510(k) Data Aggregation

    K Number
    K251208

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-09-26

    (161 days)

    Product Code
    Regulation Number
    874.3335
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K251208**
    Trade/Device Name: Tuned Self-Fitting Mobile App
    Regulation Number: 21 CFR 874.3335
    Name: Air Conduction Hearing Aid Software

    • Product Code: SCR
    • Regulation No: 21 CFR §874.3335
      SPECIAL CONTROLS

    The Tuned Self-Fitting Mobile App conforms to the special controls stated in 21 CFR 874.3335
    | 21 CFR 874.3335 | 21 CFR 874.3325 | Same as predicate |
    | Intended Use | The Tuned Self-Fitting
    predefined design input requirements and to address the special controls provided in regulation 21 CFR 874.3335

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tuned Self-Fitting Mobile App is a software-only mobile medical application that is intended to be used with compatible wireless air-conduction hearing aids. The feature is intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. The Tuned Self-Fitting Mobile App utilizes a self-fitting strategy and is adjusted by the user to meet their hearing needs without the assistance of a hearing healthcare professional. The device is intended for Over-the-Counter use.

    Device Description

    The Tuned Self-Fitting Mobile App is a software-only, over the counter (OTC) medical application designed to support adults (18+) with perceived mild to moderate hearing loss. Available on both iOS and Android platforms, the app allows users to independently tune and personalize the settings of compatible wireless air-conduction hearing aids without requiring assistance from a hearing healthcare professional.

    The Tuned Self-Fitting Mobile App utilizes calibrated pure-tone audiometry to conduct a self-administered hearing test and computes individualized amplification targets based on the NAL-NL2 prescription algorithm. It automatically configures connected hearing aids according to the results and enables additional fine-tuning by the user through intuitive controls for volume, frequency bands (low/mid/high), and preset sound environments such as everyday use, speech enhancement, music, and outdoor noise reduction.

    The reason for this submission is to expand the use of the Tuned Mobile App, which was previously cleared under K223848 for use with the Lumen155 hearing aids (Intricon, USA), to include an additional compatible wireless air-conduction hearing aid, the NoviOne (Novidan, USA).

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243150

    Validate with FDA (Live)

    Date Cleared
    2025-01-31

    (123 days)

    Product Code
    Regulation Number
    874.3335
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Italy

    Re: K243150

    Trade/Device Name: Nuance Audio Hearing Aid Software Regulation Number: 21 CFR 874.3335
    Classification Name: Air Conduction Hearing Aid Software Common Name: Hearing Aid Software Regulation: 21 CFR §874.3335
    |
    | RegulationNumber | 21 CFR 874.3335
    The Nuance Audio Hearing Aid Software also meets the special controls for 874.3335.
    Special control number 1 under 874.3335 requires clinical performance data to validate a hearing aid

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nuance Audio Hearing Aid Software is a software-only mobile medical application that is intended to be used with compatible wearable electronic products. The feature is intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. The software utilizes a preset fitting strategy and is adjusted by the user to meet their hearing needs without the assistance of a hearing healthcare professional. The device is intended for Over-the-Counter use.

    Device Description

    The Nuance Audio Hearing Aid Software is a software-only device that is comprised of a pair of software modules which operate on two separate required products: (1) Nuance Audio Application on a compatible iOS or Android product, and (2) Nuance Audio Hearing Aid Software Firmware (FW) using the compatible wearable electronic computing platform, the Nuance Audio Glasses (smart glasses). The Nuance Audio Hearing Aid Software amplifies sounds for adult users, 18 years of age and older, who have a perceived mild to moderate hearing loss. The Nuance Audio™ App, iOS or Android mobile application, quides the user through the initial onboarding and set up process providing step-by-step instructions to guide the users can select from four preset options, which are generated from typical mild to moderate hearing loss configurations (Urbanski et al., 2021) with NAL-NL2 gain settings applied. The user may also adjust the overall volume, noise reduction, and microphone directionality. The device incorporates a feedback control algorithm to reduce the likelihood of the user experiencing acoustic feedback. The same amplification settings will be applied to both the left and right ear. This device can be fitted by a user without assistance from a hearing healthcare professional.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Nuance Audio Hearing Aid Software, based on the provided text:

    Nuance Audio Hearing Aid Software: Acceptance Criteria and Study Information

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Metric/StandardAcceptance Criteria (Target)Reported Device Performance (Nuance Audio Hearing Aid Software)
    Electro-AcousticsLatency ClauseMeets 21 CFR 800.30 requirements (implicit: similar to predicate or within acceptable tolerance)8.8 ms (Predicate: 3.15 ms – performance testing evaluated differences and supports substantial equivalence)
    Frequency ResponseMeets 21 CFR 800.30 requirements (implicit: similar to predicate or within acceptable tolerance)200Hz - 6300 Hz (Predicate: 100 — 10,000 Hz – performance testing evaluated differences and supports substantial equivalence)
    Self-Generated NoiseMeets 21 CFR 800.30 requirements (implicit: similar to predicate or within acceptable tolerance)29.9 dBA (Predicate: 28 dBA – performance testing evaluated differences and supports substantial equivalence)
    Harmonic DistortionLess than or equal to 1%Less than or equal to 1%
    Maximum Output ValueMeets 21 CFR 800.30 requirements (implicit: similar to predicate or within acceptable tolerance)112.7 dB SPL (Predicate: 106 dB SPL – performance testing evaluated differences and supports substantial equivalence)
    Insertion GainWithin 5 dB of predicate device at most frequencies; within 5 dB of NAL-NL2 targets except at one frequency (5 kHz, within 8 dB).Within 5 dB of predicate device at most frequencies; within 5 dB of NAL-NL2 targets except at one frequency (5 kHz, within 8 dB).
    Usability TestingCritical ErrorsZero critical errors during simulated use scenariosAll tasks completed without critical errors.
    User SatisfactionOverall positive satisfaction on usabilityOverall positive satisfaction on the usability of the device expressed by users.
    Acceptance CriteriaAll defined acceptance criteria for EN 62366-1:2015 metAll defined acceptance criteria were met (EN 62366-1:2015).
    Clinical PerformanceReal-Ear Aided Response (REAR)Each preset (A, B, C, D) within ±6 dB of NAL-NL2 targets at 250, 500, 1000, 2000, and 4000Hz.Each Nuance Audio Hearing Aid Software preset (A, B, C, and D) profile was within the pre-specified criterion range of NAL-NL2 targets (±6 dB).
    Speech-in-Noise (SNR-50)Improvement in SNR (threshold not explicitly stated, but typically a positive improvement is expected)Average unaided vs. aided improvement was 3.49 dB.
    IOI-HA Total ScoreClinically meaningful outcome score for the intended population (threshold not explicitly stated)Mean IOI-HA total score was 29.04 (SD = 4.62), considered clinically meaningful for individuals with perceived mild to moderate hearing loss to achieve hearing benefit.
    General ComplianceSpecial ControlsMeets special controls for 874.3335 (including addressing clinical performance for fitting strategy)Meets special controls for 874.3335, and clinical data provided to address differences in technology (preset vs. self-fitting) to support substantial equivalence.
    OTC Hearing AidsMeets specifications associated with over-the-counter hearing aids (21 CFR 800.30)Nuance Audio Hearing Aid Software was shown to meet the specifications associated with over-the-counter hearing aids (21 CFR 800.30).

    2. Sample Size Used for the Test Set and Data Provenance

    • Usability Testing:
      • Sample Size: 15 representative adult users.
      • Data Provenance: Not explicitly stated, but implies prospective data collection from participants in a usability validation study.
    • Real-ear measurements:
      • Sample Size: 29 participant ears.
      • Data Provenance: Not explicitly stated, but implies prospective data collection from participants.
    • Speech in noise clinical performance:
      • Sample Size: 19 adults with mild to moderate hearing loss.
      • Data Provenance: Not explicitly stated, but implies prospective data collection from participants.
    • Patient Reported Outcome Measure (IOI-HA):
      • Sample Size: 23 adult subjects identified with self-perceived mild-to-moderate sensorineural hearing loss.
      • Data Provenance: Not explicitly stated, but implies prospective data collection after a 14-day trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not explicitly mention the use of experts to establish ground truth for the test set from these clinical studies. Instead, the studies compare the device's performance against:

    • Objective measures: NAL-NL2 targets (a commonly used clinical fitting formula).
    • Standardized clinical outcomes: SNR-50 (speech-in-noise ratio) and IOI-HA scores (patient-reported outcome).
    • Usability standards: IEC 62366-1:2015.

