Search Results
Found 35 results
510(k) Data Aggregation
(262 days)
Trade/Device Name: SnugLit(TM) Wearable Phototherapy System (SNGL-01-US)
Regulation Number: 21 CFR 880.5700
Neonatal phototherapy unit
Regulation Name: Neonatal phototherapy unit
Regulation Number: 880.5700
The SnugLit(TM) Wearable Phototherapy System is intended for use in the treatment of infant hyperbilirubinemia, commonly known as neonatal jaundice. The device can be used in a hospital or at home, by a licensed medical practitioner or by a caregiver under the supervision of a licensed medical practitioner.
The device has been designed for use with infants of weights between 1.5 – 5.5 kg (approximately 3.5 -12 lbs) and lengths between 40.5 - 57.5 cm (approximately 16 - 23 inches).
The SnugLit system is a phototherapy device that provides therapeutic blue light to infants using a Light Mat enclosed within a Disposable Swaddle. This system enables infants to receive phototherapy in a bassinet or while being held. The SnugLit is designed to deliver wrap-around phototherapy to infants while being swaddled.
The SnugLit is composed of a Light Mat, Controller, Disposable Swaddle, and Power Adapter.
N/A
Ask a specific question about this device
(211 days)
Charlotte, North Carolina 28269
Re: K243372
Trade/Device Name: BiliWrap
Regulation Number: 21 CFR 880.5700
name: BiliWrap
Classification Name: Neonatal phototherapy unit
Classification Regulation: 21 CFR §880.5700
The BiliWrap Phototherapy System is intended for use to treat infants who have been diagnosed with hyperbilirubinemia, commonly known as neonatal jaundice, which can cause a yellow discoloration of the skin and the whites of the eyes. The Phototherapy system can be used in a hospital or at home. The device is intended to be used in infants up to the age of 3 months, weighing less than 10 kg.
The BiliWrap Phototherapy System is a portable phototherapy light system that delivers a narrow band of high-intensity blue light via blue light-emitting diodes (LEDs) to provide treatment for neonatal unconjugated hyperbilirubinemia. The device is comprised of a control box and pad including a disposable pad cover and AC/DC power supply. The pad is designed to cover the torso and legs and provide phototherapy treatment on large surface area of the baby.
The BiliWrap Phototherapy System emits light in the spectrum of 450 to 475 nm, a spectrum known as effective to reduce bilirubin concertation in the body of infants. Treatment is intended to be applied until the bilirubin levels have dropped sufficiently that the child no longer suffers from jaundice. Phototherapy treatment of neonatal jaundice using the device can be applied at home or in a hospital.
The provided FDA 510(k) clearance letter for the BiliWrap Phototherapy System (K243372) clearly states: "No clinical data was required for this submission." (Page 11)
Therefore, the device's acceptance criteria and the study proving it meets these criteria are based entirely on non-clinical tests, specifically bench testing and compliance with various recognized consensus standards.
Here's a breakdown of the information requested, based on the provided document:
Acceptance Criteria and Device Performance Study for BiliWrap Phototherapy System (K243372)
The BiliWrap Phototherapy System's acceptance was established through non-clinical performance and safety testing, demonstrating its substantial equivalence to the predicate device (BiliTouch, K210289). No clinical data was required for this submission.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are primarily defined by compliance with recognized consensus standards and meeting specified performance characteristics that demonstrate substantial equivalence to the predicate device.
| Characteristic | Acceptance Criteria (Target/Standard Compliance) | Reported Device Performance (as demonstrated by testing) |
|---|---|---|
| Intended Use | Treatment of hyperbilirubinemia in infants up to 3 months, weighing <10 kg | Device supports indicated use; demonstrated through comparison to predicate and performance testing. |
| Population | Infants up to 3 months, weighing <10 kg | Device designed and tested for this population; safety and performance verified for this population. |
| Use Environment | Hospital, Home | Device demonstrated suitability for both environments through standard compliance (e.g., IEC 60601-1-11 for home healthcare). |
| Principle of Operation | Transmits blue LED light in an effective spectrum and irradiance to reduce bilirubin (similar to predicate) | Emits blue LED light (450-475 nm) in an effective spectrum and irradiance. Functionally similar despite different physical wrap design. |
| Light Source Emission Spectrum | Within AAP standard wavelength of 430-490 nm as effective for phototherapy | 450 to 475 nm (peak emission at 450 nm) - confirmed to be within AAP recommended range. |
| Light Intensity (Irradiance) | Low: 30-40 μW/cm²/nm; High: 40-50 μW/cm²/nm (meeting/exceeding recommended >30 μW/cm²/nm) | Low: 30-40 μW/cm²/nm; High: 40-50 μW/cm²/nm - confirmed to achieve recommended irradiance. |
| Light Emitting Area | Sufficient for effective treatment (larger than predicate due to body-shape design) | 1216 cm² (larger than predicate's 420 cm²), designed to provide effective phototherapy. |
| Heat Distribution | Uniform heat distribution, allowing evaporation of heat to prevent overheating | Demonstrated uniform heat distribution with design allowing heat evaporation (5 sections, arms uncovered). |
| Electrical Safety | Compliance with IEC 60601-1, IEC 60601-1-2 (EMC), IEC 60601-1-11 (Home Healthcare) | Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11. |
| Biocompatibility | Compliance with ISO 10993-1 for materials in contact with human tissues | Complies with ISO 10993-1; disposable nonwoven polypropylene cover confirmed biocompatible. |
| Photobiological Safety | Compliance with IEC 62471 | Complies with IEC 62471. |
| Software Functionality | Verification and validation in accordance with FDA guidance | Software verification and validation testing completed. |
| Durability | Device maintains functionality over intended lifespan | Durability tests completed. |
| Temperature Control | Device operates within safe temperature limits | Temperature tests completed. |
| Alarm Functionality | Alarms (visual) function as intended | Errors and alarm tests completed (note: visual alarms, no audible alarms, which was deemed acceptable). |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not applicable. No human clinical test set was used. The evaluation was based on bench testing of the device prototypes/production units. The exact number of physical units tested for each non-clinical test (e.g., durability, electrical safety) is not specified but would typically involve a statistically relevant sample as defined by internal quality systems and testing protocols.
- Data Provenance: The data is derived from retrospective (design-stage, in-house) and prospective (final product verification and validation) non-clinical laboratory testing performed by Gerium Medical, Ltd. and/or their contracted testing facilities, as per the listed industry standards. The country of origin of the data is implied to be Israel (where Gerium Medical, Ltd. is located) and potentially international testing laboratories.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. As no clinical test set was used, there was no "ground truth" derived from expert consensus on patient data. The "ground truth" for non-clinical tests is established by:
- Engineering specifications and design requirements.
- Industry consensus standards (e.g., IEC, ISO).
- Predicate device performance characteristics.
The "experts" involved would be the engineers, quality assurance personnel, and regulatory affairs specialists at Gerium Medical, Ltd. and their third-party testing partners, qualified in device design, testing, and regulatory compliance.
