Search Filters

Search Results

Found 26 results

510(k) Data Aggregation

    K Number
    K222559

    Validate with FDA (Live)

    Date Cleared
    2023-03-24

    (212 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    ----------------------------|-------------------|
    | MDB | Class I, reserved | 21 CFR 866.2560
    |
    | Regulation | 21 CFR § 866.2560

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BD BACTEC Myco/F Lytic culture medium when used with the BD BACTEC fluorescent series instruments is a nonselective culture medium to be used as an adjunct to aerobic blood culture media for the qualitative culture and recovery of mycobacteria, yeast and fungi from blood. This medium may also be used for the culture of sterile body fluids when veast or fungi are suspected.

    Device Description

    BD BACTEC Myco/F Lytic culture medium is a Middlebrook 7H9 and Brain Heart Infusion broth formulation for the recovery of mycobacteria from blood specimens, and yeast and fungi from blood and sterile body fluids. Specific modifications were made to enhance the growth and recovery of mycobacteria, yeast and fungi. These modifications include ferric ammonium citrate to provide an iron source for specific strains of mycobacteria and fungi, the addition of saponin as a blood lysing agent, and the addition of specific proteins and sugars to provide nutritional supplements. Each vial contains a sensor which can detect decreases in oxygen concentration in the vial resulting from microorganism metabolism and growth. The sensor is monitored by the BD BACTEC fluorescent series instrument for increasing fluorescence, which is due to the decrease in oxygen. A positive determination indicates the presumptive presence of viable microorganisms in the vial.

    BD BACTEC Myco/F Lytic Culture Vials are supplied in a carton containing 50 vials. It is a nonsterile product.

    AI/ML Overview

    This document is a 510(k) Substantial Equivalence Determination Decision Summary for the BD BACTEC Myco/F Lytic Culture Vials (plastic). It evaluates the device's performance against a predicate device (BD BACTEC Myco/F Lytic Culture Vials (glass)) to determine if it is substantially equivalent. The studies described are analytical performance comparisons rather than clinical trials with human subjects for diagnostic accuracy.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a bulleted or numbered list with corresponding reported performance values within a formal table. Instead, it presents performance data for the modified (plastic) device compared to the predicate (glass) device across various analytical parameters. The implicit acceptance criteria appear to be substantial equivalence, meaning the performance of the new plastic vial is "equivalent to or better than" the predicate glass vial.

    Here's a synthesized table based on the provided data, comparing the plastic device's performance to the predicate:

    Table: Summary of Device Performance Compared to Predicate (Implicit Acceptance Criteria: Equivalent to or Better)

    Performance ParameterAcceptance Criteria (Implied)Reported Device Performance (Plastic vs. Glass)
    ReproducibilityNo statistically significant differences in TTD or % recovery.Met: No statistically significant differences in organism detection times or percent recovery between three lots for blood and sterile body fluid volumes.
    Microbial Detection Limit (MDL) - BloodEquivalent or better performance (ratio of positive yields close to 1 or higher) at low inoculum levels.Met (mostly): Ratio of positive yields close to 1 or higher for most organisms, except Blastomyces dermatitidis (0.33). This organism's inconsistent growth was noted as a limitation.
    Microbial Detection Limit (MDL) - Sterile Body FluidEquivalent or better performance (ratio of positive yields close to 1 or higher) at low inoculum levels.Met (mostly): Ratio of positive yields close to 1 or higher for most organisms, except Blastomyces dermatitidis (0.86). This organism's inconsistent growth was noted as a limitation.
    Delayed Entry in blood (Mycobacteria)100% recovery showed in delayed conditions.Met: 100% recovery for all 8 mycobacterium strains evaluated across various delay times (12-96 hours) and temperatures (20-37.5°C). The Product Insert still recommends prompt placement but provides this information.
    Time to Detection (TTD) - BloodEquivalent or better performance (95% CI for median TTD difference contains zero or only negative values for shorter TTD).Met (mostly): The plastic device performed equivalently to or better than the predicate for most conditions. Exceptions noted were for Dimorphic Fungi (0-1 CFU/vial, 1 mL blood) and Mycobacteria (1-10 CFU/vial, 1mL blood) where median TTD for plastic was longer, but supporting statistical analysis found performance equivalent or better for other conditions.
    Time to Detection (TTD) - Sterile Body Fluid (SBF)Equivalent or better performance (95% CI for median TTD difference contains zero or only negative values for shorter TTD).Met (mostly): The plastic device performed equivalently to or better than the predicate for most conditions. Exceptions noted for Dimorphic Fungi under some test conditions.
    Percent Recovery (Detection) - Blood (10-100 CFU/vial)Comparable recovery rates (ratio of positive yields close to 1).Met: Plastic recovery 99.52% vs. Glass 100%. Ratio of positive yields was 1.00. One instance of negative result only in plastic vial (Blastomyces dermatitidis), noted as a limitation.
    Percent Recovery (Detection) - SBF (10-100 CFU/vial)Comparable recovery rates (ratio of positive yields close to 1).Met: Plastic recovery 100% vs. Glass 99.24%. Ratio of positive yields was 1.01. One instance of negative result only in glass vial (Candida auris).
    False Positive Rate - BloodLower or comparable to predicate.Met (superior): Plastic: 0.70% (all volumes combined) vs. Glass: 25.87%. Majority of false positives in glass were at higher blood volumes.
    False Positive Rate - SBFLower or comparable to predicate.Met (lower): Plastic: 0.47% (all SBF types combined) vs. Glass: 0.71%.
    False Negative Rate - BloodComparable rates (not statistically significant difference).Met: Plastic: 0.65% (4/619) vs. Glass: 0.48% (3/619). Difference not statistically significant (P = 1).
    False Negative Rate - SBFComparable rates (not statistically significant difference).Met: Plastic: 1.52% (6/396) vs. Glass: 1.01% (4/396). Difference not statistically significant (P = 0.7523).
    Instrument Compatibility - BloodEquivalent performance (similar TTD, comparable recovery).Met: All four instruments (FX, FX-40, 9240, 9050) showed 100% recovery for both plastic and glass vials across organisms and volumes (with one exception for plastic at 5mL on FX-40 (94.44% vs 100% for glass)). TTD results were also comparable with median differences near zero.
    Instrument Compatibility - SBFEquivalent performance (similar TTD, comparable recovery).Met: All four instruments showed 100% recovery for both plastic and glass vials across organisms and SBF volumes. TTD results were also comparable with median differences near zero.

