Search Results
Found 67 results
510(k) Data Aggregation
(130 days)
STRATAFIX™ Spiral PRONOVA™ Bidirectional Knotless Tissue Control Device
Regulation Number: 21 CFR 878.5010
suture
Class: II
Classification Name: Suture, Nonabsorbable, Synthetic, Polypropylene (21 CFR 878.5010
** | Yes | Yes | Yes | EQUIVALENT |
| FDA Regulation | 21 CFR 878.5010: Nonabsorbable polypropylene
surgical suture | 21 CFR 878.5010: Nonabsorbable polypropylene surgical suture | 21 CFR 878.5010: Nonabsorbable
The STRATAFIX™ Spiral PRONOVA™ Unidirectional Device is indicated for use in soft tissue approximation where the use of a non-absorbable barbed suture is appropriate, excluding closure of the epidermis.
The STRATAFIX™ Spiral PRONOVA™ Bidirectional Device is indicated for use in soft tissue approximation where the use of a non-absorbable barbed suture is appropriate, excluding closure of the epidermis.
The STRATAFIX™ Spiral PRONOVA™ Knotless Tissue Control Devices (KTCD) are monofilament, synthetic non-absorbable devices composed of a polymer blend of poly (vinylidene fluoride) and poly(vinylidene fluoride-co-hexafluoropropylene). The STRATAFIX™ Spiral PRONOVA™ KTCD are pigmented blue to enhance visibility in the surgical field. The STRATAFIX™ Spiral PRONOVA™ KTCD has two variations based on the orientation of the barbs and needle attachment: STRATAFIX™ Spiral PRONOVA™ Unidirectional KTCD and STRATAFIX™ Spiral PRONOVA™ Bidirectional KTCD.
The STRATAFIX™ Spiral PRONOVA™ Unidirectional KTCD consists of barbed suture material, armed with a surgical needle on one end and a fixation loop at the opposite end. The STRATAFIX™ Spiral PRONOVA™ Unidirectional Device is designed to anchor with a closed loop at one end and a unidirectional barbed section on the other end. The STRATAFIX™ Spiral PRONOVA™ Unidirectional Device barbs are oriented in one direction to allow tissue approximation without the need to tie surgical knots.
The STRATAFIX™ Spiral PRONOVA™ Bidirectional KTCD Design consists of barbed suture material, armed with a surgical needle on each end. The device also contains an unbarbed center transition zone that facilitates the initiation of the device use. The STRATAFIX™ Spiral PRONOVA™ Bidirectional Device barbs allow for tissue approximation without the need to tie surgical knots.
The subject devices are for single-use and are sterilized by Ethylene Oxide. The subject devices are available in various lengths and USP diameter sizes armed with various needle sizes. STRATAFIX™ Spiral PRONOVA™ KTCD is intended for use only by Healthcare Professionals who are trained in surgical suturing techniques.
N/A
Ask a specific question about this device
(753 days)
Polypropylene SutureProduct code: GAW – Class II |
| Regulation: | 21 CFR 878.5010
Regulation: 21 CFR 878.5030 Regulation: 21 CFR 878.5020 Regulation: 21 CFR 878.5010
suture is monofilament
Regulation: 21 CFR 878.5030 Regulation: 21 CFR 878.5020 Regulation: 21 CFR 878.5010
SafePath Suturing System - Silk Suture: The SafePath Suturing System is intended for use in placement of a silk suture in the skin and subcutaneous tissues.
SafePath Suturing System - Polyamide (Nylon) Suture: The SafePath Suturing System is intended for use in placement of a nylon suture in the skin and subcutaneous tissues.
SafePath Suturing System - Polypropylene Suture: The SafePath Suturing System is intended for use in placement of a polypropylene suture in the skin and subcutaneous tissues.
The silk suture provided with the SafePath Suturing System is a nonabsorbable, sterile, surgical suture composed of the organic protein, fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori) of the family Bombycidae. The silk suture is silicone coated, braided and dyed (black) with logwood extract. The silk suture meets all requirements established by the United States Pharmacopeia (USP) for Nonabsorbable Surgical Suture and is available in USP size 2-0. The suture is provided as a 36 in. Jength, swaged on a single-armed 1½ circle, 26 mm reverse cutting needle. The suture is pre-loaded in a hand-held, manually operated suturing device. When the suturing device is actuated a single stitch is placed through the tissue and the needle is automatically captured and reset for placement of an additional stitch.
The nylon suture provided with the SafePath Suturing System is a nonabsorbable, sterile, surgical suture composed of long chain aliphatic polymers - nylon 6,6. SafePath nylon sutures are dyed with Logwood extract, FDA approval number §73.1410. The nylon suture meets all requirements established by the United States Pharmacopeia (USP) for Nonabsorbable Surgical Suture and is available in USP sizes 2-0 and 3-0. The suture is provided as a 36 in. length, swaged to a singlearmed ½ circle. 26 mm reverse cutting needle. The suture is pre-loaded in a handheld, manually operated suturing device. When the suturing device is actuated a single stitch is placed through the tissue and the needle is automatically captured and reset for placement of an additional stitch.
The polypropylene suture provided with the SafePath Suturing System is a monofilament synthetic non-absorbable, sterile surgical suture composed of an isotactic crystalline stereoisomer of polypropylene, a synthetic linear polyolefin. The molecular formula is (C3H6)n. The Safepath suture is dyed with [Phthalocyaninato(2-)] copper, FDA approval number §74.3045. The polypropylene suture meets all requirements established by the United States Pharmacopeia (USP) for Nonabsorbable Surgical Suture and is available in USP sizes 2-0 and 3-0. The suture is provided as a 36 in. length, swaged to a single-armed ½ circle, 26 mm reverse cutting needle. The suture is pre-loaded in a hand-held, manually operated suturing device. When the suturing device is actuated a single stitch is placed through the tissue and the needle is automatically captured and reset for placement of an additional stitch.
Finished devices may be packaged individually, or in multi-unit cartons or procedure packs. The SafePath Suturing System is designed to place a suture in skin and subcutaneous tissue.
