510(k) Data Aggregation

    Why did this record match?
    510k Summary Text (Full-text Search) :

    Dri-Tech Infant Accessory Bag, HUD1631 / Infant Wye Connector with Swivel)

    Regulation Number: 21 CFR 868.5270
    BZE
    Classification Panel: Anesthesiology
    Regulatory Class: Class II
    Regulation Number: 21 CFR 868.5270
    claim does not raise any additional questions of safety and effectiveness. |
    | Regulation Number | 868.5270
    | 868.5270 | Identical. |
    | Prescription vs.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hudson RCI® Infant Dri-Tech™ Breathing Circuits are intended for neonatal, infant, and pediatric patients weighing 32kg or less in a professional healthcare environment as a conduit for warmed and humidified respiratory gases between a patient and a ventilator and includes heated wires for use with the Hudson RCI Neptune Heated Humidifier.

    Device Description

    The Hudson RCI® Infant Dri-Tech Breathing Circuits are intended for neonatal, infant, and pediatric patients weighing 32kg or less in professional healthcare environments as a conduit for respiratory gas between a patient and a ventilator and includes heated wire(s) for use with the Hudson RCI Neptune Heated Humidifier (K131912). The heated wires are intended to aid in maintaining the set patient temperature and minimize condensation in the ventilator tubing.

    The Dri-Tech Infant Breathing Circuits are designed to be used with various patient interfaces such as oral tracheal tubes, nasal tracheal tubes, endo-bronchial tubes, and masks (for noninvasive ventilation therapies such as CPAP, BiPAP, Cannula and Mask usage). As such, Dri-Tech Breathing Circuits are designed and will be manufactured to conform to the test requirements of ISO 5367:2014 "Anaesthetic and respiratory equipment -- Breathing sets and connectors".

    The dual limb configuration consists of 12mm ID inspiratory and expiratory limbs with 22mm machine side connectors that are compliant to ISO 5356-1:2015 Anaesthetic and respiratory equipment – Conical connectors – Part 1: Cones and sockets, and 12mm patient side connectors that utilize the same taper and engagement length as ISO 5356-1:2015 connectors. The infant circuits have two length configurations, a short inspiratory that is 1.12m or a long inspiratory that is 1.6m, either of which can be used with an expiratory limb that is 1.7m, and may be kitted with the accessories described below:

    • Accessory Tee with Closure Cap: Optional tubing adaptor for various connections as required.
    • Remote Port Tubing, 6" and 12": Unheated sections of tubing for use with incubators, isolettes, or other ambient heating to ensure the temperature probe is not heated by the environment.
    • Unheated Ventilator Tubing: Required tubing and adaptors for connecting the ventilator inhalation port to the heated humidifier.

    The single limb configuration is identical to the long version of the inspiratory limb of the dual limb configuration, however, is packaged with a different set of adaptors to facilitate connections and use with non-invasive therapies as described below:

    • Unheated Ventilator Tubing: Optional tubing and adaptors for connecting the ventilator inhalation port to the heated humidifier.
    • Pressure Monitoring Line: Optional accessory tubing for use with units without an Airway Pressure Monitor.

    All Hudson RCI Dri-Tech breathing circuit configurations are packaged in polyethylene (PE) bags to protect the devices from environmental influences like dust and moisture. Included in the PE bags are additional PE bags containing ConchaSmart columns (-KIT configurations only) and included accessories. The breathing circuit PE bags are packaged ten (10) to a cardboard carton for all configurations. They are sold non-sterile.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251770

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-12-15

    (188 days)

    Product Code
    Regulation Number
    868.5905
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    CPAP
    Classification Name: BZD – 868.5905 - Noncontinuous Ventilator (IPPB)
    Secondary: BZE – 868.5270
    Heated Tubing - K241268
    Common/Usual Name: Heated CPAP Hose
    Classification Name: BZE – 868.5270
    Comparison** |
    |---|---|---|---|
    | Product Code | BZD – 868.5905 Noncontinuous Ventilator (IPPB) BZE – 868.5270
    K241268** |
    |---|---|---|
    | Product Code | BZD – 868.5905 Noncontinuous Ventilator (IPPB) BZE – 868.5270
    Breathing System Heater | BZE – 868.5270 Breathing System Heater |
    | Indications for Use | The SleepRes

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SleepRes PAP system is indicated for the treatment of obstructive sleep apnea (OSA) in patients weighing more than 66 lbs. (30 kg.). The SleepRes PAP system is intended for home, hospital / institutional and sleep center settings.

    Device Description

    The SleepRes PAP system consists of a firmware-controlled console, power cable and power adapter, air filter, water tank, and heated breathing tube. The PAP console is composed of a control system, motor system, and heating system. The control system includes the main board, buttons, screen, and sensor. The motor system includes the motor, fan blade, motor outer shell, and silence box. The heating system includes a heating plate and temperature sensor. The PAP console of the subject device does not include foam.

    The SleepRes PAP system also includes the following components of PAP therapy systems:

    • Water tank
    • Air filter
    • Breathing Tube

    The SleepRes PAP system offers several modes:

    • CPAP
      • Provides continuous positive airway pressure at the prescribed pressure
    • APAP
      • Functions to titrate the pressure up (and down as well) to balance reducing or eliminating sleep events while using the lowest effective pressure. Pressure limits are pre-set by the prescribing clinician.
    • Kairos Positive Airway Pressure (KPAP)
      • This algorithm reduces airway pressure during both inspiration and during most of expiration, only returning pressure to the prescribed treatment level towards the end of expiration.

    Connectivity to a Cloud platform via Wi-Fi connection.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the SleepRes PAP System (K251770) offer some details about acceptance criteria and the study that proves the device meets these criteria. However, it's important to note that a 510(k) summary is a high-level overview and does not typically contain all the granular details of a full study report.

    Here's an attempt to extract and present the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The 510(k) summary broadly asserts that the device met performance and safety requirements. Specific, quantifiable acceptance criteria with corresponding performance outcomes are not explicitly listed in a detailed format within the provided document, beyond the comparison table with the predicate.

    Acceptance Criteria (Inferred from Predicate Comparison & Testing)Reported Device Performance (SleepRes PAP System)
    Pressure Range4 to 20 cmH2O
    Pressure Accuracy (Static)+/- 0.5 cmH2O
    Pressure Accuracy (Dynamic)+/- 1.0 -1.5 cmH2O from 10-20 BPM
    Flow Range10-60 lpm
    Flow Accuracy+/- 2 lpm or 10% of reading
    Operating Temperature Range5℃ to 32℃ (41˚F to 95˚F)
    Storage Temperature Range-20℃ to 55℃ (-13˚F to 158˚F)
    Atmospheric Pressure Range700 to 1060 hPa (0-3000m / 0-9880ft)
    Noise / Sound<35 dB(A) (Note: Higher than predicate's 27 dB(A), addressed by labeling mitigation)
    Therapy Algorithm Effectiveness (Comparative Sleep Studies)Substantially equivalent to predicate in disordered breathing events (AHI3a), sleep quality, and sleep architecture.
    BiocompatibilityMet ISO 10993 and ISO 18562 series requirements, including post-accelerated aging tests.
    Electrical Safety and EMCComplied with IEC 60601-1 and related collateral standards.
    Software Verification and ValidationConducted as per FDA guidances.
    Human Factors ValidationConducted as per FDA guidances.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 50 OSA patients.
    • Data Provenance: The document states "Each patient underwent a full night polysomnogram." It does not specify the country of origin of the data. The study involved patients "all of whom were on standard APAP in the home," implying they were existing patients.
    • Retrospective or Prospective: Implied to be prospective for the clinical validation part, as it describes a protocol where patients "underwent a full night polysomnogram" with different treatment conditions (CPAP, APAP, and APAP + KPAP) specifically for the study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not specify the number or qualifications of experts used to establish the ground truth for the test set (polysomnogram scoring). It only mentions "PSG sleep studies."