4. Adjudication Method for the Test Set
Not applicable. No clinical test set requiring expert adjudication was used. Non-clinical test results are compared against predefined technical acceptance criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. The submission explicitly states "No clinical data was required for this submission."
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
This question is not applicable to the BiliWrap Phototherapy System. The device is a physical medical device that emits light for phototherapy, not an AI algorithm or software-only diagnostic tool. Its "performance" is its ability to emit light at the specified wavelength and intensity, safely and reliably.
7. The Type of Ground Truth Used
For this device, the "ground truth" is based on:
- Engineering Specifications and Design Requirements: The device must meet its own design specifications (e.g., light intensity, dimensions, weight).
- Compliance with Recognized Consensus Standards: Many standards (e.g., IEC 60601 series for safety, ISO 10993 for biocompatibility) define quantifiable and verifiable criteria.
- Predicate Device Characteristics: Substantial equivalence often relies on demonstrating that the new device performs comparably to a legally marketed predicate device (BiliTouch in this case) on key performance parameters.
8. The Sample Size for the Training Set
Not applicable. As a physical medical device for phototherapy, there is no "training set" in the context of machine learning or AI models. The "training" for such a device is its design, engineering, and manufacturing process.
9. How the Ground Truth for the Training Set was Established
Not applicable. See point 8.
Ask a specific question about this device
(115 days)
Minnesota 55114
Re: K210289
Trade/Device Name: Infant Phototherapy Equipment Regulation Number: 21 CFR 880.5700
|
| Regulation Number | 21 CFR 880.5700
The BiliTouch™, model Motif Phototherapy Blanket and Infant Phototherapy Equipment, model BT-450 for use to treatment of infants diagnosed with hyperbilirubinemia, commonly known as neonatal jaundice, which can cause a yellow discoloration of the skin and the whites of the eyes. The devices can be used in a hospital or at home.
The device is designed to use for patient population described in the infant, who is age up to 3 months and weight less than 10 kg.
Infant phototherapy equipment is a portable phototherapy light system delivers a narrow band of high-intensity blue light via blue light-emitting diodes (LEDs) to provide treatment for neonatal unconjugated hyperbilirubinemia. The device is comprised of a control box and pad including a disposable pad cover, rechargeable battery and AC/DC power supply. The pad is designed to provide phototherapy treatment from underneath the baby.
Infant phototherapy equipment provides blue LEDs to achieve intensities by two-level with emitting light from blue LEDs (bandwidth 400 - 550 nm used).
Treatment is intended to be applied until the bilirubin levels have dropped sufficiently that the child no longer suffers from jaundice. Phototherapy treatment of neonatal jaundice using the device can be applied at home or in a hospital.
This document is an FDA 510(k) Premarket Notification summary for the BiliTouch™ (Motif Phototherapy Blanket) and Infant Phototherapy Equipment (BT-450). It details the device's characteristics and its substantial equivalence to a predicate device, focusing on non-clinical testing.
Here's a breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present "acceptance criteria" for an AI/algorithm's performance or "reported device performance" in the typical sense of metrics like accuracy, sensitivity, and specificity for diagnostics. Instead, it describes compliance with various safety and performance standards for a medical device (infant phototherapy equipment).
However, we can infer some "acceptance criteria" and "reported performance" from the "Summary table of technological characteristics of the device compared to the predicate device" and the "SUMMARY OF NON-CLINICAL TESTS" section.
| Feature/Test | Acceptance Criteria (Implied) | Reported Device Performance (Compliance/Result) |
|---|---|---|
| Biocompatibility | Compliance with ISO 10993-5 (Cytotoxicity) & ISO 10993-10 (Irritation/Sensitization) for components in contact with user. | Testing was conducted in accordance with FDA Recognition # 2-245, ISO 10993-5:2009, Biological Evaluation of Medical Devices-- Part 5: Tests for In Vitro Cytotoxicity and FDA Recognition # 2-174, ISO 10993-10:2010, Biological Evaluation of Medical Devices -Part 10: Tests for Irritation and Skin Sensitization. (Implied successful compliance). |
| Electrical Safety | Compliance with AAMI ANSI ES60601-1:2005/(R)2012 and A1:2012 (IEC 60601-1) & IEC 60601-1-11:2015. | Testing was conducted in accordance with FDA Recognition # 19-4, AAMI ANSI ES60601-1:2005/(R)2012 and A1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) and FDA Recognition # 19-14, IEC 60601-1-11:2015. (Implied successful compliance). |
| Electromagnetic Compatibility (EMC) | Compliance with IEC 60601-1-2:2014. | Testing was conducted in accordance with FDA Recognition # 19-8, IEC 60601-1-2:2014. (Implied successful compliance). |
| Alarm System | Compliance with IEC 60601-1-8: 2006 + AMD1:2012. | Testing was conducted in accordance with FDA Recognition # 5-76. IEC 60601-1-8: 2006 + AMD1:2012. (Implied successful compliance). |
| Infant Phototherapy Equipment Specifics | Compliance with IEC 60601-2-50: 2009/AMD1:2016. | Testing was conducted in accordance with FDA Recognition # 6-387, IEC 60601-2-50: 2009/AMD1:2016. (Implied successful compliance). |
| Software Verification and Validation | Compliance with FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (May 11, 2005). | Software verification and validation testing as recommended in FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (May 11, 2005). (Implied successful compliance). |
| Bench Performance Tests (Battery, Light Intensity, Temperature, Alarms) | To demonstrate specifications. | Bench performance tests including testing for battery-powered operating test, light intensity test and testing for temperature on the pad of the device and high temperature alarm were completed to verify that the infant phototherapy equipment demonstrate the specifications. (Implied successful verification). |
| Wavelength | 455~465 nm (matching predicate) | 455~465 nm |
| Light Output (Irradiance) | HIGH: 60±10 µW/cm²/nm; LOW: 30±10 µW/cm²/nm | HIGH: 60±10 µW/cm²/nm; LOW: 30±10 µW/cm²/nm |
| Thermal Safety (PAD Surface) | ≤ 40 °C | Surface of PAD ≤ 40 °C |
| Acoustic Energy | Not explicitly stated an acceptance range, but reported. | 43 ±1 dBA |
| IP Ratings | Control box: IP 21; Pad: IP 23; Power adapter: IP 22 | Control box: IP 21; Pad: IP 23; Power adapter: IP 22 |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document explicitly states: "No clinical data were submitted." This implies that the 'test set' for demonstrating the device's safety and effectiveness was based purely on non-clinical, bench testing and adherence to recognized standards, rather than patient data. Therefore, there is no information on sample size, country of origin, or retrospective/prospective nature of a clinical test set.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This section is Not Applicable as no clinical data or expert-adjudicated ground truth was used for assessing the device's performance in a diagnostic or AI context. The assessment was based on compliance with device performance standards rather than a clinical ground truth derived from expert interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This section is Not Applicable as no clinical data requiring expert adjudication was used.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC study was Not Done. The device is an Infant Phototherapy Equipment, which is a physical device delivering light therapy, not an AI or imaging diagnostic aid that would assist human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This section is Not Applicable as the device is not an algorithm for diagnosis or image interpretation. It is a phototherapy device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the purpose of this 510(k), where "No clinical data were submitted," the "ground truth" for the device's performance is established by its compliance with internationally recognized electrical, mechanical, biocompatibility, and phototherapy-specific safety and performance standards (e.g., IEC 60601 series, ISO 10993) and successful completion of internal bench performance tests.
