Search Filters

Search Results

Found 223 results

510(k) Data Aggregation

    K Number
    K242817

    Validate with FDA (Live)

    Device Name
    Jason membrane
    Date Cleared
    2025-12-12

    (450 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    15806
    GERMANY

    Re: K242817
    Trade/Device Name: Jason membrane
    Regulation Number: 21 CFR 872.3930
    December 12, 2025

    Re: K242817
    Trade/Device Name: Jason membrane
    Regulation Number: 21 CFR 872.3930
    Classification Panel:** Dental Devices
    Classification Name: NPL (Class 2) – Bone grafting material (21 CFR 872.3930

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Jason membrane alone or in combination with suitable augmentation materials (like autogenous bone, allogeneic, xenogeneic or alloplastic bone replacement materials) is indicated for immediate or delayed guided tissue and bone regeneration.

    • in case of surgical bone defects and bone wall defects
    • in the context of sinus floor augmentation and for support of the Schneiderian membrane
    • in the context of maxillary ridge augmentation
    • in the context of maxillary ridge reconstruction for prosthetic treatment
    • in the context of a treatment of fenestration defects
    • in case of periodontal bone defects (one to three-wall defects, class I and II furcation defects)
    • in case of dehiscence defects
    • after apicoectomy, cystectomy, resection of retained teeth and resection of other bone lesions
    • in extraction sockets after tooth extractions
    • in case of immediate or delayed augmentation around implants in extraction sockets
    Device Description

    Jason® membrane is a completely resorbable collagen membrane used in maxillofacial surgery, implantology, periodontology, oral surgery and endodontology as a barrier membrane to support guided tissue regeneration (GTR) and guided bone regeneration (GBR), for covering implants and for periodontal tissue regeneration. Jason® membrane is produced from porcine pericardium in a standardized, controlled purification process. When dry Jason® membrane is a white collagen matrix with a very dense fiber structure. It possesses sufficient rigidity and stability for a broad variety of intended uses. Jason® membrane possesses a physiological neutral pH-value. Jason® membrane typically resorbs within 12 weeks after implantation. The Jason® membrane is intended to be used in adults only.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250512

    Validate with FDA (Live)

    Date Cleared
    2025-12-05

    (287 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    CHINA

    Re: K250512
    Trade/Device Name: Augmented Gingival Matrix
    Regulation Number: 21 CFR 872.3930
    Common or Usual Name: Gelatin Matrix
    Classification Name: Bone Grafting Material (21 CFR 872.3930
    | Augmented Gingival Matrix | Geistlich Mucograft® | No discussion required |
    | Regulation | 21 CFR 872.3930
    | 21 CFR 872.3930 | Same as predicate |
    | Product Code | NPL | NPL | Same as predicate |
    | Device Classification

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    It is indicated for localized gingival augmentation to increase keratinized tissue (KT) around teeth.

    Device Description

    AGM is a three-layer, symmetrical composite matrix intended for oral tissue regeneration. The top and bottom of the product are gelatin layers, which are made of gelatin derived from bovine bone. The gelatin layers are structured as random fiber stacking by in situ cross-linked electrospinning technology and its average diameter is 100 - 900 nm. Users do not need to distinguish between the top and bottom of the product. The middle of the product is polycaprolactone layer (Abbreviated as "PCL layer"), which is thin polycaprolactone film with porous structure.

    When adhering to the wound surface, the "inner" gelatin layer, the one facing the wound can increase keratinized tissue (KT) around teeth. The middle of PCL layer promotes the spreading of matrix on wound surface. Along with the "outer" gelatin layer, the PCL layer also acts as physical barrier to prevent foreign objects' invasion.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252253

    Validate with FDA (Live)

    Date Cleared
    2025-11-25

    (127 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Geistlich Mucograft® /Geistlich Mucograft® Seal; Geistlich Fibro-Gide®
    Regulation Number: 21 CFR 872.3930
    Geistlich Mucograft® /Geistlich Mucograft® Seal; Geistlich Fibro-Gide®
    Regulation Number: 21 CFR 872.3930
    Matrix |
    | Classification Name: | Barrier, Animal Source, Intraoral |
    | Regulation Number: | 872.3930
    Material |
    | Classification Name: | Barrier, Animal Source, Intraoral |
    | Regulation Number: | 872.3930

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Geistlich Mucograft® and Geistlich Mucograft® Seal are indicated for:

    • covering of implants placed in immediate or delayed extraction sockets;
    • localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants;
    • alveolar ridge reconstruction for prosthetic treatment; and
    • recession defects for root coverage.

    Geistlich Fibro-Gide® is intended for soft tissue augmentation. It is indicated for:

    • Localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants
    • Alveolar ridge reconstruction for prosthetic treatment
    • Recession defects for root coverage
    Device Description

    Geistlich Mucograft® and Geistlich Mucograft® Seal are surgically implanted, fully resorbable devices intended for oral tissue regeneration. The matrices are made of collagen without further cross-linking. All configurations of the product are sterilized in a double package by gamma irradiation. Geistlich Mucograft® and Geistlich Mucograft® Seal are composed of two structures: one smooth structure and one porous structure. The device allows tissue adherence as a prerequisite for favorable wound healing. The "outer" side (i.e., turned towards the soft tissue) with a smooth surface consists of compact collagen and has a smooth texture with the appropriate elastic properties to accommodate suturing. The "inner" porous structure consists of collagen fibers in a loose, porous arrangement to allow cell invasion for soft tissue ingrowth. This roughened surface is placed next to the host tissue to facilitate tissue integration.

    The products are provided as follows:

    • Geistlich Mucograft®: 15 x 20 mm, 20 x 30 mm, and 30 x 40 mm
    • Geistlich Mucograft® Seal: 8 mm and 12 mm diameter

    Geistlich Fibro-Gide® is a fully resorbable, porous, collagen matrix of porcine origin of a spongious consistency. The collagen is extracted from veterinary certified pigs and is carefully purified to avoid antigenic reactions. The collagen scaffold is weakly crosslinked. Geistlich Fibro-Gide® is sterilized in double packaging by Gamma-irradiation.

    Geistlich Fibro-Gide® is an implantable device intended for use in soft tissue augmentation procedures. As described in more detail below, the device is indicated specifically for insufficient tissue volume at the alveolar ridge and for soft tissue recession. It has mechanical properties appropriate to withstand the mechanical stresses that occur after wound closure in soft tissue augmentation procedures, i.e., it has good volume stability and it withstands early resorption to allow the formation of new soft tissue and degrades over time. In addition, the matrix is designed with an appropriate thickness to provide sufficient space for the ingrowth of new soft tissue. Due to its good wettability, suturability and biological properties, the device becomes well integrated into the surrounding soft tissue.

