Search Results
Found 9 results
510(k) Data Aggregation
(277 days)
Einsteinstrasse 1 Sulz am Neckar, 72172 De
Re: K200190 Trade/Device Name: ScalpelCric Regulation Number: 21 CFR 868.5090
Emergencyairwayneedle. | Anesthesiology | Anesthesiology | BWC | 868.5090
| Regulation Number | 868.5090
| 868.5090
Emergency airway access via the cricothyroid membrane. Life-threatening dyspnoea that cannot be controlled in any other way.
Surgical Cricothyrotomy set for securing the airway in case of upper airway obstruction, respectively as "Ultima Ratio" if all other attempts to ventilate the patient failed. The set consists of scalpel, bougie, tube, extension tube with swivel connector, syringe and necktape.
The provided text is a 510(k) Premarket Notification for the ScalpelCric device, a cricothyrotomy set. It details the device's characteristics, intended use, and comparison to a predicate device to establish substantial equivalence.
However, this document does not contain information about a study that assesses the device's performance against specific acceptance criteria in the context of an AI/human-in-the-loop system, or a multi-reader multi-case (MRMC) study. The document focuses on the physical and biological characteristics of the medical device itself, its sterilization, biocompatibility, and bench testing, rather than a performance study involving diagnostic accuracy, expert review, or AI assistance.
Therefore, I cannot fulfill the request to describe the acceptance criteria and the study proving the device meets them in the context of an AI-assisted diagnostic system, as that information is not present in the provided text.
The closest information related to "acceptance criteria" is under section 9.4 Bench Testing, which lists features to be tested and the result (all "Passed"). This refers to the physical and functional performance of the device components, not a diagnostic or AI-related performance.
Here's what I can extract from the provided text regarding device testing, but it's important to note this is not a study proving AI performance as requested:
1. A table of acceptance criteria and the reported device performance:
The document provides a table under "9.4 Bench Testing" (page 10), which lists various tests performed on the components of the ScalpelCric set. The "Features to be tested" can be considered the acceptance criteria for those specific physical/functional aspects, and the "Result" indicates "Passed" for all.
| Object under Test | Features to be tested (Acceptance Criteria) | Reported Device Performance |
|---|---|---|
| Bonding Connector | Bonding between connector and tube | Passed |
| Bonding Inflation Line | Bonding between inflation line and tube | Passed |
| 15mm Connector | Dimensions of 15mm Connector | Passed |
| Luer Pilot Balloon | Dimensions of the female luer connector in the pilot balloon | Passed |
| Cuff | Function of the Cuff | Passed |
| Scalpel | Function and design of the scalpel | Passed |
| Bougie | Bougie which shall fulfill all relevant specification according [DI] and [MS_Bougie]. | Passed |
| Tube | Function and dimensions of the tube | Passed |
| Tube | Tube is recognizable in X-Ray images | Passed |
| Extension Tube | Function and dimension of the extension tube | Passed |
| Neck Tape | Neck tape which shall fulfill all relevant specification according [DI] and [MS_NT] after 4 times sterilization | Passed |
| Syringe | Function of the syringe | Passed |
| Tube with inflation line and cuff, | Function and features of the tube with attached inflation line and pilot balloon and with cuff | Passed |
| ScalpelCric | Biological compatibility | Passed |
| Material | Function and ingredient of the ScalpelCric materials | Passed |
| Tube and extension Tube | Oxygen resistance of Tube and Extension tube | Passed |
| Label | Device Label | Passed |
| Bougie | Device Label Bougie | Passed |
| IFU | IFU which shall contain all relevant information regarding [DI] and [ER]. | Passed |
| ScalpelCric | ScalpelCric Set can be used in Operating condition | Passed |
| ScalpelCric | Functionality after simulated transport | Passed |
| Transport Validation | Functionality after simulated transport | Passed |
| Safety | Verify implementation and effectiveness of all Risk Control Measures defined in the Risk Analysis | Passed |
| Suction Cap | Connecting tube with open suction cap during ventilation | Passed |
| User Needs | User Needs for ScalpelCric, defined in [SoP] | Passed |
| ScalpelCric | Usability | Passed |
| Headerbag | Requirements of DIN EN ISO 11607-1 and DIN EN ISO 11607-2. | Passed |
| Sterilization | Sterilization Requirements for ScalpelCric | Passed |
| Business Needs | Business Needs for ScalpelCric | Passed |
| Label Sterile Packaging | Sterile packaging label which shall contain all relevant information regarding the [SFL], [DI] and [ER]. | Passed |
| Label Outer Packaging | Outer packaging label which shall contain all relevant information regarding the [SFL], [DL] and [ER] | Passed |
| Packaging | Packaging Requirements for ScalpelCric | Passed |
| Accelerated Aging Syringe | Quality of the syringe after 5 years accelerated aging | Passed |
| Accelerated aging of Pouch | Function of the sterile packaging after 5 years accelerated aging | Passed |
| Bougie | Bougie shall fulfill all relevant specifications after 1 year accelerated aging | Passed |
| Scalpel | Function of the scalpel after accelerated aging 1 year | Passed |
| ScalpelCric | Scalpel Cric Set application oxygen resistance after 1 year accelerated aging | Passed |
| Surgicric | Function and dimension after 5 years accelerated aging of the Surgicric and the function of the sterile packaging. | Passed |
| Surgicric | are identical to the devices in ScalpelCric. | Passed |
| Surgicric | Function and dimension after 3 years real time aging of the Surgicric including Tubus and the function of the sterile packaging. The tested tube and sterile packaging are identical to the devices in ScalpelCric. | Passed |
| Syringe | Quality of the syringe after 3 years real time aging | Passed |
| Bougie | Bougie shall fulfill all relevant specifications after 5 years accelerated aging | Passed |
| Scalpel | Function of the scalpel after 5 years accelerated aging | Passed |
| ScalpelCric Set | Application oxygen resistance after 5 years accelerated aging | Passed |
| Cuff Tube Collapse included in the ScalpelCric after 5 years of accelerated aging. | Test required by DIN EN ISO 5361:2013-03, Annex C: Resistance of the VBM tracheal tube ventilation lumen collapse with cuff inflated inside a cylinder. | Passed |
The remaining requested information (sample size, data provenance, experts, adjudication, MRMC, standalone, ground truth type, training set details) is NOT available in the provided document, as it pertains to a different type of device evaluation (e.g., AI/diagnostic performance) than what is described here. This document is focused on the substantial equivalence of a physical medical device (a cricothyrotomy set) to a predicate device.
