Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K142430

    Validate with FDA (Live)

    Date Cleared
    2014-12-19

    (112 days)

    Product Code
    Regulation Number
    870.5310
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Manual Defibrillation:
    Indications: Manual defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of this energy in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and, in relatively stable patients, ventricular tachycardia.
    Contraindications: Defibrillation is contraindicated in the treatment of Pulseless Electrical Activity (PEA), such as idioventricular or ventricular escape rhythms, and in the treatment of asystole.

    Automated External Defibrillation:
    Indications: AED mode is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, pulseless, and not breathing normally before using the defibrillator to analyze the patient's ECG rhythm. In AED mode, the LIFEPAK 15 monitor/defibrillator is not intended for use on pediatric patients less than 8 years old.

    Noninvasive Pacing:
    Indications: Noninvasive pacing is indicated for symptomatic bradycardia in patients with a pulse.
    Contraindications: Noninvasive pacing is contraindicated for the treatment of ventricular fibrillation and asystole.

    12-lead Electrocardiography:
    Indications: The 12-lead electrocardiogram is used to identify, diagnose and treat patients with cardiac disorders and is useful in the early detection and prompt treatment of patients with acute ST-elevation myocardial infarction (STEMI).

    Pulse Oximetry:
    Indications: Pulse Oximetry is indicated for use in any patient who is at risk of developing hypoxemia, carboxyhemoglobinemia, or methemoglobinemia. SpO2 monitoring may be used during no motion and motion conditions, and in patients who are well or poorly perfused. SpCO and SpMet accuracies have not been validated under motion or low perfusion conditions.

    Noninvasive Blood Pressure Monitoring:
    Indications: Noninvasive blood pressure monitoring is intended for detection of hypertension or hypotension and monitoring BP trends in patient conditions such as, but not limited to, shock, acute dysrhythmia, or major fluid imbalance.

    End-Tidal CO2 monitoring:
    Indications: EtCO2 monitoring is used to detect trends in the level of expired CO2. It is used for monitoring breathing efficacy and treatment effectiveness in acute cardiopulmonary care, for example, to determine if adequate compressions are being performed during CPR or to rapidly detect whether an endotracheal tube has been placed successfully.

    Invasive Pressure Monitoring:
    Indications: Invasive pressure monitoring is indicated for use in patients who require continuous monitoring of physiological pressures in order to rapidly assess changes in the patient's condition or response to therapy. It may also be used to aid in medical diagnosis.

    Temperature Monitoring:
    Indications: Temperature monitoring is indicated for use in patients who require continuous monitoring of body temperature.

    Device Description

    The LIFEPAK 15 monitor/defibrillator is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. The LIFEPAK 15 monitor/defibrillator was designed for use in a variety of hospital and pre-hospital settings including emergency rooms, catheterization laboratories, electrophysiology laboratories, crash carts, operating rooms, and ground ambulances. Features of the LIFEPAK 15 monitor/defibrillator include manual and automated external defibrillation, noninvasive pacing, ECG monitoring (3-lead. 7-lead and interpretive 12-Lead), pulse oximetry (SpO2, SpCO, and SpMet), synchronized cardioversion, noninvasive blood pressure monitoring, end-tidal CO2 monitoring, invasive pressure monitoring, and temperature monitoring. The LIFEPAK 15 monitor/defibrillator is powered by rechargeable lithium-ion batteries or from AC power sources via an AC power adapter or DC power sources via a DC power adapter. The primary difference between the proposed LIFEPAK 15 monitor/defibrillator and the previously cleared predicate device is a combination of software and hardware modifications completed to support component obsolescence. The proposed LIFEPAK 15 monitor/defibrillator includes the same monitoring features, defibrillation waveform, pacing waveform, and Shock Advisory System™ algorithm as the previously cleared predicate device. Additionally, there are no changes to the intended use or indications for use of the previously cleared predicate device.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for the Physio-Control LIFEPAK 15 monitor/defibrillator. The document focuses on demonstrating substantial equivalence to a previously cleared device, primarily due to software and hardware modifications for component obsolescence. It does not provide detailed acceptance criteria or a specific study demonstrating performance against such criteria for the entire device's functionalities in the context of an AI/algorithm-driven medical device submission.

