Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K041627

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-09-03

    (79 days)

    Product Code
    Regulation Number
    862.1225
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For In Vitro Diagnostic Use. For Professional Use Only.

    Cystatin C Immunoparticles are intended for the quantitative determination of cystatin C in human serum, heparinized plasma and EDTA plasma by turbidimetry and nephelometry. Cystatin C measurements are used as an aid in the diagnosis and treatment of renal diseases.

    Cystatin C Control Set is an assayed bi-level control intended to monitor and evaluate the precision and accuracy of the quantitative immunological determination of human cystatin C by turbidimetry or nephelometry.

    The Cystatin C Calibrator is intended for establishing calibration curves for the quantitative immunological determination of human cystatin C by turbidimetry or nephelometry.

    Device Description

    Cystatin C Immunoparticles are purified immunoglobulin fraction of rabbit antiserum directed against cystatin C covalently coupled to uniform polystyrene particles. The Cystatin C Immunoparticles are provided as a suspension preserved with 15 mmol/L sodium azide and are supplied ready for use. Recombinant human cystatin C produced in E. coli was used as an immunogen for raising the antibody coupled to the polystyrene particles.

    In the DakoCytomation Cystatin C Assay, human serum or plasma is mixed with the Cystatin C Immunoparticles. The resulting immune complexes are measured by turbidimetry or nephelometry. The generated signal is correlated with the concentration of cystatin C in the sample. By interpolation on a standard curve, the concentration of cystatin C in the sample is calculated.

    Cystatin C Control Set is a bi-level control. The controls are liquid pools of delipidated human serum enriched with recombinant human cystatin C produced in E. coli and preserved with 15 mmol/L sodium azide. Each donor has been tested and found negative for hepatitis B virus surface antigen, antibodies to hepatitis C virus and human immunodeficiency virus 1 and 2. The Cystatin C Control Set is supplied ready for use.

    The cystatin C value assignment has been carried out by turbidimetry using the DakoCytomation Cystatin C Calibrator, code No. X 0974, as reference.

    Cystatin C Calibrator is a liquid pool of delipidated human serum enriched with recombinant human cystatin C produced in E. coli and preserved with 15 mmol/L sodium azide. Each donor has been tested and found negative for hepatitis B virus surface antigen, antibodies to hepatitis C virus and human immunodeficiency virus 1 and 2. The Cystatin C Calibrator is supplied ready for use.

    The cystatin C value assignment has been carried out by turbidimetry using a precise transfer protocol ensuring traceability to a pure recombinant human cystatin C reference preparation, where the cystatin C concentration was established by dry mass determination.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for three in vitro diagnostic devices: Cystatin C Immunoparticles, Cystatin C Control Set, and Cystatin C Calibrator. The focus of the submission is to demonstrate substantial equivalence to previously marketed predicate devices rather than proving a new performance characteristic against a fixed set of acceptance criteria. Therefore, the information regarding acceptance criteria and performance studies is presented differently than would be expected for a device with novel performance claims.

    Here's an analysis of the provided information:

    1. A table of acceptance criteria and the reported device performance

    The document does not present a formal table of distinct acceptance criteria with corresponding performance results similar to what might be found for a device with a specific diagnostic accuracy claim (e.g., sensitivity, specificity, AUC). Instead, the performance evaluations were conducted to demonstrate "substantial equivalence" to predicate devices. The performance characteristics studied include:

    Performance CharacteristicDescription / Reported Performance
    PrecisionStudied and found to contribute to substantial equivalence. Specific quantitative results (e.g., CV%) are not provided in this summary.
    AccuracyStudied and found to contribute to substantial equivalence. Specific quantitative results are not provided in this summary.
    LinearityStudied and found to contribute to substantial equivalence. Specific quantitative results are not provided in this summary.
    SensitivityStudied and found to contribute to substantial equivalence. Specific quantitative results are not provided in this summary.
    SpecificityStudied and found to contribute to substantial equivalence. Specific quantitative results are not provided in this summary.
    Security RangesStudied and found to contribute to substantial equivalence. Specific quantitative results are not provided in this summary.
    InterferencesStudied and found to contribute to substantial equivalence. Details on specific interferents and their effects are not provided.
    Stability TestingStudied and found to contribute to substantial equivalence. Details on study methods and results are not provided.
    Comparability (Turbidimetry/Nephelometry)Assay results obtained with various commercially available turbidimeters and the IMMAGE® Immunochemistry System (nephelometer) were analyzed and found to be comparable.
    Comparability to Predicate (Dade Behring)A comparison of performance demonstrated substantial equivalence to the Dade Behring, Inc. N Latex Cystatin C Test Kit (K003503) and N Protein Standard UY (K003501).
    Comparability to Predicate (Roche Diagnostics)A comparison of performance demonstrated substantial equivalence to the Roche Diagnostics Corp. Creatinine Plus assay (K953239).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for the test sets for precision, accuracy, linearity, etc. It mentions human serum or plasma samples being used. There is no information regarding the country of origin of the data or whether the studies were retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This type of information (expert review, ground truth establishment) is typically relevant for diagnostic imaging AI devices, rather than for an in vitro diagnostic (IVD) immunoassay like Cystatin C measurement. For IVD devices, the "ground truth" often refers to a reference method or validated standard, not expert human interpretation of an image. The document mentions value assignment for controls and calibrators by turbidimetry and traceability to a pure recombinant human cystatin C reference preparation where the concentration was established by dry mass determination. This constitutes the "ground truth" for the calibrators and controls.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Adjudication methods are typically associated with human reader studies for subjective interpretations (e.g., radiology). This concept does not apply to the performance testing of an automated IVD assay where results are quantitative measurements. The "adjudication" in this context would be the validation of the reference methods and calibrator value assignments.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. These are typically for AI-assisted diagnostic imaging where human readers interpret cases. This device is an in vitro diagnostic reagent and not an AI-assisted imaging device. The device itself is intended to quantitatively determine Cystatin C, and its performance is evaluated against predicate devices and analytical criteria, not human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the inherent performance of the assay itself. The performance characteristics listed (precision, accuracy, linearity, sensitivity, specificity, interferences, stability) are all "standalone" performance evaluations of the assay components (Immunoparticles, Control Set, Calibrator) as they would be run on an automated analyzer. The device's primary function is quantitative measurement, not providing AI assistance to a human interpreter.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the DakoCytomation Cystatin C Assay, the "ground truth" for the calibrators and controls was established by:

    • Turbidimetry using the DakoCytomation Cystatin C Calibrator (X 0974) as reference for the control set.
    • Traceability to a pure recombinant human cystatin C reference preparation, where the cystatin C concentration was established by dry mass determination for the calibrator.
    • For the overall performance, comparison to legally marketed predicate devices (Dade Behring N Latex Cystatin C Test Kit, Roche Diagnostics Corp. Creatinine Plus assay) served as a comparative "ground truth" in demonstrating substantial equivalence.

    8. The sample size for the training set

    The document does not describe a "training set" in the context of machine learning or AI. This device is an immunoassay, not an AI/ML product. The term "training set" is not applicable here. The development of the immunoparticles would involve reagent optimization and validation, but this is not typically referred to as a "training set" in this context.

    9. How the ground truth for the training set was established

    As there is no "training set" in the AI/ML sense, this question is not applicable. The development of the assay components would involve internal validation and standardization processes. For the purpose of regulatory substantial equivalence, the "ground truth" for comparison for the overall device performance was established by demonstrating comparable performance to the established predicate devices, which are legally marketed and are considered "safe and effective."

    Ask a Question

    Ask a specific question about this device

    K Number
    K032692

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-01-22

    (142 days)

    Product Code
    Regulation Number
    866.5630
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use.

    DakoCytomation Beta-2-Microglobulin Kit is intended for the quantitative determination of beta-2-microglobulin in human serum and plasma by rate nephelometry on IMMAGE® Immunochemistry Systems. Measurement of beta-2-microglobulin aids in the diagnosis of patients with active rheumatoid arthritis and kidney disease.

