    510k Summary Text (Full-text Search) :

    15238

    Re: K223249

    Trade/Device Name: Pro-Tx Endorectal Balloon (PROT-25) Regulation Number: 21 CFR 892.5720
    Classification Name: | Prostate Immobilizer Rectal Balloon |
    | Regulation number: | 21 CFR 892.5720
    Classification Name: | Prostate Immobilizer Rectal Balloon |
    | Regulation number: | 21 CFR 892.5720
    Both devices meet the regulatory definition for Prostate Immobilizer Rectal Balloon outlined in 21 CFR 892.5720
    bench testing of the subject device to demonstrate the ability to meet the Special Controls in 21 CFR 892.5720

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DxTx Medical Pro-Tx Endorectal Balloon is a disposable, single use non-powered inflatable positioning device intended to be used for temporary positioning of the rectal wall and adjacent structures in the male human anatomies and to assist in positioning the prostate in a more predictable and reproducible location.

    The purpose of the DxTx Medical Pro-Tx Endorectal Balloon is to stabilize the rectal wall and surrounding anatomy during all phases of radiation therapy, including treatment planning, image verification, and radiotherapy delivery. Only trained healthcare professionals are intended to operate this device.

    Device Description

    The DxTx Medical Pro-Tx Endorectal Balloon is a disposable, single use device for immobilization and localization of the rectal wall and surrounding anatomy on a daily basis for radiation therapy treatment. The Pro-Tx Endorectal Balloon includes the shaft and inflatable balloon with a gas venting feature. The Pro-Tx Endorectal Balloon is capable of being filled with air or fluid. The intended contact duration of the device is less than 60 minutes.

    No components of this device contain medicinal substances, tissues or blood products.

    The device component materials are PVC, Polycarbonate, Polyethylene, Polypropylene, Copolyester, Liquid Silicone Rubber (balloon), and brass (the radiopaque marker).

    AI/ML Overview

    This FDA 510(k) summary describes the Pro-Tx Endorectal Balloon (PROT-25) and its equivalence to a predicate device, the RadiaDyne Prostate Immobilizer Rectal Balloon.

    Here's an analysis of the provided information relating to acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly list "acceptance criteria" for each performance test in a quantitative manner. Instead, it states that the performance tests were completed "to demonstrate performance equivalent to the predicate device and in compliance with 21 CFR 892.5720." This implies that the acceptance criteria are met if the new device performs at least as well as the predicate device across these specific tests, or adheres to regulatory standards where applicable.

    Based on the "Effectiveness" section in the comparison table and the "Performance Verification (Bench Testing)" section, here's a reconstructed table reflecting the tests performed:

    Feature/TestPredicate Device Performance Criteria (Implied)DxTx Medical Device Performance (Reported)
    BioburdenPerformance achieved (from De Novo study)Performance achieved
    Leakage/Inflation IntegrityPerformance achieved (from De Novo study)Performance achieved (Inflation Integrity)
    Tensile Strength/Axial PullPerformance achieved (from De Novo study)Performance achieved (Axial Pull)
    Burst StrengthPerformance achieved (from De Novo study)Performance achieved
    Depth Stopper ResistancePerformance achieved (from De Novo study)Performance achieved
    Vent/OcclusionPerformance achieved (from De Novo study)Performance achieved (Occlusion)
    3' Drop TestNot explicitly listed for predicate (implied suitable performance)Performance achieved
    45° Shaft FlexureNot explicitly listed for predicate (implied suitable performance)Performance achieved
    BiocompatibilityNo issues identified/documented safety issues (for predicate's design lineage)No issues identified; no biocompatibility risks with the Pro-Tx device; post-market performance of legacy Pro-Tekt device had no documented safety issues.
    Shelf-lifeDemonstrated 2-year shelf lifeDemonstrated 2-year shelf life after accelerated aging

    2. Sample size used for the test set and the data provenance

    • Sample Size for Shelf-life: A box of 25 Pro-Tx devices was subjected to accelerated aging for the shelf-life study.
    • Sample Size for other Bench Tests: Not explicitly stated for individual bench tests (Axial pull, inflation integrity, burst strength, occlusion, drop test, flexure test). Typically, such tests would involve a representative sample size from a production lot.
    • Data Provenance: The accelerated aging study was conducted by an "independent ISO 17025-certified lab." Other bench test data are implied to be from internal testing by DxTx Medical, Inc. or associated labs. The predicate device's performance data likely came from its original De Novo study and post-market surveillance. The biocompatibility assessment references the design lineage from MEDRAD/Bayer Healthcare and its post-market performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the studies presented are primarily non-clinical bench and biocompatibility tests, not clinical performance studies requiring expert ground truth for classification or diagnosis. The "ground truth" for these tests relates to engineering specifications and material properties.

    4. Adjudication method for the test set

    This information is not applicable for the same reason as point 3. Adjudication methods are typically used in clinical studies involving multiple expert readers.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. This device is a physical medical device (an endorectal balloon) for positioning and stabilization during radiation therapy, not an AI-powered diagnostic or assistive tool. Therefore, the concept of human readers improving with AI assistance is not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable as the device is a physical, non-powered positioning device and does not involve any algorithm or AI for standalone performance evaluation.

    7. The type of ground truth used

    The "ground truth" for the performance tests is based on engineering specifications, material properties, and regulatory standards (21 CFR 892.5720). For biocompatibility, the ground truth refers to established biocompatibility guidelines and the absence of adverse events from the legacy device over a long period. For shelf-life, the ground truth is the maintenance of performance specifications after accelerated aging.

    8. The sample size for the training set

    This information is not applicable. There is no "training set" as this is not an AI/machine learning device. The design was based on a previously marketed device (MEDRAD/Bayer Healthcare Pro-Tekt Endorectal Balloon).

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180478

    Validate with FDA (Live)

    Date Cleared
    2018-11-13

    (264 days)

    Product Code
    Regulation Number
    892.5720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    TX 78746

    Re: K180478

    Trade/Device Name: RectalPro™ 75 Endorectal Balloon Regulation Number: 21 CFR 892.5720
    | Prostate immobilizer rectal balloon |
    | | Regulation Number: | 21 CFR 892.5720
    |
    | RegulationNumber | 21 CFR 892.5720
    | 21 CFR 892.5720

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RectalPro™15 Endorectal Balloon device is a single-use disposable, inflatable, non-powered positioning device intended for us in the temporary positioning of the rectal wall and adjacent structure in the male human anatomies. The purpose of the device is to stabilize the prostate during Computed Tomography (CT) exam, and X-Ray when these imaging techniques are used for Radiation Therapy (RT) planning.

