    510k Summary Text (Full-text Search) :

    95054

    Re: K233710

    Trade/Device Name: Symphion Operative Hysteroscopy System Regulation Number: 21 CFR 884.1710
    |
    | Classification # | 21 CFR 884.1710
    reviewed by FDA in prior submissions (K132695 and K141848), in accordance with Special Controls defined by 884.1710

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Symphion Operative Hysteroscopy System and accessories are intended for resection and coagulation of uterine tissue such as intrauterine polyps and myomas using a bipolar resecting device by distending the uterus with saline to facilitate viewing with a hysteroscope during diagnostic and operative hysteroscopy and providing fluid management through either the closed loop recirculation of filtered distension fluid or non-filtered distension fluid.

    Device Description

    The Symphion Operative Hysteroscopy System is comprised of the following:

    • Controller with IV Pole and Footswitch
    • Resecting Device
    • Fluid Management Accessories (INFINITY FMA and EXPRESS FMA)
    • Fluid Deficit Readout Optional Accessory

    The Symphion Operative Hysteroscopy System provides an integrated control system with bipolar radiofrequency outputs (cut and coagulation) and fluid management through the use of two integrated peristaltic pumps. The Resecting Device is a bipolar radiofrequency device configured for the resection and aspiration of the uterine pathology. Fluid infusion and aspiration of the uterine cavity are controlled by the Controller's peristaltic pumps, in conjunction with the Fluid Management Accessories. The Symphion Controller, with software version 3.0.0 or higher, is designed to be used in either a recirculating (closed-loop) or non-recirculating system configuration. The Fluid Manaqement Accessories are disposable tubing sets and are available in two models, INFINITY FMA and EXPRESS FMA. The FMAs are designed to assist with the delivery, monitoring, and control of the amount of distention media delivered to the uterine cavity when used with the Symphion controller. When the controller is used with the INFINITY FMA, the system components form a closed-loop recirculating system configuration. Alternatively, when the controller is used with EXPRESS FMA, the system components form an open, nonrecirculating, system configuration. The INFINITY FMA is designed to be used in all types of intrauterine pathology whereas the EXPRESS FMA, lower cost alternative, may be more suitable for softer type tissues or in procedures that require a shorter resection time such as polypectomy, visual D&C and similar.

    The integrated peristaltic pumps are operated by a software pressure control algorithm, specific to use with either INFINITY FMA or EXPRESS FMA configurations, that measures and controls intra-uterine cavity pressure by varying saline infusion and aspiration rates in response to pressure changes during aspiration and tissue resection when using the Resection Device. The Symphion Controller Graphical User Interface has two options for initial set up depending on which FMA configuration is being used during the procedure. There are fluid control settings which allow the user to toggle infusion ON/OFF and to set the target cavity pressure from 45-125mmHg. Subsequent to the initial setup GUI, all procedure and control settings are identical regardless of which FMA configuration is being used. The system includes an optional Fluid Deficit Readout accessory device which is an independent saline bag measurement scale to present the user with a digital real-time fluid loss/fluid deficit reading during the Symphion procedure.

    AI/ML Overview

    The provided document is a 510(k) premarket notification decision letter from the FDA regarding the Minerva Surgical Symphion Operative Hysteroscopy System. It details the device, its intended use, comparison to predicate devices, and a summary of non-clinical testing. However, it does not contain the specific acceptance criteria, reported device performance data, sample sizes, expert qualifications, or adjudication methods typically found in a clinical study report. It states that "Clinical Testing was not required."

    Therefore, I cannot provide the requested information about acceptance criteria, device performance, sample sizes for test or training sets, ground truth establishment, or MRMC studies, as this information is not present in the provided text.

    The document only summarizes that:

    • A range of non-clinical performance tests were conducted.
    • These tests were performed at the system, component, and subassembly levels.
    • The system met its performance specifications and performs as intended.
    • "All results have been deemed acceptable" for various tests (Biocompatibility, Software Verification, Electrical Equipment Safety, Sterilization, Shelf Life Testing, Packaging and Transit Testing).
    • For Performance Testing (Special Controls), which covers aspects like mechanical testing, pressure testing, fluid delivery, flow rate, simulated use, filtration, blood filtration capacity, tissue collection capacity, and filtrate characterization, "All results have been deemed acceptable."

