Search Results
Found 10 results
510(k) Data Aggregation
(106 days)
Redmond, Oregon 97756
Re: K252951
Trade/Device Name: Genesis Sleep
Regulation Number: 21 CFR 882.5800
Classification Name: Cranial Electrotherapy Stimulator To Treat Insomnia And/Or Anxiety
Regulation Number: 882.5800
Indications comment above. |
| Regulatory Class | Class II | Class II | Identical. |
| Classification Name | 882.5800
- Cranial electrotherapy stimulator | 882.5800 - Cranial electrotherapy stimulator | Identical. |
|
from Predicate Device (Modius Sleep, K230826)
To meet the special control requirement under 21 CFR 882.5800
Genesis Sleep is a non-invasive, clinical- and home-use neurostimulation device that is indicated to treat chronic insomnia in adults aged 22 and older.
Genesis Sleep is a non-invasive transdermal neurostimulation device to treat chronic insomnia. The device utilizes a technology known as electrical vestibular nerve stimulation (VeNS). It is indicated for clinical- and home-use to treat chronic insomnia in adults aged 22 and older.
It consists of a battery-powered device designed to transcutaneously deliver low-level electrical energy (up to 1 mA) to the skin behind the ears, over the mastoid processes. The delivery of this neurostimulation is through two single-use self-adhesive electrode pads. These pads are placed on the skin behind each ear (mastoid area).
The intensity of the electric pulse can be adjusted up or down by the user. When turned on, the device delivers a small electrical impulse which stimulates the vestibular nerve.
When the device is not being used, the battery can be charged using the charging station provided. Therefore, it is not physically possible to recharge the battery while the device is in use in stimulation mode.
The Genesis Sleep device exerts its therapeutic effect through electrical stimulation of the vestibular nerves. Electrical vestibular nerve stimulation (VeNS) is a non-invasive therapeutic method that applies pulsed, alternating microcurrent transcutaneously to the vestibular nerves via stimulation pads.
Genesis Sleep delivers this neurostimulation through two self-adhesive stimulation pads which are placed on the skin behind each ear (mastoid processes). When turned on, the device delivers a small electrical impulse which can be adjusted up or down by the user.
N/A
Ask a specific question about this device
(243 days)
Antrim BT44 8BP United Kingdom
Re: K232253
Trade/Device Name: Modius Stress Regulation Number: 21 CFR 882.5800
: Cranial Electrotherapy Stimulator (CES) Regulatory Class: II Product Code: QJQ Regulation number: 882.5800
Modius Stress is a non-invasive, home-use neurostimulation device that is indicated to treat the symptoms of Generalized Anxiety Disorder in adults aged 22 and older, when used for approximately 4 weeks. Durability of effect after 4 weeks has not been established.
Modius Stress is a non-invasive transdermal neurostimulation device to treat the symptoms of Generalized Anxiety Disorder. The proposed mechanism of the device is through a technology known as Cranial Electrotherapy Stimulation (CES). It consists of a battery-powered device designed to transcutaneously deliver low-level electrical energy (up to 1.5mA) to the skin behind the ears, over the mastoid processes. The delivery of this neurostimulation is through two self-adhesive electrode pads. These pads are placed on the skin behind each ear (mastoid area). The intensity of the electric pulse can be adjusted up or down by the user. When turned on, the device delivers a small electrical impulse, and adjustments to the stimulation level may be made using the up and down buttons on the device, which are located just above the power button. The device can also be paused by pressing the power button twice. When finished the user removes the device and disposes of the electrode pads after each use. When the device is not being used it can be charged through a micro-USB cable. For safety reasons, it is not possible to recharge the battery while the device is in use in stimulation mode. The key components of the Modius Stress include the Modius Stress device (plastic enclosure and printed circuit board assembly (PCBA), stimulation pads (K210448 and K132588), skin cleansing wipes (K121655) and charging accessories. The PCBA consists of a microcontroller, USB connector, transformer driver, IO expander and EEPROM memory. The embedded software within the device manages overall functionality of the device from Stimulation Control, Power Management, and user interaction (Indication LED and Audio tones).
The Modius Stress device is a non-invasive neurostimulation device indicated to treat symptoms of Generalized Anxiety Disorder (GAD) in adults aged 22 and older when used for approximately 4 weeks.
Here's an analysis of the acceptance criteria and supporting studies based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" in a quantitative performance table commonly seen for AI/ML devices (e.g., sensitivity, specificity, AUC thresholds). However, implicitly, the acceptance criteria for the clinical studies focused on demonstrating a significant reduction in GAD-7 scores compared to a sham control.
| Acceptance Criterion (Implicit) | Reported Device Performance (Modius Stress) |
|---|---|
| Significant reduction in GAD-7 score compared to sham control (primary effectiveness endpoint) | India Study (met): Mean GAD-7 score reduction of -7.44 in Modius Stress group vs -2.23 in sham group (p<0.001).UK Study (not statistically significant for primary endpoint): Mean GAD-7 score reduction of -5.41 in Modius Stress group vs -3.96 in sham group (p=0.139).Post-hoc Subgroup Analysis (UK): Mean GAD-7 score reduction of 7.44 in Modius Stress group (baseline GAD-7 ≥ 10) vs 5.36 in sham control. |
| Safety profile comparable to sham control (primary safety endpoint) | India Study: No causally related adverse events or serious adverse events reported in either group.UK Study: 7 minor adverse events reported (5 in Modius Stress, 2 in sham). No serious adverse events reported. |
| Device biocompatibility | Passed ISO 10993-1. |
| Electrical safety | Complies with IEC 60601-1. |
| Electromagnetic compatibility (EMC) | Complies with IEC 60601-1-2. |
| Usability | Complies with IEC 62366. |
| Maximum Current & Phase Charge Testing | Passed (detailing max current/power output, max phase charge, max current density, max power density). |
| System Verification and Validation Testing | Passed (hardware and firmware). |
| Software Verification and Validation Testing | Complies with FDA Guidance, with "moderate" level of concern. |
2. Sample Size Used for the Test Set and Data Provenance
The "test set" refers to the participant groups in the clinical trials.