    Device Description

    The RectalPro™ 75 Endorectal Balloon (ERB) is single use, disposable, inflatable, non-powered rectal device placed in the rectum to immobilize the prostate in patients undergoing radiation therapy. The device is intended to be used during all the phases of radiation therapy, including treatment planning, image verification, and radiotherapy delivery. The RectalPro75 is designed as an immobilizer to assist in positioning the prostate in a more predictable and reproducible location during Computed Tomography (CT) exam and X-ray, when these imaging techniques are used for Radiation Therapy (RT) planning. The ERB is inserted into the rectum and inflated prior to the start of a CT scan or RT therapy procedure. The device stabilizes the prostate once the device is inflated. The ERB is deflated and removed after each individual scan or therapy procedure is complete, and a new balloon is used in the next therapy session. The RectalPro ERB is designed with a slim curved, extended shaft and balloon. The device is provided non-sterile to the end user, is not intended to be sterilized by the end user. It is packaged in a kit configuration, consisting of three primary components: 1. ERB shaft This is the shaft with attached balloon and attached clear plastic tubing for filling the balloon with air / water. The end of the tubing has a female luer lock. There are white markings on the side of the ERB shaft as insertion depth markings. 2. Stopper This is the circular plastic piece with one flat side, and one side containing four prongs. There is a hole in the middle of the stopper for placement onto the ERB shaft. 3. Air / water valve This is the small plastic valve on the air tube connected to the ERB shaft. The clamp on the air / water tube stops the air / water flow by squeezing the tube closed. This component stops air / water from releasing out of the balloon after it has been filled. The kit also includes a standard 100 cc syringe with male luer lock connector, for connecting to the female luer connection of the ERB tubing, and is used to deliver the water or air to the ERB.

    AI/ML Overview

    The provided text describes the RectalPro™75 Endorectal Balloon, a device intended for prostate immobilization during radiation therapy planning. It is a 510(k) submission, meaning the manufacturer is claiming substantial equivalence to a legally marketed predicate device, not necessarily proving novel safety and effectiveness through extensive clinical trials.

    Here's an analysis of the acceptance criteria and the study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily derived from the performance of the predicate device and relevant international standards. The reported device performance is presented as meeting these criteria.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Reference/Standard)Reported RectalPro™75 Endorectal Balloon Performance
    BiocompatibilityDevice determined to be biocompatible (per ISO 10993-1)Device determined to be biocompatible. (Materials PVC with Blue colorant; Silicone; Polypropylene were tested per ISO 10993 and found to be biocompatible).
    BioburdenBioburden < 100 CFU (per ANSI/AAMI/ISO 11737-1) or comparable to predicate. Predicate: Bioburden < 100 CFU.Device met acceptance criteria. (Specifically, Bioburden < 100CFU reported).
    Leakage (Water)No leaking with 100cc water after 10 min (Predicate performance for comparison). RectalPro™75 internal protocol criteria: No leaking with 110cc water after 60 min.Device met acceptance criteria. (Specifically, no leaking with 110cc water after 60 min reported).
    Tensile (Burst) (Air)Did not burst with 100cc air under 7.1 lbs force (Predicate performance for comparison). RectalPro™75 internal protocol criteria: Did not burst with 140cc air under 7.7 lbs force for 12 min.Device met acceptance criteria. (Specifically, did not burst with 140cc air under 7.7 lbs force for 12 min reported).
    Stopper ResistanceForce required to move was 4.91 lbs average (Predicate performance for comparison). RectalPro™75 internal protocol criteria: Stopper did not move under 10 lbs of force.Device met acceptance criteria. (Specifically, stopper did not move under 10 lbs of force reported).
    TransportationMet requirements (per ASTM D4169:2016)Device met requirements.
    Risk AssessmentAcceptable residual risks (per ISO 14971)Device has acceptable residual risks.
    Shelf LifeSimilar to predicate (predicate has 2 years).2.5 years. (This is similar and actually longer than the predicate, indicating comparable or better stability).

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text discusses non-clinical bench testing. There is no mention of a "test set" in the context of clinical data or patient studies. The "test set" in this context refers to the samples of the device used for the various bench tests (e.g., individual balloons for leak, burst, and stopper resistance tests). The sample sizes for these specific bench tests are not explicitly stated in the document, only the results for a single device or a general statement of "device met acceptance criteria."

    The data provenance is bench testing conducted by QLRAD International Ltd. (the manufacturer). There is no mention of country of origin for any clinical data, as no clinical studies are reported.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This question is not applicable. The "test set" here refers to the device and its components undergoing engineering and material performance tests, not clinical performance read by experts to establish a "ground truth" for a diagnostic or imaging device. The "ground truth" for these tests would be the measurement results against predefined engineering specifications.

    4. Adjudication Method for the Test Set

    This is not applicable for bench testing. Adjudication methods (like 2+1, 3+1) are for resolving discrepancies in expert interpretations of clinical data, which is not the nature of the testing described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was mentioned or performed. The submission is a 510(k) for a device that stabilizes the prostate, not a diagnostic or imaging AI tool requiring human-in-the-loop performance evaluation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No standalone performance study, as this term typically applies to AI algorithms, was mentioned or performed. The device is a physical medical device.

    7. Type of Ground Truth Used

    For the non-clinical bench testing, the "ground truth" is established by engineering specifications and measurements performed on the device samples. For example:

    • Biocompatibility: Results of standardized in vitro/in vivo tests against ISO 10993.
    • Bioburden: Microbial count below a specified threshold (100 CFU).
    • Leakage/Burst: Absence of leaks or rupture under specific pressure/volume/time conditions.
    • Stopper Resistance: Force required to move the stopper meeting a defined threshold.
    • Transportation: Device integrity after simulated transport according to ASTM D4169.

    8. Sample Size for the Training Set

    This is not applicable. There is no mention of any AI or machine learning algorithms, and therefore, no "training set."

