Search Results
Found 128 results
510(k) Data Aggregation
(129 days)
Regulation |
|---|---|---|---|---|
| Primary | CO2 Pneumo Dissector | Cook Urological Inc | K972647 | HET, 884.1720
| Secondary #1 | Nezhat-Dorsey Hydrodissection | American Surgical Instruments Inc | K913944 | HET, 884.1720
Regulation |
|---|---|---|---|---|
| Primary | CO2 Pneumo Dissector | Cook Urological Inc | K972647 | HET, 884.1720
| Secondary #1 | Nezhat-Dorsey Hydrodissection | American Surgical Instruments Inc | K913944 | HET, 884.1720
The CO2 Pneumo-Dissector is used to dissect planes of soft tissue using pressure regulated, trigger-controlled pulses of medical grade CO2 gas. The device may be used in endoscopic, laparoscopic, and open procedures in which gentle blunt dissection of soft tissue planes is desired. Use the product only in accordance with the instructions provided.
The Pneumo Dissector Hook is a standard monopolar hook (based on design cleared under K140101) that can deliver CO2 flow on demand during general endoscopy and laparoscopic surgery. It is composed of two parts, a handle and an insert, which are linked using a nut-screw system. Also, the distal part of the instrument has the hook shape which is the monopolar electrode and the gas nozzle. It is a surgically invasive device intended for delivering pressurized CO2 gas to achieve separation of tissue layers prior to their dissection. The hook insert and tube are offered in lengths of 330mm and 200mm and diameters of 3.5mm and 5.0mm.
The handle has a connector for the CO2 as well as the banana plug or connector for electrosurgery. There is a push button to activate the CO2.
The nut-screw system is used to link the handle to the insert.
The insert consists of an insulated tube with a hook on the distal end.
The hook insert and tube are available in lengths of 330mm and 200mm and diameters of 3.5mm and 5.0mm.
N/A
Ask a specific question about this device
(219 days)
Gynecologic laparoscope and accessories
Device Classification: Class II
Regulation: 21 CFR 884.1720
It is used to connect with an optical endoscope during the endoscopic diagnosis, treatment, and observation. And to capture, process and transmit images in the human body cavity under the field of view observed by the endoscope to the monitor.
The 4K Endoscopic Camera System consists of a camera control unit and a 4K camera head.
The device is intended to be used to connect with an optical endoscope during the endoscopic diagnosis and/or treatment/surgery, and to capture, process and transmit images in the human body cavity under the field of view observed by the endoscope to the monitor.
There are 4 models of the device. The hardware configuration of the device of 4 models are same, and difference is different functions which opened up through software.
N/A
Ask a specific question about this device
(154 days)
K250939**
Trade/Device Name: SIRIUS Endoscope System (PR-SI-1230)
Regulation Number: 21 CFR 884.1720
Common Name** | Endoscope system |
| Classification Name | 876.1500 Endoscope and Accessories884.1720
Class II | Same |
| Product Code | HET, GCJ, FGB | HET, GCJ, FGB | Same |
| Regulation number | 884.1720
876.1500 | 884.1720876.1500 | Same |
| Regulation Name | Endoscope and Accessories | Endoscope
It is intended for use for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including female reproductive organs. Not suitable for use for observation and treatment of the heart and must not contact the heart or its vicinity.
The SIRIUS Endoscope System as a robotic endoscope consists of Video Processor, Laparoscope Handle, Laparoscope Head and Joystick. The SIRIUS Endoscope System is developed for minimal invasive surgery application.
The SIRIUS Endoscope System is a fully integrated compact 3D laparoscopic camera system with a flexible tip that can change its viewing direction. The Laparoscope Head is made of biocompatible materials. The articulated tip has three degrees of freedom enabling C and S shaped bending. The insertion section is 10 mm diameter and 342 mm working length. Stereo camera with 1080 high-definition resolution. It has 90 degrees field of view, 10-100 mm depth of view.
The SIRIUS Endoscope System is to be used only under the supervision of a trained surgeon and professional clinical staff with trained use of the device.
The device is intended for use in Hospital operating theatres only.
This 510(k) clearance letter focuses on establishing substantial equivalence to a predicate device, rather than proving the device meets new acceptance criteria through a specific study design typically seen for AI/ML devices or novel technologies. The "acceptance criteria" here are implicitly linked to the performance of the predicate device and relevant international standards.
Therefore, many of the specific details you requested regarding acceptance criteria and study design (like sample size for test sets, number of experts for ground truth, MRMC studies, standalone performance, training sets) are not present in this type of document because the submission is not presenting evidence of novel diagnostic or assistive AI performance. It's a submission for a new model of an endoscope system, demonstrating it performs as safely and effectively as a previously cleared version.
However, I can extract information related to product performance and testing where available, and explain why other details are absent based on the nature of this 510(k) submission.
Analysis of Acceptance Criteria and Device Performance for SIRIUS Endoscope System (K250939)
The SIRIUS Endoscope System (PR-SI-1230) received 510(k) clearance based on its substantial equivalence to a predicate device (K221642: SIRIUS Endoscope System). The "acceptance criteria" in this context are not defined as specific performance metrics for a novel AI algorithm, but rather as meeting the safety, effectiveness, and performance standards expected of an endoscope system, particularly in comparison to its predicate. The studies provided focus on verifying that the new model performs comparably to the predicate and adheres to relevant industry standards.
