    510k Summary Text (Full-text Search) :

    510(k) Summary

    K250687

    The following 510(k) summary is being submitted as required by 21 CFR 868.5120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cocoon Solo (DX-7020s) is a portable x-ray system indicated for taking diagnostic dental x-rays for both pediatric and adult patients.

    Device Description

    Cocoon Solo (DX-7020s) is a handheld, portable X-ray device designed for dental radiographic examination and diagnosis for pediatric and adult patients by exposing a X-ray image receptor to ionizing radiation.

    The X-ray tube is located inside the device body to be used with conventional film (F-speed or greater film), PSP (Phosphor plates), or digital X-ray sensors.

    The image detectors (an integral part of a fully-functional diagnostic x-ray system) are not part of the submission.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K241410

    Validate with FDA (Live)

    Device Name
    EDEN ControlCath
    Manufacturer
    Date Cleared
    2025-02-13

    (272 days)

    Product Code
    Regulation Number
    868.5120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    610 Houston, Texas 77054

    Re: K241410

    Trade/Device Name: EDEN ControlCath Regulation Number: 21 CFR 868.5120

    510(k) Summary K241410

    The following 510(k) summary is being submitted as required by 21 CFR 868.5120
    catheter |
    | Classification Name, Number | Anesthesia conduction catheter(21 CFR 868.5120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EDEN ControlCath is used for delivery of drugs that have been indicated for the epidural space.

    Device Description

    EDEN ControlCath consists of a flexible catheter, steering handle, and a port for access to the lumen. The catheter has built in steering mechanism that allows for guiding the soft tip through the epidural space and soft tissues for optimal access to the source of distress. The port facilitates the connection of syringes to deliver therapeutic agents by physicians as appropriate to their diagnosis. It is supplied sterile and it is for single use.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device called "EDEN ControlCath." This document details the device's characteristics, intended use, and its comparison to a predicate device to demonstrate substantial equivalence.

    However, the provided text DOES NOT contain information regarding:

    • Acceptance criteria beyond "test criteria" for various bench tests.
    • Specific device performance metrics that meet these criteria.
    • Study design for proving acceptance criteria, particularly regarding clinical studies or AI model performance. The document explicitly states, "Clinical testing was not required for this submission."
    • Sample sizes used for test or training sets for AI models.
    • Data provenance, expert involvement for ground truth, adjudication methods, MRMC studies, standalone AI performance, or how ground truth was established, as these are typically relevant for AI/ML-based medical devices or diagnostics.

    Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets them based on the provided text. The document focuses on demonstrating substantial equivalence to a predicate conventional medical device through bench testing and material comparisons, not an AI/ML device.

    To answer your request based on the provided text, I can only state the following:

    1. A table of acceptance criteria and the reported device performance:

    The document mentions that various properties were tested based on referenced standards, and "All the test results support substantial equivalence to the predicate devices." However, it does not provide specific numerical acceptance criteria (e.g., "tensile strength > X N") or measured device performance values. It broadly states that the device "passed the test criteria."

    Acceptance Criteria CategoryReported Device Performance
    Non-Clinical Properties (Bench Tests)The device "passed the test criteria for the side-by-side test" for Diameter and Catheter Length.
    Key Performance Specification/CharacteristicsThe device's performance for "Catheter: Tensile Load, Liquid Leakage, Corrosion Resistance, Fatigue, Flexural Strength"; "Trocar: Outer and Inner Surface, Flexural Strength, Pull Out, Elasticity, Pierce"; "Sheath: Tensile Load, Flexural Strength"; and "Guide Wire: Strength" is "Same as predicate device(*)" and passed relevant tests.
    Raw Materials (Biocompatibility)The device "passed the test criteria for the bench test according to the recognized consensus standard (ISO 10993-1 Series)." All directly/indirectly contacting materials are "compliant with ISO10993 requirements."
    SterilizationThe device "passed the test criteria for the bench test according to the recognized consensus" for E.O. gas sterilization, which is a different method than the predicate's Gamma Sterilization, but deemed not to have a "significant difference."
    General CharacteristicsThe device shows "no significant differences in most parameters" compared to the predicate, and "the differences do not affect the substantial equivalence."

    2. Sample sized used for the test set and the data provenance:
    The document does not specify sample sizes for any of the bench tests or material evaluations. It refers to "bench test results" generally. It does not involve patient data, so concepts like "country of origin" or "retrospective/prospective" are not applicable here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    Not applicable. This device is not an AI/ML diagnostic tool, and its evaluation relies on established engineering and materials testing standards, not expert consensus on medical images or clinical outcomes.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
    Not applicable. There is no human adjudication process described for the non-clinical bench testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    No. This is not an AI-assisted device, and no MRMC study was conducted or required.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable. This device is a physical epidural catheter, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):
    The "ground truth" for this device's performance is based on compliance with international standards for medical devices (e.g., ISO, ASTM) through various physical and chemical bench tests (tensile strength, leakage, biocompatibility, etc.) and comparison to a legally marketed predicate device.

    8. The sample size for the training set:
    Not applicable. There is no training set mentioned as this is not an AI/ML device.

    9. How the ground truth for the training set was established:
    Not applicable. There is no training set involved.

    In summary, the provided FDA 510(k) summary is for a traditional medical device (an anesthesia conduction catheter), not an AI/ML-driven product. Therefore, the detailed questions about AI model performance, data sets, and expert ground truth establishment are not addressed in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231242

    Validate with FDA (Live)

    Date Cleared
    2023-09-15

    (137 days)

    Product Code
    Regulation Number
    868.5120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Re: K231242

    Trade/Device Name: Perifix FX Catheter: Contiplex FX Catheter Regulation Number: 21 CFR 868.5120
    |
    | Classification Name | Catheter, Conduction, Anesthetic |
    | Regulation Number | 868.5120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PERIFIX FX Springwound Epidural Catheter is intended for administration of local anesthetic agents into the epidural space to provide continuous epidural or caudal anesthesia. The catheter should be removed or replaced every 72 hours.

    The Contiplex FX catheter is intended to provide continuous and/or intermittent infusion of local anesthetics and analgesics for peripheral plexus anesthesia and pain management. The Contiplex FX catheter may remain indwelling for up to 72 hours.