    9. How the Ground Truth for the Training Set was Established

    This is not applicable, as there is no training set for an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162954

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-06-01

    (220 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | §892.5050§892.5720

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OARtrac® System with patient specific, reusable, pre-calibrated PSD sensors are intended for use during cancer treatments to measure photon radiation therapy as an adjunct to treatment planning permitting measurement and validation of radiation dose received by the patient to the targeted area of their body, and indicated for use when adhered to the skin with a bolus, or inserted into the rectum to measure the rectal prostatic interface via a specifically designed endorectal balloon device.

    Device Description

    The OARtrac® System with patient specific, reusable, pre-calibrated PSD sensors is intended for use in photon radiation therapy to monitor and validate radiation dose during External Beam Therapy and HDR Brachytherapy to the surface of the skin or the rectal prostatic interface. This dose verification information obtained during the treatment is then used to compare with the planned dose that the Radiation Oncologists expect to provide to their patient. The OARtrac® System itself does not stop the radiation treatment to the patient, or change the radiation delivery, but only provides dose data which a trained Radiation Oncologist can decipher and use to adjust a patient's treatment plan accordingly.

    AI/ML Overview

    The provided text does not contain information about an Artificial Intelligence (AI) device or a study proving its performance against acceptance criteria in the context of AI. The document is a 510(k) premarket notification for a medical device called the "OARtrac® System with Patient Specific Reusable PSD Sensors," which is a radiation dose verification system.

    The document focuses on demonstrating substantial equivalence to previously cleared predicate devices, primarily through non-clinical performance data related to material, electrical safety, EMC, software, package shelf-life, risk analysis, dose range verification, and importantly, cleaning and disinfection validation for reusability.

    Therefore, I cannot provide the requested information regarding AI device acceptance criteria, study details, sample sizes, ground truth establishment, or multi-reader multi-case studies, as these concepts are not addressed in the provided text.

    Specifically:

    1. A table of acceptance criteria and the reported device performance: The document lists "Key Performance Specifications/Characteristics" for the device, which includes:

      • Photon Energy Based Therapies: Energy Range .37-18 MeV
      • Dose Rate Range: 1.3-17.3 cGy/s
      • Dose Range: 27-1200 cGy
      • Dose Accuracy: +/-6%, 2 σ
        It states that the subject device meets these requirements, and that "all the testing" was passed. However, it does not provide detailed performance results in a table format comparing acceptance criteria to reported performance for each specific test item in the same way an AI model's performance metrics (e.g., sensitivity, specificity, accuracy) would be presented against pre-defined thresholds. The key focus of this submission is on the reusability validation.
    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable, as this is a non-AI device. The testing described is primarily laboratory-based non-clinical performance validation (e.g., electrical safety, EMC, cleaning/disinfection validation). Specific sample sizes for these types of engineering tests are not detailed in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. "Ground truth" in the context of expert consensus is relevant for AI diagnostic or prognostic devices. For this device, "ground truth" would relate to verifiable physical measurements in a laboratory setting for dose accuracy, cleaning efficacy, etc. The document does not specify experts for establishing such ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for a non-AI device's non-clinical performance testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For the non-clinical tests mentioned, the "ground truth" would be established by standard engineering and scientific methodologies (e.g., calibrated reference instruments for dose measurement, laboratory testing protocols for cleaning/disinfection efficacy, an accredited laboratory for biocompatibility).

    8. The sample size for the training set: Not applicable, as this is not an AI device and thus has no training set in that context.

    9. How the ground truth for the training set was established: Not applicable, as this is not an AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150719

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-06-16

    (89 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    | §892.5050§892.5720

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OARtrac® System pre-calibrated skin sensors are specifically indicated for use during cancer treatments to measure photon beam therapy as an adjunct to treatment planning permitting measurement of radiation dose received on the surface of the skin. OARtrac® System pre-calibrated skin sensors are indicated for use when adhered to the skin using medical grade adhesive and with a medical grade bolus buildup placed directly on top of the sensor.

    Device Description

    The OARtrac® System with Skin Sensors provides Radiation Oncologists with near real-time, multi-point radiation-dose information obtained from two (2) Radiatrac® Plastic Scintillating Detectors (PSD) located on the surface of the patient's skin to monitor dose photon based radiation therapy for cancer treatment. This information allows the physician to monitor the dose at the skin surface, compare the actual dose relative to the planned dose, and provides graphs and dose information for the current treatment as well as a log of the dose from five previous treatments. The actual verification of the dose radiation is accomplished by the other main components of the OARtrac® System, those being the Clinical Detector Unit (CDU) with its Charged Coupled Device (CCD) camera and the system's own proprietary dose management software.

    AI/ML Overview

    The provided document, a 510(k) summary for the OARtrac® System with Skin Sensors, focuses on demonstrating substantial equivalence to a predicate device rather than detailing specific acceptance criteria and a study proving those criteria are met for a standalone device. The device is a radiation dose verification system that measures photon beam therapy on the surface of the skin.

    Here's an analysis based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics in the format typically used for a standalone effectiveness study. Instead, it relies on demonstrating equivalence to a predicate device through various non-clinical tests.

    However, based on the description of performance testing, we can infer some implied acceptance criteria and reported "performance" in the context of equivalence:

    Acceptance Criteria (Implied)Reported Device Performance
    Overall design requirements metPassed all testing in accordance with national and international standards
    BiocompatibilityPassed Biocompatibility Testing per ISO 10993-1 (Parts 5, 10 and 11)
    Package Shelf lifePassed Package Shelf-Life per ASTM F1980-07
    Electrical safetyPassed Electrical Safety per IEC 60601-1
    EMC (Electromagnetic Compatibility)Passed EMC per IEC 60601-1-2
    Software Verification and ValidationPassed Software Verifications and Validation per IEC 62304
    Device Risk AnalysisPassed Device Risk Analysis per ISO 14971
    Dose Range VerificationResults acceptable to current clinical standards when simulating treatments from a standard LINAC machine and the Accuray CyberKnife system.
    Accuracy compared to predicate device (K141154)Accurate to that within the established accuracy of the original OARtrac® System cleared under K141154.
    Ship Testing CalibrationPassed Ship Testing Calibration

    2. Sample Size used for the test set and the data provenance

    • Sample Size for Test Set: The document does not specify a "test set" in the context of a dataset of patient images or outcomes for evaluating an AI algorithm. The testing described is primarily non-clinical, involving laboratory-based simulations and physical testing of the device. Therefore, a sample size of "patients" or "cases" is not applicable in this context.
    • Data Provenance: Not applicable. The testing is described as being performed in a laboratory setting, simulating radiation treatments.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The device measures radiation dose, and its accuracy is assessed against established physical standards and the performance of a predicate device, not through expert consensus on medical images or diagnoses.

