Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K233974

    Validate with FDA (Live)

    Date Cleared
    2024-09-06

    (266 days)

    Product Code
    Regulation Number
    876.1450
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    20004

    Re: K233974

    Trade/Device Name: Mucosal Impedance Measurement System Regulation Number: 21 CFR 876.1450
    Prepared: August 6, 2024

    Name of Device: Mucosal Impedance Measurement System

    Classification Name: 876.1450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mucosal Impedance Measurement System is indicated for use by medical personnel trained in the usage of enteral feeding tubes for the treatment of adult patients requiring gastric decompression or enteral feeding for the delivery nutrition, fluids, and medications to the patient from an enteral feeding svringe or feedings set designed with ENFit connectors. The measurements of the Mucosal Impedance Measurement System are collected in patients indicated for enteral feeding. This product is single use for no longer than 24 hours.

    The Mucosal Impedance Measurement System is equipped with sensors designed to collect and display real time and continuous gastric reactance "XL" measurements. The device is not for use as a sole diagnostic screening tool and is restricted to prescription use in a hospital environment.

    Device Description

    The Mucosal Impedance Measurement System senses passive electrical features of the gastric tissue that corelate to of epithelial integrity through a technique known as 'bioimpedance spectroscopy'. Bioimpedance spectroscopy measures the ability of tissue to conduct electricity (resistance) and the ability of tissue to store charged particles (reactance).