    Device Description

    The FX catheters are regional anesthesia catheres intended to provide, via percutaneous administration, continuous and/or internittent infusion of local anesthetics and analgesics near a nerve for regional anesthesia and pain management during the preoperative, perioperative and postoperative periods associated with general and orthopedic surgery. The catheters may remain indwwelling for up to 72 hours. Routes of administration are epidural (peripheral nerve block (PNB)). The catheter is available in both open and closed tip designs. The open-tip polyamide catheter is intended to facilitate continuous delivery of anesthetic fluid via one opening at the tip, while the closed-tip polyamide catheter facilitates the administration of anesthetic through three sideports. Both catheters have ink markings, which are located in 10 mm increments along the catheter. The ink markings provide a visual indication of the depth of catheter insertion. When used according to the conditions listed on the product labeling, the FX Catheters may remain in a patient while in an MR environment.

    AI/ML Overview

    This document is a 510(k) summary for the Perifix FX Catheter and Contiplex FX Catheter manufactured by B. Braun Medical Inc. It concerns a traditional medical device (anesthesia conduction catheters) and does not involve Artificial Intelligence (AI). Therefore, the questions related to AI device performance, such as acceptance criteria for AI, training sets, test sets, ground truth establishment, expert adjudication, or MRMC studies, are not applicable to this submission.

    The 510(k) submission (K231242) focuses on demonstrating substantial equivalence to predicate devices (K991879 and K113059). The "Non-Clinical and/or Clinical Tests Summary & Conclusions" section explicitly states:

    "MRI testing was performed. Testing supports the device being labeled as MRI Conditional. No changes have been made to the device therefore, the device continues to be safe and effective and performs the same as the predicate device."

    This indicates that the primary testing performed was related to MRI compatibility, and the unchanged nature of the device from its predicates means that prior safety and efficacy data for the predicates are considered applicable.

    Therefore, I cannot provide the requested information regarding AI device acceptance criteria and study details because this 510(k) submission is not for an AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202699

    Validate with FDA (Live)

    Date Cleared
    2020-12-29

    (104 days)

    Product Code
    Regulation Number
    868.5140
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    |
    | Subsequent Classification Reference #2: | 21 CFR 868.5120
    defined in 21 CFR §868.5140 anesthesia conduction kit, product code CAZ (Subsequent Codes: 21 CFR §868.5120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The E-Cath STIM acc. Tsui System is indicated for delivery of medication for regional anesthesia and pain management. Route of administration may be intraoperative, percutaneous, or perineural. To assist the physician pinpoint the area of application an electrical stimulus can be applied to the conduction needle removal, to assist the physician pinpoint conduction catheter placement, an electrical stimulus can be applied to the conduction catheter tip via the catheter adapter.

    The E-Cath STIM acc. Tsui System is contraindicated for the epidural space.

    Device Description

    The Over-the-needle (OTN) Catheter Systems are available in different designs with cannula and catheter in different sizes. The system includes: SonoPlex cannula, permanent cannula, E-Cath catheter, locking cap.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device (E-Cath STIM acc. Tsui). It is a special 510(k) for a device modification, intending to combine features from previously cleared devices. The document mainly focuses on demonstrating substantial equivalence to predicate devices through bench testing and compliance with recognized standards, rather than clinical studies with human participants. Therefore, several of the requested categories, such as "multi-reader multi-case (MRMC) comparative effectiveness study," "sample size for the test set," "number of experts," "adjudication method," and "standalone (i.e., algorithm only) performance" are not applicable to this type of submission.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and the Reported Device Performance:

    TestAcceptance CriteriaReported Device Performance
    Needle: Bending RigidityAccording to EN ISO 9626: ≤ 0.48mm under an applied force of 15N by a span width of 17.5mm.The bending rigidity of the subject/predicate/reference device's needles is less than 0.46mm. Conclusion: In compliance.
    Needle: Breaking ResistanceAccording to EN ISO 9626: Not to break during a bending test for 20 periods at an angle defined by cannula size.The tested cannulas did not break during the test. Conclusion: In compliance.
    Needle: Bonding to HubAccording to EN ISO 7864: Pull-off force ≥ 44N.For the subject/predicate/reference device's needles, a force significantly higher than the target value has to be applied. Conclusion: In compliance.
    Needle: Penetration ForceEN 13097: No specific pass/fail criteria; objective comparison only.The subject/predicate/reference device's needles show appropriate penetration/insertion forces. Compared to the predicate device, PAJUNK's needles are in compliance. Conclusion: In compliance.
    Catheter: Leak TightnessInternal protocols based on risk assessment and clinical evaluation (standard DIN EN 1618 does not define criteria). Catheter system must be tight for a defined period and pressure.All systems are tight during the test for both the subject device and the predicate/reference devices. Conclusion: In compliance. The subject/predicate/reference device's catheters meet the acceptance criterion.
    Catheter: Tensile StrengthInternally defined acceptable value to be met without tear-off (standard DIN EN 10555-1 and DIN EN 1618 do not define criteria).The catheters of the E-Cath Systems meet the acceptance criterion (without tear-off). Conclusion: In compliance. The subject/predicate/reference device's catheters meet the acceptance criterion.
    Catheter: Flow RateInternally defined by the sponsor based on risk management and clinical evaluation (standard DIN EN 10555-1 and DIN EN 1618 do not define criteria).The flow rate of the tested E-Cath catheter meets the acceptance criterion. The subject/predicate/reference device's catheters meet the acceptance criterion and have proven to supply appropriate flow rates.

    2. Sample Size Used for the Test Set and the Data Provenance:

    • Sample Size: Not explicitly stated for each test, but implied to be sufficient for demonstrating compliance with the referenced ISO/EN standards for medical device components. The tests refer to "tested cannulas" and "tested catheters."
    • Data Provenance: Bench testing, compliance with international standards (ISO, EN, DIN). Results are compared against predicate and reference devices, implying these are internal company data.
    • Retrospective/Prospective: Not applicable in the context of bench testing for a device modification. The tests are performed on the device itself.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    • Not applicable. The ground truth for these tests is defined by international standards (e.g., ISO 9626, ISO 7864, DIN EN 1618, DIN EN 10555-1) and internal protocols for performance criteria where standards are not specific. This does not involve expert medical opinion for "ground truth" in the diagnostic sense.