    The ground truth for the "typical mild to moderate hearing loss configurations" used to generate the presets (Urbanski et al., 2021) suggests a reliance on established audiological research rather than individual expert consensus for a specific test set.

    4. Adjudication Method for the Test Set

    No adjudication method (e.g., 2+1, 3+1) is mentioned for any of the test sets. The studies appear to rely on direct measurements, standardized clinical tests, or user feedback.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is reported in relation to human readers improving with AI vs. without AI assistance. The device is a "hearing aid software" rather than an interpretative AI for medical imaging, for example, where MRMC studies are common. The clinical performance studies directly evaluate the device's efficacy.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, the studies described for "Performance Testing" (Electro-Acoustics, Real-ear measurements, Speech in Noise, IOI-HA) represent standalone performance of the algorithm integrated into the compatible hardware (Nuance Audio Glasses) without human-in-the-loop clinical decision-making. The usability testing assesses the user's ability to operate the device effectively, which is human-in-the-loop operation, but the core performance metrics are evaluating the device's output.

    7. Type of Ground Truth Used for the Test Set

    The ground truth for the test sets consisted of:

    • Objective Audiological Targets: NAL-NL2 (National Acoustics Laboratories-Non-Linear 2) fitting formula.
    • Standardized Clinical Metrics: SNR-50 (signal-to-noise ratio at 50% word recognition) and International Outcome Inventory for Hearing Aids (IOI-HA) scores.
    • Usability Standards: Compliance with IEC 62366:2015 and completion of tasks without critical errors.
    • Bench Test Standards: Adherence to ISO, ANSI/ASA, CTA, IEC standards for biocompatibility, electrical safety, EMC, battery safety, and software.

    8. Sample Size for the Training Set

    The document does not specify a sample size for a training set. The device uses "four preset options, which are generated from typical mild to moderate hearing loss configurations (Urbanski et al., 2021) with NAL-NL2 gain settings applied." This suggests that the "training" or development of these presets was based on established audiological research and fitting formulas, rather than a data-driven machine learning training set in the typical sense.

    9. How the Ground Truth for the Training Set Was Established

    The ground truth for the "training" (i.e., the generation of the preset fitting strategies) for the Nuance Audio Hearing Aid Software was established based on:

    • Published Audiological Research: "Typical mild to moderate hearing loss configurations (Urbanski et al., 2021)."
    • Standardized Fitting Formula: Application of the NAL-NL2 (National Acoustics Laboratories-Non-Linear 2) gain settings to these configurations.

    This indicates that the presets are derived from established audiological science and validated hearing loss profiles, rather than being "learned" from a distinct training dataset with a separate ground truth determination process.

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN230081

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-09-12

    (283 days)

    Product Code
    Regulation Number
    874.3335
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    NEW REGULATION NUMBER: 21 CFR 874.3335

    CLASSIFICATION: Class II

    PRODUCT CODE: SCR

    BACKGROUND

    following:

    Product Code: SCR Device Type: Air-conduction hearing aid software Class: II Regulation: 21 CFR 874.3335

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hearing Aid Feature is a software-only mobile medical application that is intended to be used with compatible wearable electronic products. The feature is intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. The Hearing Aid Feature utilizes a self-fitting strategy and is adjusted by the user to meet their hearing needs without the assistance of a hearing healthcare professional. The device is intended for Over-the-Counter use.

    Device Description

    The Hearing Aid Feature (HAF) is a software-only device that is comprised of a pair of software modules which operate on two separate required products: (1) HAF iOS Application on a compatible iOS product, and (2) HAF software (i.e., firmware) on the Apple AirPods Pro 2. Refer to Figure I, middle and right, respectively, The AirPods Pro 2, formerly named AirPods Pro (2nd generation), supported this granting and are hereafter simply referred to as "AirPods Pro" in this document.

