to the human readers. (Device Sensitivity: 0.9231 vs. Readers Sensitivity: 0.7242)
- Specificity Improvement: The device had a specificity improvement of 0.2087 (LCI: 0.1756, UCI: 0.2442) compared to the human readers. (Device Specificity: 0.9463 vs. Readers Specificity: 0.7325)
- PPV Improvement (Prevalence-adjusted): 0.4038 (0.3161 – 0.5070) in favor of the device.
- NPV Improvement (Prevalence-adjusted): 0.0250 (0.0215 – 0.0288) in favor of the device.
The study demonstrated that the device exhibited "higher sensitivity and specificity" compared to the non-ENT reader panel. This "effect size" highlights how much better the AI performed alone compared to the human readers alone for the detection task. It's not an AI-assisted human performance metric.
6. Standalone Performance (Algorithm Only)
- Was it done? Yes. The "Performance Testing - Bench" section (Page 8) explicitly details the standalone performance and accuracy (Sensitivity, Specificity, PPV, NPV) of the Tyto Insights for Eardrum Bulging Detection device.
7. Type of Ground Truth Used
- Expert Consensus: For both training and validation (test) sets, the ground truth for eardrum bulging detection was established by a panel of ENT (Ear Nose Throat) specialists (otolaryngologists) through a consensus or majority vote process.
8. Sample Size for the Training Set
- Total Training Dataset: 3,102 otoscopic recordings.
- Bulging Positive: 606 recordings
- Bulging Negative: 1,510 recordings
- Tympanic Membrane Position Not Sufficiently Captured: 986 recordings
9. How the Ground Truth for the Training Set Was Established
The ground truth for the training set was established through a multi-step process involving multiple ENT specialists:
- Step 1 (Initial Screening): Recordings were systematically annotated by one ENT specialist into "Ear Abnormal Recording" and "Ear Normal Recording." This step aimed to address the low prevalence of bulging in real-world data by enriching the "abnormal" pool.
- Step 2 (Detailed Annotation & Majority Vote): All data marked "Abnormal" from Step 1, plus a randomly selected portion of "Normal" data, were then annotated by three ENTs. They categorized recordings as "Bulging Positive," "Bulging Negative," or "Tympanic Membrane Position was not sufficiently Captured." The final ground truth for these recordings was determined by a majority vote among the three ENTs.
- Step 3 (Frame-level Annotation): Randomly selected recordings from Step 2 were further annotated on a frame level by an ENT. This highly granular annotation was used to train both the Eardrum Bulging Detection and Quality Filter Algorithms.
This comprehensive approach leveraged multiple expert opinions and a tiered annotation process to establish robust ground truth labels for training the AI algorithms.
DE NOVO CLASSIFICATION REQUEST FOR Tyto Insights for Eardrum Bulging Detection
Page 1
REGULATORY INFORMATION
FDA identifies this generic type of device as:
Ear, nose, and throat image analyzer. An ear, nose, and throat image analyzer is a device that uses software to analyze patient-specific optical images of the ear, nose, and throat area. The results are provided for healthcare providers as an aid to detection, diagnosis, or monitoring of ear, nose, and throat conditions.
NEW REGULATION NUMBER: 21 CFR 874.4775
CLASSIFICATION: Class II
PRODUCT CODE: SHL
BACKGROUND
DEVICE NAME: Tyto Insights for Eardrum Bulging Detection
SUBMISSION NUMBER: DEN250014
DATE DE NOVO RECEIVED: April 21, 2025
SPONSOR INFORMATION:
- Tyto Care Ltd.
- Stella Raizelman Perry
- 14 Beni Gaon Street
- Netanya, Israel
INDICATIONS FOR USE
The Tyto Insights for Eardrum Bulging Detection is indicated as follows:
Tyto Insights for Eardrum Bulging Detection is an over-the-counter web-based AI-enabled software analytics device intended to automatically detect the bulging of the eardrum in Otoscopic video recordings acquired by the compatible electronic otoscope, as identified in the product labeling, in patients aged 6 months and older. The device is not intended for diagnosis. A healthcare provider's advice is required to understand the meaning of the Tyto Insights for Eardrum Bulging Detection result. Healthcare providers should consider the device result in conjunction with otoscopic recordings and other relevant patient data. The device is not intended to detect other abnormal otoscopic findings.