    However, based on the information provided regarding the "Shock Advisory System™ algorithm," we can infer the aspects that would typically involve performance criteria for an automated external defibrillator (AED) algorithm.

    Here's an attempt to answer your questions based on the available information, noting that much of the specific detail for AI/algorithm performance is not present in this 510(k) submission, as it predates the widespread regulatory focus on standalone AI algorithm performance studies. The Shock Advisory System™ algorithm mentioned is likely a rule-based algorithm rather than a modern AI/ML algorithm.

    1. A table of acceptance criteria and the reported device performance

    The document states: "The proposed LIFEPAK 15 monitor/defibrillator includes the same monitoring features, defibrillation waveform, pacing waveform, and Shock Advisory System™ algorithm as the previously cleared predicate device."

    This implies that the performance of the Shock Advisory System™ algorithm is assumed to be equivalent to the predicate device. For AED algorithms, typical acceptance criteria would involve sensitivity (true positive rate for shockable rhythms) and specificity (true negative rate for non-shockable rhythms). While specific numerical acceptance criteria are not presented in this document for the current submission, for similar AED algorithms, these often fall within ranges like:

    Performance MetricAcceptance Criteria (Typical for AED Algorithms)Reported Device Performance (Implied from Substantial Equivalence)
    Shockable Rhythm Sensitivity (e.g., Ventricular Fibrillation)> 90% or >95%Substantially equivalent to predicate (K082937, K103567), which met standards at their time of clearance.
    Non-Shockable Rhythm Specificity (e.g., Asystole, Normal Sinus Rhythm)> 90% or >95%Substantially equivalent to predicate (K082937, K103567), which met standards at their time of clearance.

    Study Proving Acceptance Criteria:
    The document states: "No human clinical studies were submitted as part of this 510(k) Premarket Notification." Instead, the submission relies on "Performance Testing" including:

    • Design Requirements Testing
    • Hardware Verification
    • Software Performance
    • Electrical Safety and Electromagnetic Compatibility
    • Design Validation via Animal Studies and Simulated Use Testing

    The core claim is substantial equivalence to previously cleared LIFEPAK 15 monitor/defibrillators (K082937, K103567). The performance of the Shock Advisory System™ algorithm is explicitly stated as unchanged from the predicate device. Therefore, the "study" proving acceptance criteria for the algorithm's performance would have been conducted for the predicate devices, and the current submission relies on that prior clearance and the assertion that the algorithm itself has not changed.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not provided in this document. Information on test set size and data provenance for the Shock Advisory System™ algorithm would have been part of the original 510(k) submission for the predicate devices. This document explicitly states: "No human clinical studies were submitted as part of this 510(k) Premarket Notification."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not provided in this document. This information would be specific to the ground truth establishment for the predicate device's algorithm validation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not provided in this document. This information would be specific to the ground truth establishment for the predicate device's algorithm validation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done or reported in this 510(k) submission. The Shock Advisory System™ algorithm in an AED is typically a standalone algorithm designed to automatically detect shockable rhythms, rather than an AI-assisted interpretation tool for human readers in the context of MRMC studies. The device is intended for use by trained medical personnel, including those using the "Automated External Defibrillation" mode, which implies the algorithm is acting as a decision-maker (shock/no shock) rather than an assistant to a human interpreter in the sense of a diagnostic imaging AI.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, for the Shock Advisory System™ algorithm, its function in AED mode is inherently standalone in its determination of whether a rhythm is shockable or not. The document states: "Automated External Defibrillation: AED mode is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, pulseless, and not breathing normally before using the defibrillator to analyze the patient's ECG rhythm." This "analysis" is performed by the algorithm. The performance of this standalone algorithm would have been assessed during the predicate device's clearance. This 510(k) asserts the algorithm is unchanged.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not explicitly stated in this document. For AED algorithms, ground truth for rhythm classification (e.g., ventricular fibrillation, asystole, normal sinus rhythm) is typically established by:

    • Expert Consensus on ECG Tracings: Review by cardiologists or emergency physicians.
    • Simulated Rhythms: Using synthesized or recorded rhythm libraries with known classifications.
    • Clinical Outcomes/Events: Correlation with actual patient response to therapy, though less direct for algorithm validation.