8. The sample size for the training set
This section is Not Applicable. As stated, "No clinical data were submitted." The device is a physical medical device, not a software algorithm that requires a training set of data.
9. How the ground truth for the training set was established
This section is Not Applicable as there was no training set in the context of an AI/algorithm development.
Ask a specific question about this device
(272 days)
Trade/Device Name: Neonatal Phototherapy System, Model: KANGALITE Regulation Number: 21 CFR 880.5700
phototherapy unit
Device Class: II Product Code: LBI Panel: General Hospital Regulation Number: 21 CFR 880.5700
| 880.5700
| 880.5700
The Neonatal Phototherapy System, Model: KANGALITE is intended for the treatment of unconjugated hyperbilirubinemia in the population of neonates: gestational age ≥38 weeks and weight within 2500-4500 grams. It can be used in the clinical setting or in the home environment. It should be used for neonates whom phototherapy has been prescribed.
The Neonatal Phototherapy System, Model: KANGALITE is a wearable phototherapy system that consists of six components: Light Source with AC Power Cord, Dual Fiber Optic Panels (Standard accessory) /Single Fiber Optic Panel (optional accessory), Swaddle, Strap, Disposable Covers and Table Mounting Clip (optional accessory). The Neonatal Phototherapy System, Model: KANGALITE delivers a narrow band of high-intensity blue light emitting diode (LED), in order to provide phototherapy treatment for neonatal hyperbilirubinemia.
The Neonatal Phototherapy System, Model: KANGALITE Light Source contains a blue LED which emits light in the range 400-500 nm (peak wavelength 455±10 nm). The system gives only one blue light irradiance output level which is not adjustable. This blue light is directed through the optical fiber bundle to the illuminating area of the Fiber Optic Panel where the newborn receives blue light therapy.
The Fiber Optic Panel is inserted into the Disposable Cover and is placed in a soft and comfortable Swaddle, while the baby is wrapped inside to receive phototherapy. Baby is allowed to receive kangaroo care. Besides, compare to conventional phototherapy, no eye protection for the baby is required as blue light will not leak out when properly set up.
The light output (>30 µW/cm²/nm) is sufficient to provide intensive phototherapy which is defined as in the American Academy of Pediatrics (AAP) standard.
Average irradiance for:
Dual Fiber Optic Panels: 50 µW/cm2/nm (±25%);
Single Fiber Optic Panel: 60 uW/cm2/nm (±25%).
The spectrum range of the blue LED light matches with the spectral absorption of bilirubin. Under blue light irradiation, subcutaneous bilirubin absorbs strongly in the blue region of the light spectrum and converts into less lipophilic waste products which can be excreted through the bile or urine without the need for conjugation.
The Neonatal Phototherapy System, Model: KANGALITE can be directly connected to nominal voltages readily available throughout the world as the Light Source is affixed with a wide-range medical adaptor, rated from 100-240 VAC and 50/60 Hz. It converts the power to 15 VDC to power up the Light Source.
The provided text describes a 510(k) premarket notification for a medical device, the Neonatal Phototherapy System, Model: KANGALITE. This document primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed clinical study with specific acceptance criteria and performance metrics for a novel AI/device.
Therefore, many of the requested details regarding acceptance criteria, sample sizes, expert involvement, and ground truth establishment, which are typical for studies validating AI models or new clinical devices, are not explicitly provided in this 510(k) submission.
Here's a breakdown of what can be inferred or stated based on the given information:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not provide a table of explicit acceptance criteria with numerical performance targets typical for clinical efficacy studies (e.g., sensitivity, specificity, or improvement in bilirubin levels). Instead, the "acceptance criteria" for this 510(k) submission are met by demonstrating substantial equivalence to a previously cleared predicate device (Bili-Tx, K070180) through various non-clinical tests and comparisons of technological characteristics.
The "reported device performance" is largely qualitative, asserting that the device meets relevant safety and performance standards established for such phototherapy systems.
Here's what can be extracted regarding performance and comparison:
| Characteristic | Acceptance Criteria (Implicit for Substantial Equivalence to Predicate) | Reported Device Performance (KANGALITE) |
|---|---|---|
| Intended Use | Treatment of hyperbilirubinemia in neonates (Bili-Tx) | Treatment of unconjugated hyperbilirubinemia in neonates: gestational age ≥38 weeks and weight within 2500-4500 grams |
| Principle of Operation | Blue light phototherapy converting bilirubin to excretable products (Bili-Tx) | KANGALITE uses blue light LEDs (400-500 nm, peak 455±10 nm) for bilirubin conversion |
| Light Intensity (Irradiance) | Standard Panel-Light: 30 µW/cm²/nm; Neonatal Panel-Light: 55 µW/cm²/nm; Overhead Therapy: 30 µW/cm²/nm (Bili-Tx) | Dual Fiber Optic Panels: 50 µW/cm²/nm (±25%); Single Fiber Optic Panel: 60 µW/cm²/nm (±25%). Meets AAP intensive phototherapy standard (>30 µW/cm²/nm). |
| Wavelength | Blue LEDs: 400–550 nm (peak 450–470 nm) (Bili-Tx) | Blue LED: 400-500 nm (peak: 455 ± 10 nm) |
| LED Lifetime | Minimum lifetime of 30,000 hours (Bili-Tx) | Nominal lifetime: >45,000 hours |
| Biocompatibility | Not publicly available for predicate (Bili-Tx) | Passed ISO 10993 for Cytotoxicity, Sensitization, Skin Irritation for patient-contact materials (disposable covers, Swaddle). |
| Electrical Safety | Conforms to IEC 60601-1 (Bili-Tx) | Conforms to ES60601-1 and related amendments. |
| Electromagnetic Compatibility (EMC) | Conforms to EN 60601-1-2 (2nd edition) (Bili-Tx) | Conforms to IEC 60601-1-2 (4th edition) |
| Phototherapy Safety/Performance | Conforms to IEC 60601-2-50 (Bili-Tx) | Conforms to IEC 60601-2-50. |
| Home Use Safety | Intended for home/hospital use (Bili-Tx) | Conforms to IEC 60601-1-11 for home healthcare environment. |
| Water Ingress Protection (Light Source) | IPX0 (Bili-Tx) | IP22 (KANGALITE), fulfills IEC 60601-1-11 requirement for home use. |
| Noise | < 60 dB(A) (Bili-Tx) | < 60 dB(A) |
The acceptance criteria are implicitly met if the KANGALITE device demonstrates equivalent or superior performance based on these technical specifications and safety standards, and if any differences do not raise new questions of safety or effectiveness.
2. Sample Sizes Used for the Test Set and Data Provenance:
The document explicitly states: "No clinical study is included in this submission."
