Search Results
Found 3 results
510(k) Data Aggregation
(263 days)
Common Name: Cryptococcus Antigen EIA
Classification Name: Antigen, Elisa, Cryptococcus
Regulation: 866.3165
7 2012
Re: K120946
Trade/Device Name: ALPHA Cryptococcal Antigen EIA Regulation Number: 21 CFR $866.3165
The ALPHA Cryptococcal Antigen enzyme immunoassay (CrAg EIA) is a qualitative or semi-quantitative (titration) test system for the detection of capsular polysaccharide antigens of Cryptococcus species complex (Cryptococcus neoformans and Cryptococcus gattii) in serum and cerebrospinal fluid (CSF). The ALPHA Cryptococcal Antigen Enzyme Immunoassay is an assay which can be used as an aid in the diagnosis of cryptococcosis. Test results are to be used in conjunction with information available from the patient clinical evaluation and other diagnostic procedures.
The ALPHA Cryptococcal Antigen Enzyme Immunoassay (EIA) is a direct immunoenzymatic sandwich microplate assay which detects Cryptococcus antigens in serum and CSF. Anti-Cryptococcus antibodies bound to microwell plates are used as capture antibodies, and horseradish peroxidase (HRP)-conjugated anti-Cryptococcus antibodies are used as detect antibodies. The positive control and standard curve material are composed of cryptococcal capsular polysaccharide antigen in a buffered protein solution with a preservative. In the qualitative procedure, specimens are analyzed undiluted. In the titration procedure, specimens are analyzed after serial dilution in specimen diluent. Either serum or CSF is added to the microwells coated with the capture antibodies and incubated. If the patient specimen contains cryptococcal antigens that are recognized by the capture antibodies, those antigens will become bound to the microwells. The microwells are washed to remove unbound patient material, and HRP-conjugated detect antibody is added to the wells. If Cryptococcus antigens are bound to the microwells by the capture antibodies, the detect antibody will also become bound to the microwells. The wells are then washed to remove any unbound detect antibody. Next, tetramethylbenzidine (TMB) substrate is added to the microwells, and in the presence of HRP, a blue color will develop. The reaction is stopped by the addition of a stop solution. The optical density (OD) is determined with a microplate reader at 450 nm with reference at 630 nm (reference is optional).
Here's an analysis of the ALPHA Cryptococcal Antigen EIA based on the provided 510(k) summary, structured to address your specific points:
Acceptance Criteria and Device Performance Study
The document describes several analytical performance studies rather than clinical studies with patient outcomes or a specific "acceptance criteria" table for diagnostic accuracy. The primary criteria for device approval appear to be substantial equivalence to the predicate device (Meridian's Premier™ Cryptococcal Antigen EIA, K904393) and meeting internal laboratory performance benchmarks for precision, sensitivity, specificity, and non-interference.
However, based on the ROC analysis and method comparison studies, we can infer some performance expectations for sensitivity and specificity.
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Metric | Acceptance Criteria (Inferred from Predicate Equivalence and ROC analysis) | Reported Device Performance (ALPHA Cryptococcal Antigen EIA) |
|---|---|---|
| Qualitative Agreement vs. Commercial EIA (Serum) | High agreement (similar to predicate) | % Positive Agreement: 98.5% (95% CI: 94.7% - 99.6%) |
| % Negative Agreement: 97.7% (95% CI: 96.7% - 98.4%) | ||
| Qualitative Agreement vs. Commercial EIA (CSF) | High agreement (similar to predicate) | % Positive Agreement: 93.5% (95% CI: 79.3% - 98.2%) |
| % Negative Agreement: 99.2% (95% CI: 97.8% - 99.7%) | ||
| Qualitative Agreement vs. IMMY CrAg LFA (Serum) | High agreement (similar to companion LFA) | % Positive Agreement: 96.9% |
| % Negative Agreement: 99.8% | ||
| Qualitative Agreement vs. IMMY CrAg LFA (CSF) | High agreement (similar to companion LFA) | % Positive Agreement: 93.8% |
| % Negative Agreement: 99.5% | ||
| Analytical Sensitivity (LoD) - Serum | Not explicitly stated as acceptance criteria, but demonstrates detection capability. | 5.3 ng/ml |
| Analytical Sensitivity (LoD) - CSF | Not explicitly stated as acceptance criteria, but demonstrates detection capability. | 1.7 ng/ml |
| Analytical Specificity (Cross-Reactivity) | Total positive for fungal pathologies < 10% (excluding specific known cross-reactants) | Total fungal pathologies: 3.9% (excluding Paracoccidioides brasiliensis, which was 67%) |
| Interference | No interference from icteric, hemolyzed, or lipemic samples | All spiked samples were positive, all unspiked negative. |
| High Dose Hook Effect | Positive result still recorded at high concentrations; semi-quantitative titration able to differentiate. | Signal reduced, but still positive. Titration can differentiate. |
| Freeze-Thaw Stability | Acceptance implied to be defined by observed OD reductions. Avoid multiple freeze-thaws for very low-positive samples. | Reductions in OD values after one week at ≤ -20°C and after multiple freeze-thaws observed. |
| ROC Cutoff Sensitivity (vs. IMMY CrAg LFA) | Not explicitly stated, but target was to match LFA. | 97.4% (for blanked OD > 0.265) |
| ROC Cutoff Specificity (vs. IMMY CrAg LFA) | Not explicitly stated, but target was to match LFA. | 99.9% (for blanked OD > 0.265) |
2. Sample Size Used for the Test Set and Data Provenance
The "test set" for performance evaluation consists of several cohorts:
- Precision Studies: Spiked serum and mock CSF samples. The number of unique samples is not given, but they were tested twice per day over five days across three sites. Each panel consisted of negative, high negative, low positive, and moderate positive samples for both serum and CSF.