- UK Study:
- Sample Size: 87 eligible participants enrolled (43 Modius Stress, 44 sham control). 84 participants (42 Modius Stress, 42 sham control) were included in the ITT analysis for primary effectiveness and safety. 75 participants (36 Modius Stress, 39 sham control) were included in the complete case analysis.
- Data Provenance: Prospective, multi-site human clinical trial conducted in the United Kingdom (UK) and Republic of Ireland (ROI) (remote recruitment).
- India Study:
- Sample Size: 60 eligible participants enrolled (34 Modius Stress, 26 sham control). All 60 participants were included in the ITT and complete case analyses for both primary effectiveness and safety.
- Data Provenance: Prospective, human clinical trial conducted in India.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The ground truth in this context is the assessment of Generalized Anxiety Disorder (GAD) symptoms, primarily using the GAD-7 score. The document does not specify the number or qualifications of experts used to establish the "ground truth" for the test set in the traditional sense of medical image annotation or diagnosis. Instead, the GAD-7 scale itself is a validated self-report questionnaire, commonly used in clinical practice and research for screening and measuring the severity of GAD. The study recruited adults with a GAD-7 score of 10 or higher. While medical professionals (e.g., prescribing physicians) are involved in the overall care, the "ground truth" for GAD-7 scores is derived directly from participant self-reporting on this validated instrument.
4. Adjudication Method for the Test Set
Not applicable in the context of GAD-7 scores. The GAD-7 is a standardized questionnaire without a need for independent expert adjudication of each score. The studies were double-blinded, meaning neither participants nor the clinical study team knew the device allocation (Modius Stress vs. sham).
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance
This is not an AI/ML device in the context of image interpretation or diagnostic assistance where MRMC studies are typically performed. The device is a neurostimulation therapy device. Therefore, a MRMC comparative effectiveness study was not done.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done
The Modius Stress device is a therapeutic neurostimulation device, not an AI algorithm. Its performance is evaluated on its direct effect on physiological conditions, not on its ability to interpret data or provide diagnostic insight. Its usage is standalone in the sense that the device delivers the stimulation (algorithm) without constant human intervention during a session, but it is to be used by a human patient for a therapeutic purpose. The clinical trials evaluated the device's therapeutic effect directly.
7. The Type of Ground Truth Used
The primary ground truth used in the clinical studies was the Generalized Anxiety Disorder 7th edition (GAD-7) score, which is a validated self-report questionnaire for assessing GAD symptoms. Participants with a GAD-7 score of 10 or higher were recruited, indicating at least moderate anxiety.
8. The Sample Size for the Training Set
The Modius Stress device is a hardware neurostimulation device that applies Cranial Electrotherapy Stimulation (CES) and is not an AI/ML software algorithm that requires a "training set" in the conventional sense. Therefore, there is no training set for an AI algorithm. The device's parameters and design would have been developed based on scientific understanding of CES and prior research (including "general CES summary" referenced in the text), rather than a machine learning training dataset.
9. How the Ground Truth for the Training Set was Established
As there is no AI/ML algorithm requiring a training set, this question is not applicable. The device's underlying principles for neurostimulation are based on existing knowledge of Cranial Electrotherapy Stimulation (CES).
Ask a specific question about this device
(217 days)
BT44 8BP United Kingdom
Re: K230826
Trade/Device Name: Modius Sleep Regulation Number: 21 CFR 882.5800
Cranial Electrotherapy Stimulator (CES) Regulatory Class: II Product Code: Requested Regulation number: 882.5800
panel: Neurology Special Controls: Modius Sleep complies with the special controls as set in FDA 21 CFR 882.5800
Modius Sleep is a non-invasive, home-use neurostimulation device to treat chronic insomnia in adults aged 22 and older.
Modius Sleep is a low-risk non-invasive transdermal neurostimulation device to treat chronic insomnia. The proposed mechanism of the device is through a technology known as electrical vestibular nerve stimulation (VeNS).
It consists of a battery-powered device designed to transcutaneously deliver low-level electrical energy (up to 1mA) to the skin behind the ears, over the mastoid processes. The delivery of this neurostimulation is through two self-adhesive electrode pads. These pads are placed on the skin behind each ear (mastoid area). The intensity of the electric pulse can be adjusted up or down by the user. When turned on, the device delivers a small electrical impulse which stimulates the vestibular nerve. Adjustments to the stimulation level may be made using the up and down buttons on the device, which are located just above the power button. The device can also be paused by pressing the power button twice. When finished the user removes the device and disposes of the electrode pads after each use. When the device is not being used it can be charged through a micro-USB cable. For safety reasons, it is not possible to recharge the battery while the device is in use in stimulation mode.
The key components of the Modius Sleep include the Modius Sleep device (plastic enclosure and printed circuit board assembly (PCBA), stimulation pads (K210448 and K132588), skin cleansing wipes (K121655) and charging accessories. The PCBA consists of a microcontroller, USB connector (charging only), transformer driver, IO expander and EEPROM memory. The embedded software within the device manages overall functionality of the device from Stimulation Control, Power Management, and user interaction (Indication LED and Audio tones).
The provided text is a 510(k) Summary for the Modius Sleep device. It details a clinical study to demonstrate the device's safety and effectiveness for treating chronic insomnia. Here's a breakdown of the requested information based on the document:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" for performance in a table format. However, it defines the primary effectiveness endpoint as a "change in the Insomnia Severity Index (ISI) score at Week 0 and Week 4 between the Modius Sleep group and sham control group." The study then reports the mean change and statistical significance. The implicit acceptance criterion is a statistically significant greater reduction in ISI score for the Modius Sleep group compared to the sham control.
| Metric (Implicit Acceptance Criteria) | Modius Sleep Group (Mean ISI Change from Baseline to Week 4) | Sham Control Group (Mean ISI Change from Baseline to Week 4) | Between-Group Difference (95% CI); P-value (Model 1) | Performance Met? |
|---|---|---|---|---|
| ISI Score Reduction (Statistically Significant Improvement vs. Sham) | -5.80 (Complete Case Analysis) | -3.52 (Complete Case Analysis) | -2.28 (-3.85, -0.71); 0.005 | Yes |
Note: The table uses data from the "Complete case" analysis as it represents those who adhered to the intervention and showed a statistically significant difference in Model 1. Other models and ITT analyses are also provided in the document.