This document discusses the FDA clearance for the SafePath Suturing System, which includes silk, polyamide (nylon), and polypropylene sutures. However, the provided text does not contain information about a study based on AI/ML. All the performance criteria and testing mentioned relate to the physical and biological properties of the sutures and the suturing system, not to an AI component.
Therefore, I cannot answer your request with the specific details about acceptance criteria and studies related to AI/ML performance, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance.
The document highlights the following non-clinical performance criteria and testing for the sutures:
1. A table of acceptance criteria and the reported device performance:
| Acceptance Criteria (General) | Reported Device Performance |
|---|---|
| USP Nonabsorbable Sutures - Diameter | Met or exceeded USP criteria |
| USP Sutures - Needle Attachment | Met or exceeded USP criteria |
| USP Sutures - Tensile Strength | Met or exceeded USP criteria |
| Biocompatibility - Cytotoxicity | Met test criteria |
| Biocompatibility - Sensitization | Met test criteria |
| Biocompatibility - Irritation | Met test criteria |
| Biocompatibility - Systemic Toxicity | Met test criteria |
| Biocompatibility - Hemocompatibility | Met test criteria |
| Biocompatibility - Pyrogenicity | Met test criteria |
| Biocompatibility - Endotoxicity | Met test criteria |
2. Sample size used for the test set and the data provenance: Not applicable as this is a non-AI/ML device. The testing was physical and biological, not based on data sets.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as this is a non-AI/ML device. Ground truth as typically understood in AI/ML is not relevant here.
4. Adjudication method for the test set: Not applicable as this is a non-AI/ML device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable as this is a non-AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable as this is a non-AI/ML device.
7. The type of ground truth used: For physical testing, ground truth would be established by the specifications of the USP standards. For biocompatibility, it's about meeting pre-defined safety thresholds as per recognized standards (e.g., ISO 10993).
8. The sample size for the training set: Not applicable as this is a non-AI/ML device.
9. How the ground truth for the training set was established: Not applicable as this is a non-AI/ML device.
The study described is a non-clinical performance testing study that evaluated the physical and biocompatibility characteristics of the sutures and the suturing system. The conclusions state that the device met or exceeded USP criteria for physical properties and met the test criteria for a battery of biocompatibility evaluations. The safety evaluation for the nylon and polypropylene sutures leveraged information from a previously cleared device (K180701) and used identical raw materials, packaging, and sterilization conditions as the predicate devices.
Ask a specific question about this device
(101 days)
|
| Regulation Number | 21 CFR 878.484021 CFR 878.5010
The Cypris eXact Suture Placement Device is indicated for use in general soft tissue approximation when used with sutures.
The Cypris eXact Suture Placement Device is for use in general soft tissue approximation by surgeons, and it is single-use and provided sterile. It is a hand-held device used for the placement of a surgical suture, and the device consists of a handle, needle driver, and tissue capture port. The handle is designed for manual manipulation of the suture placement device, allowing the surgeon to advance the device to the desired deployment site. The needle driver functions to pass the suture when it is advanced and retracted. The tissue capture port consists of a capture chamber, needle, and hook assembly, and holds the tissue to be sutured during either manual palpation or vacuum-assisted palpation of tissue into the tissue capture port. Suture placement is achieved by advancing and retracting the needle driver.
{
"1. A table of acceptance criteria and the reported device performance": {
"Acceptance Criteria": [
"Biocompatibility endpoints evaluated (cytotoxicity, sensitization, irritation, acute systemic toxicity, and material-mediated pyrogenicity)",
"Integrity of the sterile barrier and functionality of the device after 3 years of real-time aging",
"Performance as expected",
"Use of vacuum-assisted palpation equivalent to manual palpation by the predicate device",
"Peak suction force measurement at 25.0 inHg"
],
"Reported Device Performance": [
"Device underwent testing for cytotoxicity, sensitization, irritation, acute systemic toxicity, and material-mediated pyrogenicity.",
"Testing demonstrated integrity of the sterile barrier and functionality after 3 years of real-time aging (shelf-life is 3 years).",
"Testing demonstrated that the subject device performs as expected.",
"Testing demonstrated that the use of vacuum-assisted palpation is equivalent to the use of manual palpation by the predicate device.",
"At a vacuum level of 25.0 inHg, the minimum, maximum, and average measured peak suction force achieved was 0.753 lbf, 1.052 lbf, and 0.904 lbf, respectively."
]
},
"2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)": "The provided text describes non-clinical testing. It does not mention specific sample sizes for a test set in the context of clinical data or human subjects. The data provenance is not specified, but the testing appears to be laboratory-based (non-clinical).",
"3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)": "Not applicable. This is a non-clinical study for a manual surgical instrument, not an AI/imaging device requiring expert ground truth establishment for a test set.",
"4. Adjudication method (e.g. 2+1, 3+1, none) for the test set": "Not applicable. This is a non-clinical study for a manual surgical instrument, not an AI/imaging device requiring an adjudication method.",
"5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance": "Not applicable. This is a non-clinical study for a manual surgical instrument, not an AI/imaging device. No MRMC study was conducted or mentioned.",
"6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done": "Not applicable. This is a non-clinical study for a manual surgical instrument. No algorithm or standalone performance testing was mentioned.",
"7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)": "Not applicable. For this non-clinical study, ground truth was established by adherence to performance specifications and established biocompatibility standards.",
"8. The sample size for the training set": "Not applicable. This is a non-clinical study for a manual surgical instrument, not an AI/machine learning device requiring a training set.",
"9. How the ground truth for the training set was established": "Not applicable. This is a non-clinical study for a manual surgical instrument, not an AI/machine learning device. No training set was used or ground truth established for one."
}
Ask a specific question about this device
(105 days)
878.4840 Absorbable polydioxanone surgical suture |
| | 878.5010
|
| Regulation Number | 21 CFR 878.5010
| 21 CFR 878.5010
The Cypris eXact Suture (PP) is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
The Cypris eXact Suture (PTFE) is indicated for use in all types of soft tissue approximation and/or ligation, including dental and general surgeries.
The Cypris eXact Suture (PDO) is indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular or neurological tissues, microsurgery or ophthalmic surgery.
