K Number
DEN160032

Validate with FDA (Live)

Date Cleared
2017-02-23

(227 days)

Product Code
Regulation Number
866.1650
Type
Direct
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Accelerate PhenoTest BC kit is a multiplexed in vitro diagnostic test utilizing both qualitative nucleic acid fluorescence in situ hybridization (FISH) identification and quantitative, antimicrobial susceptibility testing (AST) methods and is intended for use with the Accelerate Pheno system. The Accelerate PhenoTest BC kit is capable of simultaneous detection and identification of multiple microbial targets followed by susceptibility testing of the appropriate detected bacterial organisms. The Accelerate PhenoTest BC kit is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system. Results are intended to be interpreted in conjunction with Gram stain results.

The Accelerate PhenoTest BC kit identifies the following Gram-positive and Gram-negative bacteria and yeasts utilizing FISH probes targeting organism-specific ribosomal RNA sequences: Staphylococcus aureus, Staphylococcus lugdunensis, Coagulase-negative Staphylococcus species (i.e., Staphylococcus epidermidis, Staphylococcus haemolyticus, Staphylococcus hominis, Staphylococcus capitis, Staphylococcus lugdunensis, Staphylococcus warneri, not differentiated), Enterococcus faecalis, Enterococcus faecium, Streptococcus spp. (i.e., Streptococcus mitis, Streptococcus oralis, Streptococcus gallolyticus, Streptococcus agalactiae, Streptococcus pneumoniae, not differentiated), Pseudomonas aeruginosa, Acinetobacter baumannii, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated), Serratia marcescens, Candida albicans and Candida glabrata.

The Accelerate PhenoTest BC kit tests the following antimicrobial agents with the specific target organisms identified below:

Amikacin: Pseudomonas aeruginosa, Acinetobacter baumannii, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Ampicillin: Enterococcus faecalis and Enterococcus faecium
Ampicillin/Sulbactam: Escherichia coli, Klebsiella spp. (i.e., Klebsiella pneumoniae, . Klebsiella oxytoca, not differentiated), and Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated)
Aztreonam: Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated). Proteus spp. (i.e., Proteus mirabilis. Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Ceftazidime: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Ceftaroline : Staphylococcus aureus .
Cefepime: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Ceftriaxone: Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Ciprofloxacin: Pseudomonas aeruginosa, Klebsiella spp. (i.ess, Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Daptomycin: Staphylococcus aureus, Coagulase-negative Staphylococcus species (i.e., Staphylococcus epidermidis, Staphylococcus haemolyticus, Staphylococcus hominis, Staphylococcus capitis, Staphylococcus lugdunensis, Staphylococcus warneri, not differentiated), Enterococcus faecalis and Enterococcus faecium
Erythromycin: Staphylococcus aureus
Ertapenem: Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Gentamicin: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Linezolid: Staphylococcus aureus, Enterococcus faecalis and Enterococcus faecium .
Meropenem: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis. Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii. Citrobacter koseri, not differentiated) and Serratia marcescens
Piperacillin/Tazobactam: Pseudomonas aeruginosa, Acinetobacter baumannii, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Tobramycin: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella ● oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Vancomycin: Staphylococcus aureus, Staphylococcus lugdunensis, Coagulase-negative . Staphylococcus species (i.e., Staphylococcus epidermidis, Staphylococcus haemolyticus, Staphylococcus hominis, Staphylococcus capitis, Staphylococcus lugdunensis, Staphylococcus warneri, not differentiated), Enterococcus faecalis and Enterococcus faecium

The following resistance phenotypes are reported based on qualitative tests: Methicillin-resistance (S. aureus S. lugdunensis, coagulase negative staphylococci) and macrolide-lincosamide-streptogramin B resistance (MLSb) (S. lugdunensis and coagulase negative staphylococci).

The Accelerate PhenoTest BC kit is indicated as an aid in the diagnosis of bacteremia and fungemia. It is also indicated for susceptibility testing of specific pathogenic bacteria as identified above commonly associated with or causing bacteremia. Results are intended to be used in conjunction with other clinical and laboratory findings.

Standard laboratory protocols for processing positive blood cultures should be followed to ensure availability of isolates for supplemental testing as needed. Additionally, subculture of positive blood culture is necessary for the identification and susceptibility testing of: organisms not identified by the Accelerate PhenoTest BC kit, organisms present in polymicrobial samples, organisms for which species identification is critical for patient care (e.g., speciation of Streptococcus spp.), samples for which an "indeterminate" result for any probe was obtained, for testing antimicrobial agents not included on the Accelerate panel and for epidemiologic testing.

Device Description

The Accelerate Pheno system is comprised of the Accelerate Pheno instrument, software, host computer, analysis computer, and the Accelerate PhenoTest BC kit. The Accelerate PhenoTest BC Kit contains a sample vial, a 48-channel disposable test cassette and a reagent cartridge needed to test samples from a blood culture bottle that has been flagged as positive by a continuous monitoring blood culture system. All identification (ID) and antimicrobial susceptibility testing (AST) is performed in individual flowcells of the test cassette. The reagent cartridge contains gel electrofiltration (GEF) stations, fluorescence in situ hybridization (FISH) probes, antibiotics, and reagents for automated sample preparation, identification of bacterial and fungal target organisms (Table 1), and antimicrobial susceptibility testing and phenotypic resistance detection testing for bacterial target organisms (Tables 2 and 3). The user loads an aliquot of the positive blood culture into the sample vial, places the test cassette, reagent cartridge and sample vial into an Accelerate Pheno System module, and then presses the module button to close the module door and start the run. The remainder of the operations are automated as described below.

Automated Sample Preparation
Automated sample preparation is performed using gel electro-filtration (GEF) which is based on gel electrophoresis principles. The sample is automatically transferred to a gel containing pores smaller than bacterial or yeast cells. Application of an electric field causes lysed blood cells and/or other sample debris to pass into the gel while bacterial/yeast cells remain inside the gel well. The electric field is briefly reversed to dislodge bacterial/yeast cells from the gel wall prior to removal.

Cell Capture
Following sample preparation, recovered cells are automatically pipetted into multiple flowcell channels of the test cassette. Conductive layers of transparent indium tin oxide (ITO) coat the top and bottom inner surfaces of each flowcell channel and act as electrodes. An additional cationic poly-Llysine layer on the bottom of each flowcell acts as a capture surface. When a voltage is applied, the negatively-charged bacterial/yeast cells migrate to the positively-charged capture surface where they are captured prior to imaging.

Fluorescence in situ Hybridization (FISH) for Identification
Cocktails of ATTO-532 (green) fluorescently-labeled DNA probes bind to the ribosomal RNA of target organisms following permeabilization. Each cocktail also includes ATTO-647 (red) labeled universal bacterial probe that binds to the ribosomal RNA of all clinically relevant bacterial ID channels) or universal eukaryotic probe that binds to the ribosomal RNA of all clinically relevant yeast (yeast ID channels). The system images each flowcell using an epifluorescence microscope with camera at 532 nm, 647 nm and in dark-field. To exclude debris, only dark-field objects that are colocalized with universal probe signal are included in analysis. Colocalization of target probe signal and universal probe signal identifies a target organism.

The software also quantitates the total number of organisms present in a sample using a nucleic acid stain in a separate control flowcell. Comparing the relative numbers of each target organism to the number of objects lit up by the universal probes and universal nucleic acid stain allows for non-target organism and polymicrobial sample detection. FISH ID results are reported approximately 90 minutes after loading the sample, and the ID result determines the selection of appropriate antibiotics for subsequent antimicrobial susceptibility testing.

Morphokinetic Cellular Analysis (MCA) for Antimicrobial Susceptibility Testing (AST)
Sample remaining after the identification assay is initiated is combined with growth media and organisms contained in the sample undergo a pre-growth step during the FISH ID assay to normalize growth rates prior to AST. Following automated sample preparation and cell capture, growth media containing single concentrations of each test antibiotic are added to separate flowcell channels; antibiotics are selected based on the identification provided by the FISH identification (Tables 2 and 3). The bacteria in each flowcell are imaged every 10 minutes for up to 4.5 hours, creating a timelapse record of bacterial growth from individual progenitor cells into clones of daughter cells.

During this period. several microscopic features are measured through morphokinetic cellular analysis, such as cell morphology and the light intensity of a growing clone over time, and used for analysis. The precise quantitative measurement of individual clone growth rate over time is an indicator of antimicrobial efficacy. Onboard software algorithms derive minimum inhibitory concentration (MIC) values from the measured features, and apply appropriate expert rules for proper interpretation and reporting of categorical interpretations - S, I or R (susceptible, intermediate, or resistant) for MIC determinations and positive or negative for phenotypic resistance markers. AST results are reported in approximately 5 hours after ID results. The reportable ranges for each antimicrobial and phenotypic resistance markers are listed in Tables 4 and 5.

AI/ML Overview

The provided text describes the evaluation of the Accelerate PhenoTest BC Kit. Here's a breakdown of the acceptance criteria and study details:

1. Table of Acceptance Criteria & Reported Device Performance:

The document outlines performance metrics that serve as acceptance criteria for different aspects of the device.

Performance Metric CategorySpecific Metric (Acceptance Criteria Mentioned or Implied)Reported Device Performance (as stated in the text)
Identification (ID) AssayReproducibility (>95% for most probes)11 of 12 probe targets showed reproducibility > 95%. The Enterobacter probe (ENT) initially showed 87.5% but improved to 93.2% after a post-study imaging processing change and re-evaluation.
Growth & Detection (ID Consistency over time)Correct identifications obtained for all samples at both t=0 and t=8 hours post-positivity, validating the sponsor's claim for testing within 8 hours.
Analytical Inclusivity (Detection & Identification)All isolates included in the Inclusivity Study (Tables 15-17) were detected and correctly identified by the PhenoTest BC Kit.
Analytical Specificity (Exclusivity - Non-Cross Reactive)Organisms in Tables 18 and 19 (non-cross-reacting, some with indeterminate results) were tested. In silico analysis also predicted some cross-reactivity (Table 20).
Limit of Detection (LoD) (Reliable detection >95%)In most cases, the LoD is at or below the concentration of organisms present when blood cultures are determined to be positive. Average LoD: Gram-negative (4 x 10^8 CFU/mL), Gram-positive (5 x 10^8 CFU/mL), Candida sp. (2 x 10^6 CFU/mL). S. agalactiae LoD was not established. (Tables 21-23).
Interference Studies (ID)99.5% agreement with expected results; all samples returned an ID result with endogenous substances and heparin.
Blood Bottle Type (ID) (>95% detected & correctly identified)For 12 of 13 bottle types, the ID assay provided acceptable results (>95% of organisms detected and correctly identified). One bottle type (BACTEC PLUS Anaerobic F) detected and identified 94.3% of aerobic isolates.
Polymicrobial LoD (ID)Acceptable detection and identification of all isolates in various concentrations, except S. aureus (with C. albicans at high concentration) was not detected.
Biological Interference (ID)15 of 17 polymicrobial combinations resulted in 100% detection of all organisms. S. aureus and K. pneumoniae (with C. albicans at high concentration) were not detected.
Clinical Sensitivity/PPA (ID)Gram-Positive: CNS (95.3%), EFS (97.0%), EFM (98.0%), SAU (97.9%), SLU (97.5%), STR (97.2%). Gram-Negative: ABA (98.6%), CIT (96.8%), ENT (97.3%), ECO (97.3%), KLE (96.1%), PRO (97.7%), PAE (100%), SMA (100%). Candida spp.: CAL (100%), CGL (100%). (Tables 27-29)
Clinical Specificity/NPA (ID)Gram-Positive: CNS (98.2%), EFS (99.9%), EFM (99.1%), SAU (98.5%), SLU (99.9%), STR (97.6%). Gram-Negative: ABA (99.7%), CIT (99.3%), ENT (99.5%), ECO (99.7%), KLE (99.6%), PRO (99.6%), PAE (99.4%), SMA (99.9%). Candida spp.: CAL (99.6%), CGL (98.4%). (Tables 27-29)
Antimicrobial Susceptibility Testing (AST) AssayReproducibility (>95% acceptable, some exceptions)17 of 18 antimicrobials and resistance phenotypes had best case reproducibility > 95%. Erythromycin had best and worst case reproducibility of 93.6%, which was considered acceptable. (Table 7).
Growth & Detection (AST Consistency and Agreement)For parameter 1 (consistency of MIC values): 421/429 (98%) repeatability of all MIC values. For parameter 2 (EA and CA vs reference): All antimicrobials showed EA and CA ≥ 89.9% except for some specific antimicrobial/organism combinations with lower performance, addressed by limitations. (Table 14).
Interference Studies (AST)Endogenous substances/organism/antimicrobial combinations demonstrated >89.9% EA for 98% (352/360) and >89.9% CA for 94% (340/360) of combinations.
Blood Bottle Type (AST)All bottle types provided acceptable EA and CA values for most organisms. MLSb detection had insufficient data for some bottle types, leading to specific limitations.
Biological Interference (AST)AST testing of two organisms was not supported; only one organism in a polymicrobic sample will be tested.
Clinical Essential Agreement (EA)Ampicillin (100.0%), Ceftaroline (93.3%), Daptomycin-Staphylococcus (99.1%), Daptomycin-Enterococcus (95.5%), Erythromycin (98.2%), Linezolid-S. aureus (99.5%), Linezolid-Enterococcus (96.4%), Vancomycin-S. aureus (98.0%), Vancomycin-others (96.4%), Amikacin (94.2%), Ampicillin/Sulbactam (91.0%), Aztreonam (96.6%), Cefepime-Enterobacteriaceae (97.7%), Cefepime-P. aeruginosa (88.1%), Ceftazidime-Enterobacteriaceae (86.2%), Ceftazidime-P. aeruginosa (86.8%), Ceftriaxone (89.8%), Ciprofloxacin (96.7%), Ertapenem (98.9%), Gentamicin (99.2%), Meropenem-Enterobacteriaceae (97.0%), Meropenem-P. aeruginosa (88.2%), Piperacillin/Tazobactam (92.1%), Tobramycin (96.4%). (Tables 32-33)
Clinical Categorical Agreement (CA)Ampicillin (99.6%), Ceftaroline (99.7%), Daptomycin-Staphylococcus (99.7%), Daptomycin-Enterococcus (99.1%), Erythromycin (96.7%), Linezolid-S. aureus (100.0%), Linezolid-Enterococcus (98.2%), Vancomycin-S. aureus (99.0%), Vancomycin-others (96.4%), Amikacin (94.0%), Ampicillin/Sulbactam (84.2%), Aztreonam (97.7%), Cefepime-Enterobacteriaceae (96.8%), Cefepime-P. aeruginosa (92.9%), Ceftazidime-Enterobacteriaceae (93.9%), Ceftazidime-P. aeruginosa (88.7%), Ceftriaxone (96.6%), Ciprofloxacin (98.2%), Ertapenem (98.6%), Gentamicin (98.4%), Meropenem-Enterobacteriaceae (98.1%), Meropenem-P. aeruginosa (90.2%), Piperacillin/Tazobactam (92.1%), Tobramycin (96.1%). (Tables 32-33)
Clinical Phenotypic Resistance (CA)Cefoxitin (98.2%), MLSb (98.2%). (Table 34)

2. Sample Size Used for the Test Set and the Data Provenance:

  • Clinical Study Test Set:

    • Total 1850 positive blood culture samples were evaluated for clinical performance.
    • Provenance:
      • 793 fresh samples (aliquots of left-over positive blood cultures from patients suspected of bacteremia or fungemia). These were prospectively collected in the U.S.
      • 65 fresh seeded samples (blood cultures seeded with human blood and previously characterized fresh clinical isolates isolated within seven days of seeding). These were prospectively collected in the U.S.
      • 477 samples seeded with challenge isolates (obtained from culture collections).
      • 515 samples seeded with stock isolates (obtained from clinical specimens at the clinical sites and stored for longer than seven days).
      • Study conducted at 13 geographically distinct U.S. sites.
  • Analytical Studies (various test sets):

    • Reproducibility: At least 90 data points per probe target for ID, ~10 organisms with on-scale MICs for AST (total 1644 tests).
    • Growth & Detection: 21 on-panel organisms and 1 off-panel organism tested in triplicate (total 175 tests).
    • Inclusivity & Exclusivity: Three strains of each on-panel target species for inclusivity (total 253 tests for ID section). Individual representative strains for exclusivity (total 318 tests for ID section).
    • LoD: Varied by organism, usually >20 replicates per strain (e.g., 22/22, 21/22 etc. reported). Total 849 tests for monomicrobial LoD. Total 182 for polymicrobial LoD.
    • Interference: Varied by organism/substance (total 526 tests).
    • Blood Bottle Type: Minimum of 10 replicates per sample/bottle type (total 1419 tests overall for this study).
    • Polymicrobial LoD: 13 pairs of different microbial species, tested in duplicate (total 182 tests).
    • Biological Interference: 17 combinations tested (total 49 tests).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

The document does not explicitly state the number or qualifications of experts used to establish the ground truth for the test set.

  • For AST, the ground truth was established by CLSI broth microdilution reference method. The isolates were tested in triplicate by this method, with the mode MIC value serving as the reference.
  • For ID, the ground truth was VITEK2 (bioMérieux). For isolates where VITEK2 yielded an invalid result, 16S rRNA gene sequencing was used as the reference identification.
  • For phenotypic resistance markers (cefoxitin and MLSb), the ground truth was disk diffusion performed singly, with triplicate testing and modal category if the initial test failed or was near the breakpoint.

The implication is that these reference methods themselves represent an "expert consensus" or established standard, rather than a panel of human experts explicitly reviewing each case's ground truth.

4. Adjudication Method for the Test Set:

The document does not describe an explicit adjudication method (e.g., 2+1, 3+1) for discrepant results between the device and the reference methods. The reference methods (CLSI broth microdilution, VITEK2, 16S rRNA gene sequencing, and disk diffusion) are considered the definitive ground truth.

For AST, if no mode value could be determined or if a QC failure occurred for the reference method, an additional three replicates were tested to determine the mode MIC. This implies a process to ensure a robust reference result, but not an independent adjudication by experts comparing device output to reference.

5. If a Multi Reader Multi Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was NOT done. This device is an automated, in vitro diagnostic test for microbial identification and antimicrobial susceptibility, not an AI-assisted diagnostic imaging tool that would typically involve human readers. Its performance is compared directly to established laboratory reference methods.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Yes, the primary clinical and analytical performance studies were effectively standalone (algorithm only) performance studies. The Accelerate PhenoTest BC Kit is described as a "fully-integrated and fully automated system" that uses "onboard software algorithms" to derive results. The performance metrics (accuracy, sensitivity, specificity, EA, CA) are calculated by comparing the device's automated results to established reference methods, without human interpretation of the device's raw data directly influencing the reported output. The system is designed to provide final identification and AST results automatically.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The ground truth for the studies was based on:

  • Reference Laboratory Methods:
    • Identification (ID): Primarily VITEK2 (bioMérieux). For invalid VITEK2 results, 16S rRNA gene sequencing was used.
    • Antimicrobial Susceptibility Testing (AST): CLSI broth microdilution reference method.
    • Phenotypic Resistance: Disk diffusion.
  • Known Characteristics of Seeded Isolates: For a significant portion of the test set (fresh seeded, challenge, and stock isolates), the identity and susceptibility profiles were "previously characterized" or "known characteristics" of the spiked organisms. This implies a highly controlled and validated ground truth for these samples.

This is a form of reference standard ground truth, using established, highly accurate laboratory methods as the benchmark rather than direct patient outcomes or a panel of human experts interpreting raw data.

8. The sample size for the training set:

The document does not explicitly state the sample size for the training set. The evaluation focuses on the performance of the developed device using a test set against reference methods. It describes "onboard software algorithms" but doesn't detail their development or the data used for internal training/validation.

9. How the ground truth for the training set was established:

Since the training set size is not provided, the method for establishing its ground truth is also not detailed. However, given the nature of the device and the performance studies, it is highly probable that any data used for training/algorithm development would have had its ground truth established through similar rigorous microbiological reference methods (e.g., CLSI, sequencing, culture).

{0}------------------------------------------------

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR Accelerate PhenoTest BC Kit

DECISION SUMMARY

A. DEN Number:

DEN160032

B. Purpose for Submission:

De novo request for evaluation of automatic class III designation for the Accelerate PhenoTest BC Kit.

C. Measurand:

The following Gram-positive and Gram-negative bacteria and yeasts are identified using the Accelerate PhenoTest BC kit Staphylococcus aureus, Staphylococcus lugdunensis, Coaeulasenegative Staphylococcus species (i.e., Staphylococcus evidermidis, Staphylococcus haemolvticus, Staphylococcus hominis, Staphylococcus capitis, Staphylococcus lugdunensis, Staphylococcus warnerii, not differentiated), Enterococcus faecalis, Enterococcus faecium, Streptococcus spp. (i.e., Streptococcus mitis, Streptococcus oralis, Streptococcus gallolyticus, Streptococcus agalactiae, Streptococcus pneumoniae, not differentiated), Pseudomonas aeruginosa, Acinetobacter baumannii, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated), Serratia marcescens, Candida albicans and Candida glabrata.

The Accelerate PhenoTest BC kit tests the following antimicrobial agents as appropriate for the identified organism (see section H below): amikacin, ampicillin/sulbactam, aztreonam, ceftazidime, ceftaroline, ceftriaxone, ciprofloxacin, daptomycin, erythromycin, ertapenem, gentamicin, linezolid, meropenem, piperacillin/tazobactam, tobramycin and vancomycin.