Ask a specific question about this device
(443 days)
contain or be labeled for use with guidewires and intubating catheters, in addition to needles (21 CFR 868.5090
device is comprised of the following components:
Introducer Needle (legally marketed under 21 CFR 868.5090
order to provide access for wire guide placement.
Catheter Needle (legally marketed under 21 CFR 868.5090
The Retrograde Intubation Set is intended to assist in intubation during difficult or emergency airway access procedures in adult and pediatric patients.
The 6 French catheter is recommended for use with a single-lumen endotracheal tube with an inner diameter of 2.5 mm or larger.
The 11 French catheter is recommended for use with a single-lumen endotracheal tube with an inner diameter of 4 mm or larger.
The 14 French catheter is recommended for use with a single-lumen endotracheal tube with an inner diameter of 5 mm or larger.
When used for high pressure oxygenation with a Luer Lock connector, the 14 French catheter is recommended for patients older than 12 years of age.
The basic components included in the subject device, the Retrograde Intubation Set, are an introducer needle, catheter needle, syringe, hemostat, wire guide with wire guide inserter, intubation catheter, and Rapi-Fit® adapter. While the wire guide, intubation catheter, and Rapi-Fit® adapters are the subject of this De Novo request, this device is intended to be used with legally marketed needles, syringes, and hemostats.
The needle and syringe are used to puncture the patient cricothyroid membrane and confirm needle placement in the patient airway. Then, a wire guide is inserted via the needle into a patient's trachea and exits via mouth or nose (i.e., retrograde). An intubating catheter is inserted orally or nasally over the wire guide into the trachea prior to intubation.
This device offers optional Rapi-Fit adapters (attached to the intubating catheter) to facilitate supplemental oxygen delivery in adults prior to intubation with an endotracheal tube (ETT).
Here's a breakdown of the acceptance criteria and the study information for the Retrograde Intubation Set, as extracted from the provided text:
Retrograde Intubation Set: Acceptance Criteria & Study Information
1. Table of Acceptance Criteria and Reported Device Performance
Given that this is a medical device for emergency intubation assistance, the "performance" is largely assessed through successful bench testing demonstrating functionality and adherence to safety standards. Unlike diagnostic AI models, there isn't a direct "accuracy" or "sensitivity/specificity" metric reported for the overall device. Instead, the acceptance criteria relate to specific physical properties and functionalities.
| Acceptance Criteria (Special Controls) | Reported Device Performance/Evidence |
|---|---|
| A. Wire guide tensile, flex, fracture, and corrosion testing | Corrosion Test: Visual signs of corrosion observed after 3 years, but Tensile Testing (post-corrosion): peak load preceding failure ≥ 10 N for each test article (acceptable performance despite corrosion). Fracture Testing: Tested for 3 complete turns (deviation from ISO 11070's 8 turns, justified by typical clinical use not exceeding a 90° bend). Tensile & Flex Testing: Methods and results acceptable (ISO 11070). |
| B. Catheter tensile strength testing at likely points of failure | 6 & 11 Fr Catheters: Demonstrated tensile strength greater than the 15 N acceptance criterion. 14 Fr Catheter: Met tensile strength acceptance criterion at catheter and sideport. Rapi-Fit Adapter Separation Force: Demonstrated separation forces greater than 15 N (mitigates risk of disconnection). (Acceptance criterion based on EN/ISO 10555-1:2013 and prior testing of intubation catheters, with 15 N being larger than likely clinical force). |
| C. Catheter kink radius testing | 6 & 11 Fr Catheters: Acceptance criteria based on airway anatomy (48.17 mm for adults, 19 mm for pediatrics) deemed clinically appropriate. Less emphasis on patency when kinked as these are not for oxygenation. 14 Fr Catheter (for oxygenation): Conducted according to Annex H of AAMI/ISO 5361:2012. All samples met stated minimally acceptable kink radius criteria. |
| D. Compatibility of device components that interact, including compatibility in connection, disconnection, and ability to transfer fluids | Rapi-Fit Adapter Testing (Connectivity): Luer Lock connector tested per ISO 594-2. 15 mm connector tested per ISO 5356-1:2015. Methods and results acceptable. Validation Testing on Retrograde Intubation (Manikin Study): Catheters supported intubation of child and adolescent models with ETTs from 2.5 mm to 6.5 mm. |
| E. Dimensional validation | Not explicitly detailed as a separate test, but implied by adherence to specifications (e.g., wire guide diameter/length, catheter diameter/length). |
| F. Accuracy testing of markings | 14 Fr Catheter Marking Accuracy: Labeled markings (every 5 ± 0.5 cm) and non-labeled increments (1 ± 0.2 cm) were measured. Max/min distance between labeled marks was 5.1 and 5.0 cm (within specifications). Non-labeled marks were 1.0 cm for all samples. Results acceptable. |
| G. Validation of the maximum airway pressure | High Pressure Oxygenation Testing: Tested using the Luer Lock connector (worst-case scenario). In the child model (> 2 years to 12 years), some test articles of the 14 Fr intubation introducer showed average maximum airway pressure > 28 cm H2O. Limitation: Oxygenation feature on 14 Fr intubation catheter with Luer Lock Rapi-Fit adapter not permissible in patients ≤ 12 years old due to high barotrauma risk. Remaining results acceptable for other populations. |
| Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life. | Shelf-Life Claim: (b)(4) years supported by 5-year accelerated aging (ASTM F1980) of finished device packages. Tests conducted: Visual inspection, seal strength (ASTM F88-09), dye penetration (ASTM F1929-12). Results acceptable. Sterilization Validation: Ethylene Oxide (EO) sterilization via overkill approach at SAL of 10^-6 (ISO 11135-1:2007). Worst-case sample (stylet within intubating catheter) used. Sterilant residuals acceptable (ISO10993-7). |
| The device must be demonstrated to be biocompatible. | Biocompatibility Evaluation: Performed a biological risk assessment in accordance with FDA guidance (ISO 10993-1). Evaluated per ISO 10993 (subparts) for limited duration contact (≤ 24 hours) for cytotoxicity, sensitization, irritation, acute systemic toxicity, and material-mediated pyrogen testing on patient-contacting components (catheters, wire guide, Rapi-Fit adapters). Methods for extractions acceptable. Mitigates biocompatibility concerns. |
| Labeling must include: A. Instructions for use; and B. Package labels that clearly identify the minimum compatible size of endotracheal tube. | Device labeling instructs users on proper procedures and specifies minimum compatible ETT sizes on the package label. This is a labeling requirement that the sponsor would need to demonstrate they meet by providing the actual labeling. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Test Set (Bench/Performance Studies): Specific sample sizes for each bench test (tensile, kink, marking accuracy, etc.) are not always explicitly stated as a single number. For example, "all samples tested" for marking accuracy, or "each test article" for wire guide testing. The studies were prospective bench tests performed in a laboratory setting by the manufacturer (Cook Incorporated). Country of origin is not explicitly stated for individual tests but the company is based in Bloomington, Indiana, USA.
- Clinical Evidence (Literature Review): The primary "clinical evidence" is a retrospective literature review, primarily relying on a publication by Sanchez in 1993 ("The retrograde cookbook"). This publication summarized 539 patient cases up to 1993, describing techniques, indications, contraindications, complications, and failures. The provenance is global (published literature).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Bench/Performance Studies: No "clinical experts" were used to establish ground truth for the bench tests. Ground truth was based on engineering specifications, recognized international standards (ISO, ASTM, EN), and clinical understanding of airway anatomy/forces.
- Clinical Evidence (Literature Review): The "ground truth" for clinical effectiveness and safety was established by published medical literature and case reports summarized by Sanchez (1993) and subsequent reviews. The "experts" in this context are the clinicians who performed the procedures and published their findings, compiled and analyzed by Sanchez. No specific number of experts or their qualifications for establishing ground truth are provided within this document, beyond the general acceptance of the medical literature.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Bench/Performance Studies: No adjudication method like 2+1 or 3+1 described. Performance was assessed against pre-defined engineering and safety criteria from recognized standards.
- Clinical Evidence (Literature Review): No formal adjudication method described for the literature review. It was a summary and analysis of existing published data.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No. This is a physical medical device, not an AI-powered diagnostic tool. Therefore, an MRMC comparative effectiveness study with AI assistance is not applicable and was not performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
No. This is a physical medical device. The "performance" is inherently based on human operators using the device according to standard techniques. There is no standalone algorithm or AI component.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Bench/Performance Studies: Ground truth was based on engineering specifications, recognized international standards (ISO, ASTM, EN), and the successful functional demonstration in manikin models against the intended use.
- Clinical Evidence (Literature Review): Ground truth was based on patient outcomes data and clinical experience reported in published medical literature (case reports and case series).
8. The sample size for the training set
Not applicable. This is a physical medical device undergoing engineering and clinical literature review for approval, not a machine learning model requiring a training set.
9. How the ground truth for the training set was established
Not applicable. As above, there is no training set for a machine learning model.
Ask a specific question about this device
(386 days)
- Surgical, Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set Regulation Number: 21 CFR 868.5090
Cricothyrotomy Catheter Set - Surgical Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set 21 CFR §868.5090
BWC2) 21 CFR §868.5800JOH | 1) 21 CFR §868.5090BWC
BWC2) 21 CFR §868.5800JOH | 1) 21 CFR §868.5090BWC
BWC2) 21 CFR §868.5800JOH | 1) 21 CFR §868.5090BWC
The Melker Cuffed Emergency Cricothyrotomy Catheter Set - Percutaneous is intended to establish emergency airway access when endotracheal intubation cannot be performed. Airway access is achieved utilizing the percutaneous entry (Seldinger) technique via the cricothyroid membrane.
The Melker Cuffed Emergency Cricothyrotomy Catheter Set - Surgical is intended to establish emergency arrway access when endotracheal intubation cannot be performed. Airway access is achieved utilizing the surgical technique via the cricothyroid membrane.
The Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set is intended to establish emergency arrway access when endotracheal intubation cannot be performed. Airway access is achieved utilizing either the percutaneous entry (Seldinger) technique or the surgical technique via the cricothyroid membrane.