The Cypris eXact 5.1 Suture Placement Device can be used in conjunction with eXact Sutures, which are indicated for use in general soft tissue approximation and/or ligation.
The Cypris eXact Suturing System is comprised of the Cypris eXact Suture and the Cypris eXact 5.1 Suture Placement Device. The Cypris eXact Suture and the Cypris eXact 5.1 Suture Placement Device are for use in general soft tissue approximation by surgeons, are single-use, and are provided sterile.
The Cypris eXact Suture is a synthetic, monofilament suture available in three different materials: polypropylene (PP), polydioxanone (PDO), and polytetraflouroethylene (PTFE). The Cypris eXact Suture (PP) is a non-absorbable, sterile surgical suture available in two sizes, USP sizes 3-0 and 4-0. The Cypris eXact Suture (PTFE) is a non-absorbable, sterile surgical suture available in two sizes, USP sizes 2-0 and 3-0. The Cypris eXact Suture (PDO) is an absorbable, sterile surgical suture available in two sizes, USP sizes 3-0 and 4-0. Each 24-inch suture has a stainless-steel BB needle attached to one end of the suture.
Cypris eXact Sutures can be used with the Cypris eXact 5.1 Suture Placement Device, a handheld device used for the placement of a surgical suture.
The FDA 510(k) clearance document for the Cypris eXact Suturing System (K233355) indicates that the device's substantial equivalence to predicate devices is based on non-clinical testing rather than a study involving human-in-the-loop performance or diagnostic accuracy of an AI algorithm. Therefore, many of the requested criteria related to AI/MRMC studies, number of experts, adjudication methods, and training/test set ground truth for AI models are not applicable.
The document primarily focuses on demonstrating that the Cypris eXact Sutures are identical in formulation, processing, sterilization, and geometry to previously cleared sutures and that their performance meets established USP and FDA guidance standards for surgical sutures. The Cypris eXact 5.1 Suture Placement Device is evaluated for its integrity and functionality.
Here's a breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria (USP/FDA Guidance Standards) | Reported Device Performance (Cypris eXact Suture) |
|---|---|
| Suture Diameter meets USP <861> requirements | All suture sizes meet USP <861> requirements |
| Suture Tensile Strength meets USP <881> requirements | All suture sizes meet USP <881> requirements |
| Needle Attachment Strength meets USP <871> requirements | All suture sizes meet USP <871> requirements |
| Biocompatibility (as per ISO 10993-1:2018 for implanted devices with long-term contact) | All evaluated endpoints met (cytotoxicity, sensitization, acute systemic toxicity, material-mediated pyrogenicity, subacute/subchronic toxicity, chronic toxicity, genotoxicity, carcinogenicity, and implantation). |
| Sterile Barrier Integrity (for 5.1 Suture Placement Device) | Demonstrated following 3 years of real-time aging. |
| Functionality (for 5.1 Suture Placement Device) | Demonstrated following 3 years of real-time aging. |
| Shelf-life for Sutures | 5 years |
| Shelf-life for Suture Placement Device | 3 years |
2. Sample Size Used for the Test Set and Data Provenance:
The document doesn't explicitly state "sample sizes for a test set" in the context of an AI-driven study. Instead, it refers to performance testing of the physical sutures and the placement device. The data provenance is not described as medical cases or patient data, but rather as results from laboratory testing performed in accordance with USP and ISO standards. The country of origin of the data is not specified beyond being generated for an FDA submission. The studies were likely prospective in that they were designed and performed to meet regulatory requirements prior to submission.
3. Number of Experts Used to Establish Ground Truth and Qualifications:
Not applicable. This device is not an AI/diagnostic device that requires expert ground truth for image interpretation or diagnosis. The "ground truth" here is compliance with established engineering and biocompatibility standards.
4. Adjudication Method for the Test Set:
Not applicable. This is not an AI/diagnostic device.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
No. This is not an AI/diagnostic device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Not applicable. This is not an AI/diagnostic device. The "standalone" performance relates to the physical and biological characteristics of the suture materials and the mechanical functionality of the placement device as tested in laboratory settings.
7. The Type of Ground Truth Used:
The ground truth used for evaluating the performance of the sutures and placement device are established industry standards and regulatory guidelines:
- USP <861>, <881>, <871> requirements for surgical sutures (diameter, tensile strength, needle attachment strength).
- ISO 10993-1:2018 standards for biological evaluation of medical devices (biocompatibility endpoints).
- FDA's Guidance for Industry and FDA Staff: Surgical Sutures - Class II Special Controls Guidance Document for Industry and FDA Staff (issued June 3, 2003).
8. The Sample Size for the Training Set:
Not applicable. This is not an AI device that requires a training set.
9. How the Ground Truth for the Training Set Was Established:
Not applicable. This is not an AI device.
Summary of the Study Proving Device Meets Acceptance Criteria:
The study proving the device meets its acceptance criteria is a series of non-clinical, laboratory-based performance and biocompatibility tests.
- Device Tested: Cypris eXact Sutures (PP, PTFE, PDO) and Cypris eXact 5.1 Suture Placement Device.
- Tests Performed:
- Suture Testing: Measurement of suture diameter, tensile strength, and needle attachment strength for various suture sizes and materials. These tests were conducted according to the specific methods outlined in the United States Pharmacopeia (USP) standards (<861>, <881>, <871>).
- Biocompatibility Testing: Evaluation of material interactions with biological systems, including cytotoxicity, sensitization, irritation (for the placement device), acute systemic toxicity, material-mediated pyrogenicity, subacute and subchronic toxicity, chronic toxicity, genotoxicity, carcinogenicity, and implantation (for the sutures). These tests followed ISO 10993-1:2018 guidelines.
- Shelf-life Testing: Evaluation of sterile barrier integrity and device functionality for the Cypris eXact 5.1 Suture Placement Device after 3 years of real-time aging to establish its shelf-life. The sutures had a separate, longer shelf-life of 5 years.
- Acceptance Criteria Basis: The acceptance criteria for each test were defined by the adherence to the requirements specified in the aforementioned USP standards, ISO 10993-1:2018, and FDA's specific guidance for surgical sutures.