The following resistance phenotypes are reported based on qualitative tests as appropriate for the identified organism (see section H below): Methicillin-resistance and macrolide-lincosamidestreptogramin B resistance (MLSb).

D. Type of Test:

The Accelerate PhenoTest BC kit is a multiplexed in vitro diagnostic test utilizing both qualitative nucleic acid fluorescence in situ hybridization (FISH) identification and quantitative, antimicrobial susceptibility testing (AST) methods and is intended for use with the Accelerate Pheno System. The PhenoTest BC assay is performed directly on positive blood culture samples identified as positive by a continuous monitoring blood culture system.

E. Applicant:

Accelerate Diagnostics

  • F. Proprietary and Established Names: Accelerate PhenoTest BC Kit

{1}------------------------------------------------

Accelerate Pheno system

G. Regulatory Information:

    1. Regulation section: 866.1650
    1. Classification: Class II
    1. Product code: PRH, NSU, PEO, PAM, PEN, LON
    1. Panel: 83 (Microbiology)

H. Indications for Use:

    1. Indications for Use:
      The Accelerate PhenoTest BC kit is a multiplexed in vitro diagnostic test utilizing both qualitative nucleic acid fluorescence in situ hybridization (FISH) identification and quantitative, antimicrobial susceptibility testing (AST) methods and is intended for use with the Accelerate Pheno system. The Accelerate PhenoTest BC kit is capable of simultaneous detection and identification of multiple microbial targets followed by susceptibility testing of the appropriate detected bacterial organisms. The Accelerate PhenoTest BC kit is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system. Results are intended to be interpreted in conjunction with Gram stain results.

The Accelerate PhenoTest BC kit identifies the following Gram-positive and Gram-negative bacteria and yeasts utilizing FISH probes targeting organism-specific ribosomal RNA sequences: Staphylococcus aureus, Staphylococcus lugdunensis, Coagulase-negative Staphylococcus species (i.e., Staphylococcus epidermidis, Staphylococcus haemolyticus, Staphylococcus hominis, Staphylococcus capitis, Staphylococcus lugdunensis, Staphylococcus warneri, not differentiated), Enterococcus faecalis, Enterococcus faecium, Streptococcus spp. (i.e., Streptococcus mitis, Streptococcus oralis, Streptococcus gallolyticus, Streptococcus agalactiae, Streptococcus pneumoniae, not differentiated), Pseudomonas aeruginosa, Acinetobacter baumannii, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated), Serratia marcescens, Candida albicans and Candida glabrata.

The Accelerate PhenoTest BC kit tests the following antimicrobial agents with the specific target organisms identified below:

  • Amikacin: Pseudomonas aeruginosa, Acinetobacter baumannii, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens

{2}------------------------------------------------

  • . Ampicillin: Enterococcus faecalis and Enterococcus faecium
  • Ampicillin/Sulbactam: Escherichia coli, Klebsiella spp. (i.e., Klebsiella pneumoniae, . Klebsiella oxytoca, not differentiated), and Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated)
  • . Aztreonam: Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated). Proteus spp. (i.e., Proteus mirabilis. Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
  • . Ceftazidime: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
  • Ceftaroline : Staphylococcus aureus .
  • . Cefepime: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
  • . Ceftriaxone: Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
  • . Ciprofloxacin: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
  • . Daptomycin: Staphylococcus aureus, Coagulase-negative Staphylococcus species (i.e., Staphylococcus epidermidis, Staphylococcus haemolyticus, Staphylococcus hominis, Staphylococcus capitis, Staphylococcus lugdunensis, Staphylococcus warneri, not differentiated), Enterococcus faecalis and Enterococcus faecium
  • . Erythromycin: Staphylococcus aureus
  • . Ertapenem: Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
  • . Gentamicin: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
  • Linezolid: Staphylococcus aureus, Enterococcus faecalis and Enterococcus faecium .
  • . Meropenem: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii. Citrobacter koseri, not

{3}------------------------------------------------

differentiated) and Serratia marcescens

  • . Piperacillin/Tazobactam: Pseudomonas aeruginosa, Acinetobacter baumannii, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
  • Tobramycin: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella ● oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
  • Vancomycin: Staphylococcus aureus, Staphylococcus lugdunensis, Coagulase-negative . Staphylococcus species (i.e., Staphylococcus epidermidis, Staphylococcus haemolyticus, Staphylococcus hominis, Staphylococcus capitis, Staphylococcus lugdunensis, Staphylococcus warneri, not differentiated), Enterococcus faecalis and Enterococcus faecium

The following resistance phenotypes are reported based on qualitative tests: Methicillin-resistance (S. aureus S. lugdunensis, coagulase negative staphylococci) and macrolide-lincosamide-streptogramin B resistance (MLSb) (S. lugdunensis and coagulase negative staphylococci).

The Accelerate PhenoTest BC kit is indicated as an aid in the diagnosis of bacteremia and fungemia. It is also indicated for susceptibility testing of specific pathogenic bacteria as identified above commonly associated with or causing bacteremia. Results are intended to be used in conjunction with other clinical and laboratory findings.

Standard laboratory protocols for processing positive blood cultures should be followed to ensure availability of isolates for supplemental testing as needed. Additionally, subculture of positive blood culture is necessary for the identification and susceptibility testing of: organisms not identified by the Accelerate PhenoTest BC kit, organisms present in polymicrobial samples, organisms for which species identification is critical for patient care (e.g., speciation of Streptococcus spp.), samples for which an "indeterminate" result for any probe was obtained, for testing antimicrobial agents not included on the Accelerate panel and for epidemiologic testing.

  • Special conditions for use statement(s): 2. For prescription use only
    Limitations:

General Limitations

  • This product can only be used with the Accelerate Pheno system.
  • . The Accelerate PhenoTest BC kit assay has not been evaluated for specimens other than blood (e.g., sterile body fluids inoculated into blood culture bottles)
  • . The performance of this test has only been evaluated using the following blood culture bottles:
    • BD BACTEC Standard/10 Aerobic/F Medium, O
    • BD BACTEC Standard/10 Anaerobic/F Medium, о
    • BD BACTEC Lytic/10 Anaerobic/F Medium O
    • BD BACTEC PEDS PLUS/F Medium O
    • BD BACTEC Plus Aerobic/F Medium O
    • BD BACTEC Plus Anaerobic/F Medium O

{4}------------------------------------------------

  • BioMérieux BacT/ALERT SA Standard Aerobic O
  • BioMérieux BacT/ALERT SN Standard Anaerobic O
  • o BioMérieux BacT/ALERT FA Plus Aerobic
  • BioMérieux BacT/ALERT FN Plus Anaerobic O
  • BioMérieux BacT/ALERT PF Plus O
  • о Versa TREK REDOX 1 (Aerobic) Medium and
  • Versa TREK REDOX 2 (Anaerobic) Medium O
  • This product should not be used with blood culture bottles containing charcoal.
  • Positive blood culture samples must be run using the Accelerate PhenoTest BC kit on the ● Accelerate Pheno system within 8 hours of sample positivity.
  • . Positive blood culture sample must be loaded on the Accelerate Pheno system and the run must be initiated within 15 minutes of pipetting sample vial and within 1 hour of removing the assay kit from refrigerated storage.
  • . Failure to observe proper procedures for sample collection, preparation, storage, handling and/or transportation may cause incorrect results.

Identification (ID) Limitations

  • Due to the possibility of cross reactivity, all Accelerate PhenoTest BC kit results should ● be interpreted in conjunction with Gram stain.
  • . The ability of probes to detect all strains of a target species was not predicted by in-silico analysis.
  • . Additional subculture is required for the identification of S. pneumoniae in cases of a positive Streptococcus spp. call.
  • . Subculture of positive blood culture is required in the following situations:
    • O For the identification and susceptibility testing of off-panel organisms not identified by the Accelerate PhenoTest BC kit,
    • For samples that give a polymicrobial result о
    • For organisms for which species identification is critical for patient care. (e.g., о speciation of streptococci)
    • For testing antimicrobial agents not included on the Accelerate panel O
    • For testing certain antimicrobial agents as discussed in AST limitations below O
    • For testing samples for which an "indeterminate" result for any probe was O obtained
    • To obtain isolates for epidemiologic testing. o
  • Accelerate PhenoTest BC kit identification results that are discordant with the result of . the blood culture Gram stain (for example, no organism detection when the Gram stain is positive or detection of a Gram-positive cocci when Gram-positive cocci were not observed in the Gram stain) should be confirmed by an alternate technique prior to reporting the test result. For some polymicrobic calls, false positive results may not be mitigated by Gram stain analysis (for example, detection of two Enterobacteriaceae species with Gram-negative rods observed in the Gram stain). Results of such polymicrobic calls should be verified by subculture and/or an alternative identification method.

Antimicrobial Susceptibility Testing (AST) Limitations

  • Due to insufficient number of test isolates, the ability of the Accelerate PhenoTest BC kit to detect inducible MLSb resistance in coagulase-negative staphylococci is unknown when used with the following blood culture bottle types: BacT/Alert SN Standard Anaerobic, BACTEC Peds Plus/F, BACTEC Plus Anaerobic/F, BACTEC Standard Anaerobic, BACTEC Standard/10 Aerobic, VersaTrek Redox 1 Aerobic, VersaTrek

{5}------------------------------------------------

Redox 2 Anaerobic. Use an alternative method for detection of inducible MLSb resistance when using these blood culture bottle types if critical to patient care.

  • . Susceptibility testing of monomicrobial samples will only be performed when on-panel species eligible for susceptibility testing are detected. See intended use.
  • Susceptibility testing of polymicrobial samples will only be performed on one organism out of a pair of species that meet all of the following criteria:
    • One or both organisms must be on the Accelerate PhenoTest BC kit test panel and eligible for AST testing (except Proteus spp.)
    • The two organisms must have distinct growth responses or morphological o differences such that growing clones from each species can be differentiated by the software. These include the following pairs:

Polymicrobial Testing Pairs

Organism 1Organism 2
• AST-eligible organism(except Proteus spp.)• AST-ineligible organism
• Escherichia coli• Klebsiella spp.• Enterobacter spp.• Citrobacter spp.or• Serratia marcescens• Staphylococcus aureus• Coagulase-negative Staphylococcus spp.• Enterococcus spp. (E. faecalis, E. faecium)or• Pseudomonas aeruginosa
• Acinetobacter baumannii• Enterococcus spp. (E. faecalis, E. faecium)
  • Only one organism when diluted for AST must be within the required O concentration limits for AST testing (10-130 growing clones per field of view).
    If the concentration ratio between organisms is such that only one organism could be diluted to the concentration range required for AST testing, AST testing will only be performed on the higher concentration organism. If one of the organisms in the pair is eligible for AST testing and the other is not, AST testing will only be performed on the on-panel AST-eligible organism. If both organisms are eligible for AST testing and are within the required concentration limits for AST testing, AST results will not be reported.

  • . If an AST result is not provided by the Accelerate PhenoTest BC kit, susceptibility testing must be performed using an alternative method.

  • Subculturing of positive blood culture is necessary for the identification and . susceptibility testing of organisms not identified by the Accelerate PhenoTest BC kit and for antimicrobial agents not included on the Accelerate panel.

  • . Potential interference by antimicrobial agents that may be present in a patient blood specimen has not been established with the Accelerate PhenoTest BC kit.

  • . The ability of the Accelerate PhenoTest BC kit to detect resistance in the following combinations is unknown because an insufficient number of resistant isolates were encountered at the time of comparative testing:

    • o Amikacin: Citrobacter spp., Enterobacter spp., E. coli, Proteus spp., S. marcescens

{6}------------------------------------------------

  • Aztreonam: Proteus spp., S. marcescens O
  • Cefepime: Citrobacter spp., Proteus spp., S. marcescens O
  • o Ceftazidime: Proteus spp., S. marcescens
  • Ceftaroline: S. aureus O
  • Ceftriaxone: Citrobacter spp., E. cloacae, S. marcescens O
  • O Ciprofloxacin: Citrobacter spp., Proteus spp., S. marcescens
  • Daptomycin: S. aureus O
  • Ertapenem: Citrobacter spp., Proteus spp., and S. marcescens O
  • Gentamycin: Citrobacter spp., Enterobacter spp., Proteus spp., S. marcescens o
  • Meropenem: Citrobacter spp., E. coli, Proteus spp., and S. marcescens O
  • Piperacillin/Tazobactam: Proteus spp., and S. marcescens о
  • Tobramycin: Citrobacter spp., Proteus spp., S. marcescens O
  • Cefoxitin for Phenotypic Resistance: S.lugdunensis о
  • MLSb: S. lugdunensis O
  • The following antimicrobial/organism combinations may produce a resistant result that ● can be found susceptible by the reference method. If critical to patient care confirm these results with an alternate method:
    • o Meropenem: Enterobacter
    • Ceftazidime: Pseudomonas aeruginosa (Any P. aeruginosa isolate that provides O an MIC ≥16 ug/mL should be retested using an alternate method)
    • o Cefepime: Pseudomonas aeruginosa
    • Ertapenem: Enterobacter spp. o
    • Piperacillin/Tazobactam: Acinetobacter baumannii, Klebsiella spp. o
  • The ability of the Accelerate PhenoTest BC kit to provide accurate MICs with amikacin ● resistant strains of A. baumannii has not been established; isolates of this species that provide resistant results should be confirmed by an alternative method.
  • Due to a low essential agreement for Serratia marcescens with ceftriaxone, results should . be confirmed with an alternate method if critical to patient care.
  • The current absence of data on daptomycin-resistant isolates precludes defining any . categories other than "susceptible." Isolates yielding test results suggestive of a nonsusceptible category should be retested and if the result is confirmed, the isolate should be retested using the reference method.
  • . The ability of the Accelerate PhenoTest BC kit to detect vancomycin-intermediate Staphylococcus aureus isolates (VISA) is unknown because insufficient numbers of VISA isolates were evaluated at the time of comparative testing.
  • Any S. aureus isolate for which the vancomycin MIC is >= 8 ug/mL should be sent to a ● reference laboratory for reference method testing.
  • Any coagulase negative Staphylococcus isolate for which the vancomycin MIC is >= 32 . ug/mL should be sent to a reference laboratory for reference method testing.
  • The ability of the Accelerate PhenoTest BC kit to provide accurate results for S. aureus ● with MLSb as compared to the reference method has not been established; isolates of this species should be tested by an alternative method.

3. Special instrument requirements:

The Accelerate PhenoTest BC Kit is performed on the Accelerate Pheno System.

I. Device Description:

The Accelerate Pheno system is comprised of the Accelerate Pheno instrument, software, host computer, analysis computer, and the Accelerate PhenoTest BC kit. The Accelerate PhenoTest BC

{7}------------------------------------------------

Kit contains a sample vial, a 48-channel disposable test cassette and a reagent cartridge needed to test samples from a blood culture bottle that has been flagged as positive by a continuous monitoring blood culture system. All identification (ID) and antimicrobial susceptibility testing (AST) is performed in individual flowcells of the test cassette. The reagent cartridge contains gel electrofiltration (GEF) stations, fluorescence in situ hybridization (FISH) probes, antibiotics, and reagents for automated sample preparation, identification of bacterial and fungal target organisms (Table 1), and antimicrobial susceptibility testing and phenotypic resistance detection testing for bacterial target organisms (Tables 2 and 3). The user loads an aliquot of the positive blood culture into the sample vial, places the test cassette, reagent cartridge and sample vial into an Accelerate Pheno System module, and then presses the module button to close the module door and start the run. The remainder of the operations are automated as described below.

Automated Sample Preparation

Automated sample preparation is performed using gel electro-filtration (GEF) which is based on gel electrophoresis principles. The sample is automatically transferred to a gel containing pores smaller than bacterial or yeast cells. Application of an electric field causes lysed blood cells and/or other sample debris to pass into the gel while bacterial/yeast cells remain inside the gel well. The electric field is briefly reversed to dislodge bacterial/yeast cells from the gel wall prior to removal.

Cell Capture

Following sample preparation, recovered cells are automatically pipetted into multiple flowcell channels of the test cassette. Conductive layers of transparent indium tin oxide (ITO) coat the top and bottom inner surfaces of each flowcell channel and act as electrodes. An additional cationic poly-Llysine layer on the bottom of each flowcell acts as a capture surface. When a voltage is applied, the negatively-charged bacterial/yeast cells migrate to the positively-charged capture surface where they are captured prior to imaging.

Fluorescence in situ Hybridization (FISH) for Identification

Cocktails of ATTO-532 (green) fluorescently-labeled DNA probes bind to the ribosomal RNA of target organisms following permeabilization. Each cocktail also includes ATTO-647 (red) labeled universal bacterial probe that binds to the ribosomal RNA of all clinically relevant bacterial ID channels) or universal eukaryotic probe that binds to the ribosomal RNA of all clinically relevant yeast (yeast ID channels). The system images each flowcell using an epifluorescence microscope with camera at 532 nm, 647 nm and in dark-field. To exclude debris, only dark-field objects that are colocalized with universal probe signal are included in analysis. Colocalization of target probe signal and universal probe signal identifies a target organism.

The software also quantitates the total number of organisms present in a sample using a nucleic acid stain in a separate control flowcell. Comparing the relative numbers of each target organism to the number of objects lit up by the universal probes and universal nucleic acid stain allows for non-target organism and polymicrobial sample detection. FISH ID results are reported approximately 90 minutes after loading the sample, and the ID result determines the selection of appropriate antibiotics for subsequent antimicrobial susceptibility testing.

Probe SetOn-panel Species
ABAAcinetobacter baumannii
CALCandia albicans
CGLCandida glabrata
CITCitrobacter freundii

Table 1. Probe Sets and Species Identified by Each Probe Set

{8}------------------------------------------------

Probe SetOn-panel Species
CNSCitrobacter koseriStaphylococcus epidermidisStaphylococcus haemolyticusStaphylococcus hominisStaphylococcus capitisStaphylococcus lugdunensisStaphylococcus warneri
ECOEscherichia coli
EFMEnterococcus faecium
EFSEnterococcus faecalis
ENTEnterobacter aerogenesEnterobacter cloacae
KLEKlebsiella oxytocaKlebsiella pneumoniae
PAEPseudomonas aeruginosa
PROProteus mirabilisProteus vulgaris
SAUStaphylococcus aureus
SLUStaphylococcus lugdunensis
SMASerratia marcescens
STRStreptococcus mitisStreptococcus oralisStreptococcus gallolyticusStreptococcus agalactiaeStreptococcus pneumoniae

Morphokinetic Cellular Analysis (MCA) for Antimicrobial Susceptibility Testing (AST)

Sample remaining after the identification assay is initiated is combined with growth media and organisms contained in the sample undergo a pre-growth step during the FISH ID assay to normalize growth rates prior to AST. Following automated sample preparation and cell capture, growth media containing single concentrations of each test antibiotic are added to separate flowcell channels; antibiotics are selected based on the identification provided by the FISH identification (Tables 2 and 3). The bacteria in each flowcell are imaged every 10 minutes for up to 4.5 hours, creating a timelapse record of bacterial growth from individual progenitor cells into clones of daughter cells.

During this period. several microscopic features are measured through morphokinetic cellular analysis, such as cell morphology and the light intensity of a growing clone over time, and used for analysis. The precise quantitative measurement of individual clone growth rate over time is an indicator of antimicrobial efficacy. Onboard software algorithms derive minimum inhibitory concentration (MIC) values from the measured features, and apply appropriate expert rules for proper interpretation and reporting of categorical interpretations - S, I or R (susceptible, intermediate, or resistant) for MIC determinations and positive or negative for phenotypic resistance markers. AST results are reported in approximately 5 hours after ID results. The reportable ranges for each antimicrobial and phenotypic resistance markers are listed in Tables 4 and 5.

{9}------------------------------------------------

Algorithm for Performance of AST

For samples determined to be monomicrobic (only one pathogen detected), susceptibility testing will only be performed when one of the on-panel species eligible for susceptibility testing are detected. The following limitation is included in the device labeling:

Susceptibility testing of monomicrobial samples will only be performed when on-panel species eligible for susceptibility testing are detected.

For samples determined to be polymicrobic, susceptibility testing will only be performed on one organism out of a pair of species that meet all of the following criteria:

  • One or both organisms must be on the Accelerate PhenoTest BC kit test panel and . eligible for AST testing (except Proteus spp.)
  • The two organisms must have distinct growth responses or morphological differences . such that growing clones from each species can be differentiated by the software. These include the following pairs:
Organism 1Organism 2
AST-eligible organism (except Proteus spp.)AST-ineligible organism
Escherichia coli Klebsiella spp. Enterobacter spp. Citrobacter spp. orSerratia marcescensStaphylococcus aureus Coagulase-negative Staphylococcus spp. Enterococcus spp. (E. faecalis, E. faecium or other Enterococcus spp.) orPseudomonas aeruginosa
Acinetobacter baumanniiEnterococcus spp. (E. faecalis, E. faecium)
  • Only one organism when diluted for AST must be within the required concentration . limits for AST testing (10-130 growing clones per field of view).
    If the concentration ratio between organisms is such that only one organism could be diluted to the concentration range required for AST testing will only be performed on the higher concentration organism. If one of the organisms in the pair is eligible for AST testing and the other is not, AST testing will only be performed on the on-panel AST-eligible organism. If both organisms are eligible for AST testing and are within the required concentration limits for AST testing. AST results will not be reported.

Because the PhenoTest BC Kit will not be capable of performing AST testing on all organisms present in a positive blood culture, subculture is required to recover the additional organism(s) and to allow performance of AST by an alternate method. In addition, subculture is required to recover organisms not detected by the PhenoTest BC Kit and for testing antimicrobial agents not included in the kit. The following limitations are included in the device labeling:

If an AST result is not provided by the Accelerate PhenoTest BC kit, susceptibility testing must be performed using an alternative method.