    The products are provided as follows:

    • 15 x 20 x 6 mm
    • 20 x 40 x 6 mm
    • 15 x 20 x 3 mm
    • 20 x 40 x 3 mm
    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K244006

    Validate with FDA (Live)

    Device Name
    FG Bone Graft M
    Date Cleared
    2025-09-17

    (265 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    , 241407
    TAIWAN

    Re: K244006
    Trade/Device Name: FG Bone Graft M
    Regulation Number: 21 CFR 872.3930
    , 241407
    TAIWAN

    Re: K244006
    Trade/Device Name: FG Bone Graft M
    Regulation Number: 21 CFR 872.3930
    Grafting Material, Synthetic
    Common or Usual Name: Bone grafting material
    Regulation Number: 872.3930
    Name:** MBCP (K051885)
    Common or Usual Name: Bone grafting material
    Regulation Number: 872.3930

    • | MBCP | FG Bone Graft M |
      | K. number | K051885 | K244006 |
      | Regulation Number | 21 CFR 872.3930
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FG Bone Graft M is intended for use as a bone grafting material to fill, augment or reconstruct periodontal or oral/maxillofacial defects.

    These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.

    Typical uses include:

    • Periodontal/Intrabony defects filling
    • Ridge augmentation
    • Implant preparation, placement in extraction sites
    • Sinus lifts
    • Cystic cavities filling
    Device Description

    The Synthetic Bone Substitute, FG Bone Graft M, is a microporous and macroporous biphasic calcium phosphate ceramic consisting of 60% Hydroxyapatite (HA) and 40% beta-Tricalcium Phosphate (β-TCP).

    The synthetic bone substitute is presented in a porous form required for the conduction of precursor cells to the defect site to facilitate new bone growth.

    This product is provided sterile for single patient use.

    The synthetic bone substitute is progressively replaced by new bone according to the remodeling process.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K243745

    Validate with FDA (Live)

    Device Name
    FG Bone Graft B
    Date Cleared
    2025-08-29

    (267 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    , 241407
    TAIWAN

    Re: K243745
    Trade/Device Name: FG Bone Graft B
    Regulation Number: 21 CFR 872.3930
    , 241407
    TAIWAN

    Re: K243745
    Trade/Device Name: FG Bone Graft B
    Regulation Number: 21 CFR 872.3930
    M DENTAL (K051443)
    Common or Usual Name: Bone Void Filler, Synthetic
    Regulation Number: 872.3930
    | 21 CFR 872.3930 | Substantially equivalent |
    | Product code | LYC | LYC | Substantially equivalent
    | 21 CFR 872.3930 | Substantially equivalent |
    | Product code | LYC | LYC | Substantially equivalent

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FG Bone Graft B is recommended for:

    • Augmentation or reconstructive treatment of the alveolar ridge.
    • Filling of infrabony periodontal defects.
    • Filling of defects after root resection, apicoectomy, and cystectomy.
    • Filling of extraction sockets to enhance preservation of the alveolar ridge.
    • Elevation of the maxillary sinus floor.
    • Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR).
    • Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR).
    Device Description

    FG Bone Graft B is a sterile, synthetic, multi-porous biocompatible ceramic matrix in granular form for filling bone defects. The material with microporous structure supports rapid ossification with local bone. With its phase purity of >= 99%, the ceramic material complies with US standard specification ASTM F 1088-04. The validated manufacturing process guarantees batch conformity and reproducibility.

    AI/ML Overview

    The FDA 510(k) clearance letter for FG Bone Graft B indicates that the device is substantially equivalent to a predicate device (CERASORB M DENTAL). The clearance letter references non-clinical tests performed to demonstrate this equivalence, focusing on chemical composition, physical properties, and performance in vivo.

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    DescriptionAcceptance CriteriaReported Device Performance and Results
    Chemical Composition
    Complete chemical composition, summing to 100% by mass, including all additives and the Chemical Abstracts Service (CAS®) registry number of all components.Consisting of ≥ 99% beta-Tricalcium Phosphate (ß-TCP)100%
    Description of the composition, including an elemental analysis, identifying the trace impurities.Conc.(ppm) Pb ≤30, As ≤3, Cd ≤5, Hg ≤5Conc.(ppm) Pb 0, As 0.33, Cd 0.09, Hg 0
    Physical Properties
    SEM micrographs, showing particle size, shape, and porosity.The product behaves like a porous structure and is similar to the reference product.The SEM result showed the surface characteristic of the TCP sample (FG Bone Graft B) is similar in structure to the predicate device (Cerasorb) via 600X, 1000X, and 3000x SEM photos.
    A plot of the resorption of your device versus time showing the time for total clearance or integration under a representative model.Similar trend changes to the comparison products.~90% degraded by 12 weeks
    Healing time, i.e., the earliest time at which implant loading may be successfully attempted.N/A (Not explicitly defined as a numerical criterion, but evaluated in vivo).The defect fill rate was observed to be 21.5% at 4 weeks, increasing to 26.2% by 8 weeks, and reaching up to 33.9% by 12 weeks. (This implies a healing progression, though not a specific "loading time" metric).
    Phase purity, i.e., the relative mass percentages of crystalline and amorphous phases (%).Similar trend changes to the comparison products.100% β-TCP
    Calcium to phosphorus ratio (Ca/P).Ca/P ratio >1.5Ca/P ratio: 1.89 - 1.95
    Volumetric porosity (% void space).The porosity is approximately 70% ± 5% or similar to the reference product.Volumetric porosity: 68.3%
    Particle size distribution plot (μ).The mean value of the particle size distribution is within the declared specifications, or the median and mode are within the specification range.500-1000μm
    pH.Similar trend changes to the comparison products.~7.9 over 7 days
    Performance In Vivo
    New bone formation.New bone formation performance comparable to the predicate.New bone formation increased over time at comparable rates to the predicate.
    Material degradation (residual material).Material degradation rates comparable to the predicate.FG Bone Graft B degraded at comparable rates to the predicate over 12 weeks.
    Inflammatory response.Minimal to mild inflammatory response, no significant adverse reactions.Minimal to mild inflammatory response, with no significant adverse reactions.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample size for the test set: The document states that the in vivo study used a "Beagle dog" model, and the animals were "divided into groups: test group (FG Bone Graft B), positive control group (Cerasorb, a commercial β-TCP), and a negative control (empty defect)." However, the exact number of animals in each group or total animal count is not specified in the provided text.
    • Data provenance: The study was a prospective in vivo animal study performed on Beagle dogs. The location/country of origin of the study is not explicitly stated in the provided text, but the submitter "Full Golden Biotech Co., Ltd." is located in Taiwan.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of experts: Not explicitly stated. The histological and radiographic analyses were likely performed by trained professionals (e.g., veterinary pathologists, radiologists), but the number of reviewers or their specific qualifications are not detailed in the provided text.
    • Qualifications of experts: Not specified beyond the implied expertise in conducting and analyzing in vivo studies (e.g., histology, micro-CT).

    4. Adjudication Method for the Test Set

    • Adjudication method: Not explicitly stated. For animal studies, consistency and blinding are typically employed, but a formal "adjudication method" in the sense of multiple human readers for consensus is not described for this non-AI bone graft device. The results are presented as quantitative measurements and observations (e.g., "new bone formation increased," "minimal to mild inflammatory response").

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • MRMC study: No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for AI/image analysis devices where the AI's impact on human reader performance is being assessed. This document describes a traditional preclinical performance study for a bone graft material.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Standalone performance: N/A. This is a bone graft material, not an algorithm or AI device. The "performance" refers to the biological and physical properties of the material itself, not the output of a software algorithm.

