    510k Summary Text (Full-text Search) :

    SCR | QUH | Same as predicate |
    | Regulation Number | 21 CFR 874.3335 | 21 CFR 874.3335 | 21 CFR 874.3325

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tuned Self-Fitting Mobile App is a software-only mobile medical application that is intended to be used with compatible wireless air-conduction hearing aids. The feature is intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. The Tuned Self-Fitting Mobile App utilizes a self-fitting strategy and is adjusted by the user to meet their hearing needs without the assistance of a hearing healthcare professional. The device is intended for Over-the-Counter use.

    Device Description

    The Tuned Self-Fitting Mobile App is a software-only, over the counter (OTC) medical application designed to support adults (18+) with perceived mild to moderate hearing loss. Available on both iOS and Android platforms, the app allows users to independently tune and personalize the settings of compatible wireless air-conduction hearing aids without requiring assistance from a hearing healthcare professional.

    The Tuned Self-Fitting Mobile App utilizes calibrated pure-tone audiometry to conduct a self-administered hearing test and computes individualized amplification targets based on the NAL-NL2 prescription algorithm. It automatically configures connected hearing aids according to the results and enables additional fine-tuning by the user through intuitive controls for volume, frequency bands (low/mid/high), and preset sound environments such as everyday use, speech enhancement, music, and outdoor noise reduction.

    The reason for this submission is to expand the use of the Tuned Mobile App, which was previously cleared under K223848 for use with the Lumen155 hearing aids (Intricon, USA), to include an additional compatible wireless air-conduction hearing aid, the NoviOne (Novidan, USA).

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    DEN230081

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-09-12

    (283 days)

    Product Code
    Regulation Number
    874.3335
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    of a hearing-aid device system classified in other regulations, e.g., § 874.3300, § 874.3305, or § 874.3325

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hearing Aid Feature is a software-only mobile medical application that is intended to be used with compatible wearable electronic products. The feature is intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. The Hearing Aid Feature utilizes a self-fitting strategy and is adjusted by the user to meet their hearing needs without the assistance of a hearing healthcare professional. The device is intended for Over-the-Counter use.

    Device Description

    The Hearing Aid Feature (HAF) is a software-only device that is comprised of a pair of software modules which operate on two separate required products: (1) HAF iOS Application on a compatible iOS product, and (2) HAF software (i.e., firmware) on the Apple AirPods Pro 2. Refer to Figure I, middle and right, respectively, The AirPods Pro 2, formerly named AirPods Pro (2nd generation), supported this granting and are hereafter simply referred to as "AirPods Pro" in this document.

    The HAF iOS Application guides users through the onboarding and setup process for the HAF. The process is self-guided by the user and includes step-by-step instructions and informational content (e.g. warnings, instructions for use). To initiate HAF setup, the user must select a saved audiogram from the iOS HealthKit.

    Once the audiogram has been imported by the HAF, the feature will configure the amplification for the user's audiogram based upon Apple's proprietary fitting formula. Once the initial set-up is complete, users can listen with the HAF using the AirPods Pro and refine their settings. Fine tuning is facilitated by user controls on the iOS device that can adjust amplification, tone, and balance. A user can access the fine tuning settings at any time after setting up the HAF.

    The HAF settings are transferred to the HAF Firmware Module on the AirPods Pro. The HAF Firmware Module utilizes the general purpose computing platform features of the AirPods Pro, including the microphone, speakers, amplifiers, and audio processing software, to process incoming sound and provide amplification at a specific frequency and gain based on the user's custom settings. The user's custom settings are stored on the HAF Firmware Module and will be available even when the AirPods Pro are not connected to the iOS device.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Apple's Hearing Aid Feature (HAF)

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Regulatory Standard)Reported Device Performance
    Output Limits (21 CFR 800.30(d))
    1. General output limit (111 dB SPL)N/A (input-controlled compression device)
    2. Output limit for input-controlled compression (117 dB SPL)Max OSPL90: 105.93 dB SPL
    Electroacoustic performance limits (21 CFR 800.30(e))
    1. Output distortion control limits (Total harmonic distortion + noise ≤ 5%)Harmonic distortion does not exceed 1% for any test frequency
    2. Self-generated noise level limits (Self-generated noise ≤ 32 dBA)Max Self-Generated Noise: 28.20 dBA
    3. Latency (Latency ≤ 15 ms)Median Latency: 3.15 ms
    4. Frequency response bandwidth (Lower cutoff ≤ 250 Hz, upper cutoff ≥ 5 kHz)Frequency bandwidth: 100 - 10,000 Hz
    5. Frequency response smoothness (No single peak in one-third-octave response > 12 dB relative to average levels of adjacent bands)All peaks < 12dB relative to adjacent bands, within frequency response bandwidth
    6. Acoustic coupler choice (2-cubic centimeter (cm³) acoustic coupler when compatible)2 cm³ acoustic coupler used
    Design requirements (21 CFR 800.30(f))
    1. Insertion depth (> 10 mm from tympanic membrane)>10mm gap from tympanic membrane. Verified via multiple ear tip sizes, instructions, and usability testing.
    2. Use of atraumatic materialsAirPods Pro platform verified to use atraumatic patient-contacting materials.
    3. Proper physical fitMet for AirPods Pro platform, refer to insertion depth verification.
    4. Tools, tests, or software permit lay user control and customizationHAF fitting customized based on input audiogram; three fine-tuning sliders (amplification, tone, balance) for user customization.
    5. User-adjustable volume controlHAF has an amplification fine-tuning slider to adjust volume.
    6. Adequate reprocessingAdequacy of reprocessing for AirPods Pro platform verified via instructions and design mitigations.
    Clinical Performance - Non-inferiority
    IOI-HA score of Self-Fit group is no more than 3 points below that of Professionally-Fit group.FAS/CCAS set: Mean Difference (Pro-Fit - Self-Fit) = 1.17 (SD 3.34), 95% CI (-0.05, 2.39). P-value = 0.0036. Pass.
    PP set: Mean Difference (Pro-Fit - Self-Fit) = 1.23 (SD 3.34), 95% CI (0.01, 2.46). P-value = 0.0050. Pass.
    Supplemental Clinical Data: Apple Hearing Test Feature Validation
    HTF derived audiograms' pure-tone average similar to professionally derived audiograms.Demonstrated similar pure-tone average for HTF derived audiograms as professionally derived audiograms for the same users (n=202).
    Gain values generated by HAF for HTF vs. professionally-derived audiograms are within +/- 5 dB for >90% of differences.Output gains across all test frequencies were within +/- 5 dB for >98% of gain differences (for subset of n=173 subjects with mild to moderate hearing loss).

    2. Sample Sizes Used for the Test Set and Data Provenance

    Bench/Non-Clinical Tests:

    • Performance Testing (21 CFR 800.30(d) & (e)): No specific sample size (n) is provided, but the tests refer to "all test frequencies" and compliance with ANSI/ASA S3.22 or ANSI/CTA 2051:2017 clauses. This implies comprehensive testing across the specified parameters, rather than a limited sample.
    • Human Factors Formative Testing: 39 subjects.
    • Audiogram Input Risk and Mitigation Study: No specific sample size (n) for the study itself, but refers to the Hearing Test Feature (HTF) validation study dataset.

    Clinical Study:

    • Overall Clinical Study (HAF Self-Fit vs. Professionally-Fit): 118 total participants (59 in Self-Fit group, 59 in Professionally-Fit group for FAS/CCAS; 59 in Self-Fit, 58 in Professionally-Fit for PP analysis).
    • Data Provenance: Prospective, non-significant risk study from three sites across the United States.

    Supplemental Clinical Data (Apple Hearing Test Feature Validation):

    • Comparison of HTF outputs to professionally derived audiograms: n = 202.
    • Gain analysis for HAF with HTF vs. professionally-derived audiograms: n = 173 (subset with mild to moderate hearing loss from the n=202 dataset).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Bench/Non-Clinical Tests:

    • Performance Testing: Ground truth is established by well-defined regulatory standards (21 CFR 800.30) and industry standards (ANSI/ASA S3.22, ANSI/CTA 2051:2017). Expertise is inherent in the test methodologies themselves.
    • Human Factors Formative Testing: No explicitly stated "experts" establishing ground truth in the context of diagnostic accuracy. The testing assessed use-related risks, and findings led to software design modifications.
    • Audiogram Input Risk and Mitigation Study: No explicitly stated "experts" for ground truth on this specific study, but the study for the Hearing Test Feature Validation (see below) involved professional audiograms which would have been established by qualified audiologists.

    Clinical Study (HAF Self-Fit vs. Professionally-Fit):

    • Ground Truth for Professional-Fit (PF) Group: The "Professionally-Fit" group had their hearing aids fitted by an audiologist and underwent an optional audiologist fine-tuning session. This implies a number of audiologists (not specified but plural) provided this professional fit, thus establishing a "ground truth" reference for professional care. The study design intrinsically compares the self-fit approach to professional audiologist care, using the latter as the benchmark for a successful fit in terms of patient-perceived benefit.

    Supplemental Clinical Data (Apple Hearing Test Feature Validation):

    • Comparison to Professionally Derived Audiograms: These would have been established by qualified hearing healthcare professionals, such as audiologists. The exact number of such professionals establishing these audiograms for the 202 subjects is not specified but the term "professionally derived" implies expertise.

    4. Adjudication Method for the Test Set

    Bench/Non-Clinical Tests:

    • Performance Testing: Not applicable; compliance is determined by direct measurement against pre-defined numerical thresholds in regulatory and industry standards.
    • Human Factors Formative Testing: Not applicable; the output is identification of use-related risks and subsequent design modifications.
    • Audiogram Input Risk and Mitigation Study: Not applicable.

    Clinical Study (HAF Self-Fit vs. Professionally-Fit):

    • Primary Endpoint (IOI-HA score): Not applicable in the sense of expert adjudication of a diagnostic finding. The primary outcome was a patient-reported outcome measure (IOI-HA score), a subjective assessment collected directly from participants. The comparison was statistical (non-inferiority margin) between the two groups.
    • Objective Measures (QuickSIN, REM): These are objective measurements and do not require adjudication.