Therefore, there is no "test set" in the context of clinical data used for performance claims typical for AI-based devices or new therapeutic modalities. The testing described is primarily non-clinical bench testing (e.g., electrical safety, EMC, biocompatibility, light performance).
- Test Set Sample Size: N/A (no clinical test set)
- Data Provenance: Not applicable for clinical study data. The non-clinical tests were performed to established medical device standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- Number of Experts: N/A (no clinical test set requiring expert ground truth)
- Qualifications of Experts: Not applicable.
4. Adjudication Method for the Test Set:
- Adjudication Method: N/A (no clinical test set with human adjudication)
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No. This device is a phototherapy system, not an AI-assisted diagnostic tool for human readers.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done:
- Standalone Performance: Not applicable in the context of an AI algorithm. The device's performance is "standalone" in the sense that it delivers phototherapy directly, and its performance is validated through bench testing against established standards for its physical and functional characteristics.
7. The Type of Ground Truth Used:
For the non-clinical tests, the "ground truth" is defined by adherence to recognized international and national standards (e.g., ISO 10993 for biocompatibility, IEC 60601 series for electrical safety and phototherapy equipment), and the device's measured performance against the specifications deemed safe and effective for its intended use and equivalent to the predicate.
8. The Sample Size for the Training Set:
- Training Set Sample Size: N/A (no clinical data training for an AI model as this is not an AI device).
9. How the Ground Truth for the Training Set Was Established:
- Ground Truth Establishment for Training Set: N/A (Not an AI device using a training set).
Summary of the Study:
The "study" presented in this 510(k) notification is primarily a bench testing and comparative analysis rather than a clinical trial or AI model validation study. Its purpose is to demonstrate that the Neonatal Phototherapy System, Model: KANGALITE, is substantially equivalent in terms of safety and effectiveness to a legally marketed predicate device (Bili-Tx, K070180). This is achieved by showing that:
- The new device has the same intended use as the predicate device (though with a slightly more specific target population and additional contraindications explicitly stated, which are considered to not raise new safety concerns).
- The technological characteristics are sufficiently similar. Where differences exist, the submission explains why these differences do not raise new questions of safety or effectiveness (e.g., longer LED lifetime, different connection mechanism, higher IP rating for home use, larger illuminated area).
- The device conforms to relevant electrical safety, EMC, biocompatibility, and phototherapy equipment standards through detailed non-clinical bench testing.
Conclusion from the document: "Based on the performance testing, comparison, and risk analysis in this submission, the subject device Neonatal Phototherapy System, Model: KANGALITE is substantially equivalent to the predicate device K070180."
Ask a specific question about this device
(150 days)
Winchester, Massachusetts 01890
Re: K190899
Trade/Device Name: bili-hut™ Regulation Number: 21 CFR 880.5700
| LBI |
| Regulation Number: | 21 CFR 880.5700
| Manufacturer Name: | Physician Engineering Products, Inc. |
| Regulation Number: | 21 CFR 880.5700
The bill-hut™ provides phototherapy for the treatment of neonatal hyperbilirubinemia, commonly known as neonatal jaundice, during the newborn period in the clinical or home setting.
The bili-hut™ is a modular, portable blue LED-based phototherapy device intended to treat neonatal hyperbilirubinemia, commonly known as infant jaundice. The bili-hut™ consists of the assembled hut (shell with light source, base mattress, base), nest components (nest, nest liner, nest mattress, nest mattress cover), and power supply. It has an optional accessory, the perch, which is a bassinet stand insert. Features include a timer for total LED use and an Environment High Temperature Warning with automatic shutoff. The principle of operation is phototherapy using blue light to break down bilirubin. The device is powered by a 12V medical grade power supply and uses an LED light array. The materials include fabric-based, plastic, metal, and electronic components.
The bili-hut™ device is a phototherapy unit for treating neonatal hyperbilirubinemia (infant jaundice). The provided text describes the acceptance criteria and the studies performed to demonstrate substantial equivalence to a predicate device, the Ultra Bili Light™ Model 2000 (K974830).
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally established by compliance with recognized standards and performance characteristics comparable to the predicate device. The performance data is primarily non-clinical.
| Acceptance Criteria / Performance Characteristic | bili-hut™ Reported Performance | Predicate Device (Ultra Bili Light™ Model 2000) Performance | Discussion/Compliance |
|---|---|---|---|
| Intended Use | For treatment of neonatal hyperbilirubinemia in clinical or home setting. | For treatment of neonatal hyperbilirubinemia in clinical or home setting (initially Home, later clarified to include Clinical). | Similar. |
| Principle of Operation | Phototherapy with blue light to decrease bilirubin blood levels. | Phototherapy with blue light to decrease bilirubin blood levels. | Similar. |
| Energy Source | Blue light LED | Blue light fluorescent | Different technology, but both deliver therapeutic irradiance above 30 µW/cm²/nm. No new safety/effectiveness questions raised. |
| Light Source Wavelength | 430-490 nm | 430-500 nm | Similar; within AAP recommended range (430-490 nm). |
| Average Irradiance of Intended Treatment Area | 45 µW/cm²/nm | 60 µW/cm²/nm | Similar; both exceed AAP standard for high intensity phototherapy (> 30 µW/cm²/nm). |
| Output Peak Wavelength | 463 nm | Not explicitly stated, implied within 430-500 nm range | Within accepted therapeutic range. |
| Peak Irradiance | 58 µW/cm²/nm | Not explicitly stated, implied by average irradiance. | Consistent with therapeutic requirements. |
| Effective Treatment Area | 170 in² (1097 cm²) | Not explicitly stated. | Adequate for neonatal treatment. |
| Biocompatibility of Patient-Contact Materials | Passed Cytotoxicity, Sensitization (2 methods), Skin Irritation (2 methods) tests per ISO 10993 standards for polypropylene nonwovens (nest liner, nest mattress cover). | Assumed biocompatible, but no publicly available testing data for predicate. | bili-hut™ tested to current ISO 10993 standards. |
| Thermal Safety (Over-temperature shutoff) | Device material contacting newborn non-conductive, fabric-covered foam. LEDs out of reach. Over-temperature shutoff at 43°C (109.4°F) per IEC 60601-2-50. | Heating pad for cold conditions. Thermal protection circuit (beeper, flashing lights, shutoff) for over 98°F (36.7°C). | Both protect from overheating. bili-hut™ does not have a heating feature. bili-hut™ compliant with newer IEC standard, hence different cutoff temperature. No new safety/effectiveness questions raised. |
| Visible Light Radiation Safety | Requires separate eye protection (user-selected protective eye covers or goggles). | Requires BabyFace Shield or eye patches to protect eyes. | Similar; both require external eye protection. |
| Ultraviolet Light Radiation Safety | LED light source emits no UV light. | UV-blocking plastic sheet covers fluorescent bulbs. | Different. bili-hut™ inherently emits no UV; predicate filters it. No new safety/effectiveness questions raised. |
| Electrical Safety Testing | Conformance to AAMI ES60601-1:2005 + A1, IEC 60601-1-6 Ed 3.1 2013, IEC 62366-1 Ed 1.0 2015, IEC 60601-1-11 Ed 2.0 2015, IEC 60601-2-50 Ed 2.1 2016. | No publicly available information for predicate to current standards. | bili-hut™ tested to current, FDA-recognized standards. |
| Electromagnetic Compatibility (EMC) Testing | Conformance to IEC 60601-1-2 Edition 4.0 2014. | No publicly available information for predicate to current standards. | bili-hut™ tested to current, FDA-recognized standards. |
| Usability/Human Factors | Conformance to IEC 60601-1-6 and IEC 62366-1. | Not explicitly stated for predicate device, implied by general design. | bili-hut™ tested to current standards. |
| Device Power, Indicators, LED Performance, System Irradiance, System Light Intensity, Accelerated Lifetime | Bench performance tests completed to verify requirements met. | Implied functional for predicate. | bili-hut™ verified to meet device requirements and specifications. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document primarily discusses non-clinical bench testing.