    The HAF iOS Application guides users through the onboarding and setup process for the HAF. The process is self-guided by the user and includes step-by-step instructions and informational content (e.g. warnings, instructions for use). To initiate HAF setup, the user must select a saved audiogram from the iOS HealthKit.

    Once the audiogram has been imported by the HAF, the feature will configure the amplification for the user's audiogram based upon Apple's proprietary fitting formula. Once the initial set-up is complete, users can listen with the HAF using the AirPods Pro and refine their settings. Fine tuning is facilitated by user controls on the iOS device that can adjust amplification, tone, and balance. A user can access the fine tuning settings at any time after setting up the HAF.

    The HAF settings are transferred to the HAF Firmware Module on the AirPods Pro. The HAF Firmware Module utilizes the general purpose computing platform features of the AirPods Pro, including the microphone, speakers, amplifiers, and audio processing software, to process incoming sound and provide amplification at a specific frequency and gain based on the user's custom settings. The user's custom settings are stored on the HAF Firmware Module and will be available even when the AirPods Pro are not connected to the iOS device.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Apple's Hearing Aid Feature (HAF)

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Regulatory Standard)Reported Device Performance
    Output Limits (21 CFR 800.30(d))
    1. General output limit (111 dB SPL)N/A (input-controlled compression device)
    2. Output limit for input-controlled compression (117 dB SPL)Max OSPL90: 105.93 dB SPL
    Electroacoustic performance limits (21 CFR 800.30(e))
    1. Output distortion control limits (Total harmonic distortion + noise ≤ 5%)Harmonic distortion does not exceed 1% for any test frequency
    2. Self-generated noise level limits (Self-generated noise ≤ 32 dBA)Max Self-Generated Noise: 28.20 dBA
    3. Latency (Latency ≤ 15 ms)Median Latency: 3.15 ms
    4. Frequency response bandwidth (Lower cutoff ≤ 250 Hz, upper cutoff ≥ 5 kHz)Frequency bandwidth: 100 - 10,000 Hz
    5. Frequency response smoothness (No single peak in one-third-octave response > 12 dB relative to average levels of adjacent bands)All peaks < 12dB relative to adjacent bands, within frequency response bandwidth
    6. Acoustic coupler choice (2-cubic centimeter (cm³) acoustic coupler when compatible)2 cm³ acoustic coupler used
    Design requirements (21 CFR 800.30(f))
    1. Insertion depth (> 10 mm from tympanic membrane)>10mm gap from tympanic membrane. Verified via multiple ear tip sizes, instructions, and usability testing.
    2. Use of atraumatic materialsAirPods Pro platform verified to use atraumatic patient-contacting materials.
    3. Proper physical fitMet for AirPods Pro platform, refer to insertion depth verification.
    4. Tools, tests, or software permit lay user control and customizationHAF fitting customized based on input audiogram; three fine-tuning sliders (amplification, tone, balance) for user customization.
    5. User-adjustable volume controlHAF has an amplification fine-tuning slider to adjust volume.
    6. Adequate reprocessingAdequacy of reprocessing for AirPods Pro platform verified via instructions and design mitigations.
    Clinical Performance - Non-inferiority
    IOI-HA score of Self-Fit group is no more than 3 points below that of Professionally-Fit group.FAS/CCAS set: Mean Difference (Pro-Fit - Self-Fit) = 1.17 (SD 3.34), 95% CI (-0.05, 2.39). P-value = 0.0036. Pass.
    PP set: Mean Difference (Pro-Fit - Self-Fit) = 1.23 (SD 3.34), 95% CI (0.01, 2.46). P-value = 0.0050. Pass.
    Supplemental Clinical Data: Apple Hearing Test Feature Validation
    HTF derived audiograms' pure-tone average similar to professionally derived audiograms.Demonstrated similar pure-tone average for HTF derived audiograms as professionally derived audiograms for the same users (n=202).
    Gain values generated by HAF for HTF vs. professionally-derived audiograms are within +/- 5 dB for >90% of differences.Output gains across all test frequencies were within +/- 5 dB for >98% of gain differences (for subset of n=173 subjects with mild to moderate hearing loss).