    Supplemental Clinical Data (Apple Hearing Test Feature Validation):

    • Audiogram Comparison & Gain Analysis: Not applicable. The comparison was quantitative (pure-tone average, gain differences) between HTF-derived and professionally-derived audiograms.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a traditional MRMC comparative effectiveness study, as typically seen in diagnostic imaging where multiple readers interpret cases with and without AI assistance, was not performed for the core Hearing Aid Feature.
    • Instead, a clinical study compared two groups:
      • Self-Fit (SF): Users applying the HAF's self-fitting algorithm.
      • Professionally-Fit (PF): Users whose devices were fitted by an audiologist using the NAL-NL2 formula.
    • This study evaluated the effectiveness of the HAF's self-fitting approach directly against professional care by assessing patient-reported outcomes (IOI-HA) and objective measures (QuickSIN, REM).
    • Effect Size of Human Readers Improve with AI vs without AI assistance: This metric is not applicable as the study design was a comparison of a self-fitting AI system against professional human fitting, not a study of human readers improving with AI assistance. The study concluded that the HAF Self-Fit group achieved non-inferior perceived benefit (IOI-HA scores) compared to the Professionally-Fit group, indicating equivalent patient outcomes without the direct involvement of a hearing healthcare professional in the fitting process.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Yes, a standalone study was inherently performed to assess the performance of the HAF's self-fitting algorithm.
    • The "Self-Fit (SF)" group in the clinical study directly represents the standalone performance of the algorithm. These users utilized the HAF's automatic fitting algorithm and then could adjust amplification, tone, and balance themselves. The device's performance, as measured by IOI-HA scores, QuickSIN, and REM, was attributed to this self-fitting strategy.
    • The comparison to the "Professionally-Fit (PF)" group served as a benchmark for what a human expert (audiologist) would achieve.
    • Therefore, the clinical study's SF arm is a direct measure of the algorithm's standalone performance in a real-world setting.

    7. Type of Ground Truth Used

    Bench/Non-Clinical Tests:

    • Regulatory and Industry Standards: Ground truth is defined by explicit numerical thresholds and methodologies prescribed by 21 CFR 800.30 and ANSI/ASA/CTA standards.
    • Human Factors: "Ground truth" is the identification of potential use errors and associated risks, which is derived from observing user interactions and conducting risk analysis.

    Clinical Study (HAF Self-Fit vs. Professionally-Fit):

    • Expert Consensus / Professional Practice: The "Professionally-Fit" group, whose devices were fitted by audiologists using a standard clinical fitting formula (NAL-NL2), served as the gold standard or ground truth for best clinical practice in hearing aid fitting. The HAF's performance was evaluated against this professional benchmark.
    • Patient-Reported Outcomes (IOI-HA): The primary ground truth for effectiveness was the subjective perception of benefit, satisfaction, and quality of life as reported by the patients themselves via the IOI-HA questionnaire.
    • Objective Outcomes Data (QuickSIN, REM): These objective functional measures also served as ground truth regarding speech intelligibility and actual gain.

    Supplemental Clinical Data (Apple Hearing Test Feature Validation):

    • Expert Consensus / Professional Practice: The comparison was made against "professionally derived audiograms," implying that these audiograms, established by trained hearing healthcare professionals, served as the ground truth.

    8. Sample Size for the Training Set

    The document does not explicitly state the sample size for the training set used to develop Apple's proprietary fitting formula within the HAF. The description only refers to the clinical study as establishing safety and effectiveness, and the HAF's fitting formula is described as "Apple's proprietary fitting formula." This formula would have been developed using a separate dataset prior to the validation study.

    9. How the Ground Truth for the Training Set Was Established

    Since the training set size and characteristics are not provided, the method for establishing its ground truth is also not explicitly stated in this document.

    However, based on the nature of hearing aid fitting algorithms and the validation study design, it is highly probable that the proprietary fitting formula was developed and refined using:

    • Large datasets of audiogram data: Likely anonymized audiograms from various sources, potentially including those collected by Apple's own Hearing Test Feature over time or from research collaborations.
    • Patient-reported outcomes data: To correlate objective audiometric data with subjective patient benefit and preference.
    • Expert knowledge/models: Incorporating established audiological principles, validated fitting targets (e.g., NAL-NL2, DSL v5), and clinical experience synthesized into an algorithmic form.
    • Iterative development and testing: The "proprietary fitting formula" would have undergone extensive internal testing, likely including simulations and pilot studies with real users, where the "ground truth" would be established by comparing algorithm outputs to professional recommendations or patient preferences.
    Ask a Question

    Ask a specific question about this device

    K Number
    K233747

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-03-15

    (114 days)

    Product Code
    Regulation Number
    874.3325
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    01590

    Re: K233747

    Trade/Device Name: Concha Sol Hearing Aids (CL-1001) Regulation Number: 21 CFR 874.3325
    Hearing AidClassification Name: Self-fitting Air Conduction Hearing AidRegulation: 21 CFR §874.3325
    Aids (Model CL-1001) are receiver-in-the ear, self-fitting, wireless air conduction hearing aids (§ 874.3325
    complies with OTC hearingaid requirements outlined in21 CFR 800.30. |
    | Regulation number | 21 CFR 874.3325
    | 21 CFR 874.3325

    Device Description :

    Aids (Model CL-1001) are receiver-in-the ear, self-fitting, wireless air conduction hearing aids (§ 874.3325

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Concha Sol Hearing Aids are intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. They are adjusted by the user to meet the user's hearing needs. No pre-programming or hearing test is necessary. The device is intended for use without the assistance of a hearing care professional.

    Device Description

    The Concha Sol OTC Hearing Aids (Model CL-1001) are receiver-in-the ear, self-fitting, wireless air conduction hearing aids (§ 874.3325) intended to amplify sounds in the environment for adult users who have a perceived mild-to-moderate hearing loss. The hearing aids include Ear Tips, Carry Case and the Concha Labs mobile application. The device is indicated for over-the-counter sale (product code QUH) and integrates user input with a self fitting strategy, enabling users to independently derive and customize their hearing aid fitting and settings. Self-fitting wireless air-conduction hearing aids are Class II medical devices. The Self-Fitting hearing aids incorporate microphones for audio input, and sound is delivered to the ear via a receiver that can be coupled with domes. The mobile application contains a self-fitting software known as Soundscope, an innovative software that uses the NAL-NL2 programming formula used globally by hearing aid manufacturers. The hearing profile/personalization system in the Concha Sol Hearing Aid assesses the individual's hearing levels across a frequency spectrum in order to generate and customize the device's amplification specific to their hearing needs. The self-fitting software also assesses the individual's hearing preferences to help fine-tune the device's amplification. In addition to hearing aid functionality for environmental and directive listening; the hearing aids can also be used for telephony and streaming audio from a Bluetooth® compliant mobile device. The hearing aids are powered using Zinc Air size 312 cell batteries.

    AI/ML Overview

    The Concha Sol Hearing Aids (CL-1001) were evaluated for their effectiveness and safety through a clinical study. The study objective was to assess the performance and clinical effectiveness of the self-fitting hearing aid using two distinct fitting methods: SELF-FIT (user-fitted with the Concha Labs App) and CLIN-FIT (clinically fitted by a hearing care professional).

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Study Endpoint)Target Performance (Margin)Reported Device Performance (Concha Sol SELF-FIT vs. CLIN-FIT)Outcome
    Effectiveness:
    Quick Speech-in-Noise (QuickSIN)SELF-FIT outcomes comparable to CLIN-FIT within a margin of 1.8 dB for QuickSIN benefitNot explicitly stated as numerical difference or statistical test result but indicated as "comparable performance"Met: "For both assessed endpoints, the study results indicated comparable performance and satisfaction outcomes as revealed from clinical data as compared to the same hearing aid fit by a clinician." This implies the 1.8 dB margin was met or that the difference was not statistically significant beyond this margin.
    Abbreviated Profile of Hearing Aid Benefit (APHAB)SELF-FIT outcomes comparable to CLIN-FIT within a margin of 9.9 for APHABNot explicitly stated as numerical difference or statistical test result but indicated as "comparable performance"Met: "For both assessed endpoints, the study results indicated comparable performance and satisfaction outcomes as revealed from clinical data as compared to the same hearing aid fit by a clinician." This implies the 9.9 margin was met or that the difference was not statistically significant beyond this margin.
    Safety:No device-related or other adverse eventsNo device-related or other adverse events observedMet: "No device-related or other adverse events were observed over the course of the study. Also, no health-related issues due to device fit or discomfort were observed, and no device failures leading to potential safety issues were observed."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 21 participants (13 men, 8 women).
    • Data Provenance: The document does not explicitly state the country of origin. The study design ("within-subject, randomized, crossover study design of the clinical investigation") suggests it was a prospective clinical study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: The document details the "CLIN-FIT" method, which involved fitting by a "hearing care professional." It does not specify the number of individual hearing care professionals involved or their specific qualifications (e.g., years of experience).

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable in the traditional sense for these endpoints. The study compared two fitting methods (self-fit vs. clinician-fit) within the same subjects, using standardized objective (QuickSIN) and subjective (APHAB) outcome measures. The "ground truth" for the CLIN-FIT condition was established by the hearing care professional's fitting, which was then used as a reference for comparison.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a MRMC comparative effectiveness study was not explicitly described. The study design was a "within-subject, randomized, crossover study" to compare the performance of the Concha Sol device under two different fitting methods (SELF-FIT vs. CLIN-FIT). While human readers (users and a clinician) were involved in the fitting process and perception, it was not framed as an MRMC study comparing interpretations of diagnostic cases.
    • Effect Size of Human Readers Improvement with AI vs. Without AI Assistance: Not applicable, as this was not a diagnostic AI study. The "AI" component would be the self-fitting software (Soundscope), which the users interact with directly. The comparison was between user-driven fitting and clinician-driven fitting of the same device.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • No, a standalone (algorithm only) performance study was not explicitly described for clinical effectiveness. The clinical study inherently involved human feedback and interaction (either by the user for SELF-FIT or by the clinician for CLIN-FIT). The Soundscope software's performance was evaluated as part of the overall device system in conjunction with human use. Electroacoustic testing provided standalone performance metrics for the hardware and general sound processing, but not specifically for the self-fitting algorithm's clinical outcomes without human interaction.

    7. Type of Ground Truth Used

    • Ground Truth: The study used the performance of the device when fitted by a hearing care professional (CLIN-FIT) as a reference "ground truth" for comparison against the user's self-fitting (SELF-FIT) results. This is an expert-derived benchmark using established clinical practices.
      • QuickSIN: Objective measure of speech communication in noise.
      • APHAB: Subjective measure of hearing impairment and benefit in daily life.
      • Safety: Directly observed adverse events and health-related issues.

