Found 2 results
510(k) Data Aggregation
(269 days)
San Diego, California 92110
Re: K242214
Trade/Device Name: Sonu
Regulation Number: 21 CFR 874.6010
Common Name: External mechanical stimulator for the relief of congestion
Regulation Number: 21 CFR 874.6010
Sonu is indicated for the relief of moderate to severe nasal congestion due to allergic and non-allergic rhinitis. Sonu is a treatment to be used at home by individuals 12 and older.
Sonu is a non-invasive, over the counter (OTC) device designed for the relief of moderate to severe nasal congestion due to allergic and non-allergic rhinitis. Sonu uses acoustic resonance therapy to achieve its therapeutic effect. Sonu consists of an adjustable headband (Sonu Band) with integrated acoustic bone-conduction transducers (hardware), a USB-C Charging Cable (hardware), and a smartphone application (Sonu iOS App) (software) that connects to the Sonu Band. Sonu is indicated for individuals, aged 12 years old and above.
The provided text is related to an FDA 510(k) clearance for the Sonu device, which is an external mechanical stimulator for the relief of nasal congestion. The clearance is specifically for expanding the indications for use to include the pediatric population aged 12-21 years old.
However, the document does not explicitly state specific acceptance criteria in a quantitative manner (e.g., "device must achieve X accuracy"). Instead, it presents clinical study results and concludes that the device is substantially equivalent to its predicate. The "reported device performance" in this context refers to the outcomes of the clinical study, which are used to support this substantial equivalence.
Here's the information extracted and structured based on your request, with an emphasis on what's explicitly stated and what is inferred from the available text:
Acceptance Criteria and Device Performance
Since explicit, quantitative acceptance criteria are not provided in the document, the "acceptance criteria" are inferred from the demonstrated clinical improvement needed to support substantial equivalence for the expanded pediatric population. The device's performance is reported as the statistically significant improvement in the Total Nasal Symptom Score (TNSS) and its nasal congestion sub-score.
| Acceptance Criteria (Inferred from Study Success) | Reported Device Performance |
|---|---|
| Statistical evidence of improvement in nasal congestion for pediatric patients (12-21 years old). | Mean change in Total Nasal Symptom Score (TNSS): -2.42 for 12 to <16 year old group -3.03 for 16 to <22 year old group -2.74 overall for both groups |
| Mean change in nasal congestion sub-score: -0.70 for 12 to <16 year old group -0.81 for 16 to <22 year old group -0.76 overall for both groups | |
| No adverse events reported. | No adverse events were reported in both groups. |
Additional Study Information:
-
Sample size used for the test set and the data provenance:
- Sample Size: 31 subjects (15 for 12 to <16 years old, 16 for 16 to <22 years old).
- Data Provenance: Not explicitly stated, but the study was described as an "interventional study," suggesting prospective data collection. The country of origin is not specified.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the document. The study uses patient-reported scores (TNSS and nasal congestion sub-score) as outcome measures, which are subjective assessments by the subjects themselves rather than an expert-established "ground truth" in the traditional sense (like expert review of imaging).
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- This information is not provided as the study did not involve expert adjudication for establishing a ground truth.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- A multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This study design is typically used for diagnostic devices involving human interpretation (e.g., radiologists reading images) and comparing performance with and without AI assistance. The Sonu device is a therapeutic device, and the clinical study focuses on treatment effectiveness.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The Sonu device itself is the standalone therapeutic device; its performance is measured by its direct effect on patient symptoms. The study evaluates the device's therapeutic effect directly in subjects. While there is a "Sonu iOS App" (software), its function is to control the device, and the effectiveness is tied to the complete system, not just the software in isolation in a diagnostic context. Therefore, a "standalone" algorithm-only performance study in the typical diagnostic AI sense was not performed or relevant.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" (or primary outcome measures) used for the clinical study consisted of patient-reported outcomes data, specifically the Total Nasal Symptom Score (TNSS) and its nasal congestion sub-score. These are subjective measures of symptom severity reported by the subjects themselves.
-
The sample size for the training set:
- This information is not applicable and not provided. The clinical study described is an interventional study to assess the device's therapeutic effect, not a study to train a machine learning algorithm. The software component (Sonu iOS App) indicates "software development and testing was executed in compliance with IEC 62304," but this refers to software validation, not algorithm training data.
