Found 11 results

510(k) Data Aggregation

    K Number
    K253612

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2026-02-19

    (93 days)

    Product Code
    Regulation Number
    874.4180
    Age Range
    8 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Device Name:** Acclarent AERA Eustachian Tube Balloon Dilation System
    Regulation Number: 21 CFR 874.4180
    To meet special control 21 CFR 874.4180 (b)(6), simulated use cadaver training is offered to all clinicians
    suitability and equivalence of the proposed alternative training methods to meet special control 21 CFR 874.4180
    To meet special control 21 CFR 874.4180 (b)(6), simulated use cadaver training is offered to all clinicians
    head model as an alternative to a cadaver model, were provided to fulfil the special control of 21 CFR 874.4180

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACCLARENT AERA® Eustachian Tube Balloon Dilation System is intended to dilate the Eustachian tube for treatment of persistent Eustachian tube dysfunction in patients ages 18 and older.

    For patients ages 8-17 years, the ACCLARENT AERA® Eustachian Tube Balloon Dilation System, alone or in combination with adjunctive procedures, is intended to treat patients with objective signs of persistent obstructive Eustachian tube dysfunction from inflammatory pathology, resulting in chronic otitis media with effusion and are refractory to at least one surgical intervention for persistent obstructive Eustachian tube dysfunction.

    Device Description

    The ACCLARENT AERA® Eustachian Tube Balloon Dilation System includes a Eustachian Tube Balloon Catheter and Guide Catheter designed specifically for use in accessing and dilating the Eustachian Tube. The system is used with the following additional devices: the Acclarent® SE Inflation Device (or Acclarent® Balloon Inflation Device). The Acclarent® SE Inflation Device or Acclarent® Balloon Inflation Device is used to inflate the balloon. All devices are provided sterile for single-patient use. The ACCLARENT AERA® Eustachian Tube Balloon Dilation System includes a 6x16mm (inflated diameter x length) flexible Balloon Catheter with an integrated shaft and a nylon balloon at the distal tip. The non-compliant balloon is designed to dilate the cartilaginous portion of the Eustachian tube. The shaft consists of dual lumen tubing with an actuator component that is designed to assist in rotation and advancement of the balloon catheter. The balled catheter tip (aka, "blueberry tip") on the balloon catheter is designed to restrict advancement of the device into the bony portion of the Eustachian tube, known as the isthmus. There is an endoscopic marker placed at the proximal taper of the balloon to aid in positioning under direct endoscopic visualization.

    The AERA Guide Catheter is anatomically designed to facilitate AERA Balloon Catheter access to the Eustachian tube. The AERA Guide Catheter incorporates an atraumatic distal tip and distal angled tip profile that facilitates access to the Eustachian tube. The Guide Catheter supplied with the ACCLARENT AERA® Eustachian Tube Balloon Dilation System contains a lubricious inner liner to allow smooth passage for the balloon catheter and includes a hypertube for rigidity.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K230742

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-12-13

    (271 days)

    Product Code
    Regulation Number
    874.4180
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K230742

    Trade/Device Name: ACCLARENT AERA Eustachian Tube Dilation System Regulation Number: 21 CFR 874.4180
    |
    | Regulation Number: | 21 CFR 874.4180

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACCLARENT AERA® Eustachian Tube Balloon Dilation System is intended to dilate the Eustachian tube for treatment of persistent Eustachian tube dysfunction in patients ages 18 and older.

    For patients ages 8-17 years, the ACCLARENT AERA® Eustachian Tube Balloon Dilation System, alone or in combination with adjunctive procedures, is intended to treat patients with objective signs of persistent obstructive Eustachian tube dysfunction from inflammatory pathology, resulting in chronic otitis media with effusion and are refractory to at least one surgical intervention for persistent obstructive Eustachian tube dysfunction.

    Device Description

    The ACCLARENT AERA® Eustachian Tube Balloon Dilation System includes a Eustachian Tube Balloon Catheter and Guide Catheter designed specifically for use in accessing and dilating the Eustachian Tube. The system is used with the following additional devices: the Acclarent® SE Inflation Device (or Acclarent® Balloon Inflation Device). The Acclarent® SE Inflation Device or Acclarent® Balloon Inflation Device is used to inflate the balloon. All devices are provided sterile for single-patient use. The ACCLARENT AERA® Eustachian Tube Balloon Dilation System includes is a 6x16mm (inflated diameter x length) flexible Balloon Catheter with an integrated shaft and a nylon balloon at the distal tip. The non-compliant balloon is designed to dilate the cartilaginous portion of the Eustachian tube. The shaft consists of dual lumen tubing with an actuator component that is designed to assist in rotation and advancement of the balloon catheter. The balled catheter tip (aka, “blueberry tip") on the balloon catheter is designed to restrict advancement of the device into the bony portion of the Eustachian tube, known as the isthmus. There is an endoscopic marker placed at the proximal taper of the balloon to aid in positioning under direct endoscopic visualization. The AERA Guide Catheter is anatomically designed to facilitate AERA Balloon Catheter access to the Eustachian tube. The AERA Guide Catheter incorporates an atraumatic distal tip and distal angled tip profile that facilitates access to the Eustachian tube. The Guide Catheter supplied with the ACCLARENT AERA® Eustachian Tube Balloon Dilation System contains a lubricious inner liner to allow smooth passage for the balloon catheter and includes a hypotube for rigidity.

    AI/ML Overview

    The Acclarent AERA Eustachian Tube Balloon Dilation System is intended to dilate the Eustachian tube for the treatment of persistent Eustachian tube dysfunction. The expanded indication covers patients aged 8-17 years with objective signs of persistent obstructive Eustachian tube dysfunction from inflammatory pathology, resulting in chronic otitis media with effusion and refractory to at least one surgical intervention.