    2. Sample Size Used for the Test Set and Data Provenance

    The studies described are analytical performance studies performed in a laboratory setting, rather than clinical studies with patient samples. The samples used are seeded samples (i.e., known microorganisms inoculated into blood or sterile body fluid).

    • Test Set (Seeded Samples):
      • Reproducibility: Not explicitly stated as a separate "test set" size, but involved testing across three lots of media with varying blood/SBF volumes and inoculum levels.
      • Microbial Detection Limit (MDL) - Blood:
        • Sample Size: 414 paired vials (plastic vs. glass). 2 pairs discarded due to contamination, resulting in 412 paired vials. (23 organisms x 3 lots x 3 blood vol x 2 inoculum levels x 1 instrument).
        • Data Provenance: This is in vitro analytical data obtained from controlled laboratory experiments, not from human patients or a specific country of origin.
      • Microbial Detection Limit (MDL) - Sterile Body Fluid (SBF):
        • Sample Size: 264 paired vials (plastic vs. glass). (3 organisms x 3 lots x 4 SBF volumes x 2 inoculum levels x 1 instrument x 2 SBF types) + (5 organisms x 3 lots x 4 SBF volumes x 2 inoculum levels x 1 instrument x 1 SBF type).
        • Data Provenance: In vitro analytical data.
      • Delayed Entry in blood (Mycobacteria):
        • Sample Size: Not explicitly stated as a total paired set number, but involved 8 strains of mycobacterium. Total number of vials for each delay condition (e.g., 72/72 for no delay, 70/70 for 12h @ 20-25C, etc.) is provided in Table 7. Each test used 3 lots and 1, 3, or 5 mL human blood.
        • Data Provenance: In vitro analytical data using human blood matrix.
      • Percent Recovery (Detection) - Blood (10-100 CFU/vial):
        • Sample Size: 207 paired sets (plastic vs. glass). (23 organisms x 3 lots x 3 blood volumes x 1 instrument).
        • Data Provenance: In vitro analytical data using blood matrix.
      • Percent Recovery (Detection) - SBF (10-100 CFU/vial):
        • Sample Size: 132 paired sets (plastic vs. glass). (3 organisms x 3 lots x 4 SBF volumes x 1 inoculum level x 1 instrument x 2 SBF types) + (5 organisms x 3 lots x 4 SBF volumes x 1 inoculum level x 1 instrument x 1 SBF type).
        • Data Provenance: In vitro analytical data using SBF matrix.
      • False Positive Rate - Blood:
        • Sample Size: 144 paired sets initially, 143 after contamination exclusion. (8 vials x 3 blood volumes x 3 lots x 2 instruments).
        • Data Provenance: In vitro analytical data using fresh blood.
      • False Positive Rate - SBF:
        • Sample Size: 432 paired sets initially, 425/424 after contamination exclusions. (8 vials x 3 SBF volumes x 3 lots x 2 instruments x 3 SBF types).
        • Data Provenance: In vitro analytical data using sterile body fluid.
      • False Negative Rate - Blood:
        • Sample Size: 619 combined paired sets from Recovery and MDL studies.
        • Data Provenance: In vitro analytical data.
      • False Negative Rate - SBF:
        • Sample Size: 396 combined paired sets from Recovery and MDL studies.
        • Data Provenance: In vitro analytical data.
      • Instrument Compatibility - Blood:
        • Sample Size: 54 paired sets for FX and 9050 each, 53 paired sets for FX-40 and 9240 each (total 214-216 paired sets for blood). (6 organisms x 3 blood volumes x 1 inoculum x 3 lots / 4 instruments).
        • Data Provenance: In vitro analytical data.
      • Instrument Compatibility - SBF:
        • Sample Size: 48 paired sets per instrument (total 192 paired sets for SBF). (4 organisms x 4 SBF volumes x 1 inoculum x 3 lots / 4 instruments).
        • Data Provenance: In vitro analytical data.