Page 2
LIMITATIONS
Limitations on device use are included in the Instructions for Use as Contraindications, and Warnings/Precautions.
Contraindications
- Not intended for diagnostic purposes.
- Not intended for patients under the age of 6 months.
- Not suitable for use by people who have undergone ear surgical procedures on the eardrum, middle ear, and/or mastoid. These procedures encompass tympanoplasty, canal wall up tympanomastoidectomy, canal wall down mastoidectomy, modified radical mastoidectomy, radical mastoidectomy, and myringotomy with ventilation tube insertion (ear tubes).
- Not suitable for use by people suspected of having a foreign object in the ear.
Warnings
- The Tyto Insights for Eardrum Bulging Detection can detect eardrum bulging in recordings acquired only by the compatible Tyto Otoscope.
- Ears recordings must be collected only according to the guidance provided with the compatible Tyto Otoscope. Instructions for adequate data collection method are provided in the Tyto Otoscope instructions for use
- The user of the Tyto Otoscope and Tyto Insights for Eardrum Bulging Detection must be:
- An adult, aged 18 and older.
- Comfortable using mobile apps.
- The Tyto Otoscope must be used as intended according to its instructions for use. Pay special attention to recording the environment to meet the requirements defined in the instruction for use.
- Always discuss the results of the Tyto Insights for Eardrum Bulging Detection with a medical professional.
- Never rely on the result of the Tyto Insights for Eardrum Bulging Detection for diagnosis and treatment.
- Regardless of the Tyto Insights for Eardrum Bulging Detection result, if the patient has signs and symptoms such as ear pain or fever, they should always approach their healthcare professional for advice.
Page 3
DEVICE DESCRIPTION
Tyto Insights for Eardrum Bulging Detection is a web-based AI enabled software analytics device intended to automatically detect the bulging of the Eardrum in Otoscopic video recordings acquired by the compatible Tyto Otoscope in pediatrics and adults. The device is intended to process recordings only from the compatible Tyto Otoscope (Class I medical device, exempt from premarket notification procedures), manufactured by TytoCare Ltd. The acquisition of the video otoscopy data (recordings) is carried out by a professional user in a clinical environment or by a lay user in a non-clinical environment.
Bulging is a medical characteristic of the tympanic membrane that is commonly used for the diagnosis of Acute Otitis Media, but is also observed in ear trauma, barotrauma, bullous myringitis, etc. Therefore, the device can provide an aid to diagnosis when used in conjunction with otoscopic ear examinations and when considering all patient symptoms and medical history.
Device input obtained via Otoscope (Class I medical device exempt from premarket notification):
The Tyto Insights for Eardrum Bulging Detection system is a processing service and has an interface with a Telehealth system which is not part of this product. The Tyto Insights for Eardrum Bulging Detection device is responsible for the data transfer, store and read-only display of device output (otoscopic video analysis results).
Page 4
Tyto Insights for Eardrum Bulging Detection Architecture:
Tyto Insights for Eardrum Bulging Detection system is composed of the following sub-systems:
I. Application Server (APS) communicates with Algorithm Server (ALS) and implements an application programming interface (API) for communication with the telehealth server.
II. The ALS sub-system is composed of the artificial intelligence-enabled bulging detection algorithm, eardrum quality filter and logic wrapper and interface. The ALS receives a video file as input and returns as output a positive/negative/unable to analyze result.
III. The Web Server provides a graphic indication whether an eardrum bulging is detected in the recording. It can be utilized both in the patient and clinician side. All the software subsystems (servers and storage) are hosted in the cloud and communicate via IP network
When the input (.mp4 file) is received, the Tyto Insights for Eardrum Bulging Detection applies an artificial intelligence-enabled "Eardrum Quality Filter" that excludes video files with any of the following: no eardrum detected, duration less than 4 seconds, non-MP4 format, or corrupted files.
The artificial intelligence-enabled Eardrum Quality Filter Algorithm outputs a classification of:
Positive: Eardrum with sufficient quality to determine ear drum position was detected in video
Negative: Eardrum with sufficient quality to determine ear drum position was not detected in video
Page 5
Next, an artificial intelligence-enabled Bulging Detection Algorithm outputs a classification of:
Positive: Eardrum Bulging was detected in the video
Negative: Eardrum Bulging was not detected in the video
The user (lay user) view includes a warning reminding the user not to rely on the analysis results for diagnosis or treatment and to consult with a clinician.
