Found 3 results

510(k) Data Aggregation

    K Number
    K192871

    Validate with FDA (Live)

    Date Cleared
    2019-11-07

    (30 days)

    Product Code
    Regulation Number
    866.3045
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Georgia 30329

    Re: K192871

    Trade/Device Name: B. anthracis Real-time PCR Assay Regulation Number: 21 CFR 866.3045
    regulation: (if applicable) | In vitro diagnostic device for Bacillus spp. detection; 21 CFR 866.3045

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The B. anthracis Real-Time PCR Assay is an in vitro diagnostic test for the qualitative detection of plasmid and chromosomal DNA sequences from B. anthracis. The assay can be used to test human respiratory samples, whole blood, serum, plasma, swabs from lesions, cerebrospinal fluid, and bacterial culture isolates from individuals suspected of having anthrax.

    Results generated from direct specimentesting are presumptive for the identification of B. anthracis. Results generated from culture isolate testing should be used in conventional methods for identification of Bacillus anthracis isolates as part of the LRN Bacillus anthracis Testing Algorithm. The diagnosis of anthrax infection must be made based on history, signs, symptoms, exposure likelihood, and other laboratory evidences, in addition to the identification of B. anthracis from cultures or detection directly in clinical specimens.

    Use is limited to Laboratory Response Network (LRN) designated laboratories.

    The B. anthracis Real-time PCR Assay is also intended for environmental specimen testing for biothreat detection and response. FDA has not evaluated claims related to the use of this assay on environmental specimens.

    Device Description

    The B. anthracis Real-time PCR Assay uses a fluorogenic probe, consisting of an oligonucleotide with a reporter dve (FAM) attached to the 5' end and a quencher dye (BHQ1) attached at or near the 3′ end. The probe anneals to a specific target sequence located between the forward and reverse primers. During the extension phase of the PCR cycle, the 5′ nuclease activity of Taq polymerase degrades the probe causing the reporter dye to separate from the quencher dye and a fluorescent signal is generated. With each cycle, additional reporter dye molecules are cleaved from their respective probes and the fluorescence intensity is monitored during the PCR. The Tag polymerase used in this assay is inactive at room temperature. It must be activated by incubation at 95°C, which also minimizes the production of nonspecific amplification products.

    Each extracted DNA sample is tested with three B. anthracis primer and probe sets run as individual reactions. The primer and probe sets target genes encoding virulence factors as well as conserved regions of DNA from the B. anthracis chromosome. All primer and probe sets must be positive for the overall result of the B. anthracis Real-time PCR Assay to be interpreted as positive. Any result that is positive for some, but not all target regions, is still considered equivocal and follow-up laboratory investigation should be performed per the LRN Bacillus anthracis testing algorithm.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the B. anthracis Real-time PCR Assay. It indicates substantial equivalence to a predicate device and provides information on its intended use, device description, and a comparison with the predicate. However, it explicitly directs inquiries regarding performance characteristics (including analytical limit of detection, analytical sensitivity and specificity, and clinical performance) to the Centers for Disease Control and Prevention. Therefore, the acceptance criteria and the study that proves the device meets those criteria are not detailed within this document.

    Here is a summary of the information that is available in the document, and what is explicitly stated as not available:


    1. Table of Acceptance Criteria and Reported Device Performance

    • Not available in the provided document. The document directs inquiries regarding performance characteristics to the Centers for Disease Control and Prevention.

    2. Sample size used for the test set and the data provenance

    • Not available in the provided document. The document states that inquiries regarding analytical LoD, analytical sensitivity and specificity, and clinical performance should be directed to the Centers for Disease Control and Prevention. This would include information about sample sizes and data provenance for validation studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable / Not available in the provided document. For a PCR assay, the "ground truth" is typically established by definitive laboratory methods (e.g., bacterial culture, sequencing, or a reference method) rather than expert opinion on images or clinical assessments. The details of how ground truth was established for the analytical and clinical performance studies are not present.

    4. Adjudication method for the test set

    • Not applicable / Not available in the provided document. See point 3.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This device is a diagnostic PCR assay, not an AI-assisted imaging or diagnostic tool that involves human readers in the traditional sense of an MRMC study.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, implicitly. The device is a Real-time PCR Assay, which is an in vitro diagnostic test. Its performance inherently refers to its standalone analytical and clinical performance in detecting target DNA sequences. While human operators are involved in running the assay and interpreting results, the "algorithm" (the PCR chemistry and detection) itself functions stand-alone. However, specific details of "standalone performance" metrics (like sensitivity, specificity, accuracy against a reference standard) are not provided and are referred to the CDC.

