Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K153679

    Validate with FDA (Live)

    Date Cleared
    2016-01-22

    (32 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    ClassificationName | Electrosurgical, cutting & coagulation & accessories(21 CFR 884.4400
    RegulationNumber | 21 CFR 884.4400
    | 21 CFR 884.4400

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neptune E-SEP is an Integrated Smoke Evacuation Pencil) designed for general electrosurgical applications including cutting and coagulation, and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit (generator) to the operative site for the desired surgical effect.

    Device Description

    The Stryker® Neptune® E-SEPT™ Smoke Evacuation Pencil is a single use monopolar electrosurgical pencil that includes a handpiece with a smoke evacuation function and a range of electrodes (both uncoated and coated). The electrosurgical pencil is compact in size (190 mm long and 17 mm diameter) with 3 m of plastic smoke evacuation tubing and associated electrical cable housed within the tubing. The device is externally powered via an external power generator (supplied by user) connected to a monopolar receptacle using a conventional 3-pin connector. This device remains unchanged from the cleared predicate pencil.

    The Stryker® Neptune® E-SEPT™ Smoke Evacuation Pencil is available in two activation switch configurations, a rocker style and push-button style, which activates monopolar cut or coagulate functions operated by pressing the respective button. The pencil is connected to smoke evacuation tubing which features a dual connector (8 and 22 mm) to allow the user to connect to a variety of smoke evacuation systems including filtration or central vacuum systems, thus minimizing exposure of personnel to surgical smoke plume.

    Electrodes are available in Blade, Needle, Ball, T-Bar Loop, U-Bar Loop and Conization geometries and are listed in Table 6-1 below

    AI/ML Overview

    This document details the acceptance criteria and study findings for the Stryker Neptune E-SEP Smoke Evacuation Pencil, a medical device for electrosurgical applications.

    The information below is based on the provided 510(k) Summary (K153679).

    1. Table of Acceptance Criteria and Reported Device Performance

    Test AreaAcceptance Criteria (Implied by equivalence to predicate and industry standards)Reported Device Performance
    Electrical SafetyCompliance with IEC 60601-1 and IEC 60601-2-2 (General requirements for basic safety and essential performance of medical electrical equipment)Tests demonstrated compliance with IEC 60601-1 and IEC 60601-2-2.
    Functional / PerformanceDeliver intended electrosurgical cutting and coagulation, and effectively evacuate smoke. Consistent performance with predicate device.Tests demonstrated functionality, integrity, and safety/effectiveness for intended use, supporting substantial equivalence.
    Thermal SpreadThermal characteristics consistent with the predicate device and within acceptable safety limits.Testing was conducted to evaluate thermal spread, and results supported substantial equivalence. (Specific numerical criteria/results not provided beyond a general statement of sufficiency)
    BiocompatibilityMaterials used are biocompatible and do not pose undue risk to patients or users. Compliance with relevant biocompatibility standards.Biocompatibility testing was performed, and results supported substantial equivalence.
    Sterilization & PackagingDevice remains sterile, and packaging maintains sterility until point of use. Sterility Assurance Level (SAL) of $10^{-6}$.Sterilization and packaging testing confirmed sterile single use, EtO sterilization with SAL = $10^{-6}$, and maintained sterility.
    Mechanical IntegrityElectrodes and pencil maintain structural integrity during use. Suitable for electrosurgical applications.Testing demonstrated the integrity of the electrodes. (Specific criteria/results not provided beyond a general statement of sufficiency)
    Material EquivalenceNew materials (ABS Plastic, 304 Series Stainless Steel) do not negatively impact safety or performance compared to predicate materials.Performance testing (including electrical and functional) covered devices with new materials, confirming their suitability.
    Working Length EquivalenceLonger electrode working lengths (70mm - 165mm) perform comparably to predicate (70mm) in terms of safety and effectiveness.Performance testing covered various working lengths, supporting similar safety and effectiveness.
    Geometry EquivalenceNew electrode geometries (Needle, Ball, T-Bar Loop, U-Bar Loop, Conization) perform comparably to predicate (Blade, Needle-compatibility) in safety and effectiveness.Performance testing covered various geometries, supporting similar safety and effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes (number of devices or tests) for each non-clinical test conducted. It states that "Performance testing was conducted on the proposed devices as determined by the risk analysis for the products."