2. Sample size used for the test set and the data provenance
- Test Set (Participants with available data for analysis): 126 participants (61 in Modius Sleep group, 65 in Sham control group for complete case analysis). A total of 149 participants were enrolled and randomized (75 Modius Sleep, 74 Sham control).
- Data Provenance:
- Country of Origin: United Kingdom (UK) and Hong Kong (HK). The UK site also included remote recruitment from the Republic of Ireland (ROI).
- Retrospective or Prospective: Prospective. The study was a "4-week pivotal randomized controlled trial."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document does not mention the use of experts to establish a "ground truth" for the test set in the traditional sense (e.g., radiologists interpreting images). Instead, the "ground truth" for insomnia diagnosis and severity (ISI scores) was established via self-reported patient outcomes and an online survey based on clinical diagnostic criteria (AASM ICSD and DSM-V) prior to enrollment. There is no information provided about the number or specific qualifications of experts involved in the initial diagnosis or ISI score evaluation for the study participants, beyond the general statement that "further information was collected by an online survey to establish that the participants had chronic insomnia at the time of enrolment as defined by ICSD and DSM-V."
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This was a clinical trial evaluating a treatment device based on self-reported patient outcomes (ISI scores), not an AI device requiring reader adjudication of findings.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This study evaluated a medical device for treatment of insomnia in a direct patient-use scenario, not an AI-assisted diagnostic tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This was a clinical trial of a medical device used by patients, not an algorithm. Therefore, "standalone (algorithm only)" performance is not applicable. The device's performance is inherently tied to human interaction (patient use).
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for effectiveness was outcomes data based on the Insomnia Severity Index (ISI) scores, which are a validated patient-reported outcome measure for insomnia. The initial diagnosis of chronic insomnia for inclusion criteria was based on patient self-reporting via an online survey aligned with AASM (American Academy of Sleep Medicine) International Classification of Sleep Disorders (ICSD) and The Diagnostic and Statistical Manual of Mental Disorders, Fifth Edition (DSM-V) criteria.
8. The sample size for the training set
Not applicable. This clinical study evaluated the device's performance in a randomized controlled trial. There is no mention of a "training set" in the context of an algorithm or AI development.
9. How the ground truth for the training set was established
Not applicable, as there was no training set in the context of an algorithm.
Ask a specific question about this device
(231 days)
ORA Device Type: Virtual reality behavioral therapy device for pain relief Regulation Number: 21 CFR 882.5800
EaseVRx is a prescription-use immersive virtual reality system intended to provide adjunctive treatment based on cognitive behavioral therapy skills and other evidencebased behavioral methods for patients (age 18 and older) with a diagnosis of chronic lower back-pain (defined as moderate to severe pain lasting longer than three months). The device is intended for in-home use for the reduction of pain and pain interference associated with chronic lower back pain.
EaseVRx is an immersive virtual reality (VR) system which delivers 3-D VR treatment that incorporates principles of cognitive behavioral therapy (CBT), other behavioral methods, and mindfulness strategies to treat patients diagnosed with chronic pain. It is a prescription-use device which uses preloaded software content on a proprietary hardware platform to deliver treatment. The EaseVRx device, shown in the figures below, consists of an off-the-shelf VR head-mounted display (Figure 1) with added Breathing Amplifier (Figure 2) and AppliedVR developed software. The Breathing Amplifier is a mechanical attachment added to the commercially available headset which enables diaphragmatic breathing exercises designed to enhance the user's engagement by amplifying the user's exhalation into the on-board microphone. The device is also marketed with an optional hand-held controller to aid in navigating the user interface.
Acceptance Criteria and Device Performance for EaseVRx
1. Table of Acceptance Criteria and Reported Device Performance
The provided document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for device performance. However, based on the identified risks and special controls, and the clinical study outcomes presented as evidence of effectiveness, the implicit acceptance criteria are related to:
- Clinical effectiveness: Demonstrating a clinically meaningful and statistically significant reduction in pain intensity and pain interference (activity, mood, sleep, stress).
- Safety: Acceptable adverse event profile with temporary, resolvable events.
- Technical Compliance: Meeting relevant electrical, biocompatibility, and software standards.