    7. The Type of Ground Truth Used

    • Type of ground truth: The ground truth for the in vivo study (which is the primary performance study) was established through direct anatomical, histological, and radiographic assessments of the bone defects in the animal model.
      • Histological analysis: Quantified new bone formation, material degradation, and inflammatory response. This involves microscopic examination of stained tissue sections, which is considered a gold standard for assessing tissue regeneration and integration.
      • Radiographic analysis: Used micro-CT to assess bone density and bone volume, providing quantitative structural data.
      • Comparison to predicate: The "ground truth" for showing substantial equivalence was the performance of the established predicate device (Cerasorb) under the same study conditions.

    8. The Sample Size for the Training Set

    • Sample size for training set: N/A. This device is a bone graft material, not an AI or machine learning algorithm. Therefore, there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    • Ground truth for training set: N/A. As there is no AI component, there is no training set and no ground truth establishment for such a set.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    Perio; Geistlich Combi-Kit Collagen®; Geistlich Perio-System Combi Pack
    Regulation Number: 21 CFR 872.3930
    Perio; Geistlich Combi-Kit Collagen®; Geistlich Perio-System Combi Pack
    Regulation Number: 21 CFR 872.3930
    MembraneCollagen Matrix |
    | Classification Name: | Bone Grafting Material |
    | Regulation Number: | 872.3930
    MembraneCollagen Matrix |
    | Classification Name: | Bone Grafting Material |
    | Regulation Number: | 872.3930
    MembraneCollagen Matrix |
    | Classification Name: | Bone Grafting Material |
    | Regulation Number: | 872.3930

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Geistlich Bio-Gide® is intended for the following uses:

    • augmentation around implants placed in immediate extraction sockets;
    • augmentation around implants placed in delayed extraction sockets;
    • localized ridge augmentation for later implantation;
    • alveolar ridge reconstruction for prosthetic treatment;
    • filling of bone defects after root resection, cystectomy, removal of retained teeth;
    • guided bone regeneration in dehiscence defects; and
    • guided tissue regeneration procedures in periodontal defects.

    Geistlich Bio-Gide® Shape is indicated for:

    • augmentation around implants placed in immediate extraction sockets;
    • augmentation around implants placed in delayed extraction sockets;
    • localized ridge augmentation for later implantation;
    • alveolar ridge reconstruction for prosthetic treatment;
    • filling of bone defects after root resection, cystectomy, removal of retained teeth; and
    • guided bone regeneration in dehiscence defects.

    Geistlich Bio-Gide® Compressed is indicated for:

    • augmentation around implants placed in immediate extraction sockets;
    • augmentation around implants placed in delayed extraction sockets;
    • localized ridge augmentation for later implantation;
    • alveolar ridge reconstruction for prosthetic treatment;
    • filling of bone defects after root resection, cystectomy, removal of retained teeth;
    • guided bone regeneration in dehiscence defects, and
    • guided tissue regeneration procedures in periodontal defects.

    Geistlich Bio-Gide® Forte is indicated for:

    • augmentation around implants placed in immediate extraction sockets;
    • augmentation around implants placed in delayed extraction sockets;
    • localized ridge augmentation for later implantation;
    • alveolar ridge reconstruction for prosthetic treatment;
    • filling of bone defects after root resection, cystectomy, removal of retained teeth;
    • guided bone regeneration in dehiscence defects, and
    • guided tissue regeneration procedures in periodontal defects.

    Geistlich Bio-Gide® Perio is intended for the following uses:

    • augmentation around implants placed in immediate extraction sockets;
    • augmentation around implants placed in delayed extraction sockets;
    • localized ridge augmentation for later implantation;
    • alveolar ridge reconstruction for prosthetic treatment;
    • filling of bone defects after root resection, cystectomy, removal of retained teeth;
    • guided bone regeneration in dehiscence defects, and
    • guided tissue regeneration procedures in periodontal defects.

    Geistlich Bio-Oss Collagen® is intended for the following uses:

    • augmentation or reconstructive treatment of the alveolar ridge;
    • filling of periodontal defects;
    • filling of defects after root resection, apicoectomy, and cystectomy;
    • filling of extraction sockets to enhance preservation of the alveolar ridge;
    • elevation of the maxillary sinus floor;
    • filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); and
    • filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR).
    Device Description

    Geistlich Bio-Gide® is a pure collagen membrane with a bilayer structure. The porous surface (facing the bone) allows the ingrowth of bone forming cells, and the dense surface (facing the soft tissue) prevents the ingrowth of fibrous connective tissue into the bone defect. The membrane is made of collagen without further cross-linking and is sterilized by gamma irradiation.
    Geistlich Bio-Gide® is provided in the following sizes: 13 x 25 mm, 25 x 25 mm, 30 x 40 mm, 40 x 50 mm.

    Geistlich Bio-Gide® Shape is a pure collagen membrane with a bilayer structure. The porous surface (facing the bone) allows the ingrowth of bone forming cells into the membrane, and the dense surface (facing the soft tissue) prevents the ingrowth of fibrous connective tissue into the bone defect. The membrane is made of collagen without further cross-linking, and is sterilized by gamma irradiation. The Geistlich Bio-Gide® Shape membrane has a pre-shaped form with a maximum width and height of 14 mm x 24 mm, respectively.

    Geistlich Bio-Gide® Compressed is a pure collagen membrane with a bilayer structure. The porous surface (facing the bone) allows the ingrowth of bone forming cells into the membrane, and the dense surface (facing the soft tissue) prevents the ingrowth of fibrous connective tissue into the bone defect. The membrane is made of collagen without further cross-linking, and is sterilized by gamma irradiation. The Geistlich Bio-Gide® Compressed membrane is available in two different sizes, 13 x 25 mm and 20 x 30 mm.

    Geistlich Bio-Gide® Forte is a pure collagen membrane with a bilayer structure. The porous surface (facing the bone) allows the ingrowth of bone forming cells into the membrane, and the dense surface (facing the soft tissue) prevents the ingrowth of fibrous connective tissue into the bone defect. The membrane is made of collagen without further cross-linking, and is sterilized by gamma irradiation. The Geistlich Bio-Gide® Forte membrane is available in five different sizes, 13 x 25 mm, 25 x 25 mm, 20 x 30 mm, 30 x 40 mm, and 40 x 50 mm.

    Geistlich Bio-Gide® Perio is a pure collagen membrane with a bilayer structure and smoothed dense (cell-occlusive) surface. The modified surface makes the membrane somewhat stiffer in the dry state, and this facilitates cutting the membrane for periodontal applications. The porous surface (facing the bone) allows the ingrowth of bone forming cells, and the dense surface (facing the soft tissue) prevents the ingrowth of fibrous connective tissue into the defect. The membrane is made of collagen without further cross-linking, and is sterilized by gamma irradiation. Pre-formed sterile templates are provided to simplify the cutting of the respective membrane shape. Four templates (uncoated Tyvek®) are packaged with Geistlich Bio-Gide® Perio to serve as an aid to assist the clinician in trimming the Geistlich Bio-Gide® Perio membrane to fit the defect and are in varying shapes to fit the clinical need (e.g., rectangular, interproximal). The templates are packaged as an accessory product with Geistlich Bio-Gide® Perio.

