    Why did this record match?
    510k Summary Text (Full-text Search) :

    NEW REGULATION NUMBER: 21 CFR 888.3090

    CLASSIFICATION: Class II

    PRODUCT CODE: QFP

    BACKGROUND

    Product Code: OFP Device Type: Intraoperative orthopedic strain sensor Class: II Regulation: 21 CFR 888.3090

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LOADPRO™ Intraoperative Rod Strain Sensor is an intraoperative surgical tool that allows surgeons to measure unidirectional rod microstrain on posterior rods in the sagittal plane when performing spine surgery. This device is an adjunct to surgeon tactile feedback and is not intended to replace a surgeon's clinical judgment.

    The LOADPRO™ Intraoperative Rod Strain Sensor is a single use, disposable tool to be used in conjunction with X-Spine Systems Fortex Pedicle Screw System for 5.5mm diameter titanium (ASTM F136) or cobalt chrome (ASTM F1537) rod configurations.

    Device Description

    The LOADPRO™ Intraoperative Rod Strain Sensor includes a titanium/zirconia ceramic, single use strain sensing device, which includes radio-frequency identification (RFID) technology (13.56MHz), intended to enable access to strain measurement values, incorporating a passive transponder, inserter, and scanner (Figure 1 and 2). The transponder attaches to X-Spine Systems Fortex Pedicle Screw System using 5.5mm titanium alloy or cobalt chrome rods in corrective spinal surgeries (Figure 3). The transponder is used only to acquire rod microstrain values and a unique device identification code, which is read by the scanner, during the surgical correction.

    The LOADPRO™ Intraoperative Rod Strain Sensor is a titanium/zirconia ceramic, ethylene oxide (EtO) sterilized, single use device designed to provide objective readings of the change of mechanical unidirectional strain on a pedicle screw rod. The LOADPRO™ Intraoperative Rod Strain Sensor consists of the following primary components, described in the proceeding sections:

    • LOADPROTM Intraoperative Rod Strain Sensor ●
    • . Hand Held Reader (scanner)
    • . Manual Orthopaedic Surgical Instrumentation
    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the LOADPRO™ Intraoperative Rod Strain Sensor, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Note: The provided document does not explicitly present a "table of acceptance criteria and the reported device performance" in the format of a direct comparison for all performance metrics. Instead, it describes various tests and their successful outcomes against implied or stated criteria. The table below synthesizes the information available.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Implied/Stated)Reported Device Performance
    Rod DurabilityNon-abrasive clamping, does not impact mechanical integrity of the rod or rod/screw construct fatigue life.Dynamic Compression Bend Testing: No change between sensored and non-sensored rods after 15 installations/removals (runout to 5M cycles, equivalent to Fortex 510(k)).Static Cantilever Bend Testing: No meaningful difference in strength and stiffness.
    Sensor Limit TestingRod strain value > yield strain for each material (Ti > b strain, CoCr > b strain) when bridge sees 2000 u strain on the bridge.Passed (implied by "The acceptance criterion requires the rod strain value...").
    Sensor Variability TestingMeasure strain linearly and are not load dependent; acceptable sensor-to-sensor resistance variability.Resistance Variability: -50N: 9.9%, -100N: 6.2%, -150N: 5.8%, -200N: 5.2%.Linearity: Graph "Sensor Sensitivity ASTM F1717 Testing Micro Strain Vs. Load" shows linear behavior (Average Slope y = 9.9569x, R^2 = 1).
    Sensor Longevity TestAccuracy of sensor when seated on the rod construct for 8 hours with strain readings recorded every hour.Passed (implied by this being a "Test").
    Sensor Temperature Variability TestSensors will not dramatically change in strain value as the environmental temperature fluctuates within typical OR limits (65°F, 70°F, 75°F).Passed (implied by this being a "Test").
    Reader Duration TestReader can function without battery change for the entire duration of a typical surgery (5 minute intervals up to 120 minutes of total scanning time).Passed (implied by this being a "Test").
    Accuracy and Repeatability TestingAccurately and repeatably measure load on a cantilevered rod (e.g., when the mounted strain gage recorded 1400 u-strain in Titanium and Cobalt/Chrome rods); microstrain reading errors < 10% despite potential error sources (coronal curvature, off-axis loading, wetted rods, varying ambient temp).Ti Rods: All tested Sensors passed the success criteria (specific %Error values for individual sensors are redacted).CoCr Rods: All tested Sensors passed the success criteria (specific %Error values for individual sensors are redacted).Additional Testing: Individual and combined effects resulted in microstrain reading errors of less than 10%.
    Usability Study (Table Top)Adjusted total score of 70 or greater (scale of 0-100) for Instruments, Sensor, and Reader applications.Instruments: $89.5 \pm 12.8$Sensor: $84.5 \pm 12.3$Reader: $85.7 \pm 13.2$Average score for each application was above acceptance criteria.
    Usability Study (Intraoperative)No significant prolongation of operative times; negligible estimated blood loss attributed to device use.Operative Time: 16min 30sec, 7min 12sec, 11min 12sec, 8min 30sec, 11min 55sec, 13min 19sec for LOADPRO use (does not appear to significantly prolong operative times).Blood Loss: 0, <5, <5, <5, 5 to 50, 5 to 50 cc during LOADPRO use (negligible).
    Software VerificationAll features pertaining to collecting and displaying sensor readings pass testing acceptance criteria.All features passed.
    Electromagnetic Compatibility & Electrical SafetyCompliance with EN 60601-1 and IEC 60601-1-2:2007.Test results support EMC and electrical safety.
    BiocompatibilityNo evidence of causing cell lysis or toxicity.Cytotoxicity testing showed no evidence of causing cell lysis or toxicity.
    Sterilization & Shelf LifeSterility assurance level (SAL) of 10^-6; package integrity through dye penetration and burst testing over 12 months.Validated per ISO 11135 (SAL of 10^-6); 12-month shelf life verified.
    Reprocessing (Instruments)Sterility assurance level (SAL) of 10^-6.Validated per AAMI ST79 and ISO17665-1 (SAL of 10^-6).

