510(k) Data Aggregation

    K Number
    K251347

    Validate with FDA (Live)

    Date Cleared
    2026-01-28

    (273 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterilization Pouch/Roll is intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam or via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:

    • Steam Sterilization at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
    • Ethylene Oxide (EO) with a concentration of 695 mg/L at 50°C (122°F) and 30% to 90% relative humidity for 4 hour. Aeration time of 7 days at room temperature.

    The Sterilization Pouch/Roll is made with medical grade paper and medical compound film. The Sterilization Pouch/Roll maintains the sterility of the enclosed devices for up to 6 months post Steam or EO gas sterilization, and before sterilization has a maximum shelf life of 3 years from the date of manufacture.

    The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process. Sterilization indicator will change color according to the sterilization method (ethylene oxide, steam). Ethylene oxide sterilization indicator color change from Pink before sterilization to Yellow after sterilization. The steam sterilization indicator color change from Blue before sterilization to Black after sterilization.

    Device Description

    The Sterilization Pouch/Roll is composed of medical grade paper(60g/m2) and medical compound film(50μm), it is intended to be used to contain medical devices to be terminally sterilized by the EtO or Steam sterilization process.

    The recommended sterilization cycle parameter is:

    Steam: 4 minutes at 132°C(270°F); 20 minutes dry time.

    Ethylene oxide: 4 hours at 50 °C(122°F); relative humidity between 30%- 90%; ethylene oxide concentration is 695 mg/L, 7 days aeration time at room temperature.

    The medical devices are inserted into the Pouch/ Roll, sealed, and then sterilized for the EtO or Steam Sterilization Process. The pouch is self-sealed prior to sterilization processing. The roll is heat-sealed by a plastic sealing machine. After completion of the sterilization process, the Pouch/Roll maintain sterility of the enclosed medical devices until the seal is opened.

    The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 6 months post sterilization.

    The Pouch/Roll is printed with a chemical indicator bar that changes from Pink to yellow (EtO) or Blue to Black (Steam) when exposed to EtO gas or Steam vapor during process. The EtO and Steam Chemical Indicator offers an addition way to verity processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. The EtO and Steam Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to the EtO or Steam.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K180672

    Validate with FDA (Live)

    Date Cleared
    2018-06-06

    (84 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterilization Pouch/Roll Made with Tyvek® are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERRAD® 100S Sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 3 years post sterilization.

    The pouches and rolls are printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERRAD® 100S Sterilizer.

    The Sterilization Pouch/Roll Made with Tyvek® is offered in the follow 5 types:

    • Self-sealing sterilization pouches
    • Sterilization pouches, Flat
    • Sterilization pouches, Gusseted
    • Sterilization rolls, Flat
    • Sterilization rolls, Gusseted
    Device Description

    Sterilization Pouch/Roll Made with Tyvek® is intended to contain medical devices to be terminally sterilized in the STERRAD® 100S Sterilization System. The medical devices are inserted into the Pouch/Roll, sealed, and then sterilized in the STERRAD® 100S Sterilization System (The recommended cycle parameter is 6 minutes (Injection volume: 2880µL) at 50°C.). After completion of the sterilization process, the Pouch/Roll maintains sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 3 years post sterilization.

    The Pouch/Roll is printed with a chemical indicator bar that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERRAD® 100S Sterilization System.

    The pouches are constructed from Tyvek®/plastic films, with H2O2 Chemical Indicator printed onto the Tyvek® film. The Self-seal pouch permits sealing of the pouch without need of heat- sealing equipment, while the heat sealed pouches and rolls are heat sealed prior to processing in the STERRAD® Sterilization Systems.

    The H2O2 Chemical Indicator offers an additional way to verify processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. H2O2 Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to the hydrogen peroxide. The color of the Chemical Indicator changes from red to blue (or lighter) when exposed to hydrogen peroxide.

    The Sterilization Pouch/Roll Made with Tyvek® is offered in the follow 5 types:

    • Self-sealing sterilization pouches
    • Sterilization pouches, Flat
    • Sterilization pouches, Gusseted
    • Sterilization rolls, Flat
    • Sterilization rolls, Gusseted
    AI/ML Overview

    The provided text describes the regulatory submission for a medical device (Sterilization Pouch/Roll Made with Tyvek®) seeking 510(k) clearance from the FDA. This document focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than proving the safety and effectiveness of a novel AI/software-as-a-medical-device (SaMD).