    The Florence monitor model ISMO 1.0 is an electronic device that delivers electrical current to a small region of the gastric tissue through the Athena Catheter; the device measures the resultant voltage, and then calculates and displays impedance information in numeric and graphic representations. More specifically, to measure the bioimpedance of the gastric mucosa impedance, the device delivers electric current through the Athena Catheter. The resulting voltage developed between the two inner electrodes is sensed, filtered, amplified, and then digitized by the electronic circuits of the monitor. Once the measured voltage from the gastric tissue has been digitized, a proprietary algorithm processes the information and calculates the impedance spectrum and the Central Reactance parameter (XL). The device displays the information as either a function of frequency or as a function of time.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the studies performed for the Mucosal Impedance Measurement System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestMethods / Acceptance CriteriaReported Device Performance
    Flow rate"Gravity Flow Rate Testing in Enteral Tubes" as described in the "Catalog of Regulatory Science Tools to Help Assess New Medical Devices"The Mucosal Impedance Measurement System functioned as intended.
    Tensile strengthEN1615:2000 Enteral Feeding Catheters And Enteral Giving Sets For Single Use And Their Connectors. Design And Testing (Section 4.1.1) AND EN1618:1997 Catheters other than intravascular catheters - Test methods for common properties (Annex B)The Mucosal Impedance Measurement System functioned as intended.
    Liquid leakageEN1615:2000 Enteral Feeding Catheters And Enteral Giving Sets For Single Use And Their Connectors. Design And Testing (Sections 4.1.3) AND EN1618:1997 Catheters other than intravascular catheters - Test methods for common properties (Annex C)The Mucosal Impedance Measurement System functioned as intended.
    Security of connectorsEN1615:2000 Enteral Feeding Catheters And Enteral Giving Sets For Single Use And Their Connectors. Design And Testing (Sections 4.3.1) AND EN1618:1997 Catheters other than intravascular catheters - Test methods for common properties (Annex F)The Mucosal Impedance Measurement System functioned as intended.
    FlexibilityASTM F3505-21 Standard Test Method for Stent and Endovascular Prosthesis Kink ResistanceThe Mucosal Impedance Measurement System functioned as intended.
    EnFit connector (dimensional verification, rigidity, fluid leakage test, stress cracking, resistance to separation from axial load, resistance to unscrewing, resistance to overriding, and disconnection by unscrewing)ISO 80369-3:2016 Small-bore connectors for liquids and gases in healthcare applications Part 3: Connectors for enteral applicationsThe Mucosal Impedance Measurement System functioned as intended.
    Sterilization validationISO 11135:2014 Sterilization of health-care products Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devicesThe Mucosal Impedance Measurement System functioned as intended.
    Packaging validationISO 11607-1:2019 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systemsThe Mucosal Impedance Measurement System functioned as intended.
    Dimensional verificationOverall length, working length, catheter outer diameter, catheter inner diameter, sensor dimensions, and spacing of the sensors were measured.The Mucosal Impedance Measurement System functioned as intended.
    Accuracy - thermal stabilityConfirming that the measurements of the central reactance parameter (XL) remain consistent across a temperature range that covers 77-111°F (25-44°C). The PASS criterion was defined as confirming that at least 95% of all measurements are contained within a +/-1Ω tolerance.Stability of the sensors was confirmed in terms of the dispersion (standard deviation / interquartile range) of a statistically significant number of measurements across the temperature range. The device met the PASS criterion.
    Accuracy - temporal stabilityConfirming that the measurements of the central reactance parameter (XL) remain consistent across 24 hours. The PASS criterion was defined as confirming that at least 95% of all measurements are contained within a +/-1Ω tolerance (which corresponds to the accuracy stated in the Instructions for Use).Stability of the sensors was confirmed in terms of the dispersion (standard deviation / interquartile range) of a statistically significant number of measurements over the 24-hour period. The device met the PASS criterion.
    BiocompatibilityEvaluated according to FDA guidance (2023) "Use of International Standard ISO 10993-1," assessing toxicity, irritation, and sensitization.Results support the biocompatibility of the Athena Catheter.
    SoftwareA failure or flaw of the software functions does not present a hazardous situation with a probable risk of death or serious injury. (Basic documentation level identified per FDA Guidance).The software documentation was found to be at the basic level, implying it met this criterion.
    Electrical Safety and Electromagnetic Compatibility (EMC)Basic safety evaluation per the FDA consensus standard IEC-60601-1.2005, AMD1.2012. EMC testing per IEC-60601-1-2.2014.The test reports addressed basic safety and EMC testing, and passed the applicable clauses.
    Human factorsUsability validated in an actual clinical environment for intended users and environment, according to "Applying Human Factors and Usability Engineering to Medical Devices" guidance and EN 60601-1-6:2010, AMD1:2013.All evaluations have been performed according to the specified standards, supporting the safety and effectiveness of the device.
    Animal Study (Gastric Mucosal Perfusion)GLP prospective, controlled animal trial with a shock group (MAP ≤ 48 mmHg) compared to a control group, measuring XL and commonly used markers of global perfusion. Target: five continuous impedance measurements per subject/group for 5 hours.Demonstrated gastric impedance measurements were performed safely and XL measurement was an indirect and consistent marker of gastric mucosal perfusion in the subject animals, showing significant and detectable changes before commonly used markers of global perfusion under hypovolemic shock conditions. There were no adverse events.

    2. Sample Size Used for the Test Set and Data Provenance

    • Non-clinical/Bench Studies: The specific sample sizes for each bench test (e.g., flow rate, tensile strength) are not explicitly stated as numerical values in the provided text, but are implied by the use of standards that typically define such sample sizes. The text mentions "a statistically significative number of measurements" for thermal and temporal stability, indicating appropriate methodology.
    • Animal Study:
      • Control group (CG): N=5
      • Shock group (SG): N=16
      • Data Provenance: This was a GLP prospective, controlled animal trial. The country of origin of the data is not specified.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    For the non-clinical/bench studies, animal studies, and human factors, the "ground truth" is established by adherence to recognized international and FDA standards/guidances (e.g., ISO, EN, ASTM, FDA guidance documents). There is no mention of specific human experts establishing ground truth for these quantitative technical performance tests in the same way an expert would for an AI diagnostic algorithm.