{10}------------------------------------------------

Subculture of positive blood culture is required in the following situations:

  • o For testing antimicrobial agents not included on the Accelerate Panel
  • o For testing certain antimicrobial agents as discussed in AST limitations.

Table 2. Antimicrobial Agents Tested*, Gram-Positive Organisms

OrganismAmpicillinCeftarolineErythromycinDaptomycinLinezolidVancomycinCefoxitinMLSb
S. aureus-XXXXXX-
S. lugdunensis-----XXX
Coagulase negative Staphylococcus---X-XXX
Enterococcus faecalisX--XXX--
Enterococcus faeciumX--XXX--

*(X) = Tested, (-) = Not tested

Table 3. Antimicrobial Agents Tested*, Gram-Negative Organisms

OrganismAmpicillin-sulbactamPiperacillin-tazobactamCefepimeCeftazidimeCeftriaxoneErtapenemMeropenemAmikacinGentamicinTobramycinCiprofloxacinAztreonam
E. coliXXXXXXXXXXXX
Klebsiella spp.XXXXXXXXXXXX
Enterobacter spp.-XXXXXXXXXXX
Proteus spp.XXXXXXXXXXXX
Citrobacter spp.-XXXXXXXXXXX
S. marcescens-XXXXXXXXXXX
P. aeruginosa-XXX--XXXXXX
A. baumannii-X-----X----

*(X) = Tested, (-) = Not tested

{11}------------------------------------------------

AntimicrobialOrganisms TestedReportable Range inPhenoTest BC(µg/mL)FDA Breakpoints
SIR
Gram-Positive
AmpicillinEnterococcus spp.<2 to ≥ 32≤8-≥ 16
CeftarolineS. aureus≤0.25 to ≥ 8≤12≥4
DaptomycinStaphylococcus spp.≤0.25 to ≥ 4≤1-≥2
Enterococcus spp.< 1 to ≥ 16≤4-≥8
ErythromycinS. aureus<0.125 to ≥ 16≤0.51 - 4≥8
LinezolidS. aureus< 1 to ≥ 16≤4->8
Enterococcus spp.≤0.5 to ≥ 16≤24≥8
S. aureus≤0.5 to ≥ 32≤24 - 8≥16
VancomycinS. lugdunensis and CoagulaseNegative Staphylococcus< 1 to ≥ 64≤48-16≥32
Gram-Negative
AmikacinEnterobacteriaceae< 4 to ≥ 128≤1632≥ 64
P. aeruginosa
A. baumannii
Ampicillin/SulbactamaEnterobacteriaceae≤ 2 to ≥ 64≤816≥32
AztreonamEnterobacteriaceae< 1 to ≥ 32≤48≥ 16
P. aeruginosa
Enterobacteriaceae≤ 2 to ≥ 64≤816≥ 32
CefepimebP. aeruginosa
Enterobacteriaceae< 1 to ≥ 64<24 - 8≥ 16
P. aeruginosa
Enterobacteriaceae≤ 2 to ≥ 64≤8-≥ 16
CeftazidimecP. aeruginosa
Enterobacteriaceae< 1 to ≥ 32≤48≥ 16
P. aeruginosa
Enterobacteriaceae≤ 2 to ≥ 64≤8-≥ 16
CeftriaxonedP. aeruginosa
CeftriaxonedEnterobacteriaceae≤0.25 to ≥ 8≤12≥4
CiprofloxacineEnterobacteriaceae≤0.25 to ≥ 8≤12≥4
P. aeruginosa
ErtapenemEnterobacteriaceae≤0.125 to ≥4≤0.51≥2
GentamycinEnterobacteriaceae< 1 to ≥ 32≤48≥ 16
P. aeruginosa
Enterobacteriaceae≤0.25 to ≥ 8≤12≥4
MeropenemfP. aeruginosa
Enterobacteriaceae≤0.5 to ≥ 16≤24≥8
MeropenemfP. aeruginosa
Piperacillin/TazobactamgEnterobacteriaceae< 4 to ≥ 256≤1632->128
P. aeruginosa64
A. baumannii
TobramycinEnterobacteriaceae< 1 to ≥ 32≤48≥ 16
P. aeruginosa

Table 4. Reportable MIC Ranges and Organism-Specific Breakpoints for Antimicrobials Included in the PhenoTest BC

4 Truncated Range for ampicillin/sulbactam: PRO (4-64)

6 Truncated Range for cefepime: PAE (2-32)

€ Truncated Range for ceftazidime: CIT, ECO, ENT, KLE, PAE (2-32)

4 Truncated Range for ceftriaxone: PRO, SMA (0.5-8)

& Truncated Range for ciprofloxacin: ENT (0.5-8)

f Truncated Range for meropenem: ENT, PAE (1-16)

§ Truncated Range for piperacillin/tazobactam: PAE (8-256)

{12}------------------------------------------------

Resistance Phenotype TestID Target
SAUSLUCNS
Cefoxitin (MRSA)Pos/NegPos/NegPos/Neg
Erythromycin/Clindamycin (MLSB)N/APos/NegPos/Neg

Table 5. Results Applicable to Phenotypic Resistance Tests

Performing the Assay

The sample is collected from a positive blood culture using a sterile collection device. Blood culture samples should be tested within eight hours of the positive detection as determined by the continuous monitoring blood culture system. Samples may be stored at room temperature or the blood culture bottle can remain in the blood culture instrument until the sample is withdrawn for testing. The blood culture sample is vortexed and 0.5 mL is added to the sample vial. The test must be initiated within 15 minutes of placing the blood culture sample vial. The sample vial is inserted into the sample vial receptacle on the reagent cartridge, and the test cassette and reagent cartridge are placed into the Pheno instrument.

Test Results and Interpretation

Identification Assay. When the assay is complete, the Accelerate Pheno system displays the final results in a Patient Report or a Quality Control report. The "Identification Results" section lists all target groups tested and indicates which target group was detected. Target groups with positive and/or indeterminate results are printed at the top of the report in bold font. Identification results can be as follows: Positive (target group detected); Negative (target group not detected); and Indeterminate (result not defined, target group may or may not be present). An "invalid" result (a test for which too few cells are present in the control channel) is not specifically reported; for samples with invalid results the Patient Report will indicate "No ID results reported; too few cells for analysis. Perform alternate testing method for identification and susceptibility results."

When the PhenoTest BC kit detects an organism present in a positive blood culture for which there is no target probe in the PhenoTest BC kit (off-panel organism), the Patient Report indicates negative results for all target probes and instructs the user to perform an alternate testing method.

For samples for which two or more target species are identified, the Patient Report will recommend that culture should be performed "Recommend culture due to possibility of another organism being present."

The PhenoTest BC Kit does not provide speciation of Streptococcus spp. If the result indicates the presence of a Streptococcus species, and that additional testing should be performed to rule out Streptococcus pneumoniae. The following limitation is included in the device labeling:

Subculture is required for the identification of S. pneumoniae in cases of a positive Streptococcus spp. call.

AST Assay. Antimicrobial susceptibility results are listed on the Patient Report for a single target species. Only one AST result will be reported for any culture; susceptibility testing for additional organisms (if present) should be performed using an alternate method from subcultured sample. The following limitation is included in the device labeling:

{13}------------------------------------------------

If an AST result is not provided by the Accelerate PhenoTest BC kit, susceptibility testing must be performed using an alternative method.

The AST results included in the PhenoTest BC report will indicate the MIC, the breakpoints, and interpretive categories for the drugs for which testing was performed. Results for resistance phenotypes (cefoxitin and MLSb) are reported as positive or negative.

Quality Control.

For the identification assay the Quality Control Report indicates an overall status of "failed" and will list the identification results and pass/fail status for each tested species based on detection (positive or negative) and correct identification of the QC strain. For the AST assay the Quality Control report lists the QC range, MIC and pass/fail status for each tested antimicrobial agent. Pass/fail status is determined by the whether the MIC result falls within the QC range (pass) or outside of the QC range (fail).

Gram Stain Correlation

All identification results provided by the PhenoTest BC kit are intended to be interpreted in conjunction with results obtained from Gram stain of the positive bottle. Gram reaction (gram-positive or gram-negative) and cellular morphology (gram-positive cocci in clusters, pairs or chains, gramnegative rods) should be considered in the correlation with PhenoTest BC results. Gram stain results showing gram-negative cocco-bacilli may indicate the presence of Acinetobacter baumannii; however the cellular morphology is inconsistent or difficult to interpret for members of this genus. If the Gram stain results differ from the expected Gram stain morphology for the organism(s) identified by the PhenoTest BC kit, results should be confirmed with an alternate method. In addition, for any sample for which PhenoTest BC has indicated a monomicrobial result and for which the Gram stain of the positive blood culture shows multiple morphologies, results should be confirmed with an alternate method. All Patient Reports include the following footnote:

Identification results that are discordant with Gram stain should be confirmed with an alternate method.

Monomicrobial Call

The PhenoTest BC kit will report a monomicrobial call for samples for which a target organism is identified and for which there is no additional evidence of an additional organism (on- or off-panel). Results of all monomicrobial calls should be correlated with results from Gram stain of the positive blood culture bottle. Patient results for samples that are considered monomicrobial will include the following footnote:

Monomicrobial: sample positive for only one pathogen

J. Standard/Guidance Document Referenced:

  • CLSI EP-05-A3, Evaluation of Precision Performance of Quantitative Measurement, Approved ● Guideline, 2004
  • CLSI M100-S26, Performance Standards for Antimicrobial Susceptibility Testing, 2015 ●
  • CLSI M02-A12, Performance Standards for Antimicrobial Disk Susceptibility Testing, 2015 ●
  • CLSI EP07-A2, Interference Testing in Clinical Chemistry, Approved Guideline, 2005 ●

{14}------------------------------------------------

  • CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures, Approved Guideline, 2012
  • . CLSIM07-A10. Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically, Approved Standard, 2015
  • . EN 62304-2006, Medical Device Software - Software Life-Cycle Processes, 2006

K. Test Principle:

The Accelerate Pheno system uses an automated sample preparation and surface immobilization method to enable microscopy-based. single-cell analysis for identification (ID) and antimicrobial susceptibility testing (AST). Identification is accomplished via fluorescence in situ hybridization (FISH). Antimicrobial susceptibility testing uses microscopic observation of individual, live, growing bacterial cells in near real time (approximately every 10 minutes) in the presence of antimicrobial agents.

The Accelerate Pheno system employs automated sample and reagent pipetting, temperature controlled incubation, digital microscopy, image acquisition and analysis in an integrated and fully automated system.

L. Performance Characteristics:

1. Analytical performance:

a. Precision/Reproducibility:

Reproducibility studies of the Accelerate PhenoTest BC Kit for positive blood culture organism identification (ID) and antimicrobial susceptibility testing (AST) included evaluation of four replicates of the same positive blood culture sample on the same day at three testing sites for a total of 12 tests per blood culture sample. (Four replicates were initially tested to assure at least three valid results for each sample.) Samples were tested within eight hours of positivity. Primary probe targets utilized for reproducibility testing of the ID assay included representative species of each major organism group (S. aureus, S. pneumoniae, E. faecium, E. coli, A. baumannii and C. albicans). Additional species were tested in order to obtain the expected number of AST results (K. pneumoniae, P. aeruginosa, Citrobacter spp., Enterobacter spp., coagulase negative Staphylococcus, Proteus spp., and S. marcescens, and E. faecalis). A minimum of 90 data points were evaluated for each probe target.

The reproducibility of the organism identification assay was assessed on a per probe basis across all study sites and within each study site. A minimum of nine ID results (three per site) was evaluated for each sample. The ID results were classified as "correct" or "not correct" by comparing the observed ID to the known ID for each test strain.

Isolates that provided at least three ID results per site but which failed to produce at least three AST results per site were retested. All ID results were included in the analysis of probe performance. In order not to confound ID probe performance with overrepresentation of a single species, weighted percentages were calculated.

For the identification assay, 11 of the 12 probe targets showed reproducibility > 95%. The Enterobacter probe (ENT) initially showed a reproducibility of 87.5% due to false negative results. Root cause analysis resulted in a post study imaging processing change; the original data was reevaluated with regression analysis. The resulting reproducibility was 93.2%

{15}------------------------------------------------

(Table 6).

ProbeaNo. Detected/No. ValidNo. Detected/No.ValidNo. Detected/No.ValidTotalIdentifiedPercentIdentifiedWeightedPercentIdentified
Site 34Site 35Site 36
ABA62/6263/6360/64185/18997.9%98.7%
CITb17/1730/3230/3277/8195.1%94.4%
ECO55/5551/5255/55161/16299.4%99.2%
EFM51/5355/5649/54155/16395.1%97.0%
EFSb8/88/88/824/24100.0%100.0%
ENTb19/2421/2320/2460/7184.5%b87.5%c
KLE46/4846/4844/48136/14494.4%96.3%
PAE43/4348/4848/48139/139100.0%100.0%
SAU131/135138/140136/137405/41298.3%98.0%
SMAb8/88/88/824/24100.0%100.0%
STR38/3938/3835/35111/11299.1%99.2%

Table 6. Summary of Reproducibility of Identification Assay Results:

4 See Table 1 for abbreviation definitions

b Isolates of Citrobacter spp., E. faecalis, Enterobacter spp. and S. marcescens were evaluated in order to obtain sufficient results for AST Testing; total number of data points for each species <90.

6 An Image processing change was implemented post study to correct a known issue. Regression analysis of the ENT probe produced a percent identified of 93.0% and a weighted percent identified score of 93.2%.

For AST, reproducibility for all antimicrobials was assessed by evaluating approximately 10 organisms with on-scale MIC values (a value within the assay's reportable range) for each antimicrobial reported by the assay. For Daptomycin fewer than ten isolates were tested due to the lack of isolates with on-scale MICs for these antimicrobials; however, more than four replicates per isolate were tested to achieve 90 data points for those antimicrobials. AST reproducibility was determined from the total number of results that fell within 1 dilution (+/- one doubling dilution) of the mode results divided by the total number of results. On-scale AST performance for each antimicrobial agent was evaluated between sites and within each site. Both Best Case (assumes that off-scale results are within one dilution of the mode) and Worse Case (assumes that off-scale results are more than one dilution of the mode) performance was determined for each drug when the mode across all sites was on-scale for each drug. Worst case performance was not determined for the resistance phenotypes as this result is positive/negative with no off-scale results. Only those samples that met the required number of total results per site were used in the final calculations.

For the AST assay testing the reproducibility of 17 of the 18 antimicrobials and resistance phenotypes were acceptable with best case reproducibility of > 95%. However, erythromycin demonstrated best and worst case reproducibility of 93.6% which was considered acceptable (Table 7).

{16}------------------------------------------------

Indeterminate, false positive and invalid results were obtained during the course of the reproducibility study. See Tables 35 and 36 for summaries of the indeterminate results. See Tables 37 and 38 for a summary of the false positive and invalid results, respectively.

AntimicrobialOrganismsTestedNo.isolatesBest CaseNo. Within 1 ± Dil/Total Tests (%)Worst CaseNo. Within 1 ± Dil/Total Tests (%)
AmikacinA. baumannii6128/128 (100.0)118/128 (92.2)
P. aeruginosa4
S. marcescens1
AmpicillinE. faecium893/93 (100.0)88/93 (94.6)
AztreonamC. freundii3348/350 (99.4)348/350 (99.4)
K. pneumoniae10
E. coli11
P. aeruginosa6
S. marcescens1
CeftazidimeC. freundii3324/331 (100.0)317/331 (95.8)
K. pneumoniae7
E. coli11
P. aeruginosa7
S. marcescens1
CeftarolineS. aureus11151/151 (100.0)146/151 (96.7)
CiprofloxacinC. freundii1136/139 (100.0)133/139 (95.7)
K. pneumoniae2
E. coli2
P. aeruginosa5
S. marcescens1
CeftriaxoneE. aerogenes1151/151 (100.0)140/151 (92.7)
C. freundii1
K. pneumoniae2
E. coli8
S. marcescens2
DaptomycinS. aureus392/92 (100.0)82/92 (89.1)
S. haemolyticus1
E. faecalis1
ErythromycinS. aureus10117/125 (93.6)117/125 (93.6)
ErtapenemC. freundii3356/356 (100.0)354/356 (99.4)
K. pneumoniae11
E. coli12
S. marcescens2
E. aerogenes3
CefepimeC. freundii1158/162 (97.5)152/162 (93.8)
K. pneumoniae1
E. coli8
P. aeruginosa2
S. marcescens1
E. aerogenes1
AntimicrobialOrganismsTestedNo.isolatesBest CaseNo. Within 1 ± Dil/Total Tests (%)Worst CaseNo. Within 1 ± Dil/Total Tests (%)
GentamycinK. pneumoniae6233/234 (99.6)218/234 (93.2)
E. coli5
P. aeruginosa6
S. marcescens2
E. aerogenes1
LinezolidE. faecium10350/355 (98.6)350/355 (98.6)
E. faecalis2
S. aureus17
MeropenemP. aeruginosa7105/106 (99.1)104/106 (98.1)
S. marcescens2
Amp/SulbactamK. pneumoniae6157/157 (100.0)152/157 (96.8)
E. coli8
TobramycinK. pneumoniae2102/104 (98.1)101/104 (97.1)
E. coli2
P. aeruginosa3
S. marcescens2
C. freundii2
Pip/TazoK. pneumoniae4215/222 (96.9)205/222 (92.3)
E. coli4
P. aeruginosa7
S. marcescens1
E. aerogenes1
VancomycinE. faecium1268/282 (95.0)232/282 (82.3)
S. aureus7
Resistance Phenotypes
CefoxitinS. aureus21206/206 (100.0)N/Ab
MLSbS. aureus a21217/222 (97.8)N/Ab

Table 7. Summary of Reproducibility of AST Assay Results for all Antimicrobials and Resistance Phenotypes across All Sites

{17}------------------------------------------------

ª Reproducibility was performed with 21 isolates of S. aureus that were MLSb positive; however in the clinical study, major and minor errors occurred that could not be resolved; there is no claim for testing S. aureus for MLSb and results will be suppressed.

b Resistance phenotype, no worst case evaluation.

  • b. Linearity/assay reportable range: Not Applicable
  • Traceability, Stability, Expected values (controls, calibrators, or methods): C. Three internal assay controls are included in each Accelerate PhenoTest BC kit and require no action by the user:

Universal Bacterial Probe (or Universal Eukaryotic Probe for Yeast). The universal bacterial probe binds to rRNA in all bacterial cells to signal that bacterial cells are present in the flowcells and to differentiate bacteria from debris. The universal eukaryotic probe binds to rRNA in all eukaryotic cells, including yeast, to signal that yeast cells are present in the flowcells and to differentiate yeast from debris. Colocalization of universal and target probe is required for identification of a target organism.

{18}------------------------------------------------

General Nucleic Acid Stain. A cell permeant nucleic acid fluorescent dye that binds to double-stranded DNA (dsDNA) is run in a separate flowcell channel as a DNA staining control. Data from this control is used to quantify the total number of microbial cells in a sample. Based on this result the remainder of the sample is diluted in order to load the desired number of cells per microscopic field of view for susceptibility testing.

Growth Control Channel. The growth control channel is used as a positive control for AST. It measures the number of healthy clones as well as the growth density and growth rate of healthy cells that are not exposed to any antibiotics. The channel consists of the sample and Mueller-Hinton Agar as the growth media for the cells as recommended by CLSI for susceptibility testing.

External Controls. Organisms used for QC testing of the PhenoTest BC kit are included in four panels (Table 8); multiple isolates are tested in a single run by loading the members of a single panel into specified wells in the reagent cartridge. OC organism panels can be rotated to ensure that each ID and AST channel is tested on a regular basis.

Panel NameOrganismsStrain
EPEES Panel (AST)E. coliATCC 25922
E. coliATCC 35218
P. aeruginosaATCC 27853
E. faecalisATCC 29212
S. aureusATCC 29213
SES Panel (AST)S. aureusATCC 43300
S. aureusATCC BAA-977
CASKS Panel (ID)S. lugdunensisATCC 700328
K. pneumoniaeATCC 700603
A. baumanniiATCC 19606
C. freundiiATCC 6879
C. glabrataATCC 2001
S. agalactiaeATCC 12403
SPECS Panel (ID)E. faeciumATCC 19434
S. marcescensATCC 43862
E. aerogenesATCC 13048
P. vulgarisATCC 6380
C. albicansATCC 96268
S. pneumoniaeATCC 49619

Table 8. Quality Control Organism Panels

QC Reportable Range. The dilutions used for evaluation of QC strains are different than the reportable ranges used for patient isolates (Table 9). These ranges were added for the purpose of generating on-scale QC results; these ranges are not the same as those used for reporting patient samples from the PhenoTest BC kit.