    Here's the breakdown of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance and Validation
    Mechanical PerformanceDimensional attributes, cycle fatigue, balloon burst, bond separation.Previously submitted bench testing met all acceptance criteria for these attributes. No new testing for mechanical characteristics was performed for this submission as the technological characteristics of the device are identical to the predicate device.
    BiocompatibilityDevice materials are biocompatible.Testing has shown that the ACCLARENT AERA® Eustachian Tube Balloon Dilation System is biocompatible.
    SterilizationSterility assurance level (SAL) of 10^-6.The sterilization process was validated per AAMI/ANSI/ISO 11135-1: 2007, using the overkill (half-cycle approach) in a fixed chamber, and demonstrated a sterility assurance level of 10^-6.
    Ethylene Oxide ResidualsMeet ISO 10993-7:2008 requirements.Testing of ethylene oxide residuals met ISO 10993-7:2008 requirements.
    Packaging Shelf LifeOne-year shelf life.Packaging shelf life was established at one year via accelerated aging per ASTM F1980-07.
    Safety (Pediatric)Acceptable adverse event rate in pediatric patients (8-17 years old) with persistent obstructive Eustachian tube dysfunction from inflammatory pathology, resulting in chronic otitis media with effusion, and refractory to at least one surgical intervention.A review of safety data in the expanded pediatric population (8-17 years) found the device to be safe with no serious adverse events reported after ETBD. A minor AE rate of 8.2% of patients in this expanded age group was observed. This evidence was derived from real-world evidence, including published literature and data provided by established physicians.
    Efficacy (Pediatric)Improvement in tympanometry and failure-free rate in pediatric patients (8-17 years old) with persistent obstructive Eustachian tube dysfunction from inflammatory pathology, resulting in chronic otitis media with effusion, and refractory to at least one surgical intervention.Efficacy data from 20 ears in patients aged 8-17 years with Chronic Otitis Media with Effusion (COME) showed an improvement in tympanometry in 75% of procedures. A total of 84% of ears were failure-free after a mean of 2.7 years of follow-up (failure defined as requiring further surgery for ETD, e.g., tympanostomy tube insertion or revision ETBD). This evidence was derived from real-world evidence, including published literature and data provided by established physicians, and supports that ETBD is comparable, if not superior, to tympanostomy tube placement alone in the treatment of chronic otitis media with effusion in the pediatric population.
    Anatomical EquivalenceEustachian tube anatomy in the expanded patient population (8-17 years) to be similar to the predicate age group (18+ years).Equivalence of anatomy was demonstrated through a systematic literature review including several articles evaluating Eustachian tube anatomy. Based on CT scan reviews, the anatomy of the target pediatric patient population (8-17 years) was found to be similar to that of the predicate adult age group (18+) with no issues of safety raised.

    2. Sample Sizes and Data Provenance:

    • Test Set (for pediatric expansion):

      • Clinical Data: Efficacy data was reported from 20 ears of patients aged 8-17 years with Chronic Otitis Media with Effusion. Safety data encompassed a broader analysis across the expanded pediatric age group (8-17 years), without a specific patient count provided for the 8.2% minor AE rate, but it was derived from real-world evidence.
      • Anatomical Equivalence: "Several articles" based on CT scan reviews were used. No specific number of patients or scans are provided for this.
      • Provenance: "Real-world evidence, with data from published literature and data provided by established physicians" was analyzed. It is retrospective in nature, as it relies on existing published literature and real-world clinical experience. The country of origin of the data is not specified but is generally understood to be diverse given the nature of a systematic literature review and physician-provided data.
    • Training Set: The document does not explicitly describe a separate "training set" in the context of an AI/ML algorithm requiring such a set. The performance data for the device (mechanical, biocompatibility, sterilization, etc.) relies on previously submitted bench testing and validations, which are not typically referred to as training sets. The clinical data for the pediatric expansion relies on real-world evidence and a systematic literature review, which also does not fit the typical definition of a training set for an AI/ML model.

    3. Number of Experts and Qualifications for Ground Truth:

    • For Clinical Data (Pediatric Expansion): The document mentions "data provided by established physicians." It does not specify the number or specific qualifications (e.g., years of experience) of these individual physicians, nor does it detail the experts involved in conducting the systematic literature review.
    • For Anatomical Equivalence: Similarly, the CT scan reviews were used as evidence for anatomical equivalence, based on "several articles." The experts performing these CT scan reviews or evaluating the articles are not explicitly identified or qualified.

    4. Adjudication Method for the Test Set:

    • The document does not describe an explicit adjudication method (e.g., 2+1, 3+1) for the clinical data or anatomical equivalence evaluations for the pediatric expansion. The data from published literature and physician-provided data likely reflect outcomes as recorded and reported in their respective studies or practices.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The ACCLARENT AERA® Eustachian Tube Balloon Dilation System is a physical medical device for dilation, not an AI or imaging diagnostic software that would typically involve human readers interpreting cases with and without AI assistance. Therefore, an effect size of human readers improving with AI vs. without AI assistance is not applicable.

    6. Standalone (Algorithm Only) Performance Study:

    • No, a standalone (algorithm only) performance study was not done. As noted above, this device is a physical medical device (balloon dilation system), not a software algorithm that would have standalone performance.

    7. Type of Ground Truth Used:

    • For the pediatric efficacy data, the ground truth was based on:
      • Clinical Outcomes/Measurements: Improvement in tympanometry (an objective measure of middle ear function).
      • Failure-Free Rate: Defined as not requiring further surgery for ETD (e.g., tympanostomy tube insertion or revision ETBD), which is an outcome-based measure.
    • For the pediatric safety data, the ground truth was based on:
      • Reported Adverse Events: Classification of events as serious or minor, as observed in real-world clinical practice and literature.
    • For anatomical equivalence, the ground truth was based on:
      • CT Scan Reviews: Morphometric analysis of Eustachian tube anatomy as reported in published studies.

    8. Sample Size for the Training Set:

    • As mentioned in section 2, the document does not describe a "training set" in the context of AI/ML. All referred performance data are either previously established bench tests for the device itself or real-world clinical evidence/literature reviews for the expanded indication.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as a conventional "training set" with established ground truth for an AI/ML model is not described in this submission. The "ground truth" for the device's functional and safety characteristics was established through standard engineering bench testing and clinical data collection/review methodologies relevant to medical devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223542

    Validate with FDA (Live)

    Date Cleared
    2023-08-03

    (251 days)

    Product Code
    Regulation Number
    874.4180
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Germany

    Re: K223542

    Trade/Device Name: TubaVent Balloon Dilatation System Regulation Number: 21 CFR 874.4180
    | Eustachian Tube Balloon Dilation System |
    | Device Regulation Number | 21 CFR 874.4180
    | Eustachian Tube Balloon Dilation System |
    | Device Regulation Number | 21 CFR 874.4180
    A human cadaver study was executed in order to fulfil the specialcontrol of 874.4180(b)(iv)(6

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TubaVent balloon dilatation system is intended to dilate the cartilaginous portion of the Eustachian tube for treatment of persistent obstructive Eustachian tube dysfunction in patients 18 years and older.