- Conclusion: The tests demonstrated that the Cypris eXact Sutures meet all specified USP requirements and that all materials used are biocompatible. The Cypris eXact Sutures are explicitly stated as being "identical in their final finished form in formulation, processing, sterilization, and geometry to the sutures previously cleared" under the predicate devices (K131224, K160744, K063680). The Cypris eXact 5.1 Suture Placement Device also demonstrated integrity and functionality over its claimed shelf-life. This body of non-clinical evidence provided the basis for the FDA's determination of substantial equivalence.
Ask a specific question about this device
(181 days)
Trade/Device Name: Non absorbable Surgical Polypropylene Suture Regulation Number: 21 CFR 878.5010
Nonabsorbable, Synthetic, Polypropylene Classification: II Product Code: GAW Regulation Number: 21CFR 878.5010
Identification of Predicate Device 510(k) Number: K080684 Regulation Number: 21CFR 878.5010 Classification
| Same |
| Regulation Number | 21CFR 878.5010
| 21CFR 878.5010
Non absorbable Surgical Polypropylene Suture is indicated for use in general soft tissue approximation and/or ligation, but not for use in ophthalmic procedures, cardiovascular and neurological procedures. The device is limited to use where short term wound support (7-10 days) is required and can be left in place for a maximum of 10 days.
The subject device is a nonabsorbable monofilament surgical suture which is supplied sterile. The proposed device is dyed blue. The color additive is [phtalocyaninato(2-)] copper(Color Index Number 74160), and the weight percentage for the color additive is less than 0.1%. The device will be offered in diameters ranging from USP size 6-0 through 2 and is available in length 75cm or 150cm with or without needles attached. Polypropylene suture meets all the requirements of USP for Non-absorbable surgical suture.
The provided text describes the 510(k) summary for a Non Absorbable Surgical Polypropylene Suture (K230746). It focuses on demonstrating substantial equivalence to a predicate device (K080684) through non-clinical performance data.
Here's a breakdown of the requested information based on the provided document:
1. Table of acceptance criteria and the reported device performance
The document does not explicitly provide a single table listing "acceptance criteria" alongside "reported device performance" in a quantitative manner for most tests. Instead, it lists the standards the device complies with for various performance aspects and states that the test results met these requirements.
However, some specific acceptance criteria and general performance statements are given:
| Criterion Type | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Diameter of suture | Comply with USP <861> | "the test result shows that the length and diameter of the proposed device met the acceptance criteria." (General statement, implies compliance with USP <861>) |
| Needle Attachment | Comply with USP <871> | Reported as "Same" as predicate, implying compliance with USP <871> |
| Tensile Strength | Comply with USP <881> | Implied compliance with USP <881> |
| Length | Not less than 95.0% of the length stated on the label | "the test result shows that the length and diameter of the proposed device met the acceptance criteria." (General statement, implies compliance with this criterion) |
| Cytotoxicity | The viability is not reduced to less than 70% | The device passed, implying viability was not reduced to less than 70% |
| Sensitization | The Magnusson and Kligman grades is less than 1. | The device passed, implying grades less than 1. |
| Intracutaneous Reactivity | The erythema and edema grades is less than 1.0. | The device passed, implying grades less than 1.0. |
| Acute systemic toxicity | No animal died or abnormal behavior occurred. | The device passed, implying no animal died or abnormal behavior occurred. |
| Pyrogen | Temperature raise is less than 0.5℃ No behavioral change or sign of toxicity was observed. Clinical pathology parameter within the reference range. | The device passed, implying these conditions were met. |
| Subacute Systemic Toxicity | No macroscopic changes in the viscera at necropsy. Histopathology within normal histomorphological limits | The device passed, implying these conditions were met. |
| Bacterial Reverse Mutation | No obviously increase in the mean number of revertant of colonies between the test group and control group. | The device passed, implying no obvious increase. |
| Chromosome Aberration | No significant difference in the percentage of cells with chromosome aberrations between the test group and control group. | The device passed, implying no significant difference. |
| Gene Mutation | There was no significant difference in the TFT-resistant mutant frequency between the test article | The device passed, implying no significant difference. |
| Implantation | No lesion at the implantation site. The irritation score is less than 1.0. | The device passed, implying no lesion and irritation score less than 1.0. |
| Hemolysis | Hemolytic index is less than 2% | The device passed, implying hemolytic index less than 2%. |
| Seal Strength | Complies with ASTM F88/F88M-15 | "The test results demonstrated that the subject device complies with the following standards" (including ASTM F88/F88M-15). |
| Detecting Seal Leaks | Complies with ASTM F1929-15 | "The test results demonstrated that the subject device complies with the following standards" (including ASTM F1929-15). |
| Hemolytic Properties | Complies with ASTM F756-17 | "The test results demonstrated that the subject device complies with the following standards" (including ASTM F756-17). |
| EO Sterilization Residuals | Complies with ISO 10993-7:2008 | "The test results demonstrated that the subject device complies with the following standards" (including ISO 10993-7:2008). |
| Penetration Testing of Needles | Complies with ASTM F3014-14 | "The test results demonstrated that the subject device complies with the following standards" (including ASTM F3014-14). |
| Bend Testing of Needles | Complies with ASTM F1874-98 | "The test results demonstrated that the subject device complies with the following standards" (including ASTM F1874-98). |
2. Sample size used for the test set and the data provenance
The document does not specify the sample sizes for any of the non-clinical tests mentioned.
The data provenance is also not specified, as these are bench tests and biological evaluations, not human data. It is implied these tests were conducted by the manufacturer or a contracted lab to meet international standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This question is not applicable to the provided information. The tests conducted are non-clinical (bench testing, biocompatibility). There is no "test set" requiring expert ground truth establishment in the context of diagnostic performance or clinical outcomes. The "ground truth" for these tests comes from adherence to established scientific and regulatory standards (e.g., USP, ISO, ASTM).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This question is not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies or diagnostic performance evaluations involving human interpretation. The reported tests are non-clinical lab tests and biocompatibility assessments, where results are typically objective measurements or observations against predefined criteria.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is not applicable. The device is a surgical suture, not an AI-powered diagnostic tool. No MRMC study or AI assistance is mentioned or relevant to this device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable. The device is a surgical suture, not an algorithm or AI system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the non-clinical tests is established by adherence to recognized international standards and pharmacopeial monographs. For example:
- USP <861>, <871>, <881>: These are pharmacopeial standards defining requirements for suture diameter, needle attachment, and tensile strength.