    4. Adjudication Method for the Test Set

    Not applicable in the context of these device performance studies, which involve objective measurements against predefined engineering and biological performance criteria, rather than expert-based interpretation.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The provided text does not describe an MRMC comparative effectiveness study involving human readers with and without AI assistance. The device is described as measuring and displaying "XL" values, which are direct measurements, not an AI-based interpretation or diagnostic aid for human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, in essence. The internal performance evaluations (e.g., thermal stability, temporal stability, electrical safety, software validation) and the animal study tested the device's ability to accurately collect and process data and function as intended without human interpretation as the primary endpoint. The device's output (XL measurements) is for human interpretation and clinical decision-making, but the studies described focus on the accuracy and stability of the measurement itself rather than a diagnostic output of an AI algorithm. The device is "not for use as a sole diagnostic screening tool."

    7. Type of Ground Truth Used

    • Bench Studies: Established by defined physical properties, engineering specifications, and compliance with international standards (e.g., ISO, EN, ASTM).
    • Biocompatibility: Established by compliance with ISO 10993-1 and FDA guidance, indicating a lack of toxicity, irritation, and sensitization.
    • Software: Established by compliance with FDA guidance for software functions, ensuring no hazardous situations.
    • Electrical Safety & EMC: Established by compliance with IEC standards.
    • Human Factors: Established by compliance with EN 60601-1-6 and FDA guidance, ensuring usability for safety and effectiveness.
    • Animal Study: Established by measuring physiological parameters and comparing the device's XL measurements to established markers of global perfusion under induced hypovolemic shock conditions in a controlled, prospective GLP study.

    8. Sample Size for the Training Set

    The provided text does not mention a training set, as this device's performance evaluation focuses on the accuracy and stability of direct physiological measurements and its engineering compliance, rather than a machine learning model that requires training data. The "proprietary algorithm" processes measured voltage to calculate the impedance spectrum and Central Reactance parameter (XL), but there's no indication it's a machine learning algorithm requiring a separate training phase with labeled data in the context of this submission.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set for a machine learning model is described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230056

    Validate with FDA (Live)

    Date Cleared
    2023-04-25

    (106 days)

    Product Code
    Regulation Number
    876.1450
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Ranch, CO 80129

    Re: K230056 Trade/Device Name: MiVu™ Esophageal Endo Cap Regulation Number: 21 CFR 876.1450
    Classification Name: | Esophageal tissue characterization system |
    | Classification Number: | 21 CFR 876.1450
    Classification Name: | Esophageal tissue characterization system |
    | Predicate Regulation No.: | 21 CFR 876.1450
    | Identical |
    | FDAClassificationNumber | 876.1450
    | 876.1450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MiVu Mucosal Integrity Testing System is indicated for use by gastroenterologists, surgeons, and medical personnel trained in endoscopic procedures during an endoscopy to obtain real-time measurement of esophageal epithelial integrity as an adjunct for the evaluation of esophageal disorders. The device is not for use as a sole diagnostic screening tool.

    Device Description

    The Diversatek Healthcare MiVu Esophageal Endo Cap is the candidate accessory device for use with the approved predicate Diversatek Healthcare's MiVu Mucosal Integrity Testing system (MiVu System). The MiVu Esophageal Endo Cap is a new patient-contacting accessory that will be used in place of the MiVu Balloon Probe already approved as part of the MiVu System.

    The MiVu Esophageal Endo Cap device consists of a patient contacting sensor tip that is installed over the working end of an endoscope. The tip is a flexible thermoplastic elastomeric tube with a single row of four (4) gold plated pads, providing three (3) impedance sensors evenly spaced along the length of the impedance sensors are mounted on a pivoting platform to promote even contact of all sensor pads with the tissue. A polyurethane conduit tube containing interconnecting wires connects the individual impedance sensor pads on the sensor tip to the circuit board inside the connector housing. The connector is then attached to the remainder of the predicate MiVu System through the MiVu Endo Cap Cable. The MiVu System digitizes the impedance signals and passes the data to the PC-based Zvu software for display and analysis.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the MiVu™ Esophageal Endo Cap, a device intended to replace the MiVu™ Balloon Probe as an accessory for the MiVu Mucosal Integrity Testing System. The submission aims to demonstrate substantial equivalence to the predicate device.

