    K Number
    K251408

    Validate with FDA (Live)

    Date Cleared
    2025-09-02

    (119 days)

    Product Code
    Regulation Number
    892.1171
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    KINGDOM

    Re: K251408
    Trade/Device Name: OsteoSight™ Hip (v1)
    Regulation Number: 21 CFR 892.1171
    OsteoSight™ Hip (v1)
    Classification: Class II
    Product Code: SAO
    Regulation: 21 CFR 892.1171

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OsteoSight™ is a software application intended for use opportunistically with standard anteroposterior (AP) radiographs of the hip or pelvis performed in patients aged 50 years and older. OsteoSight™ provides a notification in the form of a report to aid radiologists and/or physician interpreters in identifying patients with possible low bone mineral density (BMD) at the femoral neck to prompt a clinical assessment of bone health. OsteoSight™ should not be used to rule out low BMD. Radiologists and referring clinicians should follow recommended practices for screening and assessment, regardless of the absence of an OsteoSight™ report.

    Device Description

    OsteoSight is an AI-enabled software as a Medical Device (SaMD) that processes standard anteroposterior (AP) radiographs of the hip or pelvis in patients aged 50 years and older. The device operates via integration with a host integration platform that facilitates the secure transfer of DICOM images and associated metadata from clinical imaging systems.

    OsteoSight applies an AI-enabled algorithm pipeline to estimate bone mineral density (BMD) at the femoral neck. If an estimated BMD is below the pre-specified threshold, OsteoSight generates a report indicating that the patient is at risk of having low BMD. This report is delivered to the interpreting physician via the host integration platform. OsteoSight does not provide a diagnosis or treatment recommendation. If the estimated BMD is above or equal to the pre-specified threshold, no report is generated.

    The OsteoSight report is intended to inform physicians and support further clinical evaluation. Physicians may consider additional assessments, such as a comprehensive fracture risk assessment, in conjunction with other clinical information.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study that proves the device meets them, based on the provided FDA 510(k) Clearance Letter:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as quantitative thresholds in the document. However, the study aims to demonstrate that the device's sensitivity and specificity are sufficient to aid radiologists in identifying patients with possible low BMD. The reported device performance is as follows:

    MetricAcceptance Criteria (Implicit)Reported Device Performance (Intended Use Population)Reported Device Performance (Subset that produced a result)
    SensitivitySufficient to aid identification of patients with possible low BMD0.370 (95% CI 0.349-0.391)0.441 (95% CI 0.386-0.487)
    SpecificitySufficient to aid identification of patients with possible low BMD0.951 (95% CI 0.936-0.963)0.943 (95% CI 0.922-0.961)
    AUCNot applicable (for full Intended Use Population)n/a0.837 (95% CI 0.821-0.853)

    Notes on Acceptance Criteria: The document states, "All pre-specified acceptance criteria for the co-primary and secondary endpoints in the clinical studies were met." However, the document does not explicitly list these "pre-specified acceptance criteria" as quantitative targets for sensitivity, specificity, or AUC. The general implicit criterion is that the performance metrics (sensitivity, specificity, AUC) are deemed adequate to "aid radiologists and/or physician interpreters in identifying patients with possible low bone mineral density (BMD) at the femoral neck to prompt a clinical assessment of bone health."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size (Test Set): 3082 images
    • Data Provenance:
      • Country of Origin: United States (six independent cohorts from the Midwest, Northeast, South, and West).
      • Retrospective/Prospective: Retrospective analysis. The datasets included "previously acquired X-rays that included paired DXA T-scores."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not mention the use of experts to establish the ground truth for the test set.

    4. Adjudication Method for the Test Set

    Not applicable, as expert adjudication was not used. The ground truth was established by DXA scans.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The study focuses on the standalone performance of the AI device against a gold standard (DXA), not on how human readers improve with AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance study was done. The reported sensitivity, specificity, and AUC values are for the OsteoSight device's performance in identifying low BMD as determined by DXA, without human intervention in the diagnostic process beyond the device generating a report. The study measures the device's output (notification or no notification) against the ground truth.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: Dual-Energy X-ray Absorptiometry (DXA)
    • Definition of Low BMD: Defined as a T-score of < -1.0 SD as measured at the femoral neck from a DXA scan acquired within 12 months of the X-ray.

