    510k Summary Text (Full-text Search) :

    California 94080

    Re: K251790
    Trade/Device Name: SeptAlign
    Regulation Number: 21 CFR 874.3620
    California 94080

    Re: K251790
    Trade/Device Name: SeptAlign
    Regulation Number: 21 CFR 874.3620
    Absorbable
    Regulatory Class: Class II
    Product Code: NHB
    Regulation Number: 21 CFR 874.3620

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SeptAlign is used to support and straighten deviations in septal cartilage when sufficient healthy cartilage exists, and the cartilage is appropriately mobilized utilizing standard septoplasty techniques.

    Device Description

    The SeptAlign System consists of a bioabsorbable implant and single use delivery device. The polydioxanone implant is 190 mm long and 0.65 mm thick with bi-directional anchors which enable mechanical correction of cartilaginous nasal septal deviations without cartilage resection. The implant also includes a surgical needle to enable placement which is trimmed off after use. The implant supports the cartilage in the straightened positioned as the cartilage remodels and is fully resorbed within a 6-month period.

    The implant is provided preloaded into a disposable delivery tool comprised of a non-patient contacting handle assembly and a medical grade stainless steel delivery cannula and trocar. The delivery tool enables placement of the distal portion of the implant in a minimally invasive manner. SeptAlign is provided sterile and is intended for single-use only.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K241150

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-07-24

    (90 days)

    Product Code
    Regulation Number
    874.3620
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    California 92612

    Re: K241150

    Trade/Device Name: MISTI Silicone Implant Regulation Number: 21 CFR 874.3620
    |
    | Classification: | 21CFR§874.3620
    br>21CFR§878.3550 LZK (Implant, Malar);21CFR§878.3550 FWP (Prosthesis, Chin, Internal)21CFR§874.3620

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MISTI Silicone Implant is intended for the augmentation of the nasal, malar, chin, gluteal, calf and/or pectoral contour.

    Device Description

    The MISTI Silicone Implant offers various types including Nasal, Chin, Mid-face, Forehead, Block, Temple, Gluteal, Calf and Pectoral. The Nasal, Chin, Mid-face, Forehead, Gluteal, Calf and Pectoral are used for the areas indicated in the type names. The Block type is used for various areas during plastic and reconstruction surgeries. The Temple type can be used on a juncture where four skull bones fuse together so called the temple region. It is made of implant grade silicone elastomer, and the design characteristic minimizes carving and promotes natural line.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA regarding the MISTI Silicone Implant. This document outlines the manufacturer's claim of substantial equivalence to existing predicate devices, primarily based on material, design, function, and performance data from non-clinical tests.

    However, the document does not contain information about acceptance criteria or a study proving the device meets specific acceptance criteria in the context of diagnostic accuracy, which is typically what these questions relate to for AI/Software as a Medical Device (SaMD).

    Therefore, I cannot provide the requested table and detailed information regarding acceptance criteria, device performance, sample sizes, ground truth establishment, expert qualifications, adjudication methods, or MRMC studies, as these types of studies were not conducted or reported in this 510(k) submission.

    This 510(k) focuses on demonstrating the physical and biological safety of the MISTI Silicone Implant by comparing its characteristics and performance in non-clinical tests (material properties, biocompatibility, sterilization) to legally marketed predicate devices, rather than on diagnostic accuracy or clinical effectiveness studies in humans with defined acceptance criteria for AI performance.

    The "Performance Data (Non-Clinical)" section lists various tests performed, such as:

    • Sterilization Validation
    • Appearance
    • Dimensional inspection
    • Extraction
    • Hardness (ASTM D2240 (Type A))
    • Tensile strength (ASTMD412)
    • Elongation (ASTM D412)
    • Tear strength (ASTM D624)
    • Cytotoxicity (ISO 10993-5)
    • Skin Sensitization (ISO 10993-10)
    • Intracutaneous Reactivity (ISO 10993-23)
    • Pyrogen (ISO 10993-11)
    • Acute Systemic Toxicity (ISO 10993-11)
    • Subchronic Toxicity (ISO 10993-11)
    • Implantation (ISO 10993-6)
    • Genotoxicity (Ames) (ISO 10993-3)
    • Genotoxicity (In vitro) (ISO 10993-3)