    4. Adjudication Method for the Test Set

    The document does not specify the adjudication method for the test set (PSG scoring).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    • Was an MRMC study done? No, an MRMC study was not explicitly described. The clinical study focused on patient outcomes comparing devices/algorithms, not on human readers' performance with or without AI assistance.
    • Effect size of human readers improvement: Not applicable, as this was not an MRMC study assessing human reader performance. The study compared the device's therapy algorithms directly through patient outcomes.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    While the clinical study evaluated the device's algorithms (APAP and KPAP) directly through patient polysomnography, it's not described as a "standalone algorithm-only" performance evaluation in the context of interpretation often seen in AI/ML applications. Rather, it evaluated the therapy delivered by the device to patients. The outcomes (AHI3a, sleep quality, sleep architecture) are direct measures of the device's therapeutic effect.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: Polysomnography (PSG) results, specifically metrics like AHI3a (Apnea-Hypopnea Index with 3% desaturation or arousal criterion), sleep quality, and measures of sleep architecture. These are clinical outcomes obtained from detailed sleep studies.

    8. The Sample Size for the Training Set

    The document does not specify the sample size for the training set. It only describes the clinical validation study's test set.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set details are provided, the method for establishing its ground truth is not mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K243244

    Validate with FDA (Live)

    Date Cleared
    2025-07-25

    (287 days)

    Product Code
    Regulation Number
    868.5270
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Columbia 20001

    Re: K243244
    Trade/Device Name: Heated Breathing Tube
    Regulation Number: 21 CFR 868.5270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Heated Breathing Tube is intended to be used for non-invasive respiratory therapy, such as CPAP/BiPAP therapy. It provides air connection between positive pressure ventilation, respiratory humidification therapy instrument, and ventilation mask, nasal cannula, trachea cannula etc. With the function of heating inside breathing tube, it can prevent condensation of the therapy air in the tube.

    This device is single patient use only for adults.

    Device Description

    The Compressor Nebulizer is mainly composed of the compressor and nebulizer kit(Optional). The compressor is mainly composed of shell, compressed motor, pump, fuse wire, air filter, power cord and plug (NB-1100,NB-1101,NB-1102,NB-1103 applicable), PCB (only model NB-1102,NB-1103applicable). The device is equipped with a nebulizer kit including Nebulizer Cup, Air Tube, Mouthpiece(Optional), Adult Mask(Optional), Child Mask(Optional) to easily delivery the medical aerosol. The Nebulizer kit is available in three models/configurations as below.

    • NK-101: Nebulizer Cup, Air Tube, Adult Mask and/or Child Mask, Mouthpiece;
    • NK-301: Nebulizer Cup, Air Tube, Adult Mask and/or Child Mask;
    • NK-501: Nebulizer Cup, Air Tube, Mouthpiece
      This device operates on the Venturi principle. According to the principle of Venturi, the compressed air of the motor is used to form a high-speed air flow through the small pipe mouth. The negative pressure generated drives the liquid or other fluid to spray together on the barrier, and under the high-speed impact makes the droplets turn into mist particles to spray from the outlet trachea.
    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary pertain to a "Compressor Nebulizer" and a "Heated Breathing Tube," but the main body of the 510(k) summary and the comparative particle test tables focus on the "Compressor Nebulizer" (Model NB-1100, NB-1101, NB-1102, NB-1103) and its predicate "NE-C801 Nebulizer Compressor System." The "Heated Breathing Tube" mentioned in the clearance letter is inconsistent with the detailed submission summary. Therefore, the analysis will focus on the Compressor Nebulizer (JOYTECH NB-1103).

    Based on the provided document, the acceptance criteria are implicitly derived from the comparative particle test results between the subject device (JOYTECH NB-1103) and the predicate device (OMRON NE-C801). The study aims to demonstrate substantial equivalence by showing that the differences in technological characteristics do not raise different questions of safety or effectiveness. The core of this demonstration lies in the comparative aerosol performance data.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state numerical acceptance criteria in a "pass/fail" format. Instead, it presents a comparison table and concludes that the devices are "Equivalent" or "Similar" for various parameters. For the aerosol performance metrics, the implication is that the subject device's performance should be comparable to, or within an acceptable range of, the predicate device's performance across different drug types and patient interfaces.

    Given the structure, the reported device performance is the reported data in the comparative particle test tables. The "acceptance criteria" are implicitly that the subject device's performance metrics (MMAD, GSD, Respirable Dose, Respirable fraction, Total delivered Dose, Total delivered Dose fraction, Coarse Particle Fraction, Fine Particle Fraction, Ultra-Fine Particle Fraction) are comparable to (i.e., not significantly worse than) the predicate device.

    Key Performance Metrics for Aerosol Performance (Based on Comparative Particle Test)

    Acceptance Criteria CategorySpecific MetricPredicate Device (OMRON NE-C801) Performance (Adult Mask, Albuterol Sulfate)Subject Device (JOYTECH NB-1103) Performance (Adult Mask, Albuterol Sulfate)Comparison Result/Implicit Acceptance
    Aerosol Particle Size DistributionMMAD (μm)4.173±0.1263.466±0.173Subject device has smaller MMAD, potentially better penetration. Considered "Equivalent."
    GSD1.845±0.0461.988±0.033Slightly higher GSD for subject device, indicating a wider particle size distribution. Considered "Equivalent."
    Coarse Particle Fraction (%) (>4.7μm)45.607±2.48635.924±2.212Subject device has lower coarse particle fraction. Considered "Equivalent."
    Fine Particle Fraction (%) (<4.7μm)54.393±2.48664.076±2.212Subject device has higher fine particle fraction. Considered "Equivalent."
    Ultra-Fine Particle Fraction (%) (<1.0μm)3.207±0.3515.891±0.795Subject device has higher ultra-fine particle fraction. Considered "Equivalent."
    Drug Delivery EfficiencyRespirable Dose (μg)1239.14±150.6741628.633±99.49Subject device delivers more respirable dose. Considered "Equivalent."
    Respirable fraction58.136±2.58367.273±2.066Subject device has higher respirable fraction. Considered "Equivalent."
    Total delivered Dose (μg)2131.073±210.7112418.382±90.219Subject device delivers more total dose. Considered "Equivalent."
    Total delivered Dose fraction (%)40.801±4.14346.227±1.75Subject device has higher total delivered dose fraction. Considered "Equivalent."

    Note: The table above uses Albuterol Sulfate with an adult mask as an example condition. Similar comparisons are presented in the document for other drugs (Ipratropium Bromide, Budesonide), patient interfaces (mouthpiece), and pediatric conditions. The consistent conclusion (Explicitly stated in the document: "Equivalent") across these detailed comparisons serves as the overall acceptance criteria for aerosol performance.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states, "Table: Comparative particle test comparison" and provides mean values with standard deviations (e.g., 3.466±0.173). This indicates that each reported value is the result of multiple measurements. However, the exact sample size (e.g., number of runs, 'n' for each mean and standard deviation calculation) for the particle characterization testing is not explicitly stated in the provided text.

    The data provenance is not mentioned. It is a non-clinical test conducted for regulatory submission rather than a clinical study with patient data origins.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This type of testing (particle size characterization) is generally conducted in a laboratory setting using specialized equipment (e.g., cascade impactors) and established protocols (e.g., USP <1601> or <1602>). It does not involve human experts establishing a "ground truth" in the way a diagnostic imaging study would. The data generated is objective, quantitative measurements of aerosol properties by an instrument. Therefore, this question is not applicable to the type of non-clinical study described.

    4. Adjudication Method for the Test Set

    Since the ground truth is established through objective laboratory measurements rather than human interpretation, an adjudication method (like 2+1 or 3+1 for expert review) is not applicable.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. The study described is a non-clinical laboratory comparison of aerosol performance between the subject device and a predicate device. It assesses the physical characteristics of the aerosol produced, not human reader performance with or without AI assistance. This question is not applicable to this type of device and study.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This question typically applies to AI/software as a medical device (SaMD). The device in question is a Compressor Nebulizer, a physical medical device. Therefore, a standalone algorithm-only performance assessment is not applicable. The performance tested is the physical device's ability to generate aerosols.

    7. The Type of Ground Truth Used

    The ground truth for the device's performance (specifically aerosol characteristics) is established by objective, quantitative laboratory measurements of particle size, distribution, and drug delivery efficiency, as obtained using standard analytical methods (e.g., based on FDA guidance and potentially pharmacopeial standards for aerosol drug delivery systems). This is presented as "Particle Size characterization testing" on Page 13.

    8. The Sample Size for the Training Set

    The document describes non-clinical testing for a physical device, comparing its performance to a predicate device. There is no mention of a training set as this is not a machine learning or AI-based device. This question is not applicable.

