Found 41 results
510(k) Data Aggregation
(118 days)
Korea
Re: K253796
Trade/Device Name: Lunit INSIGHT DBT (V1.2)
Regulation Number: 21 CFR 892.2090
Classification Name | Radiological Computer-Assisted Detection And Diagnosis Software |
| Regulation | 21 CFR 892.2090
Computer Assisted Detection/Diagnosis Software For Suspicious Lesions For Cancer |
| Regulation | 21 CFR 892.2090
Assisted Detection/Diagnosis Software For Suspicious Lesions For Cancer |
| Regulation | 21 CFR 892.2090
| 21 CFR 892.2090 |
| Regulatory Class | Class II | Class II |
| Product Code | QDQ | QDQ |
Lunit INSIGHT DBT is a computer-assisted detection and diagnosis (CADe/x) software intended to be used concurrently by interpreting physicians to aid in the detection and characterization of suspected lesions for breast cancer in digital breast tomosynthesis (DBT) exams from compatible DBT systems. Through the analysis, the regions of soft tissue lesions and calcifications are marked with an abnormality score indicating the likelihood of the presence of malignancy for each lesion. Lunit INSIGHT DBT uses screening mammograms of the female population.
Lunit INSIGHT DBT is not intended as a replacement for a complete interpreting physician's review or their clinical judgment that takes into account other relevant information from the image or patient history.
Lunit INSIGHT DBT is a computer-assisted detection/diagnosis (CADe/x) Software as a Medical Device that provides information about the presence, location and characteristics of lesions suspicious for breast cancer to assist interpreting physicians in making diagnostic decisions when reading digital breast tomosynthesis (DBT) images. The software automatically analyzes digital breast tomosynthesis slices via artificial intelligence technology that has been trained via deep learning.
For each DBT case, Lunit INSIGHT DBT generates artificial intelligence analysis results that include the lesion type, location, lesion-level/case-level score, and outline of the regions suspected of breast cancer. This peripheral information intends to augment the physician's workflow to better aid in detection and diagnosis of breast cancer.
Here's an analysis of the acceptance criteria and the study proving the device meets those criteria, based on the provided FDA 510(k) Clearance Letter for Lunit INSIGHT DBT (V1.2):
1. Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Primary Endpoint) | Reported Device Performance (Lunit INSIGHT DBT V1.2) |
|---|---|
| Lower bound of 95% CI of device's ROC AUC > 0.903 | 0.9388 (95% CI: 0.9304, 0.9472) - Met. The lower bound of 0.9304 is greater than 0.903. |
| p-value < 0.05 | p < 0.05 - Met. |
Secondary Endpoints and Performance:
| Secondary Endpoint | Reported Device Performance (Lunit INSIGHT DBT V1.2) |
|---|---|
| JAFROC AUC | 0.9206 (95% CI: 0.9117, 0.9295) |
| Sensitivity at default operating point (0.1) | 91.11% (95% CI: 89.66, 92.57) |
| Specificity at default operating point (0.1) | 77.62% (95% CI: 75.70, 79.54) |
| Sensitivity at supplementary operating point (0.3) | 88.38% (95% CI: 86.74, 90.02) |
| Specificity at supplementary operating point (0.3) | 83.68% (95% CI: 81.98, 85.38) |
| Sensitivity at supplementary operating point (0.6) | 81.48% (95% CI: 79.49, 83.47) |
| Specificity at supplementary operating point (0.6) | 93.44% (95% CI: 92.30, 94.58) |
| Lesion type agreement (CAD vs. ground truther) proportion (3-way classification) | 75.61% (95% CI: 73.40, 77.80) |
2. Sample Size for Test Set and Data Provenance
- Sample Size: 3,277 DBT exams of female adults.
- Data Provenance: Collected from multiple imaging facilities in US healthcare institutions to broadly cover the US population and maintain balanced demographic and cancer characteristic distributions. Data included patient demographics (age, ethnicity, race) and previous breast cancer history from the United States. DBT images were obtained from Hologic, GE Healthcare, Siemens, and FujiFilm 3D mammography equipment. The data contained various clinical subgroups and confounders (breast composition, BI-RADS categories, lesion type, cancer type, slice thickness).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts:
- In datasets where three ground truthers were involved: 3 qualified breast imaging radiologists.
- In datasets where two ground truthers were involved: 2 qualified breast imaging radiologists.
- Qualifications of Experts: Described as "expert breast imaging radiologists" and "qualified breast imaging radiologists." Within these groups, there was a hierarchy: one "final truther" or "most experienced" radiologist who made the ultimate decision.
4. Adjudication Method for the Test Set
The adjudication method varied based on the dataset:
- For datasets with three ground truthers: Two ground truthers independently performed the initial review, and the final truther (most experienced) determined the final reference standard, considering the results of the other two. This implies a 2+1 adjudication process in practice.
- For datasets with two ground truthers: The first truther independently completed the review, and the final truther (more experienced) made the final decision considering the results of the other truther. This also implies an ad-hoc 2-expert adjudication with a more experienced tie-breaker.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
-
Yes, an MRMC comparative effectiveness study was done.
-
Effect Size of Human Readers Improvement with AI vs. Without AI Assistance: The study compared AI standalone performance against the average unaided reader performance.
- AI standalone AUROC: 0.9430
- Average reader AUROC: 0.8983
- Difference: 0.0446 (95% CI: [0.0115, 0.0777], p = 0.0083)
This indicates that the AI standalone AUROC was significantly superior to the average unaided reader AUROC by 0.0446.
Further specific comparisons:
- At the highest threshold (Score 60): AI standalone sensitivity (86.2%) was equivalent to the average reader's sensitivity (85.4%, p = 0.8912).
- At the highest threshold (Score 60): AI standalone specificity (95.9%) was significantly improved over the average reader's specificity (77.3%, p < 0.001).
The MRMC study involved a testing dataset of 258 cases (128 negative, 65 benign, 65 cancer) with 4 views, and a reading panel of 15 American Board of Radiology and MQSA-certified radiologists. The Obuchowski–Rockette (OR) method for a single-treatment, random-reader random-case (RRRC) CAD-vs-radiologist design was used.
6. Standalone Performance Study
- Yes, a standalone (algorithm only without human-in-the-loop performance) study was done. The primary test to demonstrate substantial equivalence was a standalone performance test using 3,277 DBT exams. The acceptance criteria described in section 1 (ROC AUC > 0.903) were based on this standalone performance. The results of this standalone test are detailed in section 1 of this response.
7. Type of Ground Truth Used
- The ground truth used was expert consensus combined with clinical supporting data and pathology reports.
- Expert breast imaging radiologists ("Ground Truthers") classified each DBT exam as non-cancer or cancer and annotated malignant lesion locations.
- This process involved reviewing collected study exams using "relevant clinical supporting data such as radiology reports and pathology reports acquired from the investigational institution."
- For biopsy-proven cancer exams, ground truthers specifically referred to "relevant pathology report containing the cancer characteristic information (i.e., cancer location, size, shape, presence of calcification, pathologic results, etc.) for the ground truthing."
8. Sample Size for the Training Set
- The document states, "The test set used for the clinical validation was completely independent from the datasets used for training, tuning, or calibrating the algorithm." However, the sample size for the training set is not explicitly provided in the given text.
9. How the Ground Truth for the Training Set Was Established
- The document implies that training data was used ("training, tuning, or calibrating the algorithm") but does not describe how the ground truth for the training set was established.
Ask a specific question about this device
(269 days)
FRANCE
Re: K251474
Trade/Device Name: eyonis® LCS 1.1
Regulation Number: 21 CFR 892.2090
"*
FORM FDA 3881 (6/20) Page 1 of 1
Page 5
Regulation Number 21 CFR 892.2090
**Product
Software for Lesions Suspicious for Cancer |
| Class | Class II |
| Regulation Number | 21 CFR 892.2090
** |
|---|---|---|---|
| Classification Regulation | 21 CFR 892.2090 Radiological Computer Assisted
** |
|---|---|---|---|
| Classification Regulation | 21 CFR 892.2090 Radiological Computer Assisted
eyonis® LCS is indicated for use by radiologists. eyonis® LCS assists radiologists in the detection, localization and characterization of solid and part-solid probably benign, suspicious and very suspicious pulmonary parenchymal nodules with a diameter of 4-30 mm (pure ground glass, mediastinal lesions and masses - including but not limited to hilar masses - are excluded).
The eyonis® LCS - lung nodules result report provides, for each reported nodule, the following information: slice number, malignancy score, full snapshot, close-up snapshot, diameters (long/short/average) and volume.
The eyonis® LCS - lung nodules result report is indicated to aid in diagnosis as well as to aid in follow-up exam evaluations and as an adjunct to support clinical/patient management.
It cannot, however, substitute for medical experts' clinical judgment.
The target population is higher-risk patients eligible for participation in lung cancer screening programs (as per USPSTF criteria) with the exception of patients with pure ground glass cancer only and hilar/mediastinal cancer who are excluded from the intended population.
It is indicated for use with low-dose Chest CT DICOM images. The exact specifications are described in a DICOM Conformance Statement (DCS) document.
eyonis® LCS is an AI/ML technology-based end-to-end CADe/CADx Software as Medical Device (SaMD) intended to allow early detection, localization and characterization of pulmonary parenchymal nodules from LDCT DICOM images produced during Chest CT examinations.
The product consists of a Container-based Image Processing chain with no associated viewer. The software applies algorithms for detection, localization and characterization of solid and part-solid nodules. These algorithms employ proprietary AI and Machine Learning models trained with large databases containing proven examples of lung cancer lesions and benign nodules.