- ISO 10993 series: These are international standards for the biological evaluation of medical devices, covering aspects like cytotoxicity, sensitization, systemic toxicity, genotoxicity, implantation, and hemolysis.
- ASTM standards: These are industrial standards covering material properties and packaging (e.g., seal strength, dye penetration, needle bend/penetration).
The ground truth is based on the objective criteria and methodologies outlined in these standards.
8. The sample size for the training set
This question is not applicable. There is no "training set" as this is not a machine learning or AI device. The submission focuses on verifying the physical and biological characteristics of a mechanical device against established standards.
9. How the ground truth for the training set was established
This question is not applicable, as there is no training set for this device.
Ask a specific question about this device
(409 days)
Suture, Non-absorbable, Synthetic, Polypropylene (1)
- (2) Product Code: GAW
- (3) Regulation Number: 878.5010
Non-absorbable Polypropylene Suture is indicated for use in general soft tissue approximation and/or ligation. The device can be left in place for a maximum of 30 days.
Non-absorbable Polyester Suture is indicated for use in general soft tissue approximation and/or ligation. The device can be left in place for a maximum of 30 days.
Non-absorbable Polyamide suture is indicated for use in general soft tissue approximation and/or ligation. The device can be left in place for a maximum of 30 days.
The applicant devices of Non-absorbable Polypropylene Suture consist of polypropylene surgical suture made of polypropylene and a stainless steel needle. It is monofilament. It is EO sterilized, and prygon-free. It is dyed by §74.3045 [Phthalocyaninato(2-)] copper.(CI Number 74160) which is less than 0.1 % by weight.
The applicant devices of Non-absorbable Polypropylene Suture are available in 11-0, 10-0, 8 -0, 7-0, 6-0, 5-0, 4-0, 3-0, 2-0, 0, 1, 2 and 3, which are the sizes identified in the currently recognized United States Pharmacopoeia.
The applicant devices of Non-absorbable Polyester Suture consist of a polyester surgical suture made of polyester and a stainless steel needle. It is braided and coated with silicone. It is EO sterilized, and prygon-free.It is undyed (natural white) or dyed green with § 74.3206 D&C Green No. 6. (CI Number 61565), which is less than 0.1 % by weight. The applicant devices of Non-absorbable Polyester Suture are available in 9-0, 8-0, -0, 6-0, 5-0, 4-0, 3-0, 2-0, 0, 1, 2 and 3, which are the sizes identified in the currently recognized United States Pharmacopoeia.
The applicant devices of Non-absorbable Polyamide suture consist of a polyamide surgical suture made of long-chain, aliphatic polymers nylon and a stainless steel needle. It is monofilament. It is EO sterilized, and prygon-free. It is dyed by § 74.3102 FD&C Blue No. 2. ( CI Number 73015) or § 73.1410 Logwood Black (CI Number 75290), which is less than 0.1 % by weight.
The applicant devices of Non-absorbable Polyamide suture are available in 11-0, 10-0, 8-0, 7-0, 6-0, 5-0, 4-0, 3-0, 2-0, 0, 1, 2 and 3, which are the sizes identified in the currently recognized United States Pharmacopoeia.
The proposed devices are provided sterile and single use.
The proposed device is composed of suture with or without the needle.
The needle is made of stainless steel. The needles are available in six types: Taper, Cutting, Taper Cutting, Reverse Cutting, Diamond, Premium Cutting and Spatula. The Arc (Circle) of needle has 1/4 circle, 3/8circle,1/2circle,5/8circle,compound curve, straight.
The proposed devices are provided in various combinations of suture length, diameter, color (dyed or undyed), quantity of needle and needle types.
The proposed devices are sterilized by EO to achieve a SAL 10-6 and supplied in sterility maintenance package which could maintain the sterility of the device during the shelf life of 5 years.
Here's a breakdown of the acceptance criteria and study information for the "Surgical Sutures with or without Needle" device based on the provided FDA 510(k) summary.
It's important to note that this document is a 510(k) summary for a medical device that demonstrates substantial equivalence to a predicate device, not necessarily a study proving clinical effectiveness or improvement with AI. The focus is on demonstrating that the new device is as safe and effective as devices already on the market. Therefore, many standard AI/ML study components (like multi-reader multi-case studies, ground truth establishment for a training set, the role of experts for ground truth, or effects of AI on human readers) are not applicable here.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are primarily based on established standards for surgical sutures. The reported performance indicates compliance with these standards.
| Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|
| Physical Performance: | |
| USP 43-NF 38:2020 <861> Sutures - Diameter | The suture diameters of the proposed device comply with the diameter requirement listed in USP <861> Diameter. |
| USP 43-NF 38:2020 <881> Tensile Strength | The tensile strengths of the proposed device comply with the tensile requirement listed in USP <881> Tensile Strength. |
| USP 43-NF 38:2020 <871> Sutures Needle Attachment | The bond between suture and needle of the applicant device meets the requirements defined in USP <871>. |
| USP 43-NF 38:2020 Non-absorbable Surgical Suture Monograph | The proposed device has been demonstrated to comply with the requirements listed in USP Monograph of Non-Absorbable Surgical Suture. |
| Biocompatibility: | |
| ISO 10993-5:2009 (Cytotoxicity) | No cytotoxicity. |
| ISO 10993-10: 2010 (Irritation and Skin Sensitization) | No skin sensitization; No intracutaneous reactivity. |
| ISO 10993-11:2017 (Systemic Toxicity) | No acute systemic toxicity, No subacute/subchronic systemic toxicity. |
| ISO 10993-3:2014 (Genotoxicity) | No genotoxicity. |
| USP 43-NF 38:2020 <85> Bacterial Endotoxins Limit (Pyrogenicity) | No potential febrile reaction. |
| ISO 10993-4:2017 (Hemolysis) | No hemolysis (direct and indirect contact). |
| ISO 10993-6:2016 (Local effects after implantation - Muscle Implantation Study) | No reaction to the tissue. |
| Sterility & Packaging: | |
| ISO 11135:2014 (Sterilization by Ethylene Oxide) | EO sterilized to achieve a SAL 10^-6^. |
| USP 43-NF 38:2020 <71> Sterility Test | Complies with Sterility Test. |
| ASTM F88/F88M-15 (Seal Strength) | Test performed. |
| ASTM F1929-15 (Dye Penetration for Seal Leaks) | Test performed. |
| ASTM F1140-13 (Internal Pressurization Failure Resistance) | Test performed. |
| Other: | |
| Shelf Life | Performance and Package Integrity Test Report after Accelerated Aging supports a 5-year shelf life. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify exact sample sizes for each non-clinical test. The testing was conducted to verify compliance with various established standards (USP, ISO, ASTM). The data provenance is generally "non-clinical tests" performed by the manufacturer in China (Huaian Seamen Medical Technology Co., LTD) to support the 510(k) submission. This is considered retrospective in the sense that the test results are being summarized for submission, not generated as part of a prospective clinical trial.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This is not a study involving diagnostic interpretation or machine learning where "ground truth" is established by experts in an interpretative capacity. The "ground truth" here is compliance with technical specifications and safety standards as measured by laboratory tests.