    9. How the Ground Truth for the Training Set was Established

    As there is no training set for this device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241268

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-08-02

    (88 days)

    Product Code
    Regulation Number
    868.5270
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Florida 33704

    Re: K241268

    Trade/Device Name: THERMALITE CPAP Heated Tubing Regulation Number: 21 CFR 868.5270
    |
    | ClassificationRegulation: | 21 CFR §868.5270
    | K140424 | Same indications for use andsame or similar technology. | 21 CFR §868.5270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The THERMALITE® CPAP Heated Tubing is a heated wire breathing tube intended to provide warmed and/or humidified breathing gases before they enter a patient's airway. It is indicated for single-patient reuse in the home and in clinical settings, such as hospitals, institutions, sleep laboratories, and sub-acute care facilities. It may be used with noninvasive ventilation for adult patients.
    It is compatible with the Philips Respironics DreamStation 2 heated humidifiers.

    Device Description

    The Exceleron Medical, LLC (Exceleron) THERMALITE® CPAP Heated Tubing (THERMALITE) is a heated tube that connects between a continuous positive airway pressure (CPAP) machine and the patient's interface, typically a CPAP mask. The THERMALITE is compatible with the Philips Respironics DreamStation 1 and DreamStation 2 heated humidifiers.
    The THERMALITE has a unique design that helps prevent condensate from forming within the heated air stream during CPAP therapy by controlling the airflow temperature, creating an environment to maintain water in its vapor form. This condition of condensation is commonly known as "rainout," and is usually the result of temperature and humidity differentials between the device versus the surrounding air.
    The THERMALITE is available in a three (3) or two (2) wire configuration.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for a medical device called "THERMALITE CPAP Heated Tubing." It details the device's characteristics, indications for use, comparison to a predicate device, and a summary of performance testing. However, it does not contain specific acceptance criteria values or detailed study results in the format typically requested for AI/ML device performance.

    Therefore, many of the requested items cannot be fully answered with the provided text. I will answer what is available and indicate where information is missing.

    1. A table of acceptance criteria and the reported device performance

    The document lists various tests performed based on international standards and guidance documents. It states that "Test results indicated that the Exceleron THERMALITE complies with the applicable Standards and guidance documents" and "Test results indicated that the Exceleron THERMALITE complies with internal requirements, applicable Standards, and the guidance documents." However, it does not provide specific numerical acceptance criteria (e.g., "temperature must be within X +/- Y °C") or the exact reported performance values (e.g., "observed temperature was Z °C").

    Acceptance Criteria and Reported Device Performance (as far as available in the text):

    Test CategoryAcceptance Criteria (Implied by standard compliance)Reported Device Performance
    BiocompatibilityCompliance with ISO 10993-1, ISO 10993-18, ISO 18562-1, ISO 18562-2, ISO 18562-3, and FDA guidance for biocompatibility evaluation (tests for cytotoxicity, sensitization, irritation, emissions of particulate matter, volatile organic compounds, extractables and leachables)."Test results indicated that the indirect patient-contact materials in the Exceleron THERMALITE complies with the applicable Standards and guidance documents."
    Electrical SafetyCompliance with IEC 60601-1 and IEC 60601-1-11."Test results indicated that the Exceleron THERMALITE complies with the applicable Standards."
    Electromagnetic CompatibilityCompliance with IEC 60601-1-2 and FDA guidance for EMC of medical devices."Test results indicated that the Exceleron THERMALITE complies with the applicable Standards and guidance documents."
    Performance Testing (Bench)Compliance with internal requirements, IEC 60601-1-6, IEC 62366-1, ISO 5356-1, ISO 5367, ISO 80601-2-70, ISO 80601-2-74, and ISTA Procedure 3A."Test results indicated that the Exceleron THERMALITE complies with internal requirements, applicable Standards, and the guidance documents." The purpose of the device is to provide "warmed and/or humidified breathing gases" and "prevent condensate from forming within the heated air stream during CPAP therapy by controlling the airflow temperature, creating an environment to maintain water in its vapor form." The predicate device description mentions "air passing through the tubing is warmed to or above the dew point (of the air existing the humidifier) reducing or eliminating water condensation and/or pooling of water in the breathing circuit 41 °C. The raising of the gas temperature does not exceed." This implies a performance target but no specific results for the subject device are given beyond general compliance.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The document refers to "test results" and "verification and validation activities" but does not detail the sample sizes for these tests or the data provenance. These are bench tests on the device itself, not clinical studies with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable as the device is a heated tubing for CPAP machines, not an AI/ML diagnostic or therapeutic device requiring expert interpretation of results or establishment of ground truth from clinical data. The "ground truth" for this type of device is established through engineering and performance specifications and testing against those specifications and relevant international standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable for the same reasons as #3. There is no clinical imaging or diagnostic data requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. This is a physical medical device (heated tubing), not a software/AI device that would involve human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. This is a physical medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device is based on engineering specifications, compliance with international standards (e.g., ISO, IEC), and internal requirements for safety and performance (e.g., biocompatibility limits, electrical safety thresholds, EMC limits, and performance parameters for heating and humidification as defined in the standards). It is not derived from clinical expert consensus, pathology, or outcomes data in the way an AI/ML diagnostic device's ground truth would be.

    8. The sample size for the training set

    This question is not applicable. This is a physical medical device, not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    This question is not applicable. See #8.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    510k Summary Text (Full-text Search) :

    Ventilator Accessory; HUD901/Dri-Tech Ventilator Accessory Extension Line Regulation Number: 21 CFR 868.5270
    Classification Panel: Anesthesiology (Respiratory) Regulatory Class: Class II Regulation Number: 21 CFR 868.5270
    |
    | Regulation Number | 868.5270
    | 868.5270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hudson RCI Dri-Tech Breathing Circuits are intended for pediatric and adult patients in professional healthcare environments as a conduit of breathing gas between a patient and a ventilator and to maintain the temperature of humidified inspired gas.

    Device Description

    The Hudson RCI® Dri-Tech Breathing Circuits are intended for pediatric and adult patients in professional healthcare environments as a conduit for respiratory gas between a patient and includes heated wire(s) for use with the Hudson RCI Neptune Heated Humidifier (K131912). The heated wires are intended to aid in maintaining the set patient temperature and minimize condensation in the ventilator tubing.

    The Dri-Tech Breathing Circuits are designed to be used with various patient interfaces such as oral tracheal tubes, nasal tracheal tubes, endo-bronchial tubes, and masks (for noninvasive ventilation therapies such as CPAP, BiPAP, and Mask usage). As such, Dri-Tech Breathing Circuits are designed and will be manufactured to conform to the test requirements of ISO 5367:2014 "Anaesthetic and respiratory equipment -- Breathing sets and connectors".

    The adult dual limb configuration consists of 19mm ID inspiratory limbs with 22mm connectors that are compliant to ISO 5356-1.2015 Anaesthetic and respiratory equipment – Conical connectors – Part 1: Cones and sockets. The adult circuits have a length of 1.7m on the inspiratory limb, and are kitted with the accessories described below:

    • Tubing Adaptor: 22mm male to male adaptor for various connections as required.
    • . Elbow Adaptor: Adaptor for connecting to the circuit wye to the patient interface to allow the circuit to exit the patient perpendicularly for optimal placement.
    • . Tubing Hanger: Optional accessory for hanging the circuit to reduce pull on the patient interfaces and ventilator/humidifier connections.
    • . Humidifier Limb: Required tubing and adaptors for connecting the ventilator inhalation port to the heated humidifier.

    Optional Ventilator Accessory available for the Dri-Tech Dual Limb Breathing Circuit:

    • . Dri-Tech Ventilator Accessory: Optional tee accessory capable of providing the 10 LPM flow required for the expiratory star lumen tubing. This ventilator accessory is comprised of male and female medical grade air DISS connectors to connect between the air hose and ventilator, a standard Christmas tree adaptor for connecting the tubing to, and an on/off valve to start and stop the flow.
      The adult single limb configuration is identical to the inspiratory limb of the dual limb configuration, however, is packaged with a different set of adaptors to facilitate connections and use with non-invasive therapies as described below:

    • Humidifier Line: Optional tubing and adaptors for connecting the ventilator inhalation port to the heated humidifier.