    Study Details

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Rod Durability Testing (Dynamic): Not explicitly stated, but performed on "non-sensored rods and on sensored rods after fifteen (15) LOADPRO™ Intraoperative Rod Strain Sensor installations / removals were performed at the same location."
    • Rod Durability Testing (Static): 3 samples per group (non-sensored vs. sensored).
    • Sensor Variability Testing: The graph shows 6 different sensors (SN 1, SN 2, SN 4, SN 5, sn 3, SN 9).
    • Accuracy and Repeatability Testing: "All tested Sensors" passed (implied multiple sensors). Specific tables for Ti and CoCr show redacted serial numbers, implying multiple units per rod type.
    • Usability Study (Table Top): 15 users (orthopedic spine surgeons and fellows).
    • Usability Study (Intraoperative): 5 patients/surgeries (one patient had surgery over two separate days, resulting in 6 surgeries), 4 participating surgeons, 13 sensors used.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be prospective bench testing and a prospective intraoperative usability study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Bench Testing (Rod Durability, Sensor Characterization, Accuracy): Ground truth was established by comparing the LOADPRO™ sensor readings to established mechanical testing methods and reference strain gauges. No human experts were involved in establishing the "ground truth" for these physical measurements, as the measurements are objective.
    • Usability Testing (Table Top): 15 practicing orthopedic spine surgeons and fellows. Their "feedback" was the ground truth for usability, graded on a modified usability scale.
    • Usability Testing (Intraoperative): 4 participating surgeons. Their feedback, related to operative time and blood loss, contributed to the usability assessment.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Bench Testing: Not applicable, as objective physical measurements were taken. Discrepancies would be resolved by re-testing or calibration.
    • Usability Testing (Table Top): A specific scoring system was used (1-5 scale converted to 0-100, with an adjusted total score of 70 or greater as acceptable). A score less than 70 would "require a review by the sponsor and adjustment based on the user's feedback." This implies a form of internal review or adjudication if criteria were not met.
    • Usability Testing (Intraoperative): The text indicates "surgeon feedback" was used, but does not detail a formal adjudication method for that feedback.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an intraoperative strain sensor, not an AI-powered diagnostic tool. It provides objective measurements and is an "adjunct to surgeon tactile feedback," not a replacement or an AI assistance tool for image interpretation. Therefore, discussions of "human readers" or "AI assistance" are not applicable in this context.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Yes, the performance testing for the LOADPRO™ Intraoperative Rod Strain Sensor largely represents standalone performance. The various bench tests (Rod Durability, Sensor Limit, Sensor Variability, Sensor Longevity, Sensor Temperature Variability, Reader Duration, Accuracy and Repeatability) assessed the device's ability to accurately and reliably measure strain and function independently. The usability studies, while involving surgeons, focused on their interaction with the device and the device's impact on the procedure, rather than the device's performance being dependent on their interpretation for its core function. The "algorithm" here is the embedded software that processes strain data, and its direct performance was verified.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Mechanical Performance (Rod Durability, Sensor Characterization, Accuracy, etc.): The ground truth was established through objective physical measurements using established metrology and standards. This typically involves calibrated load cells, reference strain gauges, and standardized test fixtures (e.g., ASTM F1717-13, ASTM F2193-2).
    • Software Verification: Ground truth was based on the functional specifications of the software.
    • Biocompatibility, EMC, Sterilization, Shelf Life: Ground truth was based on international standards and validated test methods (e.g., ISO 10993-5, EN 60601-1, IEC 60601-1-2, ISO 11135, ASTM F1980-07, AAMI ST79, ISO17665-1).
    • Usability Testing: Ground truth was based on end-user (surgeon) feedback against a predefined usability scale and qualitative observations regarding operative time and blood loss.

    8. The sample size for the training set

    • Not applicable. The LOADPRO™ Intraoperative Rod Strain Sensor is a physical measurement device with embedded software, not a machine learning or AI algorithm that requires a separate "training set" in the conventional sense. The "training" for the device's functionality would be its design, calibration, and manufacturing process.

    9. How the ground truth for the training set was established

    • Not applicable, as there is no "training set" for an AI/ML algorithm. The device's calibration and design are based on fundamental physics and engineering principles, validated through benchmark testing against known physical standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K142847

    Validate with FDA (Live)

    Date Cleared
    2014-10-27

    (27 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Spinal Intervertebral Body HzcvkqpOrthosis, and Spinal'Ipvgtimo lpcnHkzcvkqp'Qt vj qulu'(21 CFR888.3090

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CD HORIZON® Spinal System without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to following indications: degeneralive discase (defined as back pain of discogenic origin with degencration of the disc confirmed by history and radios); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis (i.e., scoliosis, kyphosis and/or lordosis); turnor, pseudarthrosis; and/or failed previous fusion.

    Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.

    With the exception of degenerative discase, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforcementioned indications in skeletally mature patients as an adjunct to fusion. The 3,5mm rods may be used for the specific pediatric indications noted below.

    When used for posterior non-cervical pedicle serew lixation in pediatic patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to treat progressive spinal deformities (i.e., scoliosis, kyphosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scollosis, Additionally, the CD HORIZON® Spinal System is intended to treat patients diagnosed with the following conditions: spondylolistlesss / spondylolysis and fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are to be used with autograft and/or all. Pediatric pedicle screw lixation is limited to a posterior approach.

    The CD HORIZON® SPIREIM Plate is a posterior, single level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T)-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/ attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disease (as previously defined); spondylolisthesis; trauma; and/or tumor.