The Melker Cuffed Emergency Cricothyrotomy Catheter Set - Percutaneous, Melker Cuffed Emergency Cricothyrotomy Catheter Set - Surgical, and the Melker Universal Cuffed Emergency Cricothyrotomy Catheter Set consist of an airway catheter, a loading dilator, a wire guide, a catheter needle, and an introducer needle. In addition to these components, standard preparation components, such as a syringe, scalpel, dilator forceps, gauze, drape, tracheostomy tape, and tracheal hook are included. The sets consist of any combination of components appropriate for either percutaneous or surgical placement of a cricothyrotomy tube, or both.
The curved cricothyrotomy catheter is manufactured as a 22 French catheter with a length of 9 centimeters and an inner diameter of 5 millimeters. The loading dilators are either 11 centimeters or 12 centimeters long. The stainless steel wire guide has a 0.038-inch diameter and a length of 40 centimeters. The catheter needle is a two-part needle comprised of a small hubbed catheter sheath covering a hubbed needle. The sheath of the catheter needle is manufactured from fluorinated ethylene propylene tubing. The catheter needle is manufactured using an 18 gage cannula with a length of 7 centimeters. The introducer needle is also manufactured from 18 gage stainless steel and has a length of either 5 centimeters or 7 centimeters.
The provided document describes a 510(k) premarket notification for the Melker Cuffed Emergency Cricothyrotomy Catheter Set and its variations. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel effectiveness through extensive clinical trials. Therefore, the information provided relates to design validation and performance testing to ensure the device is as safe and effective as its predicate, rather than a clinical study evaluating human-in-the-loop performance or diagnostic accuracy.
Here's an analysis of the acceptance criteria and the studies that prove the device meets them, based on the provided text:
1. A table of acceptance criteria and the reported device performance:
| Test Name | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Durability of the Cuff (Zero-Time, Three-Year Accelerated Aging) | Cuff must remain fully intact and without any leaks following insertion. | The predetermined acceptance criterion was met. |
| Tensile Test of the Shaft of the Dilator (Three-Year Real-Time) | Dilator shaft must withstand a minimum force at break of 15 N. | The predetermined acceptance criterion was met. |
| Magnetically Induced Displacement Force (Per ASTM F2052) | Displacement force shall be less than or equal to the worst-case displacement force due to Earth's gravity. | The predetermined acceptance criterion was met. |
| Magnetically Induced Torque (Per ASTM F2213) | Magnetically induced torque shall be less than or equal to the worst-case torque due to Earth's gravity. | The predetermined acceptance criterion was met. |
| Radiofrequency Induced Heating | Electrical conductivity of the Melker Emergency Cricothyrotomy Catheter must be < 1 S/m. | The predetermined acceptance criterion was met. |
| Magnetic Resonance Image Artifact (Per ASTM F2119) | (Implicitly, the images should demonstrate acceptable levels of artifact, but no specific numerical criteria are given in the text.) | Image artifact was assessed according to ASTM F2119. |
2. Sample size used for the test set and the data provenance:
The document does not specify the sample sizes (number of units tested) for any of these engineering or material tests. It also does not discuss data provenance in terms of country of origin or retrospective/prospective nature, as these are in vitro or in silico material and mechanical property tests, not human clinical studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not applicable to the types of tests described. These are objective engineering and material property tests, not based on expert interpretation of data.
4. Adjudication method for the test set:
This information is not applicable. The tests described have objective pass/fail criteria based on physical measurements.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No, an MRMC comparative effectiveness study was not done. This device is a medical instrument (catheter set), not an AI diagnostic or assistance tool. The submission is for a 510(k) premarket notification, which focuses on substantial equivalence based on technical characteristics and performance, not on demonstrating clinical improvement through reader studies.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
No, a standalone algorithm performance study was not done. As mentioned, this is a physical medical device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The ground truth for these tests is based on engineering standards and physical properties. For example:
- Durability and Tensile Strength: Ground truth is the physical integrity and mechanical strength of the materials as measured against quantitative thresholds (e.g., "minimum force at break of 15 N").
- MRI Compatibility: Ground truth is established by meeting specific ASTM standards (F2052, F2213, F2119) for magnetically induced forces, torque, and image artifact.
- Radiofrequency Heating: Ground truth is based on the measured electrical conductivity compared to a specified limit (< 1 S/m).
8. The sample size for the training set:
This information is not applicable, as this is not an AI/machine learning device that requires a training set.
9. How the ground truth for the training set was established:
This information is not applicable, as this is not an AI/machine learning device.
Ask a specific question about this device
(267 days)
Re: K153761
Trade/Device Name: Emergency Transtracheal Airway Catheter Regulation Number: 21 CFR 868.5090
set against a red background.
510(k) Summary
Emergency Transtracheal Airway Catheter 21 CFR §868.5090
:
Emergency Transtracheal Airway Catheter Tracheal Catheter Needle Emergency Airway Needle 21 CFR §868.5090
Emergency Transtracheal AirwayCatheter |
| Regulation | | 21 CFR §868.5090
The Emergency Transtracheal Airway Catheter is intended for emergency airway access when conventional endotracheal intubation cannot be performed.
The Emergency Transtracheal Airway Catheter is composed of a catheter and needle. The catheter-needle assembly cannulates the trachea through the cricothyroid membrane in order to establish an airway. The catheter is designed as a 6.0 French catheters manufactured from reinforced fluorinated ethylene propylene (FEP) tubing fitted to a 15 gage stainless steel needle with a lancet bevel. The nominal length of the catheter measures 5.0 or 7.5 centimeters, depending upon device specification. The catheter French size is stamped on the proximal fitting of the catheter. The Luer hub on the proximal end of the catheter connects to an oxygen source. The needle is manufactured with a nickel plated brass hub. The lancet bevel of the needle extends from the catheter at the beel of the bevel. The needle and the hub are soldered with tin-silver alloy material. The catheter-needle assembly cannulates the trachea through the cricothyroid membrane in order to establish an airway. The needle is removed from the catheter after the catheter-needle assembly establishes an airway through the cricothyroid membrane, leaving the catheter in the trachea.