    4. Adjudication Method for the Test Set:

    • Not applicable. Performance is measured against predefined, objective, quantifiable acceptance criteria established by international standards or internal engineering/risk assessment protocols.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a medical device (anesthesia conduction kit), not an AI diagnostic or image analysis system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical medical device.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for the device's performance is established by compliance with recognized international standards for medical device components (e.g., mechanical properties, sterility, biocompatibility) and internal engineering/risk assessment protocols where standards allow for manufacturer-defined criteria (e.g., leak tightness, tensile strength, flow rate).

    8. The Sample Size for the Training Set:

    • Not applicable. This is not a machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set was Established:

    • Not applicable. As above, this is not a machine learning model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193664

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-09-09

    (253 days)

    Product Code
    Regulation Number
    868.5120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    California 92673

    Re: K193664

    Trade/Device Name: SubQKath Catheter and Needle Set Regulation Number: 21 CFR 868.5120
    Subcutaneous Catheter Classification name: Anesthesia Conduction Catheter Classification Regulation : 868.5120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SubQKath Catheter and Needle Kit is intended to provide continuous and/or intermittent infusion of local anesthetics and analgesics for subcutaneous infiltration local anesthesia, peripheral plexus anesthesia and pain management during pre-operative, perioperative and post-operative periods associated with surgical procedures.

    Device Description

    The HK SubOKath Subcutaneous Catheter is a device intended to provide a means of delivery of local analgesics into the human body via a fenestrated catheter. The device is a single lumen catheter available in a single gauge (16g) size. The distal end of the device allows for flow of analgesia into the surgical wound. The device is supplied sterile and is intended for single use.

    AI/ML Overview

    This document describes the 510(k) premarket notification for the SubQKath Catheter and Needle Set (K193664). The acceptance criteria and study proving the device meets these criteria are primarily based on non-clinical performance data demonstrating substantial equivalence to predicate devices, rather than a clinical study involving human patients or complex AI algorithm performance.

    Therefore, many of the typical questions for AI/ML-based medical devices (like sample size for test/training sets, expert consensus, MRMC studies, or standalone algorithm performance) are not applicable to this submission. This device is a physical medical device, and its safety and effectiveness are established through engineering testing and comparison to existing devices.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly list acceptance criteria in a quantitative table format for each test, but it states that "The subject device met the acceptance criteria of the above tests, and did not raise new questions of safety and effectiveness." The study design is focused on demonstrating physical and material properties, rather than diagnostic accuracy metrics.

    Acceptance Criteria (Implied)Reported Device Performance
    Pass Physical Properties VerificationMet acceptance criteria (details not specified)
    Pass Bond and Material Strength VerificationMet acceptance criteria (details not specified)
    Pass Needle Quality (Sharpness after 1-year aging)Met acceptance criteria (details not specified)
    Pass Needle Quality (Stiffness after 1-year aging per ISO 9626)Met acceptance criteria (details not specified)
    Pass Kink resistanceMet acceptance criteria (details not specified)
    Pass Flow rate characterizationPassed; "differences in fenestration pattern did not affect flow characteristics, as demonstrated by flow performance testing."
    Pass Liquid leakageMet acceptance criteria (details not specified)
    Pass Functional VerificationMet acceptance criteria (details not specified)
    Pass Biocompatibility (Cytotoxicity, Sensitization, Intracutaneous, Systemic Irritation, Toxicity, Genotoxicity, Implantation)Evaluated with a test battery addressing suggested aspects of biocompatibility per ISO 10993-1:2018; met acceptance criteria.
    Pass Sterilization Validation (Ethylene Oxide to 10-6 SAL)Successfully executed using overkill approach as referenced in ANSI/AAMI/ISO 11135.
    Pass Sterile Barrier PerformanceSuccessfully executed to demonstrate compliance with applicable elements of ISO 11607.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated in terms of a "test set" as this is a physical device and testing involves material and functional evaluations, not typically patient data sets. The sample sizes for the various engineering and biocompatibility tests would be dependent on the specific test protocols (e.g., number of catheters tested for flow rate, number of samples for material strength). These details are not provided in this summary.
    • Data Provenance: Not applicable in the context of clinical data. The data provenance refers to the results of laboratory-based physical, chemical, and biological performance testing of the device itself. This would be generated from the manufacturer's testing facilities or contract labs.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Number of Experts: Not applicable. Ground truth for a physical device is established through validated testing methodologies and adherence to standards (e.g., ISO, ANSI/AAMI). There's no "ground truth" derived from expert consensus on medical images or clinical outcomes in this type of submission.
    • Qualifications of Experts: Not applicable in the context of clinical readers. The "experts" would be engineers, material scientists, and toxicologists conducting the specified tests and interpreting the results according to relevant standards.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. This is not a study requiring reader adjudication of medical data.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • MRMC Study: No. This is not an AI/ML diagnostic device, nor is it a device that would typically undergo a human-reader comparative effectiveness study. It's a physical catheter.
    • Effect Size of Human Readers' Improvement: Not applicable.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Standalone Performance: Not applicable. This device does not involve an algorithm.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The "ground truth" for this device is based on engineering specifications, material science standards (e.g., ISO 9626, ISO 11607, ISO 10993-1), and functional performance requirements of similar, legally marketed predicate devices. For biocompatibility, it's the biological response measured against established safety thresholds.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. This device does not involve AI/ML and therefore does not have a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set Establishment: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200624

    Validate with FDA (Live)

    Date Cleared
    2020-04-27

    (48 days)

    Product Code
    Regulation Number
    868.5120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Minnesota 55114

    Re: K200624

    Trade/Device Name: Percutaneous Introducer Regulation Number: 21 CFR 868.5120
    |
    | Classification Names: | Anesthesia, Conduction Catheter,BSO (21 CFR 868.5120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Percutaneous Introducer is intended to allow for the percutaneous placement of devices, needles or probes, in close proximity to nerves and around or into surgical wound or non-surgical wound sites. It may be used to inject or aspirate the introduction area via the luer hub of the needle/introducer.