    Here's an analysis of the acceptance criteria and study information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria in a table format for performance metrics (e.g., sensitivity, specificity, accuracy). Instead, the clinical study aims to demonstrate "clinically equivalent or better results" compared to the predicate device based on "disease probabilities."

    However, we can infer some criteria based on the testing performed:

    Acceptance Criterion (Inferred)Reported Device Performance
    Functional CompatibilityThe MiVu Esophageal Endo Cap and MiVu Endo Cap Cable were tested to verify compatibility with the MiVu Mucosal Integrity Testing System. Functional testing demonstrated the acquisition of impedance data and the authentication of the Endo Cap device.
    Equivalence to Predicate Function and PerformanceBench testing compared the MiVu System with the Endo Cap to the MiVu System with the predicate Balloon Probe using a test medium for true impedance readings. The implied acceptance is that the performance is equivalent. Clinical studies found the MiVu System with the MiVu Esophageal Endo Cap device provided clinically equivalent or better results than the MiVu Balloon Probe based on the disease probabilities generated by the MiVu System.
    Biocompatibility (ISO 10993-1)Complies with all applicable parts of the standard.
    Electrical Safety and EMC (IEC 60601-1, IEC 60601-1-6, IEC 60601-1-2)System configuration complies with the applicable requirements of these standards.
    Mechanical Safety (Tensile testing, pull test)Tensile testing performed according to ISO 10555-1:2013+A1:2017. A pull test of retention on a 9.9mm endoscope was performed. The implied acceptance is that the device meets mechanical safety requirements.
    Reprocessing Effectiveness (for reusable cable)Reprocessing instructions for the MiVu Endo Cap Cable were validated by a third-party testing lab.
    Bioburden and Packaging IntegrityDevices tested for bioburden levels pre- and post-shipping simulation to verify packaging effectiveness and mechanical integrity. The implied acceptance is that bioburden levels are acceptable and packaging is effective.
    Software Level of ConcernFirmware and software classified as "minor level of concern" (a failure will not pose any risk of harm to the patient or user).
    Clinical Equivalence/UsabilityAll clinical studies found the MiVu System with the MiVu Esophageal Endo Cap device provided clinically equivalent or better results than the MiVu Balloon Probe based on the disease probabilities generated by the MiVu System. No patient adverse effects or complications were reported.

    2. Sample size used for the test set and the data provenance

    • Sample Size:
      • Comparative Test Set: 17 random patients were tested using both the MiVu Esophageal Endo Cap and the predicate MiVu Balloon Probe.
      • Endo Cap Only Test Set: An additional 7 random patients were tested using only the MiVu Esophageal Endo Cap.
      • Total Patients in Clinical Study: 24 patients.
      • Patient Demographics: Adult age (20 to 70 years), 12 males, 8 females, 4 unknowns.
    • Data Provenance: Clinical studies were performed at one United States located facility. The study was prospective as patients were enrolled and tested within the context of the study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not explicitly state how ground truth for the test set was established or the number/qualifications of experts involved in determining the actual disease state (GERD, non-GERD, EoE) for the patients in the clinical study.

    The study compared the device's generated "disease probabilities" against those generated by the predicate device, implying the predicate device's output or a physician's interpretation of it might serve as a de facto reference point for "clinically equivalent or better results."

    It does mention that the clinical studies were performed by three gastroenterologists who were familiar with both MiVu System configurations. While these gastroenterologists performed the studies and interpreted the results, it's not explicitly stated if they established a definitive "ground truth" for each patient's actual disease. The system itself "generates a probability of the tissue disease type (GERD, non-GERD, EoE) which is used by the physician as an input to the treatment plan."

