510(k) Data Aggregation

    K Number
    K183401

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-03-01

    (84 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vis-U-All Low Temperature Sterilization Pouches/Tubing are sterilization containment pouches for use by health care providers to enclose:

    • medical devices in a single or double pouch configuration
    • trays containing medical devices in a single or double pouch configuration
    • small items requiring surface sterilization in a single pouch configuration within a tray
      NOTE: Trays must be legally marketed for use in the V-PRO Low Temperature or STERRAD Sterilization Systems and contain a vent surface area to tray volume ratio ≥ 0.135 in-1 with the maximum number of instrument organizers installed.

    and to be sterilized in the:

    • Lumen, Non Lumen, Flexible, Fast Non Lumen and Fast Cycles of the V-PRO® Low Temperature Sterilization Systems
    • Default Cycle of the STERRAD 100S Sterilizer
    • Standard and Advanced Cycles of the STERRAD NX and NX with ALLClear Technology Sterilizers
    • Express, Standard, Flex Scope, and DUO Cycles of the STERRAD 100NX with ALLClear Technology Sterilizers

    The pouches maintain the sterility of the enclosed devices until used.

    When used to enclose medical devices, the pouches are intended to contain the devices in such a manner as to leave a minimum of one inch between the devices and seal on all sides. When used to enclose a tray, the tray must fit loosely within the pouch.

    Device Description

    The proposed Vis-U-All Low Temperature Sterilization Pouches/Tubing is identical to the predicate and is a Tyvek/plastic film sterilization containment pouch designed for devices to be sterilized by the health care provider in V-PRO Low Temperature Sterilization Systems. As is the predicate device, the proposed device is available as a self seal pouch, a heat seal pouch, or heat seal tubing.

    AI/ML Overview

    The provided text describes the acceptance criteria and the results of non-clinical testing for the Vis-U-All Low Temperature Sterilization Pouches/Tubing. However, it does not contain information about a study involving AI, human readers, or ground truth established by experts/pathology/outcomes data, nor does it specify sample sizes for test or training sets in the context of an AI/algorithm device.

    This document is a 510(k) summary for a medical device (sterilization pouches) and the testing described is related to the physical and microbial barrier properties, as well as the effectiveness of sterilization and residual levels of sterilant, for the pouch itself. It is not an AI/algorithm-based device and therefore does not have the types of criteria or studies you've requested.


    Based on the provided text, here is the information that is available concerning the acceptance criteria and study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance (Conclusion)
    Effective Sterilant Penetration into Pouches (including pouched trays and, if applicable, pouches placed within a tray)Worst case test articles shall be reproducibly sterilized under worst case ½ cycle conditions for the: • STERRAD 100S Default Cycle • STERRAD NX Standard Cycle • STERRAD NX Advanced Cycle • STERRAD 100NX Standard Cycle • STERRAD 100NX Flex Scope Cycle • STERRAD 100NX Express Cycle • STERRAD 100NX Duo CyclePASS
    Pouch Integrity: Physical and Microbial Barrier Properties (Tensile Strength)Pouch material tensile strength will show no statistical difference between processed and unprocessed samples.PASS
    Pouch Integrity: Physical and Microbial Barrier Properties (Whole Package Integrity - Burst)Pouch burst strength will show no statistical difference between processed and unprocessed pouches.PASS
    Pouch Integrity: Physical and Microbial Barrier Properties (Seal Strength)Pouch seal strength will show no statistical difference between processed and unprocessed pouches.PASS
    Pouch Integrity: Physical and Microbial Barrier Properties (Microbial Retention)Tyvek microbial retention will show no statistical difference between processed and unprocessed pouches.PASS
    Maintenance of Package IntegrityPackaged instruments shall remain sterile through event-related and real-time studies.PASS
    Aeration: Hydrogen Peroxide ResidualsHydrogen peroxide residuals on the pouch will be reduced to acceptable levels.PASS