    These tests are primarily focused on the safety and material properties of the implant, not on evaluating an AI algorithm's diagnostic performance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232646

    Validate with FDA (Live)

    Date Cleared
    2024-05-24

    (268 days)

    Product Code
    Regulation Number
    874.3620
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    47906

    Re: K232646

    Trade/Device Name: Biodesign Otologic Butterfly Graft Regulation Number: 21 CFR 874.3620
    | Polymer, Ent Synthetic-Polyamide (mesh or foil material) |
    | Product Code: | KHJ (21 CFR §874.3620

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Biodesign Otologic Butterfly Graft is intended for use as an implant material to aid in the natural healing process in myringoplasty and tympanoplasty procedures.

    Device Description

    The Biodesign Otologic Butterfly Graft is a self-securing butterfly-style graft structure with the same underlay component as the predicate device, attached to an external stabilizing component with an absorbable knotted thread, all made from the same SIS (small intestinal submucosa) ECM material as that of the predicate device. This self-securing structure maintains the location and close tissue approximation of the underlay component across the tympanic membrane (TM) defect as an implant material to aid in the natural healing process in myringoplasty and tympanoplasty procedures.

    AI/ML Overview

    The provided text does not describe an acceptance criteria table or a study that specifically "proves" the device meets acceptance criteria in the typical sense of a single, definitive study with a direct comparison against a set of performance metrics. Instead, it describes a Substantial Equivalence (SE) determination for a new device, the "Biodesign Otologic Butterfly Graft," by comparing it to a legally marketed predicate device (K161000).

    The concept of "acceptance criteria" here is implicitly tied to demonstrating that the new device is as safe and effective as the predicate, despite some design changes. The studies conducted are largely comparative, aiming to show that the changes do not introduce new risks or diminish performance relative to the predicate.

    Here's an attempt to extract the requested information based on the provided text, acknowledging that some details might be inferred or not explicitly stated as they would be in a direct performance study against a predefined acceptance criterion.


    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document focuses on demonstrating substantial equivalence to a predicate device rather than meeting pre-defined numerical acceptance criteria for a novel device. The "acceptance criteria" for the animal study are explicitly stated and are the closest to what was requested.

    Acceptance Criterion (from Animal Study)Reported Device Performance (from Animal Study)
    Treated tympanic membranes must (1) appear visually healed, with no perforation observed via endoscopy at follow-up.The animal study showed a healing/closure rate of 79% among the 14 tympanic membrane perforations implanted with the subject device (11/14). This rate is within the range of the clinical and animal data submitted in support of the predicate device (63-100%). Visual healing via endoscopy is implied in the "healing/closure rate" and "no perforation observed."
    Treated tympanic membranes must (2) have a waveform on the tympanogram above the baseline perforated TM (i.e. Type A graph shape).The animal study for the subject device "evaluated the status of TM closure through a functional tympanogram technique." While a direct percentage of "Type A graph shape" isn't given, the statement "This healing/closure rate is within the range of the clinical and animal data submitted in support of the predicate device (63-100%)" combined with the method of evaluation suggests this criterion was met for the healed cases.
    Treated tympanic membranes must (3) histologically show healing in progress at follow-up, as determined by the Veterinary Pathologist's review of the histologic sections of the grafted/implanted tympanic membranes."Histological evaluation of the healed TM tissue showed a tri-laminar healed TM structure in the animal studies for both the predicate and subject device." This indicates successful histological healing equivalent to the predicate.
    Implicit Acceptance Criteria for Substantial Equivalence (derived from the document's structure):
    Biocompatibility in accordance with ISO 10993 series.Biocompatibility testing for the subject device was submitted under ISO 10993, addressing various parts including cytotoxicity, irritation, sensitization, systemic toxicity, genotoxicity, implantation effects, chemical characterization, and EO residuals. It is implied these tests met the standard requirements to demonstrate safety equivalent to the predicate.
    Mechanical performance for intended use (burst strength, joint strength).Product verification testing for "mechanical performance of the subject device for its intended use" included burst strength and joint strength (specific to the new design elements). Visual inspection was also performed. The conclusion states the device "functions as intended during implantation and throughout the patient's natural healing process to locate and secure the underlay component." No specific numerical results are provided, but the statement implies successful performance meeting internal specifications.
    Usability of device handling and implantation.A Summative Usability Report utilizing "simulated-use testing... in a worst-case benchtop model" and leveraging GLP animal study data related to usability was conducted. The report verified "device handling and usability characteristics in a hydrated state." This implicitly met the usability requirements.
    Packaging system adequacy for a 6-month shelf-life.Non-clinical bench testing on accelerated-aged samples confirmed the packaging system is "adequate to support a 6-month shelf-life claim."
    Sterilization effectiveness (SAL: 10-6).The device undergoes Ethylene Oxide (EO) sterilization with a "Sterility Assurance Level (SAL): 10-6," established via an adoption analysis per AAMI TIR28:2016 for the predicate device. This implies the sterilization method and SAL are equivalent and effective.