    All data described is retrospective in the sense that it was collected as part of a pre-market submission process, likely after the plastic vial formulation was developed. It is not prospective clinical data from patient studies. The data provenance is internal laboratory studies, not patient data from a specific country.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The ground truth for these analytical studies is based on controlled, seeded experiments where the presence and identity of the microorganisms are known.

    • The "ground truth" is the known inoculum (CFU/vial) of specific ATCC strains or clinical isolates of mycobacteria, yeast, and fungi.
    • Detection by instruments (BD BACTEC fluorescent series instruments) is validated against the known presence/absence of these seeded organisms and further confirmed by subculture for discordant results (e.g., false positives/negatives), which is a standard microbiological method.
    • No human experts (like radiologists) are mentioned or typically involved in establishing ground truth for in vitro diagnostic (IVD) culture media performance studies. The ground truth is intrinsically tied to the experimental design – what was precisely inoculated and subsequently confirmed by laboratory methods.

    4. Adjudication Method for the Test Set

    Adjudication, in the context of IVD performance, would typically involve resolving discrepancies between a device's result and a reference method/truth.

    • For these studies, the primary comparison is the Time-to-Detection (TTD) and Percent Recovery of the plastic vial versus the glass predicate vial against the known seeded inoculum.
    • For false positive/negative rates, discrepancies between instrument reads and definitive culture (subculture) results are identified. The report explicitly mentions how these were defined:
      • False Positive: "instrument positive but subculture negative."
      • False Negative: "instrument-negative at the end of protocol yet contains viable organisms upon subculturing onto appropriate culture media."
    • There's no mention of a multi-observer or consensus-based adjudication process as one would see in image interpretation studies. The "adjudication" is inherent in the design of comparing instrument readouts to known inputs and standard microbiological subculture confirmation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done.

    • MRMC studies are typically performed for medical imaging devices where human readers interpret images, and the AI's impact on their performance is evaluated.
    • This submission concerns an in vitro diagnostic culture medium, which is an automated system for microbial growth detection. The "reader" is the BD BACTEC fluorescent series instrument, not a human.
    • Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable here. The comparison is between two different types of culture vials (plastic vs. glass), both processed by the same automated instruments.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in essence, standalone performance was evaluated.

    • The device being assessed is the BD BACTEC Myco/F Lytic Culture Vial (plastic), which operates when placed into the BD BACTEC fluorescent series instruments.
    • The performance metrics (TTD, Percent Recovery, False Positive/Negative rates) are generated by the instrument's automated detection algorithm based on the changes in fluorescence caused by microbial growth in the vial.
    • Humans are involved in inoculating the vials and performing subcultures for confirmation, but the "performance" described is the direct, automated output of the integrated vial-instrument system. It is not an "AI" in the sense of an image interpretation algorithm, but an automated detection system where the organism's metabolism directly causes a measurable signal change detected by the instrument's programming (analogous to an algorithm).

    7. The Type of Ground Truth Used

    The ground truth used primarily in these analytical studies is an expert-determined, controlled, seeded panel.

    • This involves known strains of mycobacteria, yeast, and fungi (e.g., ATCC strains and clinical isolates) at specified inoculum levels (CFU/vial).
    • For discordant results (e.g., instrument negative but suspected positive), subculture onto appropriate culture media plates was used to confirm the presence of viable organisms, acting as a definitive microbiological reference method for the seeded content.
    • It does not utilize pathology reports (which are for tissue diagnosis) or outcomes data (which would be clinical outcomes in patients).

    8. The Sample Size for the Training Set

    The document does not explicitly mention a "training set" or "training data" in the conventional machine learning sense.

    • This is an IVD device validation, not an AI/ML software validation.
    • The comparison is between a new device (plastic vial) and a predicate device (glass vial), demonstrating substantial equivalence based on a series of analytical performance studies.
    • The "training" of the instrument's underlying detection algorithm (which interprets the fluorescent signals) would have occurred during its initial development and prior regulatory submissions (e.g., for K970333, K970512 for the predicate device). The current submission focuses on demonstrating that the change in vial material (plastic vs. glass) does not adversely affect this established performance.