{19}------------------------------------------------

QC StrainDrugQC Reportable Range (µg/mL)
E. coliATCC 25922Ceftriaxone0.015 - 0.25
E. coliATCC 25922Ceftazidime0.03 - 1
E. coliATCC 25922Amikacin0.5 - 8
E. coliATCC 25922Aztreonam1 - 16
E. coliATCC 25922Cefepime0.25 - 8
P. aeruginosaATCC 27853Ciprofloxacin0.12 - 2
P. aeruginosaATCC 27853Ertapenem1 - 16
P. aeruginosaATCC 27853Gentamicin0.25 - 4
P. aeruginosaATCC 27853Meropenem0.25 - 2
P. aeruginosaATCC 27853Tobramycin0.25 - 2
E. coliATCC 35218Ampicillin/Sulbactam4 - 64
E. coliATCC 35218Piperacillin/Tazobactam0.25 - 4
E. coliATCC 35218Ampicillin0.25 - 4
E. faecalisATCC 29212Daptomycin0.5 - 8
E. faecalisATCC 29212Linezolid0.5 - 8
E. faecalisATCC 29212Vancomycin0.5 - 8
S. aureusATCC 29213CefoxitinNegative*
S. aureusATCC 29213Ceftaroline0.25 - 1
S. aureusATCC 29213Erythromycin0.25 - 2
S. aureusATCC 29213MLSbNegative*
S. aureusATCC 4330CefoxitinPositive*
S. aureusATCC BAA-977MLSbPositive*

Table 9. Reportable Ranges for QC organisms

*Qualitative Assay, No reportable range

Quality Control Testing. Quality Control was run each day during the clinical and analytical studies. To monitor system performance during the clinical study, QC organism panels were rotated to ensure that each ID and AST channel was tested on a regular basis. The reference laboratory performed QC each day of testing with appropriate organisms for the reference methods. QC results are shown below for the Identification assay (Table 10) and the AST assay (Table 11). Quality control results were acceptable.

Table 10. QC Results Identification Assay, Clinical Study

QC OrganismNo. TestedNo. DetectedNo. NotDetectedPercentDetected
A. baumannii ATCC 19606406398898.0
C. albicans ATCC96268390385598.7
C. glabrata ATCC 20014063881895.6
C. freundii ATCC68794063891795.8
E. aerogenes ATCC 13048390388299.5
E. faecalis ATCC 292125785671198.1
E. faecium ATCC 19434390382898.0
E. coli ATCC 259225785621697.2
K. pneumoniae ATCC 700603406400698.5
P. vulgaris ATCC 6380390382898.0
P. aeruginosa ATCC 278535785671198.1

{20}------------------------------------------------

QC OrganismNo. TestedNo. DetectedNo. NotDetectedPercentDetected
S. marcescens ATCC 43862390388299.5
S. aureus ATCC 43300354352299.4
S. lugdunensis ATCC 70028406400698.5
S. agalactiae ATCC 124034063941297.0
S. pneumoniae ATCC 49619390383798.2

Table 11. QC Results AST Assay, Clinical Study

ExpectedReferencePhenoTest BC
DrugQC OrganismRange(ug/mL)No.TestedNo. inRange (%)No. TestedNo. inRange (%)
AmikacinP. aeruginosaATCC 278531 - 46767 (100.0)284284 (100.0)
AmpicillinE. faecalisATCC 292120.5 - 26464 (100.0)278277 (99.6)
Ampicillin/SulbactamE. coliATCC 352188/4 - 32/166766 (98.5)286285 (99.7)
AztreonamP. aeruginosaATCC 278532 - 86666 (100.0)285285 (100.0)
CefepimeP. aeruginosaATCC 278530.5 - 46767 (100.0)285284 (99.7)
CeftazidimeE. coliATCC 259220.06 - 0.56969 (100.0)286286 (100.0)
CeftarolineE. faecalisATCC 292120.25 - 26464 (100.0)(100)279 (100.0)
CiprofloxacinP. aeruginosaATCC 278530.25 - 16767 (100.0)285285 (100.0)
CeftriaxoneE. coliATCC 259220.03 -0.1256464 (100.0)286286 (100.0)
DaptomycinE. faecalisATCC 292121 - 46464 (100.0)279279 (100.0)
ErythromycinS. aureusATCC 292130.25 - 16161 (100.0)284282 (99.3)
ErtapenemP. aeruginosaATCC 278532 - 86767 (100.0)285285 (100.0)
GentamycinP. aeruginosaATCC 278530.5 - 26767 (100.0)285285 (100.0)
LinezolidE. faecalisATCC 292121 - 46464 (100.0)279279 (100.0)
MeropenemP. aeruginosaATCC 278530.25 - 16666 (100.0)284282 (99.3)
Piperacillin/TazobactamE. coliATCC 352180.5/4 - 2/47069 (98.6)285285 (100.0)
TobramycinP. aeruginosaATCC 278530.25 - 16969 (100.0)285283 (99.3)
VancomycinE. faecalisATCC 292121 - 46464 (100.0)279279 (100.0)
CefoxitinS. aureusATCC 29213Neg6262 (100.0)285282 (98.9)
DrugQC OrganismExpected(µg/mL)ReferencePhenoTest BC
RangeNo.TestedNo. inRange (%)No. TestedNo. inRange (%)
S. aureusATCC 43300Pos6161 (100.0)172172 (100.0)
MLSbS. aureusATCC 29213Neg6262 (100.0)285282 (99.0)
S. aureusATCC BAA-977Pos6464 (100.0)174174 (100.0)

{21}------------------------------------------------

d. Growth and Detection Study

The Growth and Detection study was performed to evaluate two processes: 1) the validation of the inoculum concentration for seeded blood cultures and 2) to assess the equivalence and repeatability of identification and AST results at the time of positivity (t=0) and at eight hours post-positivity (t=8).

Validation of inoculum concentration. This study was performed to demonstrate that blood culture bottles inoculated with various concentrations of a specific organism and incubated until positivity in a continuous monitoring blood culture instrument showed essentially equivalent organism concentrations at the time of positivity, regardless of the starting inoculum concentration.

For each of three representative organisms (S. aureus, E. coli and C. albicans) a standardized suspension was prepared and serially diluted. Each of three dilutions was inoculated into three blood culture bottles containing the appropriate volume of human blood. Bottles were incubated until positivity on the blood culture instrument. The organism concentration for each bottle was determined using standard plate counts and an aliquot of each bottle was tested with the PhenoTest BC Kit.

Results. Inoculum validation study. The intra-organism plate count averages at positivity were generally within one log across positive blood culture samples despite seeding concentrations that spanned four logs; the concentration of organisms was within the expected range of organism concentration in positive patient blood cultures at bottle ring. All positive blood culture bottles tested at t=0 produced the expected positive ID results for each organism (Table 12). Results of this study informed the seeding concentration for analytical studies and seeding studies.

{22}------------------------------------------------

Species/IsolateSpikingConcentrationMeanConcentration atPositivityNo. Identified/NoTested (%)
S. aureusATCC 29213$1.5 X 10^3 - 3.0 X 10^3$$9.4 X 10^6$3/3 (100.0)
S. aureusATCC 29213$1.5 X 10^1 - 3.0 X 10^1$$1.0 X 10^7$3/3 (100.0)
S. aureusATCC 29213$1.5 X 10^{-1} - 3.0 X 10^{-1}$$5.5 X 10^6$5/5 (100.0)
E. coliATCC 25922$1.5 X 10^3 - 3.0 X 10^3$$5.1 X 10^8$3/3 (100.0)
E. coliATCC 25922$1.5 X 10^1 - 3.0 X 10^1$$1.7 X 10^9$3/3 (100.0)
E. coliATCC 25922$1.5 X 10^{-1} - 3.0 X 10^{-1}$$9.6 X 10^7$3/3 (100.0)
C. albicansATCC 96268$7.8 -40 X 10^3$$1.7 X 10^7$3/3 (100.0)
C. albicansATCC 96268$7.8 -40 X 10^1$$4.3 X 10^6$4/4 (100.0)
C. albicansATCC 96268$7.8 -40 X 10^{-1}$$4.7 X 10^6$4/4 (100.0)

Table 12. Spiking Concentration, Concentration at Positivity and Identification Results

Growth and Detection. The performance of the PhenoTest BC system was evaluated using seeded blood cultures prepared with 21 on-panel organisms and one off-panel organism selected to evaluate each of the 16 ID assays and 20 AST and phenotypic resistance determinants included in the PhenoTest BC Kit. Each isolate was spiked into BACTEC blood culture bottles containing the recommended quantity of human blood in concentrations equivalent to those determined in the inoculum validation study. Tests were performed in triplicate. The blood culture bottles were seeded, placed in the BACTEC blood culture instrument and incubated until positivity. The blood cultures were then tested with the PhenoTest BC kit at the time of positivity (t=0), placed back into the blood culture instrument and retested eight hours after positivity (t=8). The performance was evaluated by comparing the results to the expected result based on the known characteristics (ID and AST) of the spiked organisms. The concentration of organism in the tested blood cultures was determined at both the t=0 and t=8 time points using standard plate counts.

Results Growth and Detection Results, ID Assay. Samples tested at t=8 demonstrated organism concentrations approximately 1 log higher than samples tested at t=0 for the majority of organisms. S. lugdunensis, P. mirabilis and C. glabrata had organism concentrations that differed between the two time points at approximately 1.5 logs, 2 log and < 1 log, respectively (Table 13). Correct identifications were obtained for all samples at both t=0 and t=8 time points, validating the sponsor's claim for testing positive blood culture bottles within 8 hours of bottle ring

Organismt=0No. Identified/Total Tested(%)t=8No. Identified/Total Tested(%)ConcentrationDifferencet=0 and t=8
S. aureusS. epidermidisE. faecalisE. faeciumS. gallolyticusS. mitis21/21 (100.0)21/21 (100.0)1 log
S. lugdunensis3/3 (100.0)3/3 (100.0)~1.5 log
C. freundiiC. koseri34/34 (100.0)34/34 (100.0)1 log

Table 13. Growth and Detection Study, ID Assav

{23}------------------------------------------------

Organismt=0No. Identified/Total Tested(%)t=8No. Identified/Total Tested(%)ConcentrationDifferencet=0 and t=8
E. aerogenes
E. cloacae
E. coli
K. oxytoca
K. pneumoniae
P. vulgaris
S. marcescens
P. aeruginosa
A. baumannii
P. mirabilis3/3 (100.0)4/4 (100.0)~2 log
C. albicans4/4 (100.0)3/3 (100.0)< 1 log
C. glabrata

Results Growth and Detection, AST Assay. In the Growth and Detection Study the AST results were evaluated for two parameters: 1) Consistency and equivalence of MIC values reported for all antimicrobials and organisms at both time points t=0 and t=8, and 2) Essential agreement and categorical agreement of results obtained at t=0 and t=8 as compared to the reference method.

For parameter 1. The mode values obtained for samples tested at t=0 and t=8 were within one ± doubling dilution for all isolates tested with all antimicrobials except P. mirabilis with meropenem and C. freundii with piperacillin-tazobactam. On repeat, the mode of the P. mirabilis MIC values with meropenem was acceptable; results of C. freundii with piperacillin-tazobactam were not resolved. However, the overall repeatability of all MIC values was 421/429 (98%) indicating that the PhenoTest BC Kit will provide reproducible AST results with positive blood culture bottles tested between 0 and 8 hours post bottle ring.

For parameter 2, all results with all organisms tested with the following antimicrobials showed essential agreement and categorical agreement ≥ 89.9% as compared to results obtained with the reference method: amikacin, ampicillin-sulbactam, ceftaroline, ciprofloxacin, daptomycin, erythromycin, meropenem and vancomycin. Cefoxitin and MLSb showed category agreement of 100% with the reference method. Comparative results with the reference method obtained with the remaining antimicrobials reflected the lower EA and/or CA obtained for those antimicrobials in the clinical study; these performance issues were subsequently addressed through limitations. Results of AST testing at t=0 and t=8 are shown in Table 14:

{24}------------------------------------------------

AntimicrobialOrganismsEA(%EA)CA(%CA)EA(%EA)CA(%CA)
AmikacinEnterobacteriaceaeA. baumanniiP. aeruginosa36/36(100.0)36/36(100.0)36/36(100.0)36/36(100.0)
AmpicillinEnterococcus spp.6/6(100.0)6/6(100.0)6/6(100.0)6/6(100.0)
Ampicillin-SulbactamEnterobacteriaceae15/15(100.0)14/15(93.3)15/15(100.0)14/15(93.3)
AztreonamEnterobacteriaceaeP. aeruginosa30/33(90.0)25/33(75.7)a28/33(84.4)25/33(75.7)a
CefepimeEnterobacteriaceaeP. aeruginosa29/33(87.9)25/33(75.8)a29/33(87.9)26/33(78.8)a
CeftarolineS. aureus3/3(100.0)3/3(100.0)3/3(100.0)3/3(100.0)
CeftazidimeEnterobacteriaceaeP. aeruginosa27/33(81.8)28/33(84.8)a27/33(81.8)27/33(81.8)a
CeftriaxoneEnterobacteriaceae26/30(86.7)24/30(80.0)a27/30(90.0)24/30(80.0)a
CiprofloxacinEnterobacteriaceaeP. aeruginosa33/33(100.0)33/33(100.0)33/33(100.0)33/33(100.0)
DaptomycinStaphylococcus spp.Enterococcus spp.15/15(100.0)15/15(100.0)15/15(100.0)15/15(100.0)
ErtapenemEnterobacteriaceae24/30(80.0)21/30(70.0)a24/30(80.0)21/30(70.0)a
ErythromycinS. aureus3/3(100.0)3/3(100.0)3/3(100.0)3/3(100.0)
GentamicinEnterobacteriaceaeP. aeruginosa33/33(97.0)26/33(78.8)a33/33(100.0)28/33(84.8)a
LinezolidS. aureus9/9(100.0)6/9(66.7)b9/9(100.0)6/9(66.7)b
MeropenemEnterobacteriaceaeP. aeruginosa32/33(97.0)32/33(97.0)32/33(97.0)32/33(97.0)
Piperacillin-tazobactamEnterobacteriaceaeP. aeruginosa32/36(88.9)30/36(83.3)a27/36(75.9)27/36(75.0)a
TobramycinEnterobacteriaceaeP. aeruginosa25/33(75.8)33/33(100.0)26/33(78.8)29/33(87.9)
VancomycinStaphylococcus spp.Enterococcus spp.15/15(100.0)15/15(100.0)15/15(100.0)15/15(100.0)
CefoxitinStaphylococcus spp.9/9(100.0)9/9(100.0)9/9(100.0)9/9(100.0)
MLSbStaphylococcus spp.9/9(100.0)9/9(100.0)9/9(100.0)9/9(100.0)

Table 14. Growth and Detection Study, AST Assay, EA and CA as Compared to the Reference Method at t=0 and t=8

4 Low category agreement due to a majority of minor errors

b Low number of isolates tested

{25}------------------------------------------------

Indeterminate, false positive and invalid results were obtained during the course of the growth and detection study. See Tables 35 and 36 for summaries of the indeterminate results. See Tables 37 and 38 for a summary of the false positive and invalid results, respectively.

e. Cross Contamination/Carry Over

Evaluation of cross contamination and carryover was performed as a component of the Growth and Detection studies. In general t=0 and t=8 samples containing different organisms were run in alternating fashion to evaluate any potential carryover from samples with higher organism loads to samples with lower organism concentrations.

Results Cross Contamination/Carry-over Study. None of the alternating organism species and concentration combinations tested on the same instrument demonstrated any evidence of cross contamination.

f Analytical Inclusivity

The intent of the inclusivity study was to demonstrate the ability of the PhenoTest BC Kit to detect and identify on-panel species in positive blood culture samples. The Inclusivity (Analytical Specificity) of the PhenoTest BC Kit was evaluated by testing three strains of each of the on-panel target species. (Additional well-characterized strains were evaluated as challenge organisms for the AST assay; results obtained with the PhenoTest BC Kit with these organisms are included in the analysis of the Identification assay performance.) A standardized suspension of each organism was inoculated into blood culture bottles containing the appropriate volume of human blood and incubated until determined positive by the blood culture instrument. Testing with the PhenoTest BC Kit was then performed within approximately 1 hour. All isolates included in the Inclusivity Study are listed in Tables 15 -17 were detected and correctly identified by the PhenoTest BC Kit.

ProbeSpeciesStrain
SAUStaphylococcus aureusIHMA 166917, ATCC 10832, ATCC 29213
Staphylococcus epidermidisJMI 365, JMI 154
Staphylococcus haemolyticusATCC 29970, JMI 4510, JMI 18615*
CNSStaphylococcus lugdunensisATCC 700328, JMI 364, JMI 367
Staphylococcus capitisATCC 27840, ATCC 35661, JMI 794
Staphylococcus hominisATCC 27844*, JMI 16176, JMI 16434
SLUStaphylococcus lugdunensisATCC 700328, JMI 364, JMI 367
EFSEnterococcus faecalisIHMA 850419, ATCC 19433, ATCC 29212
EFMEnterococcus faeciumIHMA 743951, ATCC 19434, BEI N28976
STRStreptococcus agalactiae (serotype Ia)ATCC BAA 1177
Streptococcus agalactiae (serotype III)ATCC 12403
Streptococcus agalactiae (serotype V)ATCC BAA 611
Streptococcus gallolyticusATCC 43143, ATCC 51879, ATCC 700065
Streptococcus mitisIHMA 233226, HM 262, NR 34818
Streptococcus pneumoniae (serotype 19A)BEI H145
Streptococcus pneumoniae (serotype 11A)BEI N19172
Streptococcus pneumoniae (serotype 6B)BEI N19213*

Table 15. Inclusivity Study: Gram-Positive Isolates Tested for Each Probe Target

{26}------------------------------------------------

  • Indeterminate results obtained for additional probes
ProbeSpeciesStrain
ABAAcinetobacter baumanniiIHMA 770910, ATCC 19606, BEI N17785
CITCitrobacter freundiiIHMA 617760, JMI 1333*, ATCC 6879*
Citrobacter koseriIHMA 802066*, JMI 192*, JMI 631*
ENTEnterobacter aerogenesIHMA 832805, ATCC 13048, JMI 897
Enterobacter cloacaeIHMA 814234, ATCC 13047, ATCC 35030
ECOEscherichia coliIHMA 511379, ATCC 11775, ATCC 25922
KLEKlebsiella oxytocaIHMA 501913, ATCC 700324, BEI H622
Klebsiella pneumoniaeIHMA 875649, ATCC 4972, ATCC 700603
PROProteus mirabilisIHMA 1032951, BEI H752, JMI 148
Proteus vulgarisIHMA 451082, ATCC 6380, JMI 5687
PAEPseudomonas aeruginosaIHMA 761927, ATCC 10145, ATCC 27853
SAESerratia marcescensIHMA 946153, ATCC 43862, JMI 39780

Table 16. Inclusivity Study: Gram-Negative Isolates Tested for Each Probe Target

  • Indeterminate results obtained for additional probes

Table 17. Inclusivity Study: Yeast Isolates Tested for Each Probe Target

ProbeSpeciesStrain
CALCandida albicansATCC 96268, ATCC 11651, JMI 10981
CGLCandida glabrataJMI 52276, JMI 28365, JMI 21400

Analytical Specificity (Exclusivity) g.

The Exclusivity (Analytical Cross-Reactivity) of the PhenoTest BC Kit was evaluated by testing individual representative strains of on- and off-panel species. On-panel isolates were evaluated to determine if they would cross react with probes other than the expected probe. A standardized suspension of each organism was inoculated into blood culture bottles containing the appropriate volume of human blood. Testing with the PhenoTest BC Kit was performed 7.5 to 9 hours post positivity.

The exclusivity test panel for each probe cocktail was designed to challenge the specificity of the probe cocktail with species predicted to have close sequence homology to the target.

Tables 18 and 19 list the species that were determined to be non-cross reactive by analytical testing. Some organisms gave indeterminate results for certain target probes; indeterminate results are not conclusively positive or negative.

