Search Results
Found 2 results
510(k) Data Aggregation
(26 days)
The LF1212A LigaSure Sealer/Divider is a bipolar electrosurgical instrument intended for use in open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The LigaSure Sealer/ Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery and in such surgical specialties as urologic, thoracic, plastic, and reconstructive. Procedures may include, but are not limited to, bowel resections, gall bladder procedures, Nissen fundoplication, and adhesiolysis.
The instrument is also indicated for open ENT procedures in adults (thyroidectomy, radical neck dissection, parotidectomy, and tonsillectomy) for ligation and division of vessels, lymphatics and tissue bundles 2-3 mm away from unintended thermally-sensitive structures such as nerves and parathyroid glands.
The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.
The LigaSure™ Curved, Small Jaw, Open Sealer/Divider (LF1212A) is a sterile, single-use, hand-held bipolar electrosurgical instrument designed for use with compatible Covidien generators that include vessel sealing capabilities to ligate (seal) and divide (cut) vessels, tissue bundles, and lymphatics during open general surgical procedures (as indicated). Covidien electrosurgical generators that include vessel sealing capabilities deliver precise energy through the device to tissue for a controlled tissue response to achieve complete and permanent tissue fusion while producing minimal sticking, charring, and thermal spread to adjacent tissue.
The surgeon first grasps the vessel or tissue with the jaws of the instrument. Once the surgeon is confident that the instrument is placed correctly, he or she initiates a seal by pressing a handswitch or footswitch. The instrument creates a seal by application of RF electrosurgical energy to vascular structures (vessels and lymph) or tissue bundles interposed between the jaws of the instrument. Once the seal cycle is complete, the surgeon actuates a blade within the instrument to divide tissue along the seal line.
The purpose of this submission is to propose the use of an alternate dot material for the LigaSure Curved, Small Jaw, Open Sealer/Divider devices (K141153). There have been no prior submissions for the proposed change. The proposed devices do not contain software.
This document is a 510(k) summary for a medical device called the "LigaSure™ Curved, Small Jaw, Open Sealer/Divider (LF1212A)". This submission is for a modification to an existing device, specifically changing the material of ceramic spacers and adding an etched wetting ring feature. Therefore, the information provided focuses on demonstrating that the modified device is substantially equivalent to the predicate device and that the changes do not negatively impact its safety or performance.
The document does not describe a study involving AI, human readers, or a typical diagnostic device's acceptance criteria like sensitivity, specificity, or AUC. Instead, it describes hardware device performance criteria based on engineering specifications and safety standards.
Here's an analysis of the provided text in relation to the requested information, acknowledging that many requested points are not directly applicable due to the nature of this particular device submission (a modification to a non-AI surgical instrument):
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of acceptance criteria with reported device performance in the format typically seen for AI/diagnostic devices (e.g., sensitivity/specificity thresholds). Instead, it states that "design verification testing... confirmed the continued conformance to applicable technical design specifications and performance requirements". These performance requirements are listed as:
| Acceptance Criterion (Type of Testing) | Reported Device Performance |
|---|---|
| Biocompatibility evaluation | Conforms to ISO 10993-1:2009 |
| Basic safety and essential performance | Conforms to IEC 60601-1:2005, A1:2012 and IEC 60601-2-2:2009 |
| Electromagnetic compatibility | Conforms to IEC 60601-1-2:2007, CORR: July 31, 2010 |
| Design verification testing: Jaw gap | Conformance confirmed |
| Design verification testing: Force | Conformance confirmed |
| Design verification testing: Cut length | Conformance confirmed |
| Design verification testing: Hyperextension | Conformance confirmed |
| Design verification testing: Resistance | Conformance confirmed |
| Design verification testing: Renal burst pressure | Conformance confirmed |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify a "test set" in the context of data for an AI algorithm. The device is a surgical instrument, and testing involves physical evaluation and engineering measurements. The document states "design verification testing", which implies a set of physical devices were tested. However, the exact sample size (number of devices or tests performed) is not provided. No data provenance related to "country of origin for data" is relevant here, nor is retrospective or prospective as it's not a data-driven model.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This question is not applicable. The device is a surgical instrument, not an AI diagnostic tool. "Ground truth" for this device is based on objective engineering specifications and international safety standards, not expert consensus on medical images or diagnoses. The "truth" is whether the physical properties and performance characteristics of the instrument meet defined regulatory and design standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This question is not applicable as there is no "test set" in the context of medical images or diagnoses requiring adjudication by experts.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is a surgical instrument and does not involve AI or human readers for diagnostic interpretation. The document explicitly states: "This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable as the device does not contain an AI algorithm. The document explicitly states: "The proposed devices do not contain software."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this device's performance is established by engineering specifications, international safety standards (ISO, IEC), and design verification requirements. For example, "renal burst pressure" testing has an objective, measurable outcome that is compared against a predefined standard for acceptable sealing strength.
8. The sample size for the training set
This question is not applicable as the device does not involve an AI algorithm or a training set.