    4. Adjudication method for the test set

    The document does not specify any formal adjudication method (e.g., 2+1, 3+1) for the clinical study. The conclusion states that results were "clinically equivalent or better results than the MiVu Balloon Probe based on the disease probabilities generated by the MiVu System," suggesting a direct comparison of the outputs of both systems, possibly evaluated by the performing gastroenterologists.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: While the study involved three gastroenterologists and multiple cases, it was not a traditional MRMC comparative effectiveness study focused on "human reader improvement with/without AI assistance." The study's primary goal was to compare the performance of the new accessory (Endo Cap) to the predicate accessory (Balloon Probe) within the existing MiVu System, which already includes an algorithm for determining disease probabilities. The comparison was device-to-device, not human-with-AI vs. human-without-AI.
    • Effect Size of Human Readers: No such effect size is reported as this was not the objective of the study. The device's output ("disease probabilities") is an input for the physician's treatment plan, but the study doesn't isolate the impact of this input on a human reader's diagnostic accuracy.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the MiVu System, which includes the algorithm, generates "disease probabilities for GERD (Gastro-Esophageal Reflux Disease), non-GERD and EoE (Eosinophilic Esophagitis)." This is an algorithm-only output. The clinical study compares these probabilities generated when using the new Endo Cap versus when using the predicate Balloon Probe. The system's output is then "used by the physician as an input to the treatment plan," indicating the algorithm provides a standalone assessment.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The document primarily refers to comparing the "disease probabilities generated by the MiVu System" when using the new Endo Cap versus the predicate Balloon Probe. It does not explicitly state that a definitive, independent ground truth (e.g., biopsy/pathology results, long-term outcomes, or a gold-standard expert consensus) was established for each patient's condition to validate the accuracy of these probabilities themselves. The comparison is largely relative to the predicate device's performance within the same system.

    The study selected patients "based on the known or highly suspected esophageal disease type (GERD, non-GERD, EoE)," which implies some pre-existing diagnostic information, but whether this served as the formal "ground truth" for the device's output validation is not explicitly detailed.

    8. The sample size for the training set

    The document for the MiVu™ Esophageal Endo Cap does not provide information on the sample size for the training set used to develop the algorithm within the MiVu System. It only describes the clinical study for the new accessory and its comparison to the predicate. The algorithm for determining disease probabilities (GERD, non-GERD, EoE) was developed for the original MiVu System. The submission refers to the De Novo submission for the MiVu System (DEN180067) for details on the operating principle and changes in mucosal conductivity, which likely includes information about the algorithm's development.

    9. How the ground truth for the training set was established

    Similar to the training set size, the document for the MiVu™ Esophageal Endo Cap does not detail how the ground truth was established for the training set of the MiVu System's algorithm. This information would typically be found in the original De Novo submission (DEN180067) for the MiVu System, which introduced the core technology and its algorithm. The current submission focuses on the substantial equivalence of a new accessory, not the development of the underlying diagnostic algorithm itself.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN180067

    Validate with FDA (Live)

    Date Cleared
    2019-12-23

    (371 days)

    Product Code
    Regulation Number
    876.1450
    Type
    Direct
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    NEW REGULATION NUMBER: 21 CFR 876.1450

    CLASSIFICATION: Class II

    PRODUCT CODE: QIS

    BACKGROUND

    Product Code: QIS Device Type: Esophageal tissue characterization system Regulation Number: 21 CFR 876.1450

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Mucosal Integrity Conductivity Test System is indicated for use by gastroenterologists, surgeons, and medically trained personnel during an endoscopy to obtain a real time measurement of esophageal epithelial impedance. The device is not for use as a sole diagnostic screening tool.

    Device Description

    The Mucosal Integrity Conductivity (MI) Test System provides real-time measurements of alterations in esophageal evithelial integrity. The device is intended to aid in the evaluation of esophageal epithelial integrity by means of a balloon probe with direct electrical contact with the mucosal epithelium of the esophagus along with associated signal conditioning, hardware, and software for measuring and displaying information.