Processing results of eyonis® LCS are given in the form of a DICOM 'result report' which displays probably benign/suspicious/very suspicious nodules with a malignancy risk score per nodule, ranked by score and an associated malignancy rate observed in our reference population.
The result report is saved directly as a DICOM file for the purpose of being stored on a PACS system. It is transmitted back to the Median Gateway to be dispatched to a configured location, local hard drive or a DICOM Service class provider. Alternatively, Dicom Web and HL7 are supported.
eyonis trust, a license server, addresses important challenges such as authorizing or revoking customer access to eyonis® LCS 1.1, tracking software usage, and ensuring legal compliance with licensing contracts.
eyonis® LCS output is not intended to replace the clinical judgment of the interpreting physician and should only be used along with clinical interpretation.
The software can be deployed in any environment that supports Kubernetes deployment (Cloud based or on premises) as a set of orchestrated containers.
Software deployment is flexible and allows 3 installation methods, depending on the infrastructure, resources and capabilities.
Here's a breakdown of the acceptance criteria and study details for the eyonis® LCS 1.1 device, based on the provided FDA 510(k) Clearance Letter:
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Metric | Acceptance Criterion | Reported Device Performance |
|---|---|---|
| Standalone Study: | ||
| Patient-level AUROC | > 0.800 | 0.904 [0.881-0.926] p<0.0001 |
| Sensitivity at COT | > 70% | 84.50% [80.22-88.17] p<0.0001 |
| Specificity at COT | > 70% | 80.25% [77.33-82.95] p<0.0001 |
| AULROC | > 0.750 | 0.869 [0.843-0.894] p<0.0001 |
| MRMC Reader Study: | ||
| Primary Endpoint: | ||
| ΔAUC (aided – unaided) | > 0, p<0.05 | 0.0158 [0.0032-0.0288], p = 0.0277 (aided AUC = 0.8434 / unaided AUC = 0.8276) |
| Secondary Endpoints: | ||
| ΔSensitivity (aided – unaided) | > 0 | 1.25 [-1.52-4.02] (aided = 93.75% / unaided = 92.50%) (p>0.05, non-significant superior) |
| ΔSpecificity (aided – unaided) | > 0, p<0.05 | 4.14 [0.27-8.01] (aided = 53.59% / unaided = 49.45%) (p<0.05, significant) |
2. Sample Size Used for the Test Set and Data Provenance
- Standalone Study Test Set:
- Sample Size: 1,147 patients (342 cancers, 805 non-cancer cases).
- Data Provenance: Retrospective cohort study from 7 different datasets: 2 academic sites in Europe, 3 academic sites in the United States, and 2 private American data providers. The study population was high-risk individuals (50-80 years old with a history of smoking) and enriched for cancer prevalence, nodule size, and spiculation.
- MRMC Reader Study Test Set:
- Sample Size: 480 total patient images.
- Data Provenance: Retrospective study from patients across the United States and Europe.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Standalone Study: The ground truth was established by "radiologist consensus." The number of radiologists involved specific to this consensus is not explicitly stated, nor are their detailed qualifications beyond being "radiologist."
- MRMC Reader Study: The ground truth was established by a "reference standard" which the radiologists' performance was compared against. The document doesn't explicitly state the number of experts used to establish this specific reference standard, nor their qualifications. However, the readers involved in the study were 16 US Board Certified radiologists with an average of 13.31 years of experience (ranging from 2 to 32 years). These are the readers whose performance was evaluated against the ground truth.
4. Adjudication Method for the Test Set
The document does not explicitly state a specific adjudication method (e.g., 2+1, 3+1) for establishing the ground truth for either the standalone or MRMC study. It mentions "radiologist consensus" for the standalone study and "reference standard" for the MRMC study, implying a consensual agreement but without detailing the process.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
Yes, a pivotal MRMC comparative effectiveness study was done.
- Effect Size of Human Readers' Improvement with AI vs. without AI Assistance:
- Primary Endpoint (AUC): The ∆AUC (aided – unaided) was 0.0158 [0.0032-0.0288], with a p-value of 0.0277, indicating a statistically significant improvement in diagnostic accuracy.
- Sensitivity: The ∆Sensitivity (aided – unaided) was 1.25% (aided = 93.75%, unaided = 92.50%). While numerically superior, this was not statistically significant for superiority (p>0.05). However, it met the secondary non-inferiority objective.
- Specificity: The ∆Specificity (aided – unaided) was 4.14% (aided = 53.59%, unaided = 49.45%), which was statistically significant (p<0.05), demonstrating improved specificity.
- Inter-reader Agreement (Patient Score): Increased from ICC of 0.707 (unaided) to 0.830 (aided), with p<0.0001.
- Inter-reader Agreement (Patient Management): Kappa value increased from 0.3507 (unaided) to 0.4898 (aided), with p<0.05.
6. If a Standalone (i.e. Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, a standalone performance test was executed on eyonis® LCS. The results are reported in the "Non-clinical Performance Testing" section and are detailed in the table above (patient-level AUROC, sensitivity, specificity, and AULROC).
7. The Type of Ground Truth Used
- Standalone Study: The reference standard was proven via histopathology or ≥12 months stability.
- MRMC Reader Study: The document refers to a "reference standard" against which reader performance was compared, but does not explicitly detail the nature of this reference standard (e.g., specific pathology, expert consensus) within the provided text for the reader study data. However, given the context of lung nodule characterization, it is highly likely to be based on histopathology for confirmed cancers and stability for benign nodules, consistent with the standalone study.
8. The Sample Size for the Training Set
The document states that the AI and Machine Learning models were "trained with large databases containing proven examples of lung cancer lesions and benign nodules." However, the exact sample size for the training set is not provided in the given text.
9. How the Ground Truth for the Training Set Was Established
The document states that the models were trained with "large databases containing proven examples of lung cancer lesions and benign nodules." This implies that the ground truth for the training set was established through definitive diagnostic methods, likely similar to the test set (e.g., histopathology for malignancies and long-term stability or biopsy for benign cases). However, the specific methodology for establishing ground truth for the training set is not explicitly detailed in the provided text.
Ask a specific question about this device
(54 days)
Massachusetts 02144
Re: K251873
Trade/Device Name: Saige-Dx
Regulation Number: 21 CFR 892.2090
Radiological Computer Assisted Detection/Diagnosis Software for Lesions Suspicious for Cancer (21 CFR 892.2090
Radiological Computer Assisted Detection/Diagnosis Software for Lesions Suspicious for Cancer (21 CFR 892.2090
Saige-Dx analyzes digital breast tomosynthesis (DBT) mammograms to identify the presence or absence of soft tissue lesions and calcifications that may be indicative of cancer. For a given DBT mammogram, Saige-Dx analyzes the DBT image stacks and the accompanying 2D images, including full field digital mammography and/or synthetic images. The system assigns a Suspicion Level, indicating the strength of suspicion that cancer may be present, for each detected finding and for the entire case. The outputs of Saige-Dx are intended to be used as a concurrent reading aid for interpreting physicians on screening mammograms with compatible DBT hardware.
Saige-Dx is a software device that processes screening mammograms using artificial intelligence to aid interpreting radiologists. By automatically detecting the presence or absence of soft tissue lesions and calcifications in mammography images, Saige-Dx can help improve reader performance, while also reducing reading time. The software takes as input a set of x-ray mammogram DICOM files from a single digital breast tomosynthesis (DBT) study and generates finding-level outputs for each image analyzed, as well as an aggregate case-level assessment. Saige-Dx processes both the DBT image stacks and the associated 2D images (full-field digital mammography (FFDM) and/or synthetic 2D images) in a DBT study. For each image, Saige-Dx outputs bounding boxes circumscribing any detected findings and assigns a Finding Suspicion Level to each finding, indicating the degree of suspicion that the finding is malignant. Saige-Dx uses the results of the finding-level analysis to generate a Case Suspicion Level, indicating the degree of suspicion for malignancy across the case. Saige-Dx encapsulates the finding and case-level results into a DICOM Structured Report (SR) object containing markings that can be overlaid on the original mammogram images using a viewing workstation and a DICOM Secondary Capture (SC) object containing a summary report of the Saige-Dx results.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) clearance letter for Saige-Dx:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document indicates that the primary endpoint of the standalone performance testing was to demonstrate non-inferiority of the subject device (new Saige-Dx version) to the predicate device (previous Saige-Dx version). Specific quantitative acceptance criteria (e.g., AUC, sensitivity, specificity thresholds) are not explicitly stated in the provided text. However, the document states:
"The test met the pre-specified performance criteria, and the results support the safety and effectiveness of Saige-Dx updated AI model on Hologic and GE exams."
| Acceptance Criteria (Not explicitly quantified in source) | Reported Device Performance |
|---|---|
| Non-inferiority of subject device performance to predicate device performance. | "The test met the pre-specified performance criteria, and the results support the safety and effectiveness of Saige-Dx updated AI model on Hologic and GE exams." |
| Performance across breast densities, ages, race/ethnicities, and lesion types and sizes. | Subgroup analyses "demonstrated similar standalone performance trends across breast densities, ages, race/ethnicities, and lesion types and sizes." |
| Software design and implementation meeting requirements. | Verification testing including unit, integration, system, and regression testing confirmed "the software, as designed and implemented, satisfied the software requirements and has no unintentional differences from the predicate device." |
2. Sample Size for the Test Set and Data Provenance
- Sample Size for Test Set: 2,002 DBT screening mammograms from unique women.