-
How the ground truth for the training set was established:
- This information is not applicable as there was no explicit "training set" for a machine learning algorithm described in the clinical study.
Ask a specific question about this device
(171 days)
NEW REGULATION NUMBER: 21 CFR 874.6010
CLASSIFICATION: Class II
PRODUCT CODE: QZC
BACKGROUND
Sonu is indicated for the relief of moderate to severe nasal congestion due to allergic and non-allergic rhinitis. Sonu is a treatment to be used at home by individuals 22 and older.
Sonu is a non-invasive, over-the-counter (OTC) device designed for the relief of moderate to severe nasal congestion due to allergic and non-allergic rhinitis at home. Sonu consists of an adjustable headband (Sonu Band) with integrated acoustic bone-conduction transducers, a USBC Charging Cable, and a smartphone application (Sonu iOS App) that connects to the Sonu Band.
Here's an analysis of the Sonu device's acceptance criteria and the study proving it meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
|---|---|---|
| Primary Safety | Absence of Serious Adverse Events (SAEs) related to the device or treatment. | Achieved: 0 Device or Procedure-Related SAEs reported in 52 subjects over 1234 total treatments. |
| Secondary Safety | Absence of Adverse Events (AEs) related to the device or treatment. | Achieved: 0 Device or Procedure-Related AEs reported in 52 subjects over 1234 total treatments. |
| Primary Effectiveness | Statistically significant improvement in nasal congestion sub-score of TNSS, comparing baseline congestion to daily congestion averaged over 2 weeks (Sonu vs. Sham), exceeding a sponsor-defined Minimal Clinically Important Difference (MCID). | Achieved: Mean change in nasal congestion sub-score was -0.87 for Sonu Therapy vs. -0.44 for Sham (p=0.008). This difference (0.43, adjusted to 0.47 with 95% CI 0.15, 0.79, p=0.005) was statistically significant and exceeded the MCID. The nasal congestion sub-score decreased by 33% in Week 1 and 44.5% in Week 2 for Sonu Therapy vs. 20.3% and 19.3% for Sham respectively, with statistical significance by Week 1 (p=0.030) and Week 2 (p=0.008). |
| Secondary Effectiveness | Statistically significant improvement in 24-hour reflective TNSS comparing baseline TNSS to daily TNSS averaged over 2 weeks (Sonu vs. Sham). | Achieved: Mean change in composite TNSS was -2.85 for Sonu Therapy vs. -1.32 for Sham (p=0.027). The difference (1.53, adjusted to 1.80 with 95% CI 0.46, 3.13, p=0.009) was statistically significant. TNSS decreased by 38% in Week 1 and 50.1% in Week 2 for Sonu Therapy vs. 20.2% and 22.5% for Sham, reaching statistical significance by Week 2 (p=0.02). |
| Biocompatibility | Patient-contacting components must be demonstrated to be biocompatible. | Achieved: Inner headband (Polyester, polyurethane) and transducer covers (medical grade silicone rubber) listed as materials with a long history of safe use in medical devices that contact intact skin, posing very low biocompatibility risk. No exclusion characteristics apply. |
| Electromagnetic Compatibility & Safety | Device must demonstrate electromagnetic compatibility, battery safety, and electrical safety. | Achieved: EMC tests (IEC 60601-1-2, 47 CFR FCC Part 15 Subpart B) conducted by SGS Taiwan Ltd. showed Sonu is electromagnetically compatible, electrically safe, and has a safe battery. |
| Software Performance | Software verification, validation, and hazard analysis must be performed. | Achieved: Documentation provided for Software Level of Concern (Minor), description, hazard analysis, requirements specification, traceability, V&V, and revision history. Cybersecurity assessment also conducted. |
| Non-clinical Performance | Device must perform as intended under anticipated conditions of use, including verification of specified mechanical stimulation parameters. | Achieved: 100% of 60 manufactured units passed battery and functional pairing tests. Audio quality and volume changes at max and 50% settings were verified. Therapeutic resonance measurement was verified. |
| Risk Mitigation (General) | Mitigation of risks such as injury from mechanical overstimulation (skin irritation, pain, discomfort, headache, vertigo, hearing loss, tinnitus) and ineffective treatment. | Achieved: Mitigation measures include non-clinical performance testing, human factors testing, software V&V and hazard analysis, electrical safety testing, EMC testing, battery safety testing, and labeling. |
| Human Factors | Human factors testing must demonstrate that users can successfully use the device in the intended use environment based solely on its labeling and instructions for use. | Achieved: Usability study with 25 participants (22+ years old) for OTC treatment of moderate to severe nasal congestion. 100% strongly agreed or agreed that instructions were easy to follow and the device was easy to use. No neutral/negative responses or observations of difficulty by the observer. |
2. Sample Size and Data Provenance for the Test Set
- Sample Size (Clinical Study - SCORE): 52 subjects (26 Sonu Therapy, 26 Sham).