    Therefore, the information typically requested regarding acceptance criteria and study design for AI/SaMD devices (like sample size for test/training sets, data provenance, expert
    qualifications, ground truth establishment, MRMC studies, or standalone algorithm performance) is not applicable to this submission. This device is a physical product, and its performance is evaluated through material science, sterility, and packaging tests.

    However, I can extract the acceptance criteria and reported performance for the physical device based on the non-clinical testing summaries.

    1. A table of acceptance criteria and the reported device performance:

    Test CategoryTest Method / Standard (Acceptance Criteria)Reported Device Performance
    Sterilant PenetrationAAMI / ANSI / ISO 14937:2009, "Sterilization of Health Care Products - General Requirements for Characterization of a Sterilizing Agent and the Development, Validation, and Routine Control of a Sterilization Process for Medical Devices" (Acceptance: Able to penetrate and sustain direct contact with the medical instrument inside)Confirmed that the sterilant is able to penetrate the device and sustain direct contact with the medical instrument inside.
    Sterilant PenetrationHalf-Cycle Efficacy (Acceptance: 6 log reduction of Geobacillus stearothermophilus)Showed a 6 log reduction of Geobacillus stearothermophilus.
    BiocompatibilityISO 10993-10:2010(E), "Biological Evaluation of Medical Devices - Part 10: Tests for irritation and skin sensitization" (Acceptance: Meets requirements, non-irritating)Showed "negative reaction" (non-irritating) and met requirements.
    Package IntegrityASTM F2251-03, "Standard Test Method for Thickness Measurement of Flexible Packaging Material" (Acceptance: Passed)Passed
    Package IntegrityASTM D882, "Standard Test Method for Tensile Properties of Thin Plastic Sheeting" (Acceptance: Passed)Passed
    Package IntegrityASTM D 5035, "Standard Test Method for Breaking Force and Elongation of Textile Fabrics (Strip Method)" (Acceptance: Passed)Passed
    Package IntegrityASTM D1922, "Standard Test Method for Propagation Tear Resistance of Plastic Film and Thin Sheeting by Pendulum Method" (Acceptance: Passed)Passed
    Package IntegrityASTM F1140-07, "Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages" (Acceptance: Passed)Passed
    Package IntegrityASTM F88/F88M; ISO 11607-1, "Standard Test Method for Seal Strength of Flexible Barrier Materials" (Acceptance: Passed)Passed
    Package IntegrityASTM F1929-98 (04); ISO 11607-1, "Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration" (Acceptance: Passed)Passed
    Material CompatibilityAAMI / ANSI / ISO 11607-1:2006//A1:2014, "Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems" (Acceptance: Met criteria)Met material compatibility acceptance criteria.
    Sterility MaintenanceDIN 58953-6:2010-05, "Sterilization - Sterile supply - Part 6: Microbial barrier testing of packaging materials for medical devices which are to be sterilized" (Acceptance: Passed)Passed
    Sterility MaintenanceAAMI / ANSI / ISO 11607-1:2006//A1:2014 (Acceptance: Met criteria for sterility maintenance after aging)Met sterility maintenance acceptance criteria.
    Sterility MaintenanceDurability: Accelerated Aging Test (ASTM F 1980-2007; ISO 11607-1) (Acceptance: Passed for 3 years accelerated aging, maintaining seal strength)Passed (3 years accelerated aging, seal strength maintained).
    Chemical Indicator EfficacyAAMI / ANSI / ISO 11140-1:2014, "Sterilization of Health Care Products - Chemical Indicators - Part 1: General Requirements" (Acceptance: Sterilant penetrated through pouch, CI color changed to endpoint color, color stability for 3 years, completeness/uniformity of color change)The sterilant penetrated through the pouch, the CI color changed to the endpoint, indicator dye remained stable for 3 years real-time aging, and the color change was effective.

    For the following points, as explained, they are not applicable because this is a 510(k) submission for a physical medical device (sterilization pouches) based on substantial equivalence, not an AI/SaMD (Software as a Medical Device). The evaluation focuses on material properties and performance under sterilization conditions rather than algorithmic performance on complex data (like images or medical records).