- 259 cancer cases
- 1,743 non-cancer cases
- Data Provenance:
- Country of Origin: United States (cases collected from 12 diverse clinical sites).
- Retrospective or Prospective: Retrospective.
- Acquisition Equipment: Hologic (standard definition and high definition) and GE images.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The document mentions: "The case collection and ground truth lesion localization processes of the newly collected cases were the same processes used for the previously collected test dataset (details provided in K220105)."
- While the specific number and qualifications of experts for the ground truth of the current test set are not explicitly detailed in this document, it refers back to K220105 for those details. It implies that a standardized process involving experts was used.
4. Adjudication Method for the Test Set
The document does not explicitly describe the adjudication method (e.g., 2+1, 3+1) used for establishing ground truth for the test set. It states: "The case collection and ground truth lesion localization processes of the newly collected cases were the same processes used for the previously collected test dataset (details provided in K220105)." This suggests a pre-defined and presumably robust method for ground truth establishment.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was it done? Yes.
- Effect Size: The document states: "a multi-reader multi-case (MRMC) study was previously conducted for the predicate device and remains applicable to the subject device." It does not provide details on the effect size (how much human readers improve with AI vs. without AI assistance) within this document. Readers would need to refer to the K220105 submission for that information if it was presented there.
6. Standalone (Algorithm Only) Performance Study
- Was it done? Yes.
- Description: "Validation of the software was conducted using a retrospective and blinded multicenter standalone performance testing under an IRB approved protocol..."
- Primary Endpoint: "to demonstrate that the performance of the subject device was non-inferior to the performance of the predicate device."
7. Type of Ground Truth Used
- The ground truth involved the presence or absence of cancer, with cases categorized as 259 cancer and 1,743 non-cancer. The mention of "ground truth lesion localization processes" implies a detailed assessment of findings, likely involving expert consensus and/or pathology/biopsy results to confirm malignancy. Given it's a diagnostic aid for cancer, pathology is the gold standard for confirmation.
8. Sample Size for the Training Set
- Training Dataset: 161,323 patients and 300,439 studies.
9. How the Ground Truth for the Training Set Was Established
- The document states: "The Saige-Dx algorithm was trained on a robust and diverse dataset of mammography exams acquired from multiple vendors including GE and Hologic equipment."
- While it doesn't explicitly detail the method of ground truth establishment for the training set (e.g., expert consensus, pathology reports), similar to the test set, for a cancer detection AI, it is highly probable that the ground truth for the training data was derived from rigorous clinical assessments, including follow-up, biopsy results, and/or expert interpretations, to accurately label cancer and non-cancer cases for the algorithm to learn from. The implied "robust and diverse" nature of the training data suggests a comprehensive approach to ground truth.
Ask a specific question about this device
(279 days)
19702
Re: K243341
Trade/Device Name: Genius AI Detection 2.0
Regulation Number: 21 CFR 892.2090
Detection and Diagnosis Software for Lesions Suspicious for Cancer
Regulatory Number: 21 CFR 892.2090
Detection and Diagnosis Software for Lesions Suspicious for Cancer
Regulatory Number: 21 CFR 892.2090
Detection and Diagnosis Software for Lesions Suspicious for Cancer
Regulatory Number: 21 CFR 892.2090
Both the proposed and predicate device use the same technology per 21 CFR 892.2090.
Genius AI Detection is a computer-aided detection and diagnosis (CADe/CADx) software device intended to be used with compatible digital breast tomosynthesis (DBT) systems to identify and mark regions of interest including soft tissue densities (masses, architectural distortions and asymmetries) and calcifications in DBT exams from compatible DBT systems and provide confidence scores that offer assessment for Certainty of Findings and a Case Score.
The device intends to aid in the interpretation of digital breast tomosynthesis exams in a concurrent fashion, where the interpreting physician confirms or dismisses the findings during the reading of the exam.
Genius AI Detection 2.0 is a software device intended to identify potential abnormalities in breast tomosynthesis images. Genius AI Detection 2.0 analyzes each standard mammographic view in a digital breast tomosynthesis examination using deep learning networks. For each detected lesion, Genius AI Detection 2.0 produces CAD results that include:
- the location of the lesion;
- an outline of the lesion;
- a confidence score for the lesion
- Genius AI Detection 2.0 also produces a case score for the entire breast tomosynthesis exam.
Genius AI Detection 2.0 packages all CAD findings derived from the corresponding analysis of a tomosynthesis exam into a DICOM Mammography CAD SR object and distributes it for display on DICOM compliant review workstations. The interpreting physician will have access to the CAD findings concurrently to the reading of the tomosynthesis exam. In addition, a combination of peripheral information such as number of marks and case scores may be used on the review workstation to enhance the interpreting physician's workflow by offering a better organization of the patient worklist.
Here's a breakdown of the acceptance criteria and study details for Genius AI Detection 2.0, based on the provided FDA 510(k) clearance letter:
Acceptance Criteria and Device Performance for Genius AI Detection 2.0
1. Table of Acceptance Criteria and Reported Device Performance
The provided document describes a non-inferiority study to demonstrate that the performance of Genius AI Detection 2.0 on Envision (ENV) images is equivalent to its performance on the predicate's Standard of Care (SOC) images (Hologic's Selenia Dimensions systems). The primary acceptance criterion was non-inferiority of the Area Under the Curve (AUC) of the ROC curve, with a 5% margin. Secondary metrics included sensitivity, specificity, and false marker rate per view.
| Acceptance Criteria Category | Specific Metric | Predicate Device Performance (SOC Images) | Subject Device Performance (ENV Images) | Acceptance Criteria Met? |
|---|---|---|---|---|
| Primary Endpoint (Non-Inferiority) | AUC of ROC Curve (ENV-SOC) | N/A (Comparison study) | -0.0017 (95% CI -0.023 - 0.020) | Yes (p-value for difference = 0.87, indicating no significant difference, and within 5% non-inferiority margin) |
| Secondary Metrics | Sensitivity | N/A (Comparison study) | No significant difference reported between modalities | Yes |
| Specificity | N/A (Comparison study) | No significant difference reported between modalities | Yes | |
| False Marker Rate per View | N/A (Comparison study) | No significant difference reported between modalities | Yes | |
| CC-MLO Correlation | Accuracy on Malignant Lesions | N/A | 90% | Yes (Considered accurate) |
| Accuracy on Negative Cases (Correlated pairs) | N/A | 73% | Yes (Considered accurate) | |
| Implant Cases | Location-specific cancer detection sensitivity | N/A | 76% (CI 68%~84%) | Yes (Considered acceptable based on confidence intervals) |
| Specificity | N/A | 67% (CI 62%~72%) | Yes (Considered acceptable based on confidence intervals) |
(Note: The document focuses on demonstrating equivalence to the predicate's performance on a new platform rather than absolute performance against a fixed threshold for all metrics, except for the implant case where specific CIs are given and deemed acceptable.)
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size (Main Comparison Study): 1475 subjects
- 200 biopsy-proven cancer subjects
- 275 biopsy-proven benign subjects
- 78 BI-RADS 3 subjects (considered BI-RADS 1 or 2 upon diagnostic workup)
- 922 BI-RADS 1 and 2 subjects (at screening)
- Implant Case Test Set: 480 subjects
- 132 biopsy-proven cancer subjects
- 348 negative subjects (119 biopsy-proven benign, 229 screening negative)
- Data Provenance:
- Country of Origin: Not explicitly stated, but collected from a "national multi-center breast imaging network" within the U.S., implying U.S. origin.
- Retrospective or Prospective: The main comparison study data was collected for evaluating the safety and effectiveness of the Envision platform, with an IRB approved protocol. This suggests a retrospective study design, where existing images were gathered for evaluation. The implant cases were collected between 2015 and 2022, also indicating a retrospective approach.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Number of Experts: Two
- Qualifications: Both were MQSA-certified radiologists with over 20 years of experience.
4. Adjudication Method for the Test Set
The document explicitly states that the "ground truthing to evaluate performance metrics including the locations of cancer lesions was done by two MQSA-certified radiologists with over 20 years of experience."
- Adjudication Method: It does not specify a particular adjudication method (e.g., 2+1, 3+1). It simply states that ground truthing was done by two experts. This implies either consensus was reached between the two, or potentially an unstated arbitration method if they disagreed, or that their individual findings were used for analysis. Given the phrasing, expert consensus is the most likely implied method, but not explicitly detailed.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No, an MRMC comparative effectiveness study was NOT done. The study described is a standalone performance comparison of the AI algorithm on images from different modalities (Envision vs. Standard of Care), not a study involving human readers with and without AI assistance to measure effect size.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, a standalone study WAS done. The document explicitly states, "A standalone study was conducted to compare the detection performance of FDA cleared Genius AI Detection 2.0 (K221449) using Standard of Care (SOC) images acquired on the Dimensions systems against images acquired on the FDA approved Envision Mammography Platform (P080003/S009)." This study evaluated the algorithm's performance (fROC, ROC, sensitivity, specificity, false marker rate) directly against the ground truth without human intervention.
7. The Type of Ground Truth Used
- Ground Truth Type: A combination of biopsy-proven cancer and biopsy-proven benign cases, along with BI-RADS diagnostic outcomes (for negative cases). For the cancer cases, the "locations of cancer lesions" were part of the ground truth.
8. The Sample Size for the Training Set
- Not provided. The document states that the test dataset was "sequestered from any training datasets by isolating it on a secured server with controlled access permissions" and that the data for implant cases was "sequestered from the training datasets for Genius AI Detection." However, the actual sample size of the training set is not mentioned.
