    K Number
    K223182

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-05-01

    (202 days)

    Product Code
    Regulation Number
    872.3070
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    , Flexi Alloy, Hi-S Alloy, WW Dental Amalgam, Hi - Mix, Non Gamma 2 Alloy Regulation Number: 21 CFR 872.3070

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Filling material as a treatment for dental caries

    Device Description

    The subject devices are a mixture (alloy) of silver and several other metals, used by dentists to make fillings for tooth cavities. Amalgam alloys have been the most commonly used direct restorative filling material for over a 100 years.

    AI/ML Overview

    The provided text describes a 510(k) summary for several dental amalgam devices, asserting their substantial equivalence to a predicate device. This submission focuses on comparing the physical and chemical properties of the devices to establish this equivalence, rather than on the performance of an AI-powered diagnostic device.

    Therefore, many of the requested categories (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set sample size, ground truth for training set) are not applicable to this document as it does not describe a study involving an AI-powered diagnostic device that requires such metrics for validation.

    Here's the information that can be extracted relevant to acceptance criteria and performance:

    1. A table of acceptance criteria and the reported device performance:

    Product CharacteristicAcceptance Criteria (from ISO 24234)Subject Devices Performance (Reported as complying)Predicate Device PerformanceDifference & Remarks
    Chemical Composition
    Silver (Ag) (CAS 7440-22-4)≥ 40%a) 44.5%; b) 70%56%Compositions meet the requirements of ISO 24234.
    Tin (Sn) (CAS 7440-31-5)≤ 32%a) 30%; b) 18%27.9%Compositions meet the requirements of ISO 24234.
    Copper (Cu) (CAS 7440-50-8)≤ 30%a) 25.5%; b) 12%15.4%Compositions meet the requirements of ISO 24234.
    Zinc (Zn)≤ 2%Not explicitly stated, but implied to meet criteria0.2%Compositions meet the requirements of ISO 24234.
    Alloy-mercury RatioNot explicitly stated as a single criterion, but values provided1:1 (Mercury 50%)Varies between 1/0.86 and 1/0.96 (46.2% to 49.5% by weight mercury)Compositions meet the requirements of ISO 24234.
    Physical Properties
    Particle shape & sizeNot specified by a standardAdmix - spherical and lathe cut 15 μm - 35 μmAdmix - spherical and lathe cut 15 μm - 35 μmThis parameter is not specified by a technical standard; depends on product characteristics. Amalgams made from lathe-cut powders or admixed powders tend to resist condensation better.
    Compressive strength @ 1hr> 100 MPa171 MPa260 MPaData received is similar and products tested per ISO 24234. All results are within specifications and provide good performance of restoration.
    Compressive strength @ 24hr> 350 MPa443 MPa500 MPaData received is similar and products tested per ISO 24234. All results are within specifications and provide good performance of restoration.
    Working times (minutes) - CondenseNot explicitly a criterion2.5 - 52.5 - 5No significant difference.
    Working times (minutes) - CarvingNot explicitly a criterion4.5 - 75.5 - 7No significant difference.
    Corrosion products (μg/cm²)Not explicitly a criterion22.5No significant difference.
    Ions leached and mercury vapor released during corrosion (ng/cm² in 4 hrs)Not explicitly a criterion< 65< 65No significant difference.
    CreepMax. 2%0.5%0.2%All values are well within the maximum 2% criterion.
    Dimensional change-0.10 to +0.150.1-0.04%All values are within the specified range.
    Trituration time (seconds) High speed Amalgamator for capsule formNot explicitly a criterion4-86 - 8Trituration time depends on spill size, variations in amalgamator. Does not affect safety and effectiveness.
    Presentation formsNot explicitly a criterionCapsules: 1, 2 and 3 spillCapsules: 1, 2, 3 & 5 spillDifference in available spill sizes, but does not impact safety or effectiveness of the material itself.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified in terms of distinct numerical values (e.g., number of amalgam samples tested for each property). The testing was completed in accordance with ISO 24234, which would specify sample size requirements for each test.
    • Data Provenance: Not explicitly stated. The submitter is World Work Srl from Italy, so the testing was likely conducted in Europe or by contractors for the Italian company. The document does not indicate if it was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. This is a material science characterization, not a diagnostic study requiring expert ground truth establishment in the clinical sense. The "ground truth" here is the adherence to established international standards (ISO 24234) and their specified test methodologies.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. The evaluation is based on compliance with a standard (ISO 24234), not on human interpretation or adjudication of diagnostic findings.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI-powered diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not applicable. This is not an AI-powered diagnostic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for the device's performance is adherence to the physical and chemical property specifications set forth by ISO 24234. This standard provides the benchmark against which the device's characteristics are measured.

    8. The sample size for the training set

    • Not applicable. This is not an AI-powered device that undergoes a training phase.

    9. How the ground truth for the training set was established

    • Not applicable. This is not an AI-powered device that undergoes a training phase.

    Study Description (Summary of the "study" conducted):

    The "study" conducted for these dental amalgam devices was non-clinical performance testing according to ISO 24234. The purpose was to demonstrate that the physical and mechanical properties of the subject devices meet the requirements of this international standard.