Here's a table summarizing the implicit performance criteria and the reported device performance, derived from the clinical study results:
| Acceptance Criterion (Implicit) | Reported Device Performance (EaseVRx Group) |
|---|---|
| Clinical Effectiveness: | |
| 1. Reduction in Pain Intensity: Clinically meaningful (>=30%) and statistically significant reduction in average pain intensity. | 66% of EaseVRx participants achieved >30% reduction in pain intensity. 46% achieved >50% pain reduction. Average reduction: 41.6%. Statistically significant difference observed compared to sham (p-value for Condition effect is (b)(4)). |
| 2. Reduction in Pain Interference with Activity: Clinically meaningful (>=30%) and statistically significant reduction in pain interference with activity. | 71% of EaseVRx participants achieved >30% reduction in pain interference with activity. 56% achieved >50% reduction. Average reduction: 48.9%. Statistically significant difference observed compared to sham (p-value for Condition effect is (b)(4)). |
| 3. Reduction in Pain Interference with Mood: Clinically meaningful (>=30%) and statistically significant reduction in pain interference with mood. | 74% of EaseVRx participants achieved >30% reduction in pain interference with mood. 60% achieved >50% reduction. Average reduction: 55.7%. Statistically significant difference observed compared to sham (p-value for Condition effect is (b)(4)). |
| 4. Reduction in Pain Interference with Sleep: Clinically meaningful (>=30%) and statistically significant reduction in pain interference with sleep. | 70% of EaseVRx participants achieved >30% reduction in pain interference with sleep. 60% achieved >50% reduction. Average reduction: 51.7%. A statistically significant difference (condition effect p-value is (b)(4)) was observed compared to sham. |
| 5. Reduction in Pain Interference with Stress: Clinically meaningful (>=30%) and statistically significant reduction in pain interference with stress. | 76% of EaseVRx participants achieved >30% reduction in pain interference with stress. 63% achieved >50% reduction. Average reduction: 57.0%. Statistically significant difference observed compared to sham (p-value for Condition effect is (b)(4)). |
| Early Resolution/Temporary Adverse Events: Adverse events (e.g., nausea, motion sickness, discomfort) should be temporary and resolve with discontinuation or adjustment of the device. No serious adverse events. | 9.7% of EaseVRx group reported nausea and motion sickness. 20.8% reported discomfort with the headset. All reported adverse events were common and expected for VR devices, temporary, and resolved by discontinuing use or adjusting the device. No device-related serious adverse events were observed. |
| Biocompatibility: Patient-contacting components must be demonstrated to be biocompatible (though in this case, data was not evaluated due to history of safe use and low cumulative contact). | Biocompatibility data was not evaluated for EaseVRx, as the materials have a history of safe use in medical devices, total cumulative contact duration is low, and manufacturing complies with Quality System Regulations. |
| Electromagnetic Compatibility & Electrical Safety: Device must comply with relevant safety standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 62133). | Device was tested and demonstrated compliance with IEC 60601-1:2005 (Modified), IEC 60601-1-2:2014, IEC 60601-1-11:2015, and IEC 62133 Edition 2.0 2012-12 for battery. |
| Software Verification, Validation, and Hazard Analysis: Software should operate as described in specifications, with identified hazards addressed through V&V testing. | Sponsor provided documentation for software and firmware with a "Minor" Level of Concern. Hazard analysis performed and V&V testing addressed potential hazards with satisfactory results. Documentation described software, firmware, requirements, traceability, revision history, and cybersecurity. |
| Labeling Compliance: Labeling must meet 21 CFR Part 801.109, provide instructions, safety warnings (nausea, discomfort), and a summary of clinical testing. | User Manual meets requirements of 21 CFR Part 801.109. It includes instructions, appropriate warnings/precautions, and outlines cleaning methods. Provides information on safety and effectiveness. (The document states "The labeling provides information to users describing the clinical data showing the safety and effectiveness..."). |
2. Sample Size and Data Provenance
- Sample Size (Test Set): 188 subjects (patients with chronic low back pain, age 18 and older).
- Treatment group: (b)(4) participants (derived from percentages, e.g., for pain intensity 66% of 188 = approximately 124, 41% of 188 = approximately 77. The specific group counts are redacted as (b)(4) in the table but the percentages confirm a distribution of the 188 subjects between the groups.)
- Sham group: (b)(4) participants.
- Data Provenance: The document does not explicitly state the country of origin. It describes a "single-cohort, double-blinded... randomized clinical trial." This implies a prospective data collection, initiated for regulatory submission.
3. Number of Experts and Qualifications for Ground Truth
- Ground Truth for Clinical Study: No external experts were used to establish the "ground truth" for the primary clinical effectiveness endpoints. The ground truth was patient-reported outcomes (PROs) using standardized scales. Therefore, no information on the number or qualifications of experts for ground truthing is applicable here. The primary endpoints were measured using the Defense and Veterans Pain Rating Scale (DVPRS), which is a self-assessment tool.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. Since the primary endpoints were patient-reported outcomes (DVPRS scores), there was no need for expert adjudication. The study collected direct patient responses.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study: No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This type of study typically evaluates the performance of human readers (e.g., radiologists) with and without AI assistance for tasks like image interpretation. EaseVRx is a behavioral therapy device, and its primary endpoints are patient-reported pain and pain interference, not diagnostic accuracy involving expert readers.
- Effect Size of Human Readers with/without AI: Not applicable, as no MRMC study was performed.
6. Standalone (Algorithm Only) Performance Study
- Standalone Performance: Not applicable in the traditional sense of an "algorithm only" performance study. EaseVRx is an interactive virtual reality system designed for human-in-the-loop use. Its effectiveness relies on the patient's engagement with the software content and the behavioral therapy it delivers. The clinical study did evaluate the device's performance in a standalone manner relative to a sham control, demonstrating the direct effect of the EaseVRx therapeutic program on patient outcomes when used as intended by the patient. However, this is not a "standalone algorithm" performance in the context of, for example, an AI diagnostic algorithm, where the algorithm provides an output without human intervention.
7. Type of Ground Truth Used
- Ground Truth Type: Patient-Reported Outcomes (PROs).
- The primary effectiveness endpoints (average pain intensity and pain interference on activity, mood, sleep, and stress) were measured using the Defense and Veterans Pain Rating Scale (DVPRS).
- Secondary endpoints also included PROs such as Patient's Global Impression of Change (PGIC), PROMIS physical function, PROMIS sleep disturbance, Pain Self-Efficacy Questionnaire (PSEQ-2), Pain Catastrophizing Scale (PSCS-4), Chronic Pain Acceptance Questionnaire (CPAQ-8), Patient satisfaction, and Positive and Negative Affect Scale (PANAS).
8. Sample Size for the Training Set
- Training Set Sample Size: The document refers to the clinical trial as an evaluation study for the device's effectiveness. It does not describe a separate "training set" for the device's software algorithm in the context of an AI/machine learning development pipeline. EaseVRx delivers preloaded software content based on established therapeutic principles (CBT, behavioral methods, mindfulness). The device's content is "designed to minimize triggers of emotional distress and motion sickness, as well as to induce relaxation and activation of the parasympathetic nervous system." This suggests content designed by experts, rather than an algorithm trained on patient data. Therefore, the concept of a separate "training set" for an AI model that could be assessed independently is not directly addressed or implied by the provided text.
9. How Ground Truth for the Training Set Was Established
- Ground Truth for Training Set: Not applicable. As discussed above, the document does not describe the development of an adaptive AI algorithm that would require a ground truth-labeled training set. The device utilizes preloaded, expert-designed therapeutic content.