9. How the Ground Truth for the Training Set Was Established
- Not provided. Since the training set sample size and details are not disclosed, the method for establishing its ground truth is also not mentioned in this document. It is generally assumed that similar rigorous methods (e.g., biopsy-proven truth, expert review) would have been used for training data, but this specific filing does not detail it.
Ask a specific question about this device
(216 days)
Columbia 20001
Re: K243679
Trade/Device Name: MammoScreen® (4)
Regulation Number: 21 CFR 892.2090
Computer Assisted Detection/Diagnosis Software For Lesions Suspicious For Cancer
Regulation Number: 892.2090
MammoScreen 4)** |
|---|---|---|
| Manufacturer | Therapixel | Therapixel |
| Regulation number | 892.2090
| 892.2090 |
| Product Code | QDQ | QDQ |
| Intended Use | MammoScreen 3 is a concurrent reading
| 892.2090 |
| Product Code | QDQ | QDQ |
| Intended Use | MammoScreen 3 is a concurrent reading
MammoScreen® 4 is a concurrent reading and reporting aid for physicians interpreting screening mammograms. It is intended for use with compatible full-field digital mammography and digital breast tomosynthesis systems. The device can also use compatible prior examinations in the analysis.
Output of the device includes graphical marks of findings as soft-tissue lesions or calcifications on mammograms along with their level of suspicion scores. The lesion type is characterized as mass/asymmetry, distortion, or calcifications for each detected finding. The level of suspicion score is expressed at the finding level, for each breast, and overall for the mammogram.
The location of findings, including quadrant, depth, and distance from the nipple, is also provided. This adjunctive information is intended to assist interpreting physicians during reporting.
Patient management decisions should not be made solely based on the analysis by MammoScreen 4.
MammoScreen 4 is a concurrent reading medical software device using artificial intelligence to assist radiologists in the interpretation of mammograms.
MammoScreen 4 processes the mammogram(s) and detects findings suspicious for breast cancer. Each detected finding gets a score called the MammoScreen Score™. The score was designed such that findings with a low score have a very low level of suspicion. As the score increases, so does the level of suspicion. For each mammogram, MammoScreen 4 outputs the detected findings with their associated score, a score per breast, driven by the highest finding score for each breast, and a score per case, driven by the highest finding score overall. The MammoScreen Score goes from one to ten.
MammoScreen 4 is available for 2D (FFDM images) and 3D processing (FFDM & DBT or 2DSM & DBT). Optionally, MammoScreen 4 can use prior examinations in the analysis.
The results indicating potential breast cancer, identified by MammoScreen 4, are accessible via a dedicated user interface and can seamlessly integrate into DICOM viewers (using DICOM-SC and DICOM-SR). Reporting aid outputs can be incorporated into the practice's reporting system to generate a preliminary report.
Note that the MammoScreen 4 outputs should be used as complementary information by radiologists while interpreting mammograms. For all cases, the medical professional interpreting the mammogram remains the sole decision-maker.
The provided text describes the acceptance criteria and a study to prove that MammoScreen® 4 meets these criteria. Here is a breakdown of the requested information:
Acceptance Criteria and Device Performance
1. Table of Acceptance Criteria and Reported Device Performance
Rationale for using "MammoScreen 2" data for comparison: The document states that the standalone testing for MammoScreen 4 compared its performance against "MammoScreen 2 on Dimension". While MammoScreen 3 is the predicate device, the provided performance data in the standalone test section specifically refers to MammoScreen 2. The PCCP section later references performance targets for MammoScreen versions 1, 2, and 3, but the actual "Primary endpoint" results for the current device validation are given in comparison to MammoScreen 2. Therefore, the table below uses the reported performance against MammoScreen 2 as per the "Primary endpoint" section.
| Metric | Acceptance Criteria | Reported Device Performance (MammoScreen 4 vs. MammoScreen 2) |
|---|---|---|
| Primary Objective | Non-inferiority in standalone cancer detection performance compared to the previous version of MammoScreen (specifically MammoScreen 2 on Dimension). | Achieved. |
| AUC at the mammogram level | Positive lower bound of the 95% CI of the difference in endpoints between MammoScreen 4 and MammoScreen 2. | MS4: 0.894 (0.870, 0.919) MS2: 0.867 (0.839, 0.896) Δ: 0.027 (0.002, 0.052), p<0.0001 (Lower bound of difference 0.002 is positive, meeting criteria) |
| AUC at the breast level | Positive lower bound of the 95% CI of the difference in endpoints between MammoScreen 4 and MammoScreen 2. | MS4: 0.919 (0.897, 0.941) MS2: 0.895 (0.871, 0.920) Δ: 0.023 (0.002, 0.045), p<0.0001 (Lower bound of difference 0.002 is positive, meeting criteria) |
| AUC LROC at the finding level | Positive lower bound of the 95% CI of the difference in endpoints between MammoScreen 4 and MammoScreen 2. | MS4: 0.891 (0.862, 0.921) MS2: 0.837 (0.797, 0.877) Δ: 0.055 (0.032, 0.077), p<0.0001 (Lower bound of difference 0.032 is positive, meeting criteria) |
Study Details
2. Sample size used for the test set and the data provenance
- Sample Size: 1,475 patients, leading to 2,950 included studies (each patient underwent a DBT acquisition with two Hologic mammography systems).
- Data Provenance: The document explicitly mentions "Data provenance" as a considered subgroup for analysis but does not specify the country of origin. It indicates that the data for standalone performance testing only belonged to the "test group," which means it was "unseen data" from sources entirely left out during training and tuning. The study appears to be retrospective as it uses existing patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document states that for the clinical testing (MRMC studies), "MQSA-qualified and ACR-certified readers" were used. However, for the standalone performance testing (which is where the ground truth for the algorithm's performance is established), the document only describes the "Truthing process" and does not specify the number or qualifications of experts involved in establishing the ground truth.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
The document describes the "Truthing process" for the standalone performance testing but does not specify an adjudication method involving multiple readers. The ground truth establishment is described as:
- Positive cases: biopsy-proven presence of cancer.
- Benign cases: cases confirmed by biopsy result and cases confirmed by imaging follow-up.
- Negative cases: verified by imaging follow-up.
This indicates a reliance on clinical outcomes/pathology rather than reader consensus for ground truth for the standalone performance data.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Was an MRMC study done? Yes, "Clinical Testing" section explicitly states: "The clinical validation of MammoScreen 4 includes three multi-reader multi-case (MRMC) studies: One for FFDM, One for DBT, One for combined DBT and 2D mammograms (FFDM or 2DSM), and using prior examinations."
- Effect size of improvement: The document states, "The studies demonstrated the superiority of the Area Under the Receiver Operating Characteristic Curve of the radiologist using the MammoScreen algorithm compared to the unaided radiologist." However, specific effect sizes (e.g., AUC difference, confidence intervals) for the human reader performance improvement with AI assistance versus without AI assistance are not provided in the excerpt. Only the result of superiority is mentioned, not the quantitative measure of that superiority.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Was a standalone study done? Yes. The section "The standalone performance testing carried out to validate the device is summarized in what follows:" directly addresses this.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the standalone performance testing:
- Positive cases: Biopsy-proven presence of cancer.
- Benign cases: Confirmed by biopsy result and cases confirmed by imaging follow-up.
- Negative cases: Verified by imaging follow-up.
This indicates a mix of pathology (biopsy) and outcomes data (imaging follow-up) as the ground truth.
8. The sample size for the training set
The document states: "Sources in the training/tuning group may only be used for model training and tuning. Sources in the test group may only be used for external validation of the model's performances on unseen data (i.e., from sources entirely left out during training and tuning)." However, it does not provide the specific sample size for the training set. It only implies it was "very large databases."
9. How the ground truth for the training set was established
The document states: "These modules are trained with very large databases of biopsy-proven examples of breast cancer and normal tissue." This implies that the ground truth for the training set was primarily established through biopsy results for cancerous cases and likely outcomes/clinical confirmation for normal or benign cases, similar to the test set ground truth. However, detailed methodology on training set ground truth establishment is not provided beyond "biopsy-proven examples."
Ask a specific question about this device
(193 days)
District of Columbia 20004
Re: K242683
Trade/Device Name: QP-Prostate® CAD Regulation Number: 21 CFR 892.2090
CAD software for lesions suspicious for clinically significant prostate cancer
Classification Name: 892.2090
|
| Regulation number | 21 CFR 892.2090
| 21 CFR 892.2090
QP-Prostate® CAD is a Computed Aided Detection and Diagnosis (CADe/CADx) image processing software that automatically detects and identifies suspected lesions in the prostate gland based on bi-parametric prostate MRI. The software is intended to be used as a concurrent read by physicians with proper training in a clinical setting as an aid for interpreting prostate MRI studies. The results can be displayed in a variety of DICOM outputs, including identified suspected lesions marked as an overlay onto source MR images. The output can be displayed on third-party DICOM workstations and Picture Archive and Communication Systems (PACS). Patient management decisions should not be based solely on the results of QP-Prostate® CAD.
QP-Prostate® CAD is an artificial intelligence-based Computed Aided Detection and Diagnosis (CADe/CADx) image processing software. QP-Prostate® CAD uses Al-based algorithms trained with pathology data to detect suspicious lesions for clinically significant prostate cancer. The device automatically detects and identifies suspected lesions in the prostate gland based on bi-parametric prostate MRI and provides marks over regions of the images that may contain suspected lesions. There are two possible markers that are provided in different colors suggesting different levels of suspicion of clinically significant prostate cancer (moderate or high suspicion level).