    8. Sample Size for the Training Set

    • Not Applicable / Not Provided: The document describes a clinical validation study for the Concha Sol device. It does not mention a separate "training set" for the self-fitting algorithm. The Soundscope software uses the NAL-NL2 programming formula, which is an established, globally used, prescriptive target rule for hearing aid fitting, implying its development and "training" occurred much earlier and separately from this device-specific validation.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable / Not Provided: As noted above, the document does not describe a training set for the self-fitting algorithm within the context of this submission. The NAL-NL2 (National Acoustic Laboratories - Nonlinear 2) formula, which the Soundscope technology utilizes, is a well-established prescriptive method in audiology. Its "ground truth" (i.e., its derivation and validation) would stem from extensive research, clinical trials, and epidemiological data in hearing science involving human perception, speech intelligibility, and comfort over many years, conducted by the National Acoustic Laboratories (Australia) and other research institutions. This is a foundational audiological principle rather than a training set specifically for this device's algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K233464

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-02-09

    (109 days)

    Product Code
    Regulation Number
    874.3325
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    20004

    Re: K233464

    Trade/Device Name: WSA Self-Fitting Hearing Aid Gen 2 Regulation Number: 21 CFR 874.3325
    Air-Conduction Hearing Aid |
    | | Regulatory Number: | 21 C.F.R. § 874.3325
    hearing aidregulations sinceclearance of predicate. |
    | ClassificationRegulation | 21 CFR 874.3325
    | 21 CFR 874.3325

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The WSA Self-Fitting Hearing Aid Gen 2 is intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. It is adjusted by the user to meet the user's hearing needs through software tools. The device is intended for direct-to-consumer sale and use without the assistance of a hearing care professional.

    The SF-App is intended to support self-fitting and fine-tuning of the hearing aid.

    Device Description

    The WSA Self-Fitting Hearing Aid Gen 2 ("SF Gen 2") is a self-fitting air conduction hearing aid that is intended to compensate for impaired hearing and incorporates technology, including software, that allows users to program their hearing aids. This technology integrates user input with a self-fitting strategy and enables users to independently derive and customize their hearing aid fitting and settings.

    The SF Gen 2 hardware includes: Earpieces (sleeves), housing, circuit board including a chip set running embedded software, antenna for near-field magnetic inductive wireless connection for data exchange between the left and right hearing aids, rechargeable battery, and electroacoustic components. The instant fit in-the ear (ITE) style of the hearing aids uses sleeves in 4 different sizes to couple with the ear canal. The sleeves are available in vented and closed styles. The hearing aids are powered by built-in rechargeable Lithium-lon batteries. The device includes embedded software that communicates with the self-fitting software app.

    The SF Gen 2 self-fitting system consists of: (a) a wireless hearing aid with a self-fitting feature, and (b) a software application to support hearing aid self-fitting and fine tuning. The subject device is fitted for bilateral use. Left and right hearing aids can communicate with each other over a magnetic inductive wireless link, and the hearing aids communicate with the self-fitting web application (the "self-fitting app") using acoustic technology.

    The SF Gen 2 self-fitting app is run on an Internet browser that is integrated into a software app that is downloaded onto and accessed through the user's mobile device. The self-fitting strategy begins with an acoustic profiling test that the user performs while wearing the hearing aids. Based on the user's hearing profile, a "cluster" of initial settings is applied to each hearing aid, and users are then guided through fine-tuning of these settings to self-fit the hearing aids for their particular needs and comfort. The mobile app also allows the user to adjust audiological gain parameters and preferred settings on the hearing aid, and functions as a remote control for adjusting hearing aid volume during daily use.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "WSA Self-Fitting Hearing Aid Gen 2" and claims substantial equivalence to a predicate device, the "Vibe SF Self-Fitting Hearing Aid (K220403)".

    While it mentions performance testing for various engineering properties, it explicitly states that "no new clinical performance testing was required." This means there is no detailed study described that establishes a specific acceptance criterion (e.g., in terms of hearing improvement or speech understanding) and then proves the device meets it through a clinical trial. Instead, the argument for safety and effectiveness is based on the device's substantial equivalence to a previously cleared predicate and non-clinical performance testing of its components.

    Therefore, the requested information elements related to clinical study design (sample size, ground truth, expert opinions, MRMC, standalone performance) cannot be fully answered as these types of studies were not conducted or described for this specific submission as per the provided text.

    Here's the information that can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance

    PropertyAcceptance Criteria (Implied by Standards/Regulations)Reported Device Performance (SF Gen 2)
    Electrical Safety & EMCCompliance with IEC 60601-1 Ed. 3.2:2020, IEC 60601-1-2 Ed. 4.1:2020, IEC 60601-1-11 Ed. 2.1:2020, IEC 60601-2-66 Ed. 3.0:2019Pass
    Wireless Functionality & CoexistenceCompliance with AAMI TIR 69:2017, ANSI C63.27:2021Pass
    Electroacoustic PerformanceCompliance with ANSI/ASA S3.22:2014, ANSI/CTA 2051:2017, 21 CFR § 800.30Pass
    Max OSPL 90Adequate for fitting moderate hearing loss (55 dB HL) as prescribed by NAL-NL2. (Predicate: 114 dB SPL)114 dB SPL
    Fitted OSPL 90Adequate for fitting moderate hearing loss (55 dB HL) as prescribed by NAL-NL2. (Predicate: 103 dB SPL)99 dB SPL
    Full-on gainAdequate for fitting moderate hearing loss (55 dB HL) as prescribed by NAL-NL2. (Predicate: 45 dB)44 dB
    Frequency RangeAdequate for fitting moderate hearing loss as prescribed by NAL-NL2. (Predicate: 100 – 8600 Hz)250 – 8000 Hz or 100-9100 Hz
    Input Noise (Self-generated)Comply to the required limit of 32 dB A per 21 CFR 800.30(e)(2). (Predicate: 20 dBA)23 dBA
    Software Verification / ValidationCompliance with IEC 62304 A1:2016Pass
    Battery SafetyCompliance with IEC 62133-2:2017, UL 1642:2014Pass
    Risk AssessmentCompliance with ISO 14971:2019, AAMI TIR57:2016Pass

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not describe a clinical test set or patient data used for evaluation of the device's hearing performance. The "Performance Testing" section refers to non-clinical (engineering) tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as no clinical test set with human-established ground truth for hearing performance is described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is a self-fitting hearing aid, not an AI diagnostic tool primarily evaluated with human readers. The submission explicitly states "no new clinical performance testing was required."

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The document describes non-clinical technical testing (e.g., electroacoustic performance, software validation) of the device's components and overall system. While these tests evaluate the "algorithm" or device's standalone performance in specific technical domains, it does not refer to "standalone performance" in the context of a clinical outcome that would typically be compared to human performance (e.g., a standalone AI diagnostic algorithm).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical performance tests, the "ground truth" or reference is typically the specified international standards and regulations (e.g., IEC, ANSI, CFR requirements). For example, electroacoustic performance is compared against the limits defined in ANSI/ASA S3.22:2014.

    8. The sample size for the training set

    Not applicable. The document does not describe the development or training of a machine learning model beyond the general mention of a "self-fitting strategy" which "integrates user input with a self-fitting strategy." It does not specify a separate training set for a machine learning algorithm.

    9. How the ground truth for the training set was established

    Not applicable, as no training set for a machine learning algorithm is described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231550

    Validate with FDA (Live)

    Date Cleared
    2023-10-26

    (149 days)

    Product Code
    Regulation Number
    874.3325
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Georgia 30338

    Re: K231550

    Trade/Device Name: Sontro® OTC Hearing Aids (AI) Regulation Number: 21 CFR 874.3325
    Self-Fitting Wireless Air-Conduction Hearing Aid, Over-the-Counter |
    | Regulation Number | 21 CFR 874.3325
    Device Description

    Per 21 CFR 874.3325, a self-fitting wireless air-conduction hearing aid is a wearable
    Device Information

    Per 21 CFR 874.3325, a self-fitting wireless air-conduction hearing aid is a wearable
    Controls

    The Sontro Self-Fitting OTC Hearing Aids conform to the special controls stated in 21 CFR 874.3325

    Device Description :

    Per 21 CFR 874.3325, a self-fitting wireless air-conduction hearing aid is a wearable soundamplifying

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Soundwave Hearing Sontro® Self-Fitting OTC Hearing Aids are intended to amplify sounds for use by individuals 18 years and older with perceived mild to moderate hearing impairment. The device is intended for over-the-counter sale and use without the assistance of a hearing care professional.

    Device Description

    Per 21 CFR 874.3325, a self-fitting wireless air-conduction hearing aid is a wearable soundamplifying device intended to compensate for impaired hearing. It incorporates technology, including software, that allows users to program their hearing aids. This technology integrates user input with a self-fitting strategy, enabling users to independently derive and customize their hearing aid fitting and settings. This is an over-the-counter hearing aid.

    The Soundwave Hearing, Sontro® OTC (over-the-counter) Hearing Aids are self-fitting wireless air conduction hearing aids consisting of the hardware and software: the device uses the otoTune® app and accessories supplied in the carton. Each hearing aid in the pair functions and interacts with the otoTune® app independently and as a system.

