Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K202174

    Validate with FDA (Live)

    Date Cleared
    2021-02-10

    (190 days)

    Product Code
    Regulation Number
    882.1835
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    |
    | Classification | Class II - 21 CFR 882.1835, 882.1845

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Digital NeuroPort Biopotential Signal Processing System supports recording, processing, and display of biopotential signals from user-supplied electrodes. Biopotential signals include: Electrocorticography (ECoG), electroencephalography (EEG), electromyography (EMG), electrocardiography (ECG), electroculography (EOG), and Evoked Potential (EP).

    Device Description

    The Digital NeuroPort Biopotential Signal Processing System is used to acquire, process, visualize, archive/record signals as acquired from user-supplied electrodes for biopotential monitoring. Signals are acquired using a headstage relay that attaches to the pedestal interface and digitizes the signal through the hub. The Digital NeuroPort System uses preamplifiers, analog to digital converters, a signal processing unit, and software running on a personal computer to visualize and record biopotentials from electrodes in contact with the body.

    AI/ML Overview

    The document describes the Digital NeuroPort Biopotential Signal Processing System, which is a physiological signal amplifier. The device's substantial equivalence to a predicate device (K090957, NeuroPort Biopotential Signal Processing System) is affirmed based on various performance data.

    Here's an analysis of the acceptance criteria and the supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance:

      Test / CharacteristicAcceptance CriteriaReported Device Performance
      NeuroPlex E Functional Testing
      MatingScrews down on pedestal and LED turns greenPass
      CrosstalkIsolation resistance of 1kΩ at 500 V DCPass
      Label DurabilityIEC 60601-1:2005/A1:2012, Edition 3.1 7.1.3Pass
      Digital AccuracyAppropriate voltages for different filters (0.02-10 kHz Wide, 0.3-7.5 kHz Standard); Peak-to-peak of 500mV ±10%Pass
      Input Impedance≥10MΩPass
      Impedance Measurement820 ± 15% kOhms and 170 ± 15% kOhmsPass
      Current Rating<1APass
      StabilityAll channels have neural data from a simulator after 90 attachments and detachmentsPass
      AttachmentTwo-Finger TightnessPass
      Input Noise≤3 RMSPass
      Crosstalk<44mVPass
      LeakageIEC 60601-1:2005/A1:2012, Edition 3.1Pass
      Breakaway<14lbfPass
      Digital Hub Functional Testing
      Input Power SupplyExternal, medical-gradePass
      FPGA Testing from HeadstageAccommodates up to 128 channels and channel priority starts with first channel and ends with fourth channel.Pass
      Output Power Supply to Headstage4.8VPass
      Full-Scale Analog Input±8.192mV.Pass
      Burn in TestHub can run continuously for 12 hours.Pass
      Compatibility TestValidated data packets received at hub and NeuroPlex E is powered.Pass
      Digital Neural Signal Simulator (DNSS) Functional Testing
      Rechargeable BatteryBattery life is ≥10 hours.Pass
      PowerCharge battery by Digital Data Cable or USBPass
      DigitalDigital Hub recognizes DNSS connected through Data Cable.Pass
      System Functional Testing
      SynchronizationTimestamps aligned within 100 microseconds with maximum capacity of four 128-channel Es, four 128-channel hubs (only one digital data cable from one E to one hub), and two 256 NSPs.Pass
      Channel CountFacilitates up to 512 channels.Pass
      Usability Testing
      IFU ReadabilityUsers are able to configure intended settings, assemble the system, and perform maintenance activities all from instruction in the IFU.Pass
      Impedance, Reference, and Ground SwitchingUsers are able to achieve each possible configuration prompted by the facilitator.Pass
      CleaningUsers do not damage the device during cleaning. Users identify the proper cleaning solutions. Users indicate that the instructions are sufficiently clear.Pass
      Other Performance Data
      Electrical Safety/EMCCompliance with IEC 60601-1:2012, Ed 3.1 and IEC 60601-1-2:2014, Ed 4.0Compliant
      BiocompatibilityEndpoints assessed: cytotoxicity, irritation, or sensitization, per ISO 10993-1:2018 (for NeuroPlexE and pedestal)Achieved
      SterilitySterilized with 100% EtO to a SAL of 10^-6, per ISO 11135-1:2014/07/15Achieved
      Residuals (EtO)EO levels < 4mg/24hr, ECH levels < 9mg/24hr; both < 60mg/30 days, per ISO 10993-7:2008/10/15Achieved
      Shelf Life (Sterile)18 months, validated by accelerated aging per ASTM F1980-16 for representative device Patient Cable. Packaging conforms to ISO 11607-1:2019, ASTM D4332-14, ASTM D4169-16, ASTM F1886-09/(R)2013, ASTM F2096-11, F88/F88M-15.18 months achieved
    2. Sample size used for the test set and the data provenance:
      The document focuses on engineering performance testing, usability testing, and bench testing, rather than studies involving patient data or clinical datasets. Therefore, information about a "test set" in the context of clinical data (e.g., medical images, physiological recordings from patients for algorithm evaluation) is not provided. The samples used for testing are the physical components of the device (NeuroPlex E, Digital Hub, Digital Neural Signal Simulator (DNSS), and the full system). The provenance of these test samples is not explicitly stated beyond being the components of the "Digital NeuroPort Biopotential Signal Processing System." The types of tests conducted are typically performed retrospectively using manufactured device units in a lab setting.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      This information is not applicable to the type of testing described. The document describes engineering, electrical, mechanical, and usability tests for a physiological signal processing system. Ground truth in this context is established by engineering specifications, international standards (e.g., IEC, ISO, ASTM), and predefined functional requirements, not by expert interpretation of patient data. For usability testing, "users" are involved, but their specific qualifications (e.g., specific clinical experience) or number are not detailed beyond "users" being able to interact with the device and its instructions.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      Not applicable. Adjudication methods are typically used in clinical studies when establishing ground truth from multiple expert interpretations of patient data. The current document reports on bench testing, functional verification, and safety compliance, where results are measured against objective criteria from standards or specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      Not applicable. The Digital NeuroPort Biopotential Signal Processing System is a device for recording, processing, and display of biopotential signals. It is not described as an AI-powered diagnostic or interpretive tool that assists human "readers" (e.g., radiologists, cardiologists) in making diagnoses. Therefore, an MRMC study or AI-assisted performance improvement analysis is outside the scope of this submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
      Not applicable. As noted above, the device is a signal processing system, not an AI algorithm intended for standalone diagnostic performance evaluation. The "performance data" presented is about the physical and electrical functioning of the device components and the system as a whole, not about an algorithm's diagnostic accuracy.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
      For the functional and safety tests, the "ground truth" or reference values are defined by:

      • International Standards: e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26, ISO 10993-1, ISO 11135-1, ISO 10993-7, ISO 11607-1, ASTM D4169-09, ASTM D4332-14, ASTM F88/F88M-15, ASTM F1886-09/(R)2013, ASTM F1980-16, ASTM F2096-11.
      • Engineering Specifications: e.g., isolation resistance (1kΩ), digital accuracy (500mV ±10%), input impedance (≥10MΩ), current rating (<1A), input noise (≤3 RMS), Synchronization within 100 microseconds, channel count (up to 512), battery life (≥10 hours), output power supply (4.8V), full-scale analog input (±8.192mV).
      • Functional Demonstrations: e.g., "Screws down on pedestal and LED turns green" for mating, "All channels have neural data from a simulator after 90 attachments and detachments" for stability.
      • Usability Objectives: Users' ability to configure settings, assemble the system, perform maintenance, or use cleaning solutions correctly.
    8. The sample size for the training set:
      Not applicable. This device is not an AI/ML algorithm that requires a training set of data. The "training" in this context refers to the manufacturing and testing of hardware and software against pre-defined specifications.

    9. How the ground truth for the training set was established:
      Not applicable, as there is no "training set" in the context of AI/ML for this device. The development and validation of the device components and system are based on engineering principles and regulatory standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090957

    Validate with FDA (Live)

    Date Cleared
    2009-05-28

    (52 days)

    Product Code
    Regulation Number
    882.1835
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Acquisition System

    Classification Name: CFR 882.1835 Physiologic Signal Amplifier, Product Code: GWL CFR 882.1845

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Blackrock NeuroPort Biopotential Signal Processing System supports recording, processing and display of biopotential signals from user supplied electrodes. Biopotential signals include: Electrocorticography (ECoG), electroencephalography (EEG), electromyography (EMG), electrocardiography (ECG), electrooculography (EOG) and Evoked Potential (EP).