    Device Description

    The TubaVent balloon dilatation system is an Eustachian Tube Balloon Dilatation System consisting of the balloon catheter, the Tubalnsert insertion device and the inflation device. It is used for dilation of the cartilaginous portion of the Eustachian tube in persistent obstructive Eustachian tube dysfunction. The balloon catheter TubaVent is available in two versions where the balloon diameter is different: 3 and 4.5 mm. The insertion device Tubalnsert has an angle of 45° at the distal end.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the TubaVent Balloon Dilatation System:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text details performance tests that were conducted and states that the device "met all acceptance criteria." However, it does not provide a specific table outlining the acceptance criteria values themselves (e.g., "burst pressure > X bar") alongside the measured performance. Instead, it describes the types of tests performed and a general statement of compliance.

    Acceptance Criteria CategoryReported Device Performance
    Biocompatibility- All tests passed (ISO 10993-5, -10, -11, -23).
    - Not cytotoxic, does not cause sensitization, irritation, or acute systemic toxicity.
    Performance- All performance tests demonstrated device performs as intended.
    - Tests included dimensional verification, deflation time, balloon burst pressure, tensile strength, torque, and fatigue.
    - Simulation studies (head model, animal cadaver - sheep) ensured function per design specs and intended use.
    - Human factors validation testing showed safe use by intended users without serious errors.
    Shelf Life- TubaVent balloon dilatation system: 3 years
    - TubaInsert insertion device: 5 years
    - Tuba inflation devices: 3 years
    Sterilization- Achieves Sterility Assurance Level (SAL) of 10-6.
    Design Validation- Performed in a human cadaver study.

    2. Sample Size Used for the Test Set and Data Provenance

    • Human Cadaver Study (Design Validation): The text mentions a "human cadaver study" was executed for design validation and to fulfill a special control requirement (874.4180(b)(iv)(6)). It does not specify the sample size (number of cadavers or Eustachian tubes tested) or the provenance (country of origin, retrospective/prospective).
    • Simulation Studies: Acknowledges "simulation studies in a head model and in an animal cadaver model (sheep head)" but does not specify sample sizes for these.
    • Human Factors Validation Testing: No sample size or details about the participants are provided.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not describe the establishment of a "ground truth" for a test set in the context of expert consensus, pathology, or outcomes data. The studies mentioned (biocompatibility, performance, shelf life, human cadaver, human factors) are primarily engineering and safety evaluations rather than diagnostic accuracy or clinical effectiveness studies requiring expert-adjudicated ground truth.

    4. Adjudication Method for the Test Set

    Not applicable, as the clinical study type requiring such adjudication (e.g., diagnostic accuracy where experts review cases) was not performed or detailed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This device is a medical instrument (balloon dilatation system), not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was mentioned.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a physical medical instrument, not an algorithm.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The document primarily relies on engineering specifications, biological safety testing (biocompatibility), and functional performance data. For the human cadaver study, the "ground truth" would likely be the physical integrity and proper function of the device within the anatomical context, assessed through direct observation and measurements, rather than an expert consensus on a pathological finding or clinical outcome.

    8. The Sample Size for the Training Set

    Not applicable, as this device does not involve machine learning or AI models with training sets.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as this device does not involve machine learning or AI models with training sets.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230065

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-05-26

    (137 days)

    Product Code
    Regulation Number
    874.4180
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    VenSure™ Nav Balloon Dilation System, VenSure™ ET Balloon Dilation Svstem Regulation Number: 21 CFR 874.4180
    --|--------------------------------------------------|
    | Classification: | Class II per 21 CFR 874.4180
    |
    | Classification Name: | Eustachian Tube Balloon System (21 CFR 874.4180) |
    |
    predicate is the predicate device for the added Eustachian Tube indication with product code PNZ, CFR 874.4180
    Product code: PNZ, LRC | Class II21 CFR 874.442021 CFR 874.4180

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VenSure™ Balloon Dilation System is used to access and treat the frontal ostia/recesses, sphenoid sinus ostia and maxillary ostia/ethmoid infundibula in adults using a trans-nasal approach. The bony sinus outflow tracts are remodeled by balloon displacement of adjacent bone and paranasal sinus structures.
    To dilate the cartilaginous portion of the Eustachian tube for treating persistent Eustachian in adult patients using a transnasal approach.
    The VenSure™ Nav Balloon Dilation System is additionally intended for use in conjunction with the Cube Navigation System during ENT procedures when surgical navigation or image-guided surgery may be necessary to locate the Eustachian tube or to locate tissue, bone or cartilaginous tissue surrounding the drainage of frontal, maxillary, and sphenoid sinuses to facilitate dilation of the sinus ostia.
    The VenSure™ Light Balloon Dilation System is additionally used to locate, illuminate within, and transilluminate across, nasal and sinus structures in adults.

    Device Description

    The VenSure Balloon Dilation Systems combine features of a malleable suction and a malleable probe with the tissue expansion effect of balloon dilation. The distal end of the device includes an atraumatic tip and can be shaped to fit the Frontal, Maxillary, Sphenoid sinuses, and Eustachian tube using the Bending Tool provided with the device. Since the distal end of the device is re-shapeable, one balloon can be modified to work on multiple sinuses and the Eustachian tube within the same patient. The devices come in an EO sterilized tray sealed inside of a Tyvek pouch with the bending tool, inflation device, and extension line included in the trav.
    All VenSure versions enable a physician to track the device into the sinuses and eustachian tube using endoscopic visualization.
    The VenSure™ Nav allows additionally for image-guided visualization when connected to the Cube Navigation System (manufactured by Fiagon). The VenSure™ Nav contains an integrated sensor carrier that enables the use of image guidance through "blug and play" tracking capability when used with the Fiagon Navigation System. The sensor carrier containing localizer elements detects a signal within a low-energy magnetic field delivered from the navigation unit. The navigation software then displays the location of the dilation instrument's tip within multiple patient image planes and other anatomic renderings. After confirmation of placement, the balloon of the dilation device can be inflated with saline solution, using the inflation device to expand the outflow track of the targeted structure.
    The VenSure™ Light additionally allows for LED light confirmation of the VenSure balloon through transillumination across nasal and sinus structures. The VenSure™ Light has an integrated flexible light fiber with battery powered LED light source designed to emit red light from the distal end of the VenSure balloon.
    The VenSure™ ET is a 45°- pre-bent configuration of the basic VenSure (with straight balloon tip). The pre-bend to 45°bemt of the VenSure ET facilitates the use of the device when only used for treatment in the eustachian tube. The device however can also be reshaped to fit to the sinuses using the Bending tool.

    AI/ML Overview

    The provided text describes the VenSure™ Balloon Dilation System, including the VenSure™ Light, VenSure™ Nav, and VenSure™ ET variants. These devices are used for balloon dilation of frontal ostial recesses, sphenoid sinus ostia, maxillary ostia/ethmoid infundibula, and the Eustachian tube in adults.

