Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K250263

    Validate with FDA (Live)

    Date Cleared
    2025-03-31

    (61 days)

    Product Code
    Regulation Number
    874.4680
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Grasping Forceps FG-52D/54D are intended to be used to retrieve foreign bodies, calculus or tissue specimens from the tracheobronchial tree in combination with an endoscope.

    Device Description

    The Disposable Grasping Forceps FG-52D/FG-54D are designed to retrieve foreign bodies, calculus, or tissue specimens endoscopically within the tracheobronchial tree when used in conjunction with a compatible endoscope. Each device is provided sterile and consists of a handle and an insertion portion. The handle (proximal portion) of the Disposable Grasping Forceps includes a ring which can be pushed or pulled to open or close the grasping portion of the device. The distal end of the insertion portion consists of the grasping portion of the device. The grasping portion, manufactured from stainless steel, is provided in two shapes: spiral basket type (model FG-52D) and three nail type (model FG-54D) and has limited contact with the patient. The user will select the device model depending on procedural needs.

    AI/ML Overview

    The provided document is a 510(k) summary for the Olympus Disposable Grasping Forceps FG-52D/FG-54D. It outlines the device's characteristics, indications for use, comparison to a predicate device, and performance data used to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance results in a consolidated format. Instead, it lists the types of tests conducted to demonstrate safety and effectiveness. The "acceptance criteria" are implied by the successful completion and positive results of these tests, leading to the conclusion of substantial equivalence.

    Here's a re-formatted table based on the provided "Performance Data" section:

    Performance Data Category / Acceptance Criteria (Implied)Reported Device Performance (Implied by successful completion)
    BiocompatibilityMet standards per ISO 10993-1:2018, including cytotoxicity (ISO 10993-5:2009), sensitization (ISO 10993-10:2021), irritation (ISO 10993-23:2021), acute systemic toxicity (ISO 10993-11:2017), and material mediated pyrogenicity (USP <151>).
    Sterilization ValidationMet standards per ISO 11135:2014.
    Ethylene Oxide ResidualsMet standards per ISO 10993-7:2008.
    Packaging Validation & Shelf LifeMet standards in accordance with ISO 11607-1:2019 and ASTM F1980-16.
    Mechanical Testing & Comparative TestingVerified device performance for: Insertion force/Withdrawal force, Handle Operation, Grasping Performance, Strength Testing of the Connection and Distal Tip, and Device Reliability. (Implies successful performance within acceptable limits, though specific values are not provided).
    Human Factors TestingConducted. (Implies safe and effective usability, though specific findings are not detailed).

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for each test conducted. It generally states that "Performance Data" was provided. For the provenance, the data seems to be from non-clinical bench testing conducted by the manufacturer, Olympus Medical Systems Corp. The document does not mention the country of origin for the data or whether it was retrospective or prospective, as these are typically not applicable to bench testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable to this submission. The device is a physical medical instrument (graspers), not an AI/ML algorithm that requires expert labeling for ground truth. The "ground truth" for its performance is established through standardized engineering and biological tests, not expert consensus on data interpretation.

    4. Adjudication method for the test set

    This information is not applicable as the device is not an AI/ML algorithm that involves human interpretation requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study was not done. This type of study is relevant for AI-powered diagnostic devices, not for a physical grasping forceps.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable as the device is not an algorithm.

    7. The type of ground truth used

    The "ground truth" for this device's performance is established through standardized testing methodologies as referenced in the performance data section (e.g., ISO standards for biocompatibility, sterilization, packaging, and specific mechanical tests like insertion force, grasping performance, and strength testing). There is no "expert consensus," "pathology," or "outcomes data" in the typical sense of AI/ML validation for this type of medical device.

    8. The sample size for the training set

    This information is not applicable as the device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable as the device is not an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1