Gram-Positive
Bacillus cereusLactococcus lactisStaphylococcus felisa
Bacillus subtilisMacrococcus caseolyticusaStreptococcus constellatus
Clostridium perfringensMacrococcus equipercicusaStreptococcus gordonii
Corynebacterium amycolatumcMicrococcus luteusStreptococcus intermediusb
Corynebacterium jeikeiumaMycobacterium aviumStreptococcus thoraltensis
Corynebacterium striatumMycobacterium intracellulareStreptococcus vestibularis
Enterococcus saccharolyticusMycobacterium smegmatisVagococcus fluvialis

Table 18. Organisms Determined to be Non-Cross Reactive by Analytical Testing

{27}------------------------------------------------

Granulicatella adiacensNocardia novad
Gram-Negative
Achromobacter xylosoxidansCitrobacter sedlakiiProvidencia rettgeri
Acinetobacter calcoaceticusDickeya chrysanthemiPseudomonas luteola
Acinetobacter johnsoniiDickeya paradisiacaPseudomonas putida
Acinetobacter lwoffiiEnterobacter ludwigiiPseudomonas stutzeri
Acinetobacter radioresistensEscherichia albertiiRalstonia pickettii
Acinetobacter schindleriEscherichia blattaeRaoultella planticola
Aeromonas hydrophilaMoraxella catarrhalisShigella flexneri
Bacteroides fragilisMorganella morganiiStenotrophomonas maltophilia
Bordetella bronchisepticaNeisseria meningitidisYersinia enterocolitica
Bordetella parapertussisePlesiomonas shigelloidesYersinia pestise
Brenneria quercinaProteus hauseriYersinia pseudotuberculosis
Citrobacter braakii

ª Gave indeterminate results with CNS probe

b Different replicates gave indeterminate results with CNS and SAU probe

° Different replicates gave indeterminate results with CNS, SLU and SAU probes

d Gave indeterminate results with CAL probe

€ Gave indeterminate result with SLU probe

Table 19. Cross Reacting Organisms as Determined by Analytical Testing

ProbeCross Reactive Organisms
Gram-Positive
S. aureus(SAU)Candida parapsilosisPseudomonas nitroreducensCandida parapsilosisStreptococcus pneumoniaeStaphylococcus hominisPantoea agglomerans
Acinetobacter haemolyticusStaphylococcus equorum
Acinetobacter ursingiiCandida kruseiCandida parapsilosisCandida tropicalisStaphylococcus kloosiiStaphylococcus lentusStaphylococcus pasteuriStaphylococcus saprophyticus
Coagulase negativeStaphylococcus (CNS)Enterobacter hormaecheiEnterococcus canisEnterococcus dispar *Mycobacterium scrofulaceumStaphylococcus schleiferiStaphylococcus succinusStaphylococcus xylosusStreptococcus anginosus
Nocardia farcinicaStaphylococcus auricularisStaphylococcus capraeStaphylococcus cohnii *Streptococcus mutansStreptococcus pneumoniae *Streptococcus sanguinis
E. faecium (EFM)Enterobacter aerogenesEnterococcus gallinarumEnterococus duransStaphylococcus succinusStreptococcus pneumoniae *
S. lugdunensis (SLU)Nocardia farcinicaCandida krusei *Streptococcus pneumoniaeStreptococcus parasanguinis
Streptococcus spp.(STR)Staphylococcus sciuriStreptococcus constellatus *Streptococcus mutansStreptococcus pseudopneumoniaeStreptococcus pyogenes

{28}------------------------------------------------

Gram-Negative
A. baumannii (ABA)Acinetobacter pittii (A. genomospecies 3)Shigella dysenteriae
Burkholderia cepaciaStaphylococcus sciuri
P. aeruginosa (PAE)Enterobacter cloacae
Enterococcus casseliflavus
Citrobacter gilleniiEscherichia hermannii
Citrobacter spp. (CIT)Citrobacter murliniae*Pseudomonas fulva*
Citrobacter youngae*
Enterobacter asburiaeEnterobacter mori
Enterobacter spp. (ENT)Enterobacter hormaecheiEnterobacter nimipressuralis
Escherichia fergusoniiShigella dysenteriae
Escherichia coli (ECO)Pseudomonas fulva*Shigella sonnei
Shigella boydii
Brenneria goodwiniiEscherichia hermannii
Brenneria nigrifluensEscherichia vulneris
Klebsiella spp. (KLE)Brenneria salicisKlebsiella ozaenae
Cronobacter sakazakiiKlebsiella variicola*
Enterobacter kobeiPantoea agglomerans
Proteus spp. (PRO)Proteus penneri
Brenneria nigrifluensStreptococcus parasanguinis
Serratia marcescens (SMA)Pseudomonas oryzihabitans
Candida spp.
Candida albicans (CAL)Streptococcus pyogenesS. capitus
P. aeruginosa
Candida glabrata (PAE)Acinetobacter haemolyticusSalmonella enterica
Brenneria goodwiniiStreptococcus pyogenes
Enterococcus casseliflavus
  • Indeterminate results obtained for additional probes

In silico analysis was performed to assess potential cross reactivity with target probes for organisms that were not available or did not grow in blood culture media or for which valid PhenoTest BC Kit results were not available due to an insufficient cell concentration. To assess any potential cross reactivity the 16S rRNA gene sequences of the organisms were compared to the sequences of the probe targets in a sequence database. Table 20 lists additional organisms that were predicted to cross react with the target probe sequences based on in silico analysis. The following organisms were not tested for cross reactivity by either analytical or in silico analysis: Bordetella pertussis, Brenneria nigrifluens, Brenneria salicis (Erwinia salicis), Nocardia abscessus and Vibrio parahemolyticus.

{29}------------------------------------------------

OrganismProbe Which Maybe Positive Based onin silico Prediction
Acinetobacter nosocomialis (aka A. genomospecies 13)ABA
Citrobacter braakiiCIT
Citrobacter koseriCIT
Cronobacter sakazakiiKLE
Enterococcus canisEFM
Enterococcus saccharolyticusEFM
Morganella morganiiSMA
Streptococcus gordoniiSTR
Streptococcus pseudopneumoniaeSTR
Streptococcus sanguiinisSTR
Streptococcus vestibularisSTR

Table 20. Organisms Predicted to Cross React with Target Probes by In Silico Analysis® 8

ª In silico analysis did not consistently predict results obtained in analytical testing

b Inclusivity could not be predicted via in silico analysis for Klebsiella oxytoca (PPA of clinical study 90.0%) and Proteus vulgaris (PPA of clinical study 100%)

Indeterminate, false positive and invalid results were obtained during the course of the inclusivity and exclusivity studies. See Tables 35 and 36 for summaries of the indeterminate results. See Tables 37 and 38 for a summary of the false positive and invalid results, respectively.

  • h. LoD (Limit of Detection)
    The objective of the LoD study was to determine the lowest concentration of organism in blood culture bottles that could be reliably detected and identified (>95% of the time). The level of detection of the Accelerate PhenoTest BC Kit was determined for all 16 FISH probes by inoculating blood culture bottles containing 10mL human donor blood with representative strains of each target organism or group of organisms at levels below the theoretical limit of detection (LoD) of the assay and below the expected concentration at positivity. The bottles were incubated in an automated blood culture instrument for a predetermined period and removed for analysis with the Accelerate PhenoTest BC assay prior to positivity detection. Tests were repeated and incubated for a longer period of time if 95% positivity was not achieved. Colony counts were performed to determine the organism concentration at the time of testing. Additional blood culture bottles that were inoculated at the same time as those for testing with the PhenoTest BC Kit were used to determine the duration of incubation in the blood culture instrument that would result in a positive detection. The results of the study indicate that in most cases the LoD is at or below the concentration of organisms present when blood cultures are determined to be positive by the continuous monitoring blood culture system (as determined in the Growth and Detection study summarized above). The average LoD for gram-negative target bacteria, gram-positive bacteria and Candida sp. was 4 X 10°, 5 X 108 and 2 X 106, respectively. The results of the LoD studies are summarized in Tables 21 – 23 below.

{30}------------------------------------------------

ProbeSpeciesStrainTime (hours)Concentration atTesting(CFU/mL)No. Positive/No.Tested (%)
Time ToPositiveaIncubationPeriodb
ABAA. baumanniiATCC196066.24.81.51 x 10822/22 (100.0)
IHMA7582656.24.85.27 x 10822/22 (100.0)
CITC. freundiiIHMA10298754.94.52.20 x 1085/22 (22.7)
C. koseriIHMA10034844.44.33.20 x 10821/22 (95.5)
ECOE. coliATCC259224.24.01.15 x 10818/18 (100.0)
4.74.09.00 x 10722/22 (100.0)
IHMA7529614.24.07.40 x 10722/22 (100.0)
ENTE. aerogenesIHMA10064906.15.31.67 x 1071/22 (4.5)
4.65.81.59 x 10921/22 (95.5)
IHMA8328054.25.82.01 x 10922/22 (100.0)
KLEK. oxytocaIHMA10003935.44.81.14 x 10811/22 (50.0)
4.66.02.67 x 10822/22 (100.0)
K. pneumoniaeATCC7006034.54.85.33 x 10813/22 (59.1)
4.45.58.13 x 10822/22 (100.0)
4.66.03.09 x 10920/22 (90.9)
PAEP. aeruginosaATCC278535.24.37.73 x 10721/22 (95.5)
IHMA10043205.24.59.33 x 10820/22 (90.9)
6.44.37.60 x 10722/22 (100.0)
PROP. mirabilisIHMA10034715.55.32.26 x 10722/22 (100.0)
P. vulgarisIHMA10084186.24.58.93 x 10721/21 (100.0)
SMAS. marcescensIHMA10059575.34.01.30 x 1072/23 (8.7)
4.84.82.77 x 10822/22 (100.0)
IHMA9461534.64.02.65 x 10712/23 (52.2)

Table 21. Analytical Sensitivity (LoD) of Gram-Negative Species

4 Duration of incubation at the time of positivity as detected by the continuously monitored blood culture system b Duration of incubation in the continuous monitoring blood culture system prior to testing with the PhenoTest BC Kit.

Table 22. Analytical Sensitivity (LoD) for Gram-Positive Species

ProbeSpeciesStrainTime (hours)Concentration(CFU/mL)No. Positive/No.Tested (%)
Time ToPositiveaIncubationPeriodb
CNSS. epidermidisATCC7.14.83.75 x 1050/23 (0.0)
S. epidermidis149906.96.02.84 x 10621/22 (95.5)
CNSS. haemolyticusIHMA7.65.52.80 x 1064/22 (18.2)
S. haemolyticus10568776.56.43.27 x 10721/21 (100.0)
EFME. faeciumIHMA5.44.83.37 x 10720/20 (100.0)

{31}------------------------------------------------

ProbeSpeciesStrainTime (hours)Concentration(CFU/mL)No. Positive/No.Tested (%)
Time ToPositiveaIncubationPeriodb
148266
IHMA8244865.65.3$1.80 x 10^8$22/22 (100.0)
EFSE. faecalisIHMA8504194.86.0$4.00 x 10^8$21/22 (95.5)
IHMA10019734.26.0$8.33 x 10^8$22/22 (100.0)
SAUS. aureusATCC292135.14.8$3.73 x 10^6$22/22 (100.0)
IHMA6123115.04.8$1.51 x 10^7$22/22 (100.0)
SLU/CNSS. lugdunensisIHMA8455455.15.3$3.66 x 10^7$22/22 (100.0)
IHMA8673617.34.8$3.49 x 10^7$21/21 (100.0)
STRS. agalactiaeIHMA7532203.96.0$1.45 x 10^9$2/22 (9.1)c
4.35.0$1.19 x 10^9$18/22 (81.8)c
S. gallolyticusIHMA8783294.34.3$2.29 x 10^8$22/22 (100.0)
S. pneumoniaeATCC496195.05.0$4.87 x 10^9$17/22 (77.3)
5.86.5$3.81 x 10^9$21/21 (100.0)

4 Duration of incubation at the time of positivity as detected by the continuous monitoring blood culture system b Duration of incubation in the continuous monitoring blood culture system prior to testing with the PhenoTest BC Kit.

& The LoD for S. agalactiae was not established.

ProbeSpeciesStrainTime (hours)Concentration(CFU/mL)No. Positive/No.Tested (%)
Time ToPositiveaIncubationPeriodb
CALC. albicansATCC1165111.311.31.93 x 10622/22 (100.0)
CALC. albicansIHMA6932367.37.32.97 x 10620/21 (95.2)
CGLC. glabrataATCC20017.07.04.07 x 10622/22 (100.0)
CGLC. glabrataJMI 295467.64.46.13 x 10521/22 (95.5)

Table 23. Analytical Sensitivity (LoD) for Candida spp.

4 Duration of incubation at the time of positivity as detected by the continuous monitoring blood culture system b Duration of incubation in the continuous monitoring blood culture system prior to testing with the PhenoTest BC

Kit.

Indeterminate, false positive and invalid results were obtained during the course of the LoD study. See Tables 35 and 36 for summaries of the indeterminate results. See Tables 37 and 38 for a summary of the false positive and invalid results, respectively.

{32}------------------------------------------------

i. Interference Studies

The Accelerate PhenoTest BC kit ID and AST performance was compared using seeded positive blood culture samples with and without interfering substances. Organisms evaluated included one species from each of the major groups of organisms identified by the PhenoTest BC kit: E. faecalis, S. aureus, S. agalactiae, S. pneumoniae, A. baumannii, E. coli, P. aeruginosa, S. marcescens and C. albicans. Both endogenous substances and heparin were evaluated as possible interferents. Endogenous substances are listed in Table 24 below. The interference of RBCs was tested using 2 bottle types (BACTEC Plus Aerobic/F Medium and BACTEC Lytic/10 Anaerobic/F Medium). The remaining substances were tested using standard BACTEC Plus Aerobic/F Medium bottles. All interfering substances were co-spiked with the isolate and incubated until positivity in the continuous monitoring blood culture instrument. Each set included a control (contrived positive blood culture with no added endogenous substance or antibiotic) and replicates of the same contrived sample spiked with one of the interfering substances. This study was designed to allow for additional (stage 2) testing for any organism/antimicrobial/interferent combination for which discrepancies were observed.

The effect of antimicrobial agents as interfering substances was not evaluated. The following limitation is included in the device labeling:

Potential interference by antimicrobial agents that may be present in a patient blood specimen has not been established with the Accelerate PhenoTest BC Kit.

Potential InterferentConcentration TestedReference Range
RBCs(Hematocrit/Hemoglobin)20g/dL1-2 g/L
WBCs (buffy coat)12,000 WBC/µL4,500 -10,000/µL
>12,000/µL or<4000/µL
Platelets400,000/ µL150,000 -400,000/ µL
<150,000
80.000/µL
Triglycerides37 mmol/L0.34 - 3.7 mmol/L
Bilirubin (conjugated)34.2 µmol/L0 – 3.4 µmol/L
Gamma Globulin50 mg/mL6 – 13 g/L
Heparin3000350 – 1000 U/L

Table 24. Potential Interferents, Concentration Tested and Reference Range for Normal

Identification Results. Results obtained for the identification assay showed 99.5% agreement with expected results and all samples returned an identification result.

AST Results. Endogenous substances/organism/antimicrobial combinations demonstrated greater than 89.9% EA for 98% (352 out of 360) of combinations and greater than 89.9% CA for 94% (340 out of 360) of combinations as compared to controls not containing potentially interfering substances.

{33}------------------------------------------------

The results provided for the identification assay and AST assay with endogenous substances and heparin are acceptable.

Indeterminate, false positive and invalid results were obtained during the course of the interfering substances study. See Tables 35 and 36 for summaries of the indeterminate results. See Tables 37 and 38 for a summary of the false positive and invalid results, respectively.

  • j. Blood Bottle Type

Identification Assay

Thirteen different blood culture bottle types from three different blood culture systems (BacT/Alert , (bioMérieux, Inc.) BACTEC (Becton Dickinson) and VersaTREK (ThermoFisher) were evaluated analytically with the Accelerate Pheno Assay (See Table 25 below). Blood culture bottles/media were tested with the recommended ratio of blood to media. Replicates of the following organisms were inoculated into blood culture bottles: A. baumannii, E. coli, P. aeruginosa, S. aureus, S. pneumoniae, E. faecalis and C. albicans. In addition, an off-panel organism, Micrococcus luteus, was tested with all bottle types. Organisms were inoculated into the bottles at concentrations informed by results obtained in the Growth and Detection study and incubated on the appropriate automated blood culture instrument until positive detection. Positive bottles were removed from the blood culture instrument and samples were immediately run on the Accelerate PhenoTest BC Kit.

All samples were evaluated with a minimum of 10 replicates (Stage 1 testing); additional replicates were tested if the expected organism was not detected or if additional testing was required to provide a sufficient number of valid results for each antimicrobial agent (Stage 2 testing).

For the identification assay, results were considered acceptable if the organism inoculated into the bottle was detected and correctly identified. Tests that gave invalid results (too few cells in the probe channels) were not included in the evaluation. For 12 of 13 bottle types the identification assay provided acceptable results (>95% of organisms detected and correctly identified). One bottle type, BACTEC PLUS Anaerobic F detected and identified 94.3% of aerobic isolates due to non-detection of C. albicans in 3/6 replicates. The off-panel organism, M. luteus gave invalid results for the majority of tests; detection of off-panel organisms is not included in the claim for the PhenoTest BC Kit.

Results below represent cultures that were determined to be positive by the blood culture system and provided a valid result.

Blood Bottle TypeTrue PositivesNo. Detected/No.Valid Results (%)(All Organisms)Species Not DetectedNo. NotDetected/No. Tested
BacT/ALERT
SA Standard Aerobic121/121 (100.0)-
SN Standard Anaerobic108/108 (100.0)-
FA Plus Aerobic110/111 (99.1)P. aeruginosa (1/14)
FN Plus Anaerobic112/116 (96.6)C. albicans (3/13)P. aeruginosa (1/13)
PF Plus125/128 (97.7)C. albicans (3/30)

Table 25. Blood Bottle Study Results - Identification Assay

{34}------------------------------------------------

Blood Bottle TypeTrue PositivesNo. Detected/No.Valid Results (%)(All Organisms)Species Not DetectedNo. NotDetected/No. Tested
BACTEC
Standard/10 Aerobic F115/118 (97.5)P. aeruginosa (3/22)
Standard Anaerobic /F58/58 (100.0)-
Plus Aerobic F165/172 (95.9)C. albicans (1/26)E. faecalis (1/22)P. aeruginosa (5/45)
Plus Anaerobic/F66/70 (94.3)C. albicans (3/6)S. aureus (1/21)
PEDS PLUS/F137/141 (97.2)C. albicans (2/26)E. faecalis (2/21)
Lytic/10 Anaerobic/F69/72 (95.8)E. faecalis (3/23)
VersaTREK
REDOX 1 (aerobic)129/132 (97.7)E. faecalis (1/21)P. aeruginosa (2/37)
REDOX 2 (anaerobic)63/64 (98.4)A. baumannii (1/2)

AST

For the AST assay results of each bottle type with each antimicrobial/organism combination were compared to the modal MIC value obtained with the broth microdilution reference method. In addition, results for each antimicrobial/organism combination were compared across all bottles types and compared to the modal MIC obtained with the Pheno System.

For results compared to the modal MIC of the BMD reference method, all bottle types provided acceptable EA and CA values for most organisms tested.

For detection of MLSb resistance a sufficient number of tests were performed only with the following bottle types: BacT/ALERT SA Standard Aerobic, BacT/Alert FA Plus Aerobic, BacT/ALERT FN Plus Anaerobic, BacT/ALERT PF Plus, BACTEC Plus Aerobic F and BACTEC Lytic/10 Anaerobic/F. An insufficient number of tests were performed with the remaining blood culture bottle types. The sponsor included the following limitation in their device labeling:

Due to an insufficient number of test isolates, the ability of the Accelerate PhenoTest BC kit to detect inducible MLSb resistance in coagulase-negative staphylococci is unknown when used with the following blood culture bottle types: BacT/Alert SN Standard Anaerobic, BACTEC Peds Plus/F, BACTEC Plus Anaerobic/F, BACTEC Standard Anaerobic, BACTEC Standard/10 Aerobic, VersaTrek Redox 1 Aerobic, VersaTrek Redox 2 Anaerobic. Use an alternative method for detection of inducible MLSb resistance when using these blood culture bottle types if critical to patient care.

Indeterminate, false positive and invalid results were obtained during the course of the blood bottle study. See Tables 35 and 36 for summaries of the indeterminate results. See Tables 37 and 38 for a summary of the false positive and invalid results, respectively.

{35}------------------------------------------------

k. Polymicrobial LoD

The purpose of the Polymicrobial Limit of Detection Study was to characterize the levels of different microorganisms that must be present in blood cultures containing two different microbial species in order for the PhenoTest BC Kit to perform antimicrobial susceptibility testing on either or both species. Testing was performed using 13 pairs of different microbial species. Individual strains were inoculated into blood culture bottles containing whole human blood and incubated until called positive by an automated blood culture instrument. The positive blood cultures were then mixed in different ratios to obtain simulated polymicrobial cultures at different target levels. The concentration of each organism in each mixture was determined by performing viable counts on the parental monomicrobial cultures. Each mixture was tested in duplicate using the PhenoTest assay.

Results of the identification assay showed acceptable detection and identification of all isolates in the various concentrations. Only S. aureus (present in concentrations equivalent to that at positivity) with high concentrations of C. albicans (three to four log higher concentration than S. aureus) was not detected.

AST testing of two organisms was not supported by the test results. The PhenoTest BC Kit will only perform AST testing on a single isolate in a polymicrobic sample. See above for details on AST testing in polymicrobic samples.

Indeterminate, false positive and invalid results were obtained during the course of the polymicrobial LoD study. See Tables 35 and 36 for summaries of the indeterminate results. See Tables 37 and 38 for a summary of the false positive and invalid results, respectively.

l. Biological interference

The potential for biological interference of polymicrobial positive blood culture samples was evaluated by testing 17 combinations of various resistant isolates with previously established modal MICs or known resistance phenotypes in the presence of other isolates. Testing was performed at the time of positive detection by a continuous monitoring blood culture system. Testing was limited to polymicrobial samples containing 2 target species and included onpanel and off-panel species (Propionibacterium acnes, Bacteroides fragilis and Stenotrophomonas maltophilia).

For the identification assay 15 of 17 combinations resulted in 100% detection of all organisms. The results of the study are shown in Table 26 below. Results of the identification assay showed acceptable detection and identification of all isolates in the various concentrations. Only S. aureus and K. pneumoniae (present in concentrations equivalent to that at positivity) with high concentrations of C. albicans (three to four log higher concentration than S. aureus) was not detected.

AST testing demonstrated biological interference; AST results will only be reported for one organism in polymicrobial samples. See Section I (Algorithm for Performance of AST) above for details on AST testing in polymicrobic samples.