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a substantial equivalence submission for an updated hardware device (endoscope system) rather than a novel AI diagnostic, the acceptance criteria are primarily aligned with regulatory standards, predicate device performance, and international product standards. There are no specific AI performance metrics like sensitivity, specificity, or AUC mentioned.
| Category | Acceptance Criteria (Implied / Demonstrated by Compliance) | Reported Device Performance / Compliance |
|---|---|---|
| Biocompatibility | Patient contacting components meet ISO 10993 standards (e.g., cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogen). | "The evaluation included the following tests: Cytotoxicity, Sensitization, Irritation, Acute systemic toxicity, Pyrogen", confirming compliance. |
| Sterility | Sterility Assurance Level (SAL) of 10⁻⁶; EO and chlorohydrin residuals within limits per ISO 10993-7. | "Sterilization condition is validated per ISO 11135: 2014 overkill half-cycle approach. The sterility assurance level (SAL) is 10⁻⁶. The amount of ethylene oxide and chlorohydrin residual levels are within the limit and in compliance with ISO 10993-7." |
| Electrical Safety | Compliance with IEC 60601-1 and IEC 60601-2-18 standards for safety. | "The device complies with IEC 60601-1, IEC60601-2-18 standards for safety." |
| EMC | Compliance with IEC 60601-1-2 standard for EMC. | "The device complies with... the IEC 60601-1-2 standard for EMC." |
| Light Source Safety | Compliance with IEC 62471: 2006 (Risk Group 2). | "The lamp is certified as Risk group 2 and complies with IEC 62471: 2006." |
| Software V&V | Compliance with FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Basic Documentation Level). | "Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance... The software documentation level applied for this device was considered as Basic Documentation." |
| Performance Testing | Compliance with specific ISO 8600 series standards for endoscopes (e.g., field of view, direction of view, width of insertion, water resistance). | Bench testing performed verifies compliance with: ISO 8600-1, ISO 8600-3, ISO 8600-4, ISO 8600-7. |
| Usability | Acceptable residual risk regarding usability, benefits exceed possible risks of use-error, per ANSI/AAMI IEC 62366-1:2015+AMD1:2020. | "Formative and summative evaluations were conducted. Results and improvement action demonstrated that residual risk regarding usability is minimized and acceptable." |
| Technological Equivalence (Hardware Differences) | Performance equivalent for specified differences (e.g., working length, joystick design, locking mechanism). | Verified by electrical safety, EMC tests, and usability tests. Stated "no impact on the effectiveness and safety of the device." |
2. Sample Size Used for the Test Set and Data Provenance
This document does not describe a clinical "test set" in the context of an AI/ML or diagnostic performance study. The "tests" mentioned are primarily bench tests, engineering validations, and usability evaluations, not clinical studies involving patient data or images. Therefore, details like data provenance or a specific test set sample size for diagnostic performance are not provided. The comparisons are against the predicate device's design and adherence to recognized performance standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This submission is about the physical endoscope system, not an AI diagnostic algorithm requiring expert ground truth for image interpretation or diagnosis. The "ground truth" for the performance section refers to engineering specifications and compliance with international standards.
4. Adjudication Method for the Test Set
Not applicable, as there's no diagnostic test set requiring human adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC study was not conducted or reported. This type of study is typically done for AI algorithms that assist or augment human reading of medical images to demonstrate improvement in reader performance. The SIRIUS Endoscope System is a hardware device for visualization and surgery, not an AI-powered diagnostic tool in that sense.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. There is no standalone algorithm being evaluated for performance independent of human operation for diagnostic purposes. The "software" referred to in the document relates to the device's operational software, not an AI for image analysis.
7. The Type of Ground Truth Used
The "ground truth" for this submission is based on:
- International Standards: e.g., ISO 10993 series for biocompatibility, ISO 11135 for sterility, IEC 60601 series for electrical safety/EMC, IEC 62471 for light source safety, ISO 8600 series for endoscope performance, and ANSI/AAMI IEC 62366-1 for usability.
- Engineering Specifications: The design and function of the device itself are validated against its own specifications and those of the predicate.
- Predicate Device Performance: The primary "ground truth" for substantial equivalence is that the new device performs "as well as the legally marketed predicate device."
8. The Sample Size for the Training Set
Not applicable. This document does not describe an AI/ML algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for an AI/ML algorithm mentioned.
Summary of Study Type:
The submission focuses on design verification and validation testing to demonstrate that the new SIRIUS Endoscope System (PR-SI-1230) is safe, effective, and performs as well as its predicate. This includes ensuring compliance with a wide array of international standards (ISO, IEC, ANSI/AAMI) for medical device manufacturing, biocompatibility, electrical safety, sterility, and basic endoscope performance. The differences from the predicate (e.g., working length, joystick design, locking mechanism) were specifically verified through engineering and usability tests to ensure they did not negatively impact safety or effectiveness.
Ask a specific question about this device
(134 days)
Feldkirch, 6800
AUSTRIA
Re: K243821
Trade/Device Name: i-Cut
Regulation Number: 21 CFR 884.1720
Morcellator
Regulation Name: Gynecologic Laparoscope And Accessories
Regulation Number: 21 CFR § 884.1720
The i-Cut system is indicated for morcellating and extracting tissue in laparoscopic, gynecologic surgical procedures.
The i-Cut is a single use laparoscopic power morcellator. It is provided sterile.
The i-Cut consists of a rotating cutting tube with a trocar that provides protection against the cutting blade when locked in the "CLOSED" position. The device comes with an obturator for placement into the patient's body and a locking clip that keeps the trocar in the "closed" position during insertion. The activation button must be pressed to cut tissue.
The lumen of the device is designed for use with a standard grasper or Tenaculum forceps with a diameter between 10 to 14 mm. The i-Cut is designed to be used with surgical instruments of diameters between 10 to 14 mm. The lumen is fitted with a silicone valve to prevent gas loss during use of the device.
The i-Cut is electrically operated by a DC Motor which is powered by a 24V AC/DC mains adapter. The non-sterile i-Cut Power Supply with power cord is supplied separately to power the device.
This 510(k) clearance letter pertains to the i-Cut, a single-use laparoscopic power morcellator. The provided text details regulatory information, device description, indications for use, and non-clinical performance testing. However, it does not include information about clinical studies with human participants, acceptance criteria tables, sample sizes for test sets (in the context of clinical performance), expert ground truth establishment, or multi-reader multi-case (MRMC) comparative effectiveness studies.