Ask a specific question about this device
(195 days)
Suite 312 Carmel, Indiana 46032
Re: K182311
Trade/Device Name: Cervella Regulation Number: 21 CFR 882.5800
Classification Name: | Stimulator, Cranial, Electrotherapy, a preamendmentClass III device per 21 CFR §882.5800
Cervella Cranial Electrotherapy Stimulator (CES) is indicated for treatment of insomnia, depression, or anxiety.
Cervella Cranial Electrotherapy Stimulator (CES) is a device that delivers small pulses of electrical current through patient's brain. The stimulator is powered by an internal rechargeable battery which provides a low-level constant current to the cranium via a pair of conductive electrodes placed bilaterally on the mastoid process. The electrodes are incorporated into earpads of stereo over-ear noise-cancelling headphones. The patient can use the audio (through a separate dedicated Bluetooth connection) and noise cancelling features of the stereo headphones during treatment (e.g. music listening). The stimulator is controlled via a software application (app) installed on the patient's smart device (e.g. smartphone). The app communicates with the device through a dedicated Bluetooth® LE connection that is independent of the optional Bluetooth connection for audio listening purposes. The patient uses the app to adjust the intensity level, frequency, and duration of the treatment. The app also allows the patient to automatically store the treatment history for review by the patient's healthcare provider.
This submission lacks specific acceptance criteria and detailed study results to demonstrate device performance against those criteria. The provided document is a 510(k) summary, which focuses on establishing substantial equivalence to predicate devices rather than proving a device meets specific, pre-defined performance acceptance criteria through clinical studies.
Here's an analysis based on the available text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly state acceptance criteria in terms of clinical or technical performance metrics (e.g., specific reduction in anxiety scores, minimum current output accuracy, etc.) and then report the device performance against those criteria. Instead, it relies on demonstrating substantial equivalence to predicate devices based on shared design, operating principles, and compliance with general safety and performance standards.
The table presented in the "IX. PERFORMANCE DATA DEMONSTRATING SUBSTANTIAL EQUIVALENCE" and "X. SUMMARY OF SIMILARITIES AND DIFFERENCES" sections compares the Cervella device's characteristics to its predicate devices, CES Ultra and Alpha-Stim CS. This comparison serves as the de facto "performance data" that the manufacturer presented for substantial equivalence, rather than a direct measurement against pre-defined acceptance criteria for a new clinical claim.
| Property | Cervella (Reported Performance) | Predicate 1 (CES Ultra) | Predicate 2 (Alpha-Stim CS) | Differences and Comments |
|---|---|---|---|---|
| Indications for Use | Treatment of insomnia, depression, or anxiety. | Treatment of insomnia, depression, or anxiety. | Treatment of insomnia, depression, or anxiety. | None |
| Waveform | Symmetrical Biphasic Square Wave | Symmetrical Biphasic Square Wave | Symmetrical Biphasic Square Wave | None |
| Current Intensity Range | 0µA – 500µA adjustable in 50µA increments | 0µA — 1500µA continually adjustable | 0µA — 500µA adjustable in 50µA increments | Same range as Alpha-Stim, narrower range than CES Ultra |
| Pulse Width Range | 5ms - 1s (varies depending on frequency selection) | 2ms | 250ms - 1s | Pulse width varies depending on frequency selection |
| Number of electrodes | Two | Two | Two | None |
| Electrode Placement | Head (Mastoid process) | Head (Mastoid process or Earlobes) | Head (Earlobes) | Same placement as CES Ultra |
| Power Source | Li-ion Battery (3.7V) (rechargeable) | Alkaline Battery (9V) | Alkaline Battery (3V) | Cervella battery is rechargeable |
| Frequency selections | 0.5Hz, 1.5Hz, 100Hz | 100Hz | 0.5Hz, 1.5Hz, 100Hz | Same frequency selection as Alpha-Stim CS |
| Treatment Range | 10 min – 60 min in 10 min intervals | 30 min, 60min, or continuous | 10, 20, 60 minutes or continuous | Same min-max range as Alpha-Stim |
| Unit Controls | Via Bluetooth-enabled smart device featuring Cervella app | Built into the device | Built into the device | Wireless operation |
| Treatment history logging | Automatic | Not available | Not Available | New feature |
| Ability for patient to use audio and noise cancelling feature of headset during treatment | Optional via dedicated separate Bluetooth connection | Not available | Not Available | New feature |
| Dimensions and weight | 7cm x 7cm x 2cm | 13.5cm x 6.4cm x 3.3cm | 9.8cm x 6.3cm x 2cm | Cervella is slightly smaller |
| Enclosure | Plastic | Plastic | Plastic | None |
2. Sample size used for the test set and data provenance:
The document (a 510(k) summary) does not describe a clinical "test set" in the context of typical AI/ML performance evaluation (e.g., a cohort of patients whose data is used to validate an algorithm's diagnostic or therapeutic efficacy).
The performance data cited are primarily bench testing against electrical safety and performance standards for medical devices (ES60601-1:2005, IEC 60601-1-2, IEC 60601-2-10, IEC 62133). These tests do not involve patient data or clinical outcomes. Therefore, there is no information on sample size, country of origin, or whether it was retrospective or prospective data in the clinical sense.
3. Number of experts used to establish the ground truth for the test set and their qualifications:
Not applicable. As noted above, there was no clinical test set requiring expert ground truth establishment for diagnostic or therapeutic accuracy. The ground truth for electrical safety and performance standards is defined by the standards themselves, not by expert consensus on clinical data.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
Not applicable, as there was no clinical test set requiring adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. The Cervella device is a Cranial Electrotherapy Stimulator (CES), a standalone therapeutic device, not an AI-assisted diagnostic or interpretive tool that would typically be evaluated in an MRMC study with human readers. The document does not describe any such study.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
The device itself is a standalone therapeutic device. Its core function – delivering electrical stimulation – is algorithm-driven in terms of current intensity, frequency, and duration programming, but it's not an "AI algorithm" in the typical sense of a diagnostic or predictive model. The performance data provided is mainly about its physical and electrical characteristics conforming to standards, not a standalone evaluation of an AI algorithm's diagnostic accuracy.