    Device Description

    The Percutaneous Introducer is a single-use, disposable device consisting of a flexible polymer sheath with a molded plastic hub. The device incorporates a stainless steel needle inside the flexible cannula and is removed after the introducer is placed. The inserted stainless steel introducer needle has a sharp distal point. The flexible cannula acts as a conduit for percutaneous introduction of medical devices, such as catheters, needles, or probes. A stylet or porous plug (to prevent aspiration during insertion) comes seated in the standard 6 degree luer fitting of the proximal end of the needle hub. The Percutaneous Introducers are used for a short period of time and are supplied sterile and non-pyrogenic. The Percutaneous Introducer is available in four lengths and a 14 gauge.

    AI/ML Overview

    This document is a 510(k) summary for a "Percutaneous Introducer" device. It outlines the device description, intended use, technological characteristics, and non-clinical testing performed to demonstrate substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" with quantitative targets and corresponding "reported device performance" in the way a clinical study might. Instead, it describes various tests performed and states that the device "meets the requirements" or "performs comparably" or has "successful results."

    The tests and their reported outcomes are:

    Acceptance Criteria (Implicit from tests)Reported Device Performance
    Mechanical Performance:
    Penetration Force TestPerforms comparably to predicate device (K102460)
    Simulated Use Penetration/ Withdrawal/ Accordion/ Peelback TestPerforms comparably to predicate device (K102460); shown to verify resistance to collapsing and accordion/ peelback during tissue penetration.
    Tensile StrengthPerforms comparably to predicate device (K102460); proven to possess the ability to withstand a minimum acceptable tensile force.
    Material Biocompatibility:
    ISO 10993 compliant biocompatibilityMeets necessary biocompatibility requirements as per FDA guidance. Materials used support Substantial Equivalence.
    Cytotoxicity (ISO10993-5:2009)Tested with successful results.
    Sensitization (ISO10993-10:2010)Tested with successful results.
    Irritation/ Intracutaneous Reactivity (ISO10993-10:2010)Tested with successful results.
    EtO Residuals (I-CHM-2136 Rev. 0)Tested with successful results.
    Pyrogenicity (ISO10993-12:2017)Tested with successful results.
    Rabbit Blood Hemolysis (ISO10993-12:2017)Tested with successful results.
    Sterility & Packaging:
    Packaging- SterilityTested with successful results.
    Burst Testing (ASTM F 1140)Tested with successful results.
    Dye Penetration (ASTM F 1929)Tested with successful results.
    Functional Standards:
    ISO 594-1 (Luer Fittings)Tested with successful results.
    ISO 594-2 (Luer Fittings)Tested with successful results.
    ISO 10555-5 (Sterile, single-use intravascular catheters - introducers and dilators)Tested with successful results.

    The study concludes that based on these results, the Percutaneous Introducer "performs comparably to, and is substantially equivalent to predicate device (K102460)."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for any of the non-clinical tests (e.g., number of introducers tested for tensile strength, burst testing, etc.).
    The data provenance is from non-clinical bench and laboratory testing, not human or animal studies. The data would originate from the manufacturing and testing facilities of Epimed International, Inc. or their contracted labs. It is inherently prospective in the sense that the tests were conducted specifically for this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This section is not applicable in the traditional sense of clinical or diagnostic device ground truth. The "ground truth" for non-clinical performance and biocompatibility tests is established by adherence to recognized international standards (e.g., ISO, ASTM) and FDA guidance documents. The "experts" would be the qualified personnel within the testing laboratories who perform the tests according to these standards, generate the data, and interpret the results to ensure compliance. Their specific number or qualifications are not detailed here, as the focus is on adherence to standardized testing protocols.

    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving multiple human readers for diagnostic image interpretation or similar subjective assessments. The tests described are objective, standardized laboratory and bench tests where results are determined by measurement against specified criteria in the standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret results, and the AI's impact on their performance is being evaluated. This submission is for an interventional medical device (introducer) and focuses on its physical performance, material safety, and substantial equivalence, not diagnostic accuracy or human reader performance.

    6. Standalone (Algorithm Only) Performance

    No, a standalone (algorithm only) performance assessment was not done because the device is a physical medical instrument (a Percutaneous Introducer), not a software algorithm or AI-driven diagnostic tool. The concept of "algorithm only performance" does not apply to this type of device.

    7. Type of Ground Truth Used

    The "ground truth" for the non-clinical tests is based on established international standards and regulatory guidelines. For example:

    • Mechanical properties: Defined by performance criteria within ASTM and ISO standards (e.g., specific force values, absence of collapse/peelback).
    • Biocompatibility: Defined by the requirements and endpoints outlined in ISO 10993 series of standards (e.g., non-cytotoxic, non-sensitizing, non-irritating).
    • Sterility and Packaging: Defined by standards like ASTM F1140, ASTM F1929, and general sterility assurance principles.
    • Functional standards: Defined by the specific requirements of ISO 594-1, ISO 594-2, and ISO 10555-5.

    Essentially, the "ground truth" is compliance with these widely accepted benchmarks for medical device safety and performance.

    8. Sample Size for the Training Set

    Not applicable. This device is a physical medical instrument, not an AI or machine learning model. Therefore, there is no "training set" in the context of an algorithm. The development and testing process involves engineering design, material selection, and verification/validation testing, not data-driven machine learning training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191470

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-10-25

    (144 days)

    Product Code
    Regulation Number
    868.5120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Iowa 52556

    Re: K191470

    Trade/Device Name: C-CAT Anaesthesia Catheter kit Regulation Number: 21 CFR 868.5120
    Kit including models: C44-21-005

    C44-21-010 C44-21-015 C44-21-018 C44-21-020

    Anesthesia catheter

    868.5120
    020 | 868.5120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CIMPAX C-CAT Anesthesia Catheter, along with related accessory devices, is intended for the delivery of continuous or intermittent delivery of medication (such as local anesthetics or narcotics) to or around surgical wound sites for preoperative and postoperative pain management. Routes of administration may be intraoperative or percutaneous.

    Device Description

    The CIMPAX® C-CAT catheter kit is a sterile, single use, closed tip, multi-perforated catheter with a specified fenestrated length that incoporates micro-holes placed in a spiral pattern to provide 360° drug distribution. Additonally the CIMPAX® C-CAT is radiopaque and has four graduated distance markings along the catheter length for positional reference. The kit is intended for adult patients requiring the delivery of contiguous or intermittent delivery of medication (such as local anesthetics or narcotics) to or around surgical wound sites for preoperative and postoperative pain management or as perineural delivery. Routes of administration may be intraoperative or percutaneous. The device is intended to be used in healthcare /hospital facilities.