The document primarily focuses on non-clinical bench testing to demonstrate substantial equivalence to a predicate device (LiNA eXcise, K101458). Therefore, I cannot address most of the requested points related to clinical study acceptance criteria and performance as this information is not present in the provided text.
Here's an analysis of what can be extracted from the provided text, and where gaps exist:
Preamble:
The provided 510(k) clearance letter for the i-Cut device focuses on demonstrating substantial equivalence through non-clinical performance testing (bench testing) rather than clinical trials with human participants. Therefore, many of the requested details pertaining to clinical study design, human reader performance, and expert consensus for ground truth are not present in this document. The "acceptance criteria" and "device performance" discussed here apply to the bench testing conducted, not clinical outcomes in human subjects.
1. A table of acceptance criteria and the reported device performance
Based on the provided non-clinical performance testing section, we can infer some "acceptance criteria" as implied by the successful completion of each test. The "reported device performance" is generally stated as "passed" or "met requirements." No numerical data for specific criteria are provided beyond the stated specifications (e.g., cutting rate, speed).
| Acceptance Criteria (Inferred from Test Purpose) | Reported Device Performance |
|---|---|
| Sterilization Validation (ANSI/AAMI/ISO 11135:2014) | Passed sterilization validation |
| Shelf-life Validation (ASTM F1980-21:2023, etc.) | Supports 5-year shelf life (package integrity and device performance) |
| Simulated Shipping (ASTM D4332-22:2022, etc.) | Passed simulated shipping tests |
| Biocompatibility (ISO 10993-5, -10, -11, -23) | Passed biocompatibility tests (Cytotoxicity, Sensitization, Intracutaneous irritation, Acute systemic toxicity) |
| Electrical Safety (IEC 60601-1:2020) | Passed electrical safety |
| Electromagnetic Compatibility (EMC) (IEC 60601-1-2:2020) | Passed EMC |
| Load Test (Withstand expected load without failure) | Passed load test |
| Gas Tightness Test (Maintain pneumoperitoneum) | Passed gas tightness test |
| Obturator Abrasion Test (No visible wear marks) | Passed obturator abrasion test (no abrasion occurs) |
| Liquid Tightness Test (Meet requirements after load test) | Passed liquid tightness test |
| Grip of the Housing Surface Test (Surgeon's ability to properly grip) | Passed grip of the housing surface test |
| Surface Reflection Test (Cutting blade not cause reflections impeding visibility) | Passed surface reflection test |
| Trocar Sleeve – Pull-off Force Test (Connection suitable for application) | Passed pull-off force test |
| Functional Test (Morcellate and extract tissue in model specimen) | Passed functional test |
| Insertion Test (Force needed to expose cutting blade in "Closed" position) | Passed insertion test |
| Usability testing (Compatibility with containment system, integrity assessment) | Passed usability testing |
| Cutting Rate (Stated Specification) | >21.5 g/min |
| Speed (Stated Specification) | 1100 rpm |
2. Sample sizes used for the test set and the data provenance
- Sample Size: The document does not specify the sample sizes (e.g., number of devices, number of tests) for each of the non-clinical performance tests. It only states that the tests were performed and passed.
- Data Provenance: The data provenance is from non-clinical bench testing conducted by the manufacturer, A.M.I. Agency for Medical Innovations GmbH. The country of origin of the manufacturer is Austria. This data is inherently prospective as it involves new testing for the device's clearance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This was a non-clinical bench testing study. "Ground truth" in this context pertains to engineering specifications and performance metrics rather than clinical diagnoses established by medical experts. Usability testing involved "surgeons of different experience with laparoscopic morcellators," but it was to assess device compatibility and integrity, not to establish a clinical ground truth for a diagnostic AI.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This was a non-clinical bench testing study. Adjudication methods are typically used in clinical studies where multiple human readers interpret data to establish a consensus ground truth.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. No MRMC comparative effectiveness study was performed or described. The clearance is based on substantial equivalence demonstrated through non-clinical bench testing. The device is a physical surgical tool, not an AI.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a physical surgical tool and does not employ a standalone algorithm for diagnostic or interpretative purposes.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For non-clinical bench testing, the "ground truth" is defined by engineering specifications, regulatory standards (e.g., ISO, ASTM, IEC), and predefined pass/fail criteria for device performance (e.g., measuring force, checking for wear, verifying electrical safety compliance). There is no "expert consensus" or "pathology" in the medical sense for these tests.
8. The sample size for the training set
Not applicable. This device is a mechanical/electrical surgical tool, not an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for this device.
Ask a specific question about this device
(133 days)
|
| Classification Name: | 21 CFR 876.1500 Endoscope and accessories21 CFR 884.1720
The KARL STORZ suction and irrigation system (consisting of handle and suction/irrigation tubes, cannulae or needles) is intended for use by qualified surgeons to provide suction and/or irrigation function to help flush or remove blood, fluids, and tissue debris from the operation site during general, gynecologic, urologic, and thoracic laparoscopic surgical procedures.
The devices in the KARL STORZ Suction and Irrigation System are manually operated, reusable surgical devices consisting of handle, suction/irrigation tubes, cannulae or needles. The handle is intended to be used in combination with suction/irrigation cannulae and tubing and controls the suction and irrigation. Enabled by the handle with stop cock valve, the regulation of the irrigated quantity of fluids via the tubes, or rather the quantity of aspirated fluid (e.g., blood, fluids, and tissue debris) into or out of the patient is performed. Cannula and needle: Enabled by the pointed distal end, both puncture and injection/extraction cannulae and needles, perform puncturing of target organs or tissues and injecting fluids into operating site i.e. extracting blood or fluids from them. The prerequisite for suction is either a vacuum powered suction wall apparatus or a pump, which serve as impulse for the suction. The proximal end of the tube set is connected to the vacuum or pump. Either a pump is required as impulse for irrigation or the natural force of gravity, i.e., by positioning the flush bag at the appropriate height.