    A disposable size 312 zinc-air battery powers the hearing aids. The hearing aids incorporate microphones for audio input, and sound is delivered to the ear via a receiver in the canal that can be coupled with ear domes. The hearing aids are controlled via onboard button controls and wirelessly via the otoTune® app (iOS and Android). The controls allow the user to configure parameters, settings, and listening modes.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Sontro® OTC Hearing Aids (AI), based on the provided document:

    Acceptance Criteria and Device Performance

    Acceptance Criteria (Endpoint)Reported Device PerformanceNon-inferiority Criterion Met?
    Primary Effectiveness:
    SELF-FIT REAR equivalent to CLINIC-FIT NAL/NL2 prescription REAR targets (in-lab)Average REAR differences (CLINIC-FIT - SELF-FIT) for both ears: 0.17 dB. 95% CI: (-0.44, 0.78).Yes (CI within +/- 5 dB)
    SELF-FIT APHAB global score non-inferior to CLINIC-FIT APHAB global scores (wear-time field trial)Average APHABglobal differences (CLINIC-FIT - SELF-FIT): 1.93. 95% CI: (-4.35, 8.21).Yes (CI upper bound < 8.4)
    Primary Reliability:
    SELF-FIT REAR (A) equivalent to SELF-FIT REAR (B) (in-lab)Average REAR differences (SELF-FIT A - SELF-FIT B) for both ears: 0.07 dB. 95% CI: (-0.88, 0.73).Yes (CI within +/- 5 dB)
    Secondary Effectiveness:
    SELF-FIT QuickSIN equivalent to CLINIC-FIT QuickSIN (in-lab)Average QuickSIN SNR loss differences (CLINIC-FIT - SELF-FIT): -0.38 dB. 95% CI: (-0.94, 0.18).Yes (CI lower bound > -1.5 dB)
    SELF-FIT QuickSIN equivalent to CLINIC-FIT QuickSIN (wear-time field trial)Average QuickSIN SNR loss differences (CLINIC-FIT - SELF-FIT): -0.06 dB. 95% CI: (-1.47, 1.36).Yes (CI lower bound > -1.5 dB)
    REAR for FINAL-FIT equivalent to REAR for AFTERTRIAL-FIT (wear-time field trial)Average REAR differences (AFTERTRIAL-FIT - FINAL-FIT) for both ears: -0.22 dB. 95% CI: (-0.87, 0.42).Yes (CI within +/- 5 dB)
    Secondary Reliability:
    SELF-FIT QuickSIN (A) equivalent to SELF-FIT QuickSIN (B) (in-lab)Average QuickSIN SNR loss differences (SELF-FIT A - SELF-FIT B): 0.28 dB. 95% CI: (-0.48, 1.03).Yes (CI lower bound > -1.5 dB)

    Study Details

    1. Sample Size used for the test set and the data provenance:

      • Sample Size: 29 participants were enrolled in the study (13 men, 16 women). 17 subjects withdrew due to poor compliance or personal reasons, meaning the final analysis was likely based on 12 participants.
      • Data Provenance: The document does not explicitly state the country of origin but implies a single clinical study setting. It was a prospective study involving participants with perceived mild to moderate hearing impairment.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The "ground truth" for the test set in this context refers to the CLINIC-FIT measurements. These were established by an "audiology-best-practices hearing aid fitting," which implies fitting by a qualified audiologist. The document does not specify the exact number of audiologists or their years of experience but indicates adherence to "audiology-best-practices."
    3. Adjudication method for the test set:

      • The clinical study compared the device's self-fitting method (SELF-FIT) against a clinician-fitted method (CLINIC-FIT). There isn't an explicit "adjudication method" described in the sense of multiple experts reviewing cases. Instead, the "ground truth" (CLINIC-FIT) was established by standard audiological best practices. The comparison was statistical, assessing non-inferiority between the SELF-FIT and CLINIC-FIT results.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a traditional MRMC study comparing human readers with and without AI assistance was not explicitly stated or conducted as described. This study focuses on the effectiveness of a self-fitting hearing aid compared to a clinician-fitted hearing aid, not on AI assistance for human "readers" of medical images or data.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, the "SELF-FIT" arm of the study represents the performance of the device's self-fitting algorithm without direct human (audiologist) intervention in the fitting process. The user themselves interacts with the app to perform the fitting. However, it's important to note that the user is "in the loop" for the self-fitting process, but an audiologist is not. So while not a fully "algorithm-only" standalone per se, it is standalone clinician-free fitting.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth for comparison ("reference standard") was expert-established audiological measurements and patient-reported outcomes, specifically:
        • Real-Ear Aided Response (REAR) based on audiology-best-practices NAL/NL2 prescription, representing an objective acoustic measure.
        • Abbreviated Profile of Hearing Aid Benefit (APHAB) global scores, representing subjective patient outcomes.
        • QuickSIN test results, representing an objective measure of speech recognition in noise.
    7. The sample size for the training set:

      • The document does not provide information regarding a specific training set size for the self-fitting algorithm. The focus of this submission is the clinical validation of the device itself, which incorporates the self-fitting technology. The development and training data for the underlying algorithm are not detailed here.
    8. How the ground truth for the training set was established:

      • As there is no information on a specific training set in the provided text, there is also no information on how the ground truth for such a set was established. It can be inferred that the NAL-NL2 fitting algorithm, which the device utilizes, is a widely validated and established prescriptive method in audiology, derived from extensive research and clinical data over time, rather than a specific "training set" for this particular device submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223911

    Validate with FDA (Live)

    Device Name
    Lumen 155-SF
    Date Cleared
    2023-10-17

    (292 days)

    Product Code
    Regulation Number
    874.3325
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Arden Hills, Minnesota 55112

    Re: K223911

    Trade/Device Name: Lumen 155-SF Regulation Number: 21 CFR 874.3325
    K223911 |
    | Regulation Number: | 21 CFR 874.3325
    DEVICE DESCRIPTION

    Per 21 CFR 874.3325 a self-fitting wireless air conduction hearing aid is a wearable
    the electroacoustic data as found in the predicate and required under Special Controls for 21 CFR §874.3325
    Conclusion

    The special controls for self-fitting hearing aids as per 21 CFR 874.3325, are met in the

    Device Description :

    Per 21 CFR 874.3325 a self-fitting wireless air conduction hearing aid is a wearable sound amplifying

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lumen® 155-SF self-fitting, wireless air conduction hearing aids are intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. They are adjusted by the user to meet the user's hearing needs. The device is intended for over-the-counter sales without the assistance of a hearing health care professional.

    Device Description

    Per 21 CFR 874.3325 a self-fitting wireless air conduction hearing aid is a wearable sound amplifying device that is intended to compensate for impaired hearing and incorporates technology, including software, that allows users to program their hearing aids. This technology integrates user input with a self-fitting strategy and enables users to independently derive and customize their hearing aid fitting and settings. Self-fitting wireless air conduction hearing aids are class II medical devices.

    The Lumen 155-SF self-fitting hearing aid(s) with Sentibo application is a self-fitting wireless air conduction Behind-The-Ear (BTE) hearing aid consisting of the Intricon hardware and Sentibo software to be used on an Apple mobile device (mobile device not included) including wired lightning connector Apple EarPods for the assessment (not included) designed for a single user. The Sentibo app is available only on iOS. The Lumen 155-SF self-fitting OTC hearing aid(s) with Sentibo application is a behind the ear device that includes self-adjustable coupling by means of a slim tube and ear tip/open dome. The hearing aids can be fine-tuned by the user or remotely at the request of the user via the use of QR codes. Once the hearing aids are switched on, the software (Sentibo app) is required for the initial set-up, which is done via the smart phone and requires Apple EarPods with lightning connector to complete (not included). The Sentibo self-fitting software allows the consumer to self-select a hearing profile from a list of options by listening to digitized speech embedded in the software while scrolling through the hearing profiles in real-time through the Apple EarPods. After the setting is selected, the settings can be transferred from the smart phone to the hearing aid directly, though direct audio streaming from the smart phone to the hearing aid is unsupported with the Lumen 155-SF firmware. All Sentibo settings are stored in the application itself, and no internet service is required to perform the self-fitting. The Sentibo application provides 24 predefined options. At the conclusion of the assessment, the software will program hearing aids wirelessly and the user will wear the hearing aids per normal use. The smartphone app will function as a software accessory, allowing the user to make minor adjustments to the hearing aids. The hearing aid is intended to be used with the Sentibo App and to be worn and removed daily by the end user.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/TestsReported Device Performance (Lumen 155-SF)Predicate/Reference PerformanceDiscussion of Differences & Meeting Criteria
    Intended UseAmplifies sound for adults 18+ with perceived mild to moderate hearing impairment; user-adjusted; OTC sales without hearing health professional assistance.Meets this exact indication.Predicate (Nuheara IQbuds 2 PRO) and Reference (Bose SoundControl) have similar intended uses.Lumen 155-SF's intended use is substantially equivalent to the predicate and reference devices. Minor wording changes in the intended use statement are considered redundant in context of a self-fitting hearing aid.
    Electro-Acoustic Characteristics (ANSI/ASA S3.22 & ANSI/CTA 2051)Latency ≤ 15 ms5.8 msPredicate: 5 ms; Reference: 5.5 msMeets criteria of being ≤ 15 ms. Within measurement uncertainty tolerance of predicate/reference.
    Frequency Response< 200 Hz - 8000 HzPredicate: 200 Hz - 8000 Hz; Reference: < 200 - 8000 HzMeets the requirements, same as the predicate and reference.
    Input Distortion (clause 4.4.2) ≤ 5%1.2%Predicate: 0.7%; Reference: ≤ 5%Meets the requirements, same as the predicate.
    Equivalent Input Noise (EIN) (Self-generated Noise Levels) (Clause 6.12)26.4 dB SPLPredicate: 28.5 dB SPL; Reference: < 23 dB SPL typical, < 27 dB SPL maxMeets the requirements, same as the predicate.
    Harmonic Distortion (Clause 6.11) ≤ 1% max< 0.5% typical, < 1% maxPredicate: 0.2%; Reference: < 0.5% typical, < 1% maxMeets the requirements, same as the predicate. Minor differences due to physical embodiment do not affect meeting electroacoustic requirements.
    Max OSPL90 (Clause 6.2) ≤ 117 dB SPL114.4 dB SPLPredicate: 109.6 dB SPL; Reference: 113 dB SPLMeets the requirements, same as the predicate. All are < 117 dB SPL limit.
    HFA OSPL90 (Clause 6.3)109.6 dB SPLPredicate: 100.9 dB SPL; Reference: 106 dB SPLMeets the requirements, same as the predicate.
    HFA FOG (Clause 6.5)51 dBPredicate: 29.4 dB SPL; Reference: 30 dBHigher than predicate/reference but device features output limiting always enabled. No different safety/efficacy questions introduced.
    Reference Test Gain (RTG) (Clause 6.7)45 dBPredicate: 24.4 dB SPL; Reference: 29 dBMeets the requirements, same as the predicate. RTG is descriptive only, differences do not introduce different safety/efficacy questions.
    BiocompatibilityConformity to ISO 10993-1, -5, -10, -12Biocompatible; requirements met. Testing for cytotoxicity, sensitization, intracutaneous reactivity passed.Not explicitly detailed for predicate/reference, but consistency with them is stated.Materials used present no biological risk, meeting ISO standards.
    Electrical Safety, EMC, Battery SafetyConformity to IEC 60118-13 (EMC), BS EN 60601-2-66 (Basic Safety & Essential Performance), IEC 60601-1-2 (EMC), IEC 60601-1-11 (Vibration & Shock), FCC & RED Tests (Wireless), AAMI TIR69 (Wireless Coexistence Risk Analysis), IEEE / ANSI C63.27 (Wireless Coexistence), ANSI C63.19 (RF Immunity)Passed all relevant tests for these standards.Similar consistency with predicate/reference is stated.All non-clinical performance testing passed, demonstrating risk mitigation to an acceptable level and reasonable assurance of safe and effective performance.
    Software Verification & ValidationSoftware V&V must be performed.Performed.Same standards as predicate.Software V&V performed per the same standards as the predicate, providing reasonable assurance of safety and effectiveness.
    Usability Testing / Human FactorsDemonstrated usability by intended user in accordance with intended use, with adequate mitigations for training and labeling.Human factors study demonstrated that usability was analyzed, verified, and validated; implemented mitigations for training and labeling are adequate.Referenced qualitative usability of predicate/reference.The study provided evidence that the device was safe and effective to operate by the intended user.
    Clinical Performance (Non-inferiority)Non-inferiority of satisfaction outcomes for self-fit vs. professional-fit. Similar levels of perceived benefit and quality of hearing. Measured sound pressure levels at the ear canal quantitatively confirmed safety and efficacy.Self-fit and Pro-fit groups yielded similar levels of satisfaction with sound quality and fitting process. SSQ-12 and APHAB scores showed similar levels of perceived benefit. Overall Maximum Output SPL was 101.3 dB SPL (safe). Measured results in-situ minus NAL-NL2 targets were similar for both groups.Reference device (Bose De Novo DEN180026) and Bose SoundControl (K211008) studies used similar assessment approaches.The clinical study demonstrated non-inferiority to professional fitting and substantial equivalence to predicate/reference based on subjective and objective performance outcomes.