    Geistlich Combi-Kit Collagen is a convenience kit containing one unit of Geistlich Bio-Oss Collagen® and one unit of Geistlich Bio-Gide®. The two devices are packaged in double blisters in one package and then sterilized by gamma irradiation. Geistlich Bio-Oss Collagen® is a combination of purified spongiosa (cancellous) natural bone mineral granules and 10% collagen fibers in a block form (100 mg) and is sterilized by gamma irradiation. Geistlich Bio-Gide® is a pure collagen membrane with a bilayer structure. The porous surface (facing the bone) allows the ingrowth of bone forming cells, and the dense surface (facing the soft tissue) prevents the ingrowth of fibrous connective tissue into the bone defect. The membrane is made of collagen without further cross-linking and is sterilized by gamma irradiation. The size of the Geistlich Bio-Gide® bilayer membrane to be provided in the Geistlich Combi-Kit Collagen convenience kit is 16 mm x 22 mm.

    Geistlich Perio-System Combi-Pack is a convenience kit containing one unit of Geistlich Bio-Oss Collagen® and one unit of Geistlich Bio-Gide® Perio. Geistlich Bio-Oss Collagen® (sold either as an individual unit or as one of the components of Geistlich Perio-System Combi-Pack) is a combination of purified spongiosa (cancellous) natural bone mineral granules and 10% collagen fibers in a block form (100 mg) and is sterilized by gamma irradiation. Geistlich Bio-Gide® Perio (sold either as an individual unit or as one of the components of Geistlich Perio-System Combi-Pack) is a pure collagen membrane with a bilayer structure and smoothed dense (cell-occlusive) surface. The modified surface makes the membrane somewhat stiffer in the dry state, and this facilitates cutting the membrane for periodontal applications. The porous surface (facing the bone) allows the ingrowth of bone forming cells, and the dense surface (facing the soft tissue) prevents the ingrowth of fibrous connective tissue into the defect. The membrane is made of collagen without further cross-linking and is sterilized by gamma irradiation. The size of the Geistlich Bio-Gide® Perio bilayer membrane to be provided in the Geistlich Perio-System Combi-Pack convenience kit and as individual units is 16 mm x 22 mm. Preformed sterile templates are provided to simplify the cutting of the respective membrane shape. Four templates (uncoated Tyvek®) are packaged with Geistlich Bio-Gide® Perio to serve as an aid to assist the clinician in trimming the Geistlich Bio-Gide® Perio membrane to fit the defect, and are in varying shapes to fit the clinical need (e.g., rectangular, interproximal). The templates are packaged as an accessory product with Geistlich Bio-Gide® Perio.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and associated S510(k) summary documents describe a class II medical device, Geistlich Bio-Gide and its variants, which are resorbable bilayer membranes and bone grafting materials. This submission is for a determination of substantially equivalent to a predicate device.

    Crucially, this document is focused on demonstrating substantial equivalence based on material characteristics, manufacturing processes, and performance data for the device itself (a physical membrane and bone grafting material), not on the performance of an AI/ML powered device.

    Therefore, most of the requested information regarding acceptance criteria, training/test sets, expert adjudication, MRMC studies, standalone performance, and effect sizes for AI assistance are not applicable to this type of medical device submission. The "study that proves the device meets the acceptance criteria" in this context refers to the non-clinical performance testing conducted to confirm the physical and biochemical properties of the new device are equivalent to the predicate device, especially after changes to supplier and manufacturing processes.

    Here's an attempt to extract relevant information and note the inapplicable sections based on your request:

    Acceptance Criteria and Device Performance (for a physical medical device)

    1. A table of acceptance criteria and the reported device performance

    Since this is not an AI/ML device, the acceptance criteria are not typically expressed in terms of accuracy, sensitivity, or specificity. Instead, they are based on physical, chemical, and biological properties demonstrating equivalence to a predicate device. The performance data provided is primarily comparative to the predicate.

    Acceptance Criterion (Implied)Reported Device Performance (Summary from Document)
    Material Composition (Porcine Collagen)Identical to predicate device.
    Bilayer Structure (Porous and Dense Surfaces)Identical to predicate device.
    Sterilization Method (Gamma Irradiation)Identical to predicate device.
    Sizes OfferedIdentical or similar to predicate device (differences justified as non-significant, e.g., Bio-Gide Forte).
    Single-Use StatusIdentical to predicate device.
    Surface Morphology (SEM)Evaluations performed; results used to support substantial equivalence.
    Pore Characteristics (Porosity testing per ASTM F2450-18)Evaluations performed; results used to support substantial equivalence.
    Tensile Strength (Elongation measurements per ASTM F2150-19)Evaluations performed; results used to support substantial equivalence.
    Onset Temperature (DSC per ASTM F2212-20)Evaluations performed; results used to support substantial equivalence.
    Suture Pull-Out ForceEvaluations performed; results used to support substantial equivalence.
    Device Solubility (Quantification of extractable proportion)Evaluations performed; results used to support substantial equivalence.
    Collagen Degradation (Enzymatic degradation per ASTM F2212-20)Evaluations performed; results used to support substantial equivalence.
    Molecular Weight Distribution of Proteins (SDS-PAGE per ASTM F2212-20)Evaluations performed; results used to support substantial equivalence.
    Hydration Capacity (Quantification of swelling factor)Evaluations performed; results used to support substantial equivalence.
    Biocompatibility (In vitro and in vivo per ISO 10993-1:2018)Leveraged from predicate device (K212463).
    Sterilization Validation (Per ISO 11137-1,-2,-3)Leveraged from predicate device (K212463 / K171643).
    Shelf-LifeLeveraged from predicate device (K171643).
    Bench PerformanceLeveraged from predicate device (K171643).
    Clinical PerformanceLeveraged from predicate device (K212463).
    Viral Safety (Per ISO 22442-3:2007)Evaluations performed; results used to support substantial equivalence.
    Handling Properties (Only mentioned for Bio-Gide Forte & Bio-Gide Compressed)Slight modifications for Bio-Gide Compressed to improve handling, but final product specifications are equivalent. Evaluations performed for Bio-Gide Forte.

    The "analysis" column in the provided tables consistently states "The material of construction is identical," "The sizes offered are identical/similar," etc., implying the acceptance criterion is indeed identity or substantial similarity to the predicate.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not explicitly stated as a single "test set" in the context of an AI/ML model. The performance data consists of various physical, biochemical, and experimental tests. The number of samples for each specific test (e.g., number of membranes for tensile strength testing) is not provided.
    • Data Provenance: Not specified regarding country of origin. The studies are described as "in vitro and in vivo biocompatibility," "sterilization," "shelf-life," "bench," and "clinical performance studies" leveraged from previous predicate device submissions (e.g., K212463, K171643). These are likely a mix of lab-based and potentially historical clinical data. It is not specified if these are prospective or retrospective studies; however, given they are leveraged from previous clearances, they would be historical for this specific submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This is not an AI/ML diagnostic device requiring expert interpretation or ground truth establishment in that manner. The "ground truth" for a resorbable membrane involves objective physical, chemical, and biological measurements, and comparison to established standards and predicate device characteristics, not expert consensus on image interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No human adjudication method is described or relevant for the physical and chemical performance tests conducted on this medical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is not an AI-assisted device, so MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an AI algorithm. Its "standalone" performance refers to its intrinsic physical and chemical properties.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable / Different Context. The "ground truth" for this device's performance is established by:
      • Objective Measurements: Results of standardized physical and biochemical tests (e.g., SEM, porosity, tensile strength, DSC, solubility, degradation, molecular weight, hydration capacity, suture pull-out force).
      • Regulatory Standards: Compliance with ISO and ASTM standards (e.g., ISO 10993-1, ISO 11137 series, ISO 22442-3, ASTM F2450-18, ASTM F2150-19, ASTM F2212-20).
      • Predicate Device Data: Comparison and leveraging of performance data (biocompatibility, sterilization, shelf-life, bench, clinical) from previously cleared predicate devices. The claim is substantial equivalence, meaning it performs as safely and effectively as the legally marketed predicate.