The clinician view includes a view the otoscopic video recording, with a mouse-over option that displays a short video segment (up to 5 seconds) in which the eardrum is visible. The clinician should consider the results of the Tyto Insights for Eardrum Bulging Detection in conjunction with the otoscopic recording and with all other relevant patient data.
Page 6
Training Dataset
The training dataset consists of 3,102 otoscopic recordings:
- Visible Tympanic Membrane Position - Bulging Positive: 606 recordings
- Visible Tympanic Membrane Position - Bulging Negative: 1,510 recordings
- Tympanic Membrane Position Not Sufficiently Captured: 986 recordings
The training dataset was sourced from the TytoCare RWD (Real-World Data) Database, originating from individual patients who upload their otoscopic videos during routine telehealth use, as intended. The training dataset was categorized based on their ground truth annotations by multiple ENT (Ear Nose Throat) specialists (otolaryngologists):
At Step 1, the recordings were systematically annotated by one ENT specialist into two categories: "Ear Abnormal Recording" and "Ear Normal Recording." This initial classification addressed the challenge posed by the low prevalence of the targeted bulging otoscopy finding in real world data.
At Step 2, all of the data marked "Abnormal" in Step 1 and a part of randomly selected "Normal" data was transferred to a set to be annotated by 3 ENTs into the following categories: "Bulging Positive," "Bulging Negative," and "Tympanic Membrane Position was not sufficiently Captured." The final ground truth is determined by majority vote among the three truthers.
At Step 3, randomly selected recordings from Step 2 were annotated on a frame level by an ENT. This dataset was used to train both Eardrum Bulging Detection and Quality Filter Algorithms.
Page 7
Below is the training dataset distribution for recordings with a visible tympanic membrane:
N=3,102
| Age Group (Years) | Negative | Positive | Total |
|---|---|---|---|
| 0 - 2 years | 73 (42.94%) | 97 (57.06%) | 170 (8.03%) |
| 2 - 12 years | 487 (59.25%) | 335 (40.75%) | 822 (38.85%) |
| 12 - 21 years | 306 (93.01%) | 23 (6.99%) | 329 (15.55%) |
| 21+ years | 644 (81.01%) | 151 (18.99%) | 795 (37.57%) |
Gender
| Negative | Positive | Total | |
|---|---|---|---|
| Male | 642 (69.03%) | 288 (30.97%) | 930 (43.95%) |
| Female | 868 (73.19%) | 318 (26.81%) | 1186 (56.05%) |
Ear Sides
| Negative | Positive | Total | |
|---|---|---|---|
| Right | 763 (71.11%) | 310 (28.89%) | 1073 (50.71%) |
| Left | 747 (71.62%) | 296 (28.38%) | 1043 (49.29%) |
Ear cap type
| Negative | Positive | Total | |
|---|---|---|---|
| Pediatric Ear Cap | 574 (63.29%) | 333 (36.71%) | 907(42.86%) |
| Not Pediatric Ear Cap | 936 (77.42%) | 273 (22.58%) | 1209 (57.14%) |
Opacification Only
| Negative | Positive | Total | |
|---|---|---|---|
| Opacification | 276 (100.0%) | 0 (0.0%) | 276 (100.0%) |
User Type
| Negative | Positive | Total | |
|---|---|---|---|
| Lay user | 1349 (69.86%) | 582 (30.14%) | 193 (91.26%) |
| Clinician | 161 (87.03%) | 24 (12.97%) | 185 (8.74%) |
SUMMARY OF NONCLINICAL/BENCH STUDIES
SOFTWARE
Tyto Insights for Eardrum Bulging Detection was identified as requiring "Basic Documentation" as defined in the FDA Guidance "Content of Premarket Submissions for Device Software Functions," as it was concluded that failure or latent flaw of the software would not present a hazardous situation with a probable risk of death or serious injury to either the patient, user of the device, or others in the environment of use, prior to the implementation of risk control measures.