    Here's an analysis of the acceptance criteria and the study information as requested:

    1. A table of acceptance criteria and the reported device performance

    Performance TestingAcceptance Criteria (Criteria for SE justification)Reported Device Performance
    Balloon dimensional integrityThe dimensions of the balloon are as specified and in accordance with the properties of the predicates.All tests met the predefined acceptance criteria.
    Balloon pressure stability / Balloon fatigue conditioningThe burst pressure at end of lifetime is above the rated in use pressure and within the specifications of the predicates.All tests met the predefined acceptance criteria.
    Balloon Inflation/Deflation timesThe balloon inflates and deflates at end of lifetime of the device within the specifications of the predicates.All tests met the predefined acceptance criteria.
    Balloon burst pressureThe burst pressure is well above the rated in use pressure and within the specifications of the predicates.All tests met the predefined acceptance criteria.
    Mechanical integrityMechanical properties all met predefined acceptance criteria. It can be demonstrated that the minor differences in dimensions to the secondary predicate do not raise new concerns of safety and effectiveness and can be rated as substantial equivalent.All tests met the predefined acceptance criteria.
    Catheter and distal tip geometries characterizationThe specified catheter and distal tip geometries are within the range of the primary predicate devices.All tests met the predefined acceptance criteria.
    Simulated use testing in clinical modelSame design safety features and mechanically functionality as primary predicate device can be demonstrated.All tests met the predefined acceptance criteria.
    Navigation compatibility (VenSure™ Nav only)Navigation accuracy is within the rating of the secondary predicate.Test results demonstrate functionality and compatibility.
    LightGuide compatibility and performanceDimensions and light output performance are within the ratings of the secondary predicate.All tests met the predefined acceptance criteria.
    BiocompatibilityCompliance with FDA recognized consensus standard ISO 10993-1:2018 (cytotoxicity, irritation, sensitization, acute systemic toxicity testing).All tests successfully met the required acceptance criteria.
    Sterilization & StabilityCompliance with ISO 11135-1 for sterilization validation. Shelf life and packaging testing to support labeled shelf life.All tests were successfully completed.
    Electromagnetic compatibility and Electrical SafetyCompliance with IEC 60601-1 and IEC 60601-1-2.Testing was conducted per standards (implied successful).

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document mentions "Bench testing" and "Simulated use testing in clinical model." However, it does not specify sample sizes for the test sets used in performance data. It also does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. The reference to "clinical model" suggests a simulated environment rather than actual patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not provided in the document. The studies described are primarily bench testing and simulated use, not those requiring expert interpretation of medical images or patient outcomes for ground truth establishment.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not provided as the described studies do not involve expert adjudication of results in a clinical context.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study or any assessment of human reader improvement with or without AI assistance. The device in question is a physical medical device (balloon dilation system) and not an AI/software product requiring human reader performance studies.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This refers to an AI/software device. The VenSure™ Balloon Dilation System is a physical medical device. While the VenSure™ Nav includes "image-guided visualization" and "navigation software," the document does not describe "standalone" algorithm-only performance in the context of an FDA-cleared AI/software device. The navigation functionality is an aid for the physician using the physical device.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the bench testing, the "ground truth" implicitly refers to engineering specifications and established physical properties (e.g., "dimensions of the balloon are as specified," "burst pressure at end of lifetime is above the rated in use pressure"). For biocompatibility, the ground truth is established by biocompatibility testing standards (ISO 10993-1:2018). For sterilization, the ground truth is sterilization validation standards (ISO 11135-1). For navigation compatibility, it's about functionality and compatibility with the Fiagon Navigation system meeting accuracy ratings of the predicate. This is based on objective, measurable physical and engineering endpoints rather than clinical ground truth (like pathology or expert consensus on a diagnosis).

    8. The sample size for the training set

    This information is not applicable/not provided. The device is a physical medical device, not an AI/machine learning model that typically requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable/not provided as there is no mention of a training set for an AI/machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220027

    Validate with FDA (Live)

    Date Cleared
    2022-04-12

    (97 days)

    Product Code
    Regulation Number
    874.4180
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Minnesota 55447

    Re: K220027

    Trade/Device Name: Audion ET dilation system Regulation Number: 21 CFR 874.4180
    |
    | Regulation Number: | Class II 21 CFR 874.4180

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To dilate the cartilaginous portion of the Eustachian tube for treating tube dysfunction in patients 18 years and older using a transnasal approach.

    Device Description

    The Audion ET dilation system (ET is an abbreviation for Eustachian tube) is intended to dilate the Eustachian tube through use of balloon dilation and a transnasal approach. It contains a dilation device, inflation device and inflation lock. The dilation device has a 1.26 mm outer diameter (OD) curved non-malleable shaft with a balloon fixed at the distal end and a 1.91 mm atraumatic polymer ball tip. The dilation device is positioned under endoscopic visualization with the balloon fully retracted. Once the portion of the device from the ball tip to the base of the curve is seated in the Eustachian tube, the balloon is advanced and then inflated. The deployed balloon angle of the dilation device is 45° for optimal treatment of the Eustachian tubes. The distal leg length with balloon extended is 18.5 mm. When the balloon is inflated to 12 atm, the balloon diameter is 6mm and the body length is 20 mm.

    The inflation device is an accessory that consists of an inflation syringe and an extension line. The inflation device is designed to deliver a pressure of 12 atm and is used to inflate the balloon on the dilation device.

    The inflation lock is an optional accessory that is intended to interface with the Audion dilation device, inflation syringe and extension line connections to hold or release pressure during balloon dilation.

    The Audion ET dilation system is provided sterile and is for single patient use only. The dilation device, inflation device and inflation lock are sterilized using ethylene oxide.

    The Audion ET dilation system has been tested to withstand multiple inflations in a surgical case

    AI/ML Overview

    The provided text is a 510(k) summary for the Audion ET dilation system, a medical device for dilating the Eustachian tube. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving a device meets specific acceptance criteria based on diagnostic performance (like an AI/imaging device).

    Therefore, many of the requested categories (e.g., acceptance criteria table, sample size for test set, number of experts for ground truth, MRMC study, standalone performance, training set details) are not applicable or cannot be extracted from this document, as they pertain to studies typically performed for AI or diagnostic imaging devices to prove efficacy/accuracy.

    This document describes the device itself, its intended use, and its technological characteristics in comparison to a predicate device, along with bench testing to demonstrate performance as intended and meeting design specifications.