    8. The sample size for the training set

    • Not Applicable. This is a physical medical device, not an AI/ML model, so there is no "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. As there is no training set for an AI/ML model, this question does not apply.

    Summary of the Study Proving Device Meets Acceptance Criteria (in this context):

    The "study" conducted for the Geistlich Bio-Gide product family in this 510(k) submission primarily consists of a comprehensive battery of non-clinical performance tests combined with the leveraging of existing performance data from previously cleared predicate devices. The purpose of these tests was to demonstrate that modifications (e.g., new slaughterhouse, non-significant manufacturing changes) did not alter the fundamental safety and effectiveness of the device, making it substantially equivalent to its predicates.

    The non-clinical tests included:

    • Physical and Biochemical Testing: SEM (surface morphology), porosity, tensile strength, DSC (onset temperature), suture pull-out force, solubility, enzymatic degradation, SDS-PAGE (molecular weight distribution), and hydration capacity. For Geistlich Bio-Gide Forte and Compressed, handling properties were also evaluated.
    • Other Experimental Testing: Viral safety according to ISO 22442-3:2007.

    These tests, performed on the modified devices, aimed to show that their properties were consistent with a product that would continue to perform as intended and as safely and effectively as the predicate devices. The acceptance criteria were implicitly that the new devices exhibit equivalent performance characteristics to the cleared predicate devices, as supported by these various in vitro and experimental studies and referencing past clinical data from the predicates.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251786

    Validate with FDA (Live)

    Date Cleared
    2025-07-11

    (30 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K251786**
    Trade/Device Name: Geistlich Bio-Oss®; Geistlich Bio-Oss Pen®
    Regulation Number: 21 CFR 872.3930
    Material |
    | Classification Name: | Bone Grafting Material, animal source |
    | Regulation Number: | 872.3930

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Geistlich Bio-Oss® is intended for the following uses:

    • Augmentation or reconstructive treatment of the alveolar ridge;
    • Filling of infrabony periodontal defects;
    • Filling of defects after root resection, apicoectomy, and cystectomy;
    • Filling of extraction sockets to enhance preservation of the alveolar ridge;
    • Elevation of the maxillary sinus floor;
    • Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR); and
    • Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR).
    Device Description

    Geistlich Bio-Oss® is a biocompatible bone mineral matrix and is manufactured from purified spongiosa (cancellous) bovine bone mineral granules. The product is provided in granules or block form. Geistlich Bio-Oss® serves as a matrix consisting of interconnected macro- and micropores. The material is highly porous and has a large inner surface area.

    Geistlich Bio-Oss® is provided sterile in the following configurations:

    • Geistlich Bio-Oss® spongiosa (cancellous) granules (0.125 g, particle size 0.25 – 1.0 mm)
    • Geistlich Bio-Oss® spongiosa (cancellous) granules (0.25 g, particle size 0.25 – 1.0 mm)
    • Geistlich Bio-Oss® spongiosa (cancellous) granules (0.5 g, particle size 0.25 – 1.0 mm)
    • Geistlich Bio-Oss® spongiosa (cancellous) granules (1.0 g, particle size 0.25 – 1.0 mm)
    • Geistlich Bio-Oss® spongiosa (cancellous) granules (2.0 g, particle size 0.25 – 1.0 mm)
    • Geistlich Bio-Oss® spongiosa (cancellous) granules (5.0 g, particle size 0.25 – 1.0 mm)
    • Geistlich Bio-Oss® spongiosa (cancellous) granules (0.5 g, particle size 1.0 – 2.0 mm)
    • Geistlich Bio-Oss® spongiosa (cancellous) granules (1.0 g, particle size 1.0 – 2.0 mm)
    • Geistlich Bio-Oss® spongiosa (cancellous) granules (2.0 g, particle size 1.0 – 2.0 mm)
    • Geistlich Bio-Oss® spongiosa (cancellous) block (approx. 1 x 1 x 2 cm)

    Geistlich Bio-Oss Pen® is a pre-filled syringe-like applicator containing Geistlich Bio-Oss® granules. The pen allows for targeted delivery of Geistlich Bio-Oss® granules to the intended treatment site without the need for other sterile instruments.

    Geistlich Bio-Oss Pen® is provided sterile in the following configurations:

    • Geistlich Bio-Oss Pen® (filled with 0.25 g, particle size 0.25 – 1.0 mm)
    • Geistlich Bio-Oss Pen® (filled with 0.5 g, particle size 0.25 – 1.0 mm)
    • Geistlich Bio-Oss Pen® (filled with 0.5 g, particle size 1.0 – 2.0 mm)
    AI/ML Overview

    This document is a 510(k) clearance letter for a bone grafting material, Geistlich Bio-Oss® and Geistlich Bio-Oss Pen®. The core claim of the submission (K251786) is that the new devices are "substantially equivalent" to previously cleared predicate devices.

    The request asks for information typically found in an FDA submission for AI/ML-enabled devices, particularly those involving diagnostic aids. This 510(k) submission, however, is for a physical medical device (bone grafting material) and does not involve AI or algorithms for diagnostics or image analysis. Therefore, many of the requested elements are not applicable to this type of traditional medical device clearance.

    Here's a breakdown of the requested information based on the provided document, highlighting what is not applicable:


    Acceptance Criteria and Device Performance for Geistlich Bio-Oss® and Geistlich Bio-Oss Pen® (K251786)

    It is crucial to understand that this 510(k) is for a physical bone grafting material, not an AI/ML-enabled diagnostic device. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are focused on demonstrating substantial equivalence to predicate devices through material properties, manufacturing consistency, biocompatibility, sterilization, and basic handling characteristics, rather than diagnostic performance metrics (e.g., sensitivity, specificity, AUC) and reader studies typically associated with AI.

    The primary "acceptance criterion" for this 510(k) is demonstrating substantial equivalence to the predicate devices, particularly concerning the addition of an alternate raw material supplier and minor manufacturing/release testing changes, without raising new questions of safety or effectiveness.

    1. A table of acceptance criteria and the reported device performance

    For a physical bone grafting material, the "acceptance criteria" relate to material composition, physical properties, biocompatibility, sterility, and manufacturing consistency. The provided document details the comparison of characteristics to the predicate and references studies demonstrating these aspects.