    Summary of Comparative Equivalence: The overall "acceptance criterion" for the device, in the context of this 510(k) submission, is to demonstrate that its design changes do not introduce new questions of safety or effectiveness and that it performs as well as, or better than, the predicate device. The various studies and tests listed are all geared towards supporting this claim of substantial equivalence.

    2. Sample size used for the test set and the data provenance

    For the animal study:

    • Sample Size: 14 tympanic membrane perforations implanted with the subject device.
    • Data Provenance: Prospective GLP animal study in a chinchilla model. The location/country is not specified.

    For bench testing (mechanical performance, usability, packaging/shelf-life): Specific sample sizes are not provided, only that "Product verification testing was performed," "Summative Usability Report which utilized simulated-use testing," and "Non-clinical bench testing was performed on accelerated-aged representative SHS conditioned packaging system samples."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    For the animal study:

    • Number of Experts: At least one "Veterinary Pathologist."
    • Qualifications: "Veterinary Pathologist's review of the histologic sections of the grafted/implanted tympanic membranes." No specific years of experience are mentioned.

    For other testing, specific expert numbers or qualifications for ground truth establishment are not provided.

    4. Adjudication method for the test set

    For the animal study:

    • The text mentions a "Veterinary Pathologist's review of the histologic sections." This implies a single expert's determination for the histological ground truth. It does not describe a multi-reader adjudication method (e.g., 2+1, 3+1).
    • Visual healing was determined "via endoscopy," and functional healing by "typanogram technique." It's unclear if these assessments involved multiple readers or an adjudication process.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This document pertains to a medical device (graft) and its biological/mechanical performance, not an AI or imaging diagnostic device where MRMC studies involving human readers and AI assistance would be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical medical device (graft), not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the animal study:

    • Pathology: Histological assessment by a Veterinary Pathologist.
    • Clinical Observation: Visual assessment (no perforation observed via endoscopy).
    • Physiological/Functional Data: Tympanogram waveform analysis.

    8. The sample size for the training set

    • Not applicable. This document describes the evaluation of a physical medical device. There is no concept of a "training set" in the context of an animal study for a medical implant like there would be for an AI algorithm.

    9. How the ground truth for the training set was established

    • Not applicable. As there is no training set for this type of device evaluation, no ground truth was established for one.
    Ask a Question

    Ask a specific question about this device

    K Number
    K240919

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-05-03

    (30 days)

    Product Code
    Regulation Number
    874.3620
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Massachusetts 01702

    Re: K240919

    Trade/Device Name: Silk Voice (SMI-04) Regulation Number: 21 CFR 874.3620
    Device: Silk Voice Common Name: System, Vocal Cord Medialization Classification Regulation: 21 CFR 874.3620
    ------|----------------------------------------------------------|
    | ClassificationRegulation | 874.3620

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Silk Voice® is indicated for vocal fold medialization and vocal fold insufficiency that may be improved by injection of a soft tissue bulking agent. Silk Voice® injection augments the size of the displaced or deformed vocal fold so that it may meet the opposing fold at the midline for improved phonation. Vocal fold insufficiency associated with serious aspiration difficulties may be an urgent indication.