The software is intended to be used as a concurrent read by physicians with proper training in a clinical setting as an aid for interpreting prostate MRI studies. The results can be displayed in a variety of DICOM outputs, including identified suspected lesions marked as an overlay onto source MR images. The output can be displayed on third-party DICOM workstations and Picture Archive and Communication Systems (PACS). Based on biparametric input consisting of T2W and DWI series, the analysis is run automatically, and the output in standard DICOM formats is returned to PACS.
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
Table 1: Acceptance Criteria and Reported Device Performance (Standalone)
| Metric (lesion level) | Acceptance Criterion (Implicit) | Reported Device Performance |
|---|---|---|
| AUC-ROC | Evidence of good discriminatory ability (e.g., above a certain threshold) | 0.732 (95% CI: 0.668-0.791) |
| Sensitivity (high suspicion marker) | Evidence of good detection rate for clinically significant findings | 0.677 (95% CI: 0.593-0.761) |
| False Positive Rate per Case (high suspicion marker, any biopsy source) | Evidence of acceptable false positive rate | 0.417 (95% CI: 0.313-0.522) |
| Sensitivity (high and moderate suspicion markers) | Evidence of good detection rate for clinically significant findings | 0.795 (95% CI: 0.722-0.861) |
| False Positive Rate per Case (high and moderate suspicion markers, any biopsy source) | Evidence of acceptable false positive rate | 0.855 (95% CI: 0.709-0.996) |
Note: The document does not explicitly state numerical acceptance criteria thresholds for the standalone performance metrics (AUC-ROC, Sensitivity, FPR). Instead, it presents the results and implies that these values "demonstrate the safety and effectiveness" in comparison to the predicate device. The general implicit acceptance criterion for these metrics would be that they exhibit performance levels indicative of a useful diagnostic aid.
Table 2: Acceptance Criteria and Reported Device Performance (Multi-Reader Multi-Case Study)
| Metric | Acceptance Criterion (Explicit) | Reported Device Performance |
|---|---|---|
| ΔAUC (AUCaided - AUCunaided) (Primary Endpoint) | A statistically significant improvement (p-value < 0.05) | 0.019 (95% CI: 0.001-0.038) p-value: 0.039 |
| Sensitivity with/without CAD assistance (Secondary Endpoint) | Improvement when using CAD assistance | Not explicitly quantified in table, but overall improvement is stated to be demonstrated. |
| Specificity with/without CAD assistance (Secondary Endpoint) | Improvement when using CAD assistance | Not explicitly quantified in table, but overall improvement is stated to be demonstrated. |
The document stated directly that "The test results demonstrate that QP-Prostate® CAD functioned as intended and met its primary endpoint, is acceptable for clinical use, and is as safe and effective as its predicate device, without introducing new questions of safety and efficacy."
Study Details for QP-Prostate® CAD performance:
-
Sample Size and Data Provenance for Test Set:
- Sample Size: 228 cases for the clinical reader performance assessment (MRMC study) and 247 for the standalone performance assessment (lesion-level).
- Data Provenance: Retrospectively collected from multiple centers across the US. This dataset was "completely independent from the training dataset that was acquired from different institutions."
-
Number of Experts and Qualifications for Ground Truth (Test Set):
- The document states that the ground truth for the standalone performance evaluation was derived from "associated pathology reports and radiologist interpretations." It does not specify the number or qualifications of radiologists involved in these interpretations for the ground truth establishment of the test set.
- For the MRMC study, 9 readers (presumably radiologists, though specific qualifications for each weren't detailed beyond "clinical readers") participated in the study itself, but this is about their performance, not their role in establishing the ground truth.
-
Adjudication Method for Test Set:
- The document doesn't explicitly describe an adjudication method (e.g., 2+1, 3+1) for establishing the ground truth of the test set cases based on radiologist interpretations. It mentions "biopsy outcomes" and "biopsy confirmed (Gleason score ≥ 7)" for positive cases, and "biopsy-confirmed (Gleason score < 7) or non-biopsied with a clinical followup of at least one year" for negative cases. This suggests pathology and long-term clinical follow-up played a primary role in ground truth for patient status.
-
Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- Was it done? Yes, a fully crossed multi-reader multi-case study was performed.
- Effect Size of Human Reader Improvement: The primary endpoint was an improvement in AUC.
- AUC_unaided: 0.849 (95% CI: 0.814-0.884)
- AUC_aided: 0.868 (95% CI: 0.834-0.902)
- ΔAUC (AUC_aided - AUC_unaided): 0.019 (95% CI: 0.001-0.038), p-value: 0.039.
- This indicates a small but statistically significant improvement in reader performance (AUC) when assisted by the AI.
-
Standalone Performance Study (Algorithm Only):
- Was it done? Yes, a standalone performance assessment was conducted.
- The results are summarized in "Table 2: Summary of the standalone performance testing for QP-Prostate® CAD" (refer to the table above).
-
Type of Ground Truth Used:
- For the Test Set:
- Pathology: Biopsy confirmed for both positive (Gleason score ≥ 7) and negative (Gleason score < 7) cases of clinically significant prostate cancer (csPCa).
- Outcomes Data: For non-biopsied negative cases, ground truth was established by "a clinical followup of at least one year."
- Expert Consensus/Radiologist Interpretation: Used in conjunction with pathology for the standalone lesion-level evaluation ("QP-Prostate® CAD outputs were compared to ground truth diagnoses derived from associated pathology reports and radiologist interpretations, at lesion-level").
- For the Test Set:
-
Sample Size for Training Set:
- The exact sample size for the training set is not explicitly stated in numerical form. It is described as "cases acquired in the US from multiple centers."
-
How Ground Truth for Training Set Was Established:
- The AI-based algorithms were "trained with biopsy outcomes as ground truth to detect suspicious lesions for csPCa (Gleason score ≥7)." This indicates that pathology reports (biopsy outcomes) were the primary source of ground truth for the training data.
Ask a specific question about this device
(258 days)
Erlangen, 91052 Germany
Re: K241770
Trade/Device Name: Prostate MR AI (VA10A) Regulation Number: 21 CFR 892.2090
Detection/Diagnosis Software For Lesions Suspicious For Cancer Device Classification Panel Radiology 892.2090
| K181704 |
| Regulation Number | 892.2090
| 892.2090
| § 892.2090
Prostate MR AI is a plug-in Radiological Computer Assisted Detection and Diagnosis Software device intended to be used · with a separate hosting application · as a concurrent reading aid to assist radiologists in the interpretation of a prostate MRI examination acquired according to the PI-RADS standard · in adult men (40 years and older) with suspected cancer in treatment naïve prostate glands The plug-in software analyzes non-contrast T2 weighted (T2W) and diffusion weighted image (DWI) series to segment the prostate gland and to provide an automatic detection and segmentation of regions suspicious for cancer. For each suspicious region detected, the algorithm moreover provides a lesion Score, by way of PI-RADS interpretation suggestion. Outputs of the device should be interpreted consistently with ACR recommendations using all available MR data (e.g., dynamic contrast enhanced images [if available]). Patient management decisions should not be made solely based on analysis by the Prostate MR AI algorithm.
This premarket notification addresses the Siemens Healthineers Prostate MR AI (VA10A) Radiological Computer Assisted Detection and Diagnosis Software (CADe/CADx). Prostate MR AI is a Computer Assisted Detection and Diagnosis algorithm designed to plug into a hosting workflow that assists radiologists in the detection of suspicious lesions and their classification. It is used as a concurrent reading aid to assist radiologists in the interpretation of a prostate MRI examination acquired according to the PI-RADS standard. The automatic lesion detection requires transversal T2W and DWI series as inputs. The device automatically exports a list of detected prostate regions that are suspicious for cancer (each list entry consists of contours and a classification by Score and Level of Suspicion (LoS)), a computed suspicion map, and a per-case LoS. The results of the Prostate MR AI plug-in (with the case-level LoS, lesion center points, lesion diameters, lesion ADC median, lesion 10th percentile, suspicion map, and non-PZ segmentation considered optional) are to be shown in a hosting application that allows the radiologist to view the original case, as well as confirm, reject, or edit lesion candidates with their contours and Scores as generated by the Prostate MR AI plug-in. Moreover, the radiologist can add lesions with contours and PI-RADS scores and finalize the case. In addition, the outputs include an automatically computed prostate segmentation, as well as sub-segmentations of the peripheral zone and the rest of the prostate (non-PZ). The algorithm will augment the prostate workflow of currently cleared syngo.MR General Engine if activated via a separate license on the General Engine.