The provided document is an FDA 510(k) Premarket Notification for medical devices, specifically KARL STORZ suction and irrigation systems. This type of submission focuses on demonstrating substantial equivalence to legally marketed predicate devices, primarily through non-clinical performance data and does not typically include studies involving artificial intelligence (AI) or machine learning (ML) components, nor does it present "acceptance criteria" and "device performance" in the context of an AI/ML model's diagnostic or predictive capabilities.
Therefore, the requested information regarding AI/ML-related acceptance criteria and study details (such as sample sizes for test and training sets, expert adjudication methods, MRMC studies, standalone performance, ground truth establishment, etc.) cannot be extracted from this document, as it is not relevant to the type of device and submission described.
The document focuses on:
- Device Description: Manually operated, reusable surgical devices (handle, suction/irrigation tubes, cannulae, needles).
- Intended Use: Suction and/or irrigation to flush or remove blood, fluids, and tissue debris during laparoscopic surgical procedures.
- Basis for Substantial Equivalence: Comparison to predicate KARL STORZ devices (K945059 and K951190) based on technological characteristics and non-clinical performance data.
- Non-Clinical Performance Data Provided:
- Biocompatibility testing: Compliance with ISO 10993.
- Reprocessing Validation: Compliance with standards like ANSI/AAMI/ISO 14937, 11138-3, 11607-1, 11737-1, 17664, 17665-1 regarding cleaning and sterilization.
- Bench Top Performance Testing: Leak Test, Flow Incoming Test, Flow Outgoing Test.
The document explicitly states: "Clinical testing was not required to demonstrate the substantial equivalence to the predicate device. Non-clinical bench testing was sufficient to assess safety and effectiveness and to establish the substantial equivalence of the modifications."
In summary, this document does not contain information about an AI/ML device or its associated performance studies.
Ask a specific question about this device
(266 days)
|
| Regulation Numbers | 21 CFR 884.1720
Camera System: The Camera System is indicated for use in general laparoscopy, ear endoscopy, ear endoscopy, sinuscopy, and plastic surgery whenever a laparoscope/ endoscope/ arthroscope is indicated for use. A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Pleural biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement. The users of the Camera System are general surgeons, gynecologists, cardiac surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists.
NIR FI Light Source: The NIR FI Light Source and NIR FI Light Guide are indicated for use to provide real-time endoscopic visible and nearinfrared fluorescence imaging. The NIR FI Light Source and NIR FI Light Guide enable surgeons to perform minimally invasive surgery using standard endoscope visual light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct), using near-infrared imaging.
Fluorescence imaging of biliary ducts with the NIR FI Light Source and NIR FI Light Guide is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.
The individual components of the subject device, SCHOELL Y's NIR FI System, form a system to provide real-time endoscopic visible imaging (wight light imaging, WLI) and near-infrared (NIR) illumination and imaging (fluorescence imaging, FI) using indocyanine green (ICG):
- Camera System suitable for processing and recordings visible light images as well as NIR images. The Camera System consists of a Camera Control Unit (CCU) and a Camera Head for connection to a fiberoptic scope;
- . Light Source and Light Guide for use with a fiberoptic scope for emitting light within the visible spectrum as well as in the NIR spectrum to cause fluorescence;
- . Fiberoptic Laparoscope suitable for visible light and NIR light illumination and imaging;
The imaging agent (ICG) is not provided by SCHOELLY as part of the subject system.
The submitted information does not contain a study that proves the device meets the acceptance criteria in the format requested. The document primarily focuses on establishing substantial equivalence to a predicate device for regulatory clearance. It describes general performance testing conducted, but not in the context of specific acceptance criteria and detailed study results as typically found in clinical trials or dedicated performance studies for AI/ML devices.
However, based on the provided text, I can extract information related to the device and the types of testing performed to support its regulatory clearance.
Here's an attempt to structure the available information, noting where specific details (like acceptance criteria, sample sizes, ground truth establishment, or expert involvement for performance scores) are not explicitly present in the provided document:
Device Name: Near-Infrared (NIR) Fluorescence Imaging (FI) System: Camera System (Camera Control Unit, Camera Head to be coupled to a fiberoptic scope), NIR FI Light Source
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state quantitative acceptance criteria for device performance (e.g., specific sensitivity, specificity, accuracy targets). Instead, it relies on demonstrating compliance with recognized standards and substantial equivalence to a predicate device.
| Performance Aspect | Acceptance Criteria (Implicit from regulatory context) | Reported Device Performance |
|---|---|---|
| Reprocessing Validation | Compliance with FDA 2015 guidance, AAMI TIR12:2010, AAMI TIR30:2011(R)2016, ANSI/AAMI/ISO 17665-1:2006 (R)2013, ISO 17664:2017. | "These tests demonstrated that the device successfully passed cleaning, drying and sterilization validations according to the instructions in the user manual." |
| Software Documentation | Compliance with FDA's 2005 Guidance for Software, IEC 62304:2006/A1:2016 (MODERATE Level of Concern). | "Software documentation for a MODERATE Level of Concern device is provided in support of the proposed device... The software lifecycle, including software documentation and validation, is managed in accordance with IEC 62304:2006/A1:2016..." (Implies compliance). |
| Electrical Safety Testing | Compliance with IEC 60601-1:2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012 and IEC 60601-2-18:2009. | "The NIR FI System was assessed for conformity with, and was found to comply with, the relevant requirements of IEC 60601-1:2005... and IEC 60601-2-18:2009..." |
| Electromagnetic Compatibility | Compliance with IEC 60601-1-2:2014. | "The NIR FI System was assessed for conformity with, and was found to comply with, the relevant requirements of IEC 60601-1-2:2014..." |
| Non-Clinical Performance | Substantial equivalence to predicate device (Stryker AIM System) and meeting design input requirements for endoscopic white light and NIR fluorescence imaging. | "Non-Clinical performance test data demonstrate that the proposed NIR FI System performs substantially equivalent to the Stryker predicate AIM System and that the design output meets the design input requirements for endoscopic white light and near-infrared fluorescence imaging." (No specific quantitative metrics for imaging performance are disclosed in this summary). |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify a "test set" in the context of evaluating a dataset for AI performance. The performance data mentioned refers to engineering and quality system validation tests. No information is provided regarding the origin (country, retrospective/prospective) of specific data sets used for validating imaging performance beyond general statements about "non-clinical performance test data."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable based on the provided text. The document does not describe a study involving expert readers establishing ground truth for a test set to assess AI performance. The focus is on the device's technical specifications and safety/effectiveness in a comparative context to a predicate device.