    Study Details

    1. A table of acceptance criteria and the reported device performance: See table above.

    2. Sample size used for the test set and the data provenance:

      • Clinical Performance Study: 93 adults were initially enrolled, with 92 participants providing data for analysis (45 Self-Fit, 47 Pro-Fit).
        • Provenance: Prospective, multicenter, randomized, single-blind, parallel-arm study. The study was conducted adhere to COVID-19 social distancing needs. The text does not explicitly state the country of origin, but the submission is to the U.S. FDA suggesting a U.S. based or internationally recognized standard study.
      • Usability Testing (Human Factors Study): 15 adults (18 years and older) with perceived mild to moderate hearing loss.
        • Provenance: Formative and Summative Evaluations conducted in a one-on-one quiet, comfortable room. No country of origin specified.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

      • Clinical Performance Study: The "Pro-fit" arm served as a comparative ground truth, where licensed professionals (audiologists, as implied by the context of "licensed professional-fit" using "commercially available fitting software") fitted the hearing aids. The text doesn't specify the number of individual audiologists or their years of experience but indicates they used "best practice fitting algorithms" and were "licensed."
      • Usability Testing: An "independent, trained moderator" conducted each session. A "cross-functional team" reviewed outcomes, but their direct role in establishing "ground truth" (beyond analyzing usability findings) is not clarified.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Clinical Performance Study: The study design was a "multicenter, prospective, randomized, single-blind, parallel-arm study comparing outcomes of two fitting approaches: self-fitting... or licensed professional fitting." This implies a direct comparison of two defined groups rather than an adjudication process between multiple expert interpretations of a single case. There is no mention of an adjudication method in the traditional sense of resolving discrepancies in expert opinions.
      • Usability Testing: A "cross-functional team reviewed outcomes at frequent intervals and addressed changes to be considered to labeling, product packaging, and iterative minor design changes to the user interface." This suggests a collaborative review process rather than a formal adjudication method.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • A traditional MRMC study, as typically applied to image-based diagnostic AI where human readers interpret cases with and without AI assistance, was not explicitly described.
      • However, the clinical performance study does compare "Self-fit" (which uses the Sentibo mobile app for self-fitting) to "Pro-fit" (licensed professional fitting). This is a comparative effectiveness study, but it's not in the context of an "AI assistance" model for human readers. It's comparing two different fitting methodologies for a hearing aid device.
      • Effect Size: The study demonstrated non-inferiority of the self-fit approach to professional fitting.
        • Satisfaction with Sound Quality: Self-fit mean = 2.84 (SD=0.93), Pro-fit mean = 2.72 (SD=1.14) (on a 5-point scale, where higher is better).
        • Satisfaction with Fitting Process: Self-fit mean = 3.22 (SD=0.95), Pro-fit mean = 3.11 (SD=1.07) (on a 5-point scale, where higher is better).
        • SSQ-12 Benefit Score: Self-fit mean = 6.8 (SD=1.7), Pro-fit mean = 6.5 (SD=1.5) (on a 0-10 scale).
        • APHAB Global Score: Self-fit mean = 10.0 (SD=17.3), Pro-fit mean = 12.8 (SD=19.7).
        • The phrasing "similar levels of satisfaction" and "similar levels of perceived benefit" indicates that the self-fit (AI/app-driven) method performed comparably to professional fitting, demonstrating non-inferiority rather than an "improvement with AI assistance" over a human baseline.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • The "Self-fit" arm of the clinical study, which utilized the "Sentibo OTC mobile app" for fitting, represents a standalone algorithm managing the fitting process, with the user's input as part of the algorithm's execution, rather than an audiologist's direct intervention. The user makes selections based on listening to digitized speech through Apple EarPods. Therefore, the "Self-fit" group in the clinical performance study effectively assessed the standalone performance of the algorithm in guiding the user to fit the device without professional assistance.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Clinical Performance Study: The ground truth for comparative effectiveness was established by the outcomes of licensed professional fitting (Pro-fit group) using "best practice fitting algorithms" and NAL-NL2 targets, against which the self-fit outcomes were compared for non-inferiority. Subjective outcomes (satisfaction, APHAB, SSQ-12) and objective simulated real-ear measures (matching NAL-NL2 targets) were used.
      • Usability Testing: The ground truth was based on the successful completion of tasks by participants and the identification of "critical/essential tasks," reviewed by a cross-functional team.
    8. The sample size for the training set:

      • The document does not explicitly state a sample size for a training set for the Sentibo self-fitting algorithm. This type of information (how the algorithm itself was developed or trained on data) is typically for the development stage and not always detailed in an FDA 510(k) summary focused on validation. The Sentibo app offers "24 predefined options" that the user selects from, suggesting a pre-programmed or pre-trained model rather than a user-adaptive learning system that "trains" on user data during the fitting process.
    9. How the ground truth for the training set was established:

      • As the training set size is not provided, the method for establishing its ground truth is also not detailed. However, the description of the self-fitting method for the Lumen 155-SF states it "uses digitized human voice" for selection, and that its fitting method is "adequate for fitting mild to moderate hearing loss, the same as prescribed by NAL-NL2." This suggests that the underlying principles and "ground truth" for the algorithm's output (i.e., the 24 predefined options) are based on established audiological standards like NAL-NL2 or similar validated fitting algorithms for different hearing loss profiles. The predicate device uses a validated NAL-NL2 fitting algorithm, and the subject device's method is compared to this standard.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221052

    Validate with FDA (Live)

    Date Cleared
    2023-09-25

    (532 days)

    Product Code
    Regulation Number
    874.3325
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Guangdong 518000 China

    Re: K221052

    Trade/Device Name: Self-Fitting hearing aids Regulation Number: 21 CFR 874.3325
    Hearing Aid Regulation Medical Specialty: Ear Nose & Throat Product Code: QUH Regulation Number: 21 CFR 874.3325
    for use without the assistance of a hearing care professional.

    6.0 Device Information

    Per 21 CFR 874.3325
    8.0 Special controls

    The Self-Fitting Hearing Aid conforms to the special controls stated in 21 CFR 874.3325
    Self-Fitting hearing aid meets the special controls for self-fitting hearing aids outlined in 21 CFR 874.3325

    Device Description :

    Per 21 CFR 874.3325, a self-fitting wireless air conduction hearing aid is a wearable sound amplifying

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Self-Fitting hearing aids are intended to amplify sound for individuals 18 years of age or older with to moderate hearing impairment. They are adjusted by the user's hearing needs. The device is intended for use without the assistance of a hearing care professional.

    Device Description

    Per 21 CFR 874.3325, a self-fitting wireless air conduction hearing aid is a wearable sound amplifying device that is intended to compensate for impaired hearing and incorporates technology, including software, that allows users to program their hearing aids. This technology integrates user input with a self-fitting strategy and enables users to independently derive and customize their hearing aid fitting and settings. Self-fitting wireless air conduction hearing aids are Class II medical devices. The Self-Fitting hearing aids incorporate microphones for audio input and sound is delivered to the ear via a receiver that can be coupled with domes. Self-Fitting hearing aids are controlled via button controls and wirelessly via the smart hearing app (iOS and Android). The controls allow the user to adjust volume setting and customize the hearing mode program.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance for Self-Fitting Hearing Aids (K221052)

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific MetricAcceptance Criteria (Non-Inferiority Margin)Reported Device Performance (Confidence Interval / P-value)
    Co-Primary EndpointsREAG (Self-fit vs. Pro-fit) at 500Hz, 1000Hz, 2000Hz, 4000Hz (Lab)-2.5 dB (non-inferiority margin)- Right Ear: 500Hz: (-1.7976, 0.8933), P<0.0001 1000Hz: (-2.6485, 1.0217), P=0.0005 (Note: Upper bound for 1000Hz is slightly above 2.5, but text states "less than the threshold of 2.5" - this might be due to rounding or a specific interpretation of the non-inferiority test, which relies on the lower bound being above the margin for superiority. The p-value indicates non-inferiority was met.) 2000Hz: (-1.0099, 1.0003), P=<0.0001 4000Hz: (-1.6803, 2.2401), P=0.0139- Left Ear: 500Hz: (-2.4308, 0.5744), P=<0.0001 1000Hz: (-3.9191, 0.6464), P=0.0009 2000Hz: (-4.3019, -0.6981), P=<0.0001 (Negative lower bound suggests superiority of self-fit, still meets non-inferiority) 4000Hz: (-2.3433, 2.3625), P=0.0194Conclusion: Laboratory REAG of Self-fit condition was not inferior to Pro-fit condition with a threshold of 2.5 dB.
    REAG (Self-fit1 vs. Self-fit2) at 500Hz, 1000Hz, 2000Hz, 4000Hz (Lab)Similar (statistical non-significance)- Right Ear: 500Hz: (-1.0725, 0.8725), P=0.8349 1000Hz: (-0.6767, 1.9434), P=0.3310 2000Hz: (-1.9115, 1.1115), P=0.5925 4000Hz: (-1.1188, 1.8522), P=0.6175 - Left Ear: 500Hz: (-1.5968, 1.3302), P=0.8535 1000Hz: (-0.5968, 2.3302), P=0.2356 2000Hz: (-0.4443, 2.1777), P=0.1868 4000Hz: (-1.8498, 1.9831), P=0.9438 Conclusion: No statistically significant difference between the two measurements of REAG for Self-fit repeated measures in the laboratory, indicating similarity.
    APHAB global rating (Self-fit vs. Pro-fit) (Field Trial)-0.084 (non-inferiority margin)Point estimate = 0.0109 (95% CI: -0.00951, 0.0314). All values in the CI are greater than the non-inferiority margin of -0.084. P<0.0001.Conclusion: During the on-site field trial, APHAB global ratings in the Self-fit condition were not inferior to those of the Pro-fit condition with the non-inferiority margin of 0.084.
    Secondary EndpointsSNR (Self-fit vs. Pro-fit) (Lab)Not inferior (no specific margin given)Difference = 0.4450 (95% CI: -0.6703, 1.5602)
    SNR (Self-fit1 vs. Self-fit2) (Lab)Similar (statistical non-significance)Difference = 0.4000 (95% CI: -0.5020, 1.3020)
    SNR (Self-fit vs. Pro-fit) (Field Trial)Not inferior (no specific margin given)Difference = -0.1830 (95% CI: -1.0473, 0.6812)
    Safety EvaluationAbsence of device-related Adverse Events (AE) or Serious Adverse Events (SAE)No AEs or SAEs reportedAll participants were free from ear infections, pain/discomfort, dizziness, and sudden onset/rapid deterioration of tinnitus. No adverse events were reported during the clinical trial.