9. How the ground truth for the training set was established
This question is not applicable as the device does not involve an AI algorithm or a training set.
Ask a specific question about this device
The LigaSure™ Curved, Small Jaw, Open Sealer / Divider is a bipolar electrosurgical instrument intended to be used with the ForceTriad™ energy platform. The instrument is indicated for use in open general surgical procedures where ligation and division of vessels (up to 7 mm in diameter), tissue bundles, and lymphatics is performed, such as urologic, thoracic, plastic, and reconstructive; and including such procedures as bowel resections, gall bladder procedures, Nissen fundoplication, adhesiolysis, etc.
The device is also indicated for open ENT procedures in adults (thyroidectorny, radical neck dissection, parotidectorny, and tonsillectorny) for ligation and division of vessels, lymphatics and tissue bundles 2-3 mm away from unintended thermally sensitive structures such as nerves and parathyroid glands.
The LigaSure system has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure system for these procedures.
The LigaSure™ Curved, Small Jaw, Open Sealer/Divider is a sterile, single-use, hand-held bipolar electrosurgical instrument designed exclusively for use with the ForceTriad™ energy platform (generator) to ligate (seal) and divide (cut) vessels, tissue bundles, and lymphatics during open general surgical procedures (as indicated). The ForceTriad's tissue-fusion (LigaSure) mode delivers precise energy to tissue for a controlled time period to achieve complete and permanent tissue fusion, and has been designed to produce minimal sticking, charring, or thermal spread to adjacent tissue.
LigaSure instruments attach to the generator with a ten-foot containing a "smart" connector. This connector functions as a unique product identifier for device-specific recognition by the generator. The Curved, Small Jaw, Open Sealer/Divider instrument is designed to be both ergonomic and intuitive for the user. Its hemostat-style body and symmetrically placed controls facilitate handling by both left and righthanded users, and its small, curved jaws maximize visibility and access when the instrument is used in confined surgical spaces.
Ring handles function to allow the user to grasp tissue by opening and closing the jaws of the instrument. The interior surfaces of the jaws contain the electrodes, which serve to ligate by delivering energy to the grasped tissue. RF energy can be activated by the user in two ways: (1) through the use of a single button incorporated into the handle body or (2) through the use of a footswitch attached to the generator. A cutting mechanism functions to mechanically divide tissue following tissue fusion. It consists of a stainless steel blade and is controlled by the user through a trigger located on the handle body.
The device in question is the LigaSure™ Curved, Small Jaw, Open Sealer / Divider (LF1212) by Covidien, a bipolar electrosurgical instrument. The purpose of this submission (K113572) is to expand its indications for use to include ENT procedures and remove a warning regarding specific surgical procedures requiring precision. There are no physical changes to the device itself; the design remains the same as its predicate (K102470).
This 510(k) summary does not use a typical "acceptance criteria" table in the form of specific numerical thresholds for performance metrics. Instead, it demonstrates "substantial equivalence" to predicate devices by comparing various performance characteristics. For a device like an electrosurgical instrument, "acceptance criteria" are implicitly met if its performance is comparable to or better than previously cleared, substantially equivalent devices, particularly regarding safety and effectiveness.
Here's an interpretation of the performance data and how it addresses the 'acceptance criteria' in the context of substantial equivalence:
1. Table of "Acceptance Criteria" (interpreted as performance comparison to predicates) and Reported Device Performance:
| Performance Metric | Acceptance Criteria (Implied by Predicate Performance) | Reported Device Performance (LigaSure LF1212) |
|---|---|---|
| Burst Pressure | Equivalent or superior mean burst pressure on isolated vessels to predicate devices (LigaSure™ Curved, Small Jaw, Open Sealer/Divider (K102470) and LigaSure™ Precise Instrument Vessel Sealing System (K010010)). | "The subject device was shown to yield higher mean burst pressures on isolated vessels than the predicates." |
| Thermal Profile | Equivalent or superior (lower) mean temperatures after single and multiple activations compared to predicate devices. | "Thermal profile studies showed the subject device has consistently lower mean temperatures than the predicates after single and multiple activations." |
| Mean Thermal Spread | Comparable mean thermal spread to predicate devices for all vessel sizes evaluated. | "The mean thermal spread of the subject device was shown to be comparable to the predicates for all vessel sizes evaluated." |
| Usability/Usability | Adequate assurance of safety and performance (human factors/usability) in simulated operating environments, including ENT space. | "Usability was evaluated with users in simulated operating environments, including in the ENT space. These studies consisted of several formative and one summative study, which demonstrate the instrument provides adequate assurance of safety and performance (in regards to human factors/usability aspects) for the patient and operator." |
| Clinical Effectiveness/Safety in ENT Procedures | Demonstrable safety and effectiveness for use in ENT procedures, supported by existing clinical knowledge and experience. | "An assessment of published clinical literature demonstrates that there is a considerable body of knowledge and experience that shows the use of LigaSure instruments in ENT procedures is safe and effective." The submission also argues that "The use of the LigaSure Curved, Small Jaw, Open Sealer / Divider in ENT procedures does not raise any new types of questions of safety and effectiveness compared with the predicate devices currently used for these procedures." |
Study Proving Device Meets Acceptance Criteria:
The study proving the device meets the "acceptance criteria" (defined as substantial equivalence to predicates) is a preclinical bench and animal testing study combined with usability studies and a clinical literature review.