    Device Description

    The Blackrock NeuroPort Biopotential Signal Processing System (System), when connected to customer supplier bio-potential electrodes or the NeuroPort Electrode Array (K042384), supports recording and display of bio-potential signals. Bio-potential signals that can be recorded by the System include:

    1. Electrooculography (EOG),
    2. Electrocorticography (ECoG),
    3. electroencephalography (EEG),
    4. electromyography (EMG),
    5. electroencephalography (ECG) and
    6. evoked potential (EP).
      The System is not a monitoring system. No physiologic alarms are provided. The acquisition and display of bio-potential signals is for the interpretation and use of the clinician.
    AI/ML Overview

    The provided 510(k) summary for the Blackrock NeuroPort Biopotential Signal Processing System does not contain the detailed information typically found in a clinical study report that proves a device meets specific acceptance criteria using a test set. This document is a premarket notification for a medical device to demonstrate substantial equivalence to existing predicate devices.

    Instead, it focuses on comparing the new device's technical specifications and intended use with previously cleared (predicate) devices to argue that it raises "no new questions of safety or effectiveness."

    Therefore, much of the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert involvement, and comparative effectiveness studies is not available in this document.

    Here's an attempt to extract and infer information based on the provided text, while making explicitly clear what is not present:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the sense of predefined performance targets for a clinical study. Instead, it compares the device's technical specifications to those of predicate devices. The implicit acceptance criterion is that the device's performance characteristics are equivalent to or better than the predicate devices, and that it adheres to relevant safety standards.

    FeaturePredicate NeuroPort Instrument (K060523) Performance (Implicit Acceptance Criteria)Device Under Review Performance
    Intended Use, Indications for UseRecording, processing and display of biopotential signals (EEG, ECoG, ECG, EMG, EOG, EP)Equivalent
    Typical Biopotential Signals RecordedEEG, ECoG, ECG, EMG, EOG, EPEquivalent
    Number of Signal Recording Channels96Up to 128 (cascadable to 256)
    Amplifier Input Impedance1000 Megohm1000 Megohm
    A/D Conversion16 Bit16 Bit
    Sampling RateUp to 30,000 HzUp to 30,000 Hz
    CMRR> 90 dB> 90 dB
    Analysis SoftwareEmbedded, commercially available and user defined (e.g., Persyst Insight™, NeuroPlex™, Matlab™, Spike 2™)Equivalent, same examples
    Power110 VAC110 VAC
    AlarmsNoNo
    Safety Standards ComplianceIEC 60601-1, IEC 60601-1-2, IEC 60601-2-26Same

    Note on "Reported Device Performance": The reported performance is essentially a restatement of the technical specifications, demonstrating they match or exceed the predicate devices. There is no independent "study" performance data presented in this 510(k) summary that would provide statistical measures like sensitivity, specificity, accuracy, or other metrics typically seen in diagnostic device studies.

    2. Sample size used for the test set and the data provenance

    • Not Provided. This document does not describe a "test set" in the context of a clinical study with human or animal subjects, nor does it mention data provenance (country of origin, retrospective/prospective). The assessment appears to be based on direct technical specification comparison and perhaps internal engineering testing (referred to as "system testing" and "software documentation" in Section 4.0).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable/Not Provided. Since there is no "test set" or clinical study described for ground truth establishment, this information is not present.

    4. Adjudication method for the test set

    • Not Applicable/Not Provided. No adjudication method is mentioned as there's no described test set or expert evaluation process.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable/Not Provided. This device is a signal processing system for acquiring and displaying biopotential signals. It is not an AI-assisted diagnostic tool for human readers, and therefore, an MRMC study comparing human reader performance with and without AI assistance is not relevant to this device description and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable/Not Provided. The device itself is a hardware and software system for signal acquisition and display. It does not perform automated interpretations or diagnoses requiring a standalone "algorithm only" performance evaluation in the typical sense of AI/ML devices. Its performance is defined by its ability to accurately acquire and condition the biopotential signals. The "software testing" mentioned in Section 6.0 likely refers to verification and validation of the software's functionality within the system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not Applicable/Not Provided. Given the nature of a signal processing system, "ground truth" would typically relate to the accuracy of the signal acquisition and conditioning itself, which is assessed through engineering tests against known input signals, not clinical ground truth like pathology or outcomes data. The document does not specify how the performance metrics (e.g., CMRR, impedance, sampling rate) were verified, but it would logically involve calibrated electronic test equipment.

    8. The sample size for the training set

    • Not Applicable/Not Provided. This device is not an AI/ML device that requires a "training set" in the context described.