    Here's what can be extracted and addressed:

    1. A table of acceptance criteria and the reported device performance:

    The document does not present acceptance criteria in a table format with corresponding quantitative performance metrics related to diagnostic accuracy or clinical outcomes. Instead, it states that "Performance testing showed that the device meets design specifications and performs as intended." The "acceptance criteria" here are implied to be the successful completion of various bench tests demonstrating the device's functionality, safety, and compatibility with its intended use, and substantial equivalence to the predicate.

    Acceptance Criteria Category (Implied)Reported Device Performance (Summary)
    BiocompatibilityMeets (tested)
    Design Verification (Dimensional)Meets (tested)
    Design Verification (Functional)Meets (tested)
    Design Verification (Strength)Meets (tested)
    HFE/UE (Human Factors Engineering/Usability Engineering) VerificationMeets (tested with ENT physicians)
    Packaging IntegrityMeets (tested)
    Shelf LifeMeets (tested)
    Sterilization (Ethylene Oxide)Successfully sterilized
    Withstand Multiple InflationsTested to withstand multiple inflations in a surgical case
    Substantial EquivalenceDemonstrated substantial equivalence to predicate device K163509 (based on safety, efficacy, intended use, principles of operation, and technological characteristics)

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not applicable in the context of diagnostic data. The "test set" here refers to physical devices or components subjected to bench testing. The document does not specify the number of units tested for each specific bench test (e.g., how many devices were tested for strength, or how many packaging samples for integrity).
    • Data Provenance: The tests are described as "bench testing" and "performance testing," which typically occur in a lab setting. There's no mention of country of origin for clinical data (as there was no clinical study described for efficacy) or whether it was retrospective or prospective. HFE/UE verification involved "ENT physicians."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not specified. The document mentions "ENT physicians" were involved in "HFE/UE verification testing" and "design validation." Their role seems to be evaluating the usability and design effectiveness, rather than establishing a "ground truth" for a diagnostic task.
    • Qualifications of Experts: "ENT physicians." No further detail (e.g., years of experience, board certification) is provided.

    4. Adjudication method for the test set:

    Not applicable. There was no clinical study with a "ground truth" requiring adjudication. The document describes bench testing and usability verification, not multi-reader diagnostic interpretation.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is not an AI or diagnostic imaging device submission.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an AI or diagnostic imaging device submission.

    7. The type of ground truth used:

    Not applicable in the context of diagnostic ground truth. The "ground truth" for this device's performance is its ability to meet engineering specifications (e.g., balloon diameter at a certain pressure, shaft strength, sterilization efficacy), which are verified through various physical and chemical tests.

    8. The sample size for the training set:

    Not applicable. There is no AI component requiring a training set.

    9. How the ground truth for the training set was established:

    Not applicable. There is no AI component.

    Ask a Question

    Ask a specific question about this device

    K Number
    K210841

    Validate with FDA (Live)

    Date Cleared
    2021-08-16

    (147 days)

    Product Code
    Regulation Number
    874.4180
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K210841

    Trade/Device Name: NuVent Eustachian Tube Dilation Balloon Regulation Number: 21 CFR 874.4180
    Eustachian Tube Balloon Dilation System

    Classification Name: Eustachian Tube Balloon Device (21 CFR 874.4180
    )

    Classification: Class II (21 CFR 874.4180)

    Product Code: PNZ

    Primary Predicate Device: XprESS ENT
    | 21 CFR 874.4180
    to demonstrate compliance with the special controls applicable to this device, as set out in 21 CFR 874.4180

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nu Vent Eustachian Tube Dilation Balloon is indicated for use in patients 18 years and older who need treatment for persistent Eustachian tube dysfunction.

    Device Description

    The NuVent™ Eustachian Tube Dilation Balloon is composed of a sterile, single use balloon handpiece and stopcock valve. The balloon is used in conjunction with the NuVent™ inflator, which is packaged with extension tubing to connect the inflator to the stopcock valve. The NuVent™ Inflator was cleared as a part of the NuVent™ EM Sinus Balloon Dilation System, 510(k) cleared under K132297.

    A pre-angled flexible balloon section promotes easy access and insertion into the Eustachian tube from the nasal opening with the aid of endoscopic guidance. An atraumatic tip at the distal end of the flexible balloon section is meant to reduce the potential for risk of injury to the Eustachian tube and surrounding tissues during insertion. The rigid shaft and handle extending to the flexible balloon section provides stability and tactile feedback when moving the balloon through the nasal passages and into the Eustachian tube canal.

    Mounted near the distal end of the balloon handpiece is a flexible balloon section. Once the balloon section is advanced into the target Eustachian tube the 6mm diameter/16 mm long rigid balloon is inflated using the separate NuVent™ inflator, expanding the Eustachian tube canal tissue and cartilage.

    AI/ML Overview

    The provided text is a 510(k) Summary for the NuVent Eustachian Tube Dilation Balloon. This type of document is submitted to the FDA to demonstrate that a new medical device is substantially equivalent to a legally marketed predicate device. The information supplied in this document focuses on mechanical and functional performance testing to prove substantial equivalence, rather than a clinical study evaluating the device's impact on human readers or clinical outcomes in the way a diagnostic AI device might be.

    Therefore, many of the requested items (e.g., number of experts for ground truth, MRMC study, effect size of human improvement with AI, standalone performance, training set details) are not applicable to this type of device and the information provided in this 510(k) summary.

    However, I can extract the acceptance criteria and performance data for the mechanical and functional tests that were performed to support the device's substantial equivalence.

    Here's the information based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The document states that "All samples passed testing and met acceptance criteria." While the specific numerical acceptance criteria for each test are not explicitly detailed (e.g., "tensile strength must be > X N"), the general statement indicates that the device met pre-defined requirements for these performance characteristics. The table below lists the performance tests conducted, implying that the acceptance criterion for each was successful completion without failure, demonstrating performance comparable to or better than the predicate devices and ensuring safety and effectiveness.

    Performance Test CategoryReported Device Performance (Acceptance Criteria Met)
    Mechanical/Functional Testing
    Wire dislodgementPassed (met acceptance criteria)
    Balloon inflationPassed (met acceptance criteria)
    Balloon deflationPassed (met acceptance criteria)
    Stopcock holding pressure without leakingPassed (met acceptance criteria)
    Pressure indicator button functionalityPassed (met acceptance criteria)
    System does not leak or burst under use conditions in cyclic testingPassed (met acceptance criteria)
    Handpiece does not leak at the rated burst pressure in cyclic testingPassed (met acceptance criteria)
    Tensile strength of balloon, balloon bonds and tipPassed (met acceptance criteria)
    Torsional testing of balloon systemPassed (met acceptance criteria)
    Usability testing, including simulated usePassed (met acceptance criteria)
    Flexibility testing (comparing subject and predicate devices under axial, angular, and torsional loading of balloon sections)Passed (met acceptance criteria), demonstrating that the NuVent device is as safe and effective as the primary and secondary predicates.
    BiocompatibilityComplies with ISO 10993-1 and FDA guidance (met requirements for intended use)
    SterilizationAchieved minimum sterility assurance level (SAL) of 10^-6 (validated to ISO 11135, ISO 11138-2, ISO 11737-1, ISO 11737-2)
    Shelf Life TestingAll samples passed testing and met acceptance criteria of design inputs for device and packaging system following accelerated aging. Real-time aging is ongoing.