    Device Description

    DakoCytomation Beta-2-Microglobulin Kit is an in vitro diagnostic assay device for the quantitative determination of human beta-2-microglobulin. Beta-2-microglobulin (B2M), a low molecular weight polypeptide of 11,800 daltons, is the light chain component of the major histocompatibility antigen (HLA). B2M is present on the membrane surface of all cells that express major histocompatibility antigens and it is normally present in the circulation as a result of cell membrane turnover.

    The Beta-2-Microglobulin device is similar in design, materials and intended use to other 510(k) cleared devices, which are in commercial distribution.

    AI/ML Overview

    The provided document is a 510(k) summary for the DakoCytomation Beta-2-Microglobulin Kit. It describes the device's intended use and demonstrates its substantial equivalence to a predicate device. However, it does not contain the specific details required to answer all parts of your request regarding acceptance criteria and a study proving the device meets those criteria.

    Here's an breakdown of what can and cannot be answered based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document mentions "significant interference is defined as recovery within $\pm$ 10% at β2M > 6 mg/L or within $\pm$ 0.6 mg/L at β2M ≤ 6 mg/L" for interfering substances. This can be considered an acceptance criterion for interference.

    Acceptance CriteriaReported Device Performance
    Interfering Substances: No significant interference defined as recovery within $\pm$ 10% at β2M > 6 mg/L or within $\pm$ 0.6 mg/L at β2M ≤ 6 mg/L.Hemoglobin concentrations up to 1000 mg/dL, bilirubin (conjugated) up to 60 mg/dL and triglyceride up to 1500 mg/dL do not significantly interfere with the assay.
    Lower Detection Limit (Serum):0.3 mg/L
    Assay Range (Serum):0.3 to 20 mg/L

    Note: The document states that the new device meets these criteria but does not provide raw data or specific study results beyond stating that "do not significantly interfere." It primarily focuses on comparing features with the predicate device.

    2. Sample size used for the test set and the data provenance:

    • Not explicitly stated. The document describes the device and its intended use, and compares it to a predicate. It mentions interference studies but does not provide sample sizes or the origin (country) of the data, nor whether it was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable / Not stated. This device is an in vitro diagnostic kit, not an AI-powered diagnostic tool requiring expert interpretation of images or other subjective data for "ground truth" establishment in a traditional sense. The "ground truth" for this type of device would typically be established through analytical validation studies against reference methods or clinical samples with known analyte concentrations, not through expert consensus on interpretations.

    4. Adjudication method for the test set:

    • Not applicable / Not stated. See explanation for #3. Adjudication methods are relevant for subjective interpretations, which is not the primary output for this type of quantitative assay.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is an in vitro diagnostic kit, not an AI-assisted diagnostic tool for human readers. Therefore, an MRMC study is not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, in essence. The DakoCytomation Beta-2-Microglobulin Kit is a standalone diagnostic assay (an algorithm in this context refers to the chemical reactions and measurement principles), and its performance is evaluated independently. The device's output (quantitative beta-2-microglobulin levels) is directly used for diagnosis, not as an aid for human interpretation in the way AI imaging tools are used.

    7. The type of ground truth used:

    • For an in vitro diagnostic assay like this, the "ground truth" is typically established through:
      • Reference methods: Comparing results to a gold standard method for measuring beta-2-microglobulin.
      • Known concentration samples: Testing samples with precisely determined concentrations of beta-2-microglobulin.
      • Clinical samples with established clinical diagnoses: While the assay aids in diagnosis, its accuracy is validated by its ability to correctly quantify the analyte in samples from patients with and without the conditions of interest (rheumatoid arthritis, kidney disease) where the true B2M levels are well-characterized.
    • The document implies that the device accurately measures beta-2-microglobulin, which is the underlying "ground truth" it's designed to detect.

    8. The sample size for the training set:

    • Not applicable / Not stated. This device is a chemical assay kit, not a machine learning or AI model that requires a "training set" in the computational sense. Its performance is based on its chemical and immunological principles, not on being trained on a dataset.

    9. How the ground truth for the training set was established:

    • Not applicable / Not stated. (See explanation for #8).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1