Page 8
The Basic Documentation included:
- Software/Firmware Description
- Risk Management File (including Hazard Analysis)
- Software Requirement Specifications (SRS)
- System and Software Architecture Design Chart
- Software Life Cycle Process Description / Software Development, Configuration Management, and Maintenance Practices
- Software Testing as Part of Verification & Validation
- Software Version / Revision Level History
- Unresolved Software Anomalies
Regarding the cybersecurity, the documentation included all the information recommended by the FDA Guidance "Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions." This includes a Risk Management Report, Threat Model, Risk Assessment, Software Bill of Materials (SBOM), Assessment of Unresolved Anomalies for cybersecurity impact, Metrics, Controls, Architecture Views, Testing, and Management Plan.
PERFORMANCE TESTING - BENCH
Standalone performance and accuracy (Sensitivity, Specificity, Positive Predictive Value (PPV), Negative Predictive Value (NPV)) of the Tyto Insights for Eardrum Bulging Detection device in detecting eardrum bulging was evaluated on a retrospective validation dataset of Tyto Otoscope recordings sourced from real world use. There were 857 recordings (857 patients) corresponding to the intended patient population (429 recordings were eardrum bulging positive and 428 were eardrum bulging negative). The demographics of the validation dataset are shown in Table 1:
Table 1: Validation dataset demographics
N=857 Recordings
| Age Group (Years) | Positive | Negative | Total |
|---|---|---|---|
| 6 months – 2 years | 62 (66.67%) | 31 (33.33%) | 93 (10.85%) |
| 2-12 years | 296 (63.93%) | 167 (36.07%) | 463 (54.03%) |
| 12-21 years | 40 (39.6%) | 61 (60.4%) | 101 (11.79%) |
| >=22 | 31 (15.5%) | 169 (84.5%) | 200 (23.34%) |
Gender
Page 9
| Positive | Negative | Total | |
|---|---|---|---|
| Male | 225 (55.69%) | 179 (44.31%) | 404 (47.14%) |
| Female | 204 (45.03%) | 249 (54.97%) | 453 (52.86%) |
Ethnicity and Race
| Positive | Negative | Total | |
|---|---|---|---|
| Asian | 23 (60.53%) | 15 (39.47%) | 38 (4.43%) |
| Black/ African American | 11 (31.43%) | 24 (68.57%) | 35 (4.08%) |
| Caucasian | 24 (46.15%) | 28 (53.85%) | 52 (6.06%) |
| Hispanic Latino | 17 (38.64%) | 27 (61.36%) | 44 (5.13%) |
| Other or Unknown | 355(51.30%) | 337(48.69%) | 692(83.37%) |
The device demonstrated a sensitivity of 0.9231 (95% CI: 0.8902–0.9443) and a specificity of 0.9463 (95% CI: 0.9227–0.9656), corresponding to a PPV of 0.6009 (95% CI: 0.5097–0.7025) and a NPV of 0.9929 (95% CI: 0.9900–0.9948).
The standalone accuracy of the Tyto Insights for Eardrum Bulging Detection shown in Table 2:
Tyto Insights for Eardrum Bulging Detection
| Measure | n / N | Estimate | LCI | UCI |
|---|---|---|---|---|
| Sensitivity | 396/429 | 0.9231 | 0.8902 | 0.9443 |
| Specificity | 405/428 | 0.9463 | 0.9227 | 0.9656 |
| Positive Predictive Value (PPV)* | NA* | 0.6009 | 0.5097 | 0.7025 |
| Negative Predictive Value (NPV)* | NA* | 0.9929 | 0.99 | 0.9948 |
Table 2: Standalone performance among patients of age 6 months and above: PPV – positive predictive value; NPV – negative predictive value; LCI – lower bound of two-sided 95% confidence interval; UCI – upper bound of two-sided 95% confidence interval; * - calculated based on bayes theorem to account for Eardrum prevalence of 8.06%
Page 10
SUMMARY OF CLINICAL INFORMATION
In the clinical performance study, device performance was compared to an independent non otolaryngologist (non-ENT) clinical reader panel consisting of two pediatricians and one general practitioner. It should be noted that the primary care setting is the most common setting in the United States where most Acute Otitis Media cases are diagnosed, and these diagnoses are done by pediatricians and general practitioners, and not by ENTs.