    Since clinical testing was not required for this 510(k) submission, there isn't a "study that proves the device meets the acceptance criteria" in the sense of a clinical trial with human subjects and associated statistical analyses as would be typically implied by your request. The provided text only confirms that non-clinical performance and design verification tests were completed and deemed acceptable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141848

    Validate with FDA (Live)

    Device Name
    IOGYN SYSTEM
    Manufacturer
    Date Cleared
    2014-08-29

    (51 days)

    Product Code
    Regulation Number
    884.1710
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Suite 120 Cupertino, CA 95014

    Re: K141848 Trade/Device Name: Iogyn System Regulation Number: 21 CFR§ 884.1710

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IOGYN System is intended to distend the uterus by filling it with saline to facilitate viewing with a hysteroscope during diagnostic and operative hysteroscopy and provide fluid management through the closed-loop recirculation of filtered distension fluid. It is also intended for cutting and coagulation of uterine tissue such as intrauterine polyps and myomas using a bipolar resecting device.

    Device Description

    The IOGYN System is comprised of the following:

    • IOGYN Controller with Integrated Fluid Management .
      • Footswitch O
      • Fluid Management Accessories o
    • IOGYN Resecting Device ●

    The IOGYN System provides an integrated control system with bipolar radiofrequency outputs (cut and coagulation) and fluid management through the use of two integrated peristaltic pumps. The Resecting Device is a bipolar radiofrequency device configured for the resection and aspiration of uterine pathology. Fluid infusion and aspiration of the uterine cavity are controlled by the IOGYN Controller's peristaltic pumps, in conjunction with the Fluid Management Accessories; these components form a closed-loop recirculating system. These integrated peristaltic pumps are operated by a software pressure control algorithm that measures and controls intra-uterine cavity pressure by varying saline infusion and aspiration rates in response to pressure changes during aspiration and tissue resection using the Resecting Device. The IOGYN Controller Graphical User Interface (GUI) has fluid control settings which allow the user to toggle infusion ON/OFF and to set the cavity pressure from 30-125mmHg.

    The IOGYN System uses a closed-loop fluid management system with a single 3liter saline bag that continuously recirculates filtered distension fluid throughout the procedure. The fluid absorption is limited by the volume within the 3-liter saline bag minus the dead volume within the system, which limits the deliverable volume to less than 2.5L. The 2.5L of recirculated fluid is passed through two levels of filtration - the first level of filtration removes resected tissue and bulk particles; the second level of filtration removes cellular materials, hemoglobin, plasma proteins, cytokines, coagulation factors, bacteria and viruses. These two levels of filtration generate optically clear, sterile, filtered distension fluid which is returned to the 3 liter saline bag and recirculated.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called the IOGYN System. It focuses on demonstrating substantial equivalence to a predicate device after minor material changes and does not provide detailed information about acceptance criteria and comprehensive study results in the format requested.

    Therefore, much of the requested information cannot be extracted from this document.

    However, I can extract what is available:

    1. A table of acceptance criteria and the reported device performance

    The document mentions "Performance Testing" which includes:

    • Simulated Use
    • Durability Testing
    • Hi Pot Testing
    • Tensile Strength
    • Flow Rate
    • Pressure Testing
    • System undeliverable volume (dead volume)

    It states "All device bench test results were acceptable. Data demonstrate that the Resecting Device and the Fluid Management Accessories meet design specifications and are suitable for the intended use and labeled shelf-life."

    Without the specific design specifications or quantitative acceptance criteria, a table cannot be fully constructed. The document only provides a qualitative statement that results were "acceptable" and met "design specifications."