    9. How the Ground Truth for the Training Set Was Established

    As noted above, there is no explicitly defined "training set" for the vial in the context of this submission. The ground truth for the instrument's core detection capabilities would have been established historically during its development. This would also involve known, quantifiable microbial cultures and their characteristic growth patterns and metabolic activity in the media over time, likely confirmed by standard agar plate cultures or other gold standard microbiological methods.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222591

    Validate with FDA (Live)

    Date Cleared
    2023-03-24

    (210 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    21152

    Re: K222591

    Trade/Device Name: BD BACTEC Plus Aerobic/F Culture Vials Regulation Number: 21 CFR 866.2560
    -------------------------------------|-----------------|
    | MDB | Class I | 21 CFR § 866.2560

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BD BACTEC™ Plus Aerobic/F Culture Vials are used in a qualitative procedure for the aerobic culture and recovery of microorganisms (bacteria and yeast) from blood. The principal use of this medium is with the BD BACTECTM fluorescent series instruments.

    Device Description

    The BD BACTECTM Plus Aerobic/F Culture Vials contain a bacterial growth medium intended for use in the qualitative culture and recovery of aerobic microorganisms (bacteria and yeast) from blood. It has been designed for blood volumes of three (3) to ten (10) milliliters and is used specifically with the BD BACTEC™ fluorescent-series instruments in monitoring of clinical blood specimens for the presence of microorganisms.

    BD BACTECTM Plus Aerobic/F Culture Vials are supplied in a carton containing 50 vials. It is a nonsterile product.

    AI/ML Overview

    This document describes the performance assessment of the BD BACTEC™ Plus Aerobic/F Culture Vials, a modified device, against a predicate device (K113558) to establish substantial equivalence.

    Here's a breakdown of the acceptance criteria and study details:

    1. Table of acceptance criteria and the reported device performance:

    The document doesn't explicitly state quantitative acceptance criteria in a dedicated table format with precise thresholds. Instead, it describes comparative "acceptance criteria" through statements of "no statistically relevant difference" or "no significant differences" between the modified and predicate devices for various analytical performance metrics.

    However, based on the study findings, we can infer the implicit acceptance criteria:

    Acceptance Criteria (Inferred)Reported Device Performance (Summary from Tables 2, 3, 4, 5, 6)
    Instrument Time to Detection (TTD): No statistically relevant difference in TTD between the modified and predicate device.For the majority of organisms and conditions, the TTD difference between the modified and predicate device was very small, with 95% confidence intervals often spanning zero or indicating minor differences. Two isolates (Rothia mucilaginosa and Stenotrophomonas maltophilia) initially showed >10% difference in favor of the predicate, but further testing demonstrated this was strain-specific, not species-specific. The overall bootstrap analysis across all organisms showed negligible differences in TTD (e.g., -0.002 to -0.168 hours).
    Percent Recovery (Sensitivity): No statistically relevant difference in recovery between the modified and predicate device.Across all organisms, inoculum levels, and blood volumes, the percent recovery for both the predicate and modified devices was overwhelmingly 100%. In cases of lower inoculum levels (0-1 CFU), where recovery rates were expectedly lower, the differences between the two devices were generally small and varied in which device had slightly higher recovery, indicating no consistently better or worse performance. The overall recovery rates for 10-100 CFU were 100% for both devices.
    False Negative Rate: No false negative vials observed.Reported: 0% false negative vials for both devices. All instrument-negative vials did not yield growth upon terminal subculture.
    False Positive Rate: No false positive vials observed.Reported: 0% false positive vials for both devices. All vials inoculated with sterile human blood remained negative.
    Instrument Compatibility: No significant differences in TTD and recovery across different BACTEC™ Fluorescent Series Instruments.Reported: No recovery failures in any of the instruments (BACTEC™ FX, FX40, 9240, 9050) for either device. No significant TTD differences observed across the instruments (e.g., TTD differences generally close to zero with narrow 95% CIs).
    Resin Performance (Antimicrobial Neutralization Capability): No statistically significant difference in recovery between lot pairs for different antimicrobial/organism combinations.Reported: Minor variations in percent recovery between individual lots of the predicate and modified devices (e.g., Predicate Lot A: 47.77%, Modified Lot A: 47.13%). The combined data for all lots showed a very similar overall recovery (Predicate: 48.78% vs. Modified: 48.78% based on the provided table). The document explicitly states "There was no statistically significant difference observed between lot pairs."