{36}------------------------------------------------

CombinationSpecies 1NumberIdentified/Total(% Identified)Species 2NumberIdentified/Total (%)
Combination 1S. aureus8/8 (100.0%)K. pneumoniae8/8 (100.0%)
Combination 2S. aureus2/2 (100.0%)K. pneumoniae2/2 (100.0%)
Combination 3S. aureus4/4 (100.0%)E. cloacae4/4 (100.0%)
Combination 4S. aureus3/3 (100.0%)K. pneumoniae3/3 (100.0%)
Combination 5S. aureus2/2 (100.0%)P. acnesNot Detected
Combination 6S. aureus1/2a (50.0%)C. albicans2/2 (100.0%)
Combination 7S. epidermidis2/2 (100.0%)K. pneumoniae2/2 (100.0%)
Combination 8S. pneumoniae3/3 (100.0%)E. coli3/3 (100.0%)
Combination 9E. faecalis2/2 (100.0%)E. coli2/2 (100.0%)
Combination 10E. faecalis2/2 (100.0%)A. baumannii2/2 (100.0%)
Combination 11E. faecium2/2 (100.0%)E. aerogenes2/2 (100.0%)
Combination 12E. faecium6/6 (100.0%)A. baumannii6/6 (100.0%)
Combination 13E. faecium2/2 (100.0%)P. acnesOff-PanelNot Detected
Combination 14E. coli2/2 (100.0%)B. fragilisOff-PanelNot Detected
Combination 15A. baumannii2/2 (100.0%)S. maltophiliaOff-panelNot Detected
Combination 16K. pneumoniae1/3a (33.0%)C. albicans3/3 (100.0%)

Table 26. Organism Combinations Tested to Determine Biological Interference

a Candida spp. positive blood cultures contain low levels of organism relative to other on panel identification organisms. Significant dilution of species paired with Candida spp. is required to achieve 1:1 ratio of organisms.

Indeterminate, false positive and invalid results were obtained during the course of the biological interference study. See Tables 35 and 36 for summaries of the indeterminate results. See Tables 37 and 38 for a summary of the false positive and invalid results, respectively.

    1. Comparison studies:
    • Method comparison with predicate device a. Not applicable
    • b. Matrix comparison: Not Applicable
    1. Clinical studies:

The purpose of the clinical study was to demonstrate the clinical performance of the Accelerate PhenoTest BC system for positive blood culture identification and AST compared to reference results in a clinical setting. The clinical performance was established in a multi-center study conducted at 13 geographically distinct U.S. sites over approximately six months. Samples included prospectively collected fresh samples (aliquots of left-over positive blood cultures from patients suspected of bacteremia or fungemia and tested within eight hours of positivity), fresh seeded samples (blood cultures seeded with human blood and previously characterized fresh

{37}------------------------------------------------

clinical isolates isolated within seven days of seeding and incubated until positivity) and seeded samples (blood cultures seeded with human blood and previously characterized challenge and stock isolates and incubated until positivity). Challenge isolates were selected to represent organism/antimicrobial combinations that were less common and underrepresented in the fresh and fresh seeded samples and to provide resistant or on-scale MIC results for at least one antimicrobial agent. Challenge isolates were obtained from culture collections; stock isolates were obtained from clinical specimens at the clinical sites and represented isolates that had been stored for longer than seven days.

Samples were tested in the following bottle types based on the type of continuously monitored blood culture system in use at the testing site: BacT/ALERT SA Standard Aerobic, BacT/ALERT SN Standard Anaerobic, BACTEC Lytic/10 Anaerobic/F Medium, BACTEC PEDS PLUS/F Medium, BACTEC PLUS Aerobic/F Medium, BACTEC PLUS Anaerobic Medium, BACTEC Standard Anaerobic/F Medium, BACTEC Standard Aerobic/F Medium, VersaTREK REDOX 1 Aerobic Media (80 mL), VersaTREK REDOX 2 Anaerobic Media (80 mL).

Testing was initiated within eight hours of bottle ring; bottles flagged as positive were held at room temperature until testing.

Results of the identification assay were compared to results obtained with VITEK2 (bioMérieux) using isolates subcultured from the blood culture bottles tested with the Accelerate PhenoTest BC Kit. For isolates for which an invalid result was obtained by VITEK2. the reference identification was determined by 16S rRNA gene sequencing.

Results of the AST assay were compared to the CLSI broth microdilution reference method. Isolates were tested in triplicate by the broth microdilution method; the reference result was the mode MIC value determined through replicate testing. If no mode value could be determined or if a QC failure occurred, an additional three replicates were tested to determine the mode MIC. Performance of the evaluation of resistance markers (cefoxitin susceptibility and MLSb) was compared to disk diffusion performed singly; if the test failed or if the zone diameter was within one millimeter of the breakpoint, testing was repeated in triplicate with the modal category used as the reference result. All reference testing was performed at a reference laboratory.

A total of 2500 positive blood cultures were enrolled in the study. Samples were excluded from the ID assay study for the following reasons: protocol deviations, samples that halted workflow, samples enrolled but not tested, samples run outside of the eight hour post-positivity window, samples with no organism identified by Gram stain or subculture or samples for which isolates were considered unacceptable upon receipt at the reference laboratory, samples for which the Accelerate Pheno system did not complete the assay or for which results were invalid. Samples were excluded from the AST assay for the following reasons: AST testing performed after completion of collection of adequate data for AST, samples with no broth microdilution reference data collected, samples for which PhenoTest BC kit did not collect AST data, samples with discordant identifications between PhenoTest BC kit and the reference method, samples with invalid results, samples for which two isolates with the same identification were obtained.

Results were evaluated for a total of 1850 positive blood culture samples including 793 fresh samples, 65 fresh seeded samples, 477 samples seeded with challenge isolates and 515 samples seeded with stock isolates.

{38}------------------------------------------------

The mean time from assay start to ID result was 1.38 hours; the mean time from assay start to AST result was 6.69 hours.

Results obtained from the PhenoTest BC kit should be interpreted in conjunction with Gram stain results obtained from the positive blood culture sample to mitigate the occurrence of false positive results. To address the need for concurrence of PhenoTest BC kit results with Gram Stain results the following limitations were included in the device labeling:

Due to the possibility of cross reactivity, all Accelerate PhenoTest BC kit results should be interpreted in coniunction with Gram stain

Accelerate PhenoTest BC kit identification results that are discordant with the result of the blood culture Gram stain (for example, no organism detection when the Gram stain is positive or detection of a Gram-positive cocci when Gram-positive cocci were not observed in the Gram stain) should be confirmed by culture prior to reporting the test result. For some polymicrobic calls, false positive results may not be mitigated by Gram stain analysis (for example, detection of two Enterobacteriaceae species with Gramnegative rods observed in the Gram stain). Results of such polymicrobic calls should be verified by subculture and/or an alternative identification method.

The use of the PhenoTest BC kit does not eliminate the need for subculture of the positive blood culture. The following limitation was included in the device labeling to address the need for subculture for certain PhenoTest BC kit results:

Subculture of positive blood culture is required in the following situations:

  • For the identification and susceptibility testing of off-panel organisms not O identified by the Accelerate PhenoTest BC kit,
  • For samples that give a polymicrobial result o
  • For organisms for which species identification is critical for patient care, (e.g. o speciation of streptococci)
  • For testing antimicrobial agents not included on the Accelerate panel o
  • For testing certain antimicrobial agents as discussed in AST limitations below O
  • For testing samples for which an "indeterminate" result for any probe was O obtained
  • To obtain isolates for epidemiologic testing. o

In addition, the PhenoTest BC kit does not provide species identification for members of Streptococcus (STR probe). Species detected by the STR probe include S. mitis, S. oralis, S. gallolyticus, S. agalactiae and S. pneumoniae. Subculture and alternative identification methods should be employed to determine the presence of S. pneumoniae. The following limitation was included in the device labeling:

Additional subculture is required for the identification of S. pneumoniae in cases of a positive Streptococcus spp. call.

  • a. Clinical Sensitivity and Specificity, Identification Assay
    For identification results, performance was evaluated by calculating sensitivity and specificity (for prospectively collected fresh samples) or PPA and NPA (for seeded samples). The

{39}------------------------------------------------

performance of the ID assay for each of the target probes is presented in Tables 27-29 below for the 1850 analyzed samples.

ProbeSpecimenTypeSensitivity/PPASpecificity/NPA
TP/(TP +FN)%95% CITN/(TN + FP)%95% CI
CNSProspective Fresh145/15792.4(87.1-95.6)595/60997.7(96.2-98.6)
Seeded98/98100(96.2-100.0)863/87698.5(97.5-99.1)
Total243/25595.3(92.0-97.3)1458/148598.2(97.4-98.8)
EFSProspective Fresh33/3594.3(81.4-98.4)748/748100.0(99.5-100.0)
Seeded65/6698.5(91.9-99.9)978/98099.8(99.3-100.0)
Total98/10197.0(91.6-99.0)1726/172899.9(99.6-100.0)
EFMProspective Fresh20/20100(83.89-100.0)762/77099.0(98.0-99.5)
Seeded80/8297.6(91.5-99.3)962/96999.3(98.5-99.7)
Total100/10298.0(93.1-99.5)1724/173999.1(98.6-99.5)
SAUProspective fresh173/17897.2(93.6-98.8)586/59798.2(96.7-99.0)
Seeded65/65100(94.4-100.0)962/97598.7(97.7-99.2)
Total238/24397.9(95.3-99.1)1548/157298.5(97.7-99.0)
SLUProspective fresh1/250.0(2.6-97.4)784/784100(99.5-100.0)
Seeded76/7798.7(93.0-99.9)964/96599.9(99.4-100.0)
Total77/7997.5(91.2-99.3)1748/174999.9(99.7-100.0)
STRProspective fresh37/4092.5(80.1-97.4)711/74196.0(94.3-97.2)
Seeded134/13698.5(94.8-99.6)904/91399.0(98.1-99.5)
Total171/17697.2(93.5-98.8)1615/165497.6(96.8-98.3)

Table 27. Summary of Identification Assay Performance, Gram-Positive Organisms

Table 28. Summary of Identification Assay Performance. Gram-Negative Organisms
ProbeSpecimen TypeSensitivity/PPASpecificity/NPA
TP/(TP + FN)%95% CITN/(TN + FP)%95% CI
ABAProspective fresh3/3100(43.9-100.0)777/78099.6(98.9-99.9)
Seeded65/6698.5(91.9-99.9)987/99099.7(99.1-100.0)
Total68/6998.6(92.2-99.9)1764/177099.7(99.3-99.8)
CITProspective fresh5/5100(56.6-100.0)773/77899.4(98.5-99.7)
Seeded87/9096.7(90.7-98.9)904/91199.2(98.4-99.6)
Total92/9596.8(91.1-98.9)1677/168999.3(98.8-99.6)
ENTProspective fresh12/1485.7(60.1-96.0)759/76499.4(98.5-99.7)
Seeded95/9699.0(94.3-100.0)923/92799.6(98.9-99.8)
Total107/11097.3(92.29-99.07)1682/169199.5(98.99-99.72)
ECOProspective fresh118/12197.5(93.0-99.2)662/66599.6(98.7-99.9)
Seeded26/2796.3(81.7-99.8)1028/103099.8(99.3-100.0)
Total144/14897.3(93.3-98.9)1690/169599.7(99.3-99.9)
KLEProspective fresh62/6693.9(85.4-97.6)714/71599.9(99.2-100.0)
Seeded60/6198.4(91.3-99.9)937/94299.5(98.8-99.8)
Total122/12796.1(91.1-98.3)1651/165799.6(99.2-99.8)
PROProspective fresh10/1190.9(62.3-99.5)779/779100.0(99.5-100.0)
Seeded76/7798.7(93.0-99.9)972/97999.3(98.5-99.7)

{40}------------------------------------------------

Sensitivity/PPASpecificity/NPA
ProbeSpecimen TypeTP/(TP +FN)%95% CITN/(TN + FP)%95% CI
Total86/8897.7(92.1-99.4)1751/175899.6(99.2-99.8)
PAEProspective fresh19/19100(83.2-100.0)766/77199.4(98.5-99.7)
Seeded39/39100(91.0-100.0)1013/101899.5(98.9-99.8)
Total58/58100(93.8-100.0)1779/178999.4(99.0-99.7)
SMAProspective fresh10/10100(72.3-100.0)782/782100.0(99.5-100.0)
Seeded41/41100(91.4-100.0)1013/101599.8(99.3-100.0)
Total51/51100(93.0-100.0)1795/179799.9(99.6-100.0)

Table 29. Summary of Identification Assay Performance, Candida spp.

ProbeSpecimen TypeTP/(TP+FN)%95% CITN/(TN+FP)%95% CI
CALProspective fresh4/4100.0(51.0-100.0)777/78099.6(98.9-99.9)
Seeded41/41100.0(91.4-100.0)1000/100599.5(98.8-99.8)
Total45/45100.0(92.1-100.0)1777/178599.6(99.1-99.8)
CGLProspective fresh10/10100.0(72.3-100.0)752/76897.9(96.6-98.7)
Seeded40/40100.0(91.2-100.0)1002/101598.7(97.8-99.3)
Total50/50100.0(92.9-100.0)1754/178398.4(97.7-98.9)

Monomicrobic and Polymicrobic Calls

The ability of the PhenoTest BC kit to determine if a sample contains a single organism (monomicrobic) and the ability to detect polymicrobic samples was evaluated with prospectively collected fresh samples (Tables 30 and 31). For monomicrobial results the PhenoTest BC Kit uses multiple tests and an algorithm driven analysis to determine that a sample contains a single organism with no evidence of additional organisms. Result reports for samples determined to be "monomicrobic" contain the following comment: "Monomicrobial: Sample positive for only one pathogen". Of 793 fresh prospective samples evaluated with PhenoTest BC, 557 were determined to be monomicrobial by the PhenoTest BC kit. Of the samples with a PhenoTest BC monomicrobial call, 545 were also monomicrobial by the reference method giving a 97.8% agreement for the PhenoTest BC kit monomicrobial call as compared to the reference method. Twelve samples were incorrectly identified as monomicrobial by PhenoTest BC; after Gram stain mitigation (seven samples clearly mitigated by bacillus/cocci morphology, three samples with gram-positive cocci potentially mitigated by cell arrangement, cocci in chains vs. cocci in clusters) two samples remained unmitigated (Table 30).

Seven of 38 samples determined to be polymicrobic by the reference method were also found to be polymicrobic by the PhenoTest BC kit (Table 31). The reference method reported 755 monomicrobic prospectively collected samples (95.2% of all prospective samples). Of these, 664 were potentially monomicrobic by the PhenoTest BC kit (including monomicrobic, indeterminate and false positive calls.)

{41}------------------------------------------------

Table 30. Performance of Monomicrobial Calls, Fresh Prospective Samples
Total No. Prospective Fresh Samples793
Total No. Monomicrobial by PhenoTest BC Kit (with no evidence ofadditional organisms detected)557
No. Monomicrobial by PhenoTest BC Kit and Reference Method545
Percent Correct by PhenoTest BC97.8
No. False Monomicrobial by PhenoTest BC12
No. Mitigated by Gram Stain reaction (gram-positive vs. gram-negative)7
No mitigated by gram stain by cellular morphology3
No. unmitigated by Gram stain2

Table 30, Performance of Monomicrobial Calls, Evesh Prospective Sa

Table 31. Fresh Prospective Samples by Outcome

OutcomeReferencePhenoTest BC Kit
No. Fresh Prospective Samples (%)793872
No. Monomicrobic755 (95.2)664 (76.1)b
No. Polymicrobic38 (4.8)7 (0.8)b
False PositiveN/A95 (10.9)b
IndeterminateaN/A27 (3.1)b
InvalidN/A79 (9.1)b

4 Samples with only indeterminate results. Samples with a combination of valid and indeterminate results are included in the monomicrobic, polymicrobic or false positive categories

6 PhenoTest BC Kit prospective outcome rates are calculated considering valid and invalid results for a total of 872

False Positives, Invalid and Indeterminate Results

Indeterminate, false positive and invalid results were obtained during the course of the clinical study. See Tables 35 and 36 for summaries of the indeterminate results. See Tables 37 and 38 for a summary of the false positive and invalid results, respectively.

AST Results

For the AST assay, performance was determined generally based on criteria outlined in the Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems including essential agreement (EA) and categorical agreement (CA) with results determined by replicate testing of the broth microdilution reference method. The number and percent of major and very major errors was also evaluated. Essential agreement was calculated as the percentage of MIC results that fell within ± 1 doubling dilution of the broth microdilution result; CA was calculated as the percentage of interpretive categories that exactly matched the interpretive categories obtained with the broth microdilution method.

For MIC values that are on-scale, the percent EA of evaluable results was determined. This assessment was only performed for those drugs for which the number of evaluable results was at least 55% of the total number of results. Analysis was also performed to determine MIC trending (higher or lower dilution compared to the reference MIC values).

Ampicillin. A total of 238 Enterococcus spp. isolates were evaluated with ampicillin (96 E. faecium isolates, and 142 E. faecalis isolates). The combined results from clinical and challenge testing demonstrated an EA of 100% and CA of 99.6% (Table 32). There was a single major error with E. faecium that was considered a random error.

{42}------------------------------------------------

Ceftaroline. A total of 344 S. aureus isolates were evaluated with ceftaroline. The combined results from clinical and challenge testing demonstrated an EA of 93.3% and CA of 99.7% (Table 32). There were an insufficient number of resistant isolates tested which was addressed with the following limitation in the device labeling:

The ability of the Accelerate PhenoTest BC kit to detect resistance in the following combination is unknown because an insufficient number of resistant isolates were encountered at the time of comparative testing: Ceftaroline -S. aureus

Daptomycin. A total of 332 Staphylococcus isolates were evaluated with daptomycin (197 S. aureus isolates, 135 coagulase negative Staphylococcus isolates). There is only a susceptible breakpoint for daptomycin with Staphylococcus spp. (≤1 µg/mL). The combined results from clinical and challenge testing demonstrated an EA for Staphylococcus of 99.1% and CA of 99.7%. There was one very major error with a S. aureus isolate that was considered a random error; however, because an insufficient number resistant isolates of S. aureus isolates were evaluated during the clinical trial, the performance with resistant isolates is unknown. The following limitation was added to the device labeling:

The ability of the Accelerate PhenoTest BC kit to detect resistance in the following combination is unknown because an insufficient number of resistant isolates were encountered at the time of comparative testing: Daptomycin - S. aureus

A total of 111 Enterococcus isolates were evaluated with daptomycin (71 E. faecium isolates, 40 E. faecalis isolates). There is only a susceptible breakpoint for daptomycin with Enterococcus spp. (<4 ug/mL). The combined results from clinical and challenge testing demonstrated an EA for Enterococcus spp. of 95.5% and CA of 99.1% (Table 32). There was one major error with an E. faecium isolate which was considered a random error. Potential major errors with daptomycin are addressed with the following limitation in the device labeling:

The current absence of data on daptomycin-resistant isolates precludes defining any categories other than "Susceptible." Isolates yielding test results suggestive of a "Non-Susceptible" category should be retested, and if the result is confirmed, the isolate should be retested using the reference method.

Erythromycin. A total of 194 S. aureus isolates were initially tested with erythromycin. While the EA and CA were acceptable, results showed a high occurrence of very major errors (5.9%). A correction was made to the AST model and regression analysis was used to validate results from the initial testing. Final analysis of erythromycin was made using a total of 338 S. aureus isolates were evaluated with erythromycin. The combined results from clinical and challenge testing demonstrated an EA of 98.2% and CA of 96.7% (Table 32). There was one major error that was considered a random error.

Linezolid. A total of 194 S. aureus isolates were evaluated with linezolid (breakpoints ≤4, - ≥8 µg/mL). The combined results from clinical and challenge testing demonstrated an EA for S. aureus of 99.5% and CA of 100% (Table 32). The EA of evaluable results for S. aureus was 99.5%. A total of 110 Enterococcus spp. isolates were evaluated (41 E. faecalis isolates, 69 E. faecium isolates) (breakpoints ≤2, 4, ≥8 µg/mL). The combined results from clinical and challenge testing demonstrated an EA for Enterococcus spp. of 96.4% and CA of 98.2% (Table 32). The EA of evaluable results for Enterococcus spp. was 100%.

{43}------------------------------------------------

Vancomycin. A total of 275 coagulase negative Staphylococcus, and Enterococcus spp. were evaluated with vancomycin (134 coagulase negative Staphylococcus isolates, 29 S. lugdunensis isolates, 41E. faecalis isolates and 71 E. faecium isolates) (breakpoints ≤4, 8-16, ≥32) ug/mL. The combined results from clinical and challenge testing demonstrated an EA for coagulase negative Staphylococcus. S. lugdunensis and Enterococcus spp. of 96.4% and CA of 96.4%. A total of 198 S. aureus isolates were evaluated with vancomycin (breakpoints ≤2, 4-8, ≥16 ug/mL). The combined results from clinical and challenge testing demonstrated an EA for S. aureus of 98.0% and CA of 99.0% (Table 32). There were an insufficient number of vancomycinintermediate S. aureus (VISA) isolates evaluated and the following limitation was included in the device labeling:

The ability of the Accelerate PhenoTest BC kit to detect vancomycin-intermediate Staphylococcus aureus isolates (VISA) is unknown because insufficient numbers of VISA isolates were evaluated at the time of comparative testing.

Amikacin. A total of 432 isolates were evaluated with amikacin (47 A. baumannii isolates, 38 Citrobacter spp. isolates, 110 E. coli isolates, 51 Enterobacter spp. isolates, 78 Klebsiella spo. isolates, 31 Proteus spp. isolates, 35 S. marcescens isolates, 42 P. aeruginosa isolates). The combined results from clinical and challenge testing demonstrated an EA of 94.2% and CA of 94.0% (Table 33). There were an insufficient number of resistant isolates of Citrobacter sp., Enterobacter spp., E. coli, Proteus spp. and S. marcescens tested and the following limitation was included in the device labeling:

The ability of the Accelerate PhenoTest BC kit to detect resistance in the following combination is unknown because an insufficient number of resistant isolates were encountered at the time of comparative testing: Amikacin - Citrobacter spp., Enterobacter spp., E. coli, Proteus spp., S. marcescens

Testing of A. baumannii with amikacin demonstrated an EA of 80.9% and CA of 80.9%. The low CA was due to a high number of minor errors. Isolates of A. baumannii that were out of EA with the reference method were isolates that gave resistant MIC values with PhenoTest BC but gave intermediate MIC results with the reference method. To address the low EA of A. baumannii with amikacin the following limitation was included in the device labeling:

The ability of the Accelerate PhenoTest BC kit to provide accurate MICs with amikacin resistant strains of A. baumannii has not been established; isolates of this species that provide resistant results should be confirmed by an alternative method.