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Automatic Prostate Segmentation | |
| Median Dice score between AI algorithm results and ground truth masks exceeds 0.9. | The median of the Dice score between the AI algorithm results and the corresponding ground truth masks exceeds the threshold of 0.9. |
| Median normalized volume difference between algorithm results and ground truth masks is within ±5%. | The median of the normalized volume difference between the algorithm results and the corresponding ground truth masks is within a ±5% range. |
| AI algorithm results are statistically non-inferior to individual reader variability (5% margin of error, 5% significance level). | The AI algorithm results as compared to any individual reader are statistically non-inferior based on variabilities that existed among the individual readers within the 5% margin of error and 5% significance level. |
| Prostate Lesion Detection and Classification | |
| Case-level sensitivity of lesion detection ≥ 0.80 for both radiology and pathology ground truth. | The case-level sensitivity of the lesion detection is equal or greater than 0.80 for both radiology and pathology ground truth. |
| False positive rate per case of lesion detection < 1 false positive per case for radiology ground truth. | The false positive rate per case of the lesion detection is smaller than one false positive per case for radiology ground truth. |
| Accuracy of PI-RADS classification of radiology ground truth lesions (detected by algorithm) ≥ 0.8. | The accuracy of the PI-RADS classification of radiology ground truth lesions detected by the algorithm is equal or greater than 0.8. |
| Non-inferior performance in GE vs Siemens and African American vs non-African American cases, and in cases with peripheral zone vs non-peripheral lesions. | The non-inferior performance of the subject device in GE vs Siemens and African American vs non-African American cases, and in cases with peripheral zone vs non-peripheral lesions was demonstrated. (Note: Specific metrics for this non-inferiority are not explicitly stated as distinct numerical criteria but are stated as "met".) |
| Clinical Performance (Reader Study - Case-level discrimination of Gleason Grade Group ≥ 1) | |
| Statistically significant improvement in AUROC for aided reading vs unaided reading. | Fully Inclusive Analysis: AUROC improved from 0.6758 (unaided) to 0.7010 (aided), difference of 0.0252 (95% C.I. [0.0011, 0.0493]; P=0.040). Maximally Restrictive Analysis: AUROC improved from 0.6579 (unaided) to 0.6948 (aided), difference of 0.0368 (95% C.I. [0.0108, 0.0628]; P=0.006). In both analyses, the improvement was statistically significant and the primary endpoint thus met. |
| Clinical Performance (Reader Study - Lesion-level reading performance) | |
| Statistically significant improvement in AUwAFROC for aided reading vs unaided reading. | Fully Inclusive Analysis: AUwAFROC improved in aided reading by 0.0350 (95% C.I.:[0.0020, 0.0681], P=0.037). Maximally Restrictive Analysis: AUwAFROC improved in aided vs. unaided reading by 0.302 (95% C.I.: [0.0080,0.0520], P=0.008). In both analyses, the improvement was statistically significant and the secondary endpoint thus met. |
| Statistically significant improvement in Fleiss' Kappa for interreader agreement in per-case PI-RADS scores for aided reading vs unaided reading. | Fleiss' Kappa improved from 0.283 (unaided) to 0.371 (aided), with a difference of 0.087 (95% C.I. [0.051, 0.125]). The improvement was statistically significant (P<0.0001). |
Study Information
2. Sample size used for the test set and the data provenance:
- Automatic Prostate Segmentation: 222 transversal T2 series.
- Provenance: More than 10 clinical sites.
- Retrospective/Prospective: Not explicitly stated, but the description of comparing against ground truth generated implies retrospective use of existing scans.
- Prostate Lesion Detection and Classification (Standalone Performance):
- 105 cases from 6 sites (against radiology ground truth).
- 115 cases from 6 sites (against pathology ground truth).
- 340 cases from the multi-reader multi-case study (used for evaluation, implied prospective for this part of the evaluation, but the cases themselves were retrospective for the reader study).
- Provenance: 6 sites (for 105 and 115 cases), and two US sites (for 340 cases).
- Retrospective/Prospective: The cases for the lesion detection and classification evaluation were used to compare against established ground truths, suggesting retrospective analysis of existing data. The cases for the reader study were retrospectively selected.
- Multi-Reader Multi-Case (MRMC) Study: 340 cases.
- Provenance: Two US sites. Cases were consecutive and specifically included additional consecutive patient cases from men of African descent to ensure at least 13% Black or African American ethnicity.
- Retrospective/Prospective: Cases were selected retrospectively.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Automatic Prostate Segmentation: 3 expert radiologists. No specific years of experience or subspecialty beyond "radiologists" are mentioned but implied as "expert".
- Prostate Lesion Detection and Classification (Radiology Ground Truth): 3 expert radiologists in prostate MRI reading.
- MRMC Study (Lesion-level reference standard): 3 experienced radiologists acting as Truthers.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Automatic Prostate Segmentation: Pixel-wise consensus among the 3 expert radiologists.
- Prostate Lesion Detection and Classification (Radiology Ground Truth): Consensus reading of the 3 expert radiologists.
- MRMC Study (Case-level reference standard): Biopsy results (Gleason Grade Group GGG ≥ 1), or for cases without biopsy, PSA density and follow-up data.
- MRMC Study (Lesion-level reference standard): Consensus lesions with a consensus PI-RADS of at least 3 from majority voting among the 3 experienced radiologists. (This implies a form of consensus/majority vote).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Yes, an MRMC study was done with a paired split-plot design, combining two fully-crossed MRMC sub-studies.
- Case-level AUROC improvement (discriminating Gleason Score ≥ 1):
- Fully Inclusive Analysis: +0.0252 (from 0.6758 unaided to 0.7010 aided).
- Maximally Restrictive Analysis: +0.0368 (from 0.6579 unaided to 0.6948 aided).
- Lesion-level AUwAFROC improvement:
- Fully Inclusive Analysis: +0.0350.
- Maximally Restrictive Analysis: +0.0302.
- Fleiss' Kappa (interreader agreement in per-case PI-RADS scores) improvement: +0.087 (from 0.283 unaided to 0.371 aided).
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Yes, standalone performance was evaluated for:
- Automatic Prostate Segmentation: Compared algorithm results to ground truth generated by radiologists.
- Prostate Lesion Detection and Classification: Compared automatic detection and classification results to radiology ground truth and pathology ground truth.
- MRMC Study (AI Standalone reference): The ROC curves shown graphically include a "grey curve [that] denotes AI standalone performance."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For Automatic Prostate Segmentation: Pixel-wise consensus from 3 expert radiologists.
- For Prostate Lesion Detection and Classification:
- Consensus reading of 3 expert radiologists (radiology ground truth).
- Biopsy results for the same patient (pathology ground truth).
- For MRMC Study (Case-level): Biopsy results (Gleason Grade Group GGG ≥ 1), and in cases where biopsy was unavailable, PSA density and follow-up (12 months negative by PSA or MRI).
- For MRMC Study (Lesion-level): Consensus lesions with a consensus PI-RADS of at least 3 from majority voting among 3 experienced radiologists.
8. The sample size for the training set:
The document states: "The cases for the reader study were kept completely separate from those used for the training of the Prostate MR AI algorithm." However, it does not specify the sample size for the training set. It only mentions that the AI algorithm was "trained on a database of prostate MR image series acquired according to the PI-RADS standard (non-contrast T2W and DWI image series), and corresponding radiological and/or biopsy findings."
9. How the ground truth for the training set was established:
The ground truth for the training set was established based on "corresponding radiological and/or biopsy findings." Specific details on the adjudication method (e.g., number of experts, consensus process) for the training set are not provided in this document, only the source of the ground truth.
Ask a specific question about this device
(177 days)
NW Washington, DC 20004
Re: K242171
Trade/Device Name: TechCare Trauma Regulation Number: 21 CFR 892.2090
Radiological computer assisted detection/diagnosis software for fracture Regulation: Regulation: 21 CFR 892.2090
|
| ClassificationRegulation | 21 CFR 892.2090
| 21 CFR 892.2090
TechCare Trauma is intended to analyze 2D X ray radiographs using techniques to aid in the detection, localization, and characterization of fractures and/or elbow joint effusion during the review of commonly acquired radiographs of: Ankle, Foot, Knee, Leg (includes Tibia/Fibula), Femur, Wrist, Hand/Finger, Elbow, Forearm, Arm (includes Humerus), Shoulder, Clavicle, Pelvis, Hip, Thorax (includes ribs).
TechCare Trauma can provide results for fracture in neonates and infants (from birth to less than 2 years), children and adolescents (aged 2 to less than 22 years) and adults (aged 22 years and over).
TechCare Trauma can provide results for elbow joint effusions in children and adolescents (aged 2 to less than 22 years) and adults (aged 22 years and over).
The intended users of TechCare Trauma are clinicians with the authority to diagnose fractures and/or elbow joint effusions in various settings including primary care (e. g., family practice, internal medicine), emergency medicine, urgent care, and specialty care (e. g. orthopedics), as well as radiologists who review radiographs across settings.
TechCare Trauma results are not intended to be used on a stand-alone basis for clinical decision-making. Primary diagnostic and patient management decisions are made by the clinical user.
The TechCare Trauma device is a software as Medical Device (SaMD). More specifically it is defined as a "radiological computer assisted detection and diagnostic software for suspected fractures".
As a CADe/x software, TechCare Trauma is an image processing device intended to aid in the detection and localization of fractures and elbow joint effusions on acquired medical images (2D X-ray radiographs).
TechCare Trauma uses an artificial intelligence algorithm to analyze acquired medical images (2D X-ray radiographs) for features suggestive of fractures and elbow joint effusions.
TechCare Trauma can provide results for fractures in neonates and infants (from birth to less than 2 years), children and adolescents (aged 2 to less than 22 years) and adults (aged 22 years and over) regardless of their condition.
TechCare Trauma can provide results for elbow joint effusions in children and adolescents (aged 2 to less than 22 years) and adults (aged 22 years and over).The device detects and identifies fractures and elbow joint effusions based on a visual model's analysis of images and provides information about the presence and location of these prespecified findings to the user.
It relies solely on images provided by DICOM sources. Once integrated into existing networks, TechCare Trauma automatically receives and processes these images without any manual intervention. The processed results, which consist of one or more images derived from the original inputs, are then sent to specified DICOM destinations. This ensures that the results can be seamlessly viewed on any compatible DICOM viewer, allowing smooth into medical imaging workflows.