    2. Sample Sizes Used for the Test Set and Data Provenance

    Bench/Non-Clinical Tests:

    • Performance Testing (21 CFR 800.30(d) & (e)): No specific sample size (n) is provided, but the tests refer to "all test frequencies" and compliance with ANSI/ASA S3.22 or ANSI/CTA 2051:2017 clauses. This implies comprehensive testing across the specified parameters, rather than a limited sample.
    • Human Factors Formative Testing: 39 subjects.
    • Audiogram Input Risk and Mitigation Study: No specific sample size (n) for the study itself, but refers to the Hearing Test Feature (HTF) validation study dataset.

    Clinical Study:

    • Overall Clinical Study (HAF Self-Fit vs. Professionally-Fit): 118 total participants (59 in Self-Fit group, 59 in Professionally-Fit group for FAS/CCAS; 59 in Self-Fit, 58 in Professionally-Fit for PP analysis).
    • Data Provenance: Prospective, non-significant risk study from three sites across the United States.

    Supplemental Clinical Data (Apple Hearing Test Feature Validation):

    • Comparison of HTF outputs to professionally derived audiograms: n = 202.
    • Gain analysis for HAF with HTF vs. professionally-derived audiograms: n = 173 (subset with mild to moderate hearing loss from the n=202 dataset).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Bench/Non-Clinical Tests:

    • Performance Testing: Ground truth is established by well-defined regulatory standards (21 CFR 800.30) and industry standards (ANSI/ASA S3.22, ANSI/CTA 2051:2017). Expertise is inherent in the test methodologies themselves.
    • Human Factors Formative Testing: No explicitly stated "experts" establishing ground truth in the context of diagnostic accuracy. The testing assessed use-related risks, and findings led to software design modifications.
    • Audiogram Input Risk and Mitigation Study: No explicitly stated "experts" for ground truth on this specific study, but the study for the Hearing Test Feature Validation (see below) involved professional audiograms which would have been established by qualified audiologists.

    Clinical Study (HAF Self-Fit vs. Professionally-Fit):

    • Ground Truth for Professional-Fit (PF) Group: The "Professionally-Fit" group had their hearing aids fitted by an audiologist and underwent an optional audiologist fine-tuning session. This implies a number of audiologists (not specified but plural) provided this professional fit, thus establishing a "ground truth" reference for professional care. The study design intrinsically compares the self-fit approach to professional audiologist care, using the latter as the benchmark for a successful fit in terms of patient-perceived benefit.

    Supplemental Clinical Data (Apple Hearing Test Feature Validation):

    • Comparison to Professionally Derived Audiograms: These would have been established by qualified hearing healthcare professionals, such as audiologists. The exact number of such professionals establishing these audiograms for the 202 subjects is not specified but the term "professionally derived" implies expertise.

    4. Adjudication Method for the Test Set

    Bench/Non-Clinical Tests:

    • Performance Testing: Not applicable; compliance is determined by direct measurement against pre-defined numerical thresholds in regulatory and industry standards.
    • Human Factors Formative Testing: Not applicable; the output is identification of use-related risks and subsequent design modifications.
    • Audiogram Input Risk and Mitigation Study: Not applicable.

    Clinical Study (HAF Self-Fit vs. Professionally-Fit):

    • Primary Endpoint (IOI-HA score): Not applicable in the sense of expert adjudication of a diagnostic finding. The primary outcome was a patient-reported outcome measure (IOI-HA score), a subjective assessment collected directly from participants. The comparison was statistical (non-inferiority margin) between the two groups.
    • Objective Measures (QuickSIN, REM): These are objective measurements and do not require adjudication.