    4. Adjudication method for the test set

    Not applicable, as there is no test set requiring expert adjudication for ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done. This device is a measurement tool for radiation dose, not an AI-assisted diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document does not describe the device as containing an "algorithm" in the AI sense. It is a measurement system. The "standalone" performance was assessed through non-clinical laboratory testing to verify its measurement capabilities. The performance referred to is the device itself performing its measurement function.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the non-clinical performance data appears to be:

    • Physical standards and engineering specifications: For aspects like biocompatibility, electrical safety, EMC, software verification, risk analysis, and package shelf-life, the ground truth is adherence to established national and international standards (e.g., ISO, IEC, ASTM).
    • Established accuracy of the predicate device: For dose range verification, the ground truth is the established accuracy of the original OARtrac® System (K141154) and "current clinical standards" for simulating treatments from LINAC and CyberKnife machines.

    8. The sample size for the training set

    Not applicable. This device is not described as utilizing machine learning or AI that requires a "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for the device's functionality.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150234

    Validate with FDA (Live)

    Date Cleared
    2015-03-31

    (57 days)

    Product Code
    Regulation Number
    892.5720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Name: Myriad Prostate Caddy " Immobilization Rectal Balloon Item #9901.3301 Regulation Number: 21 CFR 892.5720
    Regulatory Number: 21 CFR 892.5720

    Classification: II C.

    Product Code: PCT d.

    2.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Myriad Prostate Caddy™ Immobilization Rectal Balloon is a single-use disposable, inflatable, non-powered positioning device intended for use in the temporary positioning of the rectal wall and adjacent structure in the male human anatomies. The purpose of the device is to stabilize the prostate during Computed Tomography (CT) exam and Xray, when these imaging techniques are used for Radiation Therapy (RT) planning. The placement of the balloon requires a Physician or a Physician directed healthcare professional, and is performed as a separate procedure apart from the standard CT exam and RT treatment.

    Device Description

    The Myriad Prostate Caddy™ Immobilization Rectal Balloon is designed as an immobilizer to assist in positioning the prostate in a more predictable and reproducible location during Computed Tomography (CT) exam and X-ray, when these imaging techniques are used for Radiation Therapy (RT) planning. The Prostate Immobilization Rectal Balloon is inserted into the rectum and inflated prior to the start of a CT scan or RT therapy procedure. The device stabilizes the prostate once the device is inflated. The Prostate Immobilizer Rectal Balloon is deflated and removed after each individual scan or therapy procedure is complete, and a new balloon is used in the next therapy session. Myriad's device is designed for single use, is provided non-sterile to the end user, is not intended to be sterilized by the end user, and is packaged in a kit configuration. Each kit contains the following items: Rectal Balloon (PVC); Syringe (optional accessory to the device); Locking Stopper (accessory to the device) and the Instructions for Use. The balloon consists of a single lumen shaft with a blow molded balloon attached to the distal end of the shaft. The balloon is bonded to the shaft at the proximal end of the balloon. The shaft extends into the balloon but it is not attached at the end with a rounded atraumatic tip. Several eye holes are provided to inflate/deflation of the balloon. The main section of the device consists of a balloon and single lumen tubing which are made of PVC. The remainder of the device is made of other medical grade materials consisting of the following components shown below in Table 1. The device can be inflated with either air or water, and this process is carried out with the single use disposable syringe that is supplied (or offered as an option) with the device in the accessory kit. The locking stopper supplied with the Prostate Immobilizer Rectal Balloon controls the depth at which the balloon is inserted into the rectum. The device can therefore be locked into place once inserted into place.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Myriad Prostate Caddy™ Immobilization Rectal Balloon, which aims to demonstrate substantial equivalence to a predicate device, the RadiaDyne Prostate Immobilizer Rectal Balloon RB Classic Series (K132194). This document focuses on non-clinical/bench studies to assess the device's performance, rather than a clinical study involving human subjects or AI algorithms. As such, information typically found in studies for AI/CADe devices (like sample size for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance, and ground truth types) is not applicable or available in this document.

    However, the document does detail the acceptance criteria and performance of the device based on bench testing.

    1. A table of acceptance criteria and the reported device performance:

    Test Method (Acceptance Criteria)Reported Device Performance
    Biocompatibility:
    Cytotoxicity (non-cytotoxic)Not considered to be warranted as the safety of the device has been supported by the data presented. (Implies non-cytotoxic)
    Rectal Irritation (non-irritant, irritation index = 0)For the test samples, the irritation index was 0 or no reaction. The SC (Sodium chloride) and SO (Sesame oil) test extracts were considered nonirritants to the rectal tissue of the rabbit model.
    Sensitization (no evidence of dermal reaction, no delayed dermal contact sensitization)No observed evidence of dermal reaction with all sites scoring a 0 reaction. The test article showed no evidence of causing delayed dermal contact sensitization in the guinea pig. The test article (Myriad Prostate Balloon) was not considered a sensitizer in the guinea pig maximization test.
    Acute Systemic Toxicity (no mortality, normal animal appearance, weight gain, no systemic toxicity)For the test subject and the control, there was no mortality, the animals appeared normal and all animals gained weight. There was no mortality or evidence of systemic toxicity from the extracts injected into the mice. The results showed no difference between the test and control animals.
    Shelf Life/Microbial Contamination:
    Shelf-life: 2 years (device functions properly after 2 years storage)The sponsor has provided shelf-life testing data, which ensures that the subject device functions properly after 2 years of storage.
    Bioburden (non-sterile, acceptable limits per USP <61> & USP <62>): Total Aerobic Count: ≤ 1000 CFU/g; Fungi Count: ≤ 100 CFU/gTotal Recoverable Aerobic Bioburden: average of 75 CFU (maximum 168 CFU). Total Recoverable Fungi Bioburden: average of 4 CFU (maximum 8 CFU). Bioburden results are well within acceptable limits.
    Performance Testing - Bench:
    Bioburden Testing (Maximum acceptable limit: 1000 CFU for aerobic, 100 CFU for fungi)Total Recoverable of Aerobic Bioburden was an average of 75 CFU with a maximum level of 168 CFU. The maximum acceptable limit is 1000 CFU. The Total Recoverable Fungi Bioburden was an average for 4 CFU with a maximum level of 8 CFU. The maximum acceptable limit is 100 CFU. The Bioburden results are well within acceptable limits.
    Leakage Testing (no leaks when inflated with 100cc water for 10 minutes)All units passed Leak Testing. All devices are subject to leak testing as the final quality control check (100% of Balloons inflated and held under pressure).
    Tensile-strength and Burst-strength (air): withstand 7.1 Lbs force at 100cc air inflationAll of the balloons passed the test with no signs of burst, leakage or joint failure. The test was repeated for the 2-Year Shelf Life Testing with no failures.
    Stopper Resistance Testing: withstand 3.57 Lbs tensile forceThe force to move the Stopper had an average value of 4.91 lbs. All of the units passed the test.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This document describes bench and biocompatibility testing, not a clinical study on a test set of patient data. Therefore, concepts like “test set,” “data provenance,” and “retrospective/prospective” are not applicable in this context. The samples used are devices themselves or animal models for biocompatibility.