    The CIMPAX® C-CAT catheter is a multi-perforated catheter supplied in a kit which contains components for catheter placement. The drug infusion segment is placed between the blue tip of the catheter and the first blue marking after the tip. The catheter is designed with a specified fenestrated length that incoporates micro-holes placed in a spiral pattern to provide 360° drug distribution. Additonally the CIMPAX® C-CAT is radiopaque and has four graduated distance markings along the catheter length for positional reference. Attached to the catheter is a standard female luer lock that is compatible with and can be used with infusion pumps as well as with syringes for bolus injections with standard make luer connections.

    The catheter is placed usually through an incision or wound in order to provide continuous or intermittent administration of local anesthetic to the area affected. This allows only the local nerve endings (pain receptors) to be anesthetized as opposed to giving a systemic pain relief medication to the patient.

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA, specifically concerning the CIMPAX C-CAT Anaesthesia Catheter kit. This document is a regulatory submission for medical devices, demonstrating substantial equivalence to a legally marketed predicate device, rather than a study proving the device meets specific acceptance criteria for an AI/ML product.

    Therefore, the document does not contain any information related to:

    • Acceptance criteria for an AI/ML device (e.g., performance metrics like accuracy, sensitivity, specificity, AUC).
    • A study proving an AI/ML device meets acceptance criteria.
    • Sample sizes for test sets or data provenance for AI/ML model evaluation.
    • Number of experts, qualifications of experts, or adjudication methods for establishing ground truth in the context of AI/ML.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness studies.
    • Standalone AI algorithm performance.
    • Type of ground truth used for AI/ML model evaluation (e.g., pathology, outcomes data).
    • Sample size for training sets or how ground truth was established for training sets in an AI/ML context.

    The document discusses acceptance criteria and studies, but these relate to the physical and biological safety and performance of a medical catheter, not an AI/ML device. For example, it mentions:

    • Non-Clinical Testing: Bench testing including mechanical and performance testing (e.g., peak tensile force, burst pressure, kink testing, liquid leakage under pressure, radiopacity, flow testing), biocompatibility, sterilization, packaging, and shelf-life testing.
    • Biocompatibility Testing: According to ISO 10993-1 series, covering cytotoxicity, sensitization, irritation, systemic toxicity, Ames Test, in vitro mammalian chromosome aberration, hemolytic properties, and material mediated rabbit pyrogen.
    • Flow Test: Performed according to DS/EN 1618, measuring water flow through the catheter.
    • Tensile strength: Tested in accordance with ISO 10555-1.
    • Packaging: Complies with ISO 11607-1:2014 and EN 865-5:2009.
    • Stability: Tested according to ASTM F 1980 Accelerated Aging and ISO 11607-1:2009.
    • Sterilization: Method is Ethylene Oxide (EO), complying with EN ISO 11135-1.

    In summary, the provided text describes the regulatory process for a physical medical device (an anesthesia catheter) and outlines its non-clinical performance and safety testing to demonstrate substantial equivalence, not information relevant to the performance and acceptance of an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181782

    Validate with FDA (Live)

    Date Cleared
    2019-03-04

    (244 days)

    Product Code
    Regulation Number
    868.5120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    60093

    Re: K181782

    Trade/Device Name: Medline Reinforced Epidural Catheter Regulation Number: 21 CFR 868.5120
    IL 60093

    Regulatory Class: Class II Regulation Number: 21 CFR 868.5120

    Predicate Device

    Predicate
    use up to 72 hours. | Same |
    | Regulation Number | 21 CFR 868.5120
    | 21 CFR 868.5120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medline Reinforced Epidural Catheter is indicated for the injection of local anesthetics into the epidural space for patients over 22lbs. (10kg), for a duration of use up to 72 hours.

    Device Description

    The Medline Reinforced Epidural Catheter is a single use device made of flexible polyurethane elastomer tubing. The subject device is sold sterile, and will be packaged inside of a convenience kit alongside a variety of other components that are outside the scope of this 510(k) submission. The proposed epidural catheter is designed as a closed tip device with two eyelets to provide for the dispersion of anesthetic agents. The catheter features a 19G design and is 913 millimeters (mm) long. The catheter has a marked tip, with 10mm increments up the length of the device (up to 913mm). The 100mm mark is indicated by two marks, 150mm by three marks, 200mm by 4 marks, and 250mm by 5 marks. In addition, the Medline Reinforced Epidural Catheter has an internal radiopaque inner spring that is evenly distributed inside the device, terminating 2+/-0.05mm from the liquid outlet end. As an added safety feature, a solid wide warning mark is featured on the proposed device to indicate exit of the catheter from a needle when used in conjunction with an epidural needle.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device (Medline Reinforced Epidural Catheter) and as such focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. It does not include information typical of a study that would define acceptance criteria for AI/Algorithm-based device performance and then prove the device meets those criteria using clinical data and expert input.

    Therefore, many of the requested elements (1-9) regarding acceptance criteria, reported performance, sample sizes for test/training sets, data provenance, expert involvement, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for AI/algorithm performance are not applicable or not present in this document.

    However, I can extract the information provided regarding the non-clinical testing and the company's conclusion of substantial equivalence.

    Summary of Non-Clinical Testing and Conclusion of Substantial Equivalence:

    The Medline Reinforced Epidural Catheter is seeking 510(k) clearance based on substantial equivalence to the ARROW® FlexTip Plus® Epidural Catheter (K140110). Instead of defining acceptance criteria for algorithmic performance, the submission focuses on demonstrating safety, performance, and functionality through various non-clinical tests.

    No information provided in the document refers to an AI/algorithm-based device, nor does it contain data on acceptance criteria, reported device performance (in the context of AI), sample sizes for test/training sets, data provenance, expert involvement, adjudication methods, MRMC studies, standalone performance, or ground truth establishment relevant to AI/algorithm evaluation.

    The document describes non-clinical testing for a physical medical device. Here's a summary of the non-clinical tests conducted:

    1. Table of Acceptance Criteria and Reported Device Performance:

    This document does not define quantitative acceptance criteria for algorithm performance or report such performance metrics. Instead, it lists various non-clinical tests conducted to demonstrate the safety and functionality of the physical catheter. The "reported device performance" is implied by the successful completion of these tests in accordance with relevant standards.