    The data provenance is from non-clinical bench testing performed by the manufacturer, Stryker Instruments. The country of origin for the data is not explicitly stated but is implicitly associated with the manufacturer's testing facilities. The tests are prospective in the sense that they were conducted specifically for this 510(k) submission to demonstrate the performance of the modified device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This submission relies on non-clinical performance testing (bench testing) rather than studies involving expert interpretation of medical images or diagnostic outcomes. Therefore, there were no experts used to establish ground truth in the traditional sense of clinical or diagnostic studies. The "ground truth" for these engineering and performance tests is defined by established engineering standards (e.g., IEC standards) and the performance characteristics of the predicate device.

    4. Adjudication Method for the Test Set

    As this was non-clinical bench testing, an adjudication method for a "test set" in the context of expert review or consensus (like 2+1 or 3+1) is not applicable. The outcome of each test was determined by whether the device met the pre-defined engineering and performance specifications.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This submission pertains to a medical device's electrosurgical and smoke evacuation capabilities, and the modifications are extensions of electrodes and material changes. Clinical studies, including MRMC studies, were deemed "not necessary for this 510(k)." The comparison was primarily based on non-clinical performance data and substantial equivalence to the predicate device.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

    This device is an electrosurgical pencil, a physical medical instrument. Therefore, the concept of "standalone (algorithm only without human-in-the loop performance)" as it applies to AI/software devices is not applicable. The performance testing described is inherently for the device itself.

    7. The Type of Ground Truth Used

    The "ground truth" for the non-clinical performance tests was established by:

    • Engineering and Performance Standards: Compliance with recognized international standards such as IEC 60601-1 and IEC 60601-2-2 for electrical safety.
    • Predicate Device Performance: The functional, thermal, biocompatibility, and sterilization performance of the legally marketed predicate device (SafeAir Smoke Pencil, K143145) served as the benchmark for substantial equivalence. The proposed device's performance was compared to ensure it was at least as safe and effective.

    8. The Sample Size for the Training Set

    This medical device is a physical instrument, not an AI or machine learning algorithm. Therefore, the concept of a "training set" in the context of data for model development is not applicable.

    9. How the Ground Truth for the Training Set Was Established

    As explained above, there is no training set for this type of medical device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K100517

    Validate with FDA (Live)

    Date Cleared
    2010-04-29

    (65 days)

    Product Code
    Regulation Number
    884.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    GYNECOLOGICAL DEVICES 21 CFR PART 884 Subpart E--Obstetrical and Gynecological Surgical Devices

    Sec. 884.4400

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    INSTRUMED Obstetrical Forceps are intended to grasp and apply traction to the fetal head to facilitate delivery for the following indications, provided that the fetal head is positioned appropriately in the vagina: · prolonged second stage - · suspicion of immediate or potential fetal compromise - · shortening of the second stage for maternal benefit

    Device Description

    Instrumed obstetrical forceps are instruments designed to aid in the delivery of the fetus. Many different types of forceps have been described and developed. Generally, forceps consist of 2 mirror image metal instruments that are manoeuvred to cradle the fetal head and are articulated, after which traction is applied to effect delivery. Forceps have 4 major components, as follows: - Blades: The blades grasp the fetus. Each blade has a curve to fit around the fetal head. The . blades are oval or elliptical and can be fenestrated (with a hole in the middle) or solid. Many blades are also curved in a plane 90° from the cephalic curve to fit the maternal pelvic curve). - Shanks: The shanks connect the blades to the handles and provide the length of the device. They . are either parallel or crossing. - Lock: The lock is the articulation between the shanks. Many different types have been designed. . - Handles: The handles are where the operator holds the device and applies traction to the fetal . head. The surgeon chooses the obstetrical/gynaecological forceps based on the anatomy of the site and the type desired, based on the type of the surgical procedure. Instrumed obstetrical forceps are made of the ASTM F 899-09 standardized Stainless Steel. The instruments are offered in non-sterile condition.