    2. Sample sized used for the test set and the data provenance:

    • Sample Size (Test Set):

      • Instrument Time to Detection & Percent Recovery:
        • For 44 organisms at 10-100 CFU/vial: 264 vials per device (3 lots x 2 blood volumes x 44 organisms).
        • For 15 organisms at 0-1 and 1-10 CFU/vial: 45 vials per device per inoculum level per blood volume on BACTEC™ FX, totaling 90 vials per device for these specific conditions.
        • Total vials for TTD/Recovery: Approximately 444 vials per device (264 + 90 + 90).
      • False Negative Rate: "115 instrument negative vials (57 in the modified device and 58 in the predicate device)" out of the 444 total vials from the TTD/Recovery study that were instrument negative.
      • False Positive Rate: 120 vials per device (3 lots x 8 replicates x 5 blood volumes).
      • Instrument Compatibility:
        • BACTEC™ FX and FX40: 264 paired sets per instrument (total 528 paired sets).
        • BACTEC™ 9240 and 9050: 96 paired sets per instrument (total 192 paired sets).
      • Resin Performance: 314 antibiotic/organism combinations per lot per device. With 3 lots per device, this amounts to 942 vials per device. Overall, 1884 vials for this study.
    • Data Provenance: The studies were retrospective analytical performance studies, comparing a modified device to a previously cleared predicate device. The information does not specify the country of origin of the data, but given it's an FDA submission for a Becton, Dickinson and Company product, it is likely based on studies conducted in the US or in compliance with US regulatory standards. The testing involved controlled laboratory conditions using spiked (inoculated) blood samples, not clinical patient samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    This study is an analytical performance study for a microbial growth monitor. The "ground truth" for the test set (i.e., whether growth occurred and what the TTD was) was established by:

    • Direct Inoculation: Precise quantities (CFU/vial) of known bacterial and yeast organisms were inoculated into the vials.
    • Instrument Detection: The BD BACTEC™ fluorescent series instruments automatically detect and log growth (CO2 production) as part of their function.
    • Terminal Subculture: For vials that remained negative, a terminal subculture was performed to confirm the absence of viable microorganisms.

    Therefore, this type of study does not rely on human experts to "establish ground truth" in the way a diagnostic imaging study might require radiologists. The ground truth is determined by the experimental design (known inoculum) and the objective readings of the instrument, verified by standard microbiological techniques (subculture).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. As described above, the ground truth is established by controlled inoculation and objective instrument readings/terminal subculture, not by human interpretation or adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is an analytical performance study of a microbial growth medium and instrument system, not a clinical study involving human readers interpreting diagnostic images or data with or without AI assistance. The device is not an AI-powered system that assists human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The device (BD BACTEC™ Plus Aerobic/F Culture Vials) is a component of an automated system (BD BACTEC™ fluorescent series instruments). The performance evaluated in this submission is effectively the "standalone" performance of the altered vial formulation within the automated BACTEC™ system, without direct human intervention in the detection process itself. Human involvement comes in collecting blood, inoculating vials, and interpreting the instrument's final positive/negative results. The study inherently assesses the algorithm (instrument's detection logic) as part of the overall system performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The ground truth used was controlled laboratory inoculation with known quantities of specific microorganisms, followed by objective instrument detection (fluorescence changes indicating CO2 production) and confirmation by terminal microbiological subculture for negative vials. This is a highly controlled "gold standard" for analytical performance in microbiology.

    8. The sample size for the training set:

    Not applicable. This is an analytical performance study comparing a modified in vitro diagnostic device component to a predicate device. It does not involve a "training set" in the context of machine learning model development. The study design is focused on demonstrating functional equivalence through empirical testing.

    9. How the ground truth for the training set was established:

    Not applicable, as there was no training set in the context of machine learning.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190405

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-05-15

    (85 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Missouri 63042

    Re: K190405

    Trade/Device Name: BACT/ALERT MP Reagent System Regulation Number: 21 CFR 866.2560
    Regulation Section: | 21 CFR 866.2560
    | Classification Name: Blood CulturingSystem - MicrobiologyRegulation Number: 21CFR866.2560
    | Classification Name: BloodCulturing System - MicrobiologyRegulation Number: 21CFR866.2560

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BACT/ALERT® MP Reagent System consists of the BACT/ALERT® MP culture bottle with a removable closure used in conjunction with the BACT/ALERT® MP Antimicrobial Supplement and the BACT/ALERT® MP Nutrient Supplement. The BACT/ALERT® MP Reagent System is designed for use with BACT/ALERT® 3D Mycobacteria Detection Systems for recovery and detection of mycobacteria from sterile body specimens other than blood, and from digested/decontaminated clinical specimens.

    Device Description

    BACT/ALERT® MP Reagent System consists of three reagents: BACT/ALERT® MP Culture Bottle, the BACT/ALERT® MP Antimicrobial Supplement and the BACT/ALERT® MP Nutrient Supplement.

    AI/ML Overview

    The provided text describes the performance characteristics of the BACT/ALERT® MP Reagent System, comparing it to a predicate device (BACT/ALERT® MP Culture Bottle) for the recovery and detection of mycobacteria.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state "acceptance criteria" in a table format with pass/fail thresholds. However, it presents the performance of the new device and the predicate device, implying that the new device aims to be equivalent or superior, particularly in Time To Detection. The study demonstrates that the new device has increased sensitivity and specificity and a faster time to detection for certain mycobacteria species.

    For the purpose of this response, I will interpret the performance metrics presented as the "reported device performance" and infer that the device meets implied acceptance criteria by being equivalent or better than the predicate.