    Test CategorySpecific TestAcceptance Standard/MethodReported Performance
    Chemical SafetyParticulate Matter TestingUSP <788> Particulate Matter InjectionsResults demonstrated overall safety and effectiveness (as stated in the submission). Specific particle counts or thresholds are not provided in this summary.
    Limulus Amebocyte Lysate (LAL) Bacterial Endotoxin TestingUSP <85> Bacterial Endotoxin TestingResults demonstrated overall safety and effectiveness. Specific endotoxin levels or thresholds are not provided.
    Functional PerformanceKink TestingNot explicitly stated (implied internal protocol or recognized standard)Results demonstrated overall safety and effectiveness, supporting substantial equivalence. Specific kink resistance values or criteria are not provided.
    Dimensional AnalysisNot explicitly stated (implied internal protocol)Results demonstrated overall safety and effectiveness. Specific dimensional measurements or tolerances are not provided.
    Simulated Distribution TestingISTA 3A-2018 General Simulation Performance Test Procedure for Packaged-Products for Parcel Delivery System Shipment 70 kg (150 lb) or LessResults demonstrated the ability to maintain functionality upon distribution. Specific test outcomes (e.g., integrity after simulation) are not detailed here beyond a general statement of effectiveness.
    Transportation TestingASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and SystemsResults demonstrated the ability to maintain functionality during transportation. Specific test outcomes are not detailed here.
    Visual Appearance TestingNot explicitly stated (implied internal protocol)Results demonstrated overall safety and effectiveness, presumably meeting visual inspection criteria.
    Ink Adherence TestingNot explicitly stated (implied internal protocol)Results demonstrated overall safety and effectiveness, indicating the markings adhere properly.
    Tensile Strength TestingASTM D412-16 Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers—TensionResults demonstrated overall safety and effectiveness. Specific tensile strength values are not provided.
    Flow Rate TestingASTM F623-99 Standard Performance Specification for Foley Catheter (Flow Rate) Note: While a Foley catheter standard is cited, it's used for flow rate testing of an epidural catheter, presumably for catheter-specific flow requirements.Results demonstrated overall safety and effectiveness. Specific flow rates or thresholds are not provided.
    Connector Pull Out TestingNot explicitly stated (implied internal protocol)Results demonstrated overall safety and effectiveness. Specific pull-out forces are not provided.
    Leakage TestingISO 80369-6 Small-bore connectors for liquids and gases in healthcare applications – Part 6: Connectors for neuraxial applicationsResults demonstrated overall safety and effectiveness. Specific leakage rates or criteria are not provided.
    Packaging Integrity TestingASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye PenetrationResults demonstrated overall safety and effectiveness, indicating packaging integrity.
    Seal Strength TestingASTM F88-16 Standard Test Method for Seal Strength of Flexible Barrier MaterialsResults demonstrated overall safety and effectiveness, indicating appropriate seal strength.
    X-Ray Opaque Detection TestingNot explicitly stated (implied internal protocol)Results demonstrated overall safety and effectiveness, presumably confirming the radiopaque properties for visibility under X-ray.
    BiocompatibilityCytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Material-Mediated Pyrogenicity, Hemolysis, Subacute Systemic Toxicity, Chemical CharacterizationANSI/AAMI/ISO 10993-1:2009 & specific sub-parts of ISO 10993 (e.g., -5, -10, -11, -4, -18) and USP <151>The tests were conducted and the protocols/reports are available for FDA review (Appendix E). The overall conclusion is that the device's biocompatibility supports substantial equivalence. Specific results or pass/fail criteria from these tests are not
    detailed in this summary document.
    SterilizationSterilization ValidationISO 11135:2014 Sterilization of Health Care Products - Ethylene Oxide...Demonstrated a Sterility Assurance Level (SAL) of 1 x 10^-6.
    EO and ECH ResidualsISO 10993-7:2008(R)2012 Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization ResidualsThe device was evaluated, and presumably, residuals were within acceptable limits as part of the overall safety determination. Specific residual levels are not provided.
    Shelf LifeAccelerated Aging StudiesASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical DevicesVerified a one-year shelf life, ensuring functionality and sterility are maintained.

    Regarding your specific questions:

    1. A table of acceptance criteria and the reported device performance: See the table above. Note that these are for a physical device, not an AI/algorithm.
    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable for an AI/algorithm. For the non-clinical tests, sample sizes would be defined by the selected standards and internal protocols, but are not specified in this summary. Data provenance is not specified.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable as this is not an AI/algorithm study requiring ground truth derived from expert consensus.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for non-clinical testing of a physical device.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No, this is not an AI/algorithm device submission.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: No, this is not an AI/algorithm device submission.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable as this is not an AI/algorithm study. The "ground truth" for a physical device is its ability to meet engineering specifications and safety standards through physical and chemical testing.
    8. The sample size for the training set: Not applicable; there is no training set mentioned, as this is not an AI/algorithm device.
    9. How the ground truth for the training set was established: Not applicable; there is no training set mentioned, as this is not an AI/algorithm device.

    Conclusion stated in the document:
    Based on the non-clinical testing, Medline Industries, Inc. concludes that the Medline Reinforced Epidural Catheter is substantially equivalent to the predicate device, ARROW® FlexTip Plus® Epidural Catheters (K140110).

    Ask a Question

    Ask a specific question about this device

    K Number
    K173541

    Validate with FDA (Live)

    Device Name
    EPINAUT
    Manufacturer
    Date Cleared
    2018-08-24

    (282 days)

    Product Code
    Regulation Number
    868.5120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Roosevelt, Irvine, California 92620

    Re: K173541

    Trade/Device Name: EPINAUT Regulation Number: 21 CFR 868.5120
    Name: Anesthesia conduction catheter

    • Product Code: BSO
    • Classification regulation: Class II, 21 CFR 868.5120
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For administration of anesthetic agents into the epidural space to provide epidural or caudal anesthesia.