    8. The Sample Size for the Training Set

    The document does not specify the sample size for the training set. It only describes the validation (test) set.

    9. How the Ground Truth for the Training Set was Established

    The document does not provide information on how the ground truth for the training set was established. It only discusses the ground truth for the validation (test) set, which was based on DXA scans.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN230023

    Validate with FDA (Live)

    Device Name
    Rho
    Manufacturer
    Date Cleared
    2024-04-09

    (372 days)

    Product Code
    Regulation Number
    892.1171
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    NEW REGULATION NUMBER: 21 CFR 892.1171

    CLASSIFICATION: Class II

    PRODUCT CODE: SAO

    BACKGROUND

    Radiology software for opportunistic evaluation of low bone mineral density Regulation Number: 21 CFR 892.1171

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Rho is a software application intended for use opportunistically with standard frontal radiographs of the lumbar spine, thoracic spine, chest, pelvis, knee, or hand/wrist performed in patients aged 50 years and older. Rho provides a notification in the form of a report to aid radiologists and/or physician interpreters in identifying patients with possible low bone mineral density (BMD) at L1-L4 or the femoral neck to prompt a clinical assessment of bone health. Rho should not be used to rule out low BMD. Radiologists and referring clinicians should follow recommended practices for screening and assessment, regardless of the absence of Rho report.

    Device Description

    Rho is a machine learning-based software-as-a-medical device that interfaces with institutional Picture Archiving and Communications Systems (PACS) to identify patients 50 years and older undergoing x-ray with possible low bone mineral density (BMD). Eligible x-rays are frontal projections of the lumbar spine, thoracic spine, chest, pelvis, knee or hand/wrist. Rho uses the xray DICOM and DICOM tags of age and sex as inputs into a locked machine learning algorithm. The locked machine learning algorithm is trained on a patient-based dataset (True North Imaging. TNI13). The algorithm presents a binary output to indicate whether or not the patient likely has low BMD at either the femoral neck or L1-L4. A Rho Report is generated for positive cases that can be sent back to the PACS for physician interpretation or viewed through a browser-based interface.

    Radiologists can choose to include this finding in their report to the referring physician. Inclusion may trigger a referring physician to conduct a clinical fracture risk assessment related to bone health.

    For cases where Rho algorithm outputs a negative result, no Rho Report will be sent, and neither the radiologist nor the referring physician will receive any device output.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Performance Goals Proposed)Reported Device Performance (Specificity)Reported Device Performance (AUC)Reported Device Performance (Sensitivity)
    Lower end of the two-sided 95% confidence interval of specificity >0.775 (for TN16 and OMN cohorts)Met for all datasets and most subgroups.- Overall: - TNI: 0.90 (0.88-0.92) - OMN: 0.93 (0.90-0.96) - OAI: 0.94 (0.92-0.96)- Most subgroups (specificity generally above 0.85).- Lowest reported specificity: - Hand/wrist (OMN): 0.81 (0.67-0.95) - 80+ (TNI): 0.71 (0.62-0.80)Met for all datasets and all subgroups, as per statement.- Overall: - TNI: 0.88 (0.88-0.90) - OMN: 0.85 (0.82-0.88) - OAI: 0.82 (0.79-0.85)- Most subgroups (AUC generally above 0.80).- Lowest reported AUC: - Hand/wrist (OMN): 0.73 (0.61-0.85) - 50-59 (OAI): 0.77 (0.70-0.84) - Hand/wrist (OAI): 0.77 (0.70-0.82) - 70-79 (OAI): 0.78 (0.72-0.83) - Males (OAI): 0.79 (0.75-0.84)Not met for all datasets and some subgroups.Original goal: Lower end of the two-sided 95% confidence interval of sensitivity >0.5 (for TN16 and OMN cohorts).- Overall: - TNI: 0.67 (0.66-0.69) - Met - OMN: 0.45 (0.41-0.50) - Not Met - OAI: 0.36 (0.31-0.41) - Not Met- Specific Subgroups with Sensitivity < 30% noted: - Knee subgroup (OMN and OAI) - Hand/wrist subgroup (OAI) - Male subgroup (OAI)- Examples of low sensitivity: - OMN (Males): 0.41 (0.34-0.49) - OAI (Females): 0.44 (0.37-0.50) - OAI (Males): 0.22 (0.15-0.30) - OMN (50-59): 0.43 (0.33-0.53) - OAI (50-59): 0.26 (0.15-0.38) - OAI (Knee): 0.28 (0.21-0.36) - OMN (Knee): 0.29 (0.18-0.39) - OMN (BMI ≥ 30): 0.29 (0.20-0.37) - OAI (Hand/wrist): 0.24 (0.15-0.33) - OAI (Hispanic): 0.20 (0.00-0.50) - OAI (Asian): 0.00 (0.00-0.00) - OAI (GE Manufacturer): 0.17 (0.11-0.24)