    • Exhalation Port Adaptor: Optional adaptor including an exhalation port and connector for the pressure monitoring line.

    • Pressure Monitoring Line: Optional accessory tubing for use with units without an Airway Pressure Monitor.

    • Pressure Monitoring Assembly: Optional accessory containing a tee and tubing for pressure monitoring on units with an Airway Pressure Monitor.

    • 15mm Adaptor: Optional accessory required for connecting to certain patient interfaces.

    • Male Luer: Optional accessory for required for certain setups to connect the pressure monitoring line.

    • Tubing Hanger: Tubing hanger designed for single limb tubes and pressure monitoring lines to reduce pull on the patient interfaces and ventilator/humidifier connections.

    • Bacterial/Viral Electrostatic Filter: Optional accessory in specific kit codes for user convenience.
      The Hudson RCI® Dri-Tech Breathing Circuits are sold non-sterile.

    AI/ML Overview

    The provided document describes the FDA 510(k) premarket notification for the Hudson RCI Dri-Tech Breathing Circuits. This document outlines the device's characteristics, intended use, and a comparison with a predicate device, along with a summary of non-clinical testing performed to demonstrate substantial equivalence.

    However, the document does not contain specific acceptance criteria in the form of numerical performance targets, nor does it detail a study that proves the device meets such criteria using a test set, expert adjudication, or MRMC/standalone performance metrics. The information provided heavily focuses on bench testing to established industry standards (e.g., ISO 5367:2014, ISO 80601-2-74:2017) and biocompatibility testing.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance as requested, because specific numerical acceptance criteria (e.g., a specific percentage for sensitivity or specificity) are not stated for AI/algorithm performance. The document describes a comparison to a predicate device and adherence to recognized standards, which are the basis for the FDA's substantial equivalence determination for this breathing circuit, not an AI-powered diagnostic device.

    Similarly, other requested information such as sample size for the test set, data provenance, number and qualifications of experts, adjudication methods, MRMC study details, standalone performance, and ground truth establishment are not applicable to the type of device and testing described in this FDA submission. The device is a medical accessory, not an AI/ML-driven diagnostic or assistive tool.

    Key takeaways from the document regarding "acceptance criteria" and "study":

    • Acceptance Criteria (as implied by the regulatory submission): The "acceptance criteria" for this device are implicitly tied to demonstrating substantial equivalence to a predicate device and meeting the requirements of recognized international standards for breathing circuits. These standards define performance characteristics such as resistance to flow, compliance, and leakage.

      • Resistance to Flow: Less than 1.8 cmH2O @ 30 LPM (per ISO 5367:2014)
      • Compliance: Less than 5 mL/cmH2O @ 60 cmH2O (per ISO 5367:2014)
      • Leakage: Less than 30 ml/min @ 60 cmH2O (per ISO 5367:2014)
      • Biocompatibility: Meeting the requirements of ISO 10993-1:2018 and ISO 18562-1:2017 for gas pathway evaluation (particulate matter, VOCs, leachables in condensate).
      • Safety: Meeting ISO 80601-2-74:2017 (Safety Testing), IEC 60601-1:2005 (Electrical Safety), and IEC 60601-1-2:2014 (EMC).
      • Useful Life: Meeting requirements for an intended 12-day useful life.
    • Study That Proves the Device Meets the Criteria: The "study" described is a compilation of non-clinical bench testing.

      • Performance Testing (Bench): This included verification of features like length, leakage rate, compliance, resistance to flow (with and without bend), adaptor retention, circuit weight, ventilator accessory flow rate, expiratory star lumen tubing kink and crush resistance, moisture removal, and pressures.
      • Biocompatibility Testing: Evaluated cytotoxicity, irritation/intracutaneous reactivity, sensitization, genotoxicity, material mediated pyrogenicity, and extractables/leachables. Gas pathway evaluation included particulate matter and VOCs.
      • Electrical Safety and EMC Testing: Per IEC 60601-1 and IEC 60601-1-2.
      • Human Factors and Usability Testing: Summative testing was performed.

    Therefore, I cannot populate the table or provide detailed responses to the AI/ML-specific questions as they are not relevant to the provided text.

    In summary, based on the provided document:

    1. A table of acceptance criteria and the reported device performance:

      • Acceptance Criteria (derived from standards):

        • Resistance to Flow: < 1.8 cmH2O @ 30 LPM (ISO 5367:2014)
        • Compliance: < 5 mL/cmH2O @ 60 cmH2O (ISO 5367:2014)
        • Leakage: < 30 ml/min @ 60 cmH2O (ISO 5367:2014)
        • Useful Life: 12 days
        • Biocompatibility: Pass ISO 10993-1:2018 & ISO 18562-1:2017 (Particulate matter, VOCs, etc.)
        • Electrical Safety/EMC: Pass IEC 60601-1, IEC 60601-1-2, ISO 80601-2-74:2017
      • Reported Device Performance: The document states the device met the requirements of the standards for these parameters. For instance, for resistance to flow, it states "At each flow rate the resistance to flow is below 1.8 cmH2O (1.8 hPa) as required by the standard." For compliance, "At 60 cmH2O the compliance is below... 5 mL/cmH2O (adult) as required by the standard." For leakage, "At 60 cmH2O the leakage is 40 ml/min or less as required by the standard." Specific numerical results are not provided beyond meeting the thresholds.

    2. Sample sizes used for the test set and the data provenance: Not applicable to this type of device testing. Testing was non-clinical bench testing of the physical breathing circuits.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a breathing circuit's physical performance is established by direct measurement against engineering specifications and industry standards.

    4. Adjudication method: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done: No. This is not an AI/ML-driven device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.

    7. The type of ground truth used: For physical performance testing, the ground truth is defined by the numerical limits and methodologies specified in the relevant international standards (e.g., ISO 5367, ISO 80601). For biocompatibility, it's defined by the toxicological and chemical profiling methods in ISO 10993 and ISO 18562.

    8. The sample size for the training set: Not applicable. The device is not based on AI/ML.

    9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222822

    Validate with FDA (Live)

    Date Cleared
    2023-06-14

    (268 days)

    Product Code
    Regulation Number
    868.5270
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Kit (MP02280), VentStar Helix heated (N) Plus (MP02608) Regulation Number: 21 CFR 868.5270 Regulation
    Helix Heated (N) PlusClassification Name: Breathing system heaterRegulation Number: 21 CFR 868.5270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    VentStar Helix dual heated N Exten. Kit (MP02280): Inspiratory heated disposable breathing circuit with humidifier chamber for connection to a humidifier MR850 by Fisher & Paykel, for neonatal patients with a tidal volume of up to 100 mL, for conducting humidified breathing gas from humidifier to patient.

    VentStar Helix heated (N) Plus (MP02608): Inspiratory heated disposable breathing circuit with humidifier chamber for connection to a humidifier MR850 by Fisher & Paykel, for neonatal patients with a tidal volume of up to 100 mL, for conducting humidified breathing gas from humidifier to patient.

    Device Description

    The subject devices (VentStar Helix Dual Heated N Exten. Kit and VentStar Helix Heated (N) Plus), two inspiratory (and expiratory) heated disposable breathing circuits with humidifier chamber are designed for neonatal patients with a tidal volume of up to 100 mL, for conduction of humidified breathing gas from the humidifier to the patient. VentStar Helix Dual Heated N Exten. Kit is a dual heated breathing circuit and VentStar Helix Heated (N) Plus is a single heated breathing circuit to be connected to a humidifier MR850 by Fisher and Paykel, both tested for system compatibility and released for use with Dräger Babylog VN-Series ventilators.

    Additionally, this submission comprises a neonatal humidifier chamber and a double connector for connection to the ventilators as accessories to the above mentioned medical devices.

    AI/ML Overview

    This is a 510(k) premarket notification for breathing system heaters, not an AI/ML device. Therefore, the requested information about acceptance criteria for AI performance, sample sizes for test/training sets, ground truth establishment, expert consensus, MRMC studies, or standalone algorithm performance is not applicable.

    The document discusses the substantial equivalence of the VentStar Helix Dual Heated N Exten. Kit (MP02280) and VentStar Helix Heated (N) Plus (MP02608) to previously cleared predicate devices. Substantial equivalence is determined by demonstrating that the new devices are as safe and effective as the predicate devices, despite some technological differences.