4. Adjudication Method for the Test Set
Not applicable. There is no adjudication method described as this is not a study involving human interpretation of data where disagreements need to be resolved. The test results are objective measurements against defined standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
Not applicable. This document is for a physical surgical device (sutures), not an AI/ML-driven diagnostic or interpretative device. Therefore, no MRMC study or AI assistance evaluation was performed or is relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. This is not an algorithm or AI device.
7. The Type of Ground Truth Used
The "ground truth" in this context refers to the defined acceptable parameters and thresholds within the cited standards (USP, ISO, ASTM). For example, the "ground truth" for suture diameter is the range specified in USP <861>. For biocompatibility, the "ground truth" is the absence of adverse biological reactions as defined by ISO 10993 series. These are objective, measurable standards.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/ML device that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. This is not an AI/ML device that requires a training set or its associated ground truth establishment.
Ask a specific question about this device
(220 days)
| 878.5010
The Vesseal is intended for use in the delivery of 8 interrupted sutures to assist the surgeon in the creation of a vascular anastomosis in arteries of 2-4mm in diameter. It is not intended for use in coronary artery grafts.
The Vesseal is a hand-held device, enabling simultaneous deployment of 8 double-armed 9-0 to 7-0 nylon sutures (USP certified). The sutures are delivered from the inside out, around the circumference of two ends of blood vessels (ranging in size from 2mm to 4mm) to assist in the creation of a microvascular anastomosis. The Vesseal is comprised of two main mechanisms: 1. The needle insertion mechanism contains two rod protrusions, intended for locating the blood vessel, opening, mounting, and positioning them relative to the stopper mechanism. Each rod stores, 8 curved needles that allow the transition of the double-armed nylon sutures during deployment. 2. The handle deployment mechanism features right and left switches to deploy the needles on the respective rod side, and the gear to rotate the rod and enable access to its entire perimeter.
The given text is a 510(k) summary for the medical device "Vesseal". There is no mention of acceptance criteria or a study that proves the device meets acceptance criteria. The document describes the device, its intended use, comparison to predicate devices, and the types of performance data used to support substantial equivalence.
Therefore, I cannot provide an answer based on the provided text.
Ask a specific question about this device
(500 days)
Chicago, Illinois 60611
Re: K211178
Trade/Device Name: DURAMESH™M Mesh Suture Regulation Number: 21 CFR 878.5010
Polypropylene |
| Regulation Number | 878.5010
Duramesh™ Mesh Suture is indicated for general soft tissue approximation and/or ligation in clean (CDC Class I) wounds, including tendon repair and midline laparotomy where the bowel is not entered. Duramesh™ is not for use in the skin and other epithelial surfaces, cardiovascular, ophthalmic, and/or neurological procedures.
DURAMESH™ is a polyfilament, synthetic, non-absorbable, sterile surgical suture composed of isotactic polypropylene polymer of high molecular weight. The suture is provided dyed (blue) and is uniform in appearance. The blue dye pigment is Phthalocyaninato (2-) Copper. The product meets all requirements established by the United States Pharmacopeia (USP) for nonabsorbable surgical suture, except for diameter.
DURAMESH™ is passed on either side of tissues using an attached swaged needle as an introducing agent and tied to itself with at least two knots composed of two throws each to maintain tension.
DURAMESH™ is supplied sterile and is available with two needle configurations (small and large) in four USP sizes (2-0 through 2), each with a length of approximately 36 inches (91 cm). Each box contains one DURAMESH™ housed in a Tyvek envelope within which the DURAMESH™ is packaged in a sterile protective medical grade tube fixed to a sterile card.
The provided text does not describe an AI/ML medical device, but rather a surgical suture (DURAMESH™ Mesh Suture). Therefore, the concepts of acceptance criteria for AI/ML performance metrics (like sensitivity, specificity, AUC), sample size for test/training sets in an AI/ML context, number of experts for ground truth establishment, adjudication methods, MRMC studies, standalone performance, and ground truth types related to AI/ML are not applicable to this document.
The document focuses on demonstrating substantial equivalence of the DURAMESH™ Mesh Suture to a predicate device based on material, design, performance, and biological safety. The "acceptance criteria" here refer to the product's ability to meet medical device standards and show equivalence to a predicate, not AI/ML model performance metrics.
Specifically, the "Bench Data" section on page 6 mentions some performance criteria for the suture:
- Needle attachment strength: Meets prospectively defined acceptance criteria per USP43-NF38: 2020 <871> Sutures - Needle Attachment.
- Tensile strength: Meets prospectively defined acceptance criteria per USP43-NF38: 2020 <881> Sutures - Tensile Strength. It also meets USP requirements under simulated surgical deployment, cycle loading, and dynamic loading.