4. Adjudication Method for the Test Set
Not applicable. As no human expert evaluation of a test set for AI performance is described, no adjudication method is mentioned.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size
No MRMC study is mentioned. The document primarily focuses on demonstrating substantial equivalence to a predicate device through technical comparisons and compliance with standards, rather than a study comparing human readers with and without AI assistance.
6. If a Standalone (Algorithm Only) Performance Study Was Done
The device described is a medical imaging system (hardware and integrated functionality for visible and NIR fluorescence imaging), not a standalone AI algorithm. Therefore, a standalone algorithm performance study as typically understood for AI/ML software is not applicable here. The "Non-Clinical Performance Testing" refers to the system as a whole.
7. The Type of Ground Truth Used
For the "Non-Clinical Performance Testing," the "ground truth" implicitly refers to the expected performance characteristics based on an existing predicate device and the design input requirements for endoscopic white light and near-infrared fluorescence imaging. The document does not specify an external "ground truth" like pathology, expert consensus on images, or outcomes data.
8. The Sample Size for the Training Set
Not applicable. The document does not describe the development or training of an AI algorithm with a training set. The device is an imaging system, not a machine learning model.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As there is no mention of a training set or AI algorithm training, the establishment of ground truth for such a set is not discussed.
Summary of Non-Inclusion:
The provided document is a 510(k) summary for a medical device (an imaging system) seeking clearance based on substantial equivalence. It is not a report on a clinical or performance study evaluating an AI/ML algorithm against specific performance metrics with independent test sets and expert ground truth. Therefore, many of the requested details, particularly those related to AI/ML study design (sample sizes for test/training sets, expert qualifications, ground truth establishment, MRMC studies), are not present in the given text. The "Performance Data" section details compliance with recognized safety, software, and reprocessing standards, and general non-clinical performance demonstrations for substantial equivalence.
Ask a specific question about this device
(189 days)
Kong 0000 China
Re: K221642
Trade/Device Name: SIRIUS Endoscope System Regulation Number: 21 CFR 884.1720
|
| Classification Name: | 876.1500 Endoscope and Accessories884.1720
|
| Regulation number | 876.1500884.1720
| 876.1500884.1720
It is intended for use for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including female reproductive organs. Not suitable for use for observation and treatment of the heart and must not contact the heart or its vicinity.
The SIRIUS Endoscope System as a robotic endoscope consists of Video Processor, Laparoscope Handle, Laparoscope Head and Joystick. The SIRIUS Endoscope System is developed for minimal invasive surgery application. The SIRIUS Endoscope System is a fully integrated compact 3D laparoscopic camera system with a flexible tip that can change its viewing direction. The Laparoscope Head is made of biocompatible materials. The articulated tip has three degrees of freedom enabling C and S shaped bending. The insertion is 10 mm diameter and 340 mm working length. Stereo camera with 1080 high-definition resolution. It has 120 degrees field of view, 10-100mm depth of view and bright light with 300 lumen.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Criterion (if available) | Reported Device Performance |
|---|---|---|
| Biocompatibility | Meets ISO 10993 standards | Complies with ISO 10993-1, -5, -10, -11 requirements (Cytotoxicity, Sensitization, Irritation, Acute systemic toxicity, Pyrogen testing performed). |
| Sterility | Sterility Assurance Level 10^-6 | Validated via Ethylene Oxide (EO) sterilization per ISO 11135:2014 overkill half-cycle approach. EO and chlorohydrin residuals within ISO 10993-7 limits. |
| Electrical Safety and EMC | Complies with IEC 60601 standards | Complies with IEC 60601-1, IEC 60601-2-18 (safety) and IEC 60601-1-2 (EMC). |
| Light Source Safety | Complies with IEC 62471 | Lamp certified as Risk group 2 and complies with IEC 62471:2006. |
| Software Verification and Validation | Adherence to FDA Guidance | Software Verification and Validation Testing conducted as per FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Software considered "Moderate" level of concern. |
| Usability Evaluation | Adherence to ANSI AAMI IEC 62366-1 | Conducted per ANSI AAMI IEC 62366-1:2015+AMD1:2020. Usability testing and improvement actions minimized residual risk. |
| Performance Testing (Endoscope Standards) | Meets ISO 8600-1 | Complies with ISO 8600-1:2015. |
| Meets ISO 8600-3 | Complies with ISO 8600-3:2019 (determination of field of view and direction of view). | |
| Meets ISO 8600-4 | Complies with ISO 8600-4:2014 (determination of maximum width of insertion portion). | |
| Meets ISO 8600-7 | Complies with ISO 8600-7:2012 (basic requirements for water-resistant medical endoscopes). | |
| Tip Articulation | Meets ISO 8600-1:2015 requirements | Angle of deflection for ±90° up-down and ±45° left-right joints meet the requirement of ISO 8600-1:2015. |
| Shaft Material Compatibility | Biocompatible | Different materials from predicate, but both are biocompatible, and biocompatibility tests demonstrated safety. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document primarily describes bench testing and compliance to international standards. It does not specify "test sets" in the context of patient data or clinical trials with specific sample sizes. The data provenance is not mentioned beyond the manufacturer being in Hong Kong. The testing appears to be primarily laboratory-based (bench testing) rather than clinical studies using patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided in the document. The document focuses on technical and regulatory compliance testing rather than expert-derived ground truth from clinical cases.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided as the studies described are not clinical studies requiring adjudication of diagnostic outcomes.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
There is no mention of an MRMC comparative effectiveness study, AI assistance, or human reader improvement. The device described is an endoscope system, not an AI-powered diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
There is no mention of an algorithm-only standalone performance study. The device is hardware (an endoscope system) with associated software, but not an AI algorithm performing diagnostic tasks independently.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the performance testing, the "ground truth" is adherence to international standards (e.g., ISO 8600 series) and published specifications. For biocompatibility, sterility, electrical safety, etc., the ground truth is defined by the requirements of the respective international standards (e.g., ISO 10993, ISO 11135, IEC 60601, IEC 62471).
