Found 2 results

510(k) Data Aggregation

    K Number
    K161674

    Validate with FDA (Live)

    Device Name
    QuantumCam
    Manufacturer
    Date Cleared
    2016-10-20

    (126 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    21 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QuantumCam is a gamma camera system designed to acquire data for whole body static, gated or dynamic and multislice images. The system is intended for use as diagnostic imaging device. When used with appropriate radio pharmaceuticals, images are produced representing the internal distribution of radioactivity in head or body. The system allows you to acquire data for high resolution three dimensional, static, gated or dynamic images of biochemical and metabolic processes using Tc-99m, TI-201, I-123, I-131, In-111, Ga-67, Co-57.

    Device Description

    The QuantumCam is a general purpose dual detector gamma camera system comprised of a mechanical gantry allowing the Detectors to be positioned and moved in close proximity to the patient for scanning. The Gantry provides for positioning of the detectors in location(s) suitable for patient to be brought in position for tomographic, whole body scanning as well as planar scanning in sitting or standing position.

    The Detectors are standard Nal (Sodium Iodide) / Photomultiplier based scintillation detectors designed following the Anger Camera principles first described by Hal Oscar Anger. This detector technology has been used effectively in Nuclear Medicine for decades. The Detectors are equipped with standard Collimators following the principles known to the industry and this submission is specific about the HEGP collimator. The image data from the detectors are collected by hardware in the detector and the data acquisition is controlled by software running on a standard personal computer with a suitable operating system. The data may subsequently be transferred to a Nuclear Medicine Workstation for processing and interpretation. This Workstation is not part of the QuantumCam system.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the QuantumCam 9SYS2070-B02, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily based on NEMA (National Electrical Manufacturers Association) tests for gamma camera systems. The performance reported is for the new High Energy General Purpose (HEGP) collimator, which is the main addition that expands the device's energy range.

    Feature / Performance MetricAcceptance Criteria (Specification)Reported Device Performance (NEMA Test Result)
    HEGP Collimator Hole Properties:
    ShapeHEXHEX (Measured and compared to Unicorn drawings)
    Size3.0 mm3.0 mm (Measured and compared to Unicorn drawings)
    Septa1.75 mm1.75 mm (Measured and compared to Unicorn drawings)
    Length43 mm43 mm (Measured and compared to Unicorn drawings)
    Device Performance with HEGP Collimator (NEMA Tests):
    Sensitivity183 cpm/uCiNot explicitly stated as a pass/fail. The value 183 cpm/uCi is given but not compared against a specific pass/fail threshold.
    Spatial Resolution @ 10cm, HEGP Collimator< 16.5 mm (less than 16.5 mm)Not explicitly stated whether the result meets the criteria. The value is not provided.
    Septal Penetration (Calculated)39.8 %Not explicitly stated whether the result meets the criteria. The value 39.8% is given, with the note that the official specification is based on septa thickness and
    calculated based on lead absorption (4.8%). This suggests 39.8% is the calculated value, not a measured one, and 4.8% might be the acceptable absorption.
    Detector Shielding (I-131 Source (364 keV)):
    LiNot specified6.4%
    LFiNot specified1.4%
    LSiNot specified20%

    Important Note: The document focuses on demonstrating substantial equivalence to a predicate device. The "acceptance criteria" presented are primarily the physical specifications of the HEGP collimator and results from standard NEMA tests. However, for "Spatial Resolution @ 10cm" and "Septal Penetration", the actual test result that directly compares to the specified criteria is not explicitly provided in the table. The statements indicate that these were NEMA tested, but not the explicit numerical outcome for direct comparison to the "less than 16.5 mm" or the 39.8% calculated value (which seems to be the expected performance rather than an acceptance threshold).

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated. The document mentions "NEMA test on a QuantumCam system" and measurements of collimator hole properties. This suggests a single or very small number of physical devices/collimators were tested.
    • Data Provenance: The tests were conducted internally by DDD-Diagnostic A/S as part of their quality system procedures. The country of origin for the data is implicitly Denmark, where DDD-Diagnostic A/S is located. The study is retrospective in the sense that it's a verification of a product design feature rather than a clinical trial with patients over time.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not explicitly stated. The document mentions "conducted and reported by qualified experts." No specific number or qualifications are provided beyond "qualified experts."
    • Qualifications: "Qualified experts" is the only description given. No specific details like "radiologist with 10 years of experience" are provided.

    4. Adjudication Method for the Test Set

    • The document does not describe any adjudication method. The testing appears to be objective, physical measurements and NEMA standard tests, rather than subjective interpretation requiring adjudication among multiple readers.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    • No, an MRMC comparative effectiveness study was not done. The submission is for a device modification (expanded energy range with a new collimator) based on substantial equivalence, not a clinical effectiveness study to compare diagnostic accuracy or reader performance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Yes, the tests described are standalone performance tests of the gamma camera system itself (e.g., sensitivity, spatial resolution, septal penetration), without human involvement in interpreting images or making diagnoses. This is an objective measurement of the device's physical imaging capabilities.

    7. The Type of Ground Truth Used

    • The ground truth for the device performance tests (sensitivity, spatial resolution, septal penetration) is based on NEMA standards and physical measurements. For the collimator specifications (shape, size, septa, length), the ground truth is the Unicorn drawings (presumably engineering or design drawings) that define the intended specifications, against which the manufactured collimator was measured.

    8. The Sample Size for the Training Set

    • Not applicable. This device is hardware (gamma camera system), not a machine learning or AI algorithm that requires a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no training set for this hardware device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140206

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-05-08

    (101 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the QuantumCam gamma camera system is to perform general nuclear medicine imaging procedures. This is intended to be accomplished by imaging the distribution of a radiopharmaceutical within the human body. Gamma rays emitted from the radiopharmaceutical are detected by the gamma camera system and formed into images characterizing and showing the state of organs or structures in the form of functional images. When QuantumCam is connected to a nuclear medicine workstation, these images can be used in concert with other clinical data to assist in making clinical diagnoses by authorized medical personnel.