Ampicillin/Sulbactam. A total of 322 isolates were evaluated with ampicillin/sulbactam (186 E. coli isolates, 93 Klebsiella spp. isolates, 43 Proteus spp. isolates). The combined results from clinical and challenge testing demonstrated an EA of 91.0% and CA of 84.2% (Table 33). The low CA was due to the occurrence of a high number of minor errors with E. coli and Klebsiella sp. A single very major error was observed with Proteus spp. and was considered a random error. For Proteus spp. the essential agreement was low at 81.4%. Analysis of performance of Proteus spp. using truncated reporting ranges improved the EA to 90.7%. Analysis of trending indicated that MIC values for Enterobacteriaceae tended to be one doubling dilution higher than the reference MIC value. The following statement was included as a footnote to the AST performance table:

{44}------------------------------------------------

Accelerate PhenoTest BC kit ampicillin/sulbactam MIC values for Enterobacteriaceae tended to be one doubling dilution higher than the reference MIC value.

Aztreonam. A total of 348 isolates of Enterobacteriaceae were evaluated with aztreonam (38 Citrobacter spp. isolates, 124 E. coli isolates, 39 Enterobacter spp. isolates, 73 Klebsiella spp. isolates, 38 Proteus spp. isolates, 36 S. marcescens isolates). The combined results from clinical and challenge testing demonstrated a combined EA of 96.6% and CA of 97.7% (Table 33). A single very major error with E. coli and a single major error with Enterobacter spp. were considered random errors. There were an insufficient number of resistant isolates of Proteus spp. and S. marcescens tested and the following limitation was included in the device labeling:

The ability of the Accelerate PhenoTest BC kit to detect resistance in the following combination is unknown because an insufficient number of resistant isolates were encountered at the time of comparative testing: Aztreonam - Proteus spp., S. marcescens.

Analysis of trending indicated that MIC values for Enterobacteriaceae tended to be one doubling dilution higher than the reference MIC value. The following statement was added as a footnote to the AST performance table:

Accelerate PhenoTest BC kit aztreonam MIC values for Enterobacteriaceae tended to be one dilution higher than the reference MIC value.

Cefepime. A total of 349 isolates of Enterobacteriaceae were evaluated with cefepime (37 Citrobacter spp. isolates, 124 E. coli isolates, 40 Enterobacter spp. isolates, 74 Klebsiella spo. isolates, 37 Proteus spp. isolates, 37 S. marcescens isolates) (breakpoints ≤2, 4-8, ≥16 µg/mL). The combined results from clinical and challenge testing demonstrated a combined EA of 97.7% and CA of 96.8% (Table 33). There was a single very major error observed with E. coli which considered a random error. There were an insufficient number of resistant isolates of Citrobacter spp., Proteus spp. and S. marcescens tested and the following limitation is included in the device labeling:

The ability of the Accelerate PhenoTest BC kit to detect resistance in the following combination is unknown because an insufficient number of resistant isolates were encountered at the time of comparative testing: Cefepime - Citrobacter spp., Proteus spp., S. marcescens.

A total of 42 isolates of P. aeruginosa were evaluated with cefepime (breakpoints <8. -. >16 ug/mL). The combined results from clinical and challenge testing demonstrated a combined EA of 88.1% and CA of 92.9% which was considered to be acceptable performance (Table 33). The EA of evaluable results for P. aeruginosa was 83.3%. Three major errors observed with P. aeruginosa and were addressed with the following limitation in the device labeling:

Cefepime with P. aeruginosa may produce a resistant result that can be found susceptible by the reference method. If critical to patient care confirm these results with an alternate method.

Ceftazidime. A total of 377 isolates of Enterobacteriaceae were evaluated with ceftazidime (74 Citrobacter spp. isolates. 122 E. coli isolates. 40 Enterobacter spp. isolates. 69 Klebsiella spo. isolates, 38 Proteus spp. isolates, 34 S. marcescens isolates) (breakpoints ≤4, 8, ≥16 µg/mL). The combined results from clinical and challenge testing demonstrated an EA of 86.2% and a CA of

{45}------------------------------------------------

93.9% (Table 33). Analysis of performance using truncated reporting ranges improved the EA for Enterobacteriaceae to 93.1%. Essential agreement of evaluable results was 77.4% for Enterobacteriaceae. There were an insufficient number of resistant isolates of Proteus spp. and S. marcescens tested and the following limitation was included in the device labeling:

The ability of the Accelerate PhenoTest BC kit to detect resistance in the following combination is unknown because an insufficient number of resistant isolates were encountered at the time of comparative testing: Ceftazidime-Proteus spp., S. marcescens.

Analysis of trending indicated that MIC values for Enterobacteriaceae tended to be one doubling dilution higher than the reference MIC value. The following statement was added as a footnote to the AST performance table:

Accelerate PhenoTest BC kit ceftazidime MIC values for Enterobacteriaceae tended to be one doubling dilution higher than the reference method.

A total of 53 isolates of P. aeruginosa were evaluated with ceftazidime (breakpoints ≤8, - ≥16 ug/mL). The combined results from clinical and challenge testing demonstrated an EA of 86.8% and CA of 88.7% (Table 33). Analysis of performance using truncated reporting ranges improved the EA to 90.6%. There were six major errors observed with P. aeruginosa and were addressed with the following limitation in the device labeling:

Ceftazidime with P. aeruginosa may produce a resistant result that can be found susceptible by the reference method. If critical to patient care confirm these results with an alternate method. Any P. aeruginosa isolates that provides an MIC ≥ 16ug/mL should be retested with an alternate method.

Analysis of trending indicated that MIC values for P. aeruginosa tended to be one doubling dilution lower than the reference MIC value. The following statement was added as a footnote to the AST performance table:

Accelerate PhenoTest BC kit ceftazidime MIC values for P. aeruginosa tended to be one doubling dilution lower than the reference method.

Ceftriaxone. A total of 324 isolates of Enterobacteriaceae were evaluated with ceftriaxone (30 Citrobacter spp. isolates, 111 E. coli isolates, 40 Enterobacter spp. isolates, 70 Klebsiella spo. isolates, 33 Proteus spp. isolates, 40 S. marcescens isolates). The combined results from clinical and challenge testing demonstrated an EA of 89.8% and a CA of 96.6% which was determined to be acceptable (Table 33). For Proteus spp. with ceftriaxone the essential agreement was low at 87.7%. Analysis of performance using truncated reporting ranges improved the EA for Proteus spp. to 93.9%. There were an insufficient number of resistant isolates of Citrobacter spp., Enterobacter cloacae and S. marcescens tested and the following limitation was included in the device labeling:

The ability of the Accelerate PhenoTest BC kit to detect resistance in the following combination is unknown because an insufficient number of resistant isolates were encountered at the time of comparative testing: Ceftriaxone - Citrobacter spp., E. cloacae, S. marcescens.

{46}------------------------------------------------

The essential agreement of ceftriaxone with S. marcescens was unacceptable at 45.0%; CA was 82.5% due to a high number of minor errors. Truncation of the reporting range did not improve the %EA. The low EA for S. marcescens was addressed with the following limitation in the device labeling:

Due to a low essential agreement for S. marcescens with ceftriaxone, results should be confirmed with an alternate method if critical to patient care.

Ciprofloxacin. A total of 394 isolates of Enterobacteriaceae and P. aeruginosa were evaluated with ciprofloxacin (38 Citrobacter spp. isolates, 125 E. coli isolates, 40 Enterobacter spp. isolates, 74 Klebsiella spp. isolates, 38 Proteus spp. isolates, 37 S. marcescens isolates, 42 P. aeruginosa isolates). The combined results from clinical and challenge testing demonstrated an EA of 96.7% and a CA of 98.2% (Table 33). For Enterobacter spp. with ciprofloxacin the EA was 85.0%; analysis of performance using truncated reporting ranges improved the EA for Enterobacter spp. to 100%. There were an insufficient number of resistant isolates of Citrobacter spp.. Proteus spp. and S. marcescens tested and the following limitation was included in the device labeling:

The ability of the Accelerate PhenoTest BC kit to detect resistance in the following combination is unknown because an insufficient number of resistant isolates were encountered at the time of comparative testing: Ciprofloxacin - Citrobacter spp., Proteus spp., S. marcescens.

Ertapenem. A total of 351 isolates of Enterobacteriaceae were evaluated with ertapenem (40 Citrobacter spp. isolates, 125 E. coli isolates, 40 Enterobacter spp. isolates, 71 Klebsiella spp. isolates, 38 Proteus spp., 37 S. marcescens isolates). The combined results from clinical and challenge testing demonstrated an EA of 98.9% and a CA of 98.6% (Table 33). There were two major errors with Enterobacter spp. which were addressed by the following limitation in the device labeling:

Ertapenem with Enterobacter spp. may produce a resistant result that can be found susceptible by the reference method. If critical to patient care, confirm these results with an alternate method.

There were an insufficient number of resistant isolates of Citrobacter spp. . Proteus spp. and S. marcescens tested and the following limitation was included in the device labeling:

The ability of the Accelerate PhenoTest BC kit to detect resistance in the following combination is unknown because an insufficient number of resistant isolates were encountered at the time of comparative testing: Ertapenem - Citrobacter spp., Proteus spp., S. marcescens.

Analysis of trending indicated that MIC values for Enterobacteriaceae tended to be one doubling dilution higher than the reference MIC value. The following statement was added as a footnote to the AST performance table:

Accelerate PhenoTest BC kit ertapenem MIC values for Enterobacteriaceae tended to be one doubling dilution higher than the reference MIC value.

{47}------------------------------------------------

Gentamicin. A total of 385 isolates of Enterobacteriaceae and P. aeruginosa were evaluated with gentamicin (39 Citrobacter spp. isolates, 122 E. coli isolates, 40 Enterobacter spp. isolates, 71 Klebsiella spp. isolates. 36 Proteus spp. isolates. 35 S. marcescens isolates. 42 P. aeruginosa isolates). The combined results from clinical and challenge testing demonstrated an EA of 99.2% and a CA of 98.4% (Table 33). The CA for P. geruginosa was low at 88.1% due to a high number of minor errors. There were an insufficient number of resistant isolates of Citrobacter spp., Enterobacter spp., Proteus spp. and S. marcescens tested and the following limitation is included in the device labeling:

The ability of the Accelerate PhenoTest BC kit to detect resistance in the following organisms is unknown because an insufficient number of resistant isolates were encountered at the time of comparative testing: Citrobacter spp., Enterobacter spp., Proteus spp., S. marcescens.

Meropenem. A total of 364 isolates of Enterobacteriaceae were evaluated with meropenem (39 Citrobacter spp. isolates, 124 E. coli isolates, 57 Enterobacter spp. isolates, 71 Klebsiella spp. isolates, 37 Proteus spp. isolates, 36 S. marcescens isolates) (breakpoints <1. 2. >4 ug/mL). The combined results from clinical and challenge testing with Enterobacteriaceae demonstrated an EA of 97.0% and a CA of 98.1% (Table 33). For Enterobacter spp. the EA was low at 87.7%. Analysis of performance using truncated reporting ranges improved the EA for Enterobacter spp. to 91.2%. There was a single major error with E. coli that was considered to be a random error. There were three major errors with Enterobacter spp. which were addressed by the following limitation in the device labeling:

Meropenem with Enterobacter spp. may produce a resistant result that can be found susceptible by the reference method. If critical to patient care, confirm these results with an alternate method.

There were an insufficient number of resistant isolates of Citrobacter spp., E. coli. Proteus spo. and S. marcescens tested and the following limitation is included in the device labeling:

The ability of the Accelerate PhenoTest BC kit to detect resistance in the following combination is unknown because an insufficient number of resistant isolates were encountered at the time of comparative testing: Citrobacter spp., E. coli, Proteus spp., S. marcescens.

A total of 51 isolates of P. aeruginosa were evaluated with meropenem. The combined results from clinical and challenge testing with P. aeruginosa demonstrated an EA of 88.2 and a CA of 90.2% (Table 33) Analysis of performance using truncated reporting ranges improved the EA for P. aeruginosa to 90.2%. There was a single major error that was considered to be a random error.

Piperacillin/Tazobactam. A total of 519 isolates of Enterobacteriaceae, A. baumannii and P. aeruginosa were evaluated with piperacillin/tazobactam (73 Citrobacter spp. isolates, 121 E. coli isolates, 69 Enterobacter spp. isolates, 67 Klebsiella spp. isolates, 36 Proteus spp. isolates, 36 S. marcescens isolates, 47 A. baumannii isolates, 70 P. aeruginosa isolates). The combined results from clinical and challenge testing demonstrated an EA of 92.1% and a CA of 92.1% (Table 32). There was a single very major error with E. coli that was considered a random error. For P. aeruginosa the essential agreement was low at 85.7%. Analysis of performance using truncated reporting ranges improved the EA for P. aeruginosa to 91.4%. There was one major error with A. baumannii and two major errors with Klebsiella spp. The occurrence of major errors was

{48}------------------------------------------------

addressed in the following limitation in the device labeling:

Piperacillin/Tazobactam with Acinetobacter baumannii and Klebsiella spp. may produce a resistant result that can be found susceptible by the reference method. If critical to patient care, confirm these results with an alternate method.

There were an insufficient number of resistant isolates of Proteus spp. and S. marcescens tested and the following limitation is included in the device labeling:

The ability of the Accelerate PhenoTest BC kit to detect resistance in the following combination is unknown because an insufficient number of resistant isolates were encountered at the time of comparative testing: Piperacillin /Tazobactam - Proteus spp., S. marcescens.

Tobramycin. A total of 389 isolates of Enterobacteriaceae and P. aeruginosa were evaluated with tobramycin (36 Citrobacter spp. isolates, 122 E. coli isolates, 40 Enterobacter spp. isolates, 74 Klebsiella spp. isolates, 38 Proteus spp. isolates, 37 S. marcescens isolates, 42 P. aeruginosa isolates). The combined results from clinical and challenge testing demonstrated an EA of 96.4% and a CA of 96.1% (Table 33). There were an insufficient number of resistant isolates of Citrobacter spp., Proteus spp. and S. marcescens tested and the following limitation is included in the device labeling:

The ability of the Accelerate PhenoTest BC kit to detect resistance in the following organisms is unknown because an insufficient number of resistant isolates were encountered at the time of comparative testing: Citrobacter spp., Proteus spp., S. marcescens.

Cefoxitin. A total of 398 Staphylococcus species were evaluated with cefoxitin to determine methicillin resistance (184 S. aureus isolates, 28 S. lugdunensis isolates, 180 coagulase negative Staphylococcus isolates). The combined results from clinical and challenge testing demonstrated CA of 98.2% (Table 34). Evaluation of fresh isolates of coagulase negative Staphylococcus spp. showed five very major errors; additional testing with 71 additional isolates of coagulase negative Staphylococcus demonstrated 100% CA with the reference method. Discordant analysis indicated that one isolate was falsely resistant by the reference method. Considering the lack of very major errors observed with repeat testing the performance of cefoxitin with coagulase negative Staphylococcus spp. was considered acceptable.

Macrolide-Lincosamide Streptogramim B Resistance (MLSb). A total of 164 Staphvlococcus species were evaluated with MLSb to determine inducible clindamycin resistance (29 S. lugdunensis isolates, 135 coagulase negative Staphylococcus isolates). The combined results from clinical and challenge testing demonstrated CA of 98.2% (Table 34). The ability of the PhenoTest BC kit to determine inducible clindamycin resistance in S. aureus has not been established; the following limitation is included in the device labeling:

The ability of the Accelerate PhenoTest BC kit to provide accurate results for S. aureus with MLSb as compared to the reference method has not been established; isolates of this species should be tested by an alternative method.

The ability of the PhenoTest BC kit to detect MLSb has not been established with all blood culture bottle types. The following limitation is included in the device labeling:

{49}------------------------------------------------

Due to insufficient number of test isolates, the ability of the Accelerate PhenoTest BC kit to detect inducible MLSb resistance in coagulase-negative staphylococci is unknown when used with the following blood culture bottle types: BacT/Alert SN Standard Anaerobic, BACTEC Peds Plus/F, BACTEC Plus Anaerobic/F, BACTEC Standard Anaerobic, BACTEC Standard/10 Aerobic, VersaTrek Redox 1 Aerobic, VersaTrek Redox 2 Anaerobic. Use an alternative method for detection of inducible MLSb resistance when using these blood culture bottle types if critical to patient care.

Sample TypeTotNo. EAEA %No. CACA %No. R or NSminmajvmj
Ampicillin - Enterococcus spp.
Fresh4646100.046100.015000
Fresh Seeded33100.03100.02000
Challenge189189100.018899.575010
Total238238100.023799.692010
Ceftaroline - S. aureus
Fresh14013294.3140100.00000
Fresh Seeded201891.020100.00000
Challenge18417192.918399.50100
Total34432193.334399.70100
Daptomycin - Staphylococcus spp.
Fresh25825598.825799.61001
Fresh Seeded2525100.025100.00000
Challenge4949100.049100.00000
Total33232999.133199.71001
Daptomycin - Enterococcus spp.
Fresh484797.948100.00000
Fresh Seeded33100.03100.00000
Challenge605693.35998.30010
Total11110695.511099.10010
Erythromycin – S. aureus
Fresh14013999.313596.476500
Fresh Seeded1818100.018100.010000
Challenge18017597.217496.7119410
Total33833298.232796.72051010
Linezolid S. aureus
Fresh13913899.3139100.00000
Fresh Seeded88100.08100.00000
Challenge4747100.047100.00000
Total19419399.5194100.00000
Linezolid – Enterococcus spp.
Fresh474697.94697.90100
Fresh Seeded33100.03100.00000
Challenge605795.05998.31100
Total11010696.410898.21200
Vancomycin - S. aureus
Fresh14013999.3140100.00000
Fresh Seeded8787.58100.00000
Sample TypeTotNo. EAEA%No. CACA %No. Ror NSminmajvmj
Challenge504896.04896.00200
Total19819498.019699.00200
Vancomycin - S. lugdunensis, Coagulase Negative Staphylococcus, Enterococcus sp.
Fresh16015898.815898.814200
Fresh Seeded1919100.019100.02000
Challenge888090.98090.944800
Total27526596.426596.4601000

Table 32. AST Assay Performance, Gram-Positive Organisms

{50}------------------------------------------------

EA - Essential Agreement (+/- 1 dilution)

CA - Category Agreement

min - minor discrepancies

maj — major discrepancies

R or NS – Resistant or non-susceptible isolates

vmj – very major discrepancies

Essential Agreement (EA) occurs when there is agreement between the result of the reference method and that of the PhenoTest BC Kit within plus or minus one serial two-fold dilution of the antibiotic. Category Agreement (CA) occurs when the interpretation of the reference method agrees exactly with the interpretation of the PhenoTest BC Kit.

Sample TypeTotNo. EAEA %No. CACA %No. R or NSminmajvmj
Amikacin – Enterobacteriaceae, A. baumannii, P. aeruginosa
Fresh18218098.918098.92200
Fresh Seeded2020100.020100.00000
Challenge23020790.020689.6472400
Total43240794.240694.0492600
Ampicillin/Sulbactam - Enterobacteriaceae
Fresh15513687.711976.8493411
Fresh Seeded161487.51487.58200
Challenge15114394.713891.4641300
Total32229391.027184.21214911
Aztreonam - Enterobacteriaceae
Fresh17917396.617597.826211
Fresh Seeded191894.71789.51200
Challenge15014596.714897.861200
Total34833696.634097.788611
Cefepime - Enterobacteriaceae
Fresh18017597.217496.721501
Fresh Seeded1919100.01894.72100
Challenge15014798.014697.340400
Total34934197.733896.8631001
Cefepime - P. aeruginosa
Fresh121083.31083.32020
Fresh Seeded11100.01100.00000
Challenge292689.72896.617010
Total423788.13992.919030
Ceftazidime - Enterobacteriaceae
Fresh17514784.015990.9251600
Fresh Seeded151280.01493.32100
Challenge18716688.818196.881600
Sample TypeTotNo. EAEA %No. CACA %No. R or NSminmajvmj
Total37732586.235493.91082300
Ceftazidime - P. aeruginosa
Fresh12758.3650.01060
Fresh Seeded11100.01100.00000
Challenge403895.040100.027000
Total534686.84788.728060
Ceftriaxone - Enterobacteriaceae
Fresh16615090.416297.632400
Fresh Seeded1414100.014100.02000
Challenge14412788.213795.133700
Total32429189.831396.61071100
Ciprofloxacin – Enterobacteriaceae, P. aeruginosa
Fresh19318696.418796.948600
Fresh Seeded2020100.020100.03000
Challenge18117596.718099.450100
Total39438196.738798.2101700
Ertapenem - Enterobacteriaceae
Fresh18118099.4181100.01000
Fresh Seeded1919100.019100.00000
Challenge15114898.014696.728320
Total35134798.934698.629320
Gentamycin – Enterobacteriaceae and P. aeruginosa
Fresh189189100.0189100.019000
Fresh Seeded1818100.018100.04000
Challenge17817598.317296.632510
Total38538299.237998.455510
Meropenem – Enterobacteriaceae
Fresh18017798.317999.41010
Fresh Seeded1919100.019100.00000
Challenge16515795.215996.434330
Total36435397.035798.135340
Meropenem – P. aeruginosa
Fresh12866.7995.03300
Fresh Seeded11100.01100.00000
Challenge383694.73694.722110
Total514588.24690.225410
Piperacillin/Tazobactam – Enterobacteriaceae, P. aeruginosa, A. baumannii
Fresh18916888.917089.9121630
Fresh Seeded181794.41797.40010
Challenge29331293.929193.31411911
Total51947892.147892.11533551
Tobramycin – Enterobacteriaceae, P. aeruginosa
Fresh19118496.318596.916600
Fresh Seeded191684.21789.52200
Challenge17917597.817296.145700
Total38937596.437496.1631500

Table 33. AST Assay Performance, Gram-Negative Organisms

{51}------------------------------------------------

{52}------------------------------------------------

SampleTypeTotalTestedCA%CANegative(S)Positive(R)majvmj
Cefoxitin – S. aureus, S. lugdunensis, Coagulase Negative Staphylococcus spp.
Fresh22822297.410911915
Seeded2323100.061700
Challenge14114099.33210910
Total39839198.215224625
MLSb - S. lugdunensis, Coagulase Negative Staphylococcus spp.
Fresh11911798.3625711
Seeded171694.151210
Challenge282810027100
Total16416198.2947021

Table 34. AST Assay Performance, Phenotypic Resistance

Indeterminate Results

Indeterminate results are PhenoTest BC Kit results that the target group may or may not be present. Tests resulting in an indeterminate call show characteristics that are significantly different from a negative result but do not meet positive result requirements. Indeterminate results were obtained for all target probes but most commonly with the following probes: CNS, ENT, KLE, SAU and SLU. Table 35 presents the number and percent of indeterminate results obtained in the clinical study by target probe.