    AI/ML Overview

    The provided text describes a 510(k) submission for Instrumed Obstetrical Forceps. It details the device's description, indications for use, and a comparison to predicate devices to establish substantial equivalence.

    Based on the content provided, the acceptance criteria and the study proving the device meets them are articulated in terms of equivalence to predicate devices, rather than specific performance metrics.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (Implicitly based on Predicate Device Equivalence)Reported Device Performance
    Technical CharacteristicsComparable to predicate devices (TEKNO-MEDICAL OBSTETRICAL FORCEPS K013747, V. MUELLER OBSTETRICAL (OB) FORCEPS K951529)"The Instrumed product is similar to the predicate device in terms of technical characteristics, design..."
    DesignComparable to predicate devices"The Instrumed product is similar to the predicate device in terms of technical characteristics, design..."
    Indications for UseIdentical or substantially similar to predicate devicesIdentical indications for use: "to grasp and apply traction to the fetal head to facilitate delivery for... prolonged second stage, suspicion of immediate or potential fetal compromise, shortening of the second stage for maternal benefit."
    Target PopulationComparable to predicate devices"The Instrumed product is similar to the predicate device in terms of... Target population..."
    Usage EnvironmentComparable to predicate devices"The Instrumed product is similar to the predicate device in terms of... where it is used..."
    PerformanceComparable to predicate devices"The Instrumed product is similar to the predicate device in terms of... performance..."
    BiocompatibilityAcceptable in comparison to predicate devices"Biocompatibility and sterilization studies were successfully completed." "The used materials are well researched and do not raise new questions regarding safety and effectiveness of the finished product."
    Sterilization MethodAcceptable in comparison to predicate devices"The Instrumed product is similar to the predicate device in terms of... sterilisation method..." (Offered in non-sterile condition, implying sterilization by the user is required/expected, similar to predicate).
    Mechanical SafetyComparable to predicate devices"The Instrumed product is similar to the predicate device in terms of... mechanical safety characteristics..."
    MaterialsASTM F 899-09 standardized Stainless Steel, deemed safe and effective"Instrumed obstetrical forceps are made of the ASTM F 899-09 standardized Stainless Steel." "The used materials are well researched and do not raise new questions regarding safety and effectiveness of the finished product."
    New Safety/Effectiveness QuestionsNo new questions regarding safety or effectiveness should be raised."The presented data that was conducted on the Instruments shows... that the products are absolutely safe and effective for their intended use and do not raise any new questions regarding safety and effectiveness."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a test set or data provenance in the context of clinical studies with human participants. The evaluation relies on non-clinical and bench testing to demonstrate similarity to predicate devices. Therefore, traditional "sample size" for a clinical test set is not applicable here.

    • Sample Size for Test Set: Not applicable as no human clinical test set is described. The evaluation is primarily based on non-clinical and bench testing of the device itself and comparison to predicate devices.
    • Data Provenance: Not explicitly stated as the evaluation focuses on the physical device and its properties, materials, and design in comparison to existing devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. Given that the submission is for substantial equivalence based on non-clinical and bench testing and comparison to predicate devices, there's no mention of a ground truth established by medical experts for a test set of patient data.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. As no human clinical test set requiring expert ground truth or adjudication is described, this section is not applicable.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or described in this submission. The device is a surgical instrument (forceps), and its evaluation is based on material properties, design, and functional similarity to already marketed predicate devices, not on the interpretation of medical images by multiple readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable as the device is a manual surgical instrument, not an algorithm or AI system. Its performance inherently involves a human operator (the surgeon).

    7. The Type of Ground Truth Used

    The "ground truth" in this context is implicitly established by the known safety and effectiveness of the predicate devices and the standards for material properties (e.g., ASTM F 899-09 for stainless steel). The claim is that the Instrumed Obstetrical Forceps functions equivalently to these well-established devices and uses appropriate materials, thus achieving safety and effectiveness.