TechCare Trauma can be deployed on-premises or on cloud and be connected to multiple DICOM sources / destinations (including but not limited to DICOM storage platform, PACS, VNA and radiological equipment, such as X-ray systems), ensuring easy integration into existing clinical workflows.
Here's a detailed breakdown of the acceptance criteria and study findings for the TechCare Trauma device, based on the provided text:
Acceptance Criteria and Device Performance
The acceptance criteria for the TechCare Trauma device appear to be based on achieving high diagnostic accuracy, specifically measured by the Area Under the Curve (AUC) of the Receiver Operating Characteristic (ROC) curve for both standalone performance and multi-reader multi-case (MRMC) comparative studies. The study demonstrated successful performance against these implied criteria.
Table of Acceptance Criteria and Reported Device Performance
| Metric | Acceptance Criteria (Implied/Study Goal) | Reported Device Performance (Standalone) | Reported Device Performance (MRMC with AI vs. without AI) |
|---|---|---|---|
| Standalone Performance (Image-level ROC-AUC) | High accuracy (specific threshold not explicitly stated but implied by achievement across all categories) | Fracture - Adult: 0.962 [0.957 - 0.967] Fracture - Pediatric: 0.962 [0.955 - 0.969] EJE - Adult: 0.965 [0.936 - 0.986] EJE - Pediatric: 0.976 [0.963 - 0.986] (Further detailed by anatomical regions, age, gender, image view, and imaging hardware manufacturers, all showing high AUCs.) | Not applicable (standalone algorithm only) |
| Reader Performance (MRMC ROC-AUC) | Superior to unaided reader performance (statistically significant improvement) | Not applicable (human reader performance) | Adult Fracture: Improved from 0.865 to 0.955 (Δ 0.090, p < 0.001) Adult EJE: Improved from 0.851 to 0.914 (Δ 0.064, p < 0.001) Pediatric Fracture: Improved from 0.857 to 0.931 (Δ 0.074, p < 0.001) Pediatric EJE: Improved from 0.877 to 0.941 (Δ 0.063, p = 0.002) |
| Reader Performance (MRMC Sensitivity) | Increased Sensitivity with AI aid | Not applicable (human reader performance) | Adult Fracture: Increased by 21.8% (from 0.807 to 0.983) Adult EJE: Increased by 12.7% (from 0.872 to 0.983) Pediatric Fracture: Increased by 19.9% (from 0.804 to 0.964) Pediatric EJE: Increased by 18.2% (from 0.825 to 0.975) |
| Reader Performance (MRMC Specificity) | Maintained or increased Specificity with AI aid | Not applicable (human reader performance) | Adult Fracture: Increased by 1.47% (from 0.815 to 0.827) Adult EJE: Increased by 1.08% (from 0.738 to 0.746) Pediatric Fracture: Remained the same (0.797) Pediatric EJE: Increased by 1.43% (from 0.839 to 0.851) |
Study Details
2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
Standalone Performance Test Set:
- Fracture Detection: 4109 radiographs of US adult patients and 2872 radiographs of US pediatric patients.
- EJE Detection: 280 radiographs of US adult patients and 483 radiographs of US pediatric patients.
- Data Provenance: Retrospective, obtained from various states in the US (at least 4) and various imaging hardware manufacturers (at least 14).
MRMC Comparative Effectiveness Study Test Set:
- Adult US population for fracture detection: 304 radiological cases
- Pediatric US population for fracture detection: 256 radiological cases
- Adult US population for EJE detection: 109 radiological cases
- Pediatric US population for EJE detection: 100 radiological cases
- Data Provenance: Retrospective, external multicenter anonymized datasets obtained from sites that were different from the training data sites, ensuring independence. All data from US patients.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Standalone Performance Test Set:
- Number of Experts: Three American Board of Radiology (ABR)-certified radiologists for both adult and pediatric cases.
- Qualifications: Minimum of 5 years of experience since ABR certification. Pediatric cases were annotated by pediatric radiologists, and adult cases by musculoskeletal (MSK) radiologists.
MRMC Comparative Effectiveness Study Test Set:
- Number of Experts: Three ABR-certified radiologists.
- Qualifications: At least five years of experience. Pediatric cases were annotated by a panel of three pediatric radiologists, while adult cases were reviewed by a panel of three MSK radiologists.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Both Standalone and MRMC Studies:
- Adjudication Method: Two radiologists independently assessed each case. For cases with disagreement between the first two, a third radiologist independently reviewed the case. The final reference standard (ground truth) was determined by majority consensus (referred to as "2+1" if two agree, or "3+0" if all three agree after subsequent review).
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
-
Yes, a MRMC comparative effectiveness study was done.
-
Effect Size of Improvement (AI vs. without AI assistance):
- Adult Fracture (ROC AUC delta): 0.090 (from 0.865 to 0.955)
- Adult EJE (ROC AUC delta): 0.064 (from 0.851 to 0.914)
- Pediatric Fracture (ROC AUC delta): 0.074 (from 0.857 to 0.931)
- Pediatric EJE (ROC AUC delta): 0.063 (from 0.877 to 0.941)
Additionally, significant improvements in sensitivity were observed:
- Adult Fracture Sensitivity: +21.8%
- Adult EJE Sensitivity: +12.7%
- Pediatric Fracture Sensitivity: +19.9%
- Pediatric EJE Sensitivity: +18.2%
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Yes, a standalone performance study was done.
- Fracture Detection ROC-AUC: 0.962 (Adults) and 0.962 (Pediatrics)
- EJE Detection ROC-AUC: 0.965 (Adults) and 0.976 (Pediatrics)
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The ground truth for both standalone and MRMC studies was established by expert consensus of ABR-certified radiologists.
8. The sample size for the training set
- Training Set Sample Size: 95,266 images.
9. How the ground truth for the training set was established
- The document does not explicitly describe how the ground truth for the training set was established. However, given the detailed methodology for the test set ground truth, it is highly probable that a similar expert-driven annotation process (potentially internal and/or external) was followed for the training data as well. The text states the training was performed "from various manufacturers," suggesting a diverse dataset that would necessitate robust ground truthing.
Ask a specific question about this device
(20 days)
Somerville, Massachusetts 02144
Re: K243688
Trade/Device Name: Saige-Dx (3.1.0) Regulation Number: 21 CFR 892.2090
Radiological Computer Assisted Detection/Diagnosis Software for Lesions Suspicious for Cancer (21 CFR 892.2090
Radiological Computer Assisted Detection/Diagnosis Software for Lesions Suspicious for Cancer (21 CFR 892.2090
Saige-Dx analyzes digital breast tomosynthesis (DBT) mammograms to identify the presence or absence of soft tissue lesions and calcifications that may be indicative of cancer. For a given DBT mammogram, Saige-Dx analyzes the DBT image stacks and the accompanying 2D images, including full field digital mammography and/or synthetic images. The system assigns a Suspicion Level, indicating the strength of suspicion that cancer may be present, for each detected finding and for the entire case. The outputs of Saige-Dx are intended to be used as a concurrent reading aid for interpreting physicians on screening mammograms with compatible DBT hardware.
Saige-Dx is a software device that processes screening mammograms using artificial intelligence to aid interpreting radiologists. By automatically detecting the presence or absence of soft tissue lesions and calcifications in mammography images, Saige-Dx can help improve reader performance, while also reducing time. The software takes as input a set of x-ray mammogram DICOM files from a single digital breast tomosynthesis (DBT) study and generates finding-level outputs for each image analyzed, as well as an aggregate case-level assessment. Saige-Dx processes both the DBT image stacks and the associated 2D images (full-field digital mammography (FFDM) and/or synthetic 2D images) in a DBT study. For each image, Saige-Dx outputs bounding boxes circumscribing any detected findings and assigns a Finding Suspicion Level to each finding, indicating the degree of suspicion that the finding is malignant. Saige-Dx uses the results of the finding-level analysis to generate a Case Suspicion Level, indicating the degree of suspicion for malignancy across the case. Saige-Dx encapsulates the finding and case-level results into a DICOM Structured Report (SR) object containing markings that can be overlaid on the original mammogram images using a viewing workstation and a DICOM Secondary Capture (SC) object containing a summary report of the Saige-Dx results.
The provided text describes the Saige-Dx (v.3.1.0) device and its performance testing as part of an FDA 510(k) submission (K243688). However, it does not contain specific acceptance criteria values or the quantitative results of the device's performance against those criteria. It states that "All tests met the pre-specified performance criteria," but does not list those criteria or the measured performance metrics.
Therefore, while I can extract information related to the different aspects of the study, I cannot create a table of acceptance criteria and reported device performance with specific values.
Here's a breakdown of the information available based on your request:
1. A table of acceptance criteria and the reported device performance
- Acceptance Criteria: Not explicitly stated in quantitative terms. The document only mentions that "All tests met the pre-specified performance criteria."
- Reported Device Performance: Not explicitly stated in quantitative terms (e.g., specific sensitivity, specificity, AUC values, or improvements in human reader performance).
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set Sample Size: Not explicitly stated for the validation performance study. The text mentions "Validation of the software was previously conducted using a multi-reader multi-case (MRMC) study and standalone performance testing conducted under approved IRB protocols (K220105 and K241747)." It also mentions that the tests included "DBT screening mammograms with Hologic standard definition and HD images, GE images, exams with unilateral breasts, and from patients with breast implants (on implant displaced views)."