8. The sample size for the training set
This information is not applicable or not provided. The device is an endoscope system, and the testing described is not related to machine learning model training.
9. How the ground truth for the training set was established
This information is not applicable or not provided for the same reasons as above.
Ask a specific question about this device
(26 days)
br>21 CFR §876.1500, Product Code(s) GCJ, GCQ,Gynecologic Laparoscope and Accessories21 CFR §884.1720
Obstetrical and Gynecological DiagnosticDevices - Gynecologic Laparoscope andAccessories(21 CFR §884.1720
The SaberScope5 Laparoscope is intended to be used in diagnostic procedures for endoscopy and endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs.
The Articulating Xenoscope™ System is being renamed the SaberScope5 Laparoscope. The subject device contains two separate functioning components. First is the single-use, sterile SaberScope5 Laparoscope Device, which includes a 0° camera on 5 mm rigid shaft with a ± 90° articulating tip, 10 - 36 cm long shaft, and high-definition video image. For certain procedures the shorter 10 cm laparoscope is preferred. Likewise, for other procedures, the longer 36 cm laparoscope is preferred. Except for the length difference, the scientific principles, materials of construction and design are otherwise identical. The second is the XenoBox™, which converts the digital signal from the camera to HDMI signal for display onto the HD video screen for the surgeon to view.
Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Xenocor® SaberScope5 Laparoscope, addressing your specific points:
1. A table of acceptance criteria and the reported device performance
The document provides a comparative table (Table 2) between the predicate device (Articulating Xenoscope™ Laparoscope, K193315) and the subject device (SaberScope5 Laparoscope). The acceptance criteria implicitly are that the subject device's performance attributes are "Same" as the predicate device or demonstrate equivalent performance, with some notable differences.
| Attribute | Predicate Device Performance (K193315) | SaberScope5 Laparoscope Performance | Acceptance Criteria Met? | Notes |
|---|---|---|---|---|
| Indications for Use | Diagnostic and therapeutic procedures for endoscopy and endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. | Same | Yes | Maintained same Indications for Use. |
| Classification Info. | GCJ, GCQ, HET; Class II | Same | Yes | Maintained same classification. |
| Single Use | Yes | Same | Yes | |
| Sterile | Provided Sterile (EO gas) | Same | Yes | |
| Prescription | Yes | Same | Yes | |
| Anatomical Access | Thoracic and abdominal body cavities, hollow organs, and canals, including female reproductive organs. | Same | Yes | |
| Fundamental Scientific Technology | Imaging sensor that sends digital video, converted to HDMI by Xenobox. | Same as predicate with the additional manual focus capability | Yes (with enhancement) | The addition of manual focus is highlighted as an improvement, not a deviation from acceptance. |
| Part No. | XSA-0-0510, XSA-0-0536 | Same | Yes | These are the model numbers for existing lengths. |
| Design | 0° camera angle, 5 mm rigid shaft, articulating tip with fixed focus. | Same, but with manual focus buttons on handle | Yes (with enhancement) | The addition of manual focus buttons is an enhancement. |
| Shaft Diameter (OD) | 5 mm | Same | Yes | |
| Shaft Tip | Articulating Tip (± 90°) | Same | Yes | |
| Shaft Lengths | 10 cm, 36 cm | Same | Yes | |
| Shaft Material | Carbon Fiber, covered with heat shrink sheathing | Same | Yes | |
| Field of View | 65°-75° (Nominal 69º) | Same | Yes | |
| HD Resolution | 1080p | Same | Yes | |
| Camera Focus | Fixed Focus | 3-10cm | Yes (improved) | This is a key difference. The SaberScope5 moves from "Fixed Focus" to a "3-10cm" focus range, which implies user-adjustable focus. The associated design change is the addition of focus control buttons. This is presented as an improvement that meets acceptance. |
| Focus Control Buttons | N/A | 3 buttons: forward (manual mode), backward (manual mode), manual/auto toggle | N/A | This is a new feature introduced in the SaberScope5. It's not a "met" or "not met" criteria against the predicate, but a feature leading to the "3-10cm" focus range. |
| Frame Rate | 30 fps | Same | Yes | |
| Exposure/Gain Control | Automatic (no user adjustments) | Same | Yes | |
| Latency (<100ms) | No buffering more than 3 frames | Same | Yes | |
| Light Source | 6 LED (ring) | Same | Yes | |
| White Balance | Fixed | Same | Yes | |
| LED Color Brightness Range | 2500k to 7000k | Same | Yes | |
| Handle | Handle with articulation mechanism | Handle with articulation mechanism and focus buttons | Yes (with enhancement) | The addition of focus buttons is an enhancement. |
| Packaging | Double Pouched Tyvek/ Mylar Pouch, and backer card | Same | Yes | Confirmed explicitly in the "Functional/Safety Testing" section that changes did not affect packaging. |
| Sterilization | EO Sterile (SAL 10-6) | Same | Yes | |
| Shelf Life | 1 years | Same | Yes | |
| Biocompatibility | Patient contacting components meet ISO 10993 standard | Same | Yes | Confirmed explicitly in "Functional/Safety Testing" that no new materials were introduced. |
| Applied Part Class | Type CF | Same | Yes |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document mentions "Design Verification – Performance bench testing was conducted".