    • Implicit Ground Truth: Safety and effectiveness of legally marketed predicate devices and compliance with material standards.

    8. The Sample Size for the Training Set

    This information is not applicable as there is no mention of an algorithm or AI system that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no mention of an algorithm or AI system that requires a "training set" or its associated ground truth establishment.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092827

    Validate with FDA (Live)

    Date Cleared
    2010-01-08

    (116 days)

    Product Code
    Regulation Number
    884.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Forceps / K092827

    Common name: Obstetrical forceps

    Classfication name: Obstetric forceps (21 CFR 884.4400
    78532

    JAN - 8 2010

    Re: K092827

    Trade/Name: Dimeda Obstetrical Forceps Regulation Number: 21 CFR §884.4400

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dimeda Obstetrical Forceps are intended to grasp and apply traction to the fetal head to facilitate vaginal delivery, provided that the cervix is fully dilated and the fetal head is positioned appropriately in the vagina. Use of the forceps is indicated for:

    • 0 prolonged second stage
    • 0 suspicion of immediate or potential fetal compromise
    • shortening of the second stage for maternal benefit 0
    Device Description

    The Dimeda Obstetrical Forceps is a surgical medical device for medical procedure to grasp the fetal head when it is in the vagina and bring about delivery by traction and guidance without causing injury to the mother or baby. The forceps consists of two arms which are movable. The device is manufactured from surgical stainless steel.

    AI/ML Overview

    This submission for Dimeda Obstetrical Forceps (K092827) is a 510(k) premarket notification, which establishes substantial equivalence to previously cleared devices. It does not present a study with specific acceptance criteria and detailed device performance metrics in the way a clinical trial or performance study for a novel device would. Instead, it relies on demonstrating equivalence in intended use, design, materials, and performance to predicate devices already on the market.

    Therefore, many of the requested sections (e.g., sample size for test set, number of experts, MRMC study, training set details) are not applicable or cannot be extracted from this type of submission.

    Here's a breakdown based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a substantial equivalence submission for a traditional surgical instrument, there are no explicit "acceptance criteria" for novel performance features or quantitative reported device performance in the document. The acceptance criterion is the determination of "substantial equivalence" to the predicate devices, implying that the Dimeda Obstetrical Forceps perform as safely and effectively as the predicates.

    Acceptance CriteriaReported Device Performance
    Safety and Effectiveness (implicit acceptance criterion for substantial equivalence)The Dimeda Obstetrical Forceps are "absolutely safe and effective for their intended use and do not raise any questions regarding safety and effectiveness" based on comparison to predicate devices and long market presence in Europe.
    Intended Use EquivalenceIdentical to predicate devices (grasp and apply traction to fetal head to facilitate vaginal delivery).
    Indications for Use EquivalenceIdentical to predicate devices (prolonged second stage, suspicion of fetal compromise, shortening of second stage for maternal benefit).
    Target Population EquivalenceIdentical to predicate devices.
    Hospital Use EquivalenceIdentical to predicate devices.
    Material EquivalenceManufactured from surgical stainless steel, well-researched materials, identical to predicate devices.
    Design EquivalenceConsists of two movable arms, identical to predicate devices.
    Biocompatibility EquivalenceIdentical to predicate devices.
    Sterilization Method EquivalenceIdentical to predicate devices.
    Performance EquivalenceIdentical to predicate devices (no modifications regarding design, method of operation, mechanical performance affecting safety and effectiveness).
    Mechanical Safety Characteristics EquivalenceIdentical to predicate devices.
    Sizes Equivalence"Partly" identical to predicate devices. (No further detail provided on differences).
    No Modifications Made (compared to predicates affecting safety/effectiveness)"By the Dimeda Obstetrical Forceps have been no modification made (regarding to design, method of operation, mechanical performance)."
    No Device Failures Recorded in Europe (for models covered by 510(k))"All models that are covered by this 510(k) premarket notification have been on the market in Europe for many years with no device failures."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not applicable. No specific test set with a defined sample size for performance evaluation is mentioned in the document. The submission relies on comparative data and market history.
    • Data Provenance: The statement "All models that are covered by this 510(k) premarket notification have been on the market in Europe for many years with no device failures" refers to retrospective market data from Europe.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. No ground truth establishment by experts for a specific test set is described.