The provided text describes specific acceptance criteria and the studies conducted to demonstrate that the Emergency Transtracheal Airway Catheter meets these criteria. This submission is for a Class II medical device, and the evaluation focuses on showing substantial equivalence to a predicate device, rather than a de novo effectiveness study requiring advanced statistical modeling like MRMC analyses for AI performance. Therefore, many of the requested elements pertaining to AI models, such as sample sizes for test/training sets, expert qualifications for ground truth, or MRMC studies, are not applicable to this particular device submission.
Here's a breakdown of the requested information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Design Validation (Zero Time): The test articles shall be able to access the trachea of an AirSim Advance Combo model when following the instructions provided in the Instructions for Use. | The acceptance criterion was met. |
| Bending Rigidity of the Needle (Zero Time): Peak load during compression within the specified deflection ≥ 7.5 N. | The acceptance criterion was met. |
| Insertion Force of the Needle (Zero Time): Determine the maximum compressive load. (Implicitly, this must be within acceptable clinical limits, though a specific numerical criterion is not provided explicitly in this summary). | The acceptance criterion was met. |
| Breaking Resistance (Zero Time): The cannula of the device withstands forces experienced during clinical use without breaking. | The predetermined acceptance criterion was met. |
| Air Leakage of the Catheter (Zero Time): Meet the air leakage requirements of BS EN ISO 10555-1. | The acceptance criterion was met. |
| Resistance to Overriding (Zero-Time): Meet the resistance to overriding requirements set forth in ISO 594-2. | The acceptance criterion was met. |
| Unscrewing Torque (Zero-Time): Meet the unscrewing torque requirements set forth in ISO 594-2. | The acceptance criterion was met. |
| Separation Force (Zero-Time): Meet the separation force requirements set forth in ISO 594-2. | The predetermined acceptance criterion was met. |
| Tensile Test of the Shaft of the Catheter (Three-year Accelerated Aging): Determine the peak tensile force of the shaft after accelerated aging to the real-time equivalent of 3 years. | The acceptance criterion was met. |
| Tensile Test of the Hub to Shaft Bond of the Catheter (Three-year Accelerated Aging): Determine the peak load of the hub-to-shaft bond after accelerated aging to the real-time equivalent of 3 years. | The acceptance criterion was met. |
| Tensile Test of the Hub to Cannula Bond of the Needle (Three-year Accelerated Aging): Needle cannula does not loosen from the hub when a tensile load of at least 20 N is applied (per ISO 11070) after accelerated aging. | The acceptance criterion was met. |
| Air Leakage of the Catheter (Three-year Accelerated Aging and Two-year Real Time Aging): Meet the air leakage requirements of BS EN ISO 10555-1 after accelerated aging to the real-time equivalent of 3 years. | The acceptance criterion was met. |
| Biocompatibility Testing: Meet appropriate requirements for cytotoxicity, sensitization, intracutaneous irritation, acute systemic toxicity, and material-mediated pyrogenicity (Per ISO 10993-1 and FDA guidance). | The predetermined acceptance criteria were met. |
| Five Year Accelerated Aging Report for Packaging Validation: Accelerated aging of the packaging system to the real-time equivalent of ≥ 5 years. | The acceptance criterion was met. |
| Sterility Testing: Meet appropriate requirements for bioburden, endotoxins, and EO and ECH residuals. | Testing confirms that the device meets the appropriate requirements for bioburden, endotoxins, and EO and ECH residuals. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify exact sample sizes for each test. For many of the performance tests (e.g., bending rigidity, insertion force, air leakage, tensile strength), standard engineering and quality control practices would involve testing a representative sample from a production lot. The "Zero Time" and "Three-year Accelerated Aging" indicate testing was performed on new devices and devices subjected to simulated aging, respectively. Data provenance is not described as involving human patient data, but rather simulated or laboratory testing of the physical device. The document does not indicate country of origin for the data, but the manufacturer (Cook Incorporated) is based in Bloomington, IN, USA. The studies are prospective in the sense that they were designed and executed to evaluate the device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This device is a physical medical instrument, not an AI/software device whose performance relies on established ground truth from experts interpreting images or data. The "ground truth" here is determined by engineering specifications and physical measurements against international standards.
4. Adjudication Method for the Test Set
Not applicable. As noted above, this is not an AI/software device requiring expert adjudication of results. Testing involved objective measurements against predefined engineering and material standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
Not applicable. This is not an AI-assisted device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
Not applicable. This is not an algorithm-based device. "Standalone" performance testing for this device refers to the physical device's performance against specifications without human interaction other than the execution of the test protocols (e.g., measuring forces, leakage, etc.).
7. The Type of Ground Truth Used
The ground truth used for this device is based on engineering specifications and international standards. For example:
- Design Validation: Successful access to a validated anatomical model (AirSim Advance Combo model).
- Mechanical Properties: Numerical thresholds for bending rigidity (e.5 N), tensile load (20 N), maximum compressive load, breaking resistance, unscrewing torque, and separation force.
- Fluid Dynamics: Air leakage requirements of BS EN ISO 10555-1.
- Biocompatibility: Conformance to ISO 10993-1 and FDA guidance.
- Sterility: Conformance to requirements for bioburden, endotoxins, and EO/ECH residuals.
- Aging: Performance after simulated accelerated aging for 3 years (device) and 5 years (packaging) equivalent real-time.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/software device that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As above, no training set for an AI model.