    Information Not Applicable/Not Provided in the Document:

    The product mentioned in the document (Vis-U-All Low Temperature Sterilization Pouches/Tubing) is a physical medical device (sterilization pouches/tubing), not an AI/algorithm or diagnostic imaging device. Therefore, the following requested information points are not applicable to the context of this document:

    1. Sample sizes used for the test set and the data provenance: Not applicable. This refers to test sets for algorithmic performance, not physical product testing.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI/algorithm.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K183297

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-01-03

    (37 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vis-U-All Low Temperature Sterilization Pouches/Tubing are sterilization containment pouches for use by health care providers to enclose:

    • · medical devices in a single or double pouch configuration
    • · trays* containing medical devices in a single or double pouch configuration
    • · small items requiring surface sterilization in a single pouch configuration within a tray*
      to be sterilized in the Lumen, Non Lumen and Fast Cycles of the V-PRO® Low Temperature Sterilization Systems. The pouches maintain the sterility of the enclosed devices until used.

    When used to enclose medical devices, the pouches are intended to contain the devices in such a manner as to leave a minimum of one inch between the devices and seal on all sides. When used to enclose a tray, the tray must fit loosely within the pouch.

    Device Description

    The proposed Vis-U-All Low Temperature Sterilization Pouches/Tubing is identical to the predicate and is a Tyvek/plastic film sterilization containment pouch designed for devices to be sterilized by the health care provider in V-PRO Low Temperature Sterilization Systems. As is the predicate device, the proposed device is available as a self seal pouch, a heat seal pouch, or heat seal tubing.

    AI/ML Overview

    The acceptance criteria and study proving device performance are described below:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Effective Sterilant Penetration into Pouches:
    V-PRO s2 Sterilizer Fast CycleWorst case stainless steel lumen test articles shall be reproducibly sterilized under worst case 1x10^-2 cycle conditions for the Fast Cycle.PASS
    Diffusion-restricted spaces shall be reproducibly sterilized under worst case 1x10^-2 cycle conditions for Fast Cycle.PASS
    Contact points between medical devices and tray accessories shall be reproducibly sterilized under worst case 1x10^-2 cycle conditions for the Fast Cycle.PASS
    Worst case surface sterilization test articles shall be reproducibly sterilized under worst case 1x10^-2 cycle conditions for the Fast Cycle.PASS
    V-PRO 60 & s2 Sterilizer Non Lumen Cycle, diffusion-restricted spaces claimsDiffusion-restricted spaces shall be reproducibly sterilized under worst case 1x10^-2 cycle conditions for the Non Lumen Cycle.PASS
    V-PRO 60 & s2 Sterilizer Flexible Cycle, diffusion-restricted spaces claimsDiffusion-restricted spaces shall be reproducibly sterilized under worst case 1x10^-2 cycle conditions for the Flexible Cycle.PASS
    V-PRO 60 & s2 Sterilizer Flexible Cycle extended claimsStainless steel test articles shall be reproducibly sterilized under worst case 1x10^-2 cycle conditions for the Flexible Cycle.PASS
    Pouch Integrity: Physical and Microbial Barrier Properties
    Tensile StrengthPouch material tensile strength will show no statistical difference between processed and unprocessed samples.PASS
    Whole Package Integrity (Burst)Pouch burst strength will show no statistical difference between processed and unprocessed pouches.PASS
    Seal StrengthPouch seal strength will show no statistical difference between processed and unprocessed pouches.PASS
    Microbial RetentionTyvek microbial retention will show no statistical difference between processed and unprocessed pouches.PASS
    Maintenance of Package IntegrityPackaged instruments shall remain sterile through event related and real time studies.PASS
    Aeration: Hydrogen Peroxide ResidualsHydrogen peroxide residuals on the pouch will be reduced to acceptable levels for dermal contact in Fast Non Lumen Cycle with 3 minute aeration.PASS

    Study Proving Device Meets Acceptance Criteria:

    The document describes a submission (K183297) to add claims for the Vis-U-All Low Temperature Sterilization Pouches/Tubing, including use with specific lumen configurations and diffusion-restricted spaces in various V-PRO sterilizer cycles (V-PRO 60, V-PRO s2, V-PRO 1, 1 Plus, maX & maX 2). The study is non-clinical performance testing.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the numerical sample sizes used for each specific test (e.g., number of pouches, lumens, or test articles). It refers to "worst case" scenarios for sterilant penetration tests and "samples" for physical property tests.