    • Biocompatibility testing: Conducted at NAMSA, 6750 Wales Road, Northwood, Ohio 43619.
      • Cytotoxicity: Not specified individual sample count, but "data presented" supported safety.
      • Rectal Irritation: Rabbit model. Not specified sample count.
      • Sensitization: Guinea pig model. Not specified sample count.
      • Acute Systemic Toxicity: Mice. Not specified sample count.
    • Performance Testing - Bench:
      • Bioburden Testing: The testing process was repeated 3 times. The average of 75 CFU and maximum of 168 CFU suggests multiple device samples were used.
      • Leakage Testing: "A statistically appropriate number of devices" were tested, and "100 % of the Balloons are inflated" for final QC.
      • Tensile-strength and Burst-strength: Not specified individual sample count, but performed on "the Balloons."
      • Stopper Resistance Testing: Not specified individual sample count, but "All of the units passed the test."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. This device is a physical medical device, not an AI or imaging diagnostic tool. "Ground truth" in the sense of expert consensus on medical images is not relevant to its evaluation. Biocompatibility tests were conducted by a testing facility (NAMSA) according to established standards.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. There is no test set in the context of expert adjudication. Performance was evaluated based on physical measurements and biological responses in laboratory settings.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. No MRMC study was performed as this is not an AI/CADe device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance is established by:

    • Biocompatibility standards: International Organization for Standardization (ISO) 10993 series for biological evaluation of medical devices.
    • Microbial limits: USP <61> and USP <62> for Total Aerobic Count and Fungi.
    • Physical performance derived from engineering specifications and literature: For example, the 7.1 Lbs force for tensile/burst strength and 3.57 Lbs for stopper resistance are described as simulating worst-case usage per literature with a safety factor.

    8. The sample size for the training set

    Not applicable. This device does not use a training set for machine learning.

    9. How the ground truth for the training set was established

    Not applicable. This device does not use a training set for machine learning.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141958

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-11-18

    (123 days)

    Product Code
    Regulation Number
    892.5720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Road LUTSEN MN 55612

    Re: K141958

    Trade/Device Name: ERB Endorectal Balloon Regulation Number: 21 CFR 892.5720
    |
    | CLASSIFICATION NAME: | Rectal Balloon for Prostate Immobilization,21 CFR 892.5720

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ERB is a single use, inflatable, non-powered positioning device placed in the rectum to immobilize the prostate in patients undergoing radiation therapy. The device is intended to be used during all the phases of radiation therapy, including treatment planning, image verification, and radiotherapy delivery.

    Device Description

    The ERB is designed as an immobilizer to assist in positioning the prostate in a more predictable and reproducible location during Computed Tomography (CT) exam and Xray, when these imaging techniques are used for Radiation Therapy (RT) planning. The ERB is inserted into the rectum and inflated prior to the start of a CT scan or RT therapy procedure. The device stabilizes the prostate once the device is inflated. The ERB is deflated and removed after each individual scan or therapy procedure is complete, and a new balloon is used in the next therapy session. The ERB is designed for single use, is provided non-sterile to the end user, and is not intended to be sterilized by the end user. The ERB consists of a balloon and tubing can be inflated with either water or air. The device can be locked into place once inserted into the rectum with a locking stopper.

    AI/ML Overview

    The provided text is a 510(k) summary for the ERB Endorectal Balloon, a Class II medical device. The document primarily focuses on establishing substantial equivalence to a predicate device rather than presenting a standalone study of the device's performance against specific acceptance criteria.

    However, it does mention "Non-Clinical Testing" and "Performance Data" in a general sense. Based on the provided text, a detailed table of acceptance criteria and reported device performance with specific metrics cannot be fully populated. The document only states that "Product testing was completed and met all of the acceptance criteria. Testing included dimensional, visual, mechanical and performance." and "All necessary verification and validation testing has been performed for the ERB to assure substantial equivalence to the predicate devices."

    Here's an attempt to extract the information requested, with limitations due to the nature of the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The provided document does not list specific quantitative acceptance criteria or detailed performance metrics. It generally states that testing was completed and met all acceptance criteria, and that the device's performance attributes are equivalent to the predicate. Therefore, the "Reported Device Performance" column reflects generalized statements from the document.

    Acceptance Criteria CategoryReported Device Performance
    DimensionalMet all acceptance criteria. Dimensions are equivalent to predicate (minor differences in length/diameter not affecting safety/efficacy)
    VisualMet all acceptance criteria.
    MechanicalMet all acceptance criteria.
    PerformanceMet all acceptance criteria. Performance attributes are equivalent to predicate.
    BiocompatibilityPassed ISO10993 testing (Cytotoxicity, Sensitization, Irritation). Materials are suitable for use and have been used in numerous previously cleared products.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample size used for the test set in non-clinical testing. It only mentions "Product testing" and "All necessary verification and validation testing."

    Data Provenance: The testing appears to be non-clinical (laboratory/bench testing) rather than clinical studies on human subjects. The document does not specify the country of origin for the data or if it was retrospective or prospective, beyond stating that Edge Medical, LLC performed the testing.