- Data Provenance: The data for the training set was collected from "multiple vendors including GE and Hologic equipment" and from "diverse practices with the majority from geographically diverse areas within the United States, including New York and California." For the test set, it is implied to be similar in nature as it's part of the overall "performance testing," but specific details for the test set alone are not provided regarding country of origin or retrospective/prospective nature. However, since it involves IRB protocols, it suggests a structured, likely prospective collection or at least a carefully curated retrospective collection.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not explicitly stated for the test set. The document indicates that a Multi-Reader Multi-Case (MRMC) study was performed, which implies the involvement of expert readers, but the number of experts and their qualifications are not detailed.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not explicitly stated for the test set. The involvement of an MRMC study suggests a structured interpretation process, potentially including adjudication, but the method (e.g., consensus, majority rule with an adjudicator) is not described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Yes, an MRMC study was done: "Validation of the software was previously conducted using a multi-reader multi-case (MRMC) study..."
- Effect Size: The document does not provide the quantitative effect size of how much human readers improved with AI vs. without AI assistance. It broadly states that Saige-Dx "can help improve reader performance, while also reducing time."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Yes, standalone performance testing was done: "...and standalone performance testing conducted under approved IRB protocols..."
- Results: The document states that "All tests met the pre-specified performance criteria" for the standalone performance, but does not provide the specific quantitative results (e.g., sensitivity, specificity, AUC).
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not explicitly stated. For a device identifying "soft tissue lesions and calcifications that may be indicative of cancer," ground truth would typically involve a combination of biopsy/pathology results, clinical follow-up, and potentially expert consensus on imaging in cases without definitive pathology. However, the document doesn't specify the exact method for establishing ground truth for either the training or test sets.
8. The sample size for the training set
- Training Set Sample Size: "A total of nine datasets comprising 141,768 patients and 316,166 studies were collected..."
9. How the ground truth for the training set was established
- Not explicitly stated. The document mentions the collection of diverse datasets for training but does not detail how the ground truth for these 141,768 patients and 316,166 studies was established (e.g., through radiologists' interpretations, pathology reports, clinical outcomes).
Ask a specific question about this device
(153 days)
Gelderland 6525 EC NETHERLANDS
Re: K241831
Trade/Device Name: Transpara (2.1.0) Regulation Number: 21 CFR 892.2090
Radiological Computer Assisted Detection and DiagnosisSoftware |
| Classification regulation | 21 CFR 892.2090
Radiological Computer Assisted Detection and DiagnosisSoftware |
| Classification regulation | 21 CFR 892.2090
Transpara software is intended for use as a concurrent reading aid for physicians interpreting screening full-field digital mammography exams and digital breast tomosynthesis exams from compatible FFDM and DBT systems, to identify regions suspicious for breast cancer and assess their likelihood of malignancy. Output of the device includes locations of calcifications groups and soft-tissue regions, with scores indicating the likelihood that cancer is present, and an exam score indicating the likelihood that cancer is present in the exam. Patient management decisions should not be made solely on the basis of analysis by Transpara.
Transpara is a software only application designed to be used by physicians to improve interpretation of full-field digital mammography (FFMD) and digital breast tomosynthesis (DBT). Deep learning algorithms are applied to images for recognition of suspicious calcifications and soft tissue lesions (including densities, masses, architectural distortions, and asymmetries). Algorithms are trained with a large database of biopsy-proven examples of breast cancer, benign abnormalities, and examples of normal tissue.
Transpara offers the following functions which may be used at any time in the reading process, to improve detection and characterization of abnormalities and enhance workflow:
- AI findings for display in the images to highlight locations where the device detects suspicious calcifications or soft tissue lesions, along with region scores per finding on a scale ranging from 1-100, with higher scores indicating a higher level of suspicion.
- Links between corresponding regions in different views of the breast, which may be utilized to enhance user interfaces and workflow.
- An exam-based score which categorizes exams with increasing likelihood of cancer on a scale of 1-10 or in three risk categories labeled as 'low', 'intermediate' or 'elevated'.
The concurrent use indication implies that it is up to the users to decide how to use Transpara in the reading process. Transpara functions can be used before, during or after visual interpretation of an exam by a user.
Results of Transpara are computed in a standalone processing appliance which accepts mammograms in DICOM format as input, processes them, and sends the processing output to a destination using the DICOM protocol in a standardized mammography CAD DICOM format. Common destinations are medical workstations, PACS and RIS. The system can be configured using a service interface. Implementation of a user interface for end users in a medical workstation is to be provided by third parties.
The provided text describes the acceptance criteria and a study that proves the device, Transpara (2.1.0), meets these criteria.
Here's an organized breakdown of the information requested:
Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the reported performance metrics. The study aims to demonstrate non-inferiority and superiority to the predicate device, Transpara 1.7.2. The key metrics reported are sensitivity at various specificity levels and Exam-based Area Under the Receiver Operating Characteristic Curve (AUC).
Table 1: Acceptance Criteria (Implied by Performance Goals) and Reported Device Performance (Standalone without Temporal Analysis)
| Metric | Acceptance Criteria (Implied/Target) | Reported Performance (FFDM) | Reported Performance (DBT) |
|---|---|---|---|
| Sensitivity (Sensitive Mode @ 70% Specificity) | Non-inferior & Superior to Predicate Device 1.7.2 (quantitative value not specified, but implied by comparison) | 97.4% (96.3 - 98.5) | 96.9% (95.5 - 98.3) |
| Sensitivity (Specific Mode @ 80% Specificity) | Non-inferior & Superior to Predicate Device 1.7.2 | 95.2% (93.7 - 96.7) | 95.1% (93.3 - 96.8) |
| Sensitivity (Elevated Risk @ 97% Specificity) | Non-inferior & Superior to Predicate Device 1.7.2 | 80.8% (78.0 - 83.6) | 78.4% (75.1 - 81.7) |
| Exam-based AUC | Non-inferior & Superior to Predicate Device 1.7.2 | 0.960 (0.953 - 0.966) | 0.955 (0.947 - 0.963) |
Table 2: Acceptance Criteria (Implied by Performance Goals) and Reported Device Performance (Standalone with Temporal Analysis - TA)
| Metric | Acceptance Criteria (Implied/Target) | Reported Performance (FFDM with TA) | Reported Performance (DBT with TA) |
|---|---|---|---|
| Sensitivity (Sensitive Mode @ 70% Specificity) | Superior to performance without temporal comparison | 95.7% (93.7 - 97.6) | 94.6% (91.2 - 98.0) |
| Sensitivity (Specific Mode @ 80% Specificity) | Superior to performance without temporal comparison | 95.4% (93.4 - 97.4) | 91.0% (86.7 - 95.4) |
| Sensitivity (Elevated Risk @ 97% Specificity) | Superior to performance without temporal comparison | 82.7% (79.1 - 86.4) | 74.9% (68.3 - 81.4) |
| Exam-based AUC | Superior to performance without temporal comparison | 0.958 (0.946 - 0.969) | 0.941 (0.921 - 0.958) |
Study Details
-
Sample Size Used for the Test Set and Data Provenance:
- Main Test Set (without temporal analysis): 10,207 exams (5,730 FFDM, 4,477 DBT).
- Normal: 8,587 exams
- Benign: 270 exams
- Cancer: 1,350 exams (750 FFDM, 600 DBT)
- Temporal Analysis Test Set: 5,724 exams (4,266 FFDM, 1,458 DBT).
- Normal: 4,998 exams
- Benign: 83 exams
- Cancer: 643 exams (471 FFDM, 172 DBT)
- Data Provenance: Independent dataset acquired from multiple centers in seven EU countries and the US. Retrospective in nature, as it was acquired and not used for algorithm development and included normal exams with at least one year follow-up. The data included images from various manufacturers (Hologic, GE, Philips, Siemens, Fujifilm).
- Main Test Set (without temporal analysis): 10,207 exams (5,730 FFDM, 4,477 DBT).
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:
- The document states that the cancer cases in the test set were "biopsy-proven cancer." It does not specify the number or qualifications of experts used to establish the ground truth for the entire test set (including normal and benign cases, and detailed lesion characteristics). The mechanism for establishing the "normal" and "benign" status is not explicitly detailed beyond "normal follow-up of at least one year."
-
Adjudication Method for the Test Set:
- The document does not explicitly describe an adjudication method involving multiple readers for establishing ground truth for the test set. The ground truth for cancer cases is stated as "biopsy-proven."
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No, the document does not describe a Multi-Reader Multi-Case (MRMC) comparative effectiveness study. The performance assessment is a standalone evaluation of the algorithm's performance, not a human-in-the-loop study comparing human readers with and without AI assistance.
-
If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, standalone performance tests were conducted. The results presented in Tables 2 and 5 are for the algorithm's performance only.
-
The Type of Ground Truth Used:
- The primary ground truth for cancer cases is biopsy-proven cancer. For normal exams within the test set, the ground truth was established by "a normal follow-up of at least one year," implying outcomes data (absence of diagnosed cancer over a follow-up period).
-
The Sample Size for the Training Set:
- The document does not explicitly state the sample size of the training set. It mentions "Deep learning algorithms are applied to images for recognition of suspicious calcifications and soft tissue lesions... Algorithms are trained with a large database of biopsy-proven examples of breast cancer, benign abnormalities, and examples of normal tissue."
-
How the Ground Truth for the Training Set Was Established:
- The ground truth for the training set was established using a "large database of biopsy-proven examples of breast cancer, benign abnormalities, and examples of normal tissue." This implies a similar methodology to the test set for cancer cases (biopsy verification) and likely clinical follow-up or expert consensus for benign/normal cases, though not explicitly detailed for the training set.
Ask a specific question about this device