    The data provenance is not explicitly mentioned as retrospective or prospective, nor the country of origin. However, given the context of a 510(k) submission to the U.S. FDA, it is highly likely that the testing was conducted in a controlled laboratory environment to support regulatory approval in the U.S.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This is a performance study for sterilization efficacy and package integrity. The "ground truth" for the test set is established through scientific and engineering principles of sterilization and package testing, rather than human expert interpretation of images or physiological data. Therefore, the concept of "experts" to establish ground truth in the same way as, for example, radiologists for image analysis, does not apply. The tests described (e.g., microbial retention, tensile strength) rely on objective measurements and established standards.

    4. Adjudication Method for the Test Set:

    Not applicable in the context of this type of performance study. Adjudication methods (like 2+1 or 3+1) are typically used for subjective clinical assessments where multiple human readers interpret data to reach a consensus ground truth. Here, the outcomes are objective measurements of sterilization and physical properties.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No. This is not a study involving human readers or AI assistance. It is a performance study of a medical device (sterilization pouches).

    6. Standalone (Algorithm Only Without Human-in-the Loop Performance) Study:

    Not applicable. This device is a physical product (sterilization pouches), not an AI algorithm.

    7. Type of Ground Truth Used:

    The ground truth used for these tests is based on objective, quantifiable measures related to sterilization efficacy and physical properties, typically defined by:

    • Microbial Sterilization: Defined by the absence of microbial growth after processing, often using sterility assurance levels (SAL) based on biological indicator (BI) challenge testing. The "worst case 1x10^-2 cycle conditions" refers to a reduced cycle used for BI testing to demonstrate appropriate sterilant penetration and microbial kill.
    • Physical and Barrier Properties: Defined by adherence to established engineering standards for tensile strength, burst strength, seal strength, and microbial retention, often with statistical comparisons between processed and unprocessed samples.
    • Chemical Residuals: Defined by laboratory analysis to ensure residual levels of hydrogen peroxide are below acceptable safety thresholds.

    8. Sample Size for the Training Set:

    Not applicable. This is a physical medical device, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172749

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-02-09

    (150 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vis-U-All Low Temperature Sterilization Pouches/Tubing are sterilization containment pouches for use by health care providers to enclose:

    • . medical devices in a single or double pouch configuration
    • trays* containing medical devices in a single or double pouch configuration
    • . small items requiring surface sterilization in a single pouch configuration within a tray*
      to be sterilized in the Lumen, Non Lumen, Flexible and Fast Non Lumen Cycles of the V-PRO Low Temperature Sterilization Systems. The pouches are intended to maintain the sterility of the enclosed devices until used.

    When used to enclose medical devices, the pouches are intended to contain the devices in such a manner as to leave a minimum of one inch between the devices and seal on all sides. When used to enclose a tray, the tray must fit loosely within the pouch.

    Device Description

    The proposed Vis-U-All Low Temperature Sterilization Pouches/Tubing is identical to the predicate and is a Tyvek/plastic film sterilization containment pouch designed for devices to be sterilized by the health care provider in V-PRO Low Temperature Sterilization Systems. As is the predicate device, the proposed device is available as a self seal pouch, a heat seal pouch, or heat seal tubing.