    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not applicable and not provided in the document. The testing described is non-clinical (dimensional, visual, mechanical, performance, biocompatibility) of a physical device, not an AI/software device that would require expert-established ground truth for a test set.


    4. Adjudication Method for the Test Set

    This information is not applicable and not provided in the document for the reasons stated in point 3.


    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for imaging interpretation devices, often involving AI assistance for human readers. The ERB Endorectal Balloon is a physical medical device for prostate immobilization during radiation therapy.


    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    No, a standalone (algorithm-only) performance study was not done, as the ERB is a physical device, not an algorithm.


    7. The Type of Ground Truth Used

    For the non-clinical testing mentioned (dimensional, visual, mechanical, performance), the "ground truth" would be established by engineering specifications, material standards, and benchmark measurements against known properties of the device and its predicate. For biocompatibility, the ground truth is established by adherence to ISO10993 standards.


    8. The Sample Size for the Training Set

    This information is not applicable and not provided in the document. The ERB is a physical medical device, not an AI/machine learning algorithm that requires a training set.


    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable and not provided in the document for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141154

    Validate with FDA (Live)

    Device Name
    OARTRAC SYSTEM
    Manufacturer
    Date Cleared
    2014-07-22

    (78 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Balloon (PIRB) approved by the FDA through a de novo application (K132194) under Product Code PCT and §892.5720
    | §892.5720
    | §892.5050 +§892.5720

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OARtrac® System is specifically indicated for male prostate cancer treatment to measure photon beam therapy as an adjunct to treatment planning permitting measurement of in-vivo radiation dose received on the anterior surface of a modified Prostate Immobilization Endorectal Balloon (ERB) device to monitor and verify the surrounding organs at risk, specifically the protatic rectal interface.

    Device Description

    The OARtrac® System provides radiation oncologists with near real- time, in-vivo, multipoint radiation-dose information obtained from two (2) Radiatrac® Plastic Scintillating Detectors (PSD) sensors located on the anterior surface of a modified clinically accepted OARtrac® prostate Endorectal Balloon (ERB) to monitor dose photon based radiation therapy for prostate cancer treatment. This information allows the physician to monitor the dose at the rectal prostatic interface, compare the actual dose relative to the planned dose, and provides graphs and dose information for both the current treatment as well as a log of dose from five previous treatments. The OARtrac® System also provides dose rate during actual treatment.

    The function of the OARtrac® prostate ERB is to immobilize the prostate while at the same time positioning the Radiatrac® PSD sensors so that measurement of in-vivo radiation received on the anterior surface of the balloon can be monitored to verify the radiation dose to surrounding organs at risk, specifically the protatic rectal interface. The actual verification of the dose radiation is accomplished by the other main components of the OARtrac® System, those being the Clinical Detector Unit (CDU) with its Charged Coupled Device (CCD) camera and the system's own proprietary dose management software that are further addressed in the remainder of this section.

    AI/ML Overview

    The OARtrac® System is designed to provide near real-time, in-vivo, multipoint radiation-dose information during prostate cancer treatment. It measures photon-based radiation therapy using two Plastic Scintillating Detector (PSD) sensors located on the anterior surface of a modified Endorectal Balloon (ERB). The system monitors the dose at the rectal prostatic interface, compares it to the planned dose, and provides graphs and dose information for current and past treatments. The ERB also immobilizes the prostate and positions the sensors.

    Here's a breakdown of the acceptance criteria and study information:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Dose Accuracy+/- 6% at 95% confidence interval
    Dose Range90 Gy maximum
    Dose Per Fraction1-2000 cGy at dose rates of 1 cGy/s and above
    Time to Read Dosimeter20 seconds
    Energy Range1-20 MV Photon Dose

    Study Information

    1. Sample Size and Data Provenance:

      • The document indicates that no human clinical testing was required. The "Non-Clinical Performance Data" section mentions "controlled phantom data" was used to obtain the radiation dose accuracy for the OARtrac® System.
      • No specific sample sizes for test sets (e.g., number of phantom measurements) or the data provenance (e.g., country of origin, retrospective/prospective) are explicitly provided in the provided text.
    2. Number of Experts and Qualifications for Ground Truth (Test Set):

      • Not applicable, as no human clinical testing was performed for the test set. Ground truth for the non-clinical tests would be established through a controlled phantom setup and precise dose delivery.
    3. Adjudication Method (Test Set):

      • Not applicable, as no human clinical testing requiring expert adjudication was performed.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No MRMC comparative effectiveness study was mentioned. The device is a dosimeter aimed at objectively measuring radiation dose, not interpreting images or aiding human readers in decision-making in a way that an MRMC study would typically evaluate.
    5. Standalone Performance Study:

      • Yes, a standalone performance study was done. The "Non-Clinical Performance Data" section details a series of tests, including "Dose Range Verification Testing," "Calibration Testing," and "Dose Accuracy" measurements using controlled phantom data. This implicitly demonstrates the algorithm's (or system's) standalone performance in measuring radiation dose.
    6. Type of Ground Truth Used:

      • For the non-clinical testing, the ground truth was established through controlled phantom data and presumably the known, precisely delivered radiation doses from a calibrated source.
    7. Sample Size for Training Set:

      • The document does not explicitly mention a separate "training set" or its sample size. For device calibration and verification, data would typically be collected from controlled experiments, but the text doesn't differentiate between training and testing data in a machine learning context.
    8. How Ground Truth for Training Set was Established:

      • Ground truth for calibration and verification would be established using calibrated radiation sources and dosimetric standards applied to phantoms. The document mentions the device is "Pre-Calibrated (Cobalt-60)," indicating that Cobalt-60 sources were used in establishing foundational calibration.
    Ask a Question

    Ask a specific question about this device

    K Number
    DEN130036

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-01-28

    (197 days)

    Product Code
    Regulation Number
    892.5720
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :
    • New Regulation Number: 21 CFR 892.5720 b.
    • Classification: II C.
    • Product Code: PCT d.