- Data Provenance: Prospective, multi-center, randomized, double-blinded, sham-controlled interventional study. The country of origin of the data is not explicitly stated, but the mention of a US-based sponsor ("Sound Health Systems, Inc.") and the FDA De Novo request implies the study was likely conducted in the US or under US regulatory oversight.
3. Number of Experts and their Qualifications for Ground Truth of the Test Set
The provided text describes a clinical trial where patient-reported outcomes were used as the primary and secondary endpoints.
- 0 Experts: No external experts were used to establish the "ground truth" for the test set. Instead, the ground truth was based on the subjects' self-reported symptom scores using a validated scoring system (TNSS).
4. Adjudication Method for the Test Set
- None (implicitly): Since the primary and secondary outcomes were patient-reported self-assessments (nasal congestion sub-score and TNSS), there was no mention of an independent adjudication method for these scores. The blinding of both subjects and potentially study staff (though not explicitly stated for staff beyond the "double-blinded" claim) would have been the main control against bias.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, this was not an MRMC study. The study evaluated the direct effectiveness of the device itself (Sonu Therapy vs. Sham) and did not involve human readers interpreting imaging or other data with and without AI assistance. Therefore, there is no effect size of human readers improving with AI vs. without AI.
6. Standalone (Algorithm Only) Performance Study
- No, a standalone (algorithm only) performance study was not explicitly conducted or reported for the primary effectiveness of decongestion. The device functions as a complete system (Sonu Band + App + Cloud Engine) that delivers treatment to the human user.
- The "Sonu Cloud Engine" does utilize an AI/ML algorithm to estimate patient-specific resonant frequencies based on 3D face scans. A preliminary version of the device ("Soniflow System") was evaluated in a human cadaver model to verify the resonance frequency calculation algorithm. This could be considered a form of standalone performance evaluation for that specific algorithmic component, but not for the overall decongestion claim. The text does not detail the acceptance criteria or results for this cadaver study beyond stating it was done for "verification."
7. Type of Ground Truth Used
- Patient-Reported Outcomes (via validated scoring system): The primary and secondary effectiveness endpoints relied on subjects' self-reported nasal congestion sub-scores and Total Nasal Symptom Scores (TNSS). The TNSS is described as a "validated symptom severity scoring system."
8. Sample Size for the Training Set
- Not explicitly stated for the AI/ML algorithm. The text mentions the Sonu Cloud Engine uses an "Artificial Intelligence / Machine Learning algorithm that has been trained to correlate craniofacial measurements with sinus dimensions." However, the size and nature of this training dataset are not provided.
- For the clinical effectiveness study (SCORE), no dedicated "training set" in the machine learning sense is applicable, as this was an interventional clinical trial evaluating the device in humans, not training an AI for diagnosis or treatment.
9. How the Ground Truth for the Training Set was Established
- Not explicitly stated for the AI/ML algorithm's training set. For the AI/ML algorithm that estimates patient-specific resonant frequencies, the text implies it was trained to "correlate craniofacial measurements with sinus dimensions." This would logically require a dataset where both craniofacial measurements (presumably from 3D scans) and actual sinus dimensions (e.g., from CT scans or other anatomical imaging) were available. However, the method for establishing this "ground truth" (e.g., expert anatomical measurements, verified imaging data) for the AI training is not described in the document.