Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    DEN220063

    Validate with FDA (Live)

    Date Cleared
    2023-02-24

    (149 days)

    Product Code
    Regulation Number
    892.2055
    Type
    Direct
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    NEW REGULATION NUMBER: 21 CFR 892.2055

    CLASSIFICATION: Class II

    PRODUCT CODE: QVD

    BACKGROUND

    learning-based quantitative imaging software with predetermined change control plan Class: II Regulation: 21 CFR 892.2055

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Caption Interpretation Automated Ejection Fraction software is used to process previously acquired transthoracic cardiac ultrasound images, to store images, and to manipulate and make measurements on images using an ultrasound device, personal computer, or a compatible DICOM-compliant PACS system in order to provide automated estimation of left ventricular ejection fraction. This measurement can be used to assist the clinician in a cardiac evaluation.

    The Caption Interpretation Automated Ejection Fraction Software is indicated for use in adult patients.

    Device Description

    The Caption Interpretation Automated Ejection Fraction Software ("AutoEF") applies machine learning algorithms to process two-dimensional transthoracic echocardiography images for calculating left ventricular ejection fraction.

    The current implementation of the device adds a predetermined change control plan (PCCP) to the device cleared under K210747, which allows future modifications to the device.

    The version of Caption Interpretation AutoEF cleared under K210747 performs left ventricular ejection fraction estimation using apical four chamber (A4C), apical two chamber (A2C) and the parasternal long-axis (PLAX) cardiac ultrasound images.

    The software uses an algorithm that was derived through use of deep learning and locked prior to validation. The product operates as an add-in to a DICOM PACS system, ultrasound device, or personal computer. Caption Interpretation receives imaging data either directly from an ultrasound system or from a module in a PACS system.

    The device includes the following main components:

    1. Clip Annotation and Selection: The AutoEF software includes a function that processes video clips in a study to automatically classify clips that are PLAX. AP4, and AP2 views. This view selection is based on a convolutional network. It also includes a function, Image Quality Score (IQS), that allows selection of best available PLAX, AP4, and AP2 clips within the study or provide an indication to the user if there are no clips with sufficient quality to estimate ejection fraction, based on prespecified IQS thresholds.
    2. Eiection Fraction Estimator and Confidence Metric: The automated eiection fraction estimation is performed using a machine learning model trained on apical and parasternal long-axis views. The model is trained with a dataset from a large number of patients. representative of the intended patient population and variety of contemporary cardiac ultrasound scanners. The EF calculation can be performed on a 3-view combination (PLAX, AP4 and AP2), 2-view combinations (AP4 and AP2, AP2 and PLAX, AP4 and PLAX), or single views (AP4, AP2). The confidence metric provides expected error in left-ventricle ejection fraction estimation and is based on IQS.
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    80% positive predictive value (PPV) and 80% sensitivity for correct mode, view, and minimum number of frames (Clip Annotator).Observed PPV point estimates for the Clip Annotator were greater than 97% for identification of the imaging mode and the view. Observed sensitivity point estimates were greater than 96% across views and imaging mode. (Meets Criteria)
    80% of clips meet expert criteria for suitability for EF estimation (Clip Annotator).(This specific metric for "suitability for EF estimation" is not explicitly reported with a percentage in the provided text. The Clip Annotator performance for mode, view, and frames implies suitability, but a direct percentage for "expert criteria for suitability" is not given. However, the Clip Annotator study did meet its pre-defined acceptance criteria, suggesting this was addressed indirectly or considered acceptable based on the reported PPV and sensitivity.)
    Overall (all views) and all combined views Auto EF is within 9.2% RMSD of expert EF.Overall (best available view) RMSD EF% [95% CIs]: 7.21 [6.62, 7.74] (Meets Criteria) Combined Views: - AP4 and AP2: 7.27 [6.55, 7.92] (Meets Criteria) - AP4 and PLAX: 7.50 [6.85, 8.09] (Meets Criteria) - AP4, AP2, and PLAX: 7.24 [6.64, 7.80] (Meets Criteria) - AP2 and PLAX: 8.04 [7.32, 8.7] (Meets Criteria)
    New views superior to 11.024% RMSD individually.Individual Views: - AP4 only: 7.76 [7.01, 8.45] (Meets Criteria) - AP2 only: 8.27 [7.44, 9.03] (Meets Criteria) (Note: PLAX individual view is not explicitly reported here for comparison against this criterion, but the overall context implies good performance.)
    For each standard view, the Confidence Metric must meet the equivalence to expert EF criteria as defined in PCCP.Testing of the confidence metric functionality verified successful performance of the Confidence Metric in estimating the error range of the EF estimates around the reference EF using equivalence criteria and the evidence that the difference between the estimated EF and the reference EF is normally distributed. (Meets Criteria)

    Study Details

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: 186 patient studies.
    • Data Provenance: Retrospective, multicenter study. The studies were acquired from three sites across the US: Minneapolis Heart Institute, Duke University, and Northwestern University.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Not explicitly stated as a specific number of individual experts. The text refers to "a panel of expert readers" for the Clip Annotator study and "expert cardiologists" for establishing the EF reference standard.
    • Qualifications of Experts: "Expert cardiologists." No further specific details like years of experience are provided, but the title implies appropriate medical qualifications for interpreting echocardiograms.

    4. Adjudication method for the test set

    • Adjudication Method (Clip Annotator Study): "Results of the Clip Annotator were compared to evaluation by a panel of expert readers." This implies a consensus-based or direct comparison method, but the specific adjudication (e.g., 2+1, 3+1) is not detailed.
    • Adjudication Method (EF Calculation): The reference standard for ejection fraction was established by "expert cardiologists." This suggests expert consensus or established expert interpretation, but a formal adjudication process (e.g., how disagreements between experts were resolved if multiple experts reviewed the same case) is not explicitly described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Comparative Effectiveness Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with AI assistance vs. without AI assistance was not reported in this document. The clinical validation focused on comparing the AutoEF device's performance directly against an expert-established reference standard, not on the improvement of human readers using the device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Yes, the clinical validation study assessed the standalone performance of the Caption Health AutoEF. The "test compared the Caption Health AutoEF to the expert produced and reported biplane Modified Simpson's ejection fraction." This means the algorithm's output was directly compared to the ground truth without human intervention in the device's estimation process. The clinician's ability to edit the estimation is mentioned as a feature, but the presented performance results are for the automated estimation.

    7. The type of ground truth used

    • Ground Truth Type (Clip Annotator): Evaluation by a "panel of expert readers."
    • Ground Truth Type (Ejection Fraction): Reference standard for ejection fraction was established by 2D echo using the biplane Modified Simpson's method of disks, performed by "expert cardiologists."

    8. The sample size for the training set

    • Training Set Sample Size: The text states, "The model is trained with a dataset from a large number of patients." However, a specific numerical sample size for the training set is not provided. It also mentions training occurred on "a dataset from a large number of patients, representative of the intended patient population and variety of contemporary cardiac ultrasound scanners."

    9. How the ground truth for the training set was established

    • Training Set Ground Truth Establishment: The document does not explicitly detail how the ground truth for the training set was established. It only mentions that the machine learning model was "trained on apical and parasternal long-axis views" and derived through "deep learning," and "locked prior to validation." Given the nature of EF calculation, it is highly probable that expert cardiologists also established the ground truth for the training data, likely using methods similar to the test set (e.g., biplane Modified Simpson's method).
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1