Statistically significant differences in sensitivity and specificity were observed between the device and this non-ENT reader panel, with higher sensitivity and specificity observed for the device. These comparison results are specific to the reader panel included in the study and are not intended to generalize to all non-ENT clinicians. The results demonstrate that the device can aid in the clinical assessment of otoscopic findings.
Human Performance Comparison
The reference standard was established using a panel of three U.S.-licensed non-ENT clinicians (two pediatricians and one general practitioner), who independently reviewed all recordings. These clinicians routinely perform otoscopic assessments in primary care settings in the United States and therefore represent the intended clinical users of the device in U.S. practice.
Primary Endpoint
Sensitivity and specificity of the detection of eardrum bulging by the device compared to the clinical readers
Success Criteria
Success is claimed only if two following conditions are met:
Lower-bound of two-sided 95% bootstrap confidence interval (CI) (equivalent to 97.5% one-sided CI) for the difference in sensitivity between the device and the readers is greater than 0, AND
Lower-bound of two-sided 95% bootstrap confidence interval (CI) (equivalent to 97.5% one-sided CI) for the difference in specificity between the device and the readers is greater than 0.
Secondary Endpoint
Accuracy (Sensitivity, Specificity) of the Eardrum Quality Filter algorithm in determining whether the eardrum was sufficiently captured. No success criteria was applied to the secondary endpoint.
Primary endpoints (Sensitivity, Specificity) clinical performance assessment shown in Table 3 - 4:
| Standalone Accuracy | n / N | Estimate | LCI | UCI |
|---|---|---|---|---|
| Sensitivity | 396/429 | 0.9231 | 0.8902 | 0.9443 |
| Specificity | 405/428 | 0.9463 | 0.9227 | 0.9656 |
The device's standalone performance against ground truth established by a panel of ENTs will be compared to that of non-ENT clinical readers (two pediatricians, one general practitioner):
Page 11
Sensitivity
| Sensitivity | LCI | UCI | |
|---|---|---|---|
| Readers | 0.7242 | 0.6921 | 0.7522 |
| Tyto Insights for Eardrum Bulging Detection | 0.9231 | 0.8902 | 0.9443 |
| Insights – Readers | 0.1989 | 0.1691 | 0.2315 |
Table 3: Primary Clinical Performance Endpoint Evaluation – sensitivity, LCI/UCI – lower/upper bound of two-sided 95% bootstrap (Bca) Confidence Interval.
Specificity
| Specificity | LCI | UCI | |
|---|---|---|---|
| Readers | 0.7325 | 0.7063 | 0.7671 |
| Tyto Insights for Eardrum Bulging Detection | 0.9643 | 0.9227 | 0.9656 |
| Insights – Readers | 0.2087 | 0.1756 | 0.2442 |
Table 4: Primary Clinical Performance Endpoint Evaluation – specificity, LCI/UCI – lower/upper bound of two-sided 95% bootstrap (Bca) Confidence Interval.
The study endpoints and hypothesis were met for the indicated patient population (difference in sensitivity of 0.1989 [0.1691 – 0.2315] and in specificity of 0.2087 [0.1756 – 0.2442], both in favor of the device). Prevalence-adjusted PPV and NPT (device – readers) of 0.4038 (0.3161 – 0.5070) and 0.0250 (0.0215 – 0.0288), respectively were observed, both in favor of the device.
Figure 1: ROC curve for the device (solid line) and readers (dashed lines).
TYTO – dark green, Reader 1 – dark red, Reader 2 – orange, Reader 3 - black
Page 12
Subgroup analysis
The difference in sensitivities and specificities between the device and clinical readers was in favor of the device across the evaluated subgroups defined by age (6 months-2 years*, 2-12 years, 12-21 years, and ≥ 22 years), gender, ethnicity, types of users (lay user/professional), difference type of ear cap size (Pediatrics size / No), different ear sides (left/right), and presence of opacification.