    Performance MetricBACT/ALERT® MP Reagent System (REF 419744) (Test Device)BACT/ALERT® MP Culture Bottle (REF 259797) (Predicate Device)Implied Acceptance Criterion (e.g., must be equivalent or better than predicate)
    Analytical Sensitivity: LoD
    M. avium≤ 15 CFU/bottle (100% detection)Not explicitly stated for predicate in LoD sectionLow LoD (e.g., ≤ the predicate device's LoD)
    M. kansasii≤ 4 CFU/bottle (100% detection)Not explicitly stated for predicate in LoD sectionLow LoD (e.g., ≤ the predicate device's LoD)
    M. tuberculosis≤ 14 CFU/bottle (100% detection)Not explicitly stated for predicate in LoD sectionLow LoD (e.g., ≤ the predicate device's LoD)
    Analytical Sensitivity: Growth Performance100% detection at high (10^6^ CFU/bottle) and low (10^2^ CFU/bottle) inoculum levelsNot explicitly stated for predicate100% detection for tested organisms at varying inoculum concentrations
    Within-Laboratory Precision - Recovery Rate100% recovery for all test events and concentrationsNot explicitly stated for predicate100% recovery rate
    Reproducibility - Overall Recovery Rate (Pleural Fluid)99.6%Not explicitly stated for predicateHigh recovery rate (e.g., ≥ 95%)
    Reproducibility - Overall Recovery Rate (Simulated Sputum)99.9%Not explicitly stated for predicateHigh recovery rate (e.g., ≥ 95%)
    Clinical Performance - Sensitivity86.58% (95% CI: 81.50%, 90.70%)81.39% (95% CI: 75.76%, 86.19%)Equivalent or superior to predicate; the test device showed a 5.19% increase in sensitivity.
    Clinical Performance - Specificity96.75% (95% CI: 95.49%, 97.74%)93.79% (95% CI: 92.16%, 95.18%)Equivalent or superior to predicate; the test device showed a 2.96% increase in specificity.
    Clinical Performance - Overall Recovery Rate (Positive Mycobacteria Samples)94.34%88.68%Equivalent or superior to predicate; the test device showed a 5.66% increase.
    False Negative Rate (Clinical Samples)0.13% (2/1,488)Not explicitly stated for predicateLow false negative rate (e.g., ≤ predicate's or a clinically acceptable threshold).
    False Positive Rate (Clinical Samples)2.28% (34/1,488)Not explicitly stated for predicateLow false positive rate (e.g., ≤ predicate's or a clinically acceptable threshold).
    Time to Detection (TTD) - M. tuberculosis complex speciesMean: 11.6 daysMean: 14.4 daysFaster than predicate (demonstrated -2.7 days mean difference).
    Time to Detection (TTD) - Non-tuberculous Mycobacteria Rapid GrowersMean: 7.3 daysMean: 7.4 daysFaster than or equivalent to predicate (demonstrated -0.1 days mean difference, considered faster).
    Time to Detection (TTD) - Non-tuberculous Mycobacteria Slow GrowersMean: 11.5 daysMean: 11.8 daysFaster than or equivalent to predicate (demonstrated -0.3 days mean difference, considered faster).

    2. Sample size used for the test set and data provenance:

    • Clinical Performance Studies (for Sensitivity, Specificity, Recovery Rates):

      • Sample Size: 1,488 specimens obtained from 1,162 patients.
      • Data Provenance: Prospective clinical study conducted at 3 different sites. The document does not specify the country of origin, but generally, FDA submissions involve studies conducted in or for the US market, potentially with multi-national sites.
    • Analytical Performance Studies (LoD, Growth Performance, Precision, Reproducibility):

      • Sample Size: Varies by study (e.g., LoD specified as lowest inoculum level with >95% recovery, Growth Performance at high and low inoculum levels, Precision on a minimum of 3 lots over 12 days, Reproducibility with 5 organisms at low and high concentrations in two matrices). Specific total sample numbers for these analytical studies are not given, but generally involve laboratory-controlled experiments.
      • Data Provenance: In-house seeded studies and tests at clinical trial sites (for reproducibility). The specific country of origin is not mentioned for these internal studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The ground truth for clinical samples appears to be established by laboratory results using both the test and predicate devices, and primary solid medium culture ("a sample has a positive patient infected status when one of the BACT/ALERT® MP (REF 419744 or 259797) subcultures is positive, or the primary solid medium is positive"). This implies a laboratory-based gold standard rather than expert clinical consensus or adjudication on imaging, for example.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    The document does not describe an adjudication method involving multiple human readers or a specific consensus process for reviewing diagnostic results in the test set. The ground truth seems to be determined by a combination of positive results from subcultures of both the test and predicate devices, and primary solid medium culture. This is a laboratory-based ground truth, not one requiring human expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This device is a microbial growth monitor and not an AI-powered diagnostic imaging tool or a system that requires human "readers" in the context of an MRMC study. Therefore, an MRMC comparative effectiveness study, as typically understood for AI-assisted image interpretation, was not performed and is not applicable here. The comparison is between two laboratory systems (the new device vs. the predicate).