    Here's the relevant information about the device's technical specifications and the non-clinical studies performed to demonstrate its safety and effectiveness:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document provides extensive comparison tables (Table 1 and Table 2) between the new devices and their predicate devices. These tables list various technical characteristics and performance parameters, indicating whether they are "Same," "Similar," or "Different" compared to the predicate. For characteristics marked "Different," numerical values are provided for both the subject device and the predicate device. These differences were evaluated through non-clinical testing to ensure they do not raise new questions of safety or effectiveness.

    Due to the extensive nature of the comparison, which lists numerous parameters like hose diameter, length, resistance, compliance, leakage, volume, warm-up time, humidification output, and electrical data, a comprehensive reproduction of all "acceptance criteria" (which are implicitly the performance of the predicate device or established standards) and "reported device performance" in a single table format is beyond the scope of a short answer. However, the tables clearly highlight these comparisons.

    For example, for VentStar Helix Dual Heated N Exten. Kit (MP02280) compared to predicate RT265 (K103767):

    CharacteristicSubject Device PerformancePredicate Device PerformanceComment
    Hose diameter (ID)11 mm ID10 mm IDDifferent
    ResistanceInspiratory: at 2.5 L/min < 0.1 mbar, at 5 L/min < 0.2 mbar, at 15 L/min < 1.0 mbar, at 30 L/min < 3.0 mbar. Expiratory: at 2.5 L/min < 0.1 mbar, at 5 L/min < 0.2 mbar, at 15 L/min < 0.7 mbar, at 30 L/min < 2.4 mbarInspiratory: at 2.5 L/min < 0.1 mbar, at 5 L/min < 0.4 mbar, at 15 L/min < 1.3 mbar, at 30 L/min < 9.6 mbar. Expiratory: at 2.5 L/min < 0.1 mbar, at 5 L/min < 0.2 mbar, at 15 L/min < 1.4 mbar, at 30 L/min < 5.3 mbarDifferent
    Humidification OutputInvasive ventilation at 2 to 60 L/min > 33 mg/L. Non-invasive ventilation at 2 to 60 L/min > 12 mg/L.Invasive ventilation up to 60 L/min > 33 mg/L. Non-invasive ventilation up to 120 L/min > 10 mg/L.Same / Different

    The document states that "Performance data related to each proposed modification has been tested and evaluated" and "High level summary reports included in this 510(k) demonstrate that the subject devices are substantially equivalent to the predicate device." This implies that the acceptance criteria are met if these differences do not negatively impact safety and effectiveness, as determined by the non-clinical testing against relevant standards.

    2. Sample size used for the test set and the data provenance: Not applicable. This is not an AI/ML device that uses test sets of data in the common understanding for AI performance evaluation. The "test set" here refers to the physical devices and materials subjected to non-clinical laboratory testing. The provenance of these physical samples is from the manufacturer, Drägerwerk AG & Co. KGaA (Germany). The studies are prospective in the sense that they are conducted specifically for this 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth as typically understood for AI/ML validation (e.g., expert annotations on medical images) is not relevant for this type of device submission, which relies on engineering and biocompatibility testing against established standards.

    4. Adjudication method for the test set: Not applicable. This concept is not relevant for the type of non-clinical testing performed for this device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI/ML device.

    7. The type of ground truth used: For this device, "ground truth" refers to the established requirements and performance limits defined by international consensus standards and the performance characteristics of the predicate devices. For example, specific thresholds for electrical safety, biocompatibility, resistance, and humidification output are defined by standards like IEC 60601-1, ISO 5367, and ISO 18562. The device's performance is measured against these objective, quantifiable benchmarks.

    8. The sample size for the training set: Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established: Not applicable.

    Summary of Non-Clinical Studies and Standards:

    The study that proves the device meets the acceptance criteria is a series of non-clinical tests covering various aspects:

    • Mechanical Performance: Demonstrated conformance to ISO 5367 (Breathing sets and connectors) and ISO 80601-2-12 (Critical care ventilators).
    • Thermal Safety: Implied by testing against the aforementioned standards and the device's function as a heated breathing circuit.
    • Environmental Conditions: Not explicitly detailed beyond compliance with relevant standards.
    • Electrical Safety and Electromagnetic Compatibility (EMC): Tested according to IEC 60601-1 (basic safety and performance) and IEC 60601-1-2 (EMC).
    • Functional Verification and Performance Capacity and Accuracy: Demonstrated through compliance with ISO 5367, ISO 80601-2-12, ISO 5356-1 (Conical connectors), and ISO 80601-2-74 (Respiratory humidifying equipment).
    • Biocompatibility Testing: Materials and completed products were evaluated through biological qualification safety tests as outlined in ISO 10993-1 ("Biological Evaluation of Medical Devices") and ISO 18562 series ("Biocompatibility evaluation of breathing gas pathways in healthcare"). Specific tests included:
      • Emissions of particulate matter (ISO 18562-2)
      • Emission of VOC with additional humidity (ISO 18562-3)
      • Leachables in condensate (ISO 18562-4)
      • Extraction (ISO 10993-18)
      • Material characterization (ISO 10993-18)
      • Cytotoxicity (ISO 10993-5)
      • Sensitization (ISO 10993-10)
      • Irritation (ISO 10993-10)

    The "acceptance criteria" are implicitly defined by the requirements of these recognized consensus standards and the performance characteristics of the predicate devices, which the subject devices "meet all applicable requirements of" and are found to be "substantially equivalent" to. The testing demonstrates that despite some technological differences, no new safety or effectiveness concerns are raised compared to the predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212031

    Validate with FDA (Live)

    Date Cleared
    2022-04-22

    (297 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    |
    | Device Regulation | Class II, Regulation: 21 CFR 868.5270
    | Class II, Regulation: 21 CFR 868.5270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the delivery of heated, humidified breathing gases to spontaneously breathing adult patients. This breathing set is suitable for use with Fisher & Paykel Healthcare MR850 Humidifiers in hospital and long-term care environments.

    Device Description

    The F&P 850 AirSpiral Adult NIV and NHF Circuit Kit (850A61) is designed to provide a gas conduit between respiratory support equipment and a patient interface such as a mask, nasal prongs or tracheostomy interface for a spontaneously breathing Adult patient.

    The Inspiratory Limb is an extruded bubble tube design 1.6 metres in length, and is intended to connect the gas path from humidification chamber to the patient interface. Additionally, temperature sensor clips and a gown clip are included to aid in cable management and to affix the circuit to a gown or sheet. The subject device is single use, prescription only and is provided in a non-sterile state.

    The Disposable Exhalation Port is an optional accessory for NIV applications. It features a perforated ventilation hole and a 22mm/15mm coaxial taper connectir for connecting to patient interfaces and a 22mm tapered male connector for connecting to the patient end of the Inspiratory Limb. It also features a capped port for connecting to the Pressure Line.

    AI/ML Overview

    I apologize, but the provided text does not contain information about an AI/ML-based medical device or any study related to its performance and acceptance criteria.

    The document is an FDA 510(k) clearance letter for a medical device called the "F&P 850 AirSpiral Adult NIV and NHF Circuit Kit." This device is a heated breathing tube and associated components designed to deliver heated, humidified breathing gases to spontaneously breathing adult patients.

    The 510(k) submission describes:

    • The device's intended use and indications for use.
    • A comparison of its technological characteristics with a predicate device (F&P AirSpiral Heated Breathing Tube K162553) and a secondary predicate (Philips Disposable Exhalation Port K982454).
    • A summary of non-clinical tests performed to demonstrate substantial equivalence to the predicate devices. These tests relate to:
      • Compliance with various medical device standards (e.g., ISO, IEC).
      • Exhalation port equivalence (flow rates).
      • Pressure port equivalence (inspired volume).
      • Inhalation path equivalence (tidal volumes).
      • Human factors evaluation.

    There is no mention of artificial intelligence, machine learning, deep learning, or any form of AI-based algorithm doing image analysis, risk prediction, or any other function that would require the kind of performance data usually associated with AI/ML device clearances (e.g., sensitivity, specificity, AUC).