    Acceptance CriteriaReported Device Performance
    Not specifiedAcceptable
    Not specifiedMeets design specifications
    Not specifiedSuitable for intended use
    Not specifiedSuitable for labeled shelf-life

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample sizes for any of the performance tests. There is no information regarding the provenance of the data (e.g., country of origin) or whether it was retrospective or prospective. The testing described appears to be bench/laboratory testing rather than clinical study data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This entire section is not applicable. The document describes bench testing for material changes in a medical device (IOGYN System), not a study involving human subjects or image interpretation that would require expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a study assessing diagnostic performance or requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/software device and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. Ground truth as typically defined in diagnostic studies is not relevant here. The "ground truth" for the performance tests would be the established engineering specifications and validated test methods for mechanical, electrical, and fluid dynamics properties.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that would have a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN130040

    Validate with FDA (Live)

    Device Name
    IOGYN SYSTEM
    Manufacturer
    Date Cleared
    2014-03-28

    (210 days)

    Product Code
    Regulation Number
    884.1710
    Type
    Direct
    Age Range
    All
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    .

    NEW REGULATION NUMBER: 21 CFR 884.1710

    CLASSIFICATION: CLASS II

    PRODUCT CODE: PGT

    BACKGROUND

    Loop Hysteroscopic Insufflator with Cutter-coagulator Class: II (Special Controls) Regulation: 21 CFR 884.1710

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The IOGYN System is intended to distend the uterus by filling it with saline to facilitate viewing with a hysteroscope during diagnostic and operative hysteroscopy and provide fluid management through the closed loop recirculation of filtered distension fluid. It is also intended for cutting and coagulation of uterine tissue such as intrauterine polyps and myomas using a bipolar resecting device.

    Device Description

    The IOGYN System is comprised of the following: IOGYN Controller with Integrated Fluid Management, Footswitch, Fluid Management Accessories, IOGYN Resecting Device. The IOGYN System provides an integrated control system with bipolar radiofrequency outputs (cut and coagulation) and fluid management through the use of two integrated peristaltic pumps. The Resecting Device is a bipolar radiofrequency device configured for the resection and aspiration of uterine pathology. Fluid infusion and aspiration of the uterine cavity are controlled by the IOGYN Controller's peristaltic pumps, in conjunction with the Fluid Management Accessories; these components form a closed loop recirculating system. These integrated peristaltic pumps are operated by a software pressure control algorithm that measures and controls intrauterine cavity pressure by varying saline infusion and aspiration rates in response to pressure changes during aspiration and tissue resection using the Resecting Device. The IOGYN Controller Graphical User Interface (GUI) has fluid control settings which allow the user to toggle infusion ON/OFF and to set the cavity pressure from 30-125mmHg. The IOGYN System uses a closed loop fluid management system with a single 3-liter saline bag that continuously recirculates filtered distension fluid throughout the procedure. The fluid absorption is limited by the volume within the 3-liter saline bag minus the dead volume within the system, which limits the deliverable volume to less than 2.5L. The 2.5L of recirculated fluid is passed through a tissue catch and then a molecular filter to generate optically clear, sterile, filtered distension fluid which is returned to the 3 liter saline bag and recirculated.

    AI/ML Overview

    The request is about acceptance criteria and study information for a specific medical device, IOGYN System. However, the provided text describes the device, its regulatory information, warnings, precautions, and a summary of nonclinical/bench studies, and clinical information without explicitly stating detailed acceptance criteria for each test or a comprehensive study proving specific quantitative acceptance criteria.

    The document outlines various performance tests and their results (all reported as "Pass"), but it often refers to "device specification" or "target specification" without detailing the specific numerical or qualitative targets that constitute acceptance. For example, in the "Performance Testing - Fluid Management" section, it says "Specification for flow rate was met" but doesn't state what that flow rate specification was. Similarly, for "Pressure Control," it states "Actual pressure should not exceed set pressure by (b)(4) Trade Secret/CCI for more than (b)(4) Trade Secret/CCI" but the numerical values are redacted.

    Therefore, a table of acceptance criteria and reported performance with quantitative metrics for each test cannot be fully constructed from the provided text due to the lack of explicit, unredacted numerical acceptance criteria for most tests.

    However, I can extract the available information regarding the studies and performance as much as possible.


    1. Table of Acceptance Criteria and Reported Device Performance

    As noted above, explicit numerical acceptance criteria are largely redacted or not provided in the document. The table below summarizes the tests performed and the reported results (all "Pass"), indicating that the device met its internal specifications, even if those specifications are not detailed in this public document.