    Device Description

    EPINAUT Spring Guide Epidural Catheter is constructed of a stainless steel continuous spring. EPINUAT Spring Guide Epidural Catheter offers two types of catheter. The catheter types, L330 & L600 have an uncoated distal tip which is flexible, smooth and rounded with coils slightly spread for maximum flexibility and lateral distribution of injectant. The other catheter types, C310 & C600 have a distal end which is entirely coated by tube for maximum smoothness. There are 16 models in EPINAUT: L33011320, L33013820, L33011315, L33013815, L60011320, L60013820, L60011315, L60013815, C31011320, C31013820, C31011315, C31013815, C60011320, C60013820, C60011315 and C60013815. Each model is characterized by different length of catheter, length and outer diameter of wire, inner diameter of adapter, length and diameter of introducer needle and needle cap.

    EPINUAT Spring Guided Epidural Catheter have enhanced tensile/ break strength and restrict longitudinal catheter stretch, while the spring coils make the catheter resistant to kinking and collapsing. Wire is also provided with the device which consists of a stainless steel and a molded plastic hub.

    The introducer needle is inserted into the epidural space through the sacral hiatus under intermittent fluoroscopy. After appropriate determination of the epidurogram and target area, the catheter with the wire in it is advanced through the introducer needle to the concerned area as determined by MRI or symptomatology. When the catheter is approached to target site, the wire is removed from the catheter and then the medication can be injected by attaching the adapter to a syringe.

    AI/ML Overview

    The medical device in question, EPINAUT, is an Anesthesia Conduction Catheter. The provided context does not include information about clinical studies involving human patients or AI-based performance metrics. The information focuses on non-clinical testing to establish substantial equivalence to a predicate device.

    Here's an analysis of the provided text in relation to the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes various non-clinical tests performed to evaluate the performance of EPINAUT and to demonstrate substantial equivalence to the predicate device, the Racz Epidural Catheter. The acceptance criteria are implicitly met if the test results "met the pre-set criteria" or "supported that the subject device is substantially equivalent." Specific numerical acceptance criteria are not explicitly stated for all tests; rather, the successful completion of the test according to a standard often serves as the acceptance.

    No.TestAcceptance Criteria (Implicit)Reported Device Performance
    1Visual Inspection (Appearance)Device appears as intended, free from defects.Test performed, results met pre-set criteria.
    2Size MeasurementDimensions conform to specifications.Test performed, results met pre-set criteria.
    3Leakage Test (Catheter)No leakage.Test performed, results met pre-set criteria.
    4Tensile strength (Catheter)Meets specified tensile strength (e.g., ISO 10555-1 Annex B for catheter, BS 6196:1989 Appendix D for catheter-hub/Stylet/Needle Stylet).Test performed, results met pre-set criteria. (Also see side-by-side testing results below).
    5Assemble compatibility (Catheter)Components fit together correctly.Test performed, results met pre-set criteria.
    6Radiopacity testDevice is visible under fluoroscopy.Test performed, results met pre-set criteria.
    7Corrosion resistance testNo significant corrosion.Test performed, results met pre-set criteria.
    8Flexural Rigidity (Catheter introducer and needle)Meets specified flexibility requirements.Test performed, results met pre-set criteria.
    9Pulling (Drawing) Test (Catheter introducer and needle)Meets specified pulling force requirements.Test performed, results met pre-set criteria.
    10Flexural Rigidity (Wire)Meets specified flexibility requirements.Test performed, results met pre-set criteria.
    11EO Gas residueResidue levels below accepted limits.Test performed, results met pre-set criteria.
    12Sterility TestSterility Assurance Level (SAL) of 10^-6 (in accordance with ISO 11135 and ISO 11737-1, 2).Test performed, results met pre-set criteria.
    13Endotoxin TestEndotoxin levels below specified limits.Test performed, results met pre-set criteria.
    14Particulate Matter TestParticulate matter levels below specified limits.Test performed, results met pre-set criteria.
    Side-by-Side Comparative Tests with Predicate Device:
    1Tensile strength (Catheter, Catheter-Hub, Stylet, Needle Stylet)Performance comparable to predicate device, meeting relevant standards (ISO 10555-1 Annex B, BS 6196:1989 Appendix D for various components, BS 6196:1989 Appendix F for Needle Stylet)."The test results supported that the subject device is substantially equivalent to the predicate devices." Implies comparable or better performance.
    2Dullness testPerformance comparable to predicate device, meeting BS 6196:1989 Appendix G."The test results supported that the subject device is substantially equivalent to the predicate devices." Implies comparable or better performance.
    3Fluid pressure of the catheter/adapter connectionPerformance comparable to predicate device, meeting ISO 10555-1 Annex G."The test results supported that the subject device is substantially equivalent to the predicate devices." Implies comparable or better performance.
    4Flow ratePerformance comparable to predicate device, meeting ISO 10555-1 Annex E."The test results supported that the subject device is substantially equivalent to the predicate devices." Implies comparable or better performance.
    5Luminal integrityPerformance comparable to predicate device, meeting ISO 10555-1 Annex C."The test results supported that the subject device is substantially equivalent to the predicate devices." Implies comparable or better performance.
    Biocompatibility:Biocompatible as per ISO 10993-4, 5, 10, 11 and USP 39 <151> for External communicating device (tissue/bone/dentin communicating), Limited exposure (<24 hours)."Biocompatibility tests were performed... and the test results supported that the subject devices are biocompatible."
    Shelf Life:Validated 3-year shelf life using accelerated aging (ASTM F1980)."The test results validated 3 year shelf life."

    2. Sample size used for the test set and the data provenance:

    The document describes non-clinical bench testing. It does not refer to "test sets" in the context of clinical data, human subjects, or AI performance evaluation. Therefore, information on sample sizes for a clinical test set or data provenance (country of origin, retrospective/prospective) is not applicable or provided. The tests mentioned adhere to international standards (e.g., ISO, ASTM, BS) which typically specify the number of samples required for each test.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable. The device is an Anesthesia Conduction Catheter, and its performance evaluation relies on engineering, material, and biocompatibility testing according to established standards, not on expert-adjudicated ground truth from clinical image interpretation or similar AI applications.