    7. The type of ground truth used

    • For a PCR assay detecting DNA sequences from B. anthracis, the ground truth for performance studies would typically be established by a "gold standard" laboratory method, such as:
      • Bacterial culture and identification: Confirmation of B. anthracis growth and subsequent identification using conventional microbiological methods (e.g., phenotypic differences, susceptibility testing).
      • Reference molecular methods: Other validated PCR assays or DNA sequencing for definitive identification.
    • Specific details of how ground truth was established for the validation studies are not available in the provided document.

    8. The sample size for the training set

    • Not applicable. For a PCR assay, there isn't a "training set" in the machine learning sense. The assay is developed based on known genetic targets of B. anthracis. Development involves optimizing primer/probe design and assay conditions, but not "training" an algorithm on a dataset.

    9. How the ground truth for the training set was established

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K140426

    Validate with FDA (Live)

    Date Cleared
    2014-05-22

    (92 days)

    Product Code
    Regulation Number
    866.3045
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    detection; 21 CFR 866.3045
    30333 USA

    Re: K140426

    Trade/Device Name: B. anthracis Real-Time PCR Assay Regulation Number: 21 CFR 866.3045

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The B. anthracis Real-Time PCR Assay is an in vitro diagnostic test for the qualitative detection of plasmid and chromosomal DNA sequences from B. anthracis. The assay can be used to test human respiratory samples, whole blood, serum, plasma, swabs from lesions, CSF, pleural fluid, and bacterial culture isolates from individuals suspected of having anthrax.

    Results generated from direct specimen testing are presumptive for the identification of B. anthracis. Results generated from culture isolate testing should be used in conjunction with other conventional methods for identification of Bacillus anthracis isolates as part of the LRN Bacillus anthracis Testing Algorithm. The diagnosis of anthrax infection must be made based on history, signs, symptoms, exposure likelihood, and other laboratory evidences in addition to the identification of B. anthracis from cultures or detection directly in clinical specimens.

    Use is limited to Laboratory Response Network (LRN) designated laboratories.

    The B. anthracis Real-time PCR Assay is also intended for environmental specimen testing for biothreat detection and response. FDA has not evaluated claims related to the use of this assay on environmental specimens.

    Device Description

    The B. anthracis Real-time PCR Assay uses a fluorogenic probe, consisting of an oligonucleotide with a reporter dye (FAM) attached to the 5' end and a quencher dye (BHQ1) attached at or near the 3' end. The probe anneals to a specific target sequence located between the forward and reverse primers. During the extension phase of the PCR cycle, the 5′ nuclease activity of Tag polymerase degrades the probe causing the reporter dye to separate from the quencher dye and a fluorescent signal is generated. With each cycle, additional reporter dye molecules are cleaved from their respective probes and the fluorescence intensity is monitored during the PCR. The Tag polymerase used in this assay is inactive at room temperature. It must be activated by incubation at 95°C, which also minimizes the production of nonspecific amplification products.

    Each extracted DNA sample is tested with three B. anthracis primer and probe sets run as individual reactions. The primer and probe sets target genes encoding virulence factors as well as conserved regions of DNA from the B. anthracis chromosome. All three primer and probe sets must be positive for the overall result of the B. anthracis Real-time PCR Assay to be interpreted as positive. Any result that is positive for some, but not all three target regions, is still considered equivocal and follow-up laboratory investigation should be performed per the LRN Bacillus anthracis Testing Algorithm.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the B. anthracis Real-time PCR Assay, based on the provided 510(k) summary:

    1. Acceptance Criteria and Reported Device Performance

    The provided document does not explicitly state quantitative acceptance criteria (e.g., sensitivity, specificity thresholds with confidence intervals) for the B. anthracis Real-time PCR Assay. However, the performance characteristics were established by demonstrating its ability to differentiate B. anthracis from other Bacillus species. The document details the types of studies conducted to establish performance but does not provide specific numerical outcomes.