    Acceptance Criteria Category (Implied by Submission)Reported Device Performance (as described in 510(k))
    Material CompositionIdentical to predicate: purified spongiosa (cancellous) bovine bone mineral granules. (New: alternate raw material supplier for bovine material, deemed not to raise new safety/effectiveness questions.)
    FormIdentical to predicate: Granules or block form. Geistlich Bio-Oss Pen® is granules pre-filled in a syringe-like applicator.
    Particle SizeIdentical to predicate: 0.25 – 1.0 mm, 1.0 – 2.0 mm.
    Block SizeIdentical to predicate: ~1 x 1 x 2 cm.
    Single-UseIdentical to predicate: Yes.
    Sterilization MethodIdentical to predicate: Gamma, X-ray (Geistlich Bio-Oss® only).
    Manufacturing MethodsNon-significant changes to manufacturing facilities and equipment. Deemed not to raise different questions of safety/effectiveness.
    Release Testing MethodsNon-significant changes to release testing methods. Deemed not to raise different questions of safety/effectiveness.
    BiocompatibilityLeverage results from predicate/reference devices (K120601, K240661). Conforms to ISO 10993-1:2018.
    Sterilization & Shelf-LifeLeverage results from predicate/reference devices (K120601, K240661). Conforms to ISO 11137-1/2/3 and ISO 11607-1/2.
    Viral InactivationLeverage results from reference devices (K242510, K240661). Conforms to ISO 22442-1/2/3.
    Handling and Performance (Bio-Oss Pen)Qualitative and quantitative handling and performance studies (and usability study) undertaken to support design modifications to the syringe. (Details are not in the provided text, but mentioned as having been done.)
    Clinical Performance (General)Leverage results from predicate/reference devices (K120601, K240661, K122894, K970321) for bench and non-clinical/clinical performance. No new clinical trials were explicitly required or presented in this summary for the "substantial equivalence" claim.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable in the typical AI/ML context. For a physical device like a bone graft, "test set" would typically refer to batches subjected to quality control, biocompatibility testing, or perhaps animal studies for efficacy. The document references leveraging prior studies for performance data.
    • Data Provenance: The document does not specify the country of origin for the leveraged study data or whether it was retrospective or prospective, as this level of detail is not required for a 510(k) summary focused on substantial equivalence of a material.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This pertains to establishing ground truth for diagnostic image interpretation, which is not relevant for a bone grafting material. Ground truth for a bone graft is established through material characterization, biocompatibility testing (e.g., cytotoxicity, sensitization, implantation tests analyzed by pathologists), and gross/histological evaluation in animal models. These "experts" would be materials scientists, toxicologists, and veterinary pathologists, but their number and specific qualifications are not detailed in this 510(k) summary because new studies of this nature were not performed for this submission; prior data was leveraged.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. This refers to consensus methods for establishing ground truth in diagnostic studies (e.g., by radiologists). This is irrelevant for a bone grafting material.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is specific to AI-enabled diagnostic devices assessing human reader performance. This device is a physical bone graft.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is specific to the performance of an AI algorithm in isolation. This device does not have an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For this type of device, "ground truth" for performance relates to:
      • Material Characterization: Physical and chemical properties confirmed through lab assays (e.g., composition, particle size, porosity).
      • Biocompatibility: Established through standardized in vitro and in vivo tests (e.g., cytotoxicity, irritation, sensitization, genotoxicity, implantation tests), with histological and pathological evaluations of tissues for reaction.
      • Sterility: Confirmed through microbiological testing.
      • Viral Safety: Demonstrated through validated viral inactivation processes.
      • Bench and Non-clinical/Clinical Performance: Based on previous studies (potentially animal models or human clinical data from the predicate) demonstrating the material's ability to integrate with bone, support bone formation, etc. The document generally mentions "bench and non-clinical/clinical performance" data leveraged from predicates.

    8. The sample size for the training set

    • Not Applicable. This refers to AI model training data. This device does not use an AI model.

    9. How the ground truth for the training set was established

    • Not Applicable. This refers to how data used to train an AI model was labeled or validated. This device does not use an AI model.

    In summary, the provided FDA 510(k) letter is for a traditional physical medical device. The concepts of "acceptance criteria" and "study proving device meets acceptance criteria" for such devices revolve around demonstrating that the new device (or changes to an existing one) meets established safety and performance benchmarks relevant to its physical and biological function, primarily by showing substantial equivalence to existing, cleared devices. This is a fundamentally different assessment from that of an AI/ML diagnostic tool, which would necessitate the detailed information requested in the prompt.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251613

    Validate with FDA (Live)

    Device Name
    SwissGraft X
    Date Cleared
    2025-06-26

    (30 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Indiana 46582

    Re: K251613
    Trade/Device Name: SwissGraft X
    Regulation Number: 21 CFR 872.3930
    June 26, 2025

    Re: K251613
    Trade/Device Name: SwissGraft X
    Regulation Number: 21 CFR 872.3930
    Material |
    | Classification Name: | Bone Grafting Material, animal source |
    | Regulation Number: | 872.3930

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SwissGraft X is intended for the following uses:

    • augmentation or reconstructive treatment of the alveolar ridge
    • filling of infrabony periodontal defects
    • filling of defects after root resection, apicoectomy, and cystectomy
    • filling of extraction sockets to enhance preservation of the alveolar ridge
    • elevation of the maxillary sinus floor
    • filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
    • filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR).
    Device Description

    SwissGraft X is a biocompatible bone mineral matrix and is manufactured from purified spongiosa (cancellous) bovine bone mineral granules. SwissGraft X serves as a matrix consisting of interconnected macro- and micropores. The material is highly porous, hydrophilic, and has a large inner surface area. SwissGraft X is sterilized by x-ray irradiation. SwissGraft X is provided in granule form.

    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for a bone grafting material called "SwissGraft X." It does not describe an AI medical device or present a study comparing the device's performance against acceptance criteria in the manner typically expected for AI/software-as-a-medical-device (SaMD) clearances.

    Instead, this document describes a traditional medical device (bone grafting material) seeking clearance based on substantial equivalence to a predicate device. The performance data section refers to standard biocompatibility, sterilization, shelf-life, and packaging validation, along with characterization of structural, mechanical properties, and granule size distribution – all common for a physical medical product.

    Therefore, many of the requested elements (e.g., acceptance criteria for diagnostic performance, sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, ground truth types for training/test sets) are not applicable to the information contained in this specific FDA clearance letter for SwissGraft X.

    Here's an analysis based on the available information, with notes where information is not present or not applicable to an AI device:


    1. Table of Acceptance Criteria and Reported Device Performance

    As "SwissGraft X" is a bone grafting material and not an AI or diagnostic device, the acceptance criteria relate to its physical and biological properties rather than diagnostic performance metrics (like sensitivity, specificity, AUC). The document refers to "testing against final product specifications" and "characterization of structural and mechanical properties" for the new product line's smaller granule sizes and different filling weights compared to the predicate device. Specific numerical acceptance criteria or performance values are not detailed in this summary, but the conclusion states that these evaluations support substantial equivalence.