2. Sample Size and Data Provenance (for test set, relevant for bench/animal testing and usability):
- Preclinical Bench Testing (Burst Pressure & Thermal Profile): The document states "Preclinical comparative bench testing... was conducted." It does not specify the exact sample size for isolated vessels or the number of activations for thermal profile tests. The data provenance is implied to be laboratory testing conducted by Covidien.
- Animal Testing (Thermal Spread): The document states "Animal Testing: The mean thermal spread of the subject device was shown to be comparable to the predicates for all vessel sizes evaluated." It does not specify the number of animals or the number of vessel seals performed. The data provenance is likely internal animal studies conducted by Covidien.
- Usability / Human Factors: "These studies consisted of several formative and one summative study." It does not specify the number of users (sample size) for these studies. The data provenance is internal usability studies.
All these studies appear to be prospective in nature, as they were conducted specifically for this submission to compare the device against predicates or assess usability. The country of origin for the data is not explicitly stated but is presumed to be the United States, given the submitter's location and the FDA submission.
3. Number of Experts and Qualifications (for establishing ground truth for the test set):
- Bench and Animal Testing: The establishment of ground truth for metrics like burst pressure, thermal profile, and thermal spread typically relies on objective physical measurements and established scientific methodologies, not expert consensus in the human interpretation sense. Therefore, the "experts" would be the scientists and engineers conducting and analyzing these objective tests, following standardized protocols. Their specific qualifications (e.g., "radiologist with 10 years of experience") are not provided, as this is not a diagnostic device.
- Usability Studies: "Usability was evaluated with users... in the ENT space." These users are the "experts" in terms of operating the device. While their specific qualifications (e.g., "surgeon with X years of experience") are not detailed, it can be inferred that they were medical professionals (surgeons or surgical assistants) experienced in surgical procedures, likely ENT specialists given the mention of "ENT space." No specific number of experts for "ground truth setting" is given in the context of interpretation of results, as the usability studies themselves generate the data.
- Clinical Literature Review: An "assessment of published clinical literature" was performed. This implies medical experts (likely regulatory affairs personnel or medical scientists within Covidien) evaluated existing peer-reviewed studies. No specific number or qualifications of these reviewers are provided.
4. Adjudication Method (for the test set):
Since the "acceptance criteria" are based on objective physical measurements (burst pressure, thermal profile, thermal spread) and directly observed usability, rather than subjective interpretations by multiple experts, a formal "adjudication method" like 2+1 or 3+1 for resolving discrepancies in expert opinions is not applicable or described. The results of bench and animal tests are typically analyzed statistically. Usability studies involve observations and feedback, rather than adjudication of 'ground truth'.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done and would not be applicable for this type of device (a surgical instrument). MRMC studies are typically used for diagnostic imaging devices where human readers interpret medical images with and without AI assistance to assess changes in diagnostic accuracy or efficiency. This device is a surgical tool, not an image interpretation or diagnostic aid. Therefore, no effect size of human readers improving with AI assistance is reported.
6. Standalone (Algorithm Only) Performance:
No, a standalone performance study (algorithm only without human-in-the-loop) was not done and is not applicable for this device. This device is a manual surgical instrument that requires a human operator for its function; it does not contain a standalone algorithm in the sense of AI performing a task independently. The "algorithm" here refers to the energy delivery system within the ForceTriad™ platform, which is always used with the human-controlled instrument.
7. Type of Ground Truth Used:
- Bench Test (Burst Pressure, Thermal Profile): The ground truth is based on objective physical measurements using scientific equipment and standardized test methods (e.g., pressure gauges, temperature probes).
- Animal Testing (Thermal Spread): Ground truth is based on objective pathological assessment of tissue sections (histology) or direct macroscopic measurements of the thermal damage zone in animal models.
- Usability Studies: Ground truth is based on observed user performance, task completion, and user feedback in simulated environments, assessed against predefined safety and performance metrics.
- Clinical Literature: The "ground truth" for clinical safety and effectiveness in ENT procedures is derived from the published outcomes data and expert consensus within the scientific and medical literature.
8. Sample Size for the Training Set:
This device is a hardware surgical instrument, not a machine learning or AI algorithm in the typical sense that would require a "training set" for model development. The "ForceTriad™ energy platform (generator)" delivers energy, and its design would be based on engineering principles and previous device experience, not statistical learning from a "training set" of data. Therefore, the concept of a "training set" and its sample size is not applicable or addressed in this submission.
9. How Ground Truth for Training Set Was Established:
As there is no "training set" in the context of a machine learning model, the question of how its ground truth was established is not applicable. The device's 'training' or development is based on engineering design, preclinical testing, and iterative improvements, not a data-driven model training process.
Ask a specific question about this device
Page 1 of 1