    2. Sample Sizes and Data Provenance:

    • Clinical Test Set Sample Size: 36 subjects were initially enrolled, but 6 withdrew, resulting in 30 subjects for analysis. The document states the planned minimum was 32 subjects.
    • Data Provenance: The clinical study was a prospective, randomized, cross-assigned, non-inferiority, multi-center clinical trial. The locations of the two on-site centers are implied to be within the country of the submitter (China, Huizhou Jinghao Medical Technology Co., Ltd. and Shenzhen Reanny Medical Devices Management Consulting Co., Ltd.), though not explicitly stated for the subjects' origin.

    3. Number and Qualifications of Experts for Ground Truth:

    • For the Pro-fit condition, which serves as a comparison to the self-fit device, the text refers to "audiological best practice hearing aid fittings." This implies that qualified audiologists or hearing care professionals established the "ground truth" for the professional fitting. The number and specific qualifications of these experts are not explicitly stated in the provided text.

    4. Adjudication Method for the Test Set:

    • The study design was a randomized, cross-assigned, non-inferiority, multi-center clinical trial. Participants underwent two stages: "laboratory" and "on-site cross wear field trial."
    • Within the laboratory stage, participants completed "Self-fit1", "Self-fit2", and "Pro-fit" procedures in random order.
    • The field trial involved a crossover design with participants receiving "Pro-fit" and "Self-fit" in random order.
    • This design inherently compares the different fitting strategies rather than relying on a separate adjudication method for individual cases or outcomes from human readers. The outcomes (REAG, APHAB, SNR) are objective measurements or standardized questionnaires, directly comparing the performance of the self-fitted device to a professionally fitted one. No specific adjudication method like 2+1 or 3+1 is mentioned, as it's not a study involving human reader interpretations that need consensus.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted in the traditional sense for image interpretation. This study focuses on the comparison between a self-fitting hearing aid and a professionally-fitted hearing aid, not on the performance of human readers with or without AI assistance for interpretation. Therefore, there is no effect size of how much human readers improve with AI vs. without AI assistance.

    6. Standalone (Algorithm Only) Performance Study:

    • Yes, a standalone performance study was conducted in the sense that the "Self-fit" condition represents the algorithm's performance without direct intervention from a hearing care professional during the fitting process by the user. The co-primary and secondary endpoints directly measure the performance of this self-fitting strategy against a professional fitting.

    7. Type of Ground Truth Used:

    • For the clinical study, the "ground truth" or reference for comparison was established by:
      • Audiological Best Practice Hearing Aid Fittings (Pro-fit): This involved a professionally conducted fitting, likely by qualified audiologists, measuring Real Ear Aided Gain (REAG) and collecting data on Abbreviated Profile of Hearing Aid Benefit (APHAB) and Speech Recognition Threshold in Noise (SNR). This serves as the "gold standard" or "expert ground truth" for optimal hearing aid performance.
      • Repeated Self-fitting Measurements (Self-fit1 vs. Self-fit2): This evaluated the consistency and repeatability of the device's self-fitting process, essentially using the device's own repeated output as a comparison for its internal consistency.

    8. Sample Size for the Training Set:

    • The document describes a clinical validation study for effectiveness and safety. It does not explicitly mention a training set sample size for the development of the self-fitting algorithm itself. The clinical study is a test set of real users validating the already developed self-fitting strategy.

    9. How Ground Truth for the Training Set Was Established:

    • Since there's no explicit mention of a training set or its sample size, there's no information provided on how ground truth for a training set was established. The self-fitting strategy is a feature of the device; its development would likely involve internal testing, algorithm design, and potentially proprietary data, but these details are not part of this 510(k) summary which focuses on the clinical validation of the final product.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223848

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-09-08

    (260 days)

    Product Code
    Regulation Number
    874.3325
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K223848

    Trade/Device Name: Intrisound™ Tuned Lumen® 155 Hearing Aids Regulation Number: 21 CFR 874.3325
    : Self-Fitting Air-Conduction Hearing Aid, Over The Counter Product Code: OUH Regulation No: 21 CFR 874.3325
    CONTROLS

    The Intrisound™ Tuned Lumen® 155 Hearing Aid conforms to the special controls stated in 21 CFR 874.3325
    br>Rule that became effective onOctober 17, 2022 |
    | Regulation | 21 CFR 874.3325
    safety and/or effectiveness concerns and meets the special controls provided in FDA regulation 21 CFR §874.3325

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intrisound™ Tuned Lumen® 155 are self-fitting, wireless air conduction hearing aids, intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. They are to meet the user's hearing needs. The device is intended for use without the assistance of a hearing care professional.

    Device Description

    The Intrisound™ Tuned Lumen® 155 Hearing Aid is a wireless, self-fitting air conduction hearing aid. It features digital signal processing (16 channel wide dynamic input compression, 3 layer fastacting output compression, 16 channel noise reduction, feedback cancellation), bi-directional microphone with windscreen, volume and program control, 12 band equalizer, self-selectable tube and ear tips, and customization through the Tuned Mobile Application. The Tuned Mobile Application aims to detect accurate auditory thresholds, partnering with a smart phone application to deliver user-customized sound through the Intrisound™ Tuned Lumen® 155 Hearing Aid. These thresholds are used to program the device using a proprietary algorithm.

    AI/ML Overview

    The provided text describes the acceptance criteria and the studies conducted for the Intrisound™ Tuned Lumen® 155 Hearing Aids.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not present a single, consolidated table specifically labeled "Acceptance Criteria and Reported Device Performance" in the typical format of a summary table for a medical device. However, the STUDY 1 section clearly outlines acceptance criteria for various endpoints and then immediately follows with the reported results for those criteria. Similarly, Table 2 and Table 3 detail performance characteristics against standards.

    Let's synthesize the information specifically from STUDY 1 and relevant tables:

    Acceptance Criteria and Reported Device Performance for Intrisound™ Tuned Lumen® 155 Hearing Aids

    Acceptance Criteria (Study 1, Section 10)Reported Device Performance (Study 1 Results)
    Mean Absolute Difference (MAD) of less than 10 dB for each tested frequency (500, 1000, 2000, 4000 Hz) between Tuned self-test and traditional audiometry.Total MAD (averaged over all tested frequencies) was 5.25 dB (95% CI: 4.73 to 5.78 dB), which is well within 10 dB. Frequencies were well correlated.
    Total MAD of less than 10 dB between Tuned self-test and traditional audiometry.Total MAD was 5.25 dB (95% CI: 4.73 to 5.78 dB), well within 10 dB.
    Tuned self-fitting is non-inferior to professional fitting by showing that the MAD from NAL-NL2 targets per frequency of the Tuned self-fitting is within the range of 5 dB of the MAD from NAL-NL2 of the professional fitting.The MAD from NAL-NL2 per frequency of the Tuned self-fitting is well within the range of 5 dB of the MAD from NAL-NL2 of the professional fitting.
    The total MAD from NAL-NL2 targets of the Tuned self-fitting is within the range of 5 dB.The total MAD from NAL-NL2 targets for Tuned self-fitting is 3.37 dB (95% CI: 3.2 to 3.53 dB), well within 5 dB.
    The total MAD of the Real Ear Aided Responses (REAR) of the Tuned self-fitting and the professional fitting is in the range of 5 dB.The total MAD between the REAR of professional-fitting and Tuned self-fitting is 3.72 dB (95% CI: 3.55 to 3.89 dB), well within 5 dB.
    The maximum power output (MPO) measurements do not exceed 117 dB SPL in any volume control level.All MPO measurements in the real ear were less than 117 dB SPL in all volume control levels.
    Additional Performance Characteristics (from Table 2 & Table 8: ANSI ASA S3.22 Performance Data)Reported Device Performance
    Max OSPL90 (for OTC hearing aids per 21CFR800.30)114.4 dB SPL (Lower than predicate's 120 dB SPL to comply with OTC requirements)
    High Frequency Average OSPL90 (HFA OSPL90)109.6 ± 2 dB SPL (Meets requirements of the standard)
    High Frequency Average Full-on Gain (HFA FOG)40 ± 2 dB (Meets requirements of the standard)
    Reference Test Gain (RTG)34 ± 4 dB (Meets requirements of the standard)
    Frequency Response200 Hz to 8000 Hz (Meets requirements of the standard)
    Frequency Range<200 Hz to 7100 Hz (Difference noted due to Receiver-in-Aid acoustic bandwidth limitations, but does not introduce new safety/effectiveness concerns, supported by performance testing).
    Harmonic Distortion (for OTC hearing aids per ANSI CTA 2051-2017 and 21CFR800.30)500 Hz ≤ 5.0%, 800 Hz ≤ 5.0%, 1600 Hz ≤ 5.0% (Complies with < 5% requirement).
    Equivalent Input Noise (EIN)<27 dB SPL (Meets requirements of the standard)
    Battery Current1.2 mA (Meets requirements of the standard)
    Latency (per ANSI CTA 2051-2017)Complies with ≤ 15 ms requirement.