    Type: Rectal Balloon for Prostate Immobilization Class: Class II (special controls) Regulation: 21 CFR 892.5720

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RadiaDyne Prostate Immobilizer Rectal Balloon is a single-use disposable, inflatable, non-powered positioning device intended for use in the temporary positioning of the rectal wall and adjacent structure in the male human anatomies. The purpose of the device is to stabilize the prostate during Computed Tomography (CT) exam and X-ray, when these imaging techniques are used for Radiation Therapy (RT) planning. The placement of the balloon requires a Physician or a Physician directed healthcare professional, and is performed as a separate procedure apart from the standard CT exam and RT treatment.

    Device Description

    The RadiaDyne Prostate Immobilizer Rectal Balloon is designed as an immobilizer to assist in positioning the prostate in a more predictable and reproducible location during Computed Tomography (CT) exam and X-ray, when these imaging techniques are used for Radiation Therapy (RT) planning. The Prostate Immobilizer Rectal Balloon is inserted into the rectum and inflated prior to the start of a CT scan or RT therapy procedure. The device stabilizes the prostate once the device is inflated. The Prostate Immobilizer Rectal Balloon is deflated and removed after each individual scan or therapy procedure is complete, and a new balloon is used in the next therapy session. RadiaDyne's device is designed for single use, is provided non-sterile to the end user, is not intended to be sterilized by the end user, and is packaged in a kit configuration. Each kit contains the following items: Rectal Balloon (polyurethane); Syringe (accessory to the device); Lubricant (accessory to the device); Locking Stopper (accessory to the device); and Instructions Manual. The main section of the device consists of a balloon and tubing which are made of polyurethane. The remainder of the device is made from other medical grade materials. The device can be inflated with either air or water, and this process is carried out with the single use disposable syringe that is supplied with the device in the accessory kit. The locking stopper supplied with the Prostate Immobilizer Rectal Balloon controls the depth at which the balloon is inserted into the rectum. The device can therefore be locked into place once inserted into place. RadiaDyne's Prostate Immobilizer Rectal Balloon is offered with a gas release valve (Models GRB) and without a gas release valve (Models: RB). Prior to insertion of the balloon, rectal gas can be removed from the rectum by methods not involving the balloon, such as inserting a pediatric catheter. The gas release model incorporates an open conduit through the balloon stem so that the rectal gas can be released once the balloon is inserted. After inflation, the gas release conduit continues to provide an escape path for any additional gas arriving at the superior end of the balloon, thus preventing gas accumulation during the procedure. The gas release valve is always open and does not require any action to function. If the model without the gas release valve is used, some transient gas may accumulate during the procedure, as is customary when no balloon is inserted. Aside from this feature, the GRB and RB models are identical in materials, construction and function.

    AI/ML Overview

    This document describes the acceptance criteria and supporting studies for the RadiaDyne Prostate Immobilizer Rectal Balloon.

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the RadiaDyne Prostate Immobilizer Rectal Balloon are primarily based on non-clinical performance testing and literature review. The performance data generally indicate that the device meets these criteria.

    Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance and Study
    Biocompatibility/Materials- Cytotoxicity (no reactivity).- Cytotoxicity: Test articles had a toxicity grade of 0, indicating no reactivity.
    - Skin Irritation (no significantly greater biological reaction than control sites, negligible primary irritation index).- Skin Irritation: Results showed no significantly greater biological reaction than control sites, with a primary irritation index of 0.0 (negligible).
    - Sensitization (Grade I reaction, weak allergenic potential).- Sensitization: Grade I reaction, classified as having weak allergenic potential.
    - Acute Systemic Toxicity (no significant difference between test and control animals).- Acute Systemic Toxicity: No significant difference between test and control animals.
    Shelf Life/Sterility- Device remains within acceptable bioburden limits and functions properly after 2 years of storage.- Shelf-life testing data supported a 2-year shelf-life, with tested devices well below the specified bioburden limit. The device is not sold as sterile.
    Performance Testing - Bench- Bioburden testing (within specified limits).- Bioburden testing: Performed based on (b)(4) Trade Secret Formula, results indicated acceptable bioburden levels.
    - Leakage testing (device holds pressure without leaks).- Leakage Testing: All devices are filled and held under pressure before release, passing the leak test.
    - Tensile-strength and Burst-strength (in air) (capable of withstanding a test-fill volume of (b)(4) Trade Secret Formula & Process of H2O before rupturing, specific tensile strength).- Tensile-strength and Burst-strength (in air): Balloon capable of withstanding (b)(4) Trade Secret Formula & Process of H2O before rupturing. Tensile strength (b)(4) Trade Secret Formula & Process using a standard COUPO11 at (b)(4) Trade Secret Formula & Process of travel.
    - Burst-strength testing (in vivo) (capable of inflating to designed volume plus safety factor in simulated in vivo use).- Burst-strength testing (in vivo): Each product code is designed for a (b)(4) Trade Secret Formula inflation volume and tested to inflate to this volume plus a safety factor in a simulated rectal environment.
    - Stopper resistance testing (ensures secure placement and performance).- Stopper resistance testing: Involves controlled testing of (b)(4) Trade Secret Formula & Process and the combination of (b)(4) Trade Secret Formula & Process.
    - Vent testing (if applicable, air flow greater than (b)(4) Trade Secret Formula & Process for gas release models).- Vent testing: For gas release models, air flow must be greater than (b)(4) Trade Secret Formula & Process, which was conducted by (b)(4) Trade Secret Formula & Process.
    Clinical Effectiveness/Safety (Literature Review)- Reduces prostate movements and intrafraction motion, improves target localization, and reduces rectal wall volume in high dose regions during radiation therapy. - Well-tolerated by patients, with acceptable rates of acute anorectal toxicity. - No serious complications (e.g., perforations or tissue damage) related to balloon application. - Effectiveness in releasing bowel gas and decreasing balloon displacement for gas-release models. - Risks mitigated by applying general and special controls.- Wachter et al. [1]: Observed maximum prostate AP displacements >5 mm in 2/10 patients with balloons vs. 8/10 without, concluding reduced prostate movements. 252 patients treated, no complications related to balloon application. - D'Amico et al. [2]: Reduced intrafraction motion (maximum prostate AP displacement from 4 mm to ≤1 mm), well tolerated by all patients. - Woel et al. [3]: Balloon well tolerated by all patients, none discontinued use. Expected increase in acute toxicities resolved by end of treatment or 3 months after. - Patel et al. [4]: Significantly reduced rectal wall volume in high-dose region, well tolerated without complications (some needed lidocaine). - Bastasch et al. [5]: 99.2% (393/396) patients tolerated 100 ml barium-enema type endorectal balloon for entire 35-day course. 0.8% required volume reduction, 4.3% lubricant, 11.6% anal med, 6.8% antidiarrhea med. Clinically acceptable acute anorectal toxicity, no Grade 3 or 4. - Ronson et al. [6]: 97.6% (3474/3561) patients tolerated balloon throughout treatment. Rectal balloons with fill volumes=120cc were well tolerated. Subject balloon fill volume (60-100ml) is less. - Both et al. [7]: Daily endorectal balloon placement consistently stabilized prostate, preventing clinically significant displacement (>5 mm). - Wang et al. [8]: Quantified interafraction prostate motion, showing reduced motion in the subject endorectal balloon group, especially in the anterior-posterior direction. - Wootton et al. [9]: Gas-release mechanism effectively released bowel gas, decreased balloon displacement.
    Labeling- Adequate instructions for use on proper insertion, positioning, and inflation. - Appropriate contraindications and warnings. - Expiration date supported by performance data. - Statement on sterility (non-sterile).- Labeling includes detailed instructions, specific warnings, contraindications (e.g., hemorrhoids, recent surgery), and an expiration date (2 years). States non-sterile and for single use. Consistent with clinical data.