Ask a specific question about this device
(260 days)
Name: Emergency Cricothyrotomy Kit
Classification Name: Emergency Airway Needle (Class II per 21 CFR 868.5090
endotracheal intubationand ventilation cannot be performed |
| FDA Product Code | BWC 868.5090
| BWC 868.5090
: K130405
Trade/Device Name: Rusch EasyCric Emergency Cricothyrotomy Set Regulation Number: 21 CFR 868.5090
Rusch EasyCric Emergency Cricothyrotomy Set is indicated to provide emergency airway access when conventional ventilation by intubation or face mask cannot be performed.
The Rusch EasyCric Emergency Cricothyrotomy Set is a sterile, single use emergency cricothyroidotomy set available in size 5 mm. The device employs the Seldinger technique to provide an emergency artificial airway when attempts to intubate the trachea fail. Cricothyroidotomy provides faster access in emergency situations than surgical or percutaneous tracheostomy, which are time consuming procedures. The device gains access by cutting the link between the cricoid cartilage and the thyroid cartilage into the ligamentum conicum. It is available uncuffed. Components of the set include the EasyCric tube and insertion dilator, scalpel, puncture needle, syringe, guidewire, comport neckband and saline solution.
Here's an analysis of the provided text, focusing on the acceptance criteria and study information:
The document describes the Rusch EasyCric Emergency Cricothyrotomy Set and its substantial equivalence to a predicate device. The "study" in this context refers to various performance tests conducted to demonstrate that the new device meets established safety and performance standards equivalent to the predicate. This is not a clinical study involving human patients or complex diagnostic algorithms.
Acceptance Criteria and Reported Device Performance
| Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Connector bonding strength | Must be able to sustain axial force of 15 ± 1.5N without movement | Met (Implied by "Pass/fail criteria has been met") |
| Flange (neck-plate) bonding strength | Must be able to sustain axial force of 15 ± 1.5N without movement | Met (Implied by "Pass/fail criteria has been met") |
| Connector Fit Test (ring gauge test) | The connector's leading edge shall lie between the minimum and maximum diameter steps of the gauge. | Met (Implied by "Pass/fail criteria has been met") |
| Coating consistency on the tube | Smooth and even coating on the tube | Met (Implied by "Pass/fail criteria has been met") |
| Coating lubricity | Wear value of less than 150% | Met (Implied by "Pass/fail criteria has been met") |
| Visual Inspection of Saline, Scalpel and Neckband | No torn packaging or damage to the product | Met (Implied by "Pass/fail criteria has been met") |
| Guidewire testing | Smooth insertion without restriction | Met (Implied by "Pass/fail criteria has been met") |
| Biocompatibility testing | Must meet the requirements as outlined in ISO 10993-1 | Met (Implied by "Pass/fail criteria has been met" and "All patient contacting materials are in compliance with ISO 10993-1") |
Study Details:
-
Sample size used for the test set and the data provenance:
- Sample Size: The document does not explicitly state the numerical sample size for each performance test (e.g., how many connectors were tested for bonding strength). It lists the types of tests performed.
- Data Provenance: Not specified, but these are likely laboratory-based tests conducted by the manufacturer (Teleflex Medical). No country of origin for the data or whether it's retrospective/prospective is indicated, as this typically applies to clinical or observational studies on patient data.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. These are engineering/performance bench tests, not evaluations requiring expert medical interpretation to establish ground truth such as in diagnostic imaging studies. The "ground truth" is defined by the physical or chemical properties being measured against established standards (e.g., force, dimension, visual integrity, biocompatibility standards).
-
Adjudication method for the test set:
- Not applicable in the context of expert consensus or multi-reader reviews, as these are technical performance tests against objective criteria. The pass/fail criteria for each test serves as the "adjudication."
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done:
- No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating diagnostic performance of imaging devices or algorithms with human readers, which is not the nature of this device's submission.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a medical device (cricothyrotomy set), not an AI algorithm. Its performance is assessed through its physical and functional integrity, not through algorithmic output.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" for the performance tests consists of established engineering standards, physical measurement criteria, and adherence to recognized international standards (e.g., ISO 5366-1, ISO 5356, ISO 10993-1).
-
The sample size for the training set:
- Not applicable. This device does not involve a "training set" in the context of machine learning or AI models.
-
How the ground truth for the training set was established:
- Not applicable, as there is no training set for this type of device.
Summary of Device Performance Study:
The "study" presented here is a series of bench and laboratory performance tests designed to demonstrate that the Rusch EasyCric Emergency Cricothyrotomy Set is substantially equivalent to a predicate device. The tests evaluate physical characteristics and safety aspects of the device components (e.g., connector strength, coating quality, biocompatibility). The acceptance criteria for these tests are derived from international standards (ISO) or internal specifications designed to ensure the device is safe and performs as intended. The document states that all pass/fail criteria were met, leading to the conclusion of substantial equivalence. This is primarily a regulatory submission focused on engineering and material performance.
Ask a specific question about this device
(120 days)
Cricothyroidotomy Kit
COMMON NAME Emergency Airway Needle
PRODUCT CLASS/CLASSIFICATION Class II 73, 21 CFR 868.5090
6DB
Re: K050166
Trade/Device Name: Portex Emergency Cricothyroidotomy Kit Regulation Number: 21 CFR 868.5090
Emergency airway access (when endotracheal intubation is not possible) by percutaneous insertion of a cricothyroidotomy tube through the cricothyroid membrane.