    Device Description

    Silk Voice® is a sterile, non-pyrogenic, cohesive implant provided in a prefilled syringe and is a ready to use product. Silk Voice is comprised of porous bioabsorbable silk particles suspended in an isotonic, aqueous formulation of cross-linked, high molecular weight hyaluronic acid (HA). The crosslinked HA gel acts as a carrier for the silk particles to facilitate delivery. The main component of Silk Voice is silk particles, manufactured exclusively from regenerated silk fibroin protein, isolated from purified silk fibers. When injected, Silk Voice provides immediate volume augmentation to the vocal fold tissue. The porous particles remain at the site of implantation, providing a scaffold for local tissue infiltration. This cellular infiltrated silk scaffold provides longterm restoration and augmentation.

    Silk Voice prefilled syringes are provided in a kit with a catheter, that is designed for endoscopic delivery to the vocal fold. The catheter accessory provided in the kit is specifically designed for delivery of injectable materials into tissue during endoscopic procedures.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria or specific studies that prove the device meets acceptance criteria in the context of a clinical performance study with human readers or an AI algorithm. The document is an FDA 510(k) summary for a medical device (Silk Voice, SMI-04) and primarily focuses on demonstrating substantial equivalence to a predicate device (Silk Voice, K180631) through bench testing and shelf-life data.

    Therefore, most of the requested information cannot be extracted from this document, specifically points 1, 2, 3, 4, 5, 6, 7, 8, and 9 relate to clinical or AI performance studies which are not described here.

    Here's what can be extracted:

    1. A table of acceptance criteria and the reported device performance

    The document states: "All bench testing passed the acceptance criteria." However, the specific quantitative acceptance criteria for each test and the corresponding reported performance values are not detailed in the summary. For example, for "Particle size and circularity analysis", it's mentioned that it passed, but the acceptance range and the measured value are not provided.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. The performance testing described is bench testing, not a clinical study on a test set of human data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as there is no test set involving human data or expert review described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as there is no test set involving human data or ground truth adjudication described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. No MRMC study or AI assistance is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. No AI algorithm performance is discussed.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable. The "ground truth" in this context would be the design specifications and regulatory requirements for the bench tests.

    8. The sample size for the training set

    Not applicable. No training set for an AI algorithm is mentioned.

    9. How the ground truth for the training set was established

    Not applicable. No training set for an AI algorithm is mentioned.


    Summary of available information related to performance testing:

    The device's performance was evaluated through bench testing and shelf-life testing.

    • Bench Testing:
      • Tests Performed: Particle size and circularity analysis, Rheometry, Extrusion force, HA fragment test, Particle concentration, Residual Crosslinker Content, pH, Endotoxin, Catheter leak test, Catheter tensile strength.
      • Outcome: "All bench testing passed the acceptance criteria," demonstrating that the delivery system meets pre-established design input requirements.
      • Biocompatibility: Relied on "Biocompatibility test results of the predicate device submitted as part of the original submission (K180631)," which are stated to "continue to support that Silk Voice® meets the requirements of ISO 10993 for its intended use."
    • Shelf-Life Testing: Performed to support labeled expiration dating.
    • Sterilization Validation: Conducted for steam and EtO sterilization cycles, demonstrating a Sterility Assurance Level (SAL) of 10-6.

    The primary purpose of this submission is to demonstrate substantial equivalence to an existing predicate device (Sofregen's Silk Voice K180631), highlighting that a change in silk particle size ($250 \pm 50 \mu m$ in the subject device vs. $380 \pm 46 \mu m$ in the predicate) "does not affect the safety or effectiveness assessment."

    Ask a Question

    Ask a specific question about this device

    K Number
    K233569

    Validate with FDA (Live)

    Device Name
    SeptAlign
    Manufacturer
    Date Cleared
    2024-03-22

    (137 days)

    Product Code
    Regulation Number
    874.3620
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    San Francisco, California 94080

    Re: K233569

    Trade/Device Name: SeptAlign Regulation Number: 21 CFR 874.3620
    : | NHB |
    | Regulation Number: | 21 CFR 874.3620

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SeptAlign is used to support and straighten minor deviations in septal cartilage when sufficient healthy cartilage exists, and the cartilage is appropriately mobilized utilizing standard septoplasty techniques.