    9. How the ground truth for the training set was established

    • Not Applicable/Not Provided. As there's no training set, this is not applicable.

    Summary of the "Study" (as described in the 510(k) Summary):

    The "study" described in this 510(k) summary is not a clinical trial or performance study against predefined clinical acceptance criteria involving human subjects or expert readers. Instead, it is a technical comparison and verification process.

    • Purpose: To demonstrate "substantial equivalence" of the Blackrock NeuroPort Biopotential Signal Processing System to legally marketed predicate devices (K042626, K060803, K040113).
    • Methodology:
      • Comparison of Intended Use/Indications for Use: Showing the new device's intended use is equivalent to or falls within the scope of predicate devices.
      • Comparison of Technological Characteristics: A detailed feature-by-feature comparison of specifications (e.g., number of channels, input impedance, A/D conversion, sampling rate, CMRR) between the new device and predicate devices.
      • Hardware Modifications Verification: Mention of verifying new interface cables and a passive splitter box through "continuity of the single pathway" and "continuity of the signal pathways."
      • Software Testing: Indication that "software testing" was performed for new functionalities like syncing with external EEG instruments and cascading two systems.
      • Standards Compliance: Affirmation of compliance with relevant safety standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26).
    • Conclusion: The modifications to the predicate NeuroPort System "raise no new questions of safety or effectiveness," indicating that the device is considered safe and effective based on its equivalence to existing cleared devices.

    In essence, this document is a regulatory submission demonstrating technical equivalence rather than a report of a scientific study with clinical performance endpoints.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060919

    Validate with FDA (Live)

    Device Name
    ACTICAL
    Manufacturer
    Date Cleared
    2006-09-22

    (171 days)

    Product Code
    Regulation Number
    882.1845
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Device:

    The subject device can be classified as physiological signal conditioner as described in 21 CFR 882.1845
    Aurora Road Twinsburg, Ohio 44087

    Re: K060919

    Trade/Device Name: Actical Regulation Number: 21 CFR 882.1845

    Device Description :

    The subject device can be classified as physiological signal conditioner as described in 21 CFR 882.1845

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Actical is a compact, lightweight, waist-, wrist-, or ankle-worn activity monitor that may be used to assess human gross motor activity, caloric expenditure, and estimates of energy expenditure based on motor activity in any instance where quantifiable analysis of physical motion is desirable.

    Device Description

    The subject device can be classified as physiological signal conditioner as described in 21 CFR 882.1845. Actical is a physiological data recorder consisting of a data recorder which can be worn on the waist (hip), wrist, or ankle of the subject and that detects, measures, and records physical Activity data. A Reader is used to transfer the recorded rata to a Personal Computer (PC) running the Actical Host software. The host software is used to configure the data recorder for data collection, to retrieve logged data from the recorder, to display data and trend graphs, and to provide a means to store the logged data in a PC for later use.

    AI/ML Overview

    The provided text is a 510(k) summary for the Actical Physiological Signal Recorder. This document focuses on demonstrating substantial equivalence to predicate devices and does not contain detailed information about specific acceptance criteria, performance studies, or the methodologies used to establish ground truth.

    Therefore, I cannot extract the requested information as it is not present in the provided text. The document primarily covers:

    • Contact person and date
    • Trade name, common name, classification, and product code
    • Legally marketed substantially equivalent predicate devices
    • Description of the device
    • Intended Use
    • Comparison of technical characteristics, stating that "Results of performance tests, risk analysis, and verification and validation testing demonstrate that the devices are substantially equivalent." However, it does not detail these tests, criteria, or results.
    • FDA's letter of clearance, confirming substantial equivalence.

    Without a detailed study report, specific acceptance criteria, sample sizes, expert qualifications, and ground truth methodologies cannot be described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K061870

    Validate with FDA (Live)

    Device Name
    VITALSENSE XHR
    Manufacturer
    Date Cleared
    2006-08-31

    (59 days)

    Product Code
    Regulation Number
    882.1845
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Zita.Yurko@Respironics.com |
    | Classification Reference | 21 CFR 882.1845
    Device:

    The subject device can be classified as physiological signal conditioner as described in 21 CFR 882.1845
    Murrysville, Pennsylvania 15668

    Re: K061870

    Trade/Device Name: VitalSense XHR Regulation Number: 21 CFR 882.1845

    Device Description :

    The subject device can be classified as physiological signal conditioner as described in 21 CFR 882.1845

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VitalSense XHR can be used in any setting where physiological body core temperature, skin temperature, or heart rate are used to further the understanding of body function or where quantifiable analysis of temperature or heart rate data is desirable.