    Acceptance Criteria Category (Implied from Performance Data)Reported Device Performance
    Material BiocompatibilityResults from applicant's own predicate device (K240661) were leveraged. Implied to meet established standards. Specific results not provided in this summary.
    Sterilization EfficacyResults from applicant's own predicate device (K240661) were leveraged. Implied to meet established standards for X-ray sterilization. Specific results not provided in this summary.
    Shelf-Life StabilityResults from applicant's own predicate device (K240661) were leveraged. Implied to demonstrate stability over its intended shelf life. Specific results not provided in this summary.
    Packaging ValidationResults from applicant's own predicate device (K240661) were leveraged. Implied to ensure package integrity and sterility. Specific results not provided in this summary.
    Final Product Specifications"Testing against final product specifications" undertaken. Implied to meet all defined specifications for the device. Specific numerical specifications or results not provided in this summary.
    Pore, Surface, and Internal Structure Characteristics"Characterization of structural and mechanical properties" undertaken. Implied to be comparable to the predicate device within acceptable ranges. Specific results not provided in this summary.
    Liquid Uptake"Characterization of structural and mechanical properties" undertaken. Implied to be comparable to the predicate device within acceptable ranges. Specific results not provided in this summary.
    Granule Size Distribution"Characterization of structural and mechanical properties" undertaken (including Granule size distribution). The device has a reduced granule size range (0.25-0.6 mm) compared to the predicate (0.25-1.0 mm, 1.0-2.0 mm), but this was the specific change being evaluated and found to be sufficiently similar for substantial equivalence. Specific results not provided in this summary, but the conclusion of substantial equivalence implies acceptable performance.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the 510(k) summary as it typically would be for a clinical validation or AI performance study. The "Performance Data" section states that results from the applicant's own predicate device (K240661) were "leveraged" for biocompatibility, sterilization, shelf-life, and packaging validation. For the specific differences (granule size and filling weights), "testing against final product specifications" and "characterization of structural and mechanical properties" were undertaken. The exact sample sizes for these bench tests are not disclosed.

    Data Provenance: The manufacturer is "Geistlich Pharma AG" based in Wolhusen, Switzerland. The predicate device's data was used, but details on the provenance of those original studies are not given here. The studies mentioned are primarily bench/laboratory based for material characterization, not typically clinical or retrospective/prospective data sets in the AI sense.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This concept is not applicable here. The device is a physical bone grafting material, not a diagnostic AI that requires expert-established ground truth for performance evaluation. Its performance is assessed through laboratory tests (e.g., material characterization, sterilization validation) against technical specifications, not against expert clinical diagnoses.

    4. Adjudication Method for the Test Set

    Not applicable. There is no "test set" in the context of expert adjudication for diagnostic accuracy.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. This is a clearance for a bone grafting material, not an AI device, so MRMC studies are not relevant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. "SwissGraft X" is not an algorithm or software. It is a physical bone grafting material.

    7. The Type of Ground Truth Used

    For a physical device like SwissGraft X, "ground truth" refers to established scientific/engineering principles, material standards, and validated test methods (e.g., for biocompatibility, sterility, material properties) rather than clinical ground truth (like pathology or expert consensus). The product is evaluated against its own "final product specifications" and characterization results are compared against those of the predicate device.

    8. The Sample Size for the Training Set

    Not applicable. The device is a physical product, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. The device is a physical product, not an AI model that requires a ground truth for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250833

    Validate with FDA (Live)

    Date Cleared
    2025-04-15

    (27 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    **
    Trade/Device Name: SwissMembrane X; SwissMembrane X Socket
    Regulation Number: 21 CFR 872.3930
    **
    Trade/Device Name: SwissMembrane X; SwissMembrane X Socket
    Regulation Number: 21 CFR 872.3930
    Bone Regeneration |
    | Classification Name: | Bone Grafting Material |
    | Regulation Number: | 872.3930

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SwissMembrane X is intended for the following uses:

    • augmentation around implants placed in immediate extraction sockets;
    • augmentation around implants placed in delayed extraction sockets;
    • localized ridge augmentation for later implantation;
    • alveolar ridge reconstruction for prosthetic treatment;
    • filling of bone defects after root resection, cystectomy, removal of retained teeth;
    • guided bone regeneration in dehiscence defects; and
    • guided tissue regeneration procedures in periodontal defects.

    SwissMembrane X Socket is intended for the following uses:

    • augmentation around implants placed in immediate extraction sockets;
    • augmentation around implants placed in delayed extraction sockets;
    • localized ridge augmentation for later implantation;
    • alveolar ridge reconstruction for prosthetic treatment;
    • filling of bone defects after root resection, cystectomy, removal of retained teeth;
    • guided bone regeneration in dehiscence defects.
    Device Description

    SwissMembrane X and SwissMembrane X Socket are resorbable collagen membranes made of porcine collagen. The porous surface (facing the bone) allows the ingrowth of bone forming cells, and the dense surface (facing the soft tissue) prevents the ingrowth of fibrous connective tissue into the bone defect. The membranes are made of collagen without further cross-linking and are sterilized by gamma irradiation.

    SwissMembrane X is provided in the following variants and sizes:

    • SwissMembrane X D-Line
      • 13 x 25 mm (rectangle)
      • 25 x 15/25 mm (trapezoid)
      • 30 x 25/40 mm (trapezoid)
    • SwissMembrane X Socket D-Line
      • 14 x 24 mm
    AI/ML Overview

    The provided FDA 510(k) Clearance Letter for SwissMembrane X and SwissMembrane X Socket indicates that the device has been cleared based on substantial equivalence to a predicate device, Geistlich Bio-Gide® and Geistlich Bio-Gide® Shape (K212463).

    This type of clearance (510(k)) generally means that the device does not require new detailed performance data to prove safety and effectiveness if it can demonstrate through comparison that it is as safe and effective as a legally marketed predicate device. Therefore, the document does not describe acceptance criteria for performance, nor does it detail a study proving device performance against specific metrics in the way one might expect for a novel device or a device involving an AI component.

    The document explicitly states:

    "Non-clinical data was not deemed necessary to support the extension to the product line. As the subject devices fall within the range of size configurations cleared under the predicate device, there is no new worst-case configuration."

    "Results from biocompatibility, sterilization, shelf-life, packaging validation, bench and clinical performance studies from the applicant's own predicate device (K212463) were leveraged in support of substantial equivalence."

    Because the clearance is based on substantial equivalence to an existing and already cleared device, the typical requirements for a new performance study with specific acceptance criteria, test sets, and ground truth establishment (as is common for AI/ML-driven devices or novel diagnostic tools) are not present or reported in this 510(k) summary.

    Therefore, the following information cannot be extracted from this document as no new performance study, in the sense of demonstrating a device's de novo performance against predefined metrics, was conducted or reported for this submission.


    Based on the provided document, the following information cannot be extracted for the reasons stated above:

    1. A table of acceptance criteria and the reported device performance: Not applicable. Performance data against specific acceptance criteria for a new study is not provided, as substantial equivalence was based on the predicate device's data.

    2. Sample sized used for the test set and the data provenance: Not applicable. No new test set for proving performance was described. The submission leveraged existing data from the predicate device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No new ground truth establishment process for a performance study is described.

    4. Adjudication method for the test set: Not applicable. No new test set requiring adjudication is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: Not applicable. No MRMC study described. The device is a physical bone grafting material/membrane, not one that typically involves human readers assisted by AI.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. No algorithm is involved.

    7. The type of ground truth used: Not applicable. No new performance study requiring independent ground truth is described.

    8. The sample size for the training set: Not applicable. This is not an AI/ML device; therefore, no training set is relevant.