    Device Description

    The Spirair SeptAlign implant is a bioabsorbable, polydioxanone ribbon that is intended to be used in nasal surgery. SeptAlign is 190 mm long and 0.65 mm thick with barbed features ("anchors") which enable attachment to and support of nasal septal cartilage. SeptAlign is trimmed to size by the physician to suit the anatomical conditions and clinical use case. The device includes a surgical needle to enable attachment to the tissue which is trimmed off affer use. SeptAlignis provided sterile as a single use device and when permanently implanted, is resorbed within a 6-month period.

    The SeptAlign implantis provided preloaded into a disposable delivery tool comprised of a non-patient contacting handle assembly and a medical grade stainless steel delivery cannula and trocar. The delivery tool enables placement of the distal portion of the implant in a minimally invasive manner. The SeptAlign implant and accessory delivery tool are provided sterile and are intended for single-use only.

    AI/ML Overview

    I am sorry, but the provided text from the FDA 510(k) notification for the SeptAlign device does not specify acceptance criteria for device performance or describe a study that proves the device meets such criteria in terms of accuracy, sensitivity, or specificity for any diagnostic or prognostic task.

    The document primarily focuses on the device's mechanical and material performance, biocompatibility, packaging, and shelf-life. It states that "No animal or clinical performance testing was required to support substantial equivalence." This indicates that the substantial equivalence determination for this device (SeptAlign) was not based on clinical performance studies measuring diagnostic accuracy or reader improvement.

    Therefore, I cannot fulfill your request for:

    1. A table of acceptance criteria and reported device performance related to diagnostic/prognostic metrics.
    2. Sample size, data provenance, expert ground truth establishment, adjudication, MRMC studies, standalone performance, or ground truth types. These are typically associated with studies evaluating AI/diagnostic device performance, which were not required or reported for this submission.

    The "Performance Data" section (Pages 4-5) lists the following performance elements:

    • Biocompatibility Testing: Confirms materials do not pose a risk of negative interaction. This is about safety, not diagnostic performance.
    • Distribution, Packaging and Shelf-Life Testing: Demonstrates integrity of sterile barrier and preservation of performance, related to product stability and safety.
    • Performance Testing – Bench: Focuses on the accessory delivery tool's functionality, mechanical integrity, and cannula joint strength. This is about the physical device and its delivery mechanism working as intended, not its ability to detect or diagnose anything.

    The acceptance criteria implicitly met are related to these physical and material properties (e.g., passing biocompatibility tests, maintaining sterility, successful mechanical deployment).

    In summary, the provided document does not contain the type of information needed to answer your questions regarding acceptance criteria for diagnostic performance or studies proving such performance, as the device is a physical implant (a bioabsorbable ribbon) rather than a diagnostic or AI-powered device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223167

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-08-17

    (314 days)

    Product Code
    Regulation Number
    874.3620
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    California 95138

    Re: K223167

    Trade/Device Name: Spirair Nasal Septal Strap Regulation Number: 21 CFR 874.3620
    : | NHB |
    | Regulation Number: | 21 CFR 874.3620

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Spirair Nasal Septal Strap is used to support and straighten in septal cartilage when sufficient healthy cartilage exists, and the cartilage is appropriately mobilized utilizing standard septoplasty techniques.

    Device Description

    The Spirair Nasal Septal Strap is a bioabsorbable, polydioxanone ribbon that is intended to be used in nasal surgery. The Nasal Septal Strap is 190 mm long and 0.65 mm thick with barbed features ("anchors") which enable attachment to and support of nasal septal cartilage. The Nasal Septal Strap is trimmed to size by the physician to suit the anatomical conditions and clinical use case. The device includes a surgical needle attachment to the tissue which is trimmed off after use. The Nasal Septal Strap is provided sterile as a single use device and when permanently implanted, is resorbed within a 6-month period.

    AI/ML Overview

    This document is a 510(k) summary for the Spirair Nasal Septal Strap, detailing its regulatory review and substantial equivalence to a predicate device. It primarily focuses on the device's technological characteristics, indications for use, and performance data required for regulatory clearance. It does not describe a study that proves a device meets specific acceptance criteria in the context of an AI/ML medical device, as the Spirair Nasal Septal Strap is a physical medical device.

    Therefore, I cannot provide the requested information about acceptance criteria and a study proving device performance as it pertains to an AI/ML medical device, because the provided text is for a physical medical device and lacks any mention of AI, algorithms, or complex performance metrics typical of AI/ML systems.