- Analytical Sensitivity (LoD): 80 replicates of normal human serum, 80 replicates of artificial CSF for LoB. For LoD, initially 20 replicates of 4 concentrations, then 24 replicates of serum and CSF spiked with antigen near initial LoD.
- Analytical Specificity (Cross-Reactivity): 118 serum specimens and 15 fungal culture filtrates.
- Interference Studies: Five icteric, five hemolyzed, and five lipemic serum specimens.
- High Dose Hook Effect: One negative serum specimen pool spiked with antigen.
- Freeze-Thaw Study: Three CrAg-Negative human CSF specimens and four CrAg-Negative human serum specimens, spiked.
- ROC Analysis (for Cutoff Determination): 995 combined serum and CSF specimens.
- Method Comparison (vs. Commercial EIA and IMMY LFA): 1782 specimens (CSF n=426; serum n=1356).
- Data Provenance: The studies were conducted at Immuno-Mycologics, Inc. (IMMY) and two national reference laboratories, as well as a clinical lab for precision. The specimens for the method comparison were "submitted for cryptococcal antigen EIA testing," implying they were clinical samples, likely retrospective or leftover clinical samples. The country of origin is not explicitly stated, but the locations (Norman, OK; national reference labs) suggest United States origin. The nature of the samples (spiked or "submitted for testing") suggests retrospective use of clinical samples or artificially prepared samples.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The ground truth for the test sets was primarily established through:
- Spiking: For precision, LoD, interference, high dose hook, and freeze-thaw studies, samples were spiked with known concentrations of cryptococcal antigen, establishing a de facto ground truth.
- Reference Methods:
- For ROC analysis and comparison to LFA, the IMMY CrAg Lateral Flow Assay (LFA) was used as the reference method to classify specimens as true positive or negative. The expertise behind the LFA's accuracy would pertain to its own validation.
- For comparison to another manufacturer's device, the commercial Cryptococcal antigen EIA (the predicate device, Meridian's Premier™ Cryptococcal Antigen EIA) was used as the comparator.
- For analytical specificity, known positive/negative states for cross-reactants (e.g., Aspergillus galactomannan positive confirmed by Bio-Rad's Platelia™ Aspergillus EIA) were used.
The document does not mention the use of human experts (e.g., radiologists, pathologists) to establish ground truth for the test sets in the traditional sense of medical image analysis or complex diagnostic interpretation. The focus is on analytical performance and agreement with established laboratory tests.
4. Adjudication Method for the Test Set
No explicit adjudication method (e.g., 2+1, 3+1) is mentioned for the test sets. The ground truth was primarily defined by:
- Known concentrations (for spiked samples).
- Results from a single reference laboratory test (IMMY CrAg LFA or the commercial predicate EIA).
- Known status for cross-reactivity samples (e.g., confirmed Aspergillus positive).
There is no indication of multiple human readers or a consensus process for determining the ground truth for any of these studies.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) immunoassay that provides a quantitative or qualitative result. It is not an AI-powered diagnostic tool designed to assist human readers in interpreting images or complex data, nor does it involve "human readers" in the sense of medical specialists interpreting results that the device then supports or enhances.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
This device is a standalone diagnostic test. It functions as an "algorithm only" in the sense that the assay procedure and subsequent optical density (OD) readings lead to a quantitative value or a positive/negative determination based on a defined cutoff. The performance studies described (precision, LoD, cross-reactivity, method comparison) are reflections of its standalone performance. There isn't a "human-in-the-loop" component in interpreting the EIA's direct output beyond standard laboratory practice of running the assay and reporting the result.