*Subgroup analysis of patients between 6 months and 2 years old results are shown in Table 5 - 6:
| Sensitivity | LCI | UCI | |
|---|---|---|---|
| Readers | 0.8065 | 0.7390 | 0.8624 |
| Tyto Insights for Eardrum Bulging Detection | 0.9032 | 0.8066 | 0.9655 |
| Insights - Readers | 0.0968 | 0.0108 | 0.1871 |
Table 5: Sensitivity; LCI/UCI – lower/upper bound of two-sided 95% bootstrap (BCa) Confidence Interval;
| Specificity | LCI | UCI | |
|---|---|---|---|
| Readers | 0.6344 | 0.5184 | 0.7433 |
| Tyto Insights for Eardrum Bulging Detection | 0.8710 | 0.7076 | 0.9677 |
| Insights - Readers | 0.2366 | 0.0588 | 0.3908 |
Table 6: Specificity; LCI/UCI – lower/upper bound of two-sided 95% bootstrap (BCa) Confidence Interval;
Secondary endpoints (Sensitivity, Specificity) clinical performance assessment shown in Table 7:
| n / N | Estimate | LCI | UCI | |
|---|---|---|---|---|
| Sensitivity | 833 / 854 | 0.9754 | 0.9633 | 0.9838 |
| Specificity | 350 / 389 | 0.8997 | 0.8605 | 0.9252 |
Table 7: Sensitivity/Specificity; LCI/UCI – lower/upper bound of two-sided 95% bootstrap (BCa) Confidence Interval
** calculated using prevalence of 59.6%*
The Quality Filter performance in determining whether the eardrum was sufficiently captured showed a sensitivity and specificity of 0.9754 (0.9633 – 0.9838) and 0.8997 (0.8605 – 0.9252).
Conclusion
The Tyto Insights for Eardrum Bulging Detection had a high standalone performance in detecting tympanic membrane bulging with narrow confidence intervals indicating consistent performance. The higher sensitivity and specificity for the device versus non-ENT clinicians, showed an improved detection performance without a tradeoff between true detection and false positive rates. The results demonstrate that the device can aid in the clinical assessment of otoscopic findings by analyzing recorded otoscopic videos to determine whether bulging in the eardrum is detected within the recorded video data.
Page 13
PEDIATRIC EXTRAPOLATION
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.
LABELING
The Tyto Insights for Eardrum Bulging Detection labeling is sufficient and meets labeling requirements for over-the-counter devices. It contains the indication for use, limitations, contraindication, device description, maintenance, warnings and precautions, instruction for use, and summary of clinical validation.
HUMAN FACTORS TESTING
The first human factor study included 15 U.S.-licensed healthcare providers who were tested remotely in 3 simulated scenarios (Automated Output Aligned with Clinical Findings, Device-Generated "Unable to Analyze" Output, and Discordant (False-Positive) Automated Output). Across all simulated-use scenarios, all participants successfully completed the critical task of independent clinical assessment. Participants consistently reviewed otoscopic video recordings, patient-reported symptoms, and relevant medical history, and appropriately interpreted the Tyto Insights output as supportive information only. No participant relied on the automated output as the sole basis for clinical assessment or determination of next steps.
The second human factor study was conducted in-person on 30 lay users that did not have any formalized training (15 participants aged 18-50 years and approximately 15 participants aged 50 years and older) to evaluate performance of all critical tasks of otoscopic video capture, upload, and transfer of recordings via the Tyto Care User App, comprehension of device limitations, interpretation of displayed result, and avoidance of inappropriate self-diagnosis through selection of next steps consistent with the intended use of the device. The lay users were able to use the Tyto Insights for Eardrum Bulging Detection device as part of the full operational workflow, without progression to hazardous situations.
PREDETERMINED CHANGE CONTROL PLAN (PCCP)
| Modification | Description |
|---|---|
| Modification #1[M#1]: modifications related to quantitative measures of performance specifications. | Re-training of the machine learning model with additional data to increase the sensitivity or specificity compared to the most recently validated and released device while the same type and range of input signal is used. |
Page 14
| Modification #2[M#2]: modifications related to quantitative measures – technical performance specifications. | Modification of data preprocessing methodologies /data augmentation methodologies/ Architecture and hyper-parameters to improve the performance or the efficiency of the computational resources (running time, memory consumption and CPU utilization). |
| Modification #3[M#3]: modifications related to device inputs. | Expanding the algorithm to include new sources of the same signal type, new model/version of Tyto Otoscope that meet the minimum predefined otoscopic video acquisition and file requirements for compatibility, including MP4 format or comparable, resolution ≥640×480, field of view ≥45° diagonal and frame rate ≥10 fps. * No third-party otoscopes are included within the scope of this PCCP. |
This PCCP was reviewed and authorized by FDA. A testing plan was developed to verify that the modified Tyto Insights for Eardrum Bulging Detection device continues to ensure safety and effectiveness. For instance, algorithm performance will be evaluated relative to the most recently validated and released version of the device and performance goals will be based on the lower bounds of the two-sided 95% confidence intervals (LCI).
RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of the ear, nose, and throat image analyzer and the measures necessary to mitigate these risks
| Identified Risks to Health | Mitigation Measures |
|---|---|
| Incorrect results, whether • False positives, leading to unnecessary medical procedures or treatments; or • False negatives, leading to delayed patient treatment | Clinical performance testing Labeling |
| Overreliance or misinterpretation of results leading to incorrect patient management | Labeling Human factors testing |
| Software malfunction leading to inaccurate patient diagnosis and monitoring or failure to identify the correct condition | Clinical performance testing Software verification, validation, and hazard analysis |
Page 15
SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the ear, nose, and throat image analyzer is subject to the following special controls:
(1) Clinical performance testing must demonstrate the accuracy and precision of the device output to demonstrate that the device performs as intended under anticipated conditions of use in the intended patient population for the stated indications for use.
(2) Human factors/usability testing must demonstrate that the user can correctly use the device, based solely on reading the directions for use.
(3) Software verification, validation, and hazard analysis must be performed.
(4) Labeling must include:
(i) Hardware compatibility information;
(ii) A warning that the device is not intended to be used independently for diagnosis; and
(iii) A summary of the clinical performance testing conducted with the device, including the study endpoints and statistical confidence intervals.
BENEFIT-RISK DETERMINATION
The probable benefits of this device are to automatically detect eardrum bulging in Otoscopic video recordings (which would otherwise be detected visually, relying on clinical experience and judgment to interpret the tympanic membrane position and characteristics). Bulging is a medical characteristic of the tympanic membrane that is commonly used for the diagnosis of Acute Otitis Media, but is also observed in ear trauma, barotrauma, bullous myringitis, etc. Therefore, the device can provide an aid to diagnosis when used in conjunction with otoscopic ear examinations and when considering all patient symptoms and medical history.
There is a risk of inappropriate patient management in the setting of inaccurate output from the device, which can lead to either incorrect, unnecessary, or delayed patient treatment/medication. However, labeling can mitigate these risks such as by including a warning reminding the lay user not to rely on the analysis results for diagnosis and treatment and to consult with a clinician, and to inform the clinician to consider the results of the device software output in conjunction with the otoscopic recording and with all other relevant patient data (and therefore not to solely rely on the device for diagnosis).
Based on the totality of the information provided, including the non-clinical testing/validation, clinical performance, and human factors data, there is sufficient evidence to support concluding that the probable benefit of the Tyto Insights for Eardrum Bulging Detection exceeds the probable risks for its use to automatically detect the bulging of the eardrum in otoscopic video recordings.
PATIENT PERSPECTIVES
This submission did not include specific information on patient perspectives for this device.
BENEFIT/RISK CONCLUSION
Page 16
In conclusion, given the available information above, for the following indication statement:
Tyto Insights for Eardrum Bulging Detection is an over-the-counter web-based AI-enabled software analytics device intended to automatically detect the bulging of the eardrum in Otoscopic video recordings acquired by the compatible electronic otoscope, as identified in the product labeling, in patients aged 6 months and older. The device is not intended for diagnosis. A healthcare provider's advice is required to understand the meaning of the Tyto Insights for Eardrum Bulging Detection result. Healthcare providers should consider the device result in conjunction with otoscopic recording and other relevant patient data. The device is not intended to detect other abnormal otoscopic findings.
The probable benefits outweigh the probable risks for the Tyto Insights for Eardrum Bulging Detection. The device provides benefits and the risks can be mitigated by the use of general controls and the identified special controls.
CONCLUSION
The De Novo request for the Tyto Insights for Eardrum Bulging Detection is granted and the device is classified as follows:
Product Code: SHL
Device Type: Ear, nose, and throat image analyzer
Regulation Number: 21 CFR 874.4475
Class: Class II
N/A