    It's highly probable that expert consensus on ECG tracings was used for the predicate device.

    8. The sample size for the training set

    Not provided in this document. This information would be specific to the development and validation of the Shock Advisory System™ algorithm for the predicate devices. Modern AI/ML algorithms often have very large training sets; for the kind of algorithms in AEDs cleared in 2009/2011, training sets might have been smaller, focusing on diverse representations of specific arrhythmias.

    9. How the ground truth for the training set was established

    Not provided in this document. Similar to question 7, ground truth for the training set (if applicable, as the algorithm might be rule-based rather than machine learning trained) would have been established by expert review of ECG tracings or use of labeled rhythm databases.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103567

    Validate with FDA (Live)

    Date Cleared
    2011-03-22

    (106 days)

    Product Code
    Regulation Number
    870.5310
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in outdoor and indoor emergency care settings within the environmental conditions specified. The LIFEPAK 15 monitor/defibrillator is designed to be used during ground transportation except when specified otherwise. Manual mode monitoring and therapy functions are intended for use on adult and pediatric patients.

    Automated external defibrillation mode is intended for use on patients eight years of age and older.

    Indications for Use:

    Manual Defibrillation:
    Indications
    Manual defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of this energy in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and, in relatively stable patients, ventricular tachycardia.
    Contraindications
    Defibrillation is contraindicated in the treatment of Pulseless Electrical Activity (PEA), such as indioventricular or ventricular escape rhythms, and in the treatment of asystole.

    Automated External Defibrillation:
    Indications
    AED mode is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, pulseless, and not breathing normally before using the defibrillator to analyze the patient's ECG rhythm. In AED mode, the LIFEPAK 15 monitor/defibrillator is not intended for use on pediatric patients less than 8 years old.

    Noninvasive Pacing
    Indications
    Noninvasive pacing is indicated for symptomatic bradycardia in patients with a pulse.
    Contraindications: Noninvasive pacing is contraindicated for the treatment of ventricular fibrillation and asystole.

    12-lead Electrocardiography:
    Indications
    The 12-lead electrocardiogram is used to identify, diagnose and treat patients with cardiac disorders and is useful in the early detection and prompt treatment of patients with acute ST-elevation myocardial infarction (STEMI).

    Pulse Oximetry
    Indications
    Pulse Oximetry is indicated for use in any patient who is at risk of developing hypoxemia, carboxyhemoglobinemia, or methemoglobinemia. SpO2 monitoring may be used during no motion and motion conditions, and in patients who are well or poorly perfused. SpCO and SpMet accuracies have not been validated under motion or low perfusion conditions.

    Noninvasive Blood Pressure Monitoring:
    Indications
    Noninvasive blood pressure monitoring is intended for detection of hypertension or hypotension and monitoring BP trends in patient conditions such as, but not limited to, shock, acute dysrhythmia, or major fluid imbalance.

    End-Tidal CO2 monitoring:
    Indications
    EtCO2 monitoring is used to detect trends in the level of expired CO2. It is used for monitoring breathing efficacy and treatment effectiveness in acute cardiopulmonary care, for example, to determine if adequate compressions are being performed during CPR or to rapidly detect whether an endotracheal tube has been placed successfully.

    Invasive Pressure Monitoring:
    Indications
    Invasive pressure monitoring is indicated for use in patients who require continuous monitoring of physiological pressures in order to rapidly assess changes in the patient's condition or response to therapy. It may also be used to aid in medical diagnosis.