    2. Sample sizes used for the test set and data provenance

    STUDY 1 (Clinical Performance Validation):

    • Sample Size: 30 participants (adults).
    • Data Provenance: Retrospective or Prospective is not explicitly stated as a separate point; however, the study design section states it was a "prospective, crossover, comparative study." The recruitment from "private hearing clinics" suggests the data is likely sourced from the country where these clinics are located, but the country of origin is not explicitly stated.

    STUDY 2 (Subjective Outcome Comparison - COSI/IOI-HA):

    • Sample Size: 41 participants (N=41 completed the study: 20 in Group 1, 21 in Group 2).
    • Data Provenance: "Prospective, two-arm, two-site, crossover randomized controlled and comparative study." Similar to Study 1, the country of origin is not explicitly stated. Given the context of FDA submission, it's typically assumed to be North America or Europe unless otherwise specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    STUDY 1:

    • Number of Experts: Not explicitly specified as a distinct number, but the study compared the Tuned self-test to "traditional audiometry performed by a professional" and "licensed audiologist." This implies at least one, if not multiple, qualified professionals.
    • Qualifications of Experts: "Licensed audiologist" and "professional." No specific experience (e.g., "10 years of experience") is mentioned.

    STUDY 2:

    • Number of Experts: Not explicitly stated as a distinct number. The study compared self-fitting to "licensed professional fitting." This implies the involvement of one or more licensed professionals to perform the "professional fitting."
    • Qualifications of Experts: "Licensed professional." No specific experience mentioned.

    4. Adjudication method for the test set

    The provided text does not mention any adjudication method for the test sets in either Study 1 or Study 2. The comparisons are direct between the device's output/self-fitting outcome and the professional's assessment/fitting outcome.

    5. If a Multi Reader Multi Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not explicitly described in the provided text. The studies focus on comparing the self-fitting hearing aid's performance/user outcomes against traditional audiology methods, rather than assessing how human readers (e.g., radiologists, ophthalmologists) improve with AI assistance. The device is a self-fitting hearing aid, not an AI-powered diagnostic imaging tool that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, in essence, standalone performance was assessed for the self-fitting capability.

    • Study 1 directly compared the "Tuned self-test" (algorithm-driven self-assessment) against "traditional audiometry performed by a professional" for auditory thresholds. This can be considered a form of standalone evaluation for the audiometric threshold detection aspect of the device's algorithm.
    • The "Intrisound™ Tuned Lumen® 155 Hearing Aid is a wireless, self-fitting air conduction hearing aid... customization through the Tuned Mobile Application. The Tuned Mobile Application aims to detect accurate auditory thresholds, partnering with a smart phone application to deliver user-customized sound through the Intrisound™ Tuned Lumen® 155 Hearing Aid. These thresholds are used to program the device using a proprietary algorithm." This describes the core standalone functionality of the device's self-fitting algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Study 1: The ground truth for audiometric thresholds was established by traditional audiometry performed by a licensed professional (audiologist). For fitting parameters (NAL-NL2 targets, REAR), the ground truth was based on established audiology standards and professional fitting methods.
    • Study 2: The ground truth for subjective outcomes was established using validated patient-reported outcome measures (PROMs) - COSI and IOI-HA questionnaires completed by the participants themselves, after using both self-fitting and professional fitting methods. This is a form of user-reported outcomes data. The "professional fitting" condition serves as a comparative reference for the self-fitting's effectiveness.

    8. The sample size for the training set

    The document does not specify the sample size for the training set used to develop the proprietary algorithm or the Tuned self-test. The studies described are performance validation/testing studies, not algorithm development studies.

    9. How the ground truth for the training set was established

    Since the training set size or methodology is not detailed, the document does not provide information on how the ground truth for the training set was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230538

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-06-16

    (109 days)

    Product Code
    Regulation Number
    874.3325
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    All-Day Clear ADC1 Hearing Aid Regulation Number: 21 CFR 874.3325 Regulation Name: Self-fitting air-conduction
    Self-fitting Air-Conduction Hearing Aid |
    | REGULATION NUMBER: | 21 C.F.R. § 874.3325

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The All-Day Clear Slim hearing aids are intended for individuals 18 years or older with perceived mild to moderate hearing loss. The intended use of the All-Day Clear Slim hearing aids is to amplify and transmit sound to the ears and hereby compensate for hearing loss. The hearing aids must only be used by the intended person. They should not be used by any other person as they could damage hearing. The charging case is intended to charge the battery of the rechargeable hearing aids. The All-Day Clear App is intended for individuals 18 years or older, to set-up and/or adjust All-Day Clear Slim hearing aids through Android and Apple iOS smartphones.

    The All-Day Clear hearing aids are intended for individuals 18 years or older with perceived mild to moderate hearing loss. The intended use of the All-Day Clear hearing aids is to amplify and transmit sound to the ears and hereby compensate for hearing loss. The hearing aids must only be used by the intended person. They should not be used by any other person as they could damage hearing. The charging case is intended to charge the battery of the rechargeable hearing aids. The All-Day Clear App is intended for individuals 18 years or older, to set-up and/or adjust All-Day Clear hearing aids through Android and Apple iOS smartphones.

    Device Description

    The All-Day Clear Slim ADCS1 hearing aid and All-Day Clear ADC1 hearing aid are self-fitting air-conduction hearing aids. The ADCS1 and ADC1 hearing aids are nearly identical except for minor design differences. The hearing aid system consists of hearing aids for left and right ears, charging case, charging cable with USB-A and USB-C connector, wall plug, a set of eartips (small, medium, and large sizes), cleaning brush, carrying case (for ADCS1 hearing aids alone), and wax guards. The ADCS1 and ADC1 hearing aids are indicated for over-the-counter use.

    The end user can self-fit the hearing aid by downloading the All-Day Clear App to their Android or Apple iOS device, pairing the hearing aids to the App, and following a series of guided onboarding steps:

    • Instructions on how to wear the hearing aids
    • Creation of a sound profile by listening to sample audio in which the user selects parameters for volume and sound clarity
    • Selection of sound balance

    Once the hearing aids have been set up with a sound profile, the App can continue to be used as a remote control of the hearing aids. For example, the end users can use the App to change the volume or mute the hearing aid microphones (together or independently), switch between different hearing aid sound modes, adjust the volume of a streamed signal, adjust equalizer and wind noise control, and read out the status of the hearing aid charge level.

    The ADCS1 and ADC1 hearing aid system provides an alternative optional method of fitting the device by letting the end users purchase "In-Clinic Care Package", where they can have a licensed hearing care professional fit their hearing aids using professional fitting software called "All-Day Clear Fitting Software." All-Day Clear Fitting Software is a component of Sonova's TrueFit 5.3 or higher fitting software which can only be used by a licensed hearing care professional. In both scenarios, the end users can use the App as a remote control of the hearing aid.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Endpoint)Non-inferiority MarginReported Device Performance (Self-Fit vs. Pro-Fit)Conclusion on Acceptance Criteria Met?
    Primary Effectiveness Endpoint:
    APHAB Global Benefit Score (Self-Fit)Not more than 12.5% below Pro-FitAdjusted mean treatment difference: 0.6% (95% CI: -2.919% to 4.742%) (after outlier exclusion)Met (Lower bound > -12.5%)
    Secondary Effectiveness Endpoint:
    QuickSiN Benefit Score (Self-Fit)-1.8 dB or higherAdjusted mean treatment difference: 0.103 dB SNR (95% CI: -0.867 to 1.074 dB SNR) (Period 1 data)Met (Lower bound > -1.8 dB)
    Safety Endpoint:(Implicit: No serious adverse events or unacceptable adverse events)One non-serious treatment-emergent adverse event (ear canal occlusion, not device-related). No serious adverse events.Met

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set:
      • Clinical Study: 44 participants (enrolled). Due to early discontinuations and outlier exclusion, analysis for effectiveness endpoints was based on varying numbers:
        • APHAB: 40 participants (after excluding 2 outliers from the original 42 for Period 1 analysis).
        • QuickSiN: 40 participants (Period 1 data analysis).
      • Human Factors/Usability Study: 30 participants (15 patients, 15 caregivers).
    • Data Provenance: Retrospective from clinical study performed with "HelloGO" hearing aid, which is stated to have identical hardware and almost identical firmware to the subject devices (ADCS1, ADC1). The human factors study was prospective, conducted specifically for the ADCS1/ADC1. The country of origin for the clinical study is not explicitly stated in the provided text.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • The clinical study did not involve AI assistance, therefore "ground truth" was established through direct measurement of human hearing abilities and perceptions using standardized audiological tests (APHAB and QuickSiN) with both self-fitted and professionally fitted devices.
    • Professionally Fit Devices: Implies fitting by audiologists or hearing care professionals, but the specific number or qualifications of these professionals for establishing the "ground truth" (the professional fit as a comparator) are not detailed beyond "professional fitting method."

    4. Adjudication Method for the Test Set

    • Clinical Study (Effectiveness Endpoints): Not applicable in the sense of expert adjudication of a device's output. The study design involved direct comparison of patient-reported outcomes (APHAB) and objective speech-in-noise testing (QuickSiN) under two different fitting conditions (self-fit vs. professional fit). The "ground truth" for the benefit scores was derived directly from participant responses and test results.
    • Human Factors/Usability Study: "Study staff systematically assessed device usage to identify problems that were encountered by participants, including use errors and close calls during participant performance." This suggests observation and recording of events, rather than a formal expert adjudication of discrete findings.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This study focuses on the effectiveness of a self-fitting hearing aid compared to a professionally fitted one, not on AI assistance for human readers/clinicians.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, in a sense, the "self-fitting" method can be considered a standalone performance of the algorithm (as implemented in the app) plus the user's interaction, without a hearing care professional in the loop for the fitting process itself. The study compared this "self-fitted" performance to a "professionally fitted" performance.

    7. The Type of Ground Truth Used

    • Clinical Study (Effectiveness):
      • Patient-Reported Outcomes: Abbreviated Profile of Hearing Aid Benefit (APHAB) global benefit score.
      • Objective Audiological Measure: Quick Speech-in-Noise (QuickSiN) test benefit score.
      • The "ground truth" for comparison was the outcome achieved with a professionally derived fitting.
    • Safety: Tabulation of adverse events and serious adverse events.
    • Human Factors/Usability: Observed use errors and close calls in a simulated environment, and participant feedback.

    8. The Sample Size for the Training Set

    • The document does not explicitly mention a "training set" for the clinical study parameters. The study design is a comparative effectiveness study between two fitting methods (self-fit and professional fit) for the hearing aid itself. The "self-fitting algorithm" embedded in the app would have had its own development and possible internal training process, but details of this are not provided in the context of this FDA summary.
    • The human factors study did not involve a training set; it was a usability validation.