    Supplemental Clinical Data (Apple Hearing Test Feature Validation):

    • Audiogram Comparison & Gain Analysis: Not applicable. The comparison was quantitative (pure-tone average, gain differences) between HTF-derived and professionally-derived audiograms.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a traditional MRMC comparative effectiveness study, as typically seen in diagnostic imaging where multiple readers interpret cases with and without AI assistance, was not performed for the core Hearing Aid Feature.
    • Instead, a clinical study compared two groups:
      • Self-Fit (SF): Users applying the HAF's self-fitting algorithm.
      • Professionally-Fit (PF): Users whose devices were fitted by an audiologist using the NAL-NL2 formula.
    • This study evaluated the effectiveness of the HAF's self-fitting approach directly against professional care by assessing patient-reported outcomes (IOI-HA) and objective measures (QuickSIN, REM).
    • Effect Size of Human Readers Improve with AI vs without AI assistance: This metric is not applicable as the study design was a comparison of a self-fitting AI system against professional human fitting, not a study of human readers improving with AI assistance. The study concluded that the HAF Self-Fit group achieved non-inferior perceived benefit (IOI-HA scores) compared to the Professionally-Fit group, indicating equivalent patient outcomes without the direct involvement of a hearing healthcare professional in the fitting process.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Yes, a standalone study was inherently performed to assess the performance of the HAF's self-fitting algorithm.
    • The "Self-Fit (SF)" group in the clinical study directly represents the standalone performance of the algorithm. These users utilized the HAF's automatic fitting algorithm and then could adjust amplification, tone, and balance themselves. The device's performance, as measured by IOI-HA scores, QuickSIN, and REM, was attributed to this self-fitting strategy.
    • The comparison to the "Professionally-Fit (PF)" group served as a benchmark for what a human expert (audiologist) would achieve.
    • Therefore, the clinical study's SF arm is a direct measure of the algorithm's standalone performance in a real-world setting.

    7. Type of Ground Truth Used

    Bench/Non-Clinical Tests:

    • Regulatory and Industry Standards: Ground truth is defined by explicit numerical thresholds and methodologies prescribed by 21 CFR 800.30 and ANSI/ASA/CTA standards.
    • Human Factors: "Ground truth" is the identification of potential use errors and associated risks, which is derived from observing user interactions and conducting risk analysis.

    Clinical Study (HAF Self-Fit vs. Professionally-Fit):

    • Expert Consensus / Professional Practice: The "Professionally-Fit" group, whose devices were fitted by audiologists using a standard clinical fitting formula (NAL-NL2), served as the gold standard or ground truth for best clinical practice in hearing aid fitting. The HAF's performance was evaluated against this professional benchmark.
    • Patient-Reported Outcomes (IOI-HA): The primary ground truth for effectiveness was the subjective perception of benefit, satisfaction, and quality of life as reported by the patients themselves via the IOI-HA questionnaire.
    • Objective Outcomes Data (QuickSIN, REM): These objective functional measures also served as ground truth regarding speech intelligibility and actual gain.

    Supplemental Clinical Data (Apple Hearing Test Feature Validation):

    • Expert Consensus / Professional Practice: The comparison was made against "professionally derived audiograms," implying that these audiograms, established by trained hearing healthcare professionals, served as the ground truth.

    8. Sample Size for the Training Set

    The document does not explicitly state the sample size for the training set used to develop Apple's proprietary fitting formula within the HAF. The description only refers to the clinical study as establishing safety and effectiveness, and the HAF's fitting formula is described as "Apple's proprietary fitting formula." This formula would have been developed using a separate dataset prior to the validation study.

    9. How the Ground Truth for the Training Set Was Established

    Since the training set size and characteristics are not provided, the method for establishing its ground truth is also not explicitly stated in this document.

    However, based on the nature of hearing aid fitting algorithms and the validation study design, it is highly probable that the proprietary fitting formula was developed and refined using:

    • Large datasets of audiogram data: Likely anonymized audiograms from various sources, potentially including those collected by Apple's own Hearing Test Feature over time or from research collaborations.
    • Patient-reported outcomes data: To correlate objective audiometric data with subjective patient benefit and preference.
    • Expert knowledge/models: Incorporating established audiological principles, validated fitting targets (e.g., NAL-NL2, DSL v5), and clinical experience synthesized into an algorithmic form.
    • Iterative development and testing: The "proprietary fitting formula" would have undergone extensive internal testing, likely including simulations and pilot studies with real users, where the "ground truth" would be established by comparing algorithm outputs to professional recommendations or patient preferences.