    2. Sample sizes used for the test set and the data provenance: Not applicable. The tests performed are engineering and material-science based, using samples of the device and specific test setups, not patient-derived data. The document does not specify a "test set" in the context of an AI algorithm, nor "data provenance" (e.g., country of origin) relevant to patient data. The non-clinical tests were likely conducted in a controlled laboratory environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. Ground truth as typically defined for AI/SaMD (e.g., disease diagnosis from medical images) is not relevant here. The "ground truth" for these tests are objective measurements against established engineering and biological standards (e.g., quantitative measures of tensile strength, observable microbial growth, or chemical indicator color change).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. Adjudication methods are used to resolve disagreements among human annotators/experts for ground truth establishment in AI/SaMD; this is not relevant for materials testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for the performance evaluations are based on established industry standards and scientific principles for material properties (e.g., ASTM standards for mechanical properties), biological efficacy (e.g., 6-log reduction for sterilization), and chemical reactivity (e.g., color change of indicators). No human expert consensus or outcomes data in a clinical context is used as "ground truth" for this device's acceptance.

    8. The sample size for the training set: Not applicable. There is no AI training set involved.

    9. How the ground truth for the training set was established: Not applicable. There is no AI training set involved.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133838

    Validate with FDA (Live)

    Date Cleared
    2015-03-11

    (448 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterilization Pouch/Roll Made with Tyvek® are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERRAD® 100S Sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 3 years post sterilization.

    The pouches and rolls are printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERRAD® Sterilization System.

    Device Description

    Sterilization Pouch/Roll Made with Tyvek® is intended to contain medical devices to be terminally sterilized in the STERRAD® Sterilization Systems. The recommended Gas Plasma sterilization cycle parameter is 6 minutes (Injection volume: 2880µ L) at 50°C. The medical devices are inserted into the Pouch/Roll, sealed, and then sterilized in the STERRAD® Sterilization System. After completion of the sterilization process, the Pouch/Roll maintain sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 3 years post sterilization.

    The Pouch/Roll is printed with a chemical indicator bar that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERRAD® Sterilization System.

    The proposed pouches are constructed from Tyvek® plastic films, with H2O2 Chemical Indicator printed onto the Tyvek" film. The Self-seal pouch permits sealing of the pouch without need of heat- sealing equipment, while the heat sealed pouches and rolls are heat sealed prior to processing in the STERRAD® Sterilization Systems.

    The H2O2 Chemical Indicator offers an addition way to verity processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. H2O2 Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to the hydrogen peroxide. The color of the Chemical Indicator changes from red to blue (or lighter) when exposed to hydrogen peroxide.

    The Sterilization Pouch/Roll Made with Tyvek® is offered in the follow 5 types:

    • Self-sealing sterilization pouches .
    • . Sterilization pouches, Flat
    • . Sterilization pouches, Gusseted
    • Sterilization rolls, Flat
    • Sterilization rolls, Gusseted
    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called "Sterilization Pouch/Roll Made with Tyvek®". It details the device's characteristics, intended use, and a comparison to a predicate device to establish substantial equivalence, which is the basis for its FDA clearance.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    The document presents performance data primarily in a comparative table (Table 5-2, continued on subsequent pages in the original document) comparing the new device against its predicate. It also lists the standards used for testing throughout the "Effectiveness and Safety" section.