- Biocompatibility testing: The "sample size" here refers to the materials tested (disposable covers and nest mattress). The studies were likely performed in a laboratory setting by certified testing facilities. The provenance would be the testing lab.
- Electrical Safety, EMC, and Bench Performance tests: These are performed on the device itself. The "sample size" is the device(s) tested. These are laboratory tests, not human data.
There is no mention of a clinical test set or human data used to evaluate the device's performance in treating hyperbilirubinemia.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable as there was no clinical test set for which ground truth would be established by experts. The performance evaluation was primarily through non-clinical bench testing against established engineering and medical device standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable as no clinical test set requiring expert adjudication was used.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The bili-hut™ is a phototherapy device, not an AI-assisted diagnostic or interpretative system requiring human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. The bili-hut™ is a medical device, not an algorithm. Its performance as a standalone phototherapy unit was assessed via bench testing against engineering and medical device standards, and comparison to a predicate device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical and bench performance tests, the "ground truth" or reference points were:
- Recognized international and FDA-recognized standards: (e.g., AAMI ES60601-1, IEC 60601-1-x series, ISO 10993 series).
- American Academy of Pediatrics (AAP) standards: Specifically for therapeutic irradiance (irradiance > 30 µW/cm²/nm) and peak wavelength (430-490 nm).
- Predicate device characteristics: The Ultra Bili Light™ Model 2000 (K974830) served as a benchmark for comparison to demonstrate substantial equivalence.
8. The sample size for the training set
Not applicable as this is a medical device, not an AI/machine learning algorithm requiring a training set.
9. How the ground truth for the training set was established
Not applicable.
Ask a specific question about this device
(122 days)
Re: K182178
Trade/Device Name: neoBLUE® blanket LED Phototherapy System Regulation Number: 21 CFR 880.5700
Phototherapy |
| Regulation Number: | 21 CFR 880.5700
The neoBLUE blanket LED Phototherapy System is indicated for the treatment of unconjugated hyperbilirubinemia in a hospital environment, and administered by trained professional medical staff, on the order of a physician, or in the home environment administered by a trained caregiver. The neoBlue blanket device provides intensive phototherapy underneath the patient and can be used with a bassinet, open bed, radiant warmer, incubator, or while holding the patient.
The neoBLUE blanket LED Phototherapy System is a portable phototherapy system that consists of five components: the neoBLUE blanket phototherapy light box, fiber optic blanket with attached cable, blanket mattress, disposable mattress covers, and power supply. The neoBLUE blanket LED Phototherapy System delivers a narrow band of high-intensity blue light via a blue light emitting diode (LED), in order to provide treatment for neonatal hyperbilirubinemia.
The portable neoBLUE blanket light box contains a blue LED which emits light in the range of 400 – 550 nm (peak wavelength 450-475 nm). This blue light is directed through an optical fiber cable to the fiber optic blanket pad. The fiber optic blanket pad is composed of high performance optical fibers enclosed in a vinyl material. A polyurethane mattress is placed on top of the fiber optic blanket, and a disposable polypropylene cover is placed around the blanket pad and mattress. The fiber optic blanket generates sufficient light output to provide intensive phototherapy which is defined as ≥30µW/cm²/nm.
The provided text describes a 510(k) premarket notification for a medical device called the "neoBLUE® blanket LED Phototherapy System." This document primarily focuses on demonstrating substantial equivalence to a predicate device rather than detailing specific clinical or standalone performance studies with acceptance criteria for an AI/ML device.
Therefore, many of the requested details regarding acceptance criteria, study design for AI/ML performance, sample sizes for training/test sets, expert involvement for ground truth, and MRMC studies are not present in this document. The device in question is a phototherapy system, not an AI/ML diagnostic or therapeutic device.
Here's an analysis based on the available information and an explanation of why other requested information is not applicable:
1. A table of acceptance criteria and the reported device performance:
The document describes performance specifications that the device meets, rather than acceptance criteria for an AI/ML model's output. The criteria are related to the physical and functional characteristics of the phototherapy system.
| Acceptance Criteria (Performance Specification) | Reported Device Performance |
|---|---|
| Light Spectrum Range | 400 – 550 nm (peak wavelength 450-475 nm) |
| Light Intensity (Intensive Phototherapy definition) | ≥30 µW/cm²/nm (for both blanket sizes) |
| Factory Set Intensity | 35 ± 5 µW/cm²/nm |
| Adjustable Light Output Range | 50 to 60 µW/cm²/nm (adjustable) |
| Expected LED Life | >20,000 hours |
| Operating Temperature (Light Box) | 41° to 86° F (5 to 30°C) |
| Operating Humidity (Light Box) | 10% to 90%, non-condensing |
| Operating Temperature (Blanket) | 41° to 100° F (5 to 38°C) |
| Operating Humidity (Blanket) | 10% to 90%, non-condensing |
| Storage Temperature | 32° to 122° F (0 to 50°C) |
| Storage Humidity | 10% to 90%, non-condensing |
| Altitude / Atmospheric pressure | 700 hPa to 1060 hPa |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not applicable as the document explicitly states "Clinical Tests: N/A" and focuses on non-clinical engineering and performance testing of the physical medical device (phototherapy unit), not a software or AI/ML-based diagnostic system that would require a test set of data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable. Ground truth, expert consensus, and qualifications of experts are relevant for AI/ML models that interpret medical images or data. The neoBLUE blanket LED Phototherapy System is a physical device delivering light therapy, not an interpretive AI.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable for the same reasons as point 3.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. An MRMC study is relevant for evaluating the impact of an AI system on human reader performance, typically in diagnostic imaging. The neoBLUE blanket LED Phototherapy System is a therapeutic device, not an diagnostic AI assistant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable as there is no algorithm/AI in this medical device. Its performance is based on its physical specifications and light output.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
This information is not applicable. For this physical device, "ground truth" relates to measurable physical properties and operational performance tested against engineering specifications and recognized standards (e.g., light intensity measurements, electrical safety).
8. The sample size for the training set
This information is not applicable. There is no "training set" as this is not an AI/ML device.
9. How the ground truth for the training set was established
This information is not applicable for the same reason as point 8.