    Study Information Specifics:

    1. Sample Size Used for the Test Set and Data Provenance:

      • The document states "All samples passed testing," but does not specify the sample size (N) used for each of the performance tests.
      • The data provenance is from non-clinical tests conducted by Medtronic Xomed, Inc. This testing is typically done in a lab setting, not on human subjects.
      • The data is prospective in the sense that the tests were designed and executed to evaluate the new device.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • N/A. This device's testing involves mechanical and functional performance, not diagnostic interpretation requiring expert ground truth in the context of a medical imaging or AI study. The ground truth for these tests is based on engineering specifications and physical measurements/observations of device performance against predefined criteria.
    3. Adjudication Method for the Test Set:

      • N/A. Since this involves mechanical/functional tests against engineering specifications, there is no "adjudication" in the sense of reconciling human expert opinions. Test results are objectively measured or observed.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • N/A. This device (Eustachian Tube Dilation Balloon) is not an AI-powered diagnostic tool, nor does it assist human readers in interpreting medical images. It is a surgical device.
    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • N/A. This device does not have an "algorithm-only" performance component. It is a physical medical device.
    6. The Type of Ground Truth Used:

      • The ground truth used for this type of device relies on engineering specifications, physical measurements, and adherence to established performance standards (e.g., ISO, ASTM, internal Medtronic criteria). For example, a "passed" result for tensile strength means it met a minimum force before breaking, or "passed" for inflation means it inflated to the specified dimensions without bursting within a given pressure range.
    7. The Sample Size for the Training Set:

      • N/A. There is no "training set" in the context of this device's mechanical and functional testing, as it is not an AI/machine learning model.
    8. How the Ground Truth for the Training Set was Established:

      • N/A. This question pertains to AI/machine learning model development, which is not relevant to the type of device and testing described in this 510(k) summary.
    Ask a Question

    Ask a specific question about this device

    K Number
    K201398

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-10-15

    (140 days)

    Product Code
    Regulation Number
    874.4420
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Minnesota 55447

    Re: K201398

    Trade/Device Name: SINUSPRIME Dilation System Regulation Number: 21 CFR 874.4180
    | Class I, 21 CFR 874.4420Class II 21 CFR 874.4180

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Use a single dilation device by transnasal approach to access and treat an individual anatomical location as follows:

    · Remodel by balloon displacement the adjacent bone and paranasal sinus structures in the Maxillary ostialethmoid infundibula in patients 2 years and older;

    · Remodel by balloon displacement the adjacent bone and paranasal sinus structures in the Frontal ostialrecesses in patients 12 years and older: may be used with LED Light fiber to locate, illuminate across nasal and sinus structures;

    · Remodel by balloon displacement the adjacent bone and paranasal sinus structures in the Sphenoid sinus ostia in patients 12 years and older;

    • Dilation of the cartilaginous portion of the Eustachian tube for treating persistent Eustachian tube dysfunction in patients 18 years and older.

    Device Description

    The SINUSPRIME Dilation System is intended to remodel or recreate the sinus outflow tract and dilate the Eustachian tube by transnasal balloon dilation. Each packaged SINUSPRIME System includes a balloon device configured for a specific anatomy (frontal, sphenoid, maxillary, or Eustachian tube (ET) configuration) and an inflation device. The SINUSPRIME Dilation System combines features of an ostium seeker with the tissue expansion effect of balloon dilation. The SINUSPRIME Dilation System is provided sterile and for single use only. The familiar features of this device enable a physician to track the device into the sinuses and Eustachian tubes using endoscopic visualization. The SINUSPRIME LED Light Fiber consists of a flexible illumination fiber that allow physicians to locate, illuminate within, and trans-illuminate across nasal and sinus structures. The SINUSPRIME LED Light Fiber is optional and available packaged and pre-loaded only on the SINUSPRIME Frontal with LED model.

    AI/ML Overview

    The provided text is a 510(k) summary for the Stryker ENT SINUSPRIME Dilation System. It outlines the device, its indications for use, technological characteristics, and a comparison to predicate devices to establish substantial equivalence. However, it does not contain the specific information required to answer your questions regarding acceptance criteria, device performance, study design, or ground truth establishment for an AI-based device.

    The document discusses "Performance Data" but only states: "Performance testing involved biocompatibility, design verification (dimensional, strength, HFE/UE verification testing), packaging, shelf life and design validation with physicians (HFE/UE). Performance testing showed that the device meets design specifications and performs as intended." This refers to engineering and human factors testing for a physical medical device, not a study involving AI algorithm performance with a test set, expert readers, or ground truth in the context of diagnostic or prognostic capabilities.

    Therefore, I cannot provide the requested information from this document. The SINUSPRIME Dilation System is a physical medical device (balloon catheter for dilation), not an AI/ML-based diagnostic or imaging device. The questions you've asked are typically relevant for AI/ML medical devices where performance metrics like sensitivity, specificity, AUC, or reader agreement are crucial for regulatory submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201115

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-08-27

    (122 days)

    Product Code
    Regulation Number
    874.4420
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K201115

    Trade/Device Name: Next Generation Balloon Dilation System Regulation Number: 21 CFR 874.4180
    SystemNext Generation Balloon Dilation SystemEustachian tube balloon dilation system (21 CFR 874.4180
    |
    | ClassificationSection | 874.4180
    | 874.4180

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NGB Balloon Dilation System is intended to provide a means to access the sinus space within and across nasal and sinus structures; dilate the sinus ostia and spaces associated with the paranasal sinus cavities for diagnoutic procedures; irrigate from within a target sinus; suction throughout therapeutic procedures and to facilitate diagnostic procedures.

    For children aged 17 and under, the NGB Balloon System is intended to provide a means to access the sinus space within and across nasal and sinus structures; dilate sinus ostia and spaces associated with the maxillary sinus for diagnostic and therapeutic procedures; irrigate from within the maxillary sinus; suction throughout therapeutic procedures and to facilitate diagnostic procedures.

    The NGB Balloon Dilation System is intended to dilate the cartilaginous Eustachian tube for treatment of persistent Eustachian tube dysfunction in patients ages 18 and older.