    9. How the ground truth for the training set was established: Not applicable. No training set is relevant.


    Summary of what the document DOES state regarding "performance data":

    The "Performance Data" section (Page 7) explicitly states that "Non-clinical data was not deemed necessary to support the extension to the product line." It further clarifies that "Results from biocompatibility, sterilization, shelf-life, packaging validation, bench and clinical performance studies from the applicant's own predicate device (K212463) were leveraged in support of substantial equivalence."

    This means the acceptance criteria and performance data for this particular 510(k) submission refer back to the studies and their results that supported the initial clearance of the predicate device (K212463). The current submission argues that since SwissMembrane X and SwissMembrane X Socket are essentially variants (different shapes and sizes within an existing range) of the already cleared predicate device, no new performance studies are required. They are considered "substantially equivalent" based on identical materials, manufacturing, sterilization, and similar use cases, and the safety and efficacy of the underlying technology have already been established by the predicate device's clearance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242510

    Validate with FDA (Live)

    Date Cleared
    2025-03-07

    (196 days)

    Product Code
    Regulation Number
    872.3930
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    46582

    March 07, 2025

    Re: K242510

    Trade/Device Name: Geistlich Bio-Flow® Regulation Number: 21 CFR 872.3930
    |
    | Classification Name: | Bone grafting material, animal source |
    | Regulation Number: | 872.3930

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Geistlich Bio-Flow® is intended for bone regeneration of contained defects around teeth or dental implants for the following uses:

    1. filling of extraction sockets to enhance preservation of the alveolar ridge in contained situations with entirely intact, circumferential bone walls.

    2. filling of contained periodontal defects of limited size with intact lingual and buccal walls, i.e. 3-wall intrabony defects.

    3. filling of contained peri-implant defects of limited size to include 3-wall defects of a size up to 4 mm x 5 mm x 4 mm.

    Device Description

    Geistlich Bio-Flow® is a flowable, sterile, biocompatible bone mineral plus collagen matrix consisting of Geistlich Bio-Oss® granules (K122894) and processed Geistlich Bio-Gide® collagen (K212463) in an 80:20 (dry weight) ratio. Geistlich Bio-Flow® is provided as dry granulated material pre-filled in a mixing syringe (0.2 cc or 0.5 cc fill volumes). Cannulas and a syringe for applying saline or blood to hydrate the product prior to extrusion are included with the product.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device (Geistlich Bio-Flow®), which focuses on demonstrating substantial equivalence to a predicate device rather than presenting a traditional clinical study with defined acceptance criteria and performance metrics for a diagnostic or AI-based device. This document describes the device, its intended use, and non-clinical performance data to support its safety and effectiveness. It does not contain an "acceptance criteria" table with specific thresholds or a detailed study of an AI device's performance against ground truth as would be found in a typical AI/diagnostic device submission.

    However, I can extract the relevant information from the document that best approximates the requested points based on the nature of this submission. Since this is a bone grafting material, the "performance" is assessed through non-clinical (material characterization and animal) studies rather than a multi-reader, multi-case study, or standalone AI performance.

    Here's an interpretation based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't explicitly state quantitative acceptance criteria or a performance table in the format typically used for AI/diagnostic devices (e.g., sensitivity, specificity thresholds). Instead, "performance" is demonstrated through various non-clinical tests and a non-clinical animal study, with the overarching "acceptance criterion" being comparable performance to the predicate device in relevant biological and material characteristics.

    Acceptance Criteria (Implied from the study's aim)Reported Device Performance
    Biocompatibility: Device demonstrates biological safety.Positive: Biocompatibility Studies per ISO 10993-1:2018 (Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Acute Systemic Toxicity, Material Mediated Pyrogenicity, Subacute Systemic Toxicity, Subchronic Systemic Toxicity, Chronic Systemic Toxicity, Implantation (local tissue reaction), Genotoxicity, Hemocompatibility) all performed and found acceptable (implied by clearance).
    Physical and Chemical Properties: Material characteristics are suitable for intended use.Positive: Characterization of chemical properties (Chemical composition, Amino Acid Composition, Molecular Weight Distribution of Soluble Proteins, Enzymatic Collagen Degradation, Collagen Solubility) and physical properties (Porosity and morphology, Particle (Granule) Size, Volume Changes after Incubation) were performed. Characterization of Mineral Component (Bovine Bone Mineral) via FTIR spectroscopy and X-ray diffraction was performed. Outcomes are not explicitly detailed but are presumed acceptable for substantial equivalence. Handling and extrusion studies (Extrusion Force, Material Handling with Blood, Extrudable Volume, Organic and Inorganic Content of Extruded Material) were also performed successfully.
    Sterilization and Packaging: Device maintains sterility and integrity.Positive: Sterilization Validation per ISO 11137-1:2006, ISO 11137-2:2013, and ISO 11137-3:2017. Packaging Validation per ISO 11607-1:2019, ASTM F1980:2007, ASTM F1886/F1886M:2016, ASTM F88:2015, ASTM F1929:2015, and ASTM F2096:2011.
    Stability: Device remains stable over shelf life.Positive: Product stability testing per ICH Q1A(R2) and Collagen stability studies (Molecular Weight Distribution of Soluble Protein 18-months storage, Enzymatic Collagen Degradation 18-months storage, Collagen Solubility 18-months storage) were performed.
    Biological Performance (Animal Study): Device promotes bone regeneration and exhibits comparable resorption to predicate.Positive: In a non-clinical animal performance study, no signs of adverse local tissue effects were observed with Geistlich Bio-Flow® at any time-point. At both 8 and 12 weeks, the bone substitute performance of the test and control groups (Geistlich Bio-Flow® vs. Geistlich Bio-Oss® Collagen) was comparable with respect to relevant parameters, such as Defect Fill Area, Bone Regeneration Height, and Bone-to-Granule-Contact.

    2. Sample size used for the test set and the data provenance

    • Test Set (Animal Study): The document states "a non-clinical performance study was conducted to support the indications for use for the device," and "In a study assessing new bone growth and device resorption (4, 8, and 12 weeks), at both 8 and 12 weeks, the bone substitute performance of the test and control groups was comparable."
      • Sample Size: The document does not specify the exact number of animals or defects studied.
      • Data Provenance: Non-clinical (animal study). No country of origin is specified. It is a prospective study as it involved conducting tests with the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not applicable to this type of submission. The performance study was an animal model, not a human reader study requiring expert interpretation to establish ground truth for a diagnostic output.

    4. Adjudication method for the test set

    • This is not applicable as it was not a human reader study. The animal study results would likely be evaluated by veterinary histopathologists or researchers in a blinded manner, but no specific adjudication method (like 2+1) is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This device is a bone grafting material, not a diagnostic or AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No, a standalone performance study in the context of an algorithm or AI was not done.

    7. The type of ground truth used

    • For the non-clinical performance study (animal study), the "ground truth" would be established by histological analysis (e.g., measurements of new bone formation, defect fill, and resorption characteristics) and potentially other quantitative analyses in the animal model. This falls under outcomes data simulation in an animal model.

    8. The sample size for the training set

    • This concept (training set) is not applicable to the evaluation of this bone grafting material. There is no AI model being trained.

    9. How the ground truth for the training set was established

    • This concept is not applicable as there is no AI model or training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 23