    Therefore, I cannot provide the requested information about acceptance criteria, study details, sample sizes, expert involvement, or MRMC studies, as the document clearly pertains to a non-AI medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201418

    Validate with FDA (Live)

    Date Cleared
    2021-01-29

    (245 days)

    Product Code
    Regulation Number
    868.5270
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Vegas, NV 89119 USA

    Re: K201418

    Trade/Device Name: Sunset Heated CPAP Tube Regulation Number: 21 CFR 868.5270
    |
    | Classification | 21 CFR 868.5270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sunset Heated CPAP Tube is a heated wire breathing tube intended to provide warmed and/or humidified breathing gases before they enter a patient's airway. It is indicated for single-patient reuse in the home and in clinical settings, such as hospitals, institutions, sleep laboratories, and sub-acute care facilities. It may be used with non-invasive ventilation for adult patients. It is compatible with the Philips Respironics System One Heated Humidifier and Philips Respironics Dreamstation Heated Humidifier.

    Device Description

    The Sunset Heated CPAP Tube warms air, or breathable gas, as it travels to and from the respiratory patient along the breathing circuit of a respiratory system. It reduces condensation that can form in breathing circuit. The Sunset Heated CPAP Tube has a proprietary connector with two locking tabs that makes it compatible with the Respironics System One and with the Respironics Dreamstation series.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the "Sunset Heated CPAP Tube":

    The provided document is a 510(k) Premarket Notification from the FDA, which focuses on establishing substantial equivalence to a predicate device rather than detailing specific performance acceptance criteria and study outcomes in the way a clinical trial or a more comprehensive performance study would. Therefore, much of the requested information (like specific numerical performance criteria, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, or specific ground truths) is not explicitly available in this type of regulatory submission.

    However, I can extract the general categories of testing performed and the standards met, which imply the acceptance criteria.


    1. A table of acceptance criteria and the reported device performance

    The document does not provide a table with specific numerical acceptance criteria alongside reported device performance values. Instead, it indicates that the device was tested against and meets the requirements of various international and national performance and safety standards. This implies that the acceptance criteria are embedded within these standards.

    General Categories of Bench Testing Performed:

    Acceptance Criteria (Implied by Standards Met)Reported Device Performance Statement
    Performance of sleep apnea breathing therapy equipment (per ISO 80601-2-70)Device was tested and meets the requirements of ISO 80601-2-70.
    Performance with interaction of respiratory humidifying equipment (per ISO 80601-2-74)Device was tested and meets the requirements of ISO 80601-2-74.
    Requirements for breathing sets and connectors (per ISO 5367)Device was tested and meets the requirements of ISO 5367. (ISO 5367 compliant)
    Electromagnetic compatibility (EMC) (per IEC 60601-1-2)Device complies with IEC 60601-1-2.
    Electrical safety testing (per ANSI/AAMI/ES 60601-1)Device complies with ANSI/AAMI/ES 60601-1.
    Mechanical safety testingBench testing was carried out, implying satisfactory results. (No specific standard cited for "mechanical safety testing" in the performance list, but covered by general safety standards)
    Human Factors and Usability Engineering (per FDA Guidance)Usability evaluation was conducted as per the guidance. (Implies satisfactory outcome)
    Biocompatibility (per ISO 10993-1, ISO 18562 series)Device meets the requirements of these biocompatibility standards.
    Electrical & Mechanical safety in home healthcare (per IEC 60601-1-11)Device complies with IEC 60601-1-11.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify numerical sample sizes for any test sets. The testing described (bench testing, usability testing, biocompatibility) is indicative of engineering and lab-based evaluations rather than patient studies. Therefore, data provenance in terms of country of origin or retrospective/prospective nature isn't applicable in the context typically understood for clinical data. The testing is non-clinical, likely conducted in a lab environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The type of testing described (bench, safety, biocompatibility, usability) typically relies on expert engineers, testers, and potentially human factors specialists, but the document does not detail their numbers or qualifications, nor does it explicitly mention "ground truth" establishment in the context of expert consensus as one would for image interpretation or diagnosis.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document and is generally not applicable to the types of bench and safety testing described. Adjudication methods like 2+1 or 3+1 are typically used for clinical endpoints or image interpretation where expert disagreement might arise.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study was not done. This type of study is irrelevant for a heated CPAP tube, as it is a physical medical device, not an AI-based diagnostic or assistive technology for human interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This concept is not applicable to the "Sunset Heated CPAP Tube" as it is a hardware device (a heated breathing tube), not an algorithm or software. It does not have an "algorithm only" performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the bench, safety, and biocompatibility testing, the "ground truth" implicitly refers to compliance with the specified international and national standards. For usability testing, the "ground truth" would be adherence to human factors principles and satisfactory user performance/feedback, but the specific validation method is not detailed beyond stating it was "conducted as per the Guidance."

    8. The sample size for the training set

    The concept of a "training set" is not applicable here, as the device is a hardware product, not a machine learning model.

    9. How the ground truth for the training set was established

    As there is no training set mentioned or implied, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173280

    Validate with FDA (Live)

    Date Cleared
    2018-08-02

    (290 days)

    Product Code
    Regulation Number
    868.5270
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K173280

    Trade/Device Name: Neonatal ConchaSmart Breathing Circuits Regulation Number: 21 CFR 868.5270
    |
    | Product Code: | BZE |
    | Regulation Number: | 868.5270

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neonatal ConchaSmart Breathing Circuits are intended for neonatal and infant patients in professional healthcare environments as a conduit for respiratory gas between a patient and a ventilator or Infant Flow Generator (for single limb configuration) and include heated wire(s) for use with the Hudson RCI Neptune Heated Humidifier. The heated wires are intended to aid in maintaining the set patient temperature and minimize condensation in the ventilator tubing.

    Device Description

    The Neonatal ConchaSmart Breathing Circuits are intended for neonatal and infant patients in professional healthcare environments as a conduit for respiratory gas between a patient and a ventilator or Infant Flow Generator (for single limb configuration) and includes heated wire(s) for use with the Hudson RCI Neptune Heated Humidifier. The heated wires are intended to aid in maintaining the set patient temperature and minimize condensation in the ventilator tubing.

    These devices are made of corrugated tubing (10mm in diameter), which houses the heated wires and are kitted with various adaptors and connectors to aid the respiratory care clinician in system configuration. In general these heated wire circuits are connected to both a ventilator and a Hudson RCI Neptune Heated Humidifier. The single limb configuration is compatible with the CareFusion Infant Flow system.

    All Neonatal ConchaSmart Breathing Circuits have a useful life of 21 days. Neonatal ConchaSmart Breathing Circuits are sold non-sterile.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the "Neonatal ConchaSmart Breathing Circuits." The submission aims to demonstrate substantial equivalence to legally marketed predicate devices. The document details the device's indications for use, technological characteristics, and performance data from non-clinical testing.

    However, the information provided focuses on the physical and functional performance of the breathing circuits themselves, rather than an AI/ML-driven device's performance measured against acceptance criteria in a clinical or image-based diagnostic setting. Specifically, it does not contain the following information typically found in an AI/ML device submission for establishing acceptance criteria and proving performance:

    • A table of acceptance criteria for AI/ML performance metrics (e.g., sensitivity, specificity, AUC) and reported device performance against these.
    • Sample sizes for test sets in an AI/ML context (e.g., number of images, patient cases).
    • Data provenance (country of origin, retrospective/prospective) for an AI/ML test set.
    • Number of experts and their qualifications for establishing ground truth for an AI/ML test set.
    • Adjudication methods for AI/ML ground truth.
    • Multi-reader multi-case (MRMC) comparative effectiveness study details (effect size of human reader improvement with AI assistance).
    • Standalone performance (algorithm only without human-in-the-loop) for an AI/ML device.
    • Type of ground truth used (e.g., expert consensus, pathology, outcomes data) for an AI/ML device.
    • Sample size for AI/ML training set.
    • How ground truth for the AI/ML training set was established.

    The document discusses engineering acceptance criteria related to standards for breathing circuits (e.g., resistance to flow, compliance, leakage, useful life, electrical safety, biocompatibility) and confirms that the device passed these tests. It indicates the device functions as intended with a specific humidifier.

    Therefore, I cannot extract the requested information as the document describes a traditional medical device (breathing circuits) and not an AI/ML medical device where such acceptance criteria and study designs (related to AI model performance, human-in-the-loop studies, ground truth establishment for AI) would be relevant.