    2. Sample Sizes and Data Provenance for Test Set (Clinical Studies)

    The document does not detail a single "test set" in the context of a controlled clinical trial for the device's regulatory acceptance. Instead, it relies on a review of published clinical literature (retrospective and prospective studies) that used either similar devices or the subject device. The information below summarizes the sample sizes and provenance from these cited studies.

    • Wachter et al. [1]:
      • Sample Size: 10 patients (for initial observation of AP displacements); 252 patients treated with fractionated radiation therapy.
      • Provenance: Not explicitly stated, but likely from a clinical setting where radiation therapy for prostate cancer is performed. Retrospective observations from a treated cohort.
    • D'Amico et al. [2]:
      • Sample Size: Not explicitly stated, but implies multiple patients as "all patients" tolerated placement and inflation.
      • Provenance: Clinical setting, "repeated CT-imaging with 1 min time intervals" suggests a prospective measurement during a treatment or simulation session.
    • Woel et al. [3]:
      • Sample Size: Multiple patients, as "patients treated with four-field three-dimensional (3D) conformal radiation therapy (CRT)". "all of the patients" tolerated the balloon. Specific number not given.
      • Provenance: Clinical study, evaluating acute toxicity in patients undergoing radiation therapy.
    • Patel et al. [4]:
      • Sample Size: Not explicitly stated, but refers to "patients," indicating more than one. "None of the patients in this study..."
      • Provenance: Clinical study, demonstrating rectal dose sparing.
    • Bastasch et al. [5]:
      • Sample Size: 396 patients.
      • Provenance: Baylor College of Medicine/Methodist Hospital group, study from 1998 onwards. Retrospective analysis of patient tolerance during IMRT.
    • Ronson et al. [6]:
      • Sample Size: 3561 patients.
      • Provenance: Not explicitly stated, but likely a large clinical cohort study.
    • Both et al. [7]:
      • Sample Size: Not explicitly stated, but focused on "intrafraction prostate displacements measured with electromagnetic tracking system" in a clinical setting using the subject balloon system.
      • Provenance: Clinical study, using the subject device.
    • Wang et al. [8]:
      • Sample Size: Not explicitly stated, comparing "patient groups treated with and without daily endorectal balloons".
      • Provenance: Clinical study, using the subject device.
    • Wootton et al. [9]:
      • Sample Size: Not explicitly stated, but was a "retrospectively examined" study.
      • Provenance: Clinical study, using the subject device with a gas-release mechanism.

    3. Number of Experts and Qualifications for Ground Truth (Clinical Studies)

    The document does not describe a traditional "ground truth" establishment process for a diagnostic device test set involving experts reviewing images. Instead, the "ground truth" for the device's effectiveness and safety is inferred from:

    • Clinical Outcomes: Stability of the prostate, reduction in intrafraction motion, reduction in rectal wall volume in high dose regions, patient tolerance, and lack of severe complications (e.g., perforation, tissue damage).
    • Clinical Judgement/Observations: The studies cited were conducted by medical professionals in radiation oncology. The "experts" are the physicians and healthcare professionals involved in delivering radiation therapy and assessing patient outcomes. Their qualifications are implicit in the nature of the studies (e.g., oncologists, radiation therapists performing patient care and data collection).

    There is no mention of a specific number of independent experts adjudicating ground truth for device performance in the context of AI studies or similar applications.

    4. Adjudication Method for Test Set

    Not applicable. The regulatory decision hinges on a summation of existing scientific literature and specific bench testing, not a single comparative study with a human-adjudicated test set in the manner of AI/CADe device evaluations (e.g., 2+1 adjudication).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was mentioned regarding human readers improving with AI vs. without AI assistance. This device is a physical medical device, not an AI or imaging interpretation tool. The studies focus on the physical effect of the balloon on prostate motion and patient tolerance during radiation therapy, not on human interpretive performance.

    6. Standalone (Algorithm Only) Performance

    Not applicable. The device is a physical medical device, not a software algorithm.

    7. Type of Ground Truth Used

    The ground truth for the device's effectiveness and safety is derived from:

    • Clinical Outcomes Data: This includes measurements of prostate displacement (AP displacement, intrafraction and interafraction motion), assessment of rectal toxicity (acute gastrointestinal, genitourinary, dermatological toxicity), patient tolerance, and observed complications (e.g., perforations, tissue damage).
    • Quantitative Imaging Analysis: Studies used repeated CT examinations and electromagnetic tracking systems to quantify prostate motion.
    • Biocompatibility Standards: Adherence to established standards for material safety (cytotoxicity, irritation, sensitization, acute systemic toxicity).
    • Bench Testing Standards: Verification against engineering specifications for leakage, burst strength, tensile strength, and vent performance.

    8. Sample Size for Training Set

    Not applicable. This is a physical medical device, not an AI/machine learning algorithm, so there is no "training set." The supporting information is based on literature review and physical bench tests.

    9. How Ground Truth for Training Set Was Established

    Not applicable, as there is no training set for this physical device.