K Number
DEN070014

Validate with FDA (Live)

Device Name
MANDOMETER III
Manufacturer
Date Cleared
2011-03-31

(1380 days)

Product Code
Regulation Number
882.5060
Age Range
N/A
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
Device Description
AI/ML Overview

Loading...

§ 882.5060 Conditioning tool for eating disorders.

(a)
Identification. A conditioning tool for eating disorders is a prescription device that non-invasively measures the mass of food eaten during a meal and provides feedback in the form of eating rate, patient satiety, and eating pattern information to the patient.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Nonclinical performance testing must demonstrate:
(i) Device measurement accuracy and repeatability; and
(ii) Device feedback accuracy.
(2) Software verification, validation, and hazard analysis must be performed.
(3) The patient-contacting components of the device must be demonstrated to be biocompatible.
(4) Performance testing must demonstrate the electromagnetic compatibility (EMC) and electrical safety of the device.
(5) Labeling and patient labeling must be provided which includes the following:
(i) Information identifying and explaining how to use the device and its components; and
(ii) Information on how the device operates and the typical course of treatment.