    The NGB Balloon Dilation System with the navigation guidewire may be utilized in conjunction with the TruDi™ Navigation System, to help direct access to nasal and paranasal spaces, and to confirm placement in the targeted anatomy. NGB may be utilized in conjunction with the TruDi™ NAV Wire to confirm placement of the balloon in the Eustachian tube.

    The NGB Balloon Dilation System with the illumination guidewire may be utilized to illuminate within and provide transcutaneous illumination across nasal and sinus structures.

    Device Description

    The Next Generation Balloon Dilation System (NGB) is an integrated balloon sinuplasty (BSP) and Eustachian tube (ET) dilation device, that will be available in light fiber and navigation guidewire configurations for compatibility with illumination and navigation technology. The system includes a handle with several integrated features to allow for single-handed use. The adjustable gripping feature allows for ergonomic handling. The directable guide tip enables access (i.e. manipulating/separating/dividing tissue) and placement near the targeted anatomy. There are four preset approximate positions: Sphenoid (straight), Eustachian tube (55°), Frontal (70°), Maxillary (110°). The directable guide knob allows for positioning of the balloon tip towards the target anatomy. The wire slider and spinner are used to advance, retract, and rotate the guidewire. The balloon slider is used advance and retract the balloon. The Next Generation Balloon Dilation System also allows for suction and irrigation in and around the target anatomy.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Acclarent Next Generation Balloon Dilation System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA document primarily focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing. Therefore, rather than specific acceptance criteria in terms of performance metrics with pre-defined thresholds, the document outlines the types of non-clinical tests performed and states that the device "met all acceptance criteria" and "passed all tests in accordance with appropriate test criteria and standards."

    Acceptance Criteria Category (Derived from Testing)Reported Device Performance
    Functional PerformanceMet all acceptance criteria for attributes such as:
    - Simulated Use TestingReliability to remain mechanically functional throughout anticipated conditions of use. Design features limit access to only the cartilaginous portion of the Eustachian tube.
    - Dimensional AttributesMet all acceptance criteria.
    - Cycle FatigueMet all acceptance criteria.
    - Balloon BurstMet all acceptance criteria.
    - Bond SeparationMet all acceptance criteria.
    - Irrigation Flow RateMet all acceptance criteria.
    - Device AccuracyMet all acceptance criteria.
    - Inflation and Deflation CharacterizationPerformed, including time and pressure measurements, and leak testing of the balloon.
    - Durability (fatigue and burst pressure)Performed.
    - Verification of Safety FeaturesPerformed, including characterization of catheter geometries and distal tip insertion limitation mechanisms.
    BiocompatibilityBiocompatible.
    SterilizationValidated per AAMI/ANSI/ISO 11135-1: 2007 (sterility assurance level of 10-6 via overkill (half-cycle approach) in a fixed chamber). Ethylene oxide residuals meet ISO 10993-7:2008. Not tested/labeled as “non-pyrogenic.”
    Packaging Shelf LifeEstablished to be 3 months per ASTM F1980-07.
    Mechanical TestingPerformed, including tensile and flexural testing of catheter joints and materials.

    2. Sample Size for Test Set and Data Provenance

    The document explicitly states: "Clinical data was not necessary to determine that the subject Next Generation Balloon Dilation System performs as intended." This indicates that no clinical "test set" was used as part of this submission for performance evaluation against specific clinical acceptance criteria. The performance data is derived solely from non-clinical bench testing. The provenance of this data is therefore internal laboratory testing rather than patient data from a specific country or retrospective/prospective study.

    3. Number of Experts and Qualifications to Establish Ground Truth for Test Set

    Not applicable, as no clinical test set was used to establish ground truth with expert review.

    4. Adjudication Method for the Test Set

    Not applicable, as no clinical test set was used that would require adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was done. The submission is based on demonstrating substantial equivalence through non-clinical testing.

    6. Standalone (Algorithm Only) Performance

    Not applicable. This device is a physical medical device, not an algorithm, and its performance is evaluated in relation to its physical and functional characteristics.

    7. Type of Ground Truth Used

    For the non-clinical testing, the "ground truth" would be the engineering specifications, industry standards (e.g., AAMI/ANSI/ISO for sterilization, ASTM for shelf life), and validated internal test methods that define acceptable performance.

    8. Sample Size for the Training Set

    Not applicable, as this is a physical medical device and not an AI/ML algorithm that requires a "training set" in the conventional sense. The "training" for such a device effectively comes from its design and manufacturing processes, guided by engineering principles and regulatory standards.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171761

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-01-16

    (216 days)

    Product Code
    Regulation Number
    874.4180
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Trade/Device Name: ACCLARENT AERA® Eustachian Tube Balloon Dilation System Regulation Number: 21 CFR 874.4180
    | 21 CFR 874.4180

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACCLARENT AERA® Eustachian Tube Balloon Dilation System is intended to dilate the Eustachian tube for treatment of persistent Eustachian tube dysfunction in patients ages 18 and older.

    Device Description

    The ACCLARENT AERA® Eustachian Tube Balloon Dilation System includes a Eustachian Tube Balloon Catheter and Guide Catheter designed specifically for use in accessing and dilating the Eustachian Tube. The system is used with the following additional devices: the ACCLARENT® SE Inflation Device (or ACCLARENT® Balloon Inflation Device). The ACCLARENT® SE Inflation Device or ACCLARENT® Balloon Inflation Device is used to inflate the balloon. All devices are provided sterile for single-patient use.

    The ACCLARENT AERA® Eustachian Tube Balloon Dilation System includes is a 6x16mm (inflated diameter x length) flexible Balloon Catheter with an integrated shaft and a nylon balloon at the distal tip. The non-compliant balloon is designed to dilate the cartilaginous portion of the Eustachian tube. The shaft consists of dual lumen tubing with an actuator component that is designed to assist in rotation and advancement of the balloon catheter. The balled catheter tip (aka, “blueberry tip") on the balloon catheter is designed to restrict advancement of the device into the bony portion of the Eustachian tube, known as the isthmus. There is an endoscopic marker placed at the proximal taper of the balloon to aid in positioning under direct endoscopic visualization.

    The AERA Guide Catheter is anatomically designed to facilitate AERA Balloon Catheter access to the Eustachian tube. The AERA Guide Catheter incorporates an atraumatic distal tip and distal angled tip profile that facilitates access to the Eustachian tube. The Guide Catheter supplied with the ACCLARENT AERA® Eustachian Tube Balloon Dilation System contains a lubricious inner liner to allow smooth passage for the balloon catheter and includes a hypotube for rigidity.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the ACCLARENT AERA® Eustachian Tube Balloon Dilation System, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document focuses on the substantial equivalence of the device and does not explicitly list specific numerical acceptance criteria for clinical performance (e.g., target success rates, symptom improvement percentages). Instead, the performance data section describes meeting acceptance criteria for various bench tests related to device integrity and safety.