    Note regarding "Acceptance Criteria": The text states, "The sponsor originally proposed co-primary endpoints of sensitivity and specificity (performance goals specified as the lower end of the two-sided 95% confidence interval of specificity >0.775 and lower end of the two-sided 95% confidence interval of sensitivity >0.5 for the TN16 and OMN cohorts. However, the final device performance across all analysis groups did not meet those goals." It then immediately states, "The device met its performance goals of specificity and AUC in all three datasets." This implies the final accepted performance goals were a high specificity and meeting the AUC goal, despite not fully meeting the originally proposed sensitivity goal. The approval indicates the FDA found the benefit-risk acceptable given the high specificity and AUC, and the context of use where low sensitivity is mitigated because patients continue to receive standard of care.

    Study Information:

    1. A table of acceptance criteria and the reported device performance: (See table above)

    2. Sample sizes used for the test set and the data provenance:

      • Total Clinical Validation Set: 4842 cases (3729 TNI6 + 522 OMN + 591 OAI).
      • Data Provenance:
        • TNI6 (Clinical Validation Set): 3729 cases from six True North Imaging centers, Canada (retrospective).
        • OMN (Clinical Validation Set): 522 US cases from OneMedNet (retrospective).
        • OAI (Clinical Validation Set): 591 US cases from Osteoarthritis Initiative, a multicenter, longitudinal, prospective observational study of knee osteoarthritis. This data was "previously acquired," suggesting it was used retrospectively for this device validation.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The text does not mention the use of experts for ground truth establishment.
      • Ground truth was established by Dual Energy X-ray Absorptiometry (DXA).
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable, as the ground truth was established by DXA, not by human experts requiring adjudication.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done to assess human reader improvement with AI assistance. The study described is a standalone (algorithm only) performance study against a ground truth. The device is intended to "aid radiologists and/or physician interpreters" but its direct impact on human reader performance was not evaluated in this study.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone performance study was done. The described clinical performance testing directly evaluates the algorithm's Sensitivity, Specificity, and AUC as a binary classifier (yes/no low BMD) against the DXA ground truth.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Ground Truth: Dual Energy X-ray Absorptiometry (DXA).
      • Definition of Low BMD: At least one of the femoral necks or L1-L4 vertebrae having a T-Score < -1.0 when assessed by DXA.
      • The text states: "Ground truth low BMD is defined as at least one of the femoral necks or L1-L4 having a T-Score < - 1.0 when assessed by DXA. DXA data acquired from the same patient is currently used in the patient bone mineral density screening program, and it is the current standard of care and well-established in clinical practice."
    8. The sample size for the training set:

      • The locked machine learning algorithm was trained on a patient-based dataset called True North Imaging (TNI13).
      • The specific sample size for TNI13 is not explicitly stated in the provided text. It is only mentioned that it was "used exclusively for algorithm development" and was a "split by geographically separated imaging clinics" from the TNI clinical validation set.
    9. How the ground truth for the training set was established:

      • While not explicitly detailed for the TNI13 training set, given the consistent methodology described for the validation sets, it is highly probable that the ground truth for the training set (TNI13) was also established by DXA with the same definition of low BMD (T-Score < -1.0). The device description states the algorithm is "trained on a patient-based dataset (True North Imaging. TNI13)" and the "clinical performance of the device was provided in the form of a retrospective, multi-center study to compare sensitivity and specificity of the device with the gold standard. Dual Energy X-ray Absorptiometry (DXA) with pre-specified performance goals." This implies the same ground truth approach across development and validation.