    The purpose of K172749 is to add claims for:

    • use in the V-PRO maX 2 Fast Non Lumen Cycle
    • the sterilization of titanium mated surfaces (diffusion-restricted spaces) in ● the V-PRO 1 Plus, maX, and maX 2 Sterilizers Non Lumen Cycle
    • addition of the V-PRO maX 2 Sterilizer to the indications for use in the Lumen, Non Lumen and Flexible Cycles of the V-PRO 1, 1 Plus and maX Sterilizers (the V-PRO maX 2 Sterilizer Lumen, Non Lumen, and Flexible Cycles are identical, with the exception of the condition phase of the Lumen Cycle, to the previously cleared cycles listed in Table 2-1).
    AI/ML Overview

    The provided document describes the acceptance criteria and study results for STERIS Vis-U-All Low Temperature Sterilization Pouches/Tubing (K172749).

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Effective Sterilant Penetration into Pouches:
    - V-PRO maX 2 Sterilizer Fast Non Lumen CycleWorst case test article shall be reproducibly sterilized under worst case abbreviated ½ cycle conditions for Fast Non Lumen Cycle.PASS
    - V-PRO 1 Plus, maX, and maX 2 Sterilizer Non Lumen Cycle, diffusion restricted titanium claims (first entry)Diffusion-restricted spaces (composed of stainless steel and titanium, as well as contact points between medical devices and tray accessories) shall be reproducibly sterilized under worst case abbreviated ½ cycle conditions for Fast Non Lumen Cycle.PASS
    - V-PRO 1 Plus, maX, and maX 2 Sterilizer Non Lumen Cycle, diffusion restricted titanium claims (second entry)Diffusion-restricted spaces (titanium) shall be reproducibly sterilized under worst case ½ cycle conditions for the Non Lumen Cycle.PASS
    Pouch Integrity: Physical and Microbial Barrier Properties:
    - Tensile StrengthPouch material tensile strength will show no statistical difference between processed and unprocessed samples.PASS
    - Whole Package Integrity (Burst)Pouch burst strength will show no statistical difference between processed and unprocessed pouches.PASS
    - Seal StrengthPouch seal strength will show no statistical difference between processed and unprocessed pouches.PASS
    - Microbial RetentionTyvek microbial retention will show no statistical difference between processed and unprocessed pouches.PASS
    Maintenance of Package IntegrityPackaged instruments shall remain sterile through event related and real time studies.PASS
    Aeration: Hydrogen Peroxide ResidualsHydrogen peroxide residuals on the pouch will be reduced to acceptable levels for dermal contact in Fast Non Lumen Cycle with 3 minute aeration.PASS
    CytotoxicityPouch materials shall be non-cytotoxic following worst case exposure in a V-PRO Sterilizer for Fast Non Lumen Cycle with 3 minute aeration.PASS

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not explicitly state the specific sample sizes used for each test. It refers to "worst case test article" and "processed and unprocessed samples/pouches" without providing numerical details for the test sets.

    The data provenance is not specified in terms of country of origin or whether it's retrospective or prospective. Given the nature of a 510(k) submission, these are likely laboratory-based studies conducted by the manufacturer to demonstrate product performance for regulatory approval.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable and not provided in the document. The tests performed are objective physical, chemical, and microbiological evaluations of the sterilization pouches, not interpretations of medical images or patient data that would require medical experts to establish ground truth.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies involving interpretation of data by multiple human readers, not for objective laboratory testing of device performance.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The document focuses on the performance of the sterilization pouches themselves through laboratory tests, not on human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, this is not applicable. The device is a physical sterilization pouch, not an algorithm or AI system. The tests are focused on its physical and functional capabilities.

    7. The Type of Ground Truth Used

    The "ground truth" for these tests is based on established scientific principles and regulatory standards for sterilization and material performance. For example:

    • Sterilant Penetration: Ground truth is complete sterilization, typically demonstrated by the absence of microbial growth in biological indicators after exposure to the sterilization process. The ½ cycle conditions are a recognized method to challenge sterilization efficacy.
    • Pouch Integrity (Tensile, Burst, Seal Strength): Ground truth is maintaining pre-determined physical properties within acceptable ranges after processing, ensuring the pouch remains intact.
    • Microbial Retention: Ground truth is the ability of the material to prevent microbial ingress.
    • Hydrogen Peroxide Residuals: Ground truth is residuals falling below acceptable safety limits.
    • Cytotoxicity: Ground truth is the material being non-toxic to cells.