7. The Type of Ground Truth Used
The ground truth used for these studies can be categorized as:
- Known Spiking Concentrations: For analytical sensitivity, precision, interference, high-dose hook, and freeze-thaw studies.
- Reference Laboratory Assay Results:
- IMMY CrAg Lateral Flow Assay (LFA): Used as the reference for ROC analysis and comparison studies against the LFA.
- Commercial Cryptococcal Antigen EIA (Predicate Device): Used as the reference for method comparison studies to demonstrate substantial equivalence.
- External reference assays: For confirming known states of potentially cross-reacting substances (e.g., Bio-Rad's Platelia™ Aspergillus EIA).
8. The Sample Size for the Training Set
No explicit "training set" for an AI algorithm is mentioned as this is a traditional immunoassay.
However, if we interpret "training set" in the context of establishing assay parameters or a cutoff:
- The ROC analysis to determine the cutoff (0.265 blanked OD) was performed on a dataset of 995 combined serum and CSF specimens. This dataset could be considered analogous to a "training set" for defining the assay's threshold for positivity.
9. How the Ground Truth for the Training Set was Established
For the 995 specimens used in the ROC analysis to establish the cutoff:
- The ground truth was established by comparing the ALPHA Cryptococcal Antigen EIA results to the IMMY CrAg Lateral Flow Assay (LFA). The LFA was used to classify a specimen as a "true positive or negative" for the purpose of matching the sensitivity and specificity of the new EIA to the LFA. This means the LFA's results were accepted as the ground truth for this particular analysis aiming for concordance.
Ask a specific question about this device
(218 days)
|
| Regulation: | 866.3165
73071
Re: K112422
Trade/Device Name: CrAg Lateral Flow Assay (CrAg LFA) Regulation Number: 21 CFR § 866.3165
The Cryptococcal Antigen Lateral Flow Assay (CrAg LFA) is an immunochromatographic test system for the qualitative or semi-quantitative detection of capsular polysaccharide antigens of Cryptococcus species complex (Cryptococcus neoformans and Cryptococcus gattii) in serum and cerebral spinal fluid (CSF).
The CrAg Lateral Flow Assay is a prescription-use laboratory assay, which can aid in the diagnosis of cryptococcosis.
The CrAg Lateral Flow Assay is a dipstick sandwich immunochromatographic assay, which detects cryptococcal antigen in serum and CSF. For the qualitative procedure, specimens are diluted 1:2 in 1x Specimen Diluent and analyzed. For the semi-quantitative procedure, specimens are diluted 1:5 in 1x Specimen Diluent followed by 1:2 serial dilutions. All dilutions are then analyzed as in the qualitative procedure. Specimens are placed into an appropriate reservoir, such as a test tube or microtiter plate, and the lateral flow device is then placed into the reservoir allowing the specimen to come into contact with the test membrane. The test uses specimen wicking to capture gold-conjugated, anti-cryptococcal monoclonal antibodies and gold-conjugated control antibodies that are deposited onto a membrane. If cryptococcal antigen is present in the specimen, it binds to the gold-conjugated, anti-cryptococcal antibodies. The gold-labeled antibody-antigen complex will continue to wick up the membrane where it will interact with the Test Line (T). The Test Line is immobilized anti-cryptococcal monoclonal antibodies. If the specimen contains cryptococcal antigen, a sandwich is created with the gold-labeled antibodies and the immobilized antibodies, causing a visible line to develop at the test line site (T). If proper flow occurs and the reagents are reactive at the time of use, the wicking of any specimen, positive or negative, will cause the gold-conjugated control goat IgG antibody to move to the Control Line (C) which is immobilized bovine anti-goat IgG antibody. The immobilized anti-goat antibody will bind to the gold-conjugated goat IgG Control antibody and will cause a visible line to develop (C). A positive test result will create two lines, while a negative test result will create one line (Figure 1). If the control line fails to develop a line, then the test is not valid.