    The document discusses performance data related to:

    • Biocompatibility Testing: According to ISO 10993-1 and FDA Guidance, assessing cytotoxicity, sensitization, irritation, implantation, and chemical characterization.
    • Distribution, Packaging, and Shelf-Life Testing: Successful completion of accelerated aging and integrity of sterile barrier.
    • Performance Testing - Bench: Demonstrating physical and functional requirements, including mechanical properties, material properties, migration of implant, and in vitro degradation.
    • Performance Testing - Animal: A GLP animal study for safety, degradation, and biocompatibility.
    • Performance Testing - Clinical: Interim safety data from a prospective, multi-center, multi-cohort, early feasibility clinical study in the US.

    These tests are designed to demonstrate the safety and effectiveness of a physical implantable device, not an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223051

    Validate with FDA (Live)

    Date Cleared
    2023-02-03

    (127 days)

    Product Code
    Regulation Number
    874.3620
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    California 92110

    Re: K223051

    Trade/Device Name: Pre-Formed Silicone Block Regulation Number: 21 CFR 874.3620
    | Elastomer, Silicone Block |
    | Regulation Number: | 21 CFR 874.3620
    |
    | RegulationNumber | 21 CFR 874.3620
    | 21 CFR 874.3620
    | 21 CFR 874.3620

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pre-Formed Silicone Block is intended for use in the cosmetic correction of soft tissue deformities, and is contoured at the surgeon's discretion to create a custom implant to aid in the reconstruction process.

    Device Description

    The Pre-Formed Silicone Block is an implant intended to be used in the aesthetic (cosmetic) correction of soft tissue deformities. The Pre-Formed Silicone Block comes in multiple shapes and sizes (cup-shaped in small, medium and large; crescent-shaped, and rectangular block) and one durometer to accommodate a variety of surgical techniques and implantation sites. All implants are made from medical grade silicone and can be trimmed with a knife or scissors. The trimmable feature allows the surgeon to custom fabricate, at surgery, an implantable implant for a specific surgical indication. The implants are provided either sterile or non-sterile. Devices that are provided non-sterile must be sterilized prior to use.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device (Pre-Formed Silicone Block). This document describes the device, its indications for use, and a comparison to predicate devices to demonstrate substantial equivalence, rather than detailing a study that proves the device meets specific acceptance criteria through performance evaluation.

    Therefore, the information required to answer the prompt (acceptance criteria, device performance results, sample sizes, expert ground truth establishment, MRMC studies, standalone performance, etc.) is not present in the provided text. The document states:

    • "No additional testing was provided in this submission in order to demonstrate substantial equivalence."
    • "Previous testing (i.e., biocompatibility, sterilization, and non-clinical performance testing) was leveraged to support a demonstration of substantial equivalence." (This implies that this specific submission did not include new performance studies comparing the device to acceptance criteria, but rather relied on comparisons to previously cleared devices and existing safety/biocompatibility data.)

    In summary, the provided FDA 510(k) summary does not contain the information requested about performance studies, acceptance criteria, or their results. It primarily focuses on demonstrating substantial equivalence to predicate devices based on shared characteristics.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222748

    Validate with FDA (Live)

    Date Cleared
    2022-12-23

    (102 days)

    Product Code
    Regulation Number
    874.3620
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    24262 Taiwan

    Re: K222748

    Trade/Device Name: ShiNeo Silicone Implant Regulation Number: 21 CFR 874.3620
    |
    | Additional Product code | LZK, FWP |
    | Regulation Number: | 874.3620
    |
    | Regulation Number: | 874.3620

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ShiNeo Silicone Implant is intended for the augmentation of the nasal and/or chin contour

    ShiNeo Silicone Implant is intended for the augmentation or reconstruction of the nasal and/or chin contour.

    Device Description

    ShiNeo Silicone Implant is a silicone implant used in facial surgery as nasal and/or chin implants. The ShiNeo Silicone Implant offers two major shapes to meet the needs in nasal and chin locations. The devices are also provided in various sizes and can be carved or cut to fit each patient. The ShiNeo Silicone Implant is individually packaged and sterilized by gamma radiation and is labeled for single use. ShiNeo Silicone Implant is ideal for use in soft tissue augmentation where the use of a soft silicone elastomer is appropriate.