    The tests carried out included:

    • Package & Capsule contamination
    • Chemical composition and purity of the dental amalgam alloy
    • Large particles in the dental amalgam alloy powder
    • Loss of mass from the capsule during mixing
    • Yield of amalgam from the capsule
    • Consistency of the dental amalgam from capsule to capsule
    • Physical properties (Creep, dimensional change, compressive strength, corrosion)

    Based on the data from these tests, the manufacturer concluded that the subject devices are substantially equivalent to the predicate device (Permite Dental Amalgam Alloy K801639) by demonstrating compliance with the ISO 24234 standard.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140125

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-04-29

    (103 days)

    Product Code
    Regulation Number
    872.3070
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    |
    | Regulation | 21 CFR 872.3070
    Amalgam Alloy (NU Alloy DP 40, NU Alloy DP, NU Alloy DP Active and Micronic) Regulation Number: 21 CFR 872.3070

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    New Stetic NU Alloy DP 40, NU Alloy DP, NU Alloy DP Active and Micronic Dental Amalgam Alloys are commonly used as a filling material for restoring the morphology and function of posterior teeth (cavities class I and II in molars and premolars), when these have lost their structure due to injuries such as cavities or fractures.

    Device Description

    New Stetic Dental Amalgam Alloys for dental amalgam used as a filling material for oral cavities. They are made of silver, tin and copper and free of zinc. The products are packaged in different forms: pre-dosed capsules, tablets and powder.

    AI/ML Overview

    The provided 510(k) summary for K140125 describes the New Stetic Dental Amalgam Alloys and compares them to predicate devices to establish substantial equivalence. The "study" here refers to the performance testing conducted to meet regulatory standards and demonstrate equivalence rather than a clinical trial on human subjects improved with the AI.

    Here's an analysis based on the provided document:

    Acceptance Criteria and Study Details for K140125: New Stetic Dental Amalgam Alloys

    1. Table of Acceptance Criteria and Reported Device Performance:

    The device's performance is assessed against the specifications outlined in ISO 24234:2004 for dental amalgam. The tables below compile the acceptance criteria from this ISO standard and the reported performance of the New Stetic Dental Amalgam Alloys (NU ALLOY DP 40, NU ALLOY DP, NU ALLOY DP ACTIVE, and MICRONIC).

    Acceptance Criteria and Device Performance (NU ALLOY DP 40)

    PropertyAcceptance Criteria (ISO 24234:2004)Reported Device Performance (NU ALLOY DP 40, Lot N°: 020713B)
    Compressive strength (1h)80 MPa min87 MPa (12618 PSI)
    Compressive strength (24h)300 MPa min364 MPa (52794 PSI)
    Maximum Creep (%)2.0% max0.23%
    Dimensional Change (%)-0.10 to +0.20 %-0.05%

    Acceptance Criteria and Device Performance (NU ALLOY DP)

    PropertyAcceptance Criteria (ISO 24234:2004)Reported Device Performance (NU ALLOY DP, Lot N°: 010913)
    Compressive strength (1h)80 MPa min182 MPa (26397 PSI)
    Compressive strength (24h)300 MPa min526 MPa (76290 PSI)
    Maximum Creep (%)2.0% max0.30%
    Dimensional Change (%)-0.10 to +0.20 %-0.05%

    Acceptance Criteria and Device Performance (NU ALLOY DP ACTIVE)

    PropertyAcceptance Criteria (ISO 24234:2004)Reported Device Performance (NU ALLOY DP ACTIVE, Lot N°: 010713)
    Compressive strength (1h)80 MPa min159 MPa (23061 PSI)
    Compressive strength (24h)300 MPa min500 MPa (72519 PSI)
    Maximum Creep (%)2.0% max0.18%
    Dimensional Change (%)-0.10 to +0.20 %-0.06%

    Acceptance Criteria and Device Performance (MICRONIC)

    PropertyAcceptance Criteria (ISO 24234:2004)Reported Device Performance (MICRONIC, Lot N°: 010913M)
    Compressive strength (1h)80 MPa min183 MPa (26542 PSI)
    Compressive strength (24h)300 MPa min398 MPa (57725 PSI)
    Maximum Creep (%)2.0% max0.20%
    Dimensional Change (%)-0.10 to +0.20 %-0.03%

    For all products, the report states: "Although the results are not exactly the same, our tests are carried out according to ISO 24234-2004 and all products are within specifications, which ensure the suitable performance of restoration."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document refers to "Performance /Non-Clinical Testing" which was "completed in accordance to ISO 24234:2004". The results for each alloy type are provided with "Lot N°", indicating that samples from specific production lots were tested. However, the exact sample size (number of specimens per test) used for each physical property measurement is not explicitly stated in this document. The provenance of the data is from New Stetic S.A., located in Guarne - Antioquia, Colombia, who are the submitters of the 510(k). The testing is retrospective, as it was performed on finished product lots.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    This document does not describe the establishment of a "ground truth" by experts in the context of diagnostic interpretation. Instead, the "ground truth" refers to objective physical and chemical properties measured against an international standard (ISO 24234:2004). There is no mention of human experts being involved in establishing the ground truth for these physical and chemical performance tests.

    4. Adjudication Method for the Test Set:

    Not applicable. This is not a study involving human interpretation or subjective assessment that would require an adjudication method. The performance is based on objective measurements against a standard.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    Not applicable. This is a submission for a dental material (amalgam alloy), not an AI-assisted diagnostic device. Therefore, a study comparing human reader performance with and without AI assistance is not relevant and was not conducted.

    6. Standalone (Algorithm Only) Performance Study:

    Not applicable. This device is a physical material, not an algorithm or software. Its performance is evaluated through material science testing.

    7. Type of Ground Truth Used:

    The ground truth used is the objective physical and chemical specifications defined by the international standard ISO 24234:2004. This standard sets the minimum or maximum acceptable values for properties like compressive strength, creep, and dimensional change.

    8. Sample Size for the Training Set:

    Not applicable. This is a physical dental material, not an AI model requiring a training set.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as there is no training set for this device.