Summary of the Study Proving Device Meets Acceptance Criteria (Non-Clinical):
The study proving the device meets its performance specifications (which serve as "acceptance criteria" for a physical device) was based on non-clinical tests and design verification and validation.
- Type of Study: Non-clinical tests, design verification and validation.
- Purpose: To ensure the device meets performance specifications and demonstrates equivalence to the predicate device (K103589).
- Tests Performed: Specific testing for device power, electrical safety, indicators, LED performance, intensity levels, intensity range, effective area, and usability.
- Standards Applied: Where appropriate, testing was performed to recognized international and industry standards, including:
- IEC 60601-2-50:2009 +A1 / ED 2.1: 2016-04 (Particular Requirements For The Basic Safety And Essential Performance Of Infant Phototherapy Equipment)
- IEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012 (General Requirements For Basic Safety And Essential Performance)
- IEC 60601-1-11:2015 (Second Edition) (Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment)
- IEC 60601-1-6:2010, AMD1:2013 (Usability)
- IEC 60601-1-2 ED 4.0: 2014-02 (Electromagnetic disturbances)
- An AIM Standard 7351731 Rev 2.0: 2017-02-03 (Electromagnetic disturbances)
- Conclusion: The verification and validation summary and risk analysis documentation supported the conclusion that the device is as safe and effective as the predicate device and is substantially equivalent.
No human data, expert review of images, or AI/ML model performance evaluation was part of this submission, as it is for a physical phototherapy device.
Ask a specific question about this device
(238 days)
The BiliLux phototherapy light is designed to reduce the serum bilirubin concentration in premature babies and neonates with unconjugated hyperbilirubinemia by means of phototherapeutic radiation. It is intended for use with patients up to 3 months of age who weigh less than 10 kg (22 lb).
The device BiliLux is a configurable LED phototherapy light designed to reduce serum bilirubin concentration in premature babies and neonates with unconjuqated hyperbilirubinemia by means of phototherapeutic radiation. It is intended for use with patients up to 3 months of age who weigh less than 10 kg (22 lb).
The phototherapy light houses the 20 blue lights that provide therapy as well as 2 white lights used for observation. The blue light of the phototherapy light lies within the wavelength range of 460 nm to 490 nm. The front panel has a display screen and keys to control functions and change settings. There are five intensity settings (20%, 60% 80% and 100% of nominal intensity). On the user interface, the user can see session time and any software errors that may occur.
The BiliLux phototherapy light can be placed directly on an incubator hood or be combined with a spring arm and trolley. An optional radiometer is also available to measure the phototherapeutic effective irradiance produced by the phototherapy light.
The provided document is a 510(k) premarket notification for the BiliLux® Phototherapy Light. It describes the device, its intended use, and a comparison to a predicate device (NanoBlu™ 500).
However, the document explicitly states:
- "No clinical testing was performed."
- The discussion focuses on "Non-Clinical Studies," specifically bench testing and adherence to standards.
This means that there are no acceptance criteria, performance data, or studies related to clinical effectiveness or comparison with human readers as typically found in AI/ML device submissions. The document focuses on demonstrating substantial equivalence to a predicate device through technical specifications and compliance with standards.
Therefore, I cannot provide the requested information for acceptance criteria and a study proving device performance in the context of an AI/ML device, as this device is a phototherapy light and was not evaluated based on AI/ML performance.
If this was a misunderstanding and you were looking for information about a different type of device or performance evaluation, please provide a different document.
Ask a specific question about this device
(241 days)
Bonita Springs, Florida 34134-2958
Re: K170585
Trade/Device Name: Skylife™ Regulation Number: 21 CFR 880.5700
| Classification Name: | Neonatal phototherapy |
| Regulation: | (21 CFR 880.5700
Skylife™ is intended for the treatment of neonatal unconjugated hyperbilirubinemia. It is designed to provide phototherapy treatment from underneath the baby. Skylife™ must be used within a patient enclosure, such as a bassinet, an open crib, a warming table or an incubator. Skylife™ can be used in a clinical setting or in the home.
Skylife™ is a portable phototherapy light system that delivers a narrow band of high-intensity blue light via blue light emitting diodes (LEDs) to provide treatment for neonatal unconjugated hyperbilirubinemia. Skylife™ is designed to provide phototherapy treatment from underneath the baby. The system must be used within a patient enclosure, such as a bassinet, an open crib, a warming table or an incubator. Skylife™ utilizes blue LEDs to achieve intensities from 25 uW/cm²/nm to >55 uW/cm²/nm, emitting light in a narrow bandwidth between 430-475 nm. This light bandwidth corresponds to the spectral absorption of light by bilirubin. Skylife™ utilizes blue LEDs in this range to achieve peak intensities between 25 to 35, and over 55 uw/cm²/nm at low, high settings. Skylife™ at the Very High setting provides an average intensity of 56.3 uw/cm/nm over the treatment area with peak at 72.4 uw /cm-/nm. Treatment is intended to be applied until the bilirubin levels have dropped sufficiently that the child no longer suffers from jaundice. Phototherapy treatment of neonatal jaundice using Skylife™ can be applied at home or in a hospital. Skylife™ consists of the following components: Light bed and light modules, GelMat, disposable mattress cover (Cloud Swaddle), controller, and power supply.
The NeoLight Skylife™ phototherapy unit for neonatal unconjugated hyperbilirubinemia underwent performance testing to demonstrate its substantial equivalence to predicate devices. The studies focused on electrical safety, electromagnetic compatibility, biocompatibility, and human factors usability.
Here's a breakdown of the acceptance criteria and study details:
1. Table of Acceptance Criteria and Reported Device Performance
| Test Category | Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|---|
| Electrical Safety | IEC 60601-1: 2005, 3rd Edition (Medical Electrical Equipments for basic safety and essential performance) | Device met the standard. |
| IEC 60601-1-2: 2014, 4th Edition (Electromagnetic compatibility - Requirements and tests) | Device met the standard. | |
| IEC 60601-2-50: 2009, 2nd Edition (Particular requirements for the safety of infant phototherapy equipment) | Device met the standard. | |
| IEC 60601-1-11: 2015, 2nd Edition (Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment) | Device met the standard. | |
| IEC 60601-1-8: 2006 (General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems) | Device met the standard. | |
| Biocompatibility | ISO 10993-1 (Surface Contact, Skin, Prolonged Contact (> 24 hours, < 30 days)) | Materials were found to be non-cytotoxic, non-sensitizers, and non-irritants. |
| ISO 10993-5:2009 (Tests for in vitro cytotoxicity) | Materials were found to be non-cytotoxic. | |
| ISO 10993-10:2010 (Tests for irritation and skin sensitization) | Materials were found to be non-sensitizers and non-irritants. | |
| Human Factors Usability | IEC 60601-1-6: 2010, 3rd Edition (General requirements for basic safety and essential performance Collateral standard: Usability) | The study supported the instructions for successfully using the device as intended. |
| IEC 62366:2007, 1st Edition (Medical devices -- Application of usability engineering to medical devices) | The study supported the instructions for successfully using the device as intended and substantiated the acceptability of identified risks. | |
| Light Range Intensity (Device Specification) | 25-55 μW/cm²/nm (Low and High settings); >55 μW/cm²/nm (Very High setting, avg 56.3 μW/cm²/nm, max 72.4 μW/cm²/nm) | This is a specification, not an acceptance criterion from a performance test against a standard in this section. However, the document states this range is comparable or higher than predicates and does not raise new questions of safety or effectiveness. |
| Wavelength (Device Specification) | 430 – 475 nm (453 - 460 nm typical peak) | This is a specification. Wavelength range and peak wavelengths overlap with predicate devices, thus not raising new safety/efficacy questions. |
2. Sample Sizes Used for the Test Set and Data Provenance
The provided document refers to various performance tests against international standards (IEC and ISO). These types of tests typically involve specific sample sizes prescribed by the standards for evaluating materials, electrical components, and hardware. However, the exact numerical sample sizes for each specific test (e.g., number of units tested for electrical safety, number of material samples for biocompatibility) are not explicitly stated in this summary.