7. The type of ground truth used:
For the evaluation described, the "ground truth" was compliance with mandatory and voluntary medical device safety and performance standards (e.g., IEC 60601 series). These standards define the expected electrical output, safety features, and battery performance. There is no mention of expert consensus, pathology, or outcomes data as "ground truth" for the purpose of this 510(k) submission, as it focuses on substantial equivalence of physical and electrical parameters.
8. The sample size for the training set:
Not applicable. This device is not an AI/ML algorithm that requires a "training set" of data in the typical sense to learn patterns or make predictions. Its operation is based on predefined electrical parameters.
9. How the ground truth for the training set was established:
Not applicable, as there is no training set for an AI/ML algorithm.
Ask a specific question about this device
(142 days)
TRANSCUTANEOUS ELECTRICAL NERVESTIMULATOR |
| Classification Name | : Neurology(As Per 21 CFR Sections 882.5800
K090052
Trade/Device Name: Cranial Electrical Nerve Stimulator (FM 10/C) Regulation Number: 21 CFR 882.5800
FM 10/C is indicated to be used for:
- CES for the treatment of insomnia, depression, or anxiety. .
- TNS for Symptomatic relief and management of chronic intractable pain and/or . as an adjunctive treatment in the management of post-traumatic and postsurgical acute pain.
FM 10/C is portable Cranial Electrical Nerve Stimulator. FM 10/C combines uniquely two proven therapies CES for the treatment of insomnia, depression, or anxiety and TENS for symptomatic relief of chronic intractable pain, post-traumatic and post-surgical pain. Thus FM 10/C offers a partnership between Transcutaneous Electrical Nerve Stimulation (TENS) and Cranial Electrotherapy Stimulation (CES). FM 10/C has three modes viz. (i) TENS Burst (ii) TENS Conventional and (iii) CES mode. It provides easy mode selection through slide switch and user can select any one of the available modes through it. TENS and CES modes will not operate simultaneously so, if CES Leads are inserted then TENS modes are automatically blocked for the user and vice versa. TENS mode is programmable and user will have the option to select pulse width and pulse rate using analog pots. In addition treatment time is also selectable using a slide switch options available are 30 minutes or continuous treatment time. FM 10/c is only battery operated, with low battery indication.
The provided document does not contain information about acceptance criteria or a study that proves the device meets those criteria. Instead, it is a 510(k) premarket notification summary for a medical device (Cranial Electrical Nerve Stimulator, Model FM 10/C).
This document mainly describes the device, its intended use, and compares its technical specifications to two legally marketed predicate devices (CES Ultra and ELECTRONIC PAIN RELIEVERS) to claim substantial equivalence. The FDA letter confirms the substantial equivalence determination.
Therefore, I cannot provide the requested information, including:
- A table of acceptance criteria and the reported device performance: This information is not present. The document focuses on technical specifications of the new device and predicate devices.
- Sample size used for the test set and the data provenance: No clinical testing or data is mentioned.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no test set or ground truth establishment is described.
- Adjudication method for the test set: Not applicable.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No such study is mentioned.
- If a standalone performance (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable, as this is an electrotherapy device, not an AI algorithm.
- The type of ground truth used: Not applicable.
- The sample size for the training set: Not applicable, as no AI model or training data is discussed.
- How the ground truth for the training set was established: Not applicable.
The document serves to demonstrate that the FM 10/C device is substantially equivalent to existing devices in terms of its technology and intended use, primarily through technical specification comparison and safety considerations (e.g., battery operation). It does not present results from performance studies or clinical trials to meet specific acceptance criteria.
Ask a specific question about this device
(464 days)
Baarerstrasse 8 Zug, Switzerland 6300
Re:
K070412 Trade/Device Name: Elexoma Medic Regulation Number: 21 CFR 882.5800
The "Elexoma Medic" is a battery operated portable cranial electrotherapy stimulator (CES) device that applies electrical current to a patient's head through electrodes clipped onto the patient's earlobes to treat insomnia, depression, or anxiety and The "Elexoma Medic" is a battery operated portable transcutaneous electrical nerve stimulator (TENS) device used to apply microcurrent through self-adhesive electrodes to a patient's skin for the symptomatic relief of intractable pain, chronic intractable pain, post-traumatic and postsurgical pain relief. The "Elexoma Medic" is a prescription device according to Federal law.
The "Elexoma Medic" is a battery operated portable cranial electrotherapy stimulator (CES) device that applies electrical current to a patient's head through electrodes clipped onto the patient's earlobes and a battery operated portable transcutaneous electrical nerve stimulator (TENS) device used to apply microcurrent through self-adhesive electrodes to a patient's skin.
The provided document is an FDA 510(k) clearance letter for the Elexoma Medic device. It states that the device is substantially equivalent to a legally marketed predicate device.
However, this document does not contain information about specific acceptance criteria for performance, nor does it detail a study proving the device meets particular performance metrics.
The letter focuses on the regulatory clearance process and substantial equivalence, not on the detailed technical performance study results. Therefore, I cannot provide the requested information based solely on this document.
To answer your questions thoroughly, I would need access to the 510(k) submission itself, which would include the performance data and methodologies used to demonstrate substantial equivalence.
Ask a specific question about this device
(241 days)
Classification Name: Stimulator, Cranial, Electrotherapy, a preamendment Class III device per 21 CFR § 882.5800
, Washington 98065
Re: K062284
Trade/Device Name: CES Ultra" Stimulator Regulation Number: 21 CFR 882.5800
The CES Ultra is indicated for the treatment of insomnia. depression or anxiety.
The CES Ultra is a device designed to deliver therapeutic electrical and stimulation for cranial electrotherapy stimulation (CES) applications. The stimulator is powered by a standard 9-volt alkaline battery and delivers a single channel of low-level, constant current electrical stimulation to electrodes connected to the patient's skin via conductive lead wires. Stimulation current is applied to the patient via either standard self-adhering conductive electrodes placed on the head or conductive clip-type electrodes attached to the patient's ear lobes. The stimulator's user controls include: a rotary stimulation amplitude control which allows user adjustment of stimulation current and includes an integral power on/off switch; and a pushbutton timer selector switch which allows user selection of either continuous stimulation or timed stimulation sessions with automatic stimulation shut-off.