- Sample Size: Not specified. The document only mentions "performance bench testing" was conducted, implied to be on the SaberScope5 Laparoscope.
- Data Provenance: Not specified, but generally, bench testing results for FDA submissions are generated by the manufacturer or a contracted lab. There's no indication of country of origin of data or whether it was retrospective or prospective in the clinical sense, as it was bench testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This type of information is not applicable to this submission. The submission pertains to a medical device (laparoscope) that aids human physicians, not an AI or diagnostic software that requires expert-established ground truth for its own performance evaluation in a clinical context. The performance verification here relates to engineering and functional parameters rather than diagnostic accuracy.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. As noted above, this is a hardware device undergoing performance bench testing, not an AI or diagnostic algorithm requiring clinical adjudication of results.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was done. This is not an AI-assisted diagnostic or imaging interpretation device. It is a surgical viewing device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is not an algorithm, but a hardware device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The document implicitly uses engineering specifications and performance standards as "ground truth" derived from the predicate device and relevant industry standards. For example, the "Field of View" (65°-75°) or "HD Resolution" (1080p) are objective, measurable performance characteristics. The term "ground truth" in the context of diagnostic accuracy is not directly applicable here.
8. The sample size for the training set
Not applicable. This is a hardware device; there is no "training set" in the machine learning sense.
9. How the ground truth for the training set was established
Not applicable. There is no training set.
Ask a specific question about this device
(267 days)
Trade/Device Name: Video Endoscopy System & 3D Video Endoscopy System Regulation Number: 21 CFR§ 884.1720
LPS31030 Classification Name: Gynecologic laparoscope and accessories Classification Regulation: 21 CFR 884.1720
Predicate Device
510(k) Number: K150525 Trade/Device Name: SPIES 3D System Regulation Number: 21 CFR 884.1720
| 21 CFR 884.1720
| 21 CFR 884.1720
The Video Endoscopy System & 3D Video Endoscopy System are intended to be used to provide illumination and visualization of surgical field in a wide variety of diagnostic abdominal and thoracic minimally invasive procedures, including female reproductive organs (gynecology) and urological anatomy.
The proposed Video Endoscopy Systems include the Video Endoscopy System and 3D Video Endoscopy System, Video Endoscopy System supports 2D imaqe output, 3D Video Endoscopy System supports 2D/3D image output.
Video Endoscopy System is composed of Video Endoscope (LPS21000/LPS21030) and Video Endoscopy processor (EVS100).
The Video Endoscopy Processor is a video processor, which receives the electrical signals from the Video Endoscope and process it and output the final image to the monitor. Two models 2D Video Endoscope (LPS21000/LPS21030) are available, one image sensor is located at the distal of the endoscope (LPS21000/LPS21030), they are used in conjunction with Video Endoscopy Processor (EVS100) to output 2D images.
3D Video Endoscopy System is composed of Video Endoscope (LPS31000/LPS31030) and Video Endoscopy Processor (EVS200).
The Video Endoscopy Processor is a video processor, which receives the electrical signals from the Video Endoscope and process it and output the final image to the monitor. Two models 3D Video Endoscope (LPS31000/LPS31030) are available, two image sensors are located at the distal of the endoscope (LPS31000/LPS31030), they are used in conjunction with Video Endoscopy Processor (EVS200) to output 2D/3D images.
Video Endoscopy processor is non-sterile device. The Video Endoscope is terminallysterilized device. The Video Endoscope must be sterilized by users before being used in surgery.
This submission describes the Surgnova Healthcare Technologies (Zhejiang) Co., Ltd. Video Endoscopy System & 3D Video Endoscopy System (K210116).
Here's an analysis of the acceptance criteria and study information provided:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document does not explicitly list quantitative acceptance criteria for optical performance. Instead, it states that "The device met all acceptance criteria for optical performance testing and was shown to have equivalent image quality to the predicate." This implies a comparative standard rather than absolute numerical targets.
The reported device performance is qualitative for equivalence.
| Performance Characteristic | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Resolution | Equivalent to Predicate (1920x1080 HD) | 1920x1080 HD (Same as predicate) |
| Brightness | Met acceptance criteria (not specified quantitatively) | Met acceptance criteria (shown to have equivalent image quality to predicate) |
| White Balance | Met acceptance criteria (not specified quantitatively) | Met acceptance criteria (shown to have equivalent image quality to predicate) |
| 3D-2D Mode | Met acceptance criteria (not specified quantitatively) | Met acceptance criteria (shown to have equivalent image quality to predicate) |
| Color Performance | Met acceptance criteria (not specified quantitatively) | Met acceptance criteria (shown to have equivalent image quality to predicate) |
| Field of View | Equivalent to Predicate (90°) | 90° (Same as predicate) |
| Geometric Distortion | Met acceptance criteria (not specified quantitatively) | Met acceptance criteria (shown to have equivalent image quality to predicate) |
| Signal-to-Noise Ratio (SNR) | Met acceptance criteria (not specified quantitatively) | Met acceptance criteria (shown to have equivalent image quality to predicate) |
| Dynamic Range | Met acceptance criteria (not specified quantitatively) | Met acceptance criteria (shown to have equivalent image quality to predicate) |
| Image Intensity Uniformity (IIU) | Met acceptance criteria (not specified quantitatively) | Met acceptance criteria (shown to have equivalent image quality to predicate) |
| Photobiological Safety | Compliance with IEC 62471:2006 | In compliance with IEC 62471:2006 |
| Biocompatibility | Not toxic, irritating, or sensitizing (ISO 10993 standards) | Not toxic, irritating or sensitizing |
| Software Validation | Compliance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices | In compliance with FDA Guidance |
| Sterilization Validation | Compliance with ISO 17665-1 | According to ISO 17665-1 |
| Package Verification | Compliance with ISTA 2A-11 & ASTM D 4169-16 | According to ISTA 2A-11 & ASTM D 4169-16 |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not provide details on sample sizes for the "test set" in terms of images or data points for the individual optical performance tests. The testing described is non-clinical bench testing rather than a study involving clinical data from patients. Therefore, terms like "country of origin of the data," "retrospective or prospective," or "test set" in the context of patient data do not apply here. The testing was performed on the device prototypes or production units.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The performance evaluation was based on bench testing against established standards and comparison to a predicate device's specifications, not on human expert review of clinical images to establish ground truth such as disease presence/absence.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. There was no adjudication method used, as the testing involved objective measurements and comparisons against technical specifications and standards, not subjective expert reviews requiring consensus.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was done. This submission is for a video endoscopy system as a medical device for visualization, not an AI-powered diagnostic or assistive tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, the primary evaluation was standalone performance simulation/bench testing of the device's optical and functional characteristics. The device itself (the video endoscopy system) is essentially the "algorithm only" in this context, as its performance is independently measured without human operators for the technical specifications listed.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the non-clinical testing was based on:
- Established engineering and performance standards: e.g., IEC 60601 series, ISO 10993 series, IEC 62471.