    Indications for use.
    The QuantumCam is a gamma camera system designed to acquire data for whole body static, gated or dynamic and multi-slice images. The system is intended for use as diagnostic imaging device. When used with appropriate radio pharmaceuticals, images are produced representing the internal distribution of radioactivity in head or body. The system allows you to acquire data for high resolution three dimensional, static, gated or dynamic images of biochemical and metabolic processes using Tc-99m, TI-201, I-123, In-111, Ga-67, Co-57.

    Device Description

    The QuantumCam is a general purpose dual detector gamma camera system comprised of a mechanical gantry allowing the Detectors to be positioned and moved in close proximity to the patient for scanning. The Gantry provides for positioning of the detectors in location(s) suitable for patient to be brought in position for tomographic, whole body scanning as well as planar scanning in sitting or standing position.

    Functional description:
    The Detectors are standard Nal (Sodium lodide) / Photomultiplier based scintillation detectors designed following the Anger Camera principles first described by Hal Oscar Anger. This detector technology has been used effectively in Nuclear Medicine for decades. The Detectors are equipped with standard Collimators following the principles known to the industry. The image data from the detectors are collected by hardware in the detector and the data acquisition is controlled by software running on a standard personal computer with a suitable operating system. The data may subsequently be transferred to a Nuclear Medicine Workstation for processing and interpretation. This Workstation is not part of the QuantumCam system.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study details for the QuantumCam device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    SpecificationAcceptance Criteria (NEMA Standard NU 1:2007)Reported Device Performance (from NEMA testing)
    Intrinsic Spatial Resolution, FWHM, UFOV @ surface≤ ± 3.9 mmNot explicitly stated as a single value, but compliance with NEMA standard implies meeting this.
    Spatial Resolution, FWHM, LEGP collimator @ 10cm, Tc-99m< 9.1 mmNot explicitly stated as a single value, but compliance with NEMA standard implies meeting this.
    Energy Resolution, @ Tc-99m≤ 9.6 %Not explicitly stated as a single value, but compliance with NEMA standard implies meeting this.
    Spatial Linearity, UFOV, Diff.< ± 0.2 mmNot explicitly stated as a single value, but compliance with NEMA standard implies meeting this.
    Flood Field Uniformity, UFOV Intrinsic, Diff.< ± 2.55 %Not explicitly stated as a single value, but compliance with NEMA standard implies meeting this.
    Maximum Count rate with scatter> 320 kcpsNot explicitly stated as a single value, but compliance with NEMA standard implies meeting this.
    Count rate Sensitivity, @ 20% loss> 225 kcpsNot explicitly stated as a single value, but compliance with NEMA standard implies meeting this.
    Detector shielding Sensitivity, @ 184 keV< 2.0 %Not explicitly stated as a single value, but compliance with NEMA standard implies meeting this.

    Important Note: The document states that the "test equipment used, test setup and all calculations have all been performed according to the NEMA Standard NU 1:2007 in a 20% energy." This implies that the device met the specifications outlined in the NEMA standard, even if explicit performance values are not listed next to each criterion. For the detector shielding sensitivity, it is specifically mentioned that the test was performed according to NEMA Standard NU 1:2007.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document describes non-clinical testing conducted according to NEMA Standard NU 1:2007. This standard outlines procedures for measuring the performance of gamma cameras.

    • Sample Size: The document does not specify a "sample size" in terms of patient data or clinical cases. Instead, it refers to testing of the physical device components (detectors, collimators) using standard phantoms and calibrated sources as per NEMA guidelines. Therefore, there isn't a "test set" in the traditional sense of patient data.
    • Data Provenance: The testing was "performed according to DDD quality system procedures as investigations performed according to standards... conducted and reported by qualified experts." This indicates the testing was done by the manufacturer (DDD-Diagnostic A/S) in Denmark. It is a prospective test in the sense that the device was actively tested against the NEMA standard.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Since this was a non-clinical, performance-based test against a technical standard (NEMA NU 1:2007), there was no "ground truth" established by human experts in a diagnostic sense (e.g., radiologists interpreting images). The "ground truth" was the objective measurement values obtained by following the NEMA standard, which are then compared to the specified acceptance criteria of the standard itself.

    The document mentions "qualified experts" conducted and reported the investigations, but their specific number or qualifications (e.g., "radiologist with 10 years of experience") are not provided. These "experts" would likely be engineers or physicists specializing in nuclear medicine imaging system performance.

    4. Adjudication Method for the Test Set

    Not applicable. As described above, this was a non-clinical, objective performance test against a standard, not a subjective interpretation task requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The document focuses on demonstrating the device's technical performance and substantial equivalence to a predicate device, not on how human readers perform with or without AI assistance. The QuantumCam is a gamma camera system, not an AI-assisted interpretation tool.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. The QuantumCam is a physical gamma camera system (hardware) that acquires data. It is not an algorithm that provides standalone diagnostic interpretations. While it uses software for acquisition and control, the "performance" described is about the quality of the raw data acquisition, not an automated diagnostic output from an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this study was the objective measurements obtained by following the specific protocols outlined in the NEMA Standard NU 1:2007. This standard provides a framework for quantitatively assessing the physical performance characteristics of gamma cameras. The device's performance was compared directly to the numerical acceptance criteria defined by this standard.

    8. The Sample Size for the Training Set

    Not applicable. This is a medical device (gamma camera) performance study, not an AI algorithm development study. Therefore, there is no "training set" in the context of machine learning.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set mentioned or implied by the document.