The new cricothyroidotomy kit is designed for introduction of an emergency 6mm cuffed airway tube through the cricothyroid membrane. The airway tube is provided pre-assembled onto a dilator with a Veress needle located in the centre. The kit employs two well-established and understood techniques: Firstly: The use of a Veress needle to provide visual indication of entry into a body cavity. (the trachea). (This can be confirmed by aspiration through the needle as with the predicate Cook Melker product ref. IFU in Attachment B(i)) The Veress needle is usually used for insufflation of the abdomen. It has a central sprung-loaded protective core to the needle that guards the cutting point and gives a visual indication via a "flag" at the hub end of the needle tip is passing through tissues. When the needle tip passes through into the abdominal cavity, or tractica, passure in ough assucer ------------------------------------------------------------------------------------------------------------------------------------------------------ Secondly: the use of a dilator/introducer to enlarge a stomal opening and place an airway tube within the traches. The pre-assembled needle/dilator/tube is inserted through the cricothyroid whill the tracether and the tube is then slid off the dilator fully into the trachea. This is the same principle employed by the Cook Melker Cricothyroidotomy sct. As well as the airway tube assembly, the kit provides the user with scalpel for skin incision, syringe for aspiration to confirm location of the needle in the trachea, sutures and neck-strap to secure the airway tube and a heat and moisture exchanger to reduce patient energy loss and protect the airway from blockage. These are all items that will be well understood by the user. Packaging comprises a rigid hinged blister pack for protection and a polythene/Tyvek welded pouch as the sterile barrier. pouch as ale sterilised and individually packed in a carton together with the directions for use.
The provided text is a 510(k) summary for the Portex Emergency Cricothyroidotomy Kit. This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria through a standalone study with quantitative metrics like sensitivity or specificity.
Therefore, the document does not contain the information requested in the prompt regarding acceptance criteria, study details, sample sizes, ground truth establishment, or multi-reader multi-case studies.
Here's a breakdown of why the requested information is absent based on the provided text:
-
Table of acceptance criteria and reported device performance: Not present. The document focuses on comparing the new device's technical characteristics and indications for use to those of legally marketed predicate devices to establish substantial equivalence, not to demonstrate performance against pre-defined acceptance criteria.
-
Sample sized used for the test set and data provenance: No test set or associated sample size is mentioned. Clinical studies to evaluate performance against acceptance criteria are not typically required for a 510(k) submission where substantial equivalence is being claimed. The document primarily describes the device and its operational principle.
-
Number of experts used to establish the ground truth for the test set and qualifications: Not applicable, as no test set or ground truth establishment is described.
-
Adjudication method for the test set: Not applicable, as no test set or adjudication process is described.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a mechanical medical instrument, not an AI-powered diagnostic tool. MRMC studies are used for evaluating diagnostic performance of AI or imaging interpretations.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical device, not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable, as no ground truth is established or discussed in the context of performance evaluation. The "ground truth" for a 510(k) in this context is typically the established safety and effectiveness of the predicate device.
-
The sample size for the training set: Not applicable. This device does not involve machine learning or a "training set."
-
How the ground truth for the training set was established: Not applicable for the same reason as above.
In summary, the provided 510(k) text is a regulatory document seeking market clearance based on substantial equivalence to existing devices. It does not include the detailed performance study information requested, which would be typical for a PMA submission or a de novo classification, or for devices with novel technology requiring specific quantitative performance metrics.
Ask a specific question about this device
(146 days)
Bloomington, IN 47402-0489
Re: K013252
Patil Emergency Cricothyrotomy Catheter Set Regulation Number: 868.5090
The Patil Cricothyroid Catheter is used for emergency airway access when conventional endotracheal intubation cannot be performed. It is provided in peel-open packages and is intended for one-time use.
Used for emergency airway access when conventional endotracheal intubation and ventilation cannot be performed.
The Patil Cricothyroid Catheter consists of a connector on the proximal end connected to tubing with an inner diameter. Through the catheter lumen is a dilator. The dilator over a needle is used for insertion. The catheter will be included in a set consisting of appropriately sized components.
This document (K013252) does not contain detailed information about specific acceptance criteria or a dedicated study proving that the device meets those criteria. Instead, it seems to be a 510(k) premarket notification for a medical device (Patil Cricothyrotomy Catheter), which typically focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than conducting extensive performance studies with predefined acceptance criteria.
The phrase "The results of the tests performed provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its use as a cricothyrotomy catheter" (from {1}) indicates that some testing was done, but the specifics of those tests, the acceptance criteria, and the detailed results are not included in this document.
Therefore, most of the information requested in your prompt cannot be extracted from this document. However, I can provide what is available:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Not specified in this document. | The document states, "The results of the tests performed provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its use as a cricothyrotomy catheter." No specific performance metrics or values are provided. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not specified.
- Data Provenance: Not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable/Not specified, as no study with a test set requiring expert-established ground truth is detailed.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable/Not specified.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC comparative effectiveness study is mentioned in this document. The device is a physical catheter, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. The device is a physical catheter.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not applicable/Not specified, as no study with a ground truth determination process is detailed.
8. The sample size for the training set
- Not applicable, as this device is a physical component, not an AI/algorithmic system that typically involves training sets.
9. How the ground truth for the training set was established
- Not applicable.
Ask a specific question about this device
(146 days)
Bloomington, IN 47402-0489
Re: K013241
Arndt Emergency Cricothyrotomy Catheter Set Regulation Number: 868.5090
Used for emergency airway access when conventional endotracheal . intubation and ventilation cannot be performed.
The Arndt Cricothyroid Catheter consists of a connector on the proximal end connected to tubing with an inner diameter. Through the catheter lumen is a dilator. The dilator provides a transition to a wire guide for insertion. The device is placed using the Seldinger technique. The catheter will be included in a set consisting of appropriately sized components.
The provided information does not contain any data regarding acceptance criteria or a study proving device performance in relation to such criteria. The document is a 510(k) premarket notification for the Arndt Emergency Cricothyrotomy Catheter Set, primarily focusing on its substantial equivalence to predicate devices and its intended use.