    Acceptance Criteria CategoryReported Device Performance
    Limit of Detection (LoD)Determined through in-house and multicenter sensitivity studies. (Specific numerical LoD not provided in this summary.)
    Analytical SensitivityInquiries regarding performance characteristics should be directed to the Centers for Disease Control and Prevention. (Specific numerical sensitivity not provided in this summary.)
    Analytical SpecificityDemonstrated by evaluating the BA3 marker in a historical collection of Bacillus sp. isolates and by testing known B. cereus and other bacterial isolates. (Specific numerical specificity not provided in this summary.)
    Clinical PerformanceEstablished through three evaluations: a) Evaluation of the BA3 marker to identify B. anthracis in a historical collection of Bacillus sp. isolates. b) Testing of known B. cereus isolates using the B. anthracis Real-time PCR Assay. c) Testing of known bacterial isolates using the B. anthracis Real-time PCR Assay. Repeatability in clinical matrices was determined through a multicenter study. Specific numerical results are not provided in this summary.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for the test sets in each of the clinical performance evaluations (historical collection of Bacillus sp. isolates, known B. cereus isolates, known bacterial isolates).

    • Data Provenance: The studies utilized "a historical collection of Bacillus sp. isolates," "known B. cereus isolates," and "known bacterial isolates." The direct country of origin for these specific collections is not mentioned, but the submitting organization is the Centers for Disease Control and Prevention (CDC) in the USA, suggesting the data is likely from the USA or samples curated by the CDC. The studies appear to be retrospective as they involve testing historical and known isolate collections.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not specify the "number of experts" or their "qualifications" involved in establishing the ground truth for the test set.

    4. Adjudication Method for the Test Set

    The document does not describe an adjudication method (such as 2+1 or 3+1) for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not done. This device is an in vitro diagnostic test (a PCR assay), not an AI-assisted diagnostic tool that would typically involve human readers interpreting results.

    6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

    Yes, a standalone performance study was done. The B. anthracis Real-time PCR Assay is an automated PCR-based diagnostic test. Its performance characteristics (Limit of Detection, Analytical Sensitivity, Analytical Specificity, and Clinical Performance) were established by evaluating the assay's ability to detect B. anthracis DNA in various samples. This represents the intrinsic performance of the diagnostic algorithm/assay itself, without direct human interpretation of raw data beyond reading the assay's output.

    7. Type of Ground Truth Used

    The ground truth for the clinical performance evaluations was established using a "battery of tests performed by the CDC, including the gamma phage lysis, conventional culture, microbiological, and biochemical testing." This indicates the use of expert consensus based on established microbiological and biochemical methods as the reference standard.

    8. Sample Size for the Training Set

    The document does not specify a "training set" or its sample size. As a PCR assay, its development typically involves primer and probe design and optimization, which isn't usually framed in terms of a "training set" in the same way machine learning models are. The performance characteristics were established using "in-house and multicenter sensitivity studies" and evaluations on historical and known bacterial isolate collections.

    9. How the Ground Truth for the Training Set Was Established

    Since the concept of a "training set" in the context of machine learning isn't directly applicable here, the method of establishing ground truth for development/optimization wasn't explicitly stated. However, similar to the test set, it would logically involve established microbiological and molecular techniques to confirm the identity of B. anthracis and other Bacillus species used during the assay's development and validation.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131930

    Validate with FDA (Live)

    Date Cleared
    2013-08-05

    (39 days)

    Product Code
    Regulation Number
    866.3045
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Device Name and Classification:

    Trade Name: JBAIDS Anthrax Detection Kit Regulation Number: 21 CFR 866.3045
    August 5, 2013

    Re: K131930

    Trade/Device Name: JBAIDS Anthrax Detection Kit Regulation Number: 21 CFR 866.3045

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Joint Biological Agent Identification and Diagnostic System (JBAIDS) Anthrax Detection System is a real-time polymerase chain reaction (PCR) test system intended for the qualitative in vitro diagnostic (IVD) detection of target DNA sequences on the pXO1 plasmid (Target 1) and the pXO2 plasmid (Target 2) from Bacillus anthracis. The system can be used to test the following:

    • Human whole blood collected in sodium citrate from individuals suspected of having anthrax
    • Positive blood cultures
    • Cultured organisms grown on blood agar plates.
      The JBAIDS Anthrax Target 2 assay is used as a supplementary test only after a positive result with the Target 1 Assay.
      The JBAIDS Anthrax Target 1 and Target 2 Assays are run on the JBAIDS instrument using the Diagnostic Wizard. Results are for the presumptive identification of B. anthracis, in conjunction with culture and other laboratory tests. The following considerations also apply:
    • The diagnosis of anthrax infection must be made based on history, signs, symptoms, exposure likelihood, and other laboratory evidence, in addition to the identification of pXO1 and pXO2 targets either from cultures or from direct blood specimens.
    • The assays have not been evaluated with blood from individuals without clinical signs or symptoms who were presumed exposed and who subsequently developed anthrax (inhalation or other forms of the disease), or from individuals with any form of anthrax (inhalational, cutaneous, or gastrointestinal).
    • The level of plasmid targets that would be present in blood from individuals with early systemic infection is unknown.
    • The definitive identification of B. anthracis from colony growth, liquid blood culture growth, or from blood specimens requires additional testing and confirmation procedures in consultation with public health or other authorities for whom reports are required.
      The safety and effectiveness of other types of tests or sample types (not identified as "For in vitro diagnostic use") have not been established.
    Device Description

    The Joint Biological Agent Identification and Diagnostic System (JBAIDS) Anthrax Detection System is a fully integrated IVD system composed of the portable JBAIDS instrument, laptop computer and software, the JBAIDS Anthrax Detection Kit with two different freeze-dried PCR assays for detection of pathogenic Bacillus anthracis DNA. The system has been validated using four different sample preparation kits for isolating DNA from whole blood (IT 1-2-3TM Platinum Path, QFLOWdma, FLOW Sample Purification Kits), positive blood cultures (IT I-2-3TM SWIPE Sample Purification Kit), and plate cultures (IT 1-2-3™ Platinum Path and SWIPE Sample Purification Kits). Use of the JBAIDS DNA Extraction Control Kit is also recommended.
    Prior to testing, specimens are processed using BioFire Diagnostic's IT 1-2-3 Sample Purification Kits. The resulting purified sample is added to Target 1 Unknown and Target 1 Inhibition Control vials, along with reconstitution buffer. Target 1 Positive Control and Negative Control vials are prepared using reconstitution buffer and water. When B. anthracis DNA is present, a fragment of B. anthracis DNA is amplified. The amplicon is detected by fluorescence using a specific hydrolysis probe. Each probe is labeled on one end with a fluorescent reporter moiety (6-carboxyfluorescein (6-FAM)) and elsewhere with a quencher moiety (carboxy tetramethylrhodamine (TAMRA)). When the probe is intact, the quencher absorbs the light emitted by the reporter moiety. During PCR, the probe hybridizes to the target sequence before the exonuclease activity of Taq polymerase hydrolyzes the probe, separating the fluorophore from the quencher and permitting detection of the fluorescent signal generated by the reporter. The fluorescent signal increases as additional templates are amplified and more probes are hydrolyzed.
    JBAIDS Software analyzes the fluorescence amplification curves and reports results as positive, negative, uncertain or inhibited. A failure of the Positive or Negative Control will result in the entire run being called invalid. Retesting is required to resolve uncertain, invalid or inhibited results. The Target 2 assay is used as a supplementary test only after a positive result is obtained with the Target 1 assay.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission primarily focuses on demonstrating equivalence of a new DNA extraction method (Platinum Path) to an existing one (QFLOWdna) for use with the JBAIDS Anthrax Detection Kit. Therefore, the "acceptance criteria" are implicitly defined by the performance of the predicate device with the QFLOWdna kit, and the "reported device performance" is how the new Platinum Path kit compares.

    Performance MetricAcceptance Criteria (Implicit from Predicate/QFLOWdna)Reported Device Performance (with Platinum Path)
    Clinical Performance (Surrogate Whole Blood)
    Positive Percent Agreement (PPA)Equivalent to QFLOWdna (presumably near 100% for positive samples)100% (50/50; 95% CI = 92.9-100%)
    Negative Percent Agreement (NPA)Equivalent to QFLOWdna (presumably near 100% for negative samples)100% (50/50; 95% CI = 92.9-100%)
    Limit of Detection (LoD)Previously established LoD of 1000 CFU/mL in whole blood (for Target 1 and Target 2)1000 CFU/mL (20/20 detected for both targets)
    ReproducibilityConsistent performance across sites for medium positive, low positive, and negative samples100% agreement with expected results for all sites and spike levels, for both Target 1 and Target 2
    Direct Culture Samples (B. anthracis)100% detection of B. anthracis colonies10/10 B. anthracis colonies detected
    Direct Culture Samples (Non-B. anthracis)0% detection of non-B. anthracis colonies0/10 non-B. anthracis colonies detected