    4. Adjudication method for the test set:

    Not applicable, as there is no "test set" in the context of clinical data requiring expert adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. The EPINAUT device is an anesthesia conduction catheter, a physical medical device, not an AI software. There is no mention of AI assistance or MRMC studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used:

    For non-clinical performance tests, the "ground truth" is defined by the requirements of the relevant international standards (e.g., ISO, ASTM, BS) and the pre-set internal specifications of the manufacturer. For biocompatibility, the ground truth is established by the results of tests performed according to ISO 10993 and USP standards, demonstrating the material not to be toxic or irritating. For sterilization, the ground truth is a Sterility Assurance Level (SAL) of 10^-6 demonstrated through validated methods according to ISO 11135 and ISO 11737.

    8. The sample size for the training set:

    Not applicable. This is a physical medical device, not an AI system that requires a "training set."

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172777

    Validate with FDA (Live)

    Date Cleared
    2018-05-03

    (231 days)

    Product Code
    Regulation Number
    868.5120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Adapter NRFit (ISO80369-6) and Catheter Clamping Adapter LUER (ISO80369-7) Regulation Number: 21 CFR 868.5120
    |
    | Classification Reference: | 21 CFR 868.5120
    Adapter is considered a Class II medical device according to the predicate device as defined in 21 CFR §868.5120

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Catheter Clamping Adapter, a connection device, is used to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics.

    The connector is used in conjunction with catheters for continuous administration of anesthetic agents.

    Device Description

    The Catheter Clamping Adapter with either NRFit™ Connector according to ISO80369-6 or LUER-Connector according to ISO80369-7. The device is intended to provide a specified connection (either NRFit™ Connector according to ISO80369-6 or LUER-Connector according to ISO80369-7) to an unspecified proximal end of a 20G anesthesia conduction catheter (epidural or peripheral). Therefore it may also be considered an accessory for epidural and peripheral anesthesia conduction catheters. The device is a sterile finished disposable device, supplied sterile to the end user and non-sterile intended to be sterilized prior to use to re-packagers/ medical device manufacturers.

    AI/ML Overview

    The PAJUNK Catheter Clamping Adapter NRFit (ISO80369-6) and Catheter Clamping Adapter LUER (ISO80369-7) have been determined to be substantially equivalent to a legally marketed predicate device (BBraun Perifix Catheter Connector, K022019) based on a detailed comparison and performance testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Test Detail)StandardRequired Performance (Acceptance Criteria)Reported Device PerformanceResult
    SterilizationISO 11135SAL of 10⁻⁶SAL of 10⁻⁶Pass
    Residuals (EO and ECH)ISO 10993-7Compliance with limitsLimits metPass
    BiocompatibilityISO 10993-1Evaluated (external communicating device)Evaluated and considered biocompatiblePass
    Liquid Leakage (ISO 80369-7)ISO 80369-7, 6.1PassPassPass
    Air Leakage (ISO 80369-7)ISO 80369-7, 6.2PassPassPass
    Stress Cracking (ISO 80369-7)ISO 80369-7, 6.3PassPassPass
    Axial Load (ISO 80369-7)ISO 80369-7, 6.4PassPassPass
    Unscrewing torque (ISO 80369-7)ISO 80369-7, 6.5PassPassPass
    Overriding (ISO 80369-7)ISO 80369-7, 6.6PassPassPass
    Liquid Leakage (ISO 80369-6)ISO 80369-6, 6.1PassPassPass
    Air Leakage (ISO 80369-6)ISO 80369-6, 6.2PassPassPass
    Stress Cracking (ISO 80369-6)ISO 80369-6, 6.3PassPassPass
    Axial Load (ISO 80369-6)ISO 80369-6, 6.4PassPassPass
    Unscrewing torque (ISO 80369-6)ISO 80369-6, 6.5PassPassPass
    Overriding (ISO 80369-6)ISO 80369-6, 6.6PassPassPass
    Leak TightnessInternal ProtocolLeak tight under normal conditionsComplies with internal protocolPass
    Security of connectionInternal ProtocolSecure connectionComplies with internal protocolPass
    Tensile ForcePredicate device criterion> 8N> 8NPass
    Pyroburden/EndotoxinNot specifiedLimit metLimit metPass
    Shelf Life (5 years)Internal ProtocolNo decrease in performance after 5 years (real-time aged & accelerated aged)No decrease in performance after 5 yearsPass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each individual performance test (e.g., liquid leakage, axial load). It indicates that tests were conducted according to the specified ISO standards (ISO 80369-6, ISO 80369-7) and internal protocols. The standards themselves typically outline sample size requirements.

    The data provenance is not explicitly mentioned as "country of origin" or "retrospective/prospective." However, the testing was performed by PAJUNK® GmbH Medizintechnologie (based in Germany) and its contract sterilizer (Sterigenics Germany GmbH). The testing would be considered prospective for the purpose of demonstrating substantial equivalence for this submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This section is not applicable. The device is a medical connector, and its performance is evaluated against engineering and material standards (e.g., ISO for connectivity, leakage, tensile strength, biocompatibility), not against a diagnostic "ground truth" established by experts.

    4. Adjudication Method for the Test Set

    This section is not applicable for this type of device and testing. Performance is measured objectively against predefined engineering standards and specifications.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is used for diagnostic devices involving human interpretation of data (e.g., medical images), which is not relevant for this device.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance evaluation was conducted. The device's performance (e.g., leak tightness, tensile strength, connectivity) was tested directly against established physical and material standards without human intervention in its function. The "algorithm" in this context is the physical design and manufacturing process of the device, and its performance was evaluated inherently without human-in-the-loop.

    7. The Type of Ground Truth Used

    The "ground truth" used for this device's evaluation is based on established international and internal engineering standards and specifications. This includes:

    • ISO 80369-6 (for NRFit connectors)
    • ISO 80369-7 (for LUER connectors)
    • ISO 11135 (for sterilization)
    • ISO 10993-7 (for residuals)
    • ISO 10993-1 (for biocompatibility)
    • Internal protocols for leak tightness and security of connection.

    These standards define the objective, measurable criteria for acceptable device performance for safety and effectiveness.

    8. The Sample Size for the Training Set

    This section is not applicable. This device is a manufactured medical product, not an AI or algorithm-based system that requires a "training set" in the computational sense. The "development" and "refinement" of the product involve engineering design, material selection, and manufacturing processes, culminating in validation testing.

    9. How the Ground Truth for the Training Set was Established

    This section is not applicable for the reasons stated in point 8.