- Knot Size and Knot Strength: DURAMESH™ knots were stronger than predicate knots for the same number of throws. Knot security is achieved with 4 throws as per labeling.
- Diameter: Diameter decreases with increasing axial load.
- "Stripped" samples: Maintain adequate tensile strength to satisfy USP requirements.
The document also describes:
- Biocompatibility testing: Per ISO 10993-1, ISO 14971, and 2020 FDA Biocompatibility Guidance, showing biocompatibility based on various tests (cytotoxicity, sensitization, etc.).
- Packaging testing: According to ASTM D4169-16 and ASTM F1929-15.
- Sterilization: Ethylene Oxide sterilization following AAMI TIR28:2016 and ISO 11135:2014, meeting ISO 10993-7: 2008 limits for residuals.
- Shelf-Life: Accelerated and real-time aging stability testing supports a 5-year shelf life with no loss of tensile strength.
- Animal Data: Two large-animal porcine implantation studies (1- and 3-month) assessed adverse tissue reaction and dehiscence. Outcomes showed successful closure, tissue ingrowth, and slight irritation compared to standard polypropylene suture.
- Clinical Data: A single-investigator clinical trial with 53 patients (80 surgical sites).
- Primary outcome: Surgical site occurrence (SSO). No early or late SSO observed.
- Secondary outcomes: Device-related rehospitalization, return to OR, and wounds. No such events occurred.
- Long-term assessment (12-month): 15 subjects compared to 11 patients with standard sutures, showing equivalent or slightly improved outcomes with DURAMESH™.
Given this context, I cannot generate a response that directly addresses the prompt's specific requirements for AI/ML device acceptance criteria and study design, as the provided document is not about an AI/ML device.
Ask a specific question about this device
(42 days)
COR-SUTURE OUCIK LOAD Surgical Suture COR-SUTURE QUICK LOAD Surgical Suture Regulation Number: 21 CFR 878.5010
Regulations: | 21 CFR 878.5000 (Nonabsorbable poly(ethylene terephthalate) surgical suture)21 CFR 878.5010
The RAM® COR-SUTURE® QUICK LOAD® surgical suture is indicated for use in the approximation of soft tissue and prosthetic materials.
The COR-SUTURE® OUICK LOAD® surgical suture is indicated for use in the approximation of soft tissue and prosthetic materials.
LSI SOLUTIONS® offers both the RAM® COR-SUTURE® QUICK LOAD® surgical suture, which is a non-absorbable polytetrafluoroethylene (PTFE)-coated, braided polyester surgical suture, and the COR-SUTURE® QUICK LOAD® surgical suture, which is available as either a PTFE-coated, braided polyester surgical suture or a monofilament, uncoated polypropylene surgical suture. The RAM® COR-SUTURE® QUICK LOAD® surgical suture is also supplied with an optional pre-attached PTFE pledget that is used to aid in suture buttressing and to mechanically secure and support the suture in fragile tissue, increasing the suture holding strength. A short length of modified surgical stainless-steel tubing, called a "needle cap", is attached to each end of the suture for both the RAM® COR-SUTURE® QUICK LOAD® surgical suture and the COR-SUTURE® QUICK LOAD® surgical suture. Relative to the COR-SUTURE® QUICK LOAD® device, the RAM® COR-SUTURE® QUICK LOAD® device additionally includes a needle cap holder to secure the optional pledget and needle caps in the suture tray. Both the RAM® COR-SUTURE® QUICK LOAD® surgical suture and the COR-SUTURE® QUICK LOAD® surgical suture also include a detachable clear suture tube to keep the suture from tangling. The needle caps are removed from the sutures and discarded once suturing is completed as they are not intended for implantation in the patient; only the suture and pledget are implantable of components.
Both devices are packaged for single patient use and are provided sterile by means of a validated ethylene oxide (EO) cycle. The polyester RAM® COR-SUTURE® QUICK LOAD® surgical suture is offered undyed (white), dyed green with the FDA approved colorant D&C Green No. 6. or striped green/white; all of which are offered with a surgical suture diameter of USP size 2-0 and a length of 38 inches. The polyester COR-SUTURE® QUICK LOAD® surgical suture is available undyed (white) or dyed green with the FDA approved colorant D&C Green No. 6 and will be offered with a surgical suture diameter of USP size 2-0 and a length of 38 inches. The polypropylene COR-SUTURE® QUICK LOAD® surgical suture is dyed blue with the FDA approved colorant [phthalocyaninato(2-)] copper and will be offered with a surgical suture diameter of USP size 3-0 and a length of 53 inches. For both polypropylene and polyester suture, there is no known significant change in tensile strength retention to occur in vivo. The RAM® COR-SUTURE® QUICK LOAD® and COR-SUTURE® QUICK LOAD® surgical sutures, with needle caps removed, contains implantable components which are considered MR safe. The implantable polyester and polypropylene sutures and PTFE pledget material are non-metallic and non-conducting materials and therefore, considered MR safe.
This document describes the RAM® COR-SUTURE® QUICK LOAD® and COR-SUTURE® QUICK LOAD® surgical sutures. It concerns a 510(k) premarket notification for these devices, asserting their substantial equivalence to a predicate device.
1. Table of acceptance criteria and the reported device performance:
| Acceptance Criteria (from USP Monograph for Non-Absorbable Sutures) | Reported Device Performance (RAM® COR-SUTURE® QUICK LOAD® and COR-SUTURE® QUICK LOAD® Surgical Sutures) |
|---|---|
| Conformance to USP Monograph for Non-Absorbable Sutures (specifically Physical Tests) | The subject RAM® COR-SUTURE® QUICK LOAD® and COR-SUTURE® QUICK LOAD® surgical sutures continue to conform to these requirements. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
The document does not explicitly state the sample size used for the benchtop testing test set. The data provenance is also not explicitly stated in terms of country of origin or whether it was retrospective or prospective. However, the testing was conducted as recommended by the FDA Guidance for Industry, Class II Special Control Guidance Document.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This information is not applicable. The device is a surgical suture, and "ground truth" as typically used in AI/medical imaging studies (e.g., expert consensus on diagnoses) is not relevant for this type of device and its performance testing. The performance is assessed against established physical standards and biocompatibility.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not applicable. Adjudication methods are typically employed in studies involving human interpretation or subjective assessments, which is not the case for the physical and biocompatibility testing of a surgical suture.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable. This is a medical device (surgical suture) and not an AI-powered diagnostic or assistive tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This information is not applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):
For the benchtop testing, the "ground truth" or standard for performance is the United States Pharmacopeia (USP), Monograph for Non-Absorbable Sutures, specifically sections from the USP General Chapter on Physical Tests.