Here's an analysis of the provided text regarding the acceptance criteria and study for the CrAg Lateral Flow Assay:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Implied) | Reported Device Performance (CSF) |
|---|---|
| High Sensitivity | 100% (95% CI: 94.4-100.0%) |
| High Specificity | 100% (95% CI: 96.3-100%) |
| Repeatability (Overall) | • Med. Pos: 100% positive detected • Low Pos: 100% positive detected • High Neg: 96% negative detected • Neg: 100% negative detected |
| Reproducibility | (Same as above, as "overall percent positive and percent negative detected were calculated by combining the data from all three sites") |
| Analytical Sensitivity | 1.25 ng/ml (defined as concentration where 50% positive, 50% negative) |
| No High Dose Hook Effect | High dose hook effect possible for concentrations > 200ug/ml |
Note on Acceptance Criteria: The document does not explicitly state numerical acceptance criteria (e.g., "sensitivity must be >95%"). However, it implicitly aims for "very high sensitivity and specificity" based on the background information and the excellent results presented. The repeatability and analytical sensitivity are reported as measured, indicating the performance achieved. The high dose hook effect is identified as a limitation rather than a criterion met.
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size (CSF for Method Comparison):
- Positive (Culture/India Ink): 65 specimens
- Negative (Culture/India Ink): 99 specimens
- Total: 164 specimens
- Data Provenance: The studies contained a mix of both prospective and retrospective specimens. The country of origin is not specified, but the device is indicated for use in the US (implied by FDA submission).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- The document does not specify the number of experts or their qualifications for establishing the ground truth (Culture/India Ink).
- For the repeatability and reproducibility study (CSF), "five different operators" across "three different sites" were involved. Their specific qualifications are not detailed beyond "internal" and "US reference/hospital laboratory" personnel.
4. Adjudication Method for the Test Set
- The document does not describe any adjudication method for the test set. The "gold standard" (culture and/or India Ink) is presented as the definitive truth.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
- No, an MRMC comparative effectiveness study was not explicitly described in this 510(k) summary. The study focuses on the standalone performance of the CrAg LFA compared to a gold standard, not on the improvement of human readers with or without AI assistance (as this is a point-of-care type diagnostic, not an AI-assisted interpretation tool for images).
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Yes, a standalone performance study was conducted. The CrAg Lateral Flow Assay is a rapid diagnostic test where the result (presence/absence of lines) is directly interpreted. The reported sensitivity and specificity values represent the performance of the device itself (algorithm only, in a sense, as it's a defined chemical reaction leading to a visible outcome) against the gold standard. While a human reads the lines, the core performance metrics are about the device's ability to detect the antigen.
7. The Type of Ground Truth Used
- For the method comparison study (CSF), the ground truth used was "gold standard for the diagnosis of cryptococcosis (culture and/or India Ink)."
8. The Sample Size for the Training Set
- The document does not specify a training set sample size. This is common for lateral flow assays, which are typically developed and optimized through laboratory analytical studies rather than machine learning models that require distinct training sets. The studies described are primarily performance validation studies.
9. How the Ground Truth for the Training Set Was Established
- Since a distinct "training set" in the context of machine learning is not mentioned as such, the method of establishing ground truth for development/optimization would likely involve spiking known concentrations of cryptococcal antigen into mock specimens (as seen in the analytical sensitivity sections) and testing against other validated methods or known positive/negative samples. The document refers to "mock CSF that was negative by the IMMY Latex-Cryptococcal Antigen Detection System" for repeatability/reproducibility.
Ask a specific question about this device
(342 days)
|
| Regulation: | 866.3165
| Trade/Device Name: | CrAg Lateral Flow Assay |
| Regulation Number: | 21CFR §866.3165
The CrAg Lateral Flow Assay is an immunochromatographic test system for the qualitative or semi-quantitative detection of capsular polysaccharide antigens of Cryptococcus species complex (Cryptococcus neoformans and Cryptococcus gatti) in serum.
The CrAg Lateral Flow Assay is a prescription-use laboratory assay, which can aid in the diagnosis of Cryptococcosis.
The CrAg Lateral Flow Assay is a dipstick sandwich immunochromatographic assay, which detects cryptococcal antigen in serum. For the qualitative procedure, specimens are diluted 1:2 in 1x Specimen Diluent and analyzed. For the semi-quantitative procedure, specimens are diluted 1:5 in 1x Specimen Diluent followed by 1:2 serial dilutions. All dilutions are then analyzed as in the qualitative procedure. Specimens are placed into an appropriate reservoir, such as a test tube or microtiter plate, and the lateral flow device is then placed into the reservoir allowing the specimen to come into contact with the test membrane. The test uses specimen wicking to capture gold-conjugated, anti-cryptococcal monoclonal antibodies and gold-conjugated control antibodies that are deposited onto a membrane. If cryptococcal antigen is present in the specimen, it binds to the gold-conjugated, anti-cryptococcal antibodies. The gold-labeled antibody-antigen complex will continue to wick up the membrane where it will interact with the Test Line (T). The Test Line is immobilized anti-cryptococcal monoclonal antibodies. If the specimen contains cryptococcal antigen, a sandwich is created with the gold-labeled antibodies and the immobilized antibodies, causing a visible line to develop at the test line site (T). If proper flow occurs and the reagents are reactive at the time of use, the wicking of any specimen, positive or negative, will cause the gold-conjugated control goat IgG antibody to move to the Control Line (C) which is immobilized bovine anti-goat IgG antibody. The immobilized anti-goat antibody will bind to the gold-conjugated goat IgG Control antibody and will cause a visible line to develop (C). A positive test result will create two lines, while a negative test result will create one line (Figure 1). If the control line fails to develop a line, then the test is not valid.