Table 35. Clinical Study Indeterminate Results by Probe
ProbeNo. IndeterminateIndeterminatePercent of Total Tests a
ABA00.0%
CAL60.3%
CGL00.0%
CIT00.0%
CNS1105.9%
ECO00.0%
EFM60.3%
EFS40.2%
ENT472.5%
KLE553.0%
PAE00.0%
PRO00.0%
SAU351.9%
SLU221.2%
SMA00.0%
STR90.5%

4 Total number of tests per probe = 1852

{53}------------------------------------------------

Clinical/Analytical StudyNo.TotalTestsNo.Indeterminate(%)
Clinical Performance Evaluation Study1850272 (14.7%)
Reproducibility Study1644264 (16.1%)
Blood Bottle Type Study141974 (5.2%)
Polymicrobial Biological Interference Study4910 (20.4%)
Growth and Detection Study17520 (11.4%)
Inclusivity and Exclusivity Study25337 (14.6%)
Interfering Substances Study52041 (7.9%)
Monomicrobial Limit of Detection84999 (11.7%)
Polymicrobial Limit of Detection18228 (15.4%)

Table 36. Indeterminate Results by Study Type

False Positive Results

False positive results were observed in the clinical study and in all analytical studies. Most false positive results were the result of cross reactivity, including intra-genus cross reactivity with a specific probe. Because all PhenoTest BC Kit results are correlated with results of Gram stain of the positive blood culture, many false positive results are mitigated. The number and percent of false positives and the results of Gram stain mitigation are shown in Table 37 below.

Clinical/Analytical Studies - False Positive Results
Clinical/Analytical StudyNo.TotalTestsNo. FalsePositives(%)No.MitigatedFalsePositives(%)aNo.UnmitigatedFalsePositives (%)
Clinical Performance Evaluation Study - allspecimen types1850179 (9.7)106 (59.2)73 (40.8)
Clinical - Challenge47738 (7.9)29 (76.3)9 (23.7)
Clinical - Fresh79395 (12.0)51 (53.7)44 (46.3)
Clinical - Fresh Seeded654 (6.2)1 (25.0)3 (75.0)
Clinical - Stock51542 (8.2)25 (59.5)17 (40.5)
Reproducibility Study1644131 (8.0)110 (84.0)21 (16.0)
Blood Bottle Type Study107764 (5.9)56 (87.5)8 (12.5)
Polymicrobial Biological Interference Study492 (4.1)1 (50.0)1 (50.0)
Growth and Detection Study1445 (3.5)3 (60.0)2 (40.0)
Inclusivity and Exclusivity Study31881 (25.5)24 (29.6)57 (70.4)
Interfering Substances Study52641 (7.8)33 (80.5)8 (19.5)
Monomicrobial Limit of Detection84995 (11.2)88 (92.6)7 (7.4)
Polymicrobial Limit of Detection1826 (3.3)3 (50.0)3 (50.0)

Table 37. False Positive Results in Clinical and Analytical Studies

a Gram stain mitigation includes differentiating enteric-like gram-negative rods and gram-negative cocco-bacilli suggestive of Acinetobacter spp. Acinetobacter sp. may not consistently demonstrate the gram-negative coccobacilliary morphology.

{54}------------------------------------------------

Invalid Tests.

Invalid results occur when too few cells are present in the control channel for the identification assay. No ID or AST results are reported and the Patient Report states: "Too few cells for analysis. Perform alternate testing method for identification and susceptibility results. "In the analytical studies, non-standard sample preparation designed to control for organism concentration was employed which contributed to some invalid rates. Further, the distribution of organisms in these studies was different from what would be observed in clinical samples. Some invalid results were due to off-panel species. The occurrence of invalid tests is summarized in Table 38 below.

StudyTotal No. TestsNo. Invalid (%)
Clinical Performance Evaluation Study194090a (4.6)
Reproducibility164411 (0.7)
Blood Bottle Type Stage 1115980b (6.9)
Blood Bottle Type Stage 23411 (0.3)
Polymicrobial Biological Interference490 (0)
Growth and Detection14414c (9.7)
Inclusivity and Exclusivity31865d (20.4)
Interfering Substances5266e(1.1)
Monomicrobial Limit of Detection84969f (8.0)
Polymicrobial Limit of Detection1820 (0)

Table 38. Invalid Tests in Clinical and Analytical Studies

4 52 samples invalid due to off panel organisms

6 68 samples invalid due to off-panel organisms

° 6 samples invalid due to off-panel organisms

4 53 samples invalid due to off-panel organisms

€ 6 samples invalid due to off-panel organisms

4 64 samples invalid due to off-panel organisms

  • C. Other clinical supportive data (when a. and b. are not applicable): N/A
    1. Clinical cut-off: N/A

5. Expected values/Reference range:

AntimicrobialOrganisms TestedFDA Breakpoints
SIR
AmikacinEnterobacteriaceae, P. aeruginosa and A. baumannii≤1632≥64
AmpicillinEnterococcus spp.≤8-≥16
CeftarolineS. aureus≤12≥4
Ampicillin/SulbactamEnterobacteriaceae≤816≥32
AztreonamEnterobacteriaceae≤48≥16
CefepimeEnterobacteriaceae≤24-8≥16
P. aeruginosa≤8-≥16

{55}------------------------------------------------

AntimicrobialOrganisms TestedFDA Breakpoints
SIR
CeftazidimeEnterobacteriaceae≤48≥16
P. aeruginosa≤8-≥16
CeftriaxoneEnterobacteriaceae≤12≥4
CiprofloxacinEnterobacteriaceae and P. aeruginosa≤12≥4
DaptomycinStaphylococcus spp.≤1-≥2
Enterococcus spp.≤4-≥8
ErtapenemEnterobacteriaceae≤0.51≥2
ErythromycinS. aureus≤0.51 - 4≥8
GentamycinEnterobacteriaceae and P. aeruginosa≤48≥16
LinezolidS. aureus≤4-≥8
Enterococcus spp.≤24≥8
MeropenemEnterobacteriaceae≤12≥4
P. aeruginosa≤24≥8
Piperacillin/TazobactamEnterobacteriaceae≤1632 - 64≥128
TobramycinEnterobacteriaceae and P. aeruginosa≤48≥16
VancomycinS. aureus≤24 - 8≥16
S. lugdunensis andCoagulase Negative Staphylococcus≤48-16≥32

M. Instrument Name:

Accelerate Pheno System

N. System Descriptions:

    1. Modes of Operation:
      Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?

Yes

Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?

Yes ______________ or No _____________________________________________________________________________________________________________________________________________________

Instrument System

The Accelerate Pheno system is comprised of the Accelerate Pheno instrument, software, host computer, analysis computer, and the Accelerate PhenoTest BC kit. This fully-integrated in vitro diagnostic system utilizes the following components for test samples:

  • One to four modules (where each module can run one patient sample independently) ●
  • Automated pipetting system for transfer of sample and reagents to appropriate cassette . flowcell channels in a predetermined sequence

{56}------------------------------------------------

  • Incubation system to support nucleic acid hybridization and microbiological growth ●
  • Optical system comprised of a digital microscope, camera. and illuminator (both laser . and darkfield light sources) for image capture of fluorescent signals and dark-field images
  • . Embedded control software, Graphical User Interface, and network connectivity
  • Graphical data management, processing, analysis hardware, and software algorithms . (designed to interpret images) in order to report identification and susceptibility results

The Accelerate PhenoTest BC kit contains a sample vial, a 48-channel disposable test cassette, and a reagent cartridge. Each module contains the electronics, motion control components, camera, illumination sources, and optical components necessary to process a patient sample.

    1. Software:
      FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes

Level of Concern: Moderate

Software Description:

A detailed software description of the Accelerate Pheno system software, v1.0 was provided, which included a description of device features controlled by the software. A summary of the features also detailed the software operating environment (e.g. programming. language, hardware platform, operating system, use of off-the-shelf software).

The Accelerate Pheno system software is comprised of three (3) major architectural components:

Major ArchitecturalComponentsFunction
Instrument ControlResponsible for execution of assay steps through modules that gothrough physical procedures and capture/deliver images foranalysis. It also provides the user and LIS interfaces, managementcapabilities, and test result reporting for the modules
Module FirmwareReal-time control of hardware components inside a module(heaters, motors, fluidics, illumination, etc.) and communicateswith instrument control software
Image Processing andAnalysisAnalyzes images captured by the instrument to identify microbialspecies, quantify growth, and provide susceptibility results

Device Hazard Analysis

Risk analysis tables were provided that detailed software/hardware hazards related to the Accelerate Pheno system and the potential impact of each hazard on the performance of FISH ID and AST or possible injury to the user. The risk analysis listed each potential hazard, possible causes of the hazard, severity of the hazard, likelihood of occurrence, risk severity, and corrective measures. After applying mitigations, all potential risks were reduced to "As Far as Possible" or were considered "Broadly Acceptable."

{57}------------------------------------------------

For each hazard, identified lines provided software traceability to each Software Requirements Specification (including the Software Requirement Items) and Software Test Plan/Procedure. Mitigations included design, manufacturing controls, supplier management, product labeling, verification/validation and quality control testing.

Architecture Design Chart:

Diagrams of the system software architecture were provided to cover the Physical and Logical Architecture of the Accelerate Pheno system.

Software Requirements Specification (SRS) and Software Design Specification (SDS):

SRS documents were provided that described the requirements for various components of the Accelerate Pheno system. Each Software Requirement Item (SRI) specified the software requirement to be fulfilled for the product, including appropriate performance, function, reliability, and timing. New SRIs were assigned for new features, enhancements, and for any issue resolution. The SRS were under revision control and were updated with each newly assigned SRI. Requirements defined in the SRS document were implemented according to the design specifications described.

Traceability Analysis:

A traceability matrix that linked requirements, specifications, hazards, mitigations and verification & validation testing for the software was acceptable.

Software Development Environment Description:

The software development life cycle plan and software configuration management plan for the Accelerate Pheno system was acceptable.

Verification and Validation Testing (V&V):

Based on the SRS and SDS, software verification test plans were devised to verify that software met the requirements. Accelerate Pheno system software testing was successfully completed at unit, integration, and system levels. Pass/fail criteria were reported along with test results in the Software Validation Summary Reports.

Revision Level History:

The software revision history record for the Accelerate Pheno system software was acceptable.

Unresolved Anomalies:

A report of V&V testing throughout the software life cycle also included a list of all unresolved anomalies (as a result of V&V testing), known issues (before V&V testing), and resolved issues (after V&V testing). A risk analysis for each issue was performed, and results indicated that all remaining issues were those of low severity (Broadly Acceptable) and would not impact device performance. Any remaining issues, along with a work-around, will either be communicated to user through labeling or fixed in future software versions. The impact analyses of the unresolved anomalies on device safety and effectiveness were considered acceptable.

ES and EMC Testing:

The Accelerate Pheno system was tested and certified to the applicable electrical safety standards IEC 61010-1:2010 (3rd edition)—Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use - Part 1: General Requirements. The device also complied with electromagnetic compatibility, emission and immunity and the requirements of IEC 61326-2-6 and other applicable standards. Operations, limitations hazards, and warnings were listed in the User Guide.

{58}------------------------------------------------

3. Specimen Identification:

Sample vial can be labeled with barcode or written label. If the barcode is unreadable, the user can manually enter sample information to continue the run.

4. Specimen Sampling and Handling:

The labeled Accelerate Pheno system sample vial is filled with an aliquot of positive blood culture. The sample vial cap is screwed on tightly. The test must be initiated within 15 minutes of placing the positive blood culture media in the Accelerate sample vial. A labeled sample vial containing positive blood culture is placed into the sample vial receptacle on the reagent cartridge, where the vial label is facing outward.

    1. Calibration:
      There is no operator scheduled maintenance or operator replaceable parts or materials on the Accelerate Pheno system. An annual preventative maintenance visit by an authorized Accelerate Diagnostics service representative is recommended. Module relocation by the customer is not recommended. Relocation and recalibration should be performed by Accelerate trained service personnel before resuming normal operation. Contact Accelerate Diagnostics Technical Support to arrange for Accelerate trained service personnel to relocate the module.

Quality Control (OC): 6.

Three internal process controls are included in each assay:

  • Universal Bacterial Probe and Universal Eukaryotic Probe for yeast ●
  • General Nucleic Acid Stain
  • Growth Control Channel .

Instructions for OC testing with external controls are provided in the instructions for use. For OC panels used to monitor performance of ID and AST function of the Accelerate Pheno system. only one OC panel should be run on a given module at a time. Organisms for the OC panel being run should be inoculated into specified individual empty wells of the reagent cartridge. QC testing should be rotated between the modules such that all modules are used to perform OC at approximately the same frequency. External controls should be tested in accordance with the appropriate accrediting organization requirements.

O. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

P. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable, and the special controls for this device type.

O. Identified Risks to Health and Identified Mitigations

Identified Risks to HealthIdentified Mitigations
If identification assay is included, false positiveGeneral controls and special controls (1), (2),

{59}------------------------------------------------

or false negative results or incorrect(3), (4), and (5)
identifications can lead to
• a delay in determining the true cause ofthe infection
• unnecessary, ineffective or lack ofantimicrobial therapy
• delayed or skipped treatments ordiagnostic procedures
• inappropriate infection prevention andcontrol measures/and or public healthprocedures
• interference with antimicrobialstewardship efforts
Failure to perform appropriate AST testing mayresult in
• unnecessary, ineffective or lack ofantimicrobial therapy
• interference with antimicrobialstewardship efforts
• development of antimicrobial resistance
An organism determined to be resistant when it issusceptible may lead to
• treatment with an ineffective antibiotic
• administration of unnecessary broadspectrum drugs
• adverse effects from antimicrobials
• costly implementation of infectioncontrol measures
An organism determined to be susceptible when itis resistant may lead to
• treatment with an ineffective antibiotic
• increased morbidity or death
Errors in InterpretationGeneral controls and special control (6)
Failure to correctly operate the test systemGeneral controls and special control (7)

{60}------------------------------------------------

R. Benefit/Risk Analysis

Summary
Summary ofthe Benefit(s)The PhenoTest BC Kit is the first multiplexed in vitro diagnostic platform to use rapid nucleic acid fluorescence in situ hybridization (FISH) identification and quantitative AST to provide identification and susceptibility testing for sixteen pathogenic species of bacteria and yeast to aid in the diagnosis of bacteremia and fungemia. The PhenoTest BC Kit returns results to patients and healthcare providers within six and a half hours of positive blood culture result, compared to current methods which may take 1-3 days to return results. The PhenoTest BC Kit demonstrated PPA greater than 95%, NPA greater than 99%, and EA/CA greater than 90% for all microbial and drug targets.
Summary ofthe Risk(s)False positive results, false negative results, misidentifications and incorrect AST results are the primary risks associated with use of the PhenoTest BC Kit. A false positive result may lead to unnecessary antimicrobial therapy or incorrect antimicrobial therapy with subsequent delay in diagnosis. Patients may receive unnecessary infection control measures, such as contact isolation, if a highly resistant organism is thought to be present. A false negative result may result in the inappropriate discontinuation or delay of antimicrobial therapy, with subsequent worsening of infection. Patients may also not receive sufficient infection control measures, such as contact isolation. Misidentifications may occur if the device identifies the wrong microorganism and could result in the ineffective antimicrobial therapy and/or delay to appropriate therapy. Misidentifications could also result in incorrect infection control measures. Errors in AST result could result in the wrong antibiotic choice with subsequent worsening of infection, or unnecessarily broad antibiotic treatment with the potential for the development of antimicrobial resistance or increased drug side effects.
Summary ofOther FactorsNone
ConclusionsDo theprobablebenefitsoutweigh theprobable risks?The probable benefits of the PhenoTest BC Kit outweigh the potential risks in light of the listed special controls and applicable general controls, including design controls. The PhenoTest BC Kit is the first of its kind and is likely to benefit patients through more timely identification and susceptibility testing of bloodstream infections, with the potential for improved patient outcomes and enhanced antimicrobial stewardship. The high performance observed during the pivotal clinical trials and the proposed special controls suggest that errors will be uncommon and will be well mitigated by laboratory practices and product labeling. The PhenoTest BC Kit could provide substantial benefits to the management of bloodstream infections as a complementary technology to current standard of care blood culture technology.

{61}------------------------------------------------

Patient Perspectives

This submission did not include specific information on patient perspective for this device.

S. Conclusion:

The information provided in this de novo submission is sufficient to classify this device into class II under regulation 21 CFR 866.1650.

FDA believes that the stated special controls, and applicable general controls, including design controls, provide reasonable assurance of the safety and effectiveness of the device type. The device is classified under the following:

Product Code:PRH, NSU, PEO, PAM, PEN, LON
Device Type:A cellular analysis system for multiplexed antimicrobial susceptibility testing
Class:II (special controls)
Regulation:21 CFR 866.1650
  • (a) Identification. A cellular analysis system for multiplexed antimicrobial susceptibility testing is a multiplex qualitative and/or quantitative in vitro device intended for the identification and determination of the antimicrobial susceptibility results of organisms detected in samples from patients with suspected microbial infections. This device is intended to aid in the determination of antimicrobial susceptibility or resistance when used in conjunction with other laboratory findings.
  • (b) Classification. Class II (special controls). A cellular analysis system for multiplexed antimicrobial susceptibility testing must comply with the following special controls:
      1. Premarket notification submissions must include detailed device description documentation, including the device components, ancillary reagents required but not provided, a detailed explanation of the methodology including primer/probe sequence, design, rationale for sequence selection and details of the antimicrobial agents, as applicable.
      1. Premarket notification submissions must include detailed documentation from the following analytical and clinical performance studies: limit of detection, inclusivity, precision, reproducibility, interference, cross reactivity, carry-over, and cross contamination, quality control and additional studies as applicable to specimen type and assay claims.
      1. Premarket notification submissions must include detailed documentation from an appropriate clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted reference methods.
      1. Premarket notification submissions must include detailed documentation for device software, including, but not limited to, software applications and hardware-based devices that incorporate software.

{62}------------------------------------------------

    1. The 21 CFR 809.10(b) compliant labeling must include limitations and protocols regarding the need for correlation of results by standard laboratory procedures as applicable.
    1. A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
    1. A detailed explanation of the principles of operation and procedures for assay performance and troubleshooting must be included in the device's 21 CFR 809.10(b) compliant labeling.

§ 866.1650 A cellular analysis system for multiplexed antimicrobial susceptibility testing.

(a)
Identification. A cellular analysis system for multiplexed antimicrobial susceptibility testing is a multiplex qualitative and/or quantitative in vitro diagnostic device intended for the identification and determination of the antimicrobial susceptibility results of organisms detected in samples from patients with suspected microbial infections. This device is intended to aid in the determination of antimicrobial susceptibility or resistance when used in conjunction with other laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) Detailed device description documentation, including the device components, ancillary reagents required but not provided, a detailed explanation of the methodology, including primer/probe sequence, design, rationale for sequence selection, and details of the antimicrobial agents, as applicable.
(ii) Detailed documentation from the following analytical and clinical performance studies: limit of detection, inclusivity, precision, reproducibility, interference, cross-reactivity, carryover, and cross-contamination, quality control and additional studies, as applicable to specimen type and assay intended use.
(iii) Detailed documentation from an appropriate clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted reference methods.
(iv) Detailed documentation for device software, including software applications and hardware-based devices that incorporate software.
(2) The labeling required under § 809.10(b) of this chapter must include:
(i) Limitations and protocols regarding the need for correlation of results by standard laboratory procedures, as applicable.
(ii) A detailed explanation of the interpretation of results and acceptance criteria.
(iii) A detailed explanation of the principles of operation and procedures for assay performance and troubleshooting.