    The acceptance criteria and study described in the document are limited to the physical and functional aspects of a non-AI/ML medical device.

    The closest information provided related to "acceptance criteria" and "study proving it meets criteria" is:

    Acceptance Criteria (Extrapolated from "Test Objective" and ISO Standards):

    Acceptance Criteria (related to breathing circuits)Reported Device Performance
    Useful Life: Maintain performance after 21 days of use post-1-year accelerated aging.PASS
    ISO 8185:2007 (Respiratory tract humidifiers for medical use): Perform as intended with Hudson RCI Neptune Heated Humidifier post-aging/useful life testing.PASS
    IEC 60601-1:1988 + A1:1991 + A2:1995 (Medical Electrical Equipment Safety): Perform as intended with Hudson RCI Neptune Heated Humidifier post-aging/useful life testing.PASS
    ISO 5367:2014 (Breathing sets and connectors): Perform as intended post-aging/useful life testing.PASS
    ISO 5356-1:2015 (Conical connectors): Perform as intended post-aging/useful life testing.PASS
    Additional Design Testing (incl. component break/slip): Perform as intended post-aging/useful life testing.PASS
    Biocompatibility (ISO 10993 series): Cytotoxicity, sensitization, irritation, extractable/leachable testing.PASS
    Leakage: Less than 30 ml/min at 60 cmH2O (per ISO 5367:2014)Less than 30 ml/min @ 60 cmH2O (Implication of PASS)
    Resistance to Flow (inspiratory/expiratory limb): Less than 1.89 cmH2O @ 2.5LPM (per ISO 5367:2014)Less than 1.89 cmH2O @ 2.5LPM (Implication of PASS)
    Compliance: Less than 1.5 ml/cmH2O @ 60 cmH2O (per ISO 5367:2014)Less than 1.5 ml/cmH2O @ 60 cmH2O (Implication of PASS)

    Study Proving Device Meets Acceptance Criteria:

    The study involved a series of non-clinical, laboratory-based tests.

    1. Sample Size Used for Test Set and Data Provenance: The document does not specify a "sample size" in terms of number of patient cases or images, as it is not an AI/ML study. It refers to testing "the proposed devices," implying physical units of the Neonatal ConchaSmart Breathing Circuits. The tests were performed in a lab setting, not using human data/patients, so "provenance" in the sense of country of origin or retrospective/prospective is not applicable.
    2. Number of Experts and Qualifications for Ground Truth: Not applicable for this type of device where laboratory measurements against international standards establish "truth."
    3. Adjudication Method: Not applicable. Lab testing against defined physical and electrical parameters.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not applicable. This is for AI/ML devices involving human readers.
    5. Standalone Performance: Not applicable in the context of an AI/ML algorithm. The device performance itself is evaluated (e.g., its resistance to flow, electrical safety, etc.).
    6. Type of Ground Truth Used: The "ground truth" here is adherence to specified parameters and performance limits defined by international standards (e.g., ISO 5367, IEC 60601-1, ISO 8185, ISO 5356-1, ISO 10993) and the manufacturer's design verification testing for compatibility and durability.
    7. Sample Size for Training Set & How Ground Truth for Training Set was Established: Not applicable, as this is not an AI/ML device relying on a "training set" of data.
    Ask a Question

    Ask a specific question about this device

    K Number
    K163283

    Validate with FDA (Live)

    Date Cleared
    2017-09-20

    (303 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Circuit Set, with water chamber |
    | Classification Name: | Breathing system heater (21 CFR § 868.5270
    &Product code | 868.5270,BZE
    | 868.5270,BZE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended Use HAMILTON-H900

    The HAMILTON-H900 humidifier is intended to add moisture and to warm breathing gases during invasive and non-invasive mechanical ventilation. The intended area of use is the intensive care ward or the recovery room.

    The HAMILTON-H900 humidifier is a medical device intended for use by qualified, trained personnel under the direction of a physician and within the limits of its stated technical specifications.

    Intended Use HAMILTON-BC8010

    The HAMILTON-BC8010 breathing set is intended to be used together with compatible Hamilton Medical respiratory gas humidifiers during invasive and noninvasive mechanical ventilation of pediatric and neonatal patients.

    Intended Use HAMILTON-BC4010

    The HAMILTON-BC4010 breathing set is intended to be used together with compatible Hamilton Medical respiratory gas humidifiers during invasive and noninvasive mechanical ventilation of pediatric and neonatal patients.

    Device Description

    HAMILTON-H900

    The HAMILTON-H900 respiratory gas humidifier is designed to add heat and moisture to respiratory gases, which are then administered to patients who need invasive or noninvasive ventilation. The breathing gas is passed through a humidifier chamber, where the gas is heated and humidified using an external heat source.

    The HAMILTON-H900 humidifier uses breathing circuits that are recognized automatically when connected to the humidifier. The humidifier uses two heating systems as heat sources:

    • -A heating plate that has contact with the metal plate of a special humidifier chamber inserted into the base of the humidifier prior to application, thus heating the water contained inside the chamber.
    • -The controlled heating within the breathing circuits.

    The respiratory gas exiting the humidifier chamber is continuously monitored using temperature sensors that are integrated into the humidifier base. Additionally, a temperature probe inside the breathing tube is used to measure the temperature of the gas delivered to the patient airway.

    The design of the breathing tubes offers heating performance over the entire length by providing constant heating of the entire tube wall, from the chamber exit to the patient connection. A float mechanism, together with detection of the water level inside the humidifier chamber and the autofill mechanism, keeps a constant water level.

    HAMILTON-BC8010

    The HAMILTON-BC8010 is a single-use, dual-limb breathing circuit comprising heated inspiratory and heated expiratory tubes. The humidifier chamber uses a float mechanism to keep a constant water level inside the auto-feed chamber.

    HAMILTON-BC4010

    The HAMILTON-BC4010 is a single-imb breathing circuit comprising a heated inspiratory tube. The humidifier chamber uses a float mechanism to keep a constant water level inside the auto-feed chamber.

    HAMILTON-HC322/HC310

    The HAMILTON-HC322/HC310 are single-use humidifier chambers with a float mechanism to keep a constant water level inside the auto-feed chamber. The HAMILTON-HC322 humidifier chamber is part of the HAMILTON-BC8022/BC4022 breathing sets and the HAMILTON-HC310 humidifier chamber is part of the HAMILTON-BC8010/BC4010 breathing sets, as cleared in K152029.

    AI/ML Overview

    This is a 510(k) summary for a medical device (respiratory gas humidifier and breathing circuits), not an AI/ML medical device. Therefore, the information requested about acceptance criteria, study details, and ground truth establishment, which are typical for AI/ML device submissions, are not primarily available or applicable in the provided text.

    The document discusses substantial equivalence to predicate devices based on technological characteristics and performance, focusing on non-clinical data. It does not describe an AI/ML algorithm that predicts or diagnoses based on medical images or other data, nor does it detail a study that evaluates the performance of such an algorithm against a ground truth established by experts.

    However, I can provide the acceptance criteria and performance as described for this specific device, based on the provided text, recognizing that it's not in the context of an AI/ML model.

    Here's an analysis of the provided text, focusing on the closest equivalents to your requested information, understanding the context is a medical device, not an AI/ML diagnostic:


    The provided text is a 510(k) summary for the Hamilton-H900 Humidifier and Hamilton-BC series breathing circuits. This document outlines the rationale for the device's substantial equivalence to previously cleared predicate devices. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to engineering and performance specifications and non-clinical testing, rather than the type of clinical performance study typically conducted for AI/ML diagnostic or predictive devices.

    Therefore, many of the requested points (e.g., number of experts for ground truth, adjudication methods, MRMC study, standalone performance for an algorithm) are not applicable to this type of device submission.

    Here is the information extractable from the document, focusing on performance characteristics and non-clinical testing rather than AI/ML specific criteria:

    Acceptance Criteria and Reported Device Performance (Non-AI/ML Context)

    The acceptance criteria are implied by the "Characteristic used for comparison" and the "HAMILTON-H900 (Subject device)" column in Table 1 for the humidifier and Table 2 for the breathing circuits. The "reported device performance" is the statement that the subject device meets these established characteristics, which are largely identical to the predicate device.