The CrAg Lateral Flow Assay is an immunochromatographic test system for the qualitative or semi-quantitative detection of capsular polysaccharide antigens of Cryptococcus species complex in serum. The study provided compares the CrAg Lateral Flow Assay to a Cryptococcal Antigen EIA and to culture/India Ink as gold standards.
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly state pre-defined acceptance criteria in terms of specific sensitivity and specificity thresholds for the de novo approval. However, the performance is reported as follows:
Comparison with Cryptococcal Antigen EIA (Serum):
| Metric | Acceptance Criteria (Implied) | Reported Performance (95% CI) |
|---|---|---|
| % Agreement Positive | Not explicitly stated | 100% (96-100%) |
| % Agreement Negative | Not explicitly stated | 93% (86-97%) |
Comparison with Culture/India Ink (Gold Standard for Serum):
| Metric | Acceptance Criteria (Implied) | Reported Performance (95% CI) |
|---|---|---|
| Sensitivity | Not explicitly stated | 100% (96.0%-100%) |
| Specificity | Not explicitly stated | 100% (97.0%-100%) |
Semi-Quantitative Serum Correlation (CrAg LFA vs. Latex Agglutination):
- Correlation (R²): Not explicitly stated | 0.905 |
2. Sample Size and Data Provenance:
- Test Set (Method Comparison with Cryptococcal Antigen EIA): 197 serum specimens. The data provenance is from a "US reference laboratory" and the study was retrospective (collected and stored frozen until studies were performed).
- Test Set (Method Comparison with Culture/India Ink): The exact number of specimens is not clearly enumerated, but the table indicates a 2x2 contingency table for "Culture/India Ink" with "NO ANDRE & FOR A LE SE SE SE SE SE SE SERVED AN A SE" in one row and other unclear annotations. However, the sample size can be inferred from the 95% Confidence Intervals for Sensitivity and Specificity, which suggest a reasonable sample size for a diagnostic test. The document states "These studies contained a mix of both prospective and retrospective specimens."
- Test Set (Predicate Device Comparison - Latex-Cryptococcal Antigen Detection System): 197 serum specimens. The data provenance is from a "US reference laboratory" and the study was retrospective (collected and stored frozen until studies were performed).
3. Number of Experts and Qualifications for Ground Truth:
The document does not specify the number or qualifications of experts used to establish the ground truth for the test set.
4. Adjudication Method for the Test Set:
The document does not describe an adjudication method for the test set results. The ground truth (EIA or Culture/India Ink) was likely established by the reference laboratory or standard clinical procedures without explicit mention of an additional adjudication process for this study.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No multi-reader multi-case (MRMC) comparative effectiveness study was done. The CrAg Lateral Flow Assay is a diagnostic device that provides an objective result (presence/absence of line, or semi-quantitative titer), and its performance is assessed against established laboratory methods or clinical gold standards, not by comparing the performance of human readers with and without AI assistance.
6. Standalone Performance Study:
Yes, a standalone performance study was done. The method comparison studies (against Cryptococcal Antigen EIA and Culture/India Ink) and the precision, analytical sensitivity, and analytical specificity studies all evaluate the intrinsic performance of the CrAg Lateral Flow Assay device without human-in-the-loop assistance.
7. Type of Ground Truth Used:
- For Method Comparison: Cryptococcal Antigen EIA results from a US reference laboratory.
- For Gold Standard Comparison: Culture and/or India Ink results. These are considered the "gold standard for the diagnosis of cryptococcosis."
- For Training Set: Not applicable as this is a diagnostic assay, not a machine learning model requiring a separate training set. The performance studies evaluate the assay itself.
8. Sample Size for the Training Set:
Not applicable. This is a conventional in vitro diagnostic assay, not an AI/ML-based device that typically requires a distinct training set for model development. The described studies are all for performance evaluation of the developed assay.
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as there is no mention of a training set for an AI/ML model. The "ground truth" in the context of this device refers to the established diagnostic methods used to evaluate the CrAg Lateral Flow Assay's accuracy.
Ask a specific question about this device
Page 1 of 1