For biocompatibility testing, the standards are ANSI AAMI ISO 10993-1:2009/(R)2013 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process and the FDA Guidance for Industry, Use of International Standard ISO 10993-1.
8. The sample size for the training set:
This information is not applicable. There is no concept of a "training set" for physical and biocompatibility testing of a surgical suture.
9. How the ground truth for the training set was established:
This information is not applicable, as there is no training set for this type of device testing.
Ask a specific question about this device
(39 days)
SUTURE, 2-0 Polyester RD QUCIK LOAD SUTURE, 2-0 Polypropylene Regulation Number: 21 CFR 878.5000 21 CFR 878.5010
| Nonabsorbable polypropylene surgical suture(Regulation: 21 CFR 878.5010
The RD® QUICK LOAD® suture is intended for use in general soft tissue approximation and/or ligation.
LSI SOLUTIONS® RD® QUICK LOAD® SUTURE is a non-absorbable surgical suture available as monofilament polypropylene suture (diameter of USP size 2-0) and polytetrafluoroethylene (PTFE)-coated, braided polyester suture (diameter of USP sizes 0 and 2-0). A short length of surgical stainless-steel tubing, called a "needle cap," is attached to one end of the suture. The needle cap is removed from the suture after suturing is completed and not meant for implantation into the patient. The RD® QUICK LOAD® SUTURE includes a detachable clear suture tube to keep the suture from tangling. The RD® QUICK LOAD® SUTURE is packaged for single patient use and is provided sterile by means of a validated ethylene oxide (EO) cycle. The polypropylene suture is dyed blue with the FDA approved colorant [phthalocyaninato(2-)] copper and will be offered with a surgical suture diameter of USP size 2-0 and in a length of 53 inches. The polyester suture is offered dyed green with the FDA approved colorant D&C Green No. 6 and is offered in a surgical suture diameter of USP size of 0 and 2-0 and a length of 53 inches. For both polypropylene and polyester suture, there is no known significant change in tensile strength retention to occur in vivo. The RD® QUICK LOAD® SUTURE products, with needle caps removed, contain implantable components which are considered MR (magnetic resonance) safe. The implantable polyester and polypropylene suture materials are non-metallic and nonconducting materials and therefore, considered MR safe.
This document does not describe a study involving an AI/Medical Device that requires the establishment of ground truth, expert consensus, or multi-reader multi-case studies. Instead, it is a 510(k) premarket notification for a surgical suture, the RD QUICK LOAD SUTURE, asserting its substantial equivalence to a previously cleared predicate device.
The acceptance criteria and proof of substantial equivalence for this suture device are based on non-clinical performance testing and biocompatibility assessment, not on clinical performance studies involving a test set, expert consensus, or multi-reader studies as one would expect for an AI/Medical Device.
Here's why and what information is provided:
1. Acceptance Criteria and Device Performance (for a Suture)
The acceptance criteria for the RD QUICK LOAD SUTURE are based on established standards for surgical sutures. The performance of the device is evaluated against these standards.
| Acceptance Criteria (for Suture) | Reported Device Performance (Summary) |
|---|---|
| Conformance to USP 42-NF37:2019 Monograph for Non-Absorbable Sutures (Physical Tests) | The RD® QUICK LOAD® SUTURE continues to conform to these requirements and test methods. |
| Biocompatibility in accordance with ISO 10993-1:2009(R)2013 | All appropriate biological endpoints for consideration per ISO 10993-1 were assessed with passing results. |
2. Sample Size and Data Provenance for Testing:
- Data provenance: The document states "Biocompatibility results from Teleflex® Medical OEM were leveraged." This implies testing was conducted by the raw material supplier.
- The specific sample sizes for the USP and biocompatibility testing are not explicitly stated in this document. These would typically be defined by the relevant standards and internal test protocols. The nature of this 510(k) summary means highly detailed test parameters are usually omitted, with the FDA reviewing the full test reports.
- The tests are non-clinical (benchtop and lab-based biocompatibility), not clinical.
3. Number of Experts/Qualifications (Not Applicable for Suture Evaluation):
This information is not relevant to the evaluation of a surgical suture's substantial equivalence based on physical and biocompatibility properties. There are no "experts" establishing ground truth in the context of image interpretation or diagnosis.
4. Adjudication Method (Not Applicable):
Not applicable for a medical device cleared through substantial equivalence based on physical and biological testing.
5. MRMC Comparative Effectiveness Study (Not Applicable):
- No MRMC study was done. This is a surgical suture, not a diagnostic AI device. There are no "human readers" to improve with AI assistance. Clinical studies were deemed unnecessary.
6. Standalone Performance (Not Applicable as AI):
This is not an AI/algorithm. Its performance is its tensile strength, knot security, and biocompatibility, which are tested through benchtop methods defined by USP standards.
7. Type of Ground Truth (for Suture Properties, not a Diagnostic):
- The "ground truth" for the suture's performance is established by validated physical test methods (USP standards) and biocompatibility testing against ISO standards.
- For the claim of substantial equivalence, the "ground truth" is also the performance of the predicate device, against which the new device is compared.
8. Sample Size for Training Set (Not Applicable - No AI/Machine Learning):
This device does not involve a "training set" in the context of AI or machine learning. It is a manufactured physical good.
9. How Ground Truth for Training Set was Established (Not Applicable):
Not applicable, as there's no training set for an AI model.
In summary, this FDA clearance document is for a traditional medical device (surgical suture) demonstrating chemical and physical equivalence to a predicate device, not an AI-powered diagnostic tool. Therefore, many of the requested criteria, which are standard for AI/ML device evaluations, are not applicable here.
Ask a specific question about this device
Page 1 of 7