    1. Table of Acceptance Criteria and the Reported Device Performance

    For HAMILTON-H900 (Humidifier):

    Characteristic Used for ComparisonAcceptance Criteria (Implied by Subject Device Spec)Reported Device Performance
    Indication for UseAdd moisture and warm breathing gases during invasive/non-invasive mechanical ventilation for adult, pediatric, and neonatal patients in ICU/recovery room.Device intended for this use, within limits of stated technical specifications. Patient population expanded from predicate to include pediatric and neonatal.
    EnvironmentHospital use by trained personnel.Confirmed.
    Type of VentilationInvasive and non-invasive.Confirmed.
    Modes of OperationAutomatic Mode (Invasive & Noninvasive), Manual Mode (Invasive & Noninvasive), Standby Mode, Default Mode (Invasive after power on).Confirmed.
    Flow RangeNon-invasive: up to 120 L/min; Invasive: up to 60 L/min.Confirmed.
    Maximum Operating DurationContinuously.Confirmed.
    Accuracy Displayed TemperatureDisplayed Range: 10°C to 60°C; Chamber exit: 10°C to 60°C ± 1°C, 30°C to 41°C ± 0.5°C; Y-piece: 28°C to 43°C ± 0.5°C.Confirmed.
    Resolution of Displayed Temp.0.1 °C.Confirmed.
    Temp Control Setting: AirwayInvasive: 33°C to 42°C; Noninvasive: 28°C to 38°C; in steps of 0.5°C (manually).Confirmed.
    Temp Control Setting: ChamberInvasive: 35°C to 41°C; Noninvasive: 30°C to 35°C; in steps of 0.5°C (manually).Confirmed.
    Disabling Heater WiresYes; expiratory tube disabled with Single Limb Circuits; temp alarm can disable all heating wires.Confirmed.
    Humidity PerformanceInvasive mode: >33 mg H2O/L @ 60 L/min and 26°C ambient; Noninvasive mode: >10 mg H2O/L @ 120 L/min and 26°C ambient.Confirmed.
    Alarm DisplayIndicators for Chamber Temp, Heater Wire, Humidity Alarm, Patient-end Probe (Y-Piece), water level (Low/High), device inclination, tube detection/recognition/connection, chamber insertion.Confirmed.
    Maximum Power283 VA (230V), 293 VA (115V), 268 VA (100V).Confirmed.
    Power Heated Tube22V; 60 W (dual limb), 30 W (single limb).Confirmed.
    Power Heating Plate150 W ± 5 %.Confirmed.
    Heating Plate Overheat Prot.130°C ± 4°C.Confirmed.
    Chamber Insertion DetectionAble to detect if a chamber is inserted.Confirmed.
    Water Level DetectionHigh and Low water level detected via optical sensors.Confirmed.
    Breathing Circuit RecognitionVia electrical connectors inside the breathing tubes.Confirmed.
    Single Use / ReuseCompatible with single use and reusable breathing circuits.Confirmed.
    ClassificationClass I (IEC 60601-1).Confirmed.
    Applied PartType BF.Confirmed.
    Standards / PerformanceCompliance with IEC 60601-1:2012, IEC 60601-1-2:2007, IEC 60601-1-8:2005+Amd1:2012, ISO 8185:2007, MIL-STD-461F:2007, IEC 62304:2006, IEC 62366:2007+Amd1:2014.Confirmed.

    For HAMILTON-BC8010 / BC4010 (Breathing Circuits):

    Characteristic Used for ComparisonAcceptance Criteria (Implied by Subject Device Spec)Reported Device Performance
    DescriptionBC8010: dual limb, heated with water chamber; BC4010: single limb, heated, with water chamber.Confirmed.
    Classification No. & Product868.5270, BZE.Confirmed.
    Intended UseUse with compatible Hamilton Medical respiratory gas humidifiers during invasive and noninvasive mechanical ventilation of pediatric and neonatal patients.Patient population expanded from predicate to include pediatric and neonatal.
    Patient PopulationPediatric and neonatal patients.Confirmed.
    EnvironmentHospital use by trained personnel.Confirmed.
    Indications for UseInvasive and noninvasive ventilation.Confirmed.
    Compatibility with other DevicesHamilton Medical compatible humidifier (e.g., HAMILTON-H900).Confirmed.
    Where UsedDuring invasive and noninvasive mechanical ventilation, hospital use by trained personnel.Confirmed.
    Target/Patient PopulationPediatric and neonatal patients requiring mechanical ventilation or positive pressure breathing assistance.Confirmed.
    Compressible Volume800 ml (BC8010), 600 ml (BC4010).Confirmed.
    Maximum Chamber Operating Press.20 kPa.Confirmed.
    LengthInspiration: 1.65m total (1.3m heated, 0.35m unheated); Expiration: 1.65m.Confirmed.
    Inner Diameter12 mm.Confirmed.
    Flow Range1 to 30 L/min.Confirmed.
    Flow Resistance@ 2.5 l/min: Inspiration: 0.03 cmH2O/(l/min), Expiration: 0.02 cmH2O/(l/min). @ 15 l/min: Inspiration: 0.06 cmH2O/(l/min), Expiration: 0.04 cmH2O/(l/min).Confirmed.
    Gas Leakage15 mL/min (BC8010), 4.0 mL/min (BC4010).Confirmed.
    Compliance< 1 ml/cmH2O/m.Confirmed.
    Wire ResistanceInspiration: 14.6±0.9Ω; Expiration: 18.2±1.2Ω.Confirmed.
    ConnectorConical according to ISO 5356-1; electrical connector part of breathing tube.Confirmed.
    Breathing Circuit RecognitionTubes recognized by Hamilton Medical Humidifier.Confirmed.
    SterilityNon-sterile, ready for use.Confirmed.
    Single Use / ReuseSingle Use.Confirmed.
    Power Heated Tube33.15W inspiratory tube @100% power (BC8010/BC4010); 26.59W expiratory tube @100% power (BC8010).Confirmed.
    Standards / PerformanceCompliance with various ISO and IEC standards listed (e.g., IEC 60601-1:2012, ISO 8185:2007, ISO 10993 series).Confirmed.

    2. Sample Size Used for the Test Set and Data Provenance

    As this is a medical device submission based on substantial equivalence and non-clinical testing rather than an AI/ML study with a "test set" of patient data, specific "sample sizes" in the sense of clinical cohorts are not detailed. The testing involved:

    • Electrical Safety and Electromagnetic Compatibility: Conducted on the HAMILTON-H900 and HAMILTON-BC-series breathing circuits.
    • Mechanical Testing: Conducted on the HAMILTON-H900 and HAMILTON-BC-series breathing circuits.
    • Software Verification and Validation Testing: Conducted for the software within the HAMILTON-H900.
    • Biocompatibility Testing: Performed on "relevant parts of the breathing circuit."
    • Human Factors / Usability Study: Conducted on the HAMILTON-H900.

    The document does not specify the number of units tested for each category, which is typical for non-clinical testing reports summarized in 510(k)s.

    Data Provenance: The document does not specify the country of origin of the data or explicitly state whether the tests were retrospective or prospective. Given these are physical device tests, they would be prospective (performed specifically for this submission). The applicant is based in Switzerland (Hamilton Medical AG, Via Crusch 8, 7402 Bonaduz, Switzerland).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. "Ground truth" in the context of an AI/ML algorithm with expert adjudication is not relevant here. The "ground truth" for this device's performance is objective measurements against engineering specifications and international standards (e.g., temperature accuracy, flow resistance, humidity output).

    4. Adjudication Method for the Test Set

    Not applicable. There is no expert adjudication for this type of device testing. Performance is measured against predefined engineering and safety standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is for evaluating human reader performance, often with and without AI assistance, typically in diagnostic imaging. This device is a respiratory gas humidifier and breathing circuit, not a diagnostic AI/ML system.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable directly. This device is a physical medical device, not a standalone algorithm. Its performance is evaluated against its physical and electronic specifications.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is established by engineering specifications, adherence to international standards (e.g., IEC 60601 series, ISO 8185, ISO 5367, ISO 5356-1, ISO 10993 series), and objective measurements during non-clinical testing. For example, temperature accuracy is measured against a calibrated standard, not against a human expert's assessment.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is not an AI/ML device that requires a training set with established ground truth.