    Performance MetricAcceptance Criteria (Standard Reference and/or Implied Criterion)Reported Device Performance
    Sterilant PenetrationMeet the requirement of SAL 10-6 (ANSI/AAMI/ISO 14937:2009)The test met the requirement of SAL 10-6
    Package Integrity (Physical Properties)(Tested before and after Gas Plasma Sterilization)
    Thickness Variations (mm)(ASTM F 2251-03) No specific numerical acceptance criteria provided, but "Small" values for both new and predicate are similar.Small: 0.145 (before/after); Large: 0.146 (before/after)
    Tear Resistance (g)(ASTM D1922) No specific numerical acceptance criteria provided, but similar to predicate.CD: 259 (before), 258 (after); MD: 282 (before), 280 (after)
    Tensile strength of plastic film (kgf/mm2)(ASTM D882) No specific numerical acceptance criteria provided, but similar to predicate.CD: 575 (before), 531 (after); MD: 577 (before), 531 (after)
    Tensile strength of Tyvek® (N/2.54cm)(ASTM D 5035) No specific numerical acceptance criteria provided, but similar to predicate.CD: 174 (before), 170 (after); MD: 163 (before), 162 (after)
    Burst Strength (kPa)(ASTM F1140-07) No specific numerical acceptance criteria provided, but similar to predicate.Small: 21.4 (before), 17.95 (after); Large: 4.49 (before), 2.03 (after)
    Seal Peel Test (g/15mm)Pass (ASTM F88/F88M-09; ISO 11607-1) - The predicate also reports "Pass".Upper: 340.3 (before), 493.6 (after); Down: 506.7 (before), 709.9 (after); Left: 345.7 (before), 436.0 (after); Right: 316.5 (before), 518.5 (after). Result: Pass
    Dye penetration TestNo Channels identified on package (ASTM F 1929-12; ISO 11607-1)No Channels identified on package
    Microbial Barrier Test< 1 (Result: Pass) (DIN 58953-6:2010-05)< 1 (Result: Pass)
    Toxicological PropertiesNegative reaction (ANSI/AAMI/ISO 10993-10)Negative reaction
    Durability: Accelerated Aging Test3 Years (ASTM F 1980-2007; ISO 11607-1 2006) - The predicate also states "3 Years"3 Years
    Shelf Life3 Years3 Years
    Chemical Indicator EfficacyAbility to effectively demonstrate stability of ink before use, lasting quality (color stability) of color change, completeness and uniformity of color change, and all or none color change (AAMI / ANSI / ISO 11140-1:2005)Meets the requirements of ISO 11140-1:2005

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific sample sizes used for each test (e.g., number of pouches/rolls tested for peel strength, burst strength, etc.). It mentions that "the test items were performed on materials of the products" for some physical property tests, implying samples of the materials were used.

    The data provenance is from Taiwan (R.O.C.), as the manufacturer is Sigma Medical Supplies Corporation located in New Taipei City, Taiwan. The studies appear to be prospective as they involve testing the newly developed device against established standards and a predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This section is not applicable to this type of device and study. The "ground truth" for this medical device (sterilization pouches/rolls) is established through adherence to recognized international and national consensus standards for physical properties, sterilization efficacy, biocompatibility, and product performance, rather than through expert clinical assessment or interpretation of medical images. The studies are laboratory-based and measure objective physical and biological parameters. Therefore, there are no "experts" in the traditional sense of medical image interpretation (like radiologists) establishing ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This section is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving human interpretation (e.g., radiologists reviewing images) where disagreements need to be resolved. For laboratory-based testing of physical and chemical properties of a medical device like a sterilization pouch, direct measurements are taken against predefined acceptance criteria from standards. There isn't a process of expert adjudication for these types of objective measurements.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This section is not applicable. MRMC studies are related to the evaluation of AI-powered diagnostic devices that involve human readers (e.g., radiologists, pathologists). This document is for a physical medical device (sterilization pouches/rolls) and does not involve AI or human "readers" in the context of diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This section is not applicable. This device is a physical product, not an algorithm related to AI or software.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance of the Sterilization Pouch/Roll Made with Tyvek® is based on established international and national consensus standards for medical device packaging, sterilization processes, and biocompatibility. These include:

    • Sterilant Penetration: Adherence to SAL 10-6 as per ANSI/AAMI/ISO 14937:2009.
    • Package Integrity and Material Properties: Defined by standards like AAMI/ANSI/ISO 11607-1:2006/(R) 2010, ASTM D882, ISO 1924-2, ASTM D 5035, ASTM F 2251-03, ASTM D 1922, ASTM D 1004, ASTM F 1140-07, ASTM F 1929-98 (04), ASTM F88/F88M-09, ASTM F 1980-2007.
    • Microbial Barrier: DIN 58953-6:2010-05.
    • Biocompatibility: ISO 10993-10 Third Edition 2010-08-01.
    • Chemical Indicator Efficacy: AAMI / ANSI / ISO 11140-1:2005.

    These standards provide objective, measurable criteria for the device's performance, thereby defining the "ground truth" for its acceptability.

    8. The sample size for the training set

    This section is not applicable. This submission is for a physical medical device and does not involve machine learning algorithms that require a "training set." The materials and manufacturing processes are validated, not "trained."

    9. How the ground truth for the training set was established

    This section is not applicable as there is no training set involved for this device submission.