    Temperature Monitoring:
    Indications
    Temperature monitoring is indicated for use in patients who require continuous monitoring of body temperature.

    Device Description

    The LIFEPAK 15 monitor/defibrillator is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. The LIFEPAK 15 monitor/defibrillator was designed for use in a variety of hospital and pre-hospital settings including emergency rooms, catheterization laboratories, electrophysiology laboratories, crash carts, operating rooms, and ground ambulances.

    Features of the LIFEPAK 15 monitor/defibrillator include manual and automated external defibrillation, noninyasive pacing, ECG monitoring (3-lead, 7-lead and interpretive 12-Lead), pulse oximetry (SpO2, SpCO. and SpMet), synchronized cardioversion, noninvasive blood pressure monitoring, end-tidal CO2 monitoring and invasive pressure monitoring. The existing LIFEPAK 15 monitor/defibrillator is powered by rechargeable lithium-ion batteries.

    The proposed LIFEPAK 15 monitor/defibrillator includes all of the same features and functions as the predicate device plus 2 new features: 1) temperature monitoring and 2) auxiliary power. The temperature monitoring feature allows continuous monitoring of body temperature with the use of temperature probes. The temperature reading can be trended and displayed on the LIFEPAK 15 monitor/defibrillator screen. The temperature trend graph can also be printed or saved and transmitted to the Physio-Control CODE-STAT" software for post event review.

    The new auxiliary power feature gives users an additional option for powering the LIFEPAK 15 monitor/defibrillator. In addition to the existing rechargeable lithium-ion batteries, users will have the option to operate their device from AC power sources via an AC power adapter or from DC power sources via a DC power adapter. The LIFEPAK 15 monitor/defibrillator charges installed batteries when connected to auxiliary power.

    AI/ML Overview

    The provided text does not contain detailed information about specific acceptance criteria or a dedicated study proving the device meets these criteria for the new features (temperature monitoring and auxiliary power). Instead, it states: "The information in this 510(k) notification demonstrates that the LIFEPAK 15 monitor/defibrillator is substantially equivalent to the predicate LIFEPAK 15 device with respect to safety, effectiveness, and performance." This implies that the new features are considered substantially equivalent to existing, cleared technologies and thus a comprehensive study with acceptance criteria might not have been performed or documented in this summary.

    Here's a breakdown of what can be inferred and what is missing, based on your requested information:

    1. A table of acceptance criteria and the reported device performance:

    This information is not provided in the document. The document primarily focuses on describing the device, its intended use, and its equivalence to a predicate device.

    2. Sample size used for the test set and the data provenance:

    This information is not provided. As no specific study proving the device meets acceptance criteria for the new features is detailed, there's no mention of a test set, its sample size, or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable as the device is a monitor/defibrillator adding temperature monitoring and auxiliary power, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    This is not applicable as the device is a medical monitoring and intervention device, not an algorithm being tested in a standalone capacity.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    This information is not provided.

    8. The sample size for the training set:

    This information is not provided.

    9. How the ground truth for the training set was established:

    This information is not provided.

    Summary of available information regarding acceptance criteria and studies:

    The document concludes with: "The information in this 510(k) notification demonstrates that the LIFEPAK 15 monitor/defibrillator is substantially equivalent to the predicate LIFEPAK 15 device with respect to safety, effectiveness, and performance." This statement suggests that the new features (temperature monitoring and auxiliary power) were likely evaluated against established standards for vital sign monitoring and power supply systems, rather than requiring a new, extensive clinical study with specific acceptance criteria outlined in this summary. The "substantial equivalence" argument is the primary method used to demonstrate that the device with these new features is safe and effective.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082937

    Validate with FDA (Live)

    Date Cleared
    2009-03-11

    (161 days)

    Product Code
    Regulation Number
    870.5310
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in out-of doors and indoor emergency care settings within the environmental conditions specified. The LIFEPAK 15 monitor/defibrillator is designed to be used during ground transportation except when specified otherwise. Manual mode monitoring and therapy functions are intended for use on adult and pediatric patients. Automated external defibrillation mode is intended for use on patients eight years of age and older.