    9. How the Ground Truth for the Training Set Was Established

    • As a training set for the specific self-fitting algorithm is not detailed, the method for establishing its "ground truth" (e.g., what data it was trained on to achieve optimal fitting) is not provided in this document. The clinical study established ground truth for effectiveness comparison by using a professionally fitted condition as a benchmark.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223137

    Validate with FDA (Live)

    Date Cleared
    2023-03-14

    (161 days)

    Product Code
    Regulation Number
    874.3325
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Name: Lexie Lumen Self-Fitting OTC Hearing Aid(s) with Lexie Application Regulation Number: 21 CFR 874.3325
    Name: Lexie Lumen Self-Fitting OTC Hearing Aid(s) with Lexie Application Regulation Number: 21 CFR 874.3325
    Hearing Aid Classification Name: Self-Fitting, Air-Conduction Hearing Aid, Over the Counter (21 CFR 874.3325
    DEVICE DESCRIPTION

    Per 21 CFR 874.3325 a self-fitting wireless air conduction hearing aid is a wearable
    Summary - Conclusion

    The special controls for self-fitting hearing aids as per 21 CFR 874.3325, are

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lexie Lumen Self-Fitting OTC Hearing Aid(s) with Lexie Application is a self-fitting air-conduction hearing aid, intended to amplify sound for individuals 18 years of age or older with perceived mild to moderate hearing impairment. The device is adjusted by the user to meet the user's hearing needs. The device is intended for use without the assistance of a hearing care professional.

    Device Description

    The Lexie Lumen Self-Fitting OTC Hearing Aid(s) with Lexie Application is a self-fitting, air conduction hearing aid consisting of the IntriCon hardware, Lexie Software, Lexie Application and accessories supplied in the carton. The Lexie App is available on Android and iOS. The Lumen self-fitting OTC hearing aid is the only model applicable to this 510(k). The Lexie Lumen Self-Fitting OTC Hearing Aid(s) with Lexie Application is a behind the ear device that includes self-adjustable coupling by means of a slim tube and ear tip/dome. The hearing aids can be fine-tuned remotely by trained hearing experts, at the request of the user, in the Lexie contact center. The Lexie app receives custom remote settings as performed through the Lexie adjustment wizard and fitting portal (internet service).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Lexie Lumen Self-Fitting OTC Hearing Aid(s) with Lexie Application, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    CategoryAcceptance Criteria (from study)Reported Device Performance (Lexie SF Group vs. AF Group)Conclusion
    Clinical Performance (Effectiveness)
    Primary Endpoint: APHAB Global Benefit(Implicit: Outcomes equivalent to AF group)2-weeks: SF group had less difficulty in background noise (p < .05) than AF group. 6-weeks: No significant differences between SF and AF groups for any APHAB scores (unaided, aided, benefit). Scores fell within the 16.3 acceptance criteria for equivalence.Met
    Secondary Endpoint: IOI-HA Outcomes1-point margin for subscales (ordinal scales)2-weeks: SF group reported significantly longer duration of hearing aid use per day (p < .05). Significant difference in total score (p < .05). All subscales within 1-point margin. Total score within 5-point margin.Met
    5-point margin for total score6-weeks: No significant differences between groups for any subscales or total score (p > .05). All subscales within 1-point margin. Total score within 5-point margin.Met
    Secondary Endpoint: Speech-in-Noise (DIN)1.8 dB SNR acceptance criteria2-weeks: DIN aided scores poorer for AF group than SF (p < .05), but within 1.8 dB SNR criteria. Benefit scores not significantly different.Met
    (Implicit: Outcomes equivalent to AF group)6-weeks: No significant differences for aided QuickSIN or DIN tests (p > .05) or calculated benefit scores.Met
    SafetyNo Serious Adverse Events (SAEs)One participant withdrew due to a middle ear infection (unrelated to device). No SAEs occurred.Met
    Human Factors TestingSuccess rate > 79% on all use-related critical tasks> 79% success rate on all use-related critical tasks. High user satisfaction (65.5% found it "easy to use" and "user-friendly"). One minor issue with slim tube size/domes not impacting safety.Met
    Non-clinical Performance Testing(Each standard has its specific pass/fail criteria)All tested standards (IEC 60601-1, IEC 60601-1-11, IEC 60601-2-66, IEC 60601-1-2, IEC 60118-13, ANSI/ASA S3.22, ANSI CTA 2051, ANSI ASA S3.6, IEC 62304, ISO 10993 series, IEC 60601-1-6, FDA Cybersecurity Guidance, IEEE / ANSI C63.27, ANSI C63.19)Pass
    ANSI ASA S3.22 (Acoustic Performance)Max OSPL90: <120dB SPLMeasured 114.4dB SPLMet
    HFA-OSPL90: 111 ± 2dB SPL110dB SPLMet
    HFA-FOG: 40 ± 2dB45dBMet
    Reference Test Gain (RTG): 34 ± 4dB34dBMet
    Frequency Response: 200Hz to 8000Hz (lower cutoff to 200Hz, upper to 5kHz+)200Hz to 7000HzMet
    Harmonic Distortion: ≤ 1.5% @ 500Hz, ≤ 2.0% @ 800Hz, ≤ 3.0% @ 1600Hz (overall <5%)THD@ 500Hz: 0.4%, THD@ 800Hz: 0.2%, THD@ 1600Hz: 0.4% (all <5%)Met
    EIN: <32 dB SPLActual measured value 26.4dB SPL (<32dB SPL)Met
    Latency: <15ms<15msMet

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Study:
      • Sample Size: 64 participants (32 in the Audiologist Fit (AF) group, 32 in the Self-Fit (SF) group).
      • Data Provenance: Prospective, Randomized Controlled Trial conducted at the University of Pretoria, Gauteng, South Africa.
    • In-situ Hearing Test Accuracy Validation (separate study):
      • Sample Size: 45 participants (90 ears).
    • Human Factors Testing:
      • Sample Size: 29 adult participants.
      • Data Provenance: Simulated-use session, likely within a controlled environment (e.g., laboratory).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Clinical Study:
      • The "ground truth" for the Audiologist Fit (AF) group was established by audiologists who performed standard audiometric evaluations and fitted the hearing aids to match NAL-NL2 prescriptive targets. The text doesn't specify the exact number of audiologists, but refers to "the audiologist" (singular) in descriptions of procedures, implying at least one qualified audiologist was involved in the fitting and orientation for the AF group. The context of a university setting (University of Pretoria) suggests these would be qualified professionals.
      • For the in-situ hearing test validation, "the audiologist-performed audiometry" was used as a reference, again implying qualified audiologist(s).
    • Human Factors Testing:
      • No specific experts are mentioned as establishing "ground truth" in the same way as the clinical study. Instead, the "effectiveness and accuracy of the tasks" were assessed by observing participants' performance during simulated use, implying assessment by human factors specialists or researchers.

    4. Adjudication Method for the Test Set

    • The clinical study was a randomized controlled trial (RCT) comparing two groups (SF vs. AF).
    • No explicit "adjudication method" (like 2+1 or 3+1 for expert review) is mentioned for the outcomes of the clinical study, as the primary and secondary endpoints (APHAB, IOI-HA, QuickSIN, DIN) are quantitative measures.
    • For the in-situ hearing test accuracy, performance (Mean Absolute Difference) was compared against "audiologist-performed audiometry," where the audiologist's results served as the reference.
    • For Human Factors Testing, an "observer" evaluated task success, and a "semi-structured interview" was used for user perceptions.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • Yes, a comparative effectiveness study was done, comparing the Lexie Lumen in a self-fit (SF) condition against an audiologist-fit (AF) condition. This is a direct comparison of the device's performance with and without expert assistance in fitting.
    • Effect Size of Human Readers Improve with AI vs. without AI assistance:
      • The study compares self-fitting (device with its AI/software-driven fitting strategy) to audiologist-fitting (human expert). It doesn't present an effect size of "human readers improve with AI vs. without AI assistance" in the context of interpretation of medical images.
      • Instead, it evaluates if the self-fitting strategy (which is AI/software-driven) is non-inferior to an audiologist's fitting. The results show that the SF group was equivalent or even slightly better in some aspects (background noise difficulty at 2 weeks, duration of use at 2 weeks) compared to the AF group, and overall achieved equivalence within the defined acceptance criteria for the primary and secondary endpoints. This implies that the device's self-fitting capability largely negates the need for a human professional's direct fitting intervention for achieving comparable outcomes for mild to moderate hearing loss.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, the "Self-Fit (SF) group" in the clinical study represents the standalone performance of the device's self-fitting algorithm without human-in-the-loop fitting assistance. Participants in this group were explicitly stated to "set up and manage the devices using the Lexie app, entirely without professional support." Any assistance sought in Phase II was through the Lexie online hearing experts, but the initial fitting and core performance were self-directed via the app's algorithm.

    7. The Type of Ground Truth Used

    • Clinical Study (Effectiveness):
      • The ground truth was established by comparison with best-practice audiological fitting (NAL-NL2 targets) performed by audiologists. The aim was to demonstrate non-inferiority or equivalence against this established clinical standard.
      • For the in-situ hearing test, the "Reference" was audiologist-performed audiometry.
    • Safety:
      • "Ground truth" for safety was the occurrence and nature of Adverse Events (AE) and Serious Adverse Events (SAE), which are reportable objective events.
    • Human Factors Testing:
      • "Ground truth" was the observational assessment of task completion and accuracy by study personnel, corroborated by user self-reports and satisfaction.
    • Non-clinical Performance Testing:
      • "Ground truth" was defined by the acceptance criteria within the cited national and international consensus standards (e.g., IEC, ISO, ANSI, CTA), against which the device's physical and electrical properties were measured.

    8. The Sample Size for the Training Set

    • The document does not provide information on the sample size for the training set used to develop the Lexie self-fitting algorithm or its core software. This information is typically proprietary and not usually disclosed in a 510(k) summary, which focuses on validation rather than development.

    9. How the Ground Truth for the Training Set Was Established

    • The document does not provide information on how the ground truth for the training set was established. The text only states that the fitting algorithm (Lexie Comfort) is "based on National Acoustics Laboratories' Non-Linear Version 2 (NAL-NL2), with additional adjustments aimed for a greater listening comfort." This implies that the NAL-NL2 prescriptive targets likely form a significant part of the foundational "ground truth" or reference for the algorithm's development, but details of specific training data and its ground truth are absent.