    4. Adjudication Method for the Test Set

    Not applicable. No test set requiring adjudication is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, an MRMC comparative effectiveness study was not done. This type of study is more typical for diagnostic imaging AI systems, not generally for surgical instruments demonstrating substantial equivalence.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. This device is a manual surgical instrument and does not involve an algorithm or AI.

    7. The Type of Ground Truth Used

    Not applicable in the usual sense of a diagnostic or algorithmic device. The "ground truth" for demonstrating safety and effectiveness relies on:

    • Predicate Device Performance: The established safety and efficacy of the legally marketed predicate devices (K013747 - Tekno-Medical Obstetrical Forceps and K951529 - V. MUELLER Obstetrical (OB) Forceps).
    • Market History: The lack of device failures over many years of use in Europe.

    8. The Sample Size for the Training Set

    Not applicable. This device is a manual surgical instrument and does not involve a training set for an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013747

    Validate with FDA (Live)

    Date Cleared
    2002-02-11

    (90 days)

    Product Code
    Regulation Number
    884.4400
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    |
    | Product classification: | 884.4400
    GERMANY

    Re: K013747

    Trade/Device Name: Techno-Medical Obstetrical Forceps Regulation Number: 21 CFR 884.4400

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    tekno-medical obstetrical forceps are used for helping delivering the baby. The forceps is consisting of two parts and you are introducing every part from the forceps(blades) seperately into the vagina.

    Device Description

    The Obstetrical forceps consist of two parts. Each of these two parts is introduced from the forceps (blades) seperately into the vagina.

    AI/ML Overview

    This document is a 510(k) premarket notification decision letter from the FDA regarding "Techno-Medical Obstetrical Forceps." It communicates that the device is substantially equivalent to legally marketed predicate devices.

    This type of document (a 510(k) clearance letter for a physical medical device like obstetrical forceps) typically does not contain the detailed information about acceptance criteria and a specific study proving device performance as would be found for a software-as-a-medical-device (SaMD) or AI-based device.

    For a physical device like obstetrical forceps, "acceptance criteria" generally refer to design specifications, material properties, sterilization methods, and manufacturing quality control, rather than performance metrics like sensitivity, specificity, or AUC as seen in diagnostic algorithms. The "study" would typically involve biocompatibility testing, mechanical testing (e.g., tensile strength, fatigue), and potentially cadaveric or animal studies for safety and function, rather than human subject clinical trials focused on diagnostic accuracy.

    Therefore, based solely on the provided text, I cannot extract the information required for your specific prompt. The document indicates that the device is "substantially equivalent" to predicate devices, meaning its safety and effectiveness are considered comparable to devices already on the market, but it does not detail the specific studies and acceptance criteria that led to this determination in the way you've outlined for an AI/software product.

    Here's a breakdown of why each specific point you requested cannot be answered from the provided text:

    1. A table of acceptance criteria and the reported device performance: Not present. The document only states the device is "substantially equivalent."
    2. Sample size used for the test set and the data provenance: Not applicable/not present. This is a physical device, and the focus is on equivalence, not diagnostic test performance.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable/not present. Ground truth as typically defined for AI/diagnostic algorithms is not relevant here.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable/not present.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical device.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable/not present.
    8. The sample size for the training set: Not applicable/not present. This is a physical device.
    9. How the ground truth for the training set was established: Not applicable/not present.

    To find the kind of information you are looking for (acceptance criteria and detailed study results), you would need to examine the actual 510(k) submission document (if publicly available) which would contain engineering specifications, test reports, and comparisons to predicate devices, but these would not be structured in the same way as studies for an AI/SaMD product.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1