    AI/ML Overview

    This document is a 510(k) summary for the ShiNeo Silicone Implant. It asserts that the device is "identical/substantially equivalent" to a predicate device, the BioSiCar Silicone Implant (K193392).

    Because the submission claims identity between the proposed device and the predicate device, it explicitly states:

    • No performance data was submitted, and no new testing was provided.
    • No acceptance criteria were established or studies conducted to demonstrate performance against such criteria for the ShiNeo Silicone Implant.

    Therefore, the requested information regarding acceptance criteria, performance studies, sample sizes, expert qualifications, ground truth establishment, MRMC studies, or standalone performance simply does not exist within this 510(k) summary, as it wasn't required given the claim of identity. The regulatory strategy relies on showing the new device is the same as an already cleared device, not that it independently meets specific performance metrics through new testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200073

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-09-30

    (990 days)

    Product Code
    Regulation Number
    874.3620
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    BlockClassification Name: Ear, Nose, and Throat Synthetic Polymer MaterialRegulation: 21 CFR §874.3620

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Augmenta Penile Implantable device intended for use in the cosmetic correction of soft tissue deformities, and is trimmed to length at the surgeon's discretion for a custom implant.

    Device Description

    The Augmenta Penile Implant is an implantable device intended for use in the cosmetic correction of soft tissue deformities, and is trimmed to length at the surgeon's discretion for a custom implant. The Augmenta Penile Implant is made of silicone. The Augmenta Penile Implant comes in 175 sizes with a range of dimensions. The proximal end of the device may be trimmed by the physician to further customize the fit. The device is provided sterile and is intended for single-use only.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria and a study that proves a device meets those criteria, as typically seen for AI/ML-based medical devices or diagnostic tools. Instead, it is an FDA 510(k) clearance letter for a physical medical device, the "Augmenta Penile Implant," which is a silicone implant for cosmetic correction of soft tissue deformities.

    The document focuses on demonstrating substantial equivalence to a predicate device (Pre-Formed Penile Silicone Block, K181387) rather than proving performance against specific acceptance criteria for an AI/ML algorithm.

    Therefore, I cannot extract the requested information regarding acceptance criteria, device performance tables, sample sizes for test sets, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, ground truth types, or training set details. These concepts are not applicable to the type of device and clearance described in the provided text.

    The closest analogue to "acceptance criteria" in this document are the non-clinical performance data listed on page 5, which are primarily related to biocompatibility, material integrity, and sterilization for an implantable device, usually demonstrated through adherence to recognized standards like ISO 10993 and ASTM standards.

    Non-Clinical Performance Data (related to material and sterilization safety, not algorithmic performance):

    • Biocompatibility Assessments (in accordance with ISO 10993-1):
      • Cytotoxicity
      • Sensitization
      • Irritation/intracutaneous reactivity
      • Acute Systemic Toxicity
      • Material-Mediated Pyrogenicity
      • Implantation
      • Genotoxicity
      • Extractables and Leachables
      • Toxicology Risk Assessment
    • Product and Packaging Integrity Assessments (in accordance with recognized standards):
      • Visual integrity
      • Seal strength
      • Shipping
      • Burst test
      • Sterilization validation (ISO 111135: 2014)
      • EO residual (ISO 10993-7: 2008)
      • Image artifact evaluations (F2119-07 (2013))
      • ASTM D4169-16
      • ASTM F88/F88M-15
      • ASTM F1140/F1140M-13
    • Conclusion: The claim is that these assessments provide evidence that the Augmenta Penile Implant "performs comparable to the legally marketed predicate device."

    In summary, none of the requested information regarding AI/ML device performance and validation can be provided from this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220760

    Validate with FDA (Live)

    Date Cleared
    2022-05-13

    (59 days)

    Product Code
    Regulation Number
    874.3620
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    92110

    Re: K220760

    Trade/Device Name: Pre-Formed Penile Silicone Block Regulation Number: 21 CFR§ 874.3620
    Penile Silicone Block

    COMMON NAME

    Elastomer, Silicone Block

    DEVICE CLASSIFICATION

    21 CFR 874.3620

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pre-Formed Penile Silicone Block is intended for use in augmentation, reconstructive and cosmetic surgery, and is contoured at the surgeon's discretion to create a custom implant. When used in augmentation procedures, the device provides cosmetic augmentation of the penis and is intended for aesthetic purposes.