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The device (BACT/ALERT® MP Reagent System) is an automated system for detecting microbial growth. Its performance is inherently standalone in the sense that the instrument provides the detection signal without continuous human intervention for interpretation in the same way an AI algorithm for image analysis would. The "algorithm" here is the system's ability to detect CO2 production. The study evaluates the performance of this system directly.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The ground truth for the clinical performance studies (sensitivity, specificity, recovery) in the document is defined as follows:

    • Positive Patient Infected Status: "A sample has a positive patient infected status when one of the BACT/ALERT® MP (REF 419744 or 259797) subcultures is positive, or the primary solid medium is positive."
    • Negative Patient Infected Status: "A sample has a negative patient infected status when both BACT/ALERT® MP (REF 419744 and 259797) subcultures are negative and the primary solid media is negative (i.e. there is no growth of mycobacteria)."

    This is a laboratory-based and culture-confirmed ground truth, relying on the isolation and identification of mycobacteria from subcultures and primary solid media.

    8. The sample size for the training set:

    The document does not mention a training set because this is a medical device for in-vitro diagnostics based on chemical reactions and optical detection, not typically a machine learning or AI algorithm that requires a separate training set. The "formulation changes" and "analytical performance studies" represent the development and validation of the reagent system itself.

    9. How the ground truth for the training set was established:

    As no training set is mentioned or applicable in the context of this device's type, there is no information on how ground truth for a training set was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183166

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-02-11

    (88 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    63042

    Re: K183166

    Trade/Device Name: BacT/ALERT FA Plus; BacT/ALERT PF Plus Regulation Number: 21 CFR 866.2560
    |
    | Classification Name: | Blood Culturing System, Microbiology21 CFR 866.2560

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BacT/ALERT® FA Plus Culture Bottles are used with BacT/ALERT® Microbial Detection Systems in qualitative procedures for recovery and detection of aerobic microorganisms (bacteria and yeast) from blood and other normally sterile body fluids.

    BacT/ALERT® PF Plus Culture Bottles are used with BacT/ALERT® Microbial Detection Systems in qualitative procedures for recovery and detection of aerobic microorganisms (bacteria and yeast) from blood.

    Device Description

    The BacT/ALERT® Microbial Detection Systems utilizes a colorimetric sensor and reflected light to monitor the presence and production of carbon dioxide (CO2) dissolved in the culture medium. If microorganisms are present in the test sample, carbon dioxide is produced as the microorganisms metabolize the substrates in the culture medium. When growth of the microorganisms produces CO2 the color of the gas-permeable sensor installed in the bottom of each culture bottle changes from blue-green to yellow. The lighter color results in an increase of reflectance units monitored by the system. Bottle reflectance is monitored and recorded by the instrument every 10 minutes.

    BacT/ALERT® Microbial Detection Systems are used to determine if microorganisms are present in blood or other normally sterile body fluid samples taken from a patient suspected of having bacteremia/fungemia. The BacT/ALERT® Systems and culture bottles provide both a microbial detection system and a culture medium with suitable nutritional and environmental conditions for organisms commonly encountered in blood infections and other normally sterile body fluid infections. An inoculated bottle is placed into the instrument where it is incubated and continuously monitored for the presence of microorqanisms that will grow in the BacT/ALERT® FA Plus or PF Plus Culture Bottle (The BacT/ALERT® PF Plus Culture Bottle provides for detection of microorganisms when a small volume of blood is available.).

    AI/ML Overview

    The provided text describes the performance characteristics of the BacT/ALERT® FA Plus and BacT/ALERT® PF Plus culture bottles, which are microbial detection systems. The study aims to demonstrate the equivalency of a modified formulation of these devices to their predicate devices.