The provided text is a 510(k) summary for the CES Ultra™ Stimulator, which is a Cranial Electrotherapy Stimulator (CES). Based on the information provided, no specific acceptance criteria or a study proving the device meets those criteria were required or conducted for this 510(k) submission.
Here's a breakdown of why and what information is available based on your requested points:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Not Applicable | Not Applicable |
Reasoning: The document explicitly states: "Neither nonclinical nor clinical testing was required to demonstrate the substantial equivalence of the CES Ultra to the designated predicate devices." This means no specific performance criteria linked to clinical outcomes were set for this submission. The basis for clearance was substantial equivalence to already marketed devices.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not applicable, as no clinical or nonclinical testing was required or conducted to demonstrate substantial equivalence against specific performance criteria.
- Data Provenance: Not applicable.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable, as no clinical or nonclinical testing was required or conducted.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable, as no clinical or nonclinical testing was required or conducted.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC study was done. The device is a direct therapeutic device (Cranial Electrotherapy Stimulator), not an AI-assisted diagnostic or interpretation tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Yes, in spirit, the substantial equivalence determination is a form of standalone evaluation. The FDA reviewed the device's technological characteristics (frequency, pulse width, amplitude) and compared them to those of legally marketed predicate devices to determine that the CES Ultra was "substantially equivalent." This means the device itself, based on its design and parameters, was deemed equivalent without needing a human-in-the-loop study.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Ground Truth Type: The ground truth for this submission was the established safety and effectiveness of the predicate devices (NF-1 Mindpeace CES, K895175, and HP-1 Healthpax CES, K883812). The "ground truth" for the CES Ultra's substantial equivalence was that its stimulation parameters (frequency, pulse width, and amplitude) were "well within the established range of such parameters that are generally accepted as safe and effective for CES" as demonstrated by the predicate devices.
8. The sample size for the training set
- Not applicable. This device is not an AI/ML algorithm that requires a training set in the conventional sense. The "training" here refers to existing knowledge about CES devices from prior marketing clearances.
9. How the ground truth for the training set was established
- Not applicable in the context of an AI/ML training set. The "ground truth" for establishing substantial equivalence was based on the prior regulatory clearances and established safety and effectiveness of the predicate CES devices. These predicates had previously gone through their own approval processes, which may have involved clinical studies or other data to establish their safety and effectiveness. The CES Ultra leveraged this existing "ground truth" by demonstrating its similarity to these already-approved devices.
Ask a specific question about this device
(266 days)
2000 Microcurrent Stimulator Common Name: Stimulator, Cranial Electrotherapy Classification: 21 CFR 882.5800
Treats anxiety,depression, and insomnia |
| Classification | CES, Class IIIPrescription, 882.5800
| CES, Class IIIPrescription, 882.5800 |
| Contraindications | None
Lewis Ward
Re: K060158
Trade/Device Name: Net-2000 Microcurrent Stimulator Regulation Number: 21 CFR 882.5800
Application of electrical current to the head to treat insomnia, depression, or anxiety.
The NET-2000 Microcurrent Stimulator is a precision electronic instrument used for the treatment of anxiety, depression, or insomnia. The treatment is simple and can easily be selfadministered. The device consists of a microprocessor controller box and skin contact electrodes. The small controller box is a software controlled low-intensity output. Current ranges from 0-600 microamperes typically set at 0.5 Hz. Additional frequencies at 1.5 and 100 Hz are available for the physician's use. The waveform is a bipolar asymmetric rectangular shape. Duty cycle is 50% with a 0 net current. Timed treatment is set at 16.5 minutes.
The provided text describes the NET-2000 Microcurrent Stimulator, a device used for the treatment of anxiety, depression, or insomnia. The submission is a 510(k) premarket notification, which seeks to demonstrate substantial equivalence to a legally marketed predicate device.
Here's an analysis of the acceptance criteria and the study (or lack thereof) based on the provided information:
1. A table of acceptance criteria and the reported device performance
The provided information does not explicitly state "acceptance criteria" in terms of performance metrics like sensitivity, specificity, or accuracy. Instead, the submission relies on a comparison table to demonstrate the NET-2000's features are equivalent to its predicate device, the Alpha-Stim 100. This is the primary method of showing "performance" in this context.
| Feature | NET-2000 | Alpha-Stim 100 (Predicate) | Equivalence Justification/Reported Performance |
|---|---|---|---|
| Indication | Treats anxiety, depression, and insomnia | Treats anxiety, depression, and insomnia | Identical |
| Classification | CES, Class III Prescription, 882.5800 | CES, Class III Prescription, 882.5800 | Identical |
| Contraindications | None | None | Identical |
| Power Source | 9 volt battery | 9 volt battery | Identical |
| Current | 0-600 microamperes | 10-600 microamperes | Within comparable range |
| Frequency | 0.5, 1.5, 100 Hz | 0.5, 1.5, 100 Hz | Identical |
| Waveform | Bipolar asymmetric rectangular waves, 50% duty cycle, 0 net current | Bipolar asymmetric rectangular waves, 50% duty cycle, 0 net current | Identical |
| Electrodes | Silver, self-adhesive pads with conduction solution, clip style. Applied to earlobes. | Silver, self-adhesive pads with conducting solution, clip style. Applied to earlobes. | Identical |
| Timer Treatment Settings | 16.5 minutes | 10, 20, 60 minutes and continuous | Similar (device offers a specific timed setting, predicate offers options) |
| Electrical Safety | Confirmed by testing and passing requirements under EN60601-1-2 and EN60601-1 | (Implied predicate meets similar standards for marketing) | Demonstrated through compliance with standards |
| Biocompatibility | For the electrode patient contact is confirmed by testing. | (Implied predicate materials are biocompatible) | Demonstrated through testing |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not describe any clinical study with a test set involving human subjects or data. The "performance testing" mentioned is comparative testing to the Alpha-Stim 100 and electrical safety and biocompatibility testing. This implies bench testing rather than clinical data from a sample size.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. No clinical data or human evaluation with a "ground truth" established by experts is mentioned for this device. The assessment is based on technical specifications and safety/performance standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. As no clinical study with a test set is described, there is no adjudication method.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a microcurrent stimulator, not an AI-powered diagnostic tool. Therefore, an MRMC study related to AI assistance for human readers is irrelevant and not mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This section is typically relevant for AI/ML-based algorithms. The NET-2000 is a physical medical device; its "performance" is assessed through its electrical characteristics, safety, and functional equivalence to a predicate, not through an algorithm's standalone performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable. For this type of 510(k) submission and device, the "ground truth" for demonstrating substantial equivalence is primarily the technical specifications and established safety and effectiveness of the predicate device, alongside adherence to recognized electrical safety and biocompatibility standards. There is no mention of clinical outcomes data in terms of "ground truth" for the device's efficacy. The claim of "safe and effective" is based on the equivalence to a device already deemed safe and effective.