    Device Description

    The Pre-Formed Penile Silicone Block is made from medical grade silicone with an embedded polyester mesh. The device comes in variations that include three sizes (L, XL, and XXL) and one durometer ("Soft"). Size "L" is 12 cm in length with a maximum thickness of 0.5 cm and a height of 2 cm. Size "XL" is 15 cm in length with a maximum thickness of 0.8 cm and a height of 3 cm. Size "XXL" is 18 cm in length with a maximum thickness of 1.1 cm and a height of 3.5 cm. The device is used in visual augmentation, reconstructive and cosmetic surgery for the penis, and may be trimmed to allow the surgeon to tailor the device to the needs of a specific patient.

    AI/ML Overview

    The device in question is the Pre-Formed Penile Silicone Block.

    Here's an analysis of the provided text regarding acceptance criteria and the supporting study:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria in a quantitative format for the clinical study. Instead, it describes observed outcomes and improvements. Based on the "Clinical effectiveness" section, we can infer the measures of performance.

    Performance Metric (Implied Acceptance Criteria)Reported Device Performance (Achieved)
    Penile Circumference IncreaseIncreased from 8.5 ± 1.2 cm to 13.4 ± 1.9 cm (56.7% increase, P<.001)
    Self-confidence and Self-esteem Improvement83% of patients noted a 2-category improvement 6-8 weeks postoperatively
    APPSSI Score Improvement (Long-term)72% patients maintained a 2-category improvement upon long-term follow-up (2-6 years, mean 4 years)
    Patient Satisfaction81% of subjects reported "high" or "very high" levels of satisfaction
    Adverse Events requiring device removal3% of patients
    Impact on Sexual FunctionNo reports of any changes to sexual function, erections, or ejaculations
    Most Frequent Postoperative ComplicationsSeroma (4.8%), Scar formation (4.5%), Infection (3.3%)

    2. Sample Size and Data Provenance for the Test Set

    • Sample Size (Test Set):
      • Initial retrospective study included 526 patients.
      • 400 patients responded to the request for consent and were included in the analysis.
    • Data Provenance: Retrospective study. The country of origin is not explicitly stated, but the submission is from "International Medical Devices, Inc." in Beverly Hills, CA, USA, suggesting the study likely took place in the USA.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not mention the use of experts to establish ground truth for the clinical outcomes. The clinical effectiveness was evaluated via patient-reported outcomes (APPSSI, self-confidence/self-esteem, satisfaction) and objective measurements (penile circumference) taken by clinical staff, and complication rates observed by the surgeons.

    4. Adjudication Method for the Test Set

    No adjudication method (e.g., 2+1, 3+1) is mentioned or implied, as the study relies on patient-reported outcomes, objective measurements, and observed complications rather than expert consensus on interpretations.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study is a clinical effectiveness study of the device itself, not an evaluation of an AI system with or without human assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No, a standalone (algorithm only) study was not done. The device is a physical implant, and the study evaluates its clinical performance in human patients. This document does not describe an AI medical device.

    7. Type of Ground Truth Used

    The ground truth used for evaluating the device's clinical effectiveness was a combination of:

    • Objective Measurements: Penile circumference measurements taken before, immediately after, and 30-90 days after surgery.
    • Patient-Reported Outcomes: Changes in the Augmentation Phalloplasty Patient Selection and Satisfaction Inventory (APPSSI), changes in patient self-confidence and self-esteem, and patient-reported satisfaction levels.
    • Clinical Observations: Incidence of adverse events (seroma, scar formation, infection) and events necessitating device removal, as well as absence of changes to sexual function.

    8. Sample Size for the Training Set

    The document describes a clinical study evaluating the performance of the device; it does not mention a training set in the context of machine learning or AI. This is a study of a physical medical device.

    9. How the Ground Truth for the Training Set was Established

    As there is no mention of a training set or an AI algorithm, this question is not applicable to the provided document.