    Here's an analysis of the acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance CharacteristicAcceptance CriteriaReported Device Performance and Notes
    Growth PerformanceRecovery Rate: Equivalent recovery rates based on Newcombe's Hybrid Score. Time to Detection (TTD): Equivalent TTD based on the mean percent difference in TTD.Recovery Rate: Met criteria for equivalency for all 39 microorganisms in both presence and absence of blood. TTD (with blood): Met criteria for 38 of 39 microorganisms. A delay was observed for Haemophilus parainfluenzae, though both adjusted and predicate had <100% detection. TTD (without blood - FA Plus): Met criteria for 32 of 34 microorganisms. S. aureus and S. pneumoniae showed TTD delays of 1.7h and 5h, respectively (no delay with blood). C. albicans and C. glabrata showed faster TTD (3.6h and 36.6h, respectively).
    Antimicrobial NeutralizationNot explicitly stated as a general criterion, but implied to demonstrate neutralization of claimed drug categories.Neutralized penicillins, glycyclines, polyenes, macrolides, triazoles, echinocandins, aminoglycosides, fluoroquinolones, lincosamides, glycopeptides, and oxazolidinones. One echinocandin/microorganism combination (C. albicans/caspofungin) had <100% recovery (77.8% without blood, 97% with blood). Substantial equivalency for ceftaroline, cefoxitin, and cefazolin was observed to the previous formulation. S. aureus/cefoxitin initially showed 66% recovery but 100% on retest. E. coli/cefoxitin showed 100% recovery. Less than complete neutralization for cefazolin (neutralized at 50% peak serum level in PBS). Ceftazidime, cefepime, cefotaxime, and ceftriaxone were not evaluated due to prior incomplete neutralization.
    Limit of Detection (LoD)For each microorganism tested, ≥95% recovery and detection within the 5-day test period.Met LoD criterion for equivalency for all 11 microorganisms tested, with ≥95% recovery and detection within 5 days.
    Delayed EntryPercent recovery for delayed entry-adjusted devices was 100% within 5 days for each claimed holding temperature. False positive rate for uninoculated bottles (without organism) was <3.4%.Met Delayed Entry recovery for all 11 microorganisms tested and the false positive rate criteria.
    Within Laboratory Precision (Repeatability)100% recovery for all organisms/variables tested.Met repeatability criterion with 100% recovery for all organisms/variables tested.

    2. Sample Size Used for the Test Set and Data Provenance

    • Growth Performance: A panel of 39 clinically significant microorganisms.
    • Antimicrobial Neutralization: Representative antimicrobials from claimed drug categories. Specific numbers of combinations are mentioned (e.g., five echinocandin/microorganism combinations, specific antimicrobial/organism pairs).
    • Limit of Detection: A panel of 11 microorganisms listed in the IFU.
    • Delayed Entry: A panel of 11 microorganisms.
    • Repeatability: Multiple organisms and multiple organism/antimicrobial combinations, replicates from three bottle lots.

    Data Provenance: The studies were described as "internal (bioMérieux) verification studies." This suggests the data was generated internally by the device manufacturer. The text does not specify the country of origin of the data or explicitly state if the studies were retrospective or prospective, but "verification studies" implies prospective testing of the modified devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This document describes the performance of a microbial detection system (culture bottles). The "ground truth" here is the actual presence and growth of microorganisms in the culture media, which is determined by the inherent biological phenomenon and observed by the instrument. This type of device does not involve human expert interpretation for "ground truth" establishment in the same way an imaging or diagnostic AI device would. The performance metrics rely on comparison to established microbiological methods or the predicate device's performance.

    4. Adjudication Method for the Test Set

    Not applicable. As described in point 3, the ground truth is based on the biological growth of microorganisms detected by the instrument, not on expert human interpretation that would require adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    Not applicable. This is a microbial detection system, not an AI-assisted diagnostic tool that involves human readers. Therefore, an MRMC study and a human-in-the-loop performance evaluation are not relevant in this context.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The performance data presented here is effectively standalone performance of the device (BacT/ALERT® FA Plus and PF Plus culture bottles with the BacT/ALERT® Microbial Detection Systems). The device functions autonomously to detect microbial growth.

    7. The Type of Ground Truth Used

    The ground truth used in these studies is the microbiological growth and detection of specific microorganisms in the culture bottles. This is established by direct observation of microbial growth (either visually or primarily, through the instrument's detection of metabolic byproducts like CO2) and comparison to expected growth profiles for known inocula.

    8. The Sample Size for the Training Set

    The provided text describes verification studies for a modification to a microbial culture medium. It does not mention a "training set" in the context of machine learning or AI. The modified device's performance is compared against a predicate device and established criteria, not against a model trained on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" mentioned or implied for this device type.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172412

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-05-01

    (264 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Massachusetts 01915

    Re: K172412

    Trade/Device Name: BacterioScan 216Dx System Regulation Number: 21 CFR 866.2560
    BacterioScan™ 216Dx™ System

    Common Name

    Microbial Growth Monitor System

    Product Classification

    866.2560

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BacterioScan 216Dx System is a semi-automated, in vitro diagnostic system (consisting of an instrument, software, and disposable Multicuvettes) that analyzes light scattering to measure any bacterial growth directly from urine sample incubated in trypticase soy broth. The BacterioScan 216Dx is for qualitative determination (presumptive positive or presumptive negative) of bacteriuria at a density of ≥5x104 CFU/mL. The BacterioScan 216Dx results are intended for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of Urinary Tract Infections (UTIs).

    The 216Dx System is not intended to provide bacteriuria levels, bacterial identification, or differentiation. The system does not distinguish between growth of infecting, colonizing or contaminating bacteria, or if mixed urogenital flora are present. Presumptive positive urine samples must be cultured. Presumptive negative urine samples must be cultured if a low level of bacteriuria is suspected and is clinically relevant. The system also does not detect