    AI/ML Overview

    The provided text describes the "Mucosal Integrity Conductivity (MI) Test System," an esophageal tissue characterization system. However, the document does not contain information about a study that assesses the device's performance against specific acceptance criteria using a test set of patient data or
    human readers in a comparative effectiveness study. The "Performance Testing - Bench" section details benchtop tests of the device's physical properties and electrical measurement capabilities, but this is distinct from a clinical performance study using patient data.

    The "Summary of Clinical Information" section briefly mentions two prospective clinical studies where the device was used in patients, noting its ability to measure esophageal mucosal impedance and the occurrence of few adverse events. However, these descriptions do not provide specific acceptance criteria, performance metrics (like sensitivity, specificity, accuracy), or details on how ground truth was established for diagnosis or classification of esophageal conditions based on the impedance measurements, which would typically be part of a study proving the device meets clinical acceptance criteria for its intended use as an aid in evaluation.

    Therefore,Based on the provided text, I cannot answer all parts of your request as it does not detail a study that proves the clinical performance of the device against specific acceptance criteria for diagnostic accuracy or human-in-the-loop improvement for clinical decision making. The information focuses on bench testing, safety, and general functionality.

    However, I can extract the acceptance criteria and performance for the benchtop performance tests.


    Acceptance Criteria and Reported Device Performance (Benchtop Testing)

    Here's a table summarizing the acceptance criteria and reported "Pass" results for the benchtop performance testing:

    TestAcceptance CriteriaReported Device Performance
    Balloon Functionality Testing
    Double Wall Thickness.00045" +/- .00025"Pass
    Compliance Diameter @ 0.5atm20 mm +/- .65 mmPass
    Burst Pressure2.04 atm minimumPass
    Dimensional Design Verification Testing
    Probe Overall Length55.25" +/- 2.0"Pass
    Probe, Balloon Diameter when Inflated2.0 cm +/- 0.1 cmPass
    Probe Weight2.8 oz. +/- 0.5 oz.Pass
    Bond between Sensors and Balloon Testing
    Adhesive Strength after Exposure to Saline≥ adhesive strength without saline exposure2.267 lbfPass
    Tensile Strength, Distal Tip to Catheter of MI Probe
    Tensile Strength Distal Tip Joint MI Probe (Initial)≥ 3.0 lbf based on independent studyPass
    Tensile Strength Distal Tip Joint MI Probe (Added to Submission)≥ 4.5 lbf from EN 1617 (20 Newtons)Pass
    Impedance Measurements of MI Probe Sensors
    Impedance Measurement Accuracy174+/- 100 OhmsPass
    Impedance Signal Noise< 20 OhmsPass
    Impedance Thermal Stability+/- 100 OhmsPass
    Impedance Temporal Stability+/- 100 OhmsPass

    Since the document primarily describes benchtop performance and safety testing, and general clinical use observations rather than a clinical performance study with diagnostic accuracy outcomes, the following information is not available or not applicable based on the provided text:

    • Sample size used for the test set: Not specified for a clinical performance test set. The bench tests specify n=10 for balloon functionality, n=33 for some dimensional tests, n=1 for impedance accuracy tests, and n=4 or n(g) for tensile strength.
    • Data provenance (e.g. country of origin of the data, retrospective or prospective): The clinical studies mentioned are stated to be "prospective," but no country of origin is specified for the data.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not specified for any clinically relevant ground truth.
    • Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not specified.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No such study is mentioned or implied, as the device gives a direct measurement and is not described as an AI assistance tool for human readers.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The device directly measures impedance, it's not described as an algorithm making a diagnostic decision in a standalone fashion. Its purpose is to obtain a real time measurement to aid in the evaluation.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc): While one clinical study mentions patients with "active eosinophilic esophagitis (EoE, confirmed with pathology)," the document doesn't explicitly state that pathology was used as the ground truth for evaluating the device's performance against specific clinical endpoints derived from impedance measurements. The device simply measured impedance in these patients.
    • The sample size for the training set: Not applicable as there's no mention of a machine learning model requiring a training set for diagnostic classification in the provided text.
    • How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1