Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K993799

    Validate with FDA (Live)

    Date Cleared
    1999-12-10

    (31 days)

    Product Code
    Regulation Number
    870.2920
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CG-7000DX Recorder/Transmitter is a 12 Lead ambulatory electrocardiograph capable of recording and transmitting up to 40 standard ECGs for the purpose of cardiac monitoring and diagnosis, which incorporates a recording and transmitting circuitry, graphic LCD, a package of firmware tools and is intended for use by a medical professional:
    a. The ECG is recorded and transmitted to a remote receiving station for consultation with a cardiologist.
    b. The ECG is recorded and transferred to a PC for viewing and processing.

    Device Description

    CG-7000DX Recorder/Transmitter is a 12 Lead ambulatory electrocardiograph capable of recording and transmitting up to 40 standard ECGs for the purpose of cardiac monitoring and diagnosis, which incorporates a recording and transmitting circuitry, graphic LCD, a package of firmware tools and is intended for use by a medical professional.

    AI/ML Overview

    The provided text describes the CG-7000DX Recorder/Transmitter, a 12-lead ambulatory electrocardiograph. However, the document is a 510(k) Summary of Safety and Effectiveness for a medical device and does not contain a study that proves the device meets specific acceptance criteria in the typical format of a clinical or performance study report.

    Instead, the document focuses on demonstrating substantial equivalence to a predicate device (CG-7000D) and lists various verification, validation, and environmental tests conducted as part of design controls. These tests ensure the device's compliance with design specifications and relevant standards, but they are not presented as a study with reported performance against quantitative acceptance criteria in the way a clinical trial or performance study would be.

    Therefore, many of the requested fields cannot be directly extracted from the provided text. I will fill in what can be inferred and indicate where information is not present.


    Description of Acceptance Criteria and Device Performance

    The CG-7000DX Recorder/Transmitter is indicated for recording and transmitting up to 40 standard 12-lead ECGs for cardiac monitoring and diagnosis. The primary "acceptance criterion" for this 510(k) submission is substantial equivalence to its predicate device, the CG-7000D ECG Recorder/Transmitter (K942704). The document attests that the CG-7000DX is as safe, effective, and performs as well as or better than the predicate device.

    The acceptance criteria are implicitly defined by compliance with the following standards and successful completion of the listed verification, validation, and environmental tests:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Inferred from Standards & Tests)Reported Device Performance (as stated in the document)
    Standards Compliance:
    ANSI/AAMI EC38, "Ambulatory Electrocardiographs" 1994Met
    ANSI/AAMI EC13 Cardiac Monitors, Heart Rate Meters and Alarms, 200 edition 1992Met
    ANSI/AAMI EC11 Diagnostic Electrocardiographic Devices, 2nd edition 1991Met
    IEC-601-1-4 Medical Electrical Equipment 1996Met
    Design Control Tests:
    SW Unit/Module Testing (19 tests)Conducted, device compliant with design specifications
    User Interface Verification TestConducted, device compliant with design specifications
    Reference Voltage Verification TestConducted, device compliant with design specifications
    Clock Oscillator Verification TestConducted, device compliant with design specifications
    Flash Memory Verification TestConducted, device compliant with design specifications
    LPF Verification TestConducted, device compliant with design specifications
    QRS Detection TestConducted, device compliant with design specifications
    PM Pulse Detection TestConducted, device compliant with design specifications
    Patient Cable TestConducted, device compliant with design specifications
    Common Mode Rejection TestConducted, device compliant with design specifications
    Frequency Response TestConducted, device compliant with design specifications
    Input Dynamic Range TestConducted, device compliant with design specifications
    Overall System Error TestConducted, device compliant with design specifications
    Step Response TestConducted, device compliant with design specifications
    System Noise TestConducted, device compliant with design specifications
    Safe Current TestConducted, device compliant with design specifications
    High and Low Temperature and Humidity TestConducted, device compliant with design specifications
    Surface Temperature TestConducted, device compliant with design specifications
    Leakage Current TestConducted, device compliant with design specifications
    Dielectric Strength TestConducted, device compliant with design specifications
    Mechanical Vibration Shock TestConducted, device compliant with design specifications
    Ingress of Liquids TestConducted, device compliant with design specifications
    EMC TestConducted, device compliant with design specifications
    Biocompatibility EvaluationSatisfactory
    System Safety and Risk AnalysisConducted, rigorous design and structural evaluation
    Level of ConcernCharacterized as a moderate level of concern system
    Clinical Performance:"As safe, as effective and performs as well as or better than the legally marketed predicate device, CG-7000D ECG Recorder/Transmitter (K942704)."

    Study Details:

    1. Sample size used for the test set and the data provenance:

      • The document mentions "laboratory testing" and "clinical and laboratory testing" but does not specify sample sizes for any test sets related to performance evaluation (e.g., number of ECGs, number of patients, etc.).
      • The provenance of any data (e.g., country of origin, retrospective/prospective) is not mentioned.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable/Not mentioned. The document describes engineering/design control tests and claims substantial equivalence, but it does not detail a study involving expert-established ground truth for diagnostic performance. The intended use includes "consultation with a cardiologist" but this refers to the use of the device, not the validation of its performance in a clinical study as described.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable/Not mentioned. As detailed above, there is no described study requiring an adjudication method.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No such study was done or reported. The device is an ECG recorder/transmitter, not an AI-assisted diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable in the context of diagnostic AI. The device itself has internal algorithms (e.g., QRS detection, PM Pulse Detection), which were tested as part of the "SW Unit/Module Testing" and other verification tests. However, these are part of the device's basic functionality, not a "standalone performance" study in the sense of an independent diagnostic algorithm. The device's primary function is data acquisition and transmission, not automated diagnosis.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Not explicitly stated for diagnostic accuracy. For the functional tests (e.g., QRS detection, frequency response), the "ground truth" would be the expected technical specifications and performance characteristics, typically established by design requirements and engineering standards. For the overall claim of "safe and effective," the ground truth is established by compliance with recognized standards and successful completion of the described verification and validation tests, implying that the device accurately records and transmits ECG signals as per its design.
    7. The sample size for the training set:

      • Not applicable/Not mentioned. This information would typically be relevant for machine learning-based devices. The CG-7000DX is a hardware device with firmware, not a learning algorithm product.
    8. How the ground truth for the training set was established:

      • Not applicable/Not mentioned. (See point 7).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1