For the Human Factors Usability study:
- Number of participants: 2 user groups
- Data provenance: Not explicitly stated, but based on the nature of human factors usability, it would be prospective data collection (users interacting with the device).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
For the technical performance tests (electrical safety, EMC, biocompatibility), the "ground truth" is established by adherence to the specified international standards. The evaluation is typically performed by trained engineers and technicians in accredited testing facilities. The document does not specify the number or qualifications of experts explicitly.
For the Human Factors Usability study:
- The document implies that the study was conducted by human factors experts. However, the number of experts used to establish "ground truth" (e.g., identifying usability issues or defining successful use) and their specific qualifications are not provided. The "ground truth" for usability is generally derived from user performance and expert observation against established usability principles.
4. Adjudication Method for the Test Set
The document does not describe specific adjudication methods for the performance tests in the context of clinical endpoints or expert review. For objective technical tests, the results either meet the standard's criteria or they don't, often without a need for "adjudication" in the sense of resolving inter-rater disagreement.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance
No, the provided document does not describe an MRMC comparative effectiveness study, nor does it mention AI in any context. This device is a phototherapy unit, not an AI-powered diagnostic or assistive tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No, as mentioned, this device is a phototherapy unit and does not involve an algorithm or AI for standalone performance evaluation in that manner.
7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)
- For Electrical Safety, EMC, and Safety of Infant Phototherapy Equipment: The ground truth is established by the requirements and methodologies defined within the respective IEC standards. Compliance with these standards indicates safety and essential performance.
- For Biocompatibility: The ground truth is established by the requirements and testing protocols defined within the ISO 10993 series of standards. Non-cytotoxicity, non-sensitization, and non-irritation are the "ground truth" criteria.
- For Human Factors Usability: The ground truth is established by the requirements of IEC 60601-1-6 and IEC 62366, which focus on preventing use errors and ensuring that the device can be used safely and effectively as intended. User performance during the study, observed errors, and successful task completion against defined criteria form the basis of the ground truth.
8. The Sample Size for the Training Set
This information is not applicable as the device is a phototherapy unit and does not rely on a "training set" in the context of machine learning.
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for the same reason as above.
Ask a specific question about this device
(294 days)
69100 FRANCE
Re: K163526
Trade/Device Name: Bilicocoon™ Phototherapy System Regulation Number: 21 CFR 880.5700
|
| Regulation Name:Regulation Number:Product Code: | Unit, Neonatal Phototherapy21 CFR 880.5700
The BiliCocoon™ Phototherapy System is intended for the treatment of unconjugated hyperbilirubinemia in the population of neonates and infants under 3 months old and weighing less than 10kg. It can be used in the clinical setting or in the home.
The BiliCocoon™ Phototherapy System is a phototherapy system designed to treat unconjugated hyperbilirubinemia in newborns and infants under 3 months old and weighing less than 10kg. It is a phototherapy device, which emits light in the absorption spectrum of the bilirubin, from 430 to 490 nm. A custom optic directs light from the LEDs into the fiberoptic Pad. The system is composed of a blue light electronic generator - the Light Box - and of a light emitting fabric – the Pad – which transmits the blue light to the newborn. The Pad is provided in two versions: Bag Pad. The Light Box has a user interface to set the session time and inform the user about eventual errors. The other parameters of the therapy (irradiance level, light wavelength, etc.) are fixed.
The BiliCocoon™ Phototherapy System is used with the BiliCocoon™ Disposable, a non-woven disposable designed to fit the Pad which interfaces between the newborn. The disposable is the only part which contacts the newborn skin.
The BiliCocoon™ Phototherapy System can be fixed on the BiliCocoon™ Fixation system.
Here's an analysis of the provided text regarding the acceptance criteria and study for the BiliCocoon™ Phototherapy System (K163526).
It's important to note that this document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device, not necessarily proving efficacy through a clinical trial. Therefore, the information provided primarily pertains to bench testing and comparison, rather than comprehensive clinical study data as might be expected for novel devices or PMAs.
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a table of acceptance criteria with corresponding device performance for a specific statistical endpoint (e.g., accuracy, sensitivity, specificity) relevant to a diagnostic or AI-driven device. Instead, the "acceptance criteria" are implied by the performance of the predicate device and relevant industry standards. The "reported device performance" is largely described through a comparison against the predicate and adherence to safety and performance standards.
The closest equivalent to "acceptance criteria" for a phototherapy device would be its light output specifications and safety characteristics.
| Acceptance Criteria (Implied from Predicate/Standards) | Reported Device Performance (BiliCocoon™) |
|---|---|
| Light Characteristics | |
| Peak Emission Range | Blue LED Light with peak emission between 430 and 490 nm |
| Intensity (Irradiance) average | 30 - 40 µW/cm²/nm |
| Recommended by AAP for intensive phototherapy | Meets recommendation (>30 µW/cm²/nm) |
| Safety and Electrical | |
| Leakage Current | <100µA |
| Electrical Safety | Compliant with IEC 60601-1 (3rd edition A1:2012) |
| Electromagnetic Compatibility (EMI/EMC) | Compliant with IEC 60601-1-2 (4th edition) |
| Phototherapy Safety and Performance | Compliant with IEC 60601-2-50 (2nd edition) |
| Home Use Assessment | Assessed per IEC 60601-1-11 (2nd edition:2015) and FDA Guidance, "Applying Human Factors and Usability Engineering to Medical Devices" |
| Photobiological Safety | Assessed with IEC 62471 |
| Biocompatibility | |
| Disposable cover skin-contacting (prolonged) | Cytotoxicity (ISO 10993-5, USP 87) - Met requirements Intracutaneous Irritation (ISO 10993-10) - Met requirements Sensitization (ISO 10993-10, USP 88) - Met requirements |
| Other | |
| Noise | 35dB (Similar to predicate's 25dB; considered similar) |
| Alarms (bad optical connection, LED overheating, LED control) | Yes, for all listed. |
2. Sample Size Used for the Test Set and Data Provenance
This document primarily describes bench testing for hardware and software components, as well as biocompatibility. There is no "test set" in the context of clinical data (e.g., images, patient records) or a machine learning model.