    Test CategorySpecific Test / CriterionAcceptance Criteria (as implied/stated)Reported Device Performance
    BiocompatibilityCytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Acute Systemic Toxicity, Hemocompatibility (Hemolysis, complement activation, coagulation), Reproductive Toxicity, Pyrogen testing (Filter and Aspiration/Infusion tubing)Compliance with ISO 10993-1-2009 for patient-contacting materials and filtrate. Filtrate not posing significant reproductive toxicity risk. Filtration for bacteria, endotoxins, viruses meeting applicable test standards.All tests passed; sufficient information provided.
    Shelf Life/SterilitySterility Assurance Level (SAL), Packaging Integrity (Visual, Dye Penetration, Seal Peel Strength), Device Performance (RF connector, vapor/bubble formation, oscillations of inner tube, pressure sensor, tissue resection/aspiration without clogging, discoloration).SAL of 10^-6 (VDmax method). Met ASTM standards for packaging integrity. Met specific functional criteria for device components (e.g., vapor/bubble formation, target oscillations, specific pressure sensor reading).All tests passed. 7-month labeled shelf-life supported.
    EMC and Electrical SafetyElectrical Safety TestingCompliance with IEC 60601-1, UL 60601-1, CSA C22.2 No 601.1, EN 60601-1.Pass (in both active and standby modes)
    System Electromagnetic Compatibility (Radiated/Conducted Emissions, Current Harmonics, Voltage Fluctuations, Electrostatic Discharge, Radiated/Conducted Immunity, Surge, Power Frequency Magnetic Immunity, Voltage Dips and Interrupts)Compliance with IEC 60601-1-2 and IEC 60601-2-2, CISPR 11, IEC 61000 series.Pass (in both active and standby modes)
    SoftwareSoftware development, verification, validation, hazard analysisCompliance with FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005) for 'MAJOR' LOC devices; adequate mitigation of key hazards (fluid overload, overpressurization, air/gas embolism); software performs as intended.Documentation provided; adequate and acceptable.
    Mechanical TestingResecting Device (RD) Shaft Compression/Tensile, RD Cutting Window Bend, RD Cable Tensile, FMA Tubing to Connectors TensileRD shaft resists buckling under specified force; tensile force for RD shaft and cable meets device specification; bending force for cutting window meets device specification; tensile force for FMA tubing meets device specification.All met the acceptance criteria ("Pass").
    Performance Testing (FMA)Temperature and Coagulation TestingMaximum temperature and coagulation depth meet predefined target specifications (ex vivo testing).Pass
    Maximum Aspiration Flow Rate (Diagnostic Mode, Resection Mode)Specification for flow rate met.Pass
    Maximum Infusion Flow RateSpecification for flow rate met.Pass
    Maximum VacuumSpecification for vacuum met.Pass
    Pressure ControlActual pressure should not exceed set pressure by (b)(4) for more than (b)(4) of the time.Pass
    Over Pressure Relief TestingPressure relief activates within (b)(4) of overpressure condition.Pass
    Relief ValvePressure maintained (b)(4) mmHg.Pass
    Backflow ValveValve activates at (b)(4) pressure.Pass
    Dead VolumeVolume less than 2.5L.Pass
    Filtration Testing (Molecular Filter, Tissue Catch Container and Capacity)Molecular filter capable of capturing minimum blood specified while maintaining target pressure. Tissue catch meets filtration ability and maximum tissue capacity specifications.All specifications met ("Pass").
    Durability TestingNo device issues or rust/corrosion after (b)(4) hours of operation and tissue resection.Pass
    Simulated Use TestingTissue coagulation, cutting, aspiration, and visualization were evaluated and met requirements.Pass
    Usability VerificationUsers able to appropriately follow IFU to setup system and perform procedure under simulated use conditions.Pass
    Clinical Performance (Primary Endpoint)Ability to resect tissue with the IOGYN system (Percentage of Polyp/Myoma removed)Myoma: 75-100% removal for the majority of cases. Polyp: 75-100% removal for all cases. (Implied acceptance based on positive presentation of results).Myoma (n=19): 16 (75-100% removed), 3 (0-49% removed). Polyp (n=29): 29 (100% removed).
    Clinical SafetyFluid deficit, Serious adverse events, Unanticipated adverse device effects, Histopathology of removed tissue.Fluid deficit not exceeding 2500 mL. No serious adverse events or unanticipated adverse device effects. Histopathology showing consistent zone of thermal damage and sufficient tissue viability.Fluid deficit did not exceed 2500 mL in any case. No serious adverse events or unanticipated adverse device effects reported. Histopathology showed consistent zone of thermal damage and sufficient tissue viability.