- Predicate device specifications: The comparison table explicitly states characteristics like resolution, field angle, and depth of field that are identical or comparable to the predicate.
- Objective measurements: The various optical performance tests (Resolution, Brightness, White Balance, Color Performance, etc.) would have involved objective measurement techniques with specific targets or ranges derived from industry standards or performance requirements for such devices.
8. The sample size for the training set
Not applicable. This device is a video endoscopy system, not an AI or machine learning algorithm that requires a training set of data.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for this type of medical device submission.
Ask a specific question about this device
(442 days)
Trade/Device Name: ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-5 Regulation Number: 21 CFR§ 884.1720
|
| ■ Regulation Number | 21 CFR 884.1720
This instrument is intended to be used with a video system center, light source, documentation equipment, monitor. hand instruments, electrosurgical unit, and other ancillary equipment for endoscopic surgery within the thoracic and abdominal cavities including female reproductive organs.
The LTF-S190-5 is a video endoscope consisting of three parts: the control section, the insertion section, and the connector section. The device uses an objective lens and light guide lens for capturing images/videos from the distal end of the endoscope.
Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-5:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion | Device Performance / Result |
|---|---|
| Reprocessing | All reprocessing methods (Sterilizer/Sterilization process) were validated following FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" (March 17, 2015). The acceptance criteria were met. |
| Biocompatibility | Tested in accordance with FDA's Guidance for Industry and FDA Staff, Use of International Standard ISO 10993-1, and specific sub-parts: |
| - Cytotoxicity (ISO 10993-5:2009) | |
| - Intracutaneous (ISO 10993-10:2010) | |
| - Sensitization (ISO 10993-10:2010) | |
| - Acute Systemic Toxicity (ISO 10993-11:2006) | |
| - Rabbit Pyrogen (USP <151>, USP <37>) | |
| The device was shown to be non-cytotoxic, non-irritating, non-sensitizing, not systemically toxic, and non-pyrogenetic. | |
| Software V&V | Software verification and validation testing conducted following FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices". The device met all requirements. |
| Electrical Safety | Tested in accordance with ANSI/AAMI ES 60601-1:2005/(R)2012 and A1:2012, and IEC 60601-2-18:2009. The system complies with these standards. |
| EMC | Tested in accordance with IEC 60601-1-2:2014. The system complies with this standard. |
| Bench Performance | Bench testing included: |
| - Thermal safety test | |
| - Composite durability test | |
| - Color (imaging) performance | |
| - Photobiological Safety | |
| Device performance was assessed against design requirements, including process verification, design verification, and design validation. The device met all acceptance criteria. |
2. Sample Size for Test Set and Data Provenance
The provided text describes various engineering and performance tests (reprocessing, biocompatibility, software V&V, electrical safety, EMC, and bench performance). These types of tests typically do not involve "test sets" in the clinical sense with human data. Instead, they involve testing the physical device, its materials, and its software against established technical and safety standards.
- No "sample size" for a clinical test set is mentioned.
- No "data provenance" (e.g., country of origin, retrospective/prospective) for clinical data is mentioned, as clinical data was not part of this submission's performance data section. The submission focuses on substantial equivalence based on technical characteristics and non-clinical performance data.
3. Number of Experts and Qualifications for Ground Truth Establishment
Not applicable. The performance data presented are based on engineering tests and adherence to recognized standards, not on expert-established ground truth from clinical cases.
4. Adjudication Method for Test Set
Not applicable. There is no clinical test set requiring adjudication in the provided performance data.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No multi-reader multi-case (MRMC) comparative effectiveness study is mentioned. This evaluation focuses on the safety and effectiveness of the device itself, rather than comparing human reader performance with and without AI assistance. The device is a videoscope, not an AI diagnostic tool.
6. Standalone (Algorithm Only) Performance Study
Not applicable. This device is a physical videoscope, not an algorithm, so a standalone algorithm-only performance study is not relevant.
7. Type of Ground Truth Used
The "ground truth" for the various performance tests is defined by:
- Established national and international standards (e.g., FDA guidance documents, ISO 10993, ANSI/AAMI ES 60601-1, IEC 60601-1-2)
- Manufacturer's design specifications and requirements.
8. Sample Size for Training Set
Not applicable. This device is a physical videoscope, not a machine learning model, so there is no concept of a "training set."
9. How Ground Truth for Training Set Was Established
Not applicable, as there is no training set mentioned or implied for this device.