The text describes the device, its indications for use, and a comparison to existing predicate devices (Melker Cuffed Cricothyroid Catheter Set and The Pertrach). The FDA letter confirms the substantial equivalence determination, allowing the device to be marketed. However, no performance metrics, acceptance criteria, or study results are presented.
Therefore, I cannot fulfill the request to provide:
- A table of acceptance criteria and reported device performance.
- Sample sizes used for the test set or data provenance.
- Number of experts or their qualifications for establishing ground truth.
- Adjudication method.
- MRMC comparative effectiveness study results or effect sizes.
- Standalone performance results.
- Type of ground truth used.
- Sample size for the training set.
- How ground truth for the training set was established.
Ask a specific question about this device
(23 days)
needle |
| Proposed Classification Name: | Emergency Airway Needle21 CFR Part 868.5090
Bloomington, IN 47402-0489
Re: K013916
Melker Emergency Cricothyrotomy Catheter Set Regulation Number: 868.5090
The Melker Emergency Cricothyrotomy Catheter Set is used for emergency airway access.
. In patients whom conventional endotracheal intubation and ventilation cannot be performed.
The device will be supplied sterile and is intended for one-time use.
Used for emergency airway access when conventional endotracheal intubation and ventilation cannot be performed.
The Melker Emergency Cricothyrotomy Catheter with a standard inner diameter of 3.5, 4, or 6 mm and a length of either 3.8, 4.2, 7.5 cm with a proximal fitting and a coaxial dilator tapered to a 0.038-inch wire guide.
This 510(k) premarket notification for the Melker Emergency Cricothyrotomy Catheter Set (K013916) does not contain the detailed information necessary to fully address all aspects of your request. The submission is primarily focused on demonstrating substantial equivalence to a predicate device (K010016) based on indications for use, material, and physical characteristics, and adherence to quality system procedures.
Here's an analysis based on the provided text, highlighting what is available and what is missing:
Acceptance Criteria and Study for K013916: Melker Emergency Cricothyrotomy Catheter Set
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria Category | Reported Device Performance | Comments |
|---|---|---|
| Specific Performance Criteria | Not explicitly stated in terms of measurable thresholds or pass/fail criteria. | The document states "Specific performance of the device subject of this submission is assured through adherence to the firm's quality system procedures for device quality. Various device design specifications and quality specifications are used to produce a device that meets predetermined acceptance criteria." However, the specific acceptance criteria themselves and the performance metrics against those criteria are not detailed in this 510(k) summary. |
| Substantial Equivalence | "This device is the same with respect to indications for use, material and physical characteristics to the Melker Cuffed Cricothyrotomy Catheter Set cleared under K010016 on 10/09/01." | This is the primary "performance" being demonstrated for regulatory clearance. The device is deemed to perform acceptably because it is equivalent to a previously cleared device. |
| Sterility | "The Melker Emergency Cricothyrotomy Catheter Set will be sterilized using an Ethylene Oxide (ETO) gas cycle validated to assure a 10⁻⁶ sterility assurance level." | This is a specific performance criterion and the reported result (validation to 10⁻⁶ SAL) meets the industry standard for sterile medical devices. |
| Packaging Integrity | "Each unit is packaged and sealed within a Tyvek-Poly pouch and properly labeled." | Implies that packaging integrity is a criterion, though specific test results (e.g., seal strength, barrier properties) are not provided. |
2. Sample Size Used for the Test Set and Data Provenance:
- The document does not describe a specific "test set" in the context of a clinical or performance study with human or simulated cases for K013916.
- The comparison is based on the characteristics of the device being "the same" as a predicate device (K010016).
- Therefore, no sample size for a test set is provided, and data provenance for such a test set is not applicable in this submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:
- Not applicable. As no specific "test set" or clinical study requiring ground truth establishment is described for K013916, there is no mention of experts or their qualifications for this purpose.
4. Adjudication Method for the Test Set:
- Not applicable. No test set requiring an adjudication method is described.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
- No MRMC comparative effectiveness study was done or reported in this submission. The submission relies on substantial equivalence to the predicate device, not on demonstrating improved human reader performance with or without AI.
6. Standalone Performance Study (Algorithm Only):
- Not applicable. This device is a catheter set, not an algorithm or AI-driven system. Therefore, a standalone (algorithm only) performance study is not relevant.
7. Type of Ground Truth Used:
- Not applicable. As there is no described test set or study involving diagnostic performance, the concept of "ground truth" (e.g., pathology, outcomes data) is not relevant to this 510(k) submission for K013916. The "ground truth" for the clearance is essentially the established performance and safety profile of the predicate device (K010016).
8. Sample Size for the Training Set:
- Not applicable. As this is a physical medical device (catheter set) and not an AI/ML algorithm requiring a training set, the concept of a "training set" is not relevant to this submission.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable. As there is no training set mentioned, the establishment of its ground truth is also not applicable.
Summary of Study (as presented in the 510(k) for K013916):
The "study" for K013916 is a substantial equivalence comparison to a predicate device (Melker Cuffed Cricothyrotomy Catheter Set, K010016). The manufacturer asserts that the new device is "the same with respect to indications for use, material and physical characteristics" as the predicate device. Therefore, the "acceptance criteria" in this context are that the new device replicates the key attributes of the cleared predicate device. The "study" largely involves documentation of device design, materials, manufacturing processes, and sterilization validation, aligning them with those of the predicate device and the firm's quality system.
The submission does not involve clinical trials, comparative studies against human performance, or studies typically associated with AI/ML devices that would require test sets, ground truth, or statistical power calculations for diagnostic accuracy.
Ask a specific question about this device
Page 1 of 1