    2. Sample Sizes and Data Provenance for Test Sets

    • Nonclinical/Bench Studies:
      • Shelf Life Testing: Resection device (n=30) and fluid management accessories (n=30) were tested at baseline and after 7 months of real-time aging. Packaging integrity (peel strength, dye penetration) involved n=10 samples each. Device functionality tests involved n=10 samples.
      • Mechanical Testing: 10 samples following simulated use and corrosion exposure.
      • Performance Testing (Fluid Management): 10 samples of FMA and RDs.
      • Durability Testing: Devices operated for over (b)(4) Trade Secret/CCI (number redacted) hours.
      • Simulated Use Testing: Used (b)(4) Trade Secret/CCI (description redacted).
      • Data Provenance: Bench testing, ex-vivo testing (for Temperature and Coagulation Testing) in a lab setting.
    • Clinical Testing:
      • Sample Size: 36 subjects (age 25 to 65 years with endometrial polyps and/or submucosal myomas) enrolled. 34 subjects were treated.
      • Data Provenance: Prospective (implied by "clinical studies were conducted...between October 2012 and May 2013" with follow-up), multi-country (Mexico, Hungary, Canada).

    3. Number of Experts and Qualifications for Establishing Ground Truth (Clinical Test Set)

    The provided text does not specify the number of experts used to establish ground truth for the clinical test set, nor their specific qualifications (e.g., radiologist with X years of experience). The clinical study focused on the "ability to resect tissue" and observed "histopathology examination of the removed tissue," which implies a pathologist would be involved in ground truth, but no details are given. The report states "The sponsor concluded that a large intramural myoma migrated into the visual field secondary to incision of the myoma capsule," suggesting clinical observation and potentially a surgeon's assessment.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for the clinical test set or any other test set. Clinical outcomes appear to be reported as observed within the study.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed as described in the provided text. The clinical study evaluated the device's performance directly, not human readers' improvement with or without AI assistance. The IOGYN System is a medical device for hysteroscopic procedures, not an AI diagnostic tool that assists human readers.

    6. Standalone (Algorithm Only) Performance Study

    As the IOGYN System is a closed-loop hysteroscopic insufflator with cutter-coagulator, it is a physical device that performs surgical functions and fluid management. It is not an "algorithm only" device in the sense of a diagnostic AI. Its "performance" is inherently linked to its physical operation, and thus, a "standalone algorithm only" study is not applicable in the typical AI context. However, the software components underwent standalone verification and validation testing as part of the overall software safety and performance assessment for a 'MAJOR' level of concern device.

    7. Type of Ground Truth Used

    • Nonclinical/Bench Studies: Ground truth was established by engineering specifications, physical measurements, and defined performance criteria (e.g., flow rates, pressure ranges, mechanical strength limits, bacterial/viral removal percentages, durability hours).
    • Clinical Testing: Ground truth for the primary endpoint (tissue resection) was based on observed surgical outcomes (percentage of myoma/polyp resected) and histopathology results of the removed tissue. Safety ground truth included clinical observation of adverse events and fluid deficit measurements.

    8. Sample Size for the Training Set

    The document does not specify a "training set" in the context of an AI/machine learning algorithm, as the IOGYN System is a physical device with integrated software for control, not a learning algorithm that requires a traditional training set. The software itself underwent hazard analysis, verification, and validation, but this is distinct from training an AI model.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable as there is no mention of a "training set" for an AI/machine learning algorithm in the provided document. The software's "ground truth" or correct functionality was established through software requirements specifications, design documents, and subsequent verification and validation testing against those specifications.

    Ask a Question