    These are objective, empirical measures determined by laboratory testing, not by expert consensus, pathology, or outcomes data in a medical context.

    8. The Sample Size for the Training Set

    This information is not applicable. The device is a physical product and does not involve AI or machine learning, which would require a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160908

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-07-22

    (112 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vis-U-All Low Temperature Sterilization Pouches/Tubing are sterilization containment pouches for use by health care providers to enclose:

    • medical devices in a single or double pouch configuration ●
    • trays* containing medical devices in a single or double pouch configuration
    • . small items requiring surface sterilization in a single pouch configuration within a tray*
      to be sterilized in the Lumen. Non Lumen and Flexible Cycles of the V-PRO Low Temperature Sterilization Systems. The pouches maintain the sterility of the enclosed devices until used.

    When used to enclose medical devices, the pouches are intended to contain the devices in such a manner as to leave a minimum of one inch between the devices and seal on all sides. When used to enclose a tray, the tray must fit loosely within the pouch.

    Device Description

    The proposed Vis-U-All Low Temperature Sterilization Pouch/Tubing is identical to the predicate and is a Tyvek/plastic film sterilization containment pouch designed for devices to be sterilized by the health care provider in V-PRO Low Temperature Sterilization Systems. As is the predicate, the proposed device is available as a self seal pouch, a heat seal pouch, or heat seal tubing.

    The purpose of this submission is to add claims for: use in the V-PRO maX Flexible cycle; pouching of sterilization trays and for sterilization of pouched items within a sterilization tray.

    AI/ML Overview

    While the document provided details about the device's intended use and the tests performed, it does not contain the kind of detailed information about acceptance criteria and study design typically associated with a medical AI/ML device as outlined in the prompt. The device, "Vis-U-All Low Temperature Sterilization Pouch/Tubing," is a physical sterilization containment pouch, not an AI/ML device.

    Therefore, the following information cannot be extracted from the provided text as it pertains to AI/ML device evaluation, which is not applicable here:

    • Sample size used for the test set and data provenance
    • Number of experts used and their qualifications
    • Adjudication method
    • Multi-reader multi-case (MRMC) comparative effectiveness study
    • Standalone (algorithm only) performance
    • Type of ground truth used (expert consensus, pathology, outcomes data)
    • Sample size for the training set
    • How the ground truth for the training set was established

    However, I can provide the available information regarding the device's acceptance criteria and reported performance from the document.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Effective Sterilant PenetrationWorst case test article shall be reproducibly sterilized under worst case ½ cycle conditions.PASS
    Package Integrity: Event Related Package Integrity Pouching a Qualified TrayNo visual damage to pouch material (no holes or tears). Seals remain intact.PASS

    2-9. Information not applicable to this device type.

    The provided document describes a 510(k) submission for a sterilization pouch, which is a physical medical device. The evaluation criteria are therefore focused on physical performance, sterility maintenance, and material integrity, rather than AI/ML performance metrics like sensitivity, specificity, reader studies, or training/test set provenance common in the context of AI/ML medical devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140487

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-07-22

    (145 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vis-U-All Low Temperature Sterilization Pouches/Tubing are sterilization containment pouches for use by health care providers to enclose medical devices to be sterilized in Lumen and Non Lumen Cycles in the V-PRO 60 Low Temperature Sterilization System. The pouches are intend to maintain the sterility of the enclosed devices until used.

    Pouches are intended to contain devices in such a manner as to leave a minimum of one inch between the devices and seal on all sides. Devices with stainless steel lumens and the following minimum internal diameter (ID) and maximum length can be processed in Vis-U-All Low Temperature Sterilization Pouch/Tubing using the Lumen Cycle.