    Manual defibrillation is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of this energy in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and, in relatively stable patients, ventricular tachycardia.

    AED mode is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, pulseless, and not breathing normally before using the defibrillator to analyze the patient's ECG rhythm. In AED mode, the LIFEPAK 15 monitor/defibrillator is not intended for use on pediatric patients less than eight years old.

    Noninvasive pacing is indicated for symptomatic bradycardia in patients with a pulse.

    The 12-lead electrocardiogram is used to identify, diagnose and treat patients with cardiac disorders and is useful in the early detection and prompt treatment of patients with acute ST-elevation myocardial infarction (STEMI).

    Pulse oximetry is indicated for use in any patient who is at risk of developing hypoxemia, carboxyhemoglobinemia, or methemoglobinemia. SpO2 monitoring may be used during no motion and motion conditions, and in patients who are well or poorly perfused. SpCO and SpMet accuracies have not been validated under motion or low perfusion conditions.

    Noninvasive blood pressure monitoring is intended for detection of hypertension or hypotension and monitoring BP trends in patient conditions such as, but not limited to, shock, acute dysrhythmia, or major fluid imbalance.

    EtCO2 monitoring is used to detect trends in the level of expired CO2. It is used for monitoring breathing efficacy and treatment effectiveness in acute cardiopulmonary care, for example, to determine if adequate compressions are being performed during CPR or to rapidly detect whether an endotracheal tube has been placed successfully.

    Invasive pressure monitoring is indicated for use in patients who require continuous monitoring of physiological pressures in order to rapidly assess changes in the patient's condition or response to therapy. It may also be used to aid in medical diagnosis.

    Device Description

    The LIFEPAK 15 monitor/defibrillator is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. The LIFEPAK 15 monitor/defibrillator product is a next generation product to the currently legally marketed LIFEPAK 12 defibrillator/ monitor product. The LIFEPAK 15 monitor/defibrillator design is significantly based upon the existing LIFEPAK 12 architecture and when compared to the LIFEPAK 12 predicate it has similar performance characteristics, features, and accessories.

    AI/ML Overview

    The provided text from K082937 describes a 510(k) submission for the LIFEPAK® 15 Monitor/Defibrillator. It states that the device's features and functions are "substantially equivalent" to previously cleared predicate devices (specifically the LIFEPAK 12 defibrillator/monitor). This type of submission relies on demonstrating that the new device is as safe and effective as a legally marketed predecessor, rather than presenting new clinical study data with specific acceptance criteria and performance metrics for an AI algorithm.

    Therefore, the requested information regarding acceptance criteria, device performance tables, sample sizes for test and training sets, expert qualifications, and adjudication methods for a study proving an AI device meets acceptance criteria cannot be extracted from this document, as it outlines a substantial equivalence claim for a traditional medical device, not an AI/ML software.

    The document highlights:

    • Predicate Device Comparison: The core of the submission is the comparison to the LIFEPAK 12 for demonstrating substantial equivalence.
    • Verification and Validation: It mentions "documentation related to the verification and validation of the LIFEPAK 15 monitor/defibrillator," which typically refers to engineering and system-level testing, not clinical studies in the context of AI performance.
    • Intended Use and Indications/Contraindications: These are provided for the various monitoring and therapy functions of the device (e.g., manual defibrillation, AED, noninvasive pacing, 12-lead ECG, pulse oximetry, NIBP, EtCO2, invasive pressure monitoring).

    In summary, this document does not contain the information needed to fill out the requested table and answer the specific questions related to the performance of an AI/ML algorithm. The regulatory pathway chosen (510(k) based on substantial equivalence to a predicate device) for this product typically does not require the extensive clinical efficacy studies with detailed acceptance criteria and performance metrics that would be associated with a novel AI-powered diagnostic or therapeutic device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1