Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K223158

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-01-23

    (108 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LigaSure XP Maryland Jaw Sealer/Divider is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation and division of vessels, thick tissue (tissue bundles), and lymphatics is desired. The LigaSure XP Maryland Jaw Sealer/Divider can be used on vessels (arteries and veins) up to and including 7 mm. It is indicated for use in general surgery and such surgical specialties as colorectal, bariatric, urologic, vascular, thoracic, and gynecologic. These may include, but are not limited to, such procedures as Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, hysterectomy, and so forth.

    The LigaSure XP Maryland Jaw Sealer/Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the LigaSure XP Maryland Jaw Sealer/Divider for these procedures.

    Device Description

    The proposed LigaSure™ XP Maryland Jaw Sealer/Divider is a sterile, single-use bipolar vessel sealer. It is labeled as prescription use only. The device connects to the Valleylab™ FT10 Electrosurgical Platform generator for tissue ligation. Energy is applied to tissue interposed between the nano-coated Maryland-style jaws creating a ligation. The jaws contain an independent cutting blade for division of tissue. A distinctive characteristic is the new continuous rotation capability of the jaws and shaft. The shaft is a common 5 mm diameter available in three lengths (23 cm, 37 cm, 44 cm) for various general surgical procedures in both open and minimally invasive (laparoscopic) approaches. The instrument body is a pistol grip design which can be used by right or left-handed users to access the controls. The new device is offered with two handle body designs, One-Step Sealing (LXMJ23S, LXMJ37S, LXMJ44S) or Latching Handle (LXMJ23L, LXMJ37L, LXMJ44).

    AI/ML Overview

    The provided text does not contain information about the acceptance criteria and study proving a digital health device's performance. Instead, it describes a substantial equivalence determination for a surgical instrument, the LigaSure™ XP Maryland Jaw Sealer/Divider.

    Therefore, many of the requested categories in your prompt, such as "number of experts used to establish the ground truth," "adjudication method," "MRMC comparative effectiveness study," "standalone performance," "training set size," and "ground truth for the training set," are not applicable to this medical device submission, which focuses on hardware and electrosurgical performance rather than an AI-powered diagnostic or assistive tool.

    However, I can extract the relevant performance testing and acceptance criteria information that is present in the document for the LigaSure™ XP Maryland Jaw Sealer/Divider.

    Here's the information structured to the best of my ability given the provided document:


    Acceptance Criteria and Study for LigaSure™ XP Maryland Jaw Sealer/Divider

    This submission details the substantial equivalence of an electrosurgical cutting and coagulation device, the LigaSure™ XP Maryland Jaw Sealer/Divider. The performance testing described focuses on various engineering, biological, and functional aspects of the surgical instrument, not on the performance of a digital health or AI device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Predetermined Device Requirements)Reported Device Performance
    SterilizationCompliance with ISO 11135, ISO 11737-1, ISO 11737-2, ISO 10993-7Met acceptance criteria for sterilization by ethylene oxide (EO).
    Shelf LifeCompliance with ISO 11607-1, ISO 11607-2, ASTM D4169, ASTM F1980Packaging and product integrity acceptable for five-year shelf life.
    BiocompatibilityCompliance with ISO 10993-1 for: cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, and material-mediated pyrogenicity.Met requirements for all specified endpoints.
    SoftwareN/A (Device does not contain software)N/A
    Electromagnetic Compatibility (EMC) & Electrical SafetyCompliance with IEC 60601-1-2, IEC 60601-2-2, and IEC 60601-2-18Met applicable clauses of standards.
    Bench PerformancePredetermined device requirements for: mechanical/functional performance, visual inspection, device reliability, ex vivo vessel sealing/burst, and ex vivo lymphatic sealing/burst testing.Met all predetermined device requirements.
    Animal PerformancePredetermined device requirements for: in vivo acute hemostasis, acute thermal spread, and chronic hemostasis tissue testing.Met all predetermined device requirements.
    Clinical PerformanceSufficiency of clinical literature to support safety and effectiveness for stated indications.Clinical literature studies were evaluated to further support safety and effectiveness.
    Technological CharacteristicsComparison to predicate devices for aspects like indications for use, sterility, packaging, instrument design, lever latching mechanism, energy activation, shaft/jaws rotation, jaw type/shape, electrical characteristics, compatible energy platform, operating amperage/voltage, direct tissue contacting materials, patient contacting materials, jaw coating, and shaft material.Differences (e.g., continuous 360° rotation, removal of heat shrink on shaft) were deemed not to affect safety and effectiveness and not to raise different questions of safety and effectiveness as demonstrated through performance testing.

    2. Sample size used for the test set and the data provenance
    The document does not specify exact sample sizes for each test set (e.g., number of vessels, lymphatic segments, or animals).

    • Provenance: The tests described (bench, animal) are laboratory-based performance studies, likely conducted by the manufacturer Covidien, llc. The document does not mention data provenance in terms of country of origin or whether human data was retrospective or prospective, as clinical trials for efficacy/safety in humans were not the primary evaluation method shown (literature review was used for clinical support).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    This information is not applicable and not provided. The "ground truth" for this device's performance would be objective measurements in engineering, biological models, and animal studies (e.g., burst pressure, hemostasis, thermal spread), not expert consensus on interpretations.

    4. Adjudication method for the test set
    This is not applicable and not provided. Performance outcomes are likely based on direct measurement against predefined physical/biological thresholds.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    This is not applicable. This device is a surgical instrument, not an AI or diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    This is not applicable. This device is a surgical instrument, not an algorithm.

    7. The type of ground truth used
    The "ground truth" for this device's performance is based on:

    • Engineering standards (e.g., ISO, ASTM, IEC for sterilization, packaging, electrical safety).
    • Biocompatibility standards (ISO 10993-1).
    • Predefined physical and biological thresholds/measurements from ex vivo (vessel/lymphatic sealing/burst) and in vivo (hemostasis, thermal spread) studies.
    • "Clinical literature studies were evaluated to further support the safety and effectiveness use of the proposed LigaSure™ XP Maryland Jaw Sealer/Divider indications for use." This implies that existing clinical evidence for similar devices and principles contributes to the overall justification, but no new clinical study data from the sponsor is explicitly referenced.

    8. The sample size for the training set
    This is not applicable. This is a hardware device; there is no "training set" in the context of machine learning.

    9. How the ground truth for the training set was established
    This is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1