8. The sample size for the training set
Not applicable. There is no mention of a "training set" as this is not an AI/ML device.
9. How the ground truth for the training set was established
Not applicable. As there is no training set, this question is not relevant.
In summary:
This 510(k) submission for the NET-2000 Microcurrent Stimulator focuses on "substantial equivalence" to a predicate device (Alpha-Stim 100) based on technical specifications, indications for use, and a demonstration of electrical safety and biocompatibility. It does not present evidence from clinical trials or studies involving human subjects with a "test set" and "ground truth" as typically seen in submissions for diagnostic algorithms or devices requiring new clinical evidence of efficacy. The "study that proves the device meets the acceptance criteria" in this context refers to the comparative analysis of features, electrical safety testing, and biocompatibility testing, all demonstrating the NET-2000's similarity to the predicate and compliance with relevant standards.
Ask a specific question about this device
(202 days)
Trade/Device Name: Transcranial Electrotherapy Stimulator-A, Model TESA-1 Regulation Number: 21 CFR 882.5800
Treatment with the TESA device is intended to reduce the symptoms of depression, anxiety and insomnia.
The TESA is a transcranial electrotherapy stimulation device. Three electrodes are applied, one to the forehead and one behind each ear. TESA delivers low amplitude AC current for a desired time of 0 to 99:59 minutes. Treatment can be stopped by pressing the "current off" button. Either the operator or patient may press this button at any time. During treatment, peak current amplitude can be adjusted over the available range of 0 - 4 mA. Treatment parameters are monitored and controlled by a built in microprocessor. The display on the front of TESA indicates remaining treatment time and shows the current setting.
During treatment, a modulated high frequency square wave is delivered to the patient electrodes. The amplitude and duration of the positive and negative going portions of this waveform have been carefully designed to deliver no DC component. Moreover, the electrodes are AC coupled to the patient to insure no DC component in the event of device failure.
Additional safety features of the TESA device include rapid measurement of electrode impedance to insure good electrode contact and to maintain controlled current output over changing load conditions (e.g. when electrode contact changes). Detection of poor electrode contact causes the device to stop treatment and illuminate an electrode fault LED. Further treatment is blocked until electrode/skin impedance is restored to nominal range.
The provided document is a 510(k) Summary for the Transcranial Electrotherapy Stimulator-A (TESA). It describes the device, its intended use, and its substantial equivalence to predicate devices. However, it does not contain information about acceptance criteria or a study demonstrating the device meets such criteria in terms of clinical performance metrics (e.g., sensitivity, specificity, accuracy for a diagnostic device, or efficacy for a therapeutic device).
The "Testing" section only states that:
- "Various tests of the hardware and software have been performed to verify that the device works as described in this document."
- "Verification procedures with pass/fail critcria were developed to ensure that the product met all the specified requirements."
- "As part of this vorification, a certified body will be used to determine that the device conforms to the recognized safety standards listed below."
- "The device will not be marketed until it has been tested and complies with the standards."
The listed standards (UL2601-1, IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-10) are safety and electromagnetic compatibility (EMC) standards, not clinical performance standards for therapeutic efficacy.
Therefore, I cannot populate the requested table and information beyond acknowledging what is present in the document.
Here's what can be extracted based on the provided text, and what cannot:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Clinical Performance: (e.g., reduction in symptoms of depression, anxiety, insomnia) | Not provided in this document. The document states the device's intended use is to reduce these symptoms, but no performance metrics, acceptance criteria for these metrics, or study results are presented. |
| Hardware & Software Functionality: Device works "as described," meeting "specified requirements." | "Various tests of the hardware and software have been performed to verify that the device works as described in this document." "Verification procedures with pass/fail critcria were developed to ensure that the product met all the specified requirements." (Specific criteria and results not detailed). |
| Safety Standards Conformance: | Compliance with UL2601-1, IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-10. (Study results for these are not detailed, only the intent to conform). |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not provided. The document does not describe any clinical test set or data provenance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable/Not provided. No clinical performance study using a test set requiring expert-established ground truth is described.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable/Not provided. No clinical performance study using a test set is described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This device is a therapeutic stimulation device, not an AI-assisted diagnostic or interpretation tool. No such study is mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a hardware therapeutic device, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not applicable/Not provided. No clinical performance study requiring ground truth is described.
8. The sample size for the training set
- Not applicable/Not provided. No machine learning or AI training set is mentioned for this device.
9. How the ground truth for the training set was established
- Not applicable/Not provided.
Summary of what is present:
The document details the device's technical specifications, indications for use, and declares its substantial equivalence to other legally marketed cranial electrotherapy stimulators based on similar current amplitudes and frequencies. It also lists the safety and EMC standards it intends to conform to. It does not provide data on clinical efficacy or specific performance criteria related to its stated indications for reducing symptoms of depression, anxiety, and insomnia. This type of information is typically required for higher-risk devices or when a new mechanism of action is being introduced, often necessitating a PMA (Premarket Approval) rather than a 510(k) which focuses on substantial equivalence to existing devices. Given this device is Class III, a PMA would generally be expected for demonstrating efficacy, but a 510(k) was allowed due to substantial equivalence to older, pre-amendment devices or reclassified devices.