    Acceptance Criteria CategoryReported Device Performance
    Bench Testing
    Dimensional AttributesMet all acceptance criteria
    Cycle FatigueMet all acceptance criteria
    Balloon BurstMet all acceptance criteria
    Bond SeparationMet all acceptance criteria
    BiocompatibilityTesting has shown the device is biocompatible
    SterilizationValidated per AAMI/ANSI/ISO 11135:2014, achieving a sterility assurance level of 10^-6
    Ethylene Oxide ResidualsMet ISO 10993-7:2008 requirements
    Packaging Shelf LifeEstablished at one year via accelerated aging per ASTM F1980-07

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Clinical Data):
      • The document references "real-world clinical data collected from 25 patients from two sites."
      • Data Provenance: The document cites a publication by Luukkainen et al. (2017) regarding the feasibility of balloon Eustachian tuboplasty under local anesthesia. One of the authors, Rautiainen M, is from Finland (based on a quick search for the publication reference, not explicitly stated in the provided text). It's likely this smaller clinical data set is retrospective or a case series/pilot observational study, given the description as "real-world clinical data" supporting use under local/topical anesthetic. The document does not describe a formal, large-scale clinical trial used as a "test set" for the purpose of establishing primary effectiveness or safety endpoints in a randomized controlled manner.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • The document does not specify the number of experts used or their qualifications for establishing ground truth for the 25-patient clinical data set. The clinical data appears to be observational and intended to support the feasibility of the device under certain anesthetic conditions rather than being a formal "test set" with pre-defined ground truth established by experts for performance evaluation.

    4. Adjudication Method for the Test Set

    • No adjudication method is described for the 25-patient clinical data.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not explicitly mentioned or described in the provided document. The device is a medical device for dilation, not an imaging or diagnostic AI tool where MRMC studies are typically performed.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This question is not applicable. The ACCLARENT AERA® Eustachian Tube Balloon Dilation System is a physical medical device, not an algorithm or AI system. Its performance is evaluated through bench testing, biocompatibility, sterilization, and then clinical use with human operators.

    7. The Type of Ground Truth Used

    • For Bench Testing: The "ground truth" for bench testing would be predefined engineering specifications, material properties, and regulatory standards (e.g., dimensional tolerances, burst pressure limits, sterilization efficacy, biocompatibility assay results).
    • For Clinical Data (25 patients): The clinical data from the 25 patients likely relied on clinical endpoints, such as resolution of Eustachian tube dysfunction symptoms, patency of the Eustachian tube, or patient satisfaction, as assessed by the treating physicians. The document does not detail how "ground truth" was established for these clinical outcomes in this specific dataset.

    8. The Sample Size for the Training Set

    • This question is not applicable as the ACCLARENT AERA® Eustachian Tube Balloon Dilation System is a physical medical device, not an AI model requiring a "training set" in the computational sense.
    • The document mentions that the device has "expanded indications for use" and that "additional clinical testing was not required to demonstrate safety and efficacy of the device" because it is substantially equivalent to predicate devices. This indicates that prior clinical data from the predicate devices and existing literature inform its use, rather than a new "training set" for an algorithm.

    9. How the Ground Truth for the Training Set was Established

    • This question is not applicable for the reasons stated in point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K163509

    Validate with FDA (Live)

    Date Cleared
    2017-04-05

    (111 days)

    Product Code
    Regulation Number
    874.4180
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    510k Summary Text (Full-text Search) :

    XprESS ENT Dilation System

    Balloon Sinus Dilation System

    Class I, 21 CFR 874.4420

    Class II, 21 CFR 874.4180

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    To access and treat the maxillary ostia/ethmoid infundibula in patients 2 years and older, and frontal ostia/recesses and sphenoid sinus ostia in patients 12 years and older using a transnasal approach. The bony sinus outflow tracts are remodeled by balloon displacement of adjacent bone and paranasal sinus structures.

    To dilate the cartilaginous portion of the Eustachian tube for treating persistent Eustachian tube dysfunction in patients 18 years and older using a transnasal approach.

    Device Description

    The XprESS ENT Dilation System is intended to recreate the sinus outflow tract and dilate the Eustachian tube by transnasal balloon dilation. The XprESS device combines features of a curved suction tip and an ostium seeker with the tissue expansion effect of balloon dilation. The familiar features of this device enable a physician to track the device into the sinuses and Eustachian tubes using endoscopic visualization. Since the distal end of the device is re-shapeable, one balloon can be modified to work on multiple sinuses and Eustachian tubes within the same patient.

    The XprESS device curved suction tip has an atraumatic ball tip. A suction tube may be connected to the proximal barbed fitting to provide active suction by covering the suction vent. An Extension Line connected to a syringe may be connected to the proximal barbed fitting to provide irrigation. The device was designed to prevent fluid from exiting the suction vent during irrigation. The XprESS ENT Dilation System is provided sterile and for single use only.

    The Xpress ENT Dilation System includes the XprESS device, Inflation Syringe, Bending Tool, and two Extension Lines. The XprESS LoProfile and Ultra ENT Dilation Systems also include the PathAssist LED Light Fiber. The XprESS Pro ENT Dilation System also includes a Tuohy Adapter.

    AI/ML Overview

    I am sorry, but I cannot fulfill your request to describe the acceptance criteria and study proving device performance based on the provided text. The document is a 510(k) summary for a medical device (XprESS ENT Dilation System) and does not contain acceptance criteria or a detailed study description in the specified format.

    The document states that "The primary efficacy endpoint was met with Eustachian tube balloon dilation demonstrating superiority over medical management for improvement in ETDQ-7 symptom scores (p<0.0001). The primary safety endpoint was met with a 0% complication rate." However, it does not explicitly define acceptance criteria as a numerical threshold or target that the device needed to meet for these endpoints.

    It only summarizes the results of a clinical study, but does not provide the detailed information requested in points 1-9 of your prompt, such as:

    • A table of acceptance criteria with reported performance.
    • Specific sample sizes for test sets, data provenance, expert numbers/qualifications, or adjudication methods for establishing ground truth for a test set.
    • Information on MRMC comparative effectiveness or standalone algorithm performance.
    • Details on the training set or how its ground truth was established, as this is a device and not an AI algorithm.

    Therefore, I cannot extract the requested information from the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2