    • o single or dual lumen devices
      • 0.77 mm ID and 410 mm in length I
    • o triple lumen devices
      • 1.2 mm ID and 275 mm in length D
      • . 1.8 mm ID and 310 mm in length

    or

    • 트 2.8 mm ID and 317 mm in length
    Device Description

    The proposed Vis-U-All Low Temperature Sterilization Pouch/Tubing is a Tyvek/plastic film sterilization containment pouch designed for devices to be sterilized by the health care provider in the V-PRO 60 Low Temperature Sterilization System. The proposed device is available as a self seal pouch, a heat seal pouch, or heat seal tubing.

    AI/ML Overview

    The STERIS® Vis-U-All Low Temperature Sterilization Pouch/Tubing underwent several tests to demonstrate its safety and substantial equivalence to a predicate device. The information provided outlines the acceptance criteria and a summary of the test results for each.

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance
    Effective Sterilant PenetrationWorst case test article shall be reproducibly sterilized under worst case ½ cycle conditions.PASS
    Tensile StrengthPouch material tensile strength will show no statistical difference between processed and unprocessed samples.PASS
    Whole Package Integrity (Burst)Pouch burst strength will show no statistical difference between processed and unprocessed pouches.PASS
    Seal StrengthPouch seal strength will show no statistical difference between processed and unprocessed pouches.PASS
    Microbial RetentionTyvek microbial retention will show no statistical difference between processed and unprocessed pouches.PASS
    Maintenance of Package IntegrityPackaged instruments shall remain sterile through event related and real time studies.PASS
    Aeration: Hydrogen Peroxide ResidualsHydrogen peroxide residuals on the pouch will be reduced to acceptable levels for dermal contact.PASS
    CytotoxicityPouch materials shall be non-cytotoxic following worst case exposure in a V-PRO 60 Sterilizer.PASS

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the sample sizes for each specific test or the data provenance (e.g., country of origin, retrospective or prospective) for the test set. It broadly mentions "Testing of the Vis-U-All Low Temperature Sterilization Pouch/Tubing" and provides summarized conclusions.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The tests appear to be laboratory-based performance evaluations with objective, measurable criteria, rather than those requiring expert human interpretation for ground truth establishment.

    4. Adjudication method for the test set:

    This information is not provided in the document. Given the nature of the tests (physical and microbial barrier properties, chemical residuals), it's unlikely that an adjudication method typically used for subjective assessments (like 2+1 or 3+1 for expert consensus) would be relevant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is relevant for evaluating the performance of AI-assisted diagnostic tools with human readers, which is not applicable to a sterilization pouch.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    The device is a physical sterilization pouch, not an algorithm. Therefore, "standalone" algorithm performance is not applicable. The tests performed are inherent to the physical and chemical properties of the pouch itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth for these tests is based on objective, quantifiable measurements and established scientific/engineering principles for material performance, microbial viability, and chemical safety. For example:

    • Effective Sterilant Penetration: Ground truth is achieving sterilization, likely verified by biological indicators or sterility testing methods.
    • Tensile Strength, Burst Strength, Seal Strength, Microbial Retention: Ground truth is determined by comparing measured values against statistical equivalence to unprocessed samples or defined material specifications.
    • Maintenance of Package Integrity: Ground truth is confirmed by maintaining sterility through various stress tests and real-time aging, likely using microbial challenge tests.
    • Hydrogen Peroxide Residuals: Ground truth is the reduction of residuals to acceptable, predefined safety levels.
    • Cytotoxicity: Ground truth is the absence of cytotoxic effects, determined by established biological evaluation methods (e.g., ISO 10993).

    8. The sample size for the training set:

    A "training set" is not applicable as this is a physical medical device (sterilization pouch) undergoing performance validation, not a machine learning algorithm.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for a machine learning algorithm in this context.