    Device Description

    The subject device can be classified as physiological signal conditioner as described in 21 CFR 882.1845. VitalSense XHR is a physiological data recorder that is worn on the body. Up to ten sensor inputs may be used with a single VitalSense XHR recorder. The device senses and records physiological data and displays the data on its LCD screen. Recorded data may be transferred later to a PC for display and conversion for export to other programs.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the device:

    Summary of Device Performance Study

    The provided documentation for the Respironics VitalSense XHR (K061870) describes a submission for a modified device based on a predicate device (VitalSense K033534). The core of the study is a demonstration of substantial equivalence to this predicate device by showing that modifications (primarily to software for heart rate display) have no impact on safety and effectiveness.

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly list quantitative acceptance criteria for specific physiological parameters (e.g., accuracy ranges for temperature or heart rate) alongside reported device performance. Instead, it states:

    Acceptance CriteriaReported Device Performance
    All tests verified to meet required acceptance criteria as a result of risk analysis and product requirements."Results of performance tests, risk analysis, and verification and validation testing demonstrate that the devices are substantially equivalent." "Respironics has determined that the modifications have no impact on the safety and effectiveness of the device."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not specified. The document refers to "performance tests, risk analysis, and verification and validation testing" but does not give a sample size for human subjects or data points used in these tests.
    • Data Provenance: Not specified. The document does not indicate the country of origin of any data or whether the data was retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not specified.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No, an MRMC comparative effectiveness study was not conducted or described. This device is a physiological data recorder, not an AI-assisted diagnostic tool for human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: The documentation implies that "performance tests, risk analysis, and verification and validation testing" were done on the device itself. Given that it's a physiological recorder, its performance fundamentally relates to its ability to accurately sense and record data in a standalone manner. However, specific details of these "performance tests" are not provided. The focus is on the substantial equivalence of the modified software within the existing hardware.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: Not explicitly stated. The document implies that the device's measurements (body core temperature, skin temperature, heart rate) are considered the "truth" for the purpose of demonstrating substantial equivalence to the predicate device. It's likely that established measurement standards or comparison to other validated devices would serve as the ground truth during performance testing, though this is not detailed.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. This device is a physiological recorder, not an AI/ML device that requires a training set in the typical sense. The software modification described (allowing heart rate display) would not involve a machine learning training set.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable, as no training set for AI/ML is described.

    In summary:

    The submission for the VitalSense XHR focuses on demonstrating substantial equivalence to an existing predicate device after a software modification. The provided text outlines a regulatory submission process rather than a detailed scientific study with quantitative performance metrics, test set descriptions, or expert involvement in establishing ground truth for a diagnostic algorithm. The "acceptance criteria" are broad statements about meeting regulatory requirements and ensuring the modifications have no impact on safety and effectiveness, rather than specific numerical targets for physiological measurement accuracy.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052489

    Validate with FDA (Live)

    Device Name
    ACTIHEART
    Manufacturer
    Date Cleared
    2005-09-27

    (15 days)

    Product Code
    Regulation Number
    882.1845
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Classification

    Predicate Devices

    Physiological Signal Recorder Device Classification Name

    Neurology

    21 CFR 882.1845
    Boxborough, Massachusetts 01719

    Re: K052489 Trade/Device Name: Actiheart Regulation Number: 21 CFR 882.1845

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Actiheart is an ambulatory Heart Rate and Activity Recorder. Actiheart may be used to quantifiably measure Heart Rate, Activity, and estimates of Caloric Expenditure. This device is not intended for use as an ECG monitor.

    Device Description

    Actiheart is a compact, ambulatory, physiological Heart Rate and Activity data recorder. Actiheart may be attached to the chest surface through the use of standard ECG electrodes. Recorded data may be transferred later to a PC for display and conversion for export to other programs.

    AI/ML Overview

    The document provides details of the Actiheart device, an ambulatory Heart Rate and Activity Recorder. It includes information on its substantial equivalence to a predicate device, indications for use, device description, and performance testing results against acceptance criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    Figure of MeritRequirementActiheart Performance
    Tall T-wave Rejection CapabilityReject Tall T-Waves up to 1.2 mV with ±10% accuracy or ±5 BPM (greater of two)Sensitivity is better than 98%Specificity is 100%
    MIT-BIH Normal Sinus Rhythm waveform libraryDetect HR with accuracy of ±10% or ±5 BPM (greater of two)Sensitivity is better than 99%Specificity is better than 98%
    European ST-T waveform libraryDetect HR with accuracy of ±10% or ±5 BPM (greater of two)Sensitivity is better than 99%Specificity is better than 97%
    AAMI Normal Sinus Rhythm waveform libraryDetect HR with accuracy of ±10% or ±5 BPM (greater of two)Sensitivity is better than 99%Specificity is better than 99%
    Line frequency tolerance and drift toleranceDetect HR with accuracy of ±10% or ±5 BPM (greater of two) when drift or line noise presentNo difference between HR measurement for waveforms with and without superimposed noise
    Range and accuracy of HR meterDetect HR with accuracy of ±10% or ±5 BPM (greater of two) over specified range of 32 to 255 BPMWithin limits for all values of HR
    Amplitude of QRS waveformDetect HR with accuracy of ±10% or ±5 BPM (greater of two) for amplitudes 0.5 mV to 5 mVWithin limits for all values of amplitude between 0.45 mV and 9 mV.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size of unique patients or recordings for the test sets. Instead, it refers to "libraries of ECG waveforms" such as the MIT-BIH Normal Sinus Rhythm waveform library, European ST-T waveform library, and AAMI Normal Sinus Rhythm waveform library. These are standardized, publicly available databases of ECG recordings, often used for testing algorithms and devices. The provenance of these publicly available libraries is generally known, but not detailed in this document. The testing described is a "bench test comparison report" and relies on "library recordings of known Heart Rate." This suggests a retrospective analysis using pre-recorded data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number or qualifications of experts used to establish the ground truth for the test set. Ground truth for these waveform libraries is typically established through a combination of expert cardiological review and often includes reference annotations that have been meticulously validated.

    4. Adjudication Method for the Test Set

    The document does not specify an adjudication method for the test set. Given that it relies on established ECG waveform libraries, the ground truth (e.g., true heart rate, QRS complex locations) from these libraries would serve as the reference, precluding the need for real-time expert adjudication of the device's output.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No multi-reader multi-case (MRMC) comparative effectiveness study is mentioned. The study focuses purely on the standalone performance of the Actiheart device against established benchmarks and in comparison to a predicate device in a bench test setting. It does not evaluate human readers' improvement with or without AI assistance.

    6. Standalone Performance

    Yes, a standalone performance study was conducted. The "Actiheart to Mini Logger Series 2000 bench test comparison report" directly assesses "the Heart Rate detection performance characteristics of both devices side-by-side against a set of library recordings of known Heart Rate." This is further supported by "Additional testing with the use of simulated recordings per ASTM EC13:2002," and the detailed "Assessment of non-clinical performance data" which lists various tests applied to the "submitted Recorder" (Actiheart) against specific performance requirements. These tests evaluate the device's ability to detect and measure heart rate independently.

    7. Type of Ground Truth Used

    The ground truth used for the performance testing cited in Table 10D is based on "known Heart Rate" from "libraries of ECG waveforms" such as MIT-BIH, European ST-T, and AAMI Normal Sinus Rhythm waveform libraries. These libraries contain reference annotations (e.g., R-peak locations, true heart rates) typically established through expert consensus and automated validation. The document also mentions testing with "simulated recordings per ASTM EC13:2002," implying a ground truth derived from controlled, simulated ECG signals.

    8. Sample Size for the Training Set

    The document does not provide information on the sample size used for the training set for any algorithms within the Actiheart device. It mentions "proprietary code transforms digitized signal and detects location of features in ECG complex" and "proprietary algorithms to detect motion" and "proprietary algorithm" for caloric expenditure, but does not detail the development or training of these algorithms.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set (if any was used for the proprietary algorithms) was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K971740

    Validate with FDA (Live)

    Device Name
    AMPLAID MK12
    Manufacturer
    Date Cleared
    1998-03-16

    (308 days)

    Product Code
    Regulation Number
    882.1870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    The electrodiagnostic system is a combination of devices which are Class II per regulations 882.1835, 882.1845

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Amplaid MK12 is a multi-channel system for electrodiagnosis capable of performing multisensory evoked potential tests including:

    • Electronystagmography
    • Visual evoked potentials,
    • Auditory evoked potentials, and
    • Electrical (somatosensory) evoked potentials.
    Device Description

    The Amplaid MK12 Multi-channel system for electrodiagnosis consists of a two or four channel signal acquisition system coupled to a microcomputer which can perform signal averaging, storage and display, along with microcomputer controlled multiple mode evoked potential stimulators: auditory, visual, and electrical. A thermal printer is built in, and laser printer connection is supported.

    AI/ML Overview

    The provided text describes the Amplaid MK12, an evoked potential electrodiagnostic system, and its substantial equivalence to a predicate device, the Amplaid MK15. However, the document does not contain details about specific acceptance criteria, a dedicated study proving device performance against those criteria, or the methodology typically used for such studies in the context of AI/ML or a new, novel medical device.

    The submission is a 510(k) for a device that is stated to be "similar in design and function" to a previously cleared device (K861014). The primary focus of the 510(k) summary is to demonstrate substantial equivalence, rather than to present a de novo performance study against explicit acceptance criteria.

    Therefore, many sections of your request cannot be fulfilled by the provided text.

    Here is an attempt to address your request based on the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria for performance metrics (e.g., sensitivity, specificity, accuracy). Instead, it relies on demonstrating that the new device is "as safe and effective as the predicate device" through a comparison of characteristics.

    CharacteristicAcceptance Criteria (Implied)Reported Device Performance (Amplaid MK12)
    Intended UseSame as predicate deviceElectronystagmography, Visual Evoked Potentials, Auditory Evoked Potentials, Electrical (Somatosensory) Evoked Potentials (Same as MK15)
    Physical Characteristics (Size/Weight)Comparable to predicate device21.25"H x 19.7"D x 7.9"H, 28.6 lbs (Different from MK15, but acceptable)
    Energy SourceSame as predicate device115/230 Vac, ± 10%, 50-60 Hz (Same as MK15)
    Hardcopy OutputEquivalent or improved functionality to predicate deviceBuilt-in via 640 point thermal printer or laser printer connection (Improved/added functionality compared to MK15's computer interface output)
    Electrical SafetyCompliance with UL-544, IEC 601UL-544, IEC 601 (Same as MK15)
    Safety and EffectivenessAs safe and effective as predicate deviceReported as "as safe and effective as the predicate devices" based on "bench and user testing."
    Technological DifferencesFew and do not raise new questions of safety or effectivenessHas "few technological differences" compared to the predicate.
    New Indications for UseNoneNo new indications for use.

    2. Sample size used for the test set and the data provenance

    The document mentions "bench and user testing" but does not provide details on:

    • The specific sample size (number of cases, subjects, or data points) used for this testing.
    • The data provenance (e.g., country of origin, retrospective or prospective nature).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. Given that the testing mainly aimed to establish substantial equivalence based on safety and effectiveness comparable to a predicate, it's unlikely a formal "ground truth" establishment by multiple experts, as understood in an AI/ML context, was performed or detailed.

    4. Adjudication method for the test set

    This information is not provided in the document.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study is not mentioned. The device is an evoked potential electrodiagnostic system and does not appear to be an AI-assisted diagnostic tool for human readers in the way an AI-powered image analysis system would be.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This concept is not applicable to the device described. The Amplaid MK12 is a multi-channel system for electrodiagnosis, implying human operation and interpretation of the evoked potentials. It's not a standalone algorithm in the sense of AI/ML.

    7. The type of ground truth used

    The document does not specify the type of ground truth used. For an evoked potential system, "ground truth" would generally relate to the accuracy, reliability, and precision of the physiological signal acquisition and processing compared to established medical benchmarks or the performance of the predicate device.

    8. The sample size for the training set

    The concept of a "training set" in the context of machine learning is not applicable here, as the device is not described as using AI/ML algorithms that require training data.

    9. How the ground truth for the training set was established

    Not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K972959

    Validate with FDA (Live)

    Device Name
    AMPLAID MK22
    Manufacturer
    Date Cleared
    1998-02-06

    (179 days)

    Product Code
    Regulation Number
    882.1870
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    The electrodiagnostic system is a combination of devices which are Class II per regulations 882.1835, 882.1845

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Amplaid MK22 is a multi-channel system for electrodiagnosis capable of performing multisensory evoked potential tests including:
    Electronystagmography
    Visual evoked potentials,
    Auditory evoked potentials, and
    Electrical (somatosensory) evoked potentials.

    Device Description

    The Amplaid MK22 Multi-channel system for electrodiagnosis consists of a two channel signal acquisition system coupled to a microcomputer which can perform signal averaging, storage and display, along with microcomputer controlled multiple mode evoked potential stimulators: auditory, visual, and electrical. A thermal printer is built in, and laser printer connection is supported.

    AI/ML Overview

    The Amplaid MK22 is an evoked potential electrodiagnostic system. The submission establishes substantial equivalence to the Amplaid MK15 (K861014).

    1. Table of Acceptance Criteria and Reported Device Performance

    For this 510(k) submission, the acceptance criteria are based on substantial equivalence to the predicate device, Amplaid MK15. This means the new device must be as safe and effective as the predicate device and have the same intended uses, with no new indications for use. Performance is implicitly assessed by demonstrating that the key characteristics of the new device are either identical or comparable to the predicate device and that any differences do not raise new questions of safety or effectiveness.

    CharacteristicAcceptance Criteria (Predicate Device: Amplaid MK15)Reported Device Performance (New Device: Amplaid MK22)
    Intended Uses:Electronystagmography, Electroneurography, Visual evoked potentials, Auditory evoked potentials, Electrical (somatosensory) evoked potentials(Same) Electronystagmography, Electroneurography, Visual evoked potentials, Auditory evoked potentials, Electrical (somatosensory) evoked potentials
    Physical characteristics:
    Number of channels2 or 41 or 2 (Difference noted, but acceptable as it's a reduction in channels, not an increase in complexity or new functionality requiring more channels than available)
    Size/weight20.8" W x 18.9" D x 13.4" H, 44 lbs.10" W x 14" D x 14" H, 22.2 lbs (Smaller and lighter, considered an improvement or neutral difference)
    Energy Source:115/230 Vac, ± 10%, 50-60 Hz95-250 VAC, 50-60 Hz 100 VA (Wider voltage range, acceptable as it signifies wider compatibility without compromising safety)
    DisplayCRT monochrome (green), 9"CRT, monochrome (green), 9", 640 x 400 (Same)
    Hardcopy Output:Built in via 640 point thermal printer or laser printer connection(Same) Built in via 640 point thermal printer or laser printer connection
    Standards and Safety characteristics:
    Electrical safety:UL-544, IEC 601, Type BF(Same) UL-544, IEC 601, Type BF
    Conclusion:The predicate device is safe and effective, with established performance for its intended uses.The new device is as safe and effective as the predicate device, has fewer channels (which is not a new safety/effectiveness concern), and has no new indications for use, thus rendering it substantially equivalent.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly mention a "test set" in the context of a clinical performance study with human subjects or a defined set of data for algorithm evaluation. The assessment for this 510(k) is based on bench and user testing and a comparative analysis of specifications against the predicate device.

    • Sample Size: Not applicable in the context of a traditional performance study. The evaluation focuses on the device hardware and software specifications, and its equivalence to a previously cleared device.
    • Data Provenance: Not specified as there isn't a clinical data set being referenced for performance evaluation. The "bench and user testing" information is general and does not refer to specific patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. This 510(k) submission is for an electrodiagnostic system, and its clearance is based on substantial equivalence to a predicate device, not on diagnostic accuracy against a ground truth established by experts.

    4. Adjudication Method for the Test Set

    Not applicable. There is no specific "test set" and corresponding adjudication process described for evaluating device performance in this submission.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No. An MRMC comparative effectiveness study was not performed or referenced in this 510(k) submission. The device is an electrodiagnostic system and the clearance pathway relies on demonstrating equivalent technical specifications and intended uses to a predicate device, not on improving human reader performance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. The Amplaid MK22 is a medical device (an electrodiagnostic system) that acquires and processes physiological signals for medical diagnosis by a clinician. It is not an AI algorithm that produces a diagnostic output on its own. While it involves signal averaging, storage, and display, this is part of standard electrodiagnostic equipment functionality, not an AI algorithm performing diagnostic interpretation without human input.

    7. The Type of Ground Truth Used

    Not applicable. The submission does not refer to a ground truth for diagnostic accuracy, as it focuses on demonstrating the device's technical specifications and intended uses are substantially equivalent to a predicate.

    8. The Sample Size for the Training Set

    Not applicable. The Amplaid MK22 is a hardware and software system for electrodiagnosis; it is not an AI algorithm that undergoes a "training" phase with a dataset.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1