- Sample Size: Not applicable in the traditional sense for a test set of clinical data. Testing involved samples of materials (e.g., disposable cover), components, and the device itself.
- Data Provenance: Not applicable. The testing is based on laboratory measurements and adherence to international standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable. There is no clinical "ground truth" derived from expert consensus mentioned in this 510(k) summary for the device's technical performance. Ground truth for biocompatibility is established by standardized lab tests, and for electrical/optical performance by instrumentation and adherence to published standards.
4. Adjudication Method for the Test Set
Not applicable, as there isn't a "test set" requiring expert adjudication in the context of this device's bench testing.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study was mentioned. This device is a phototherapy system, not an AI-assisted diagnostic tool for human readers.
6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)
Not applicable. The BiliCocoon™ Phototherapy System is a medical device, not an AI algorithm. Its performance is directly tied to its physical characteristics and light output.
7. Type of Ground Truth Used
The "ground truth" for the various performance and safety tests conducted were:
- Biocompatibility: Established by reference to ISO 10993 standards and USP 87/88.
- Spectral Output, Light Irradiance, Effective Surface Treatment Area: Established by direct measurement using calibrated equipment.
- Electrical Safety, EMC, Phototherapy Safety/Performance, Home Use, Photobiological Safety: Established by adherence to and compliance with relevant IEC standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-50, IEC 60601-1-11, IEC 62471) and FDA guidance.
8. Sample Size for the Training Set
Not applicable. This device does not employ a machine learning algorithm; therefore, there is no "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for a machine learning algorithm.
Summary regarding the nature of the study:
The provided document describes a bench testing study designed to demonstrate the substantial equivalence of the BiliCocoon™ Phototherapy System to a legally marketed predicate device (K103589, neoBLUE® blanket LED Phototherapy System). The study focuses on comparing technological characteristics and ensuring adherence to relevant safety and performance standards for neonatal phototherapy units. It is not a clinical study or an AI/software performance study in the way the input questions imply.
Ask a specific question about this device
(269 days)
Re: K160745
Trade/Device Name: neoBLUE Compact LED Phototherapy System Regulation Number: 21 CFR 880.5700
|
| Classification Name andNumber: | Neonatal Phototherapy Unit21 CFR 880.5700
The neoBLUE compact LED Phototherapy System is intended for the treatment of neonatal hyperbilirubinemia. The light can be used for infants in a bassinet, incubator, open bed, or radiant warmer.
The neoBLUE compact LED Phototherapy System utilizes a portable phototherapy light that delivers a narrow band of high-intensity blue light emitting diodes (LEDs) to provide treatment for neonatal hyperbilirubinemia.
The neoBLUE compact LED Phototherapy System light consists of a lightweight plastic light enclosure which can be used independently by placing directly on top of an incubator, or with the arm, or with the arm and roll stand together. The arm can be attached to the pole mount accessory on most incubators and radiant warmers. The arm, when attached to the roll stand, can be used for infants in a bassinet, incubator, open bed, or radiant warmer.
The neoBLUE compact LED Phototherapy System light is equipped with a treatment timer to track the total number of treatment hours per patient, a device timer to track the total number of operating hours for the blue LEDs, and an exam light.
The provided text describes the neoBLUE Compact LED Phototherapy System and its substantial equivalence to predicate devices, not a study evaluating its performance against specific acceptance criteria in a clinical setting. Therefore, many of the requested categories for a study are "Not Applicable" or "Not Provided" in this document.
Here's an analysis based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not present a table of specific quantitative acceptance criteria for clinical performance or a clinical study to prove the device meets them. Instead, it focuses on demonstrating substantial equivalence to predicate devices through technical characteristics and compliance with safety standards.
| Feature/Test Category | Acceptance Criteria (Implied from Predicate/Standards) | Reported Device Performance (neoBLUE compact LED Phototherapy System) |
|---|---|---|
| Intended Use | Treatment of neonatal hyperbilirubinemia | Treatment of neonatal hyperbilirubinemia |
| Treatment Method | Overhead phototherapy | Overhead phototherapy |
| User | Healthcare professionals in a clinical setting | Healthcare professionals in a clinical setting |
| Sites of Use | Bassinet, incubator, open bed, radiant warmer | Bassinet, incubator, open bed, radiant warmer |
| Patient Population | Neonates | Neonates |
| Light Source | LEDs (blue) | Blue and white Light Emitting Diodes (LEDs) |
| Wavelength | Peak @ 440-470 nm (combined range of predicates) | Peak @ 450 to 470 nm |
| Intensity | Range overlapping with predicates (>20 to 50 μW/cm²/nm @ 40cm, 40-49 μW/cm²/nm @ 30-40cm) | 10 – 55 μW/cm²/nm @ 35 cm |
| Intensity Settings | 1 or 3 settings (from predicates) | 2 settings (high, low) |
| Effective Treatment Area | Range overlapping with predicates (660 to 800 cm² @ 40cm) | 700 cm² @ 35 cm height |
| Exam Light | Present on one predicate, absent on other | White Light Emitting Diodes (LEDs) (present) |
| Electrical Safety | Compliance with IEC 60601-1, IEC 60601-1-2 | Compliance with IEC 60601-1, IEC 60601-1-2 |
| Thermal Safety | Fan to cool, thermal protection circuit | Fan to cool circuitry, Thermal protection circuit |
| Radiation Safety | Minimal UV/IR light | LED light source produces minimal UV and IR light |
| Software Verification | Verified | Performed |
| Design Verification & Validation | Performed | Performed (includes device power, electrical safety, indicators, LED performance, intensity levels, intensity range, effective treatment area, usability) |
| Hazard Analysis | Performed | Performed |
| Human Factors Validation Testing | Performed | Performed |
Note: The "acceptance criteria" here are implied from the comparison to predicate devices and the mention of standards compliance for safety and performance testing. There are no explicit, quantifiable clinical efficacy acceptance criteria provided within this document for a clinical study.
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not Applicable. This document describes a premarket notification (510(k)) seeking substantial equivalence, not a clinical trial with a patient test set. Performance testing mentioned (electrical safety, LED performance, etc.) would be conducted on units of the device, not a patient sample.
- Data Provenance: Not Applicable. No patient data or clinical data is mentioned for a test set.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not Applicable. No clinical ground truth or expert adjudication for patient data is mentioned.
4. Adjudication method for the test set
- Not Applicable. No patient test set or adjudication process is described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This device is a phototherapy system, not an AI-assisted diagnostic or interpretative tool for human readers. No MRMC study was conducted or is relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is a medical device (phototherapy system), not a standalone algorithm.
7. The type of ground truth used
- Not Applicable. For the technical performance aspects, the "ground truth" would be established by reference standards, engineering specifications, and validated measurement methods, rather than clinical ground truth (pathology, expert consensus, outcomes data).
8. The sample size for the training set
- Not Applicable. This is not an AI/machine learning device that requires a training set.
9. How the ground truth for the training set was established
- Not Applicable. This is not an AI/machine learning device.
Ask a specific question about this device
Page 1 of 4