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Performance (Surrogate Whole Blood):

      • Sample Size: 100 surrogate whole blood specimens (50 spiked with B. anthracis, 50 not spiked). Each sample was processed by both new (Platinum Path) and old (QFLOWdna) extraction methods and then tested.
      • Data Provenance: Prospectively collected from febrile volunteers in the USA (November 2012 to April 2013). These were surrogate specimens spiked with inactivated B. anthracis, not true clinical anthrax samples.
    • Limit of Detection:

      • Sample Size: 20 independent whole blood specimens, each spiked with B. anthracis at the LoD level.
      • Data Provenance: Not explicitly stated, but implied to be laboratory-prepared samples for analytical validation.
    • Reproducibility:

      • Sample Size: A panel of 12 blood samples (4 medium positive, 4 low positive, 4 negative). Each sample was tested twice a day for four days at each of three testing sites. Total individual tests for each target would be 12 samples * 2 tests/day * 4 days * 3 sites = 288 for each target (Target 1 and Target 2).
      • Data Provenance: Multicenter study, locations are "Site 1, Site 2, Site 3." Country of origin not explicitly stated, but typically assumed to be domestic for FDA submissions unless otherwise specified.
    • Detection of Direct Culture Samples:

      • Sample Size: 10 B. anthracis colonies and 10 non-B. anthracis colonies.
      • Data Provenance: Not explicitly stated, but implied to be laboratory-prepared cultures.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The ground truth for the surrogate clinical specimens and analytical studies was established by:

    • Spiking Status: Knowing which samples were spiked with inactivated B. anthracis and at what concentration, and which were not. This is a controlled laboratory process, not dependent on human expert interpretation for definitive classification.
    • Predicate Device Performance: For the surrogate clinical specimens, the "correct result" was defined by the JBAIDS result when processed with the QFLOWdna kit, which is the predicate/established method. This avoids the need for external expert consensus on the B. anthracis status of a sample that has already been spiked.

    Therefore, no external clinical experts were used to establish the ground truth in the traditional sense for these studies. The ground truth was based on the controlled spiking of samples and comparison to an established method.

    4. Adjudication Method for the Test Set

    Not applicable in the traditional sense. The "ground truth" for the surrogate clinical specimens was determined by the results from the predicate extraction method (QFLOWdna), and for the analytical studies, it was determined by the known spiking status of the samples. There was no need for expert adjudication of conflicting results as the comparison was against a predetermined "truth."

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This device is a molecular diagnostic test for B. anthracis DNA, not an imaging device that would typically involve human readers interpreting images. The study focuses on the analytical and clinical (surrogate) performance of the extraction method and PCR kit.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies presented are standalone (algorithm only). The JBAIDS system, with its software, analyzes the fluorescence amplification curves and reports results (positive, negative, uncertain, inhibited). The operators were blinded to the analyte content, meaning their human "interpretation" was limited to running the instrument and following its automated output. The performance metrics (PPA, NPA, LoD, reproducibility) are based directly on the device's output.

    7. The Type of Ground Truth Used

    • Clinical Performance (Surrogate Whole Blood): Known spiking status of the samples from a laboratory perspective, and the performance of the predicate DNA extraction method (QFLOWdna) as the "correct result" for comparison.
    • Limit of Detection: Known concentration of spiked B. anthracis in laboratory-prepared whole blood specimens.
    • Reproducibility: Known concentration of spiked B. anthracis (medium, low, or unspiked).
    • Detection of Direct Culture Samples: Known identity of the bacterial colonies (either B. anthracis or non-B. anthracis) through laboratory culture and identification.

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" for the device itself. This submission is for modifications to an existing device (the JBAIDS Anthrax Detection Kit) by adding compatibility with a new sample purification kit (IT 1-2-3 Platinum Path). The JBAIDS system and its algorithm were presumably developed and validated previously. The studies described here are a validation of the new sample preparation method, not the training of a new algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no explicit mention of a "training set" for a new algorithm within this submission, this question is not directly applicable. The JBAIDS Anthrax Detection System already existed and was previously cleared (K051713 and K071188). The ground truth for its initial development and validation would have followed similar principles of known positive and negative samples, likely through spiking and confirmed cultures, as is typical for molecular diagnostic assays.