    In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.

    Device Description

    The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® Plates, staples and connecting components, as well as implant components from other Medtronic spinal systems, which can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.

    A subset of CD HORIZON® Spinal System components may be used for posterior pedicle screw fixation in pediatric cases. These constructs may be comprised of a variety of shapes and sizes of rods (ranging in diameter from 3.5mm to 6.35mm), hooks, screws, CROSSLINK® Plates, and connecting components. Similarly to the CD HORIZON® implants used in adult cases, these components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.

    Certain components within the CD HORIZON® Spinal System are specifically excluded for use in pediatric patients. These include PEEK rods, Shape Memory Alloy Staples, SPIRE™ Plates and DYNALOK® bolts. All screws used in pediatric cases are only cleared for use via a posterior approach. All of the components used in pediatric cases are fabricated from medical grade stainless steel, medical grade titanium, titanium alloy, and medical grade cobalt-chromiummolybdenum alloy.

    Certain implant components from other Medtronic spinal systems can be used with the CD HORIZON® Spinal System in non-pediatric cases. These components include TSRH® rods, hooks, screws, plates, CROSSLINK® plates, connectors, staples, washers, GDLH® rods, hooks, connectors and CROSSLINK® bar and connectors; LIBERTY® rods and screws; DYNALOK® PLUS and DYNALOK CLASSIC® bolts along with rod/bolt connectors; and Medtronic Multi-Axial rods and screws. Please note that certain components are specifically designed to connect to specific rod diameters, while other components can connect to multiple rod diameters. Care should be taken so that the correct components are used in the spinal construct.

    CD HORIZON® hooks are intended for posterior use only. CD HORIZON® staples and CD HORIZON® ECLIPSE® rods and associated screws are intended for anterior use only. However, for patients of smaller stature and pediatric patients, CD HORIZON® 4.5mm rods and associated components may be used posteriorly.

    The CD HORIZON® Spinal System implant components are fabricated from medical grade stainless steel, medical grade titanium alloy, medical grade cobalt-chromiummolybdenum alloy, or medical grade PEEK OPTIMA-LT1. Certain CD HORIZON® Spinal System components may be coated with hydroxyapatite.

    The CD HORIZON® Spinal System also includes anterior staples made of Shape Memory Alloy (Nitinol - NiTi). Shape Memory Alloy is compatible with titanium alloy, and cobaltchromium-molybdenum alloy. Do not use with stainless steel. These staples are not to be used in pediatric patients.

    PEEK OPTIMA-LT1 implants may be used with stainless steel, titanium, or cobalt-chromiummolybdenum alloy implants. CD HORIZON® PEEK Rods are not to be used with CROSSLINK® Plates or in pediatric patients.

    The purpose of this 510(k) is to add additional sterile rods to the CD HORIZON® Spinal System.

    The subject rods, along with other components such as metal screws, hooks and other connecting components, are intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of thoracic, lumbar, and/or sacral spine. The subject rods are manufactured out of Commercially Pure (CP) Titanium.

    AI/ML Overview

    This document is a 510(k) summary for the CD HORIZON® Spinal System, specifically addressing the addition of sterile rods. It does not contain information about acceptance criteria or a study proving the device meets those criteria, as it is a premarket notification for substantial equivalence, not a performance study report.

    Therefore, I cannot provide the requested information from the given text.

    The document primarily focuses on:

    • Device Description: What the CD HORIZON® Spinal System is composed of.
    • Indications for Use: The medical conditions for which the device is intended.
    • Technological Characteristics: How the device is similar to its predicate.
    • Substantial Equivalence: The argument that the new sterile rods are substantially equivalent to previously cleared versions and predicates.
    • Non-Clinical Testing: A mention of a sterilization assessment and shelf-life rationale, which are part of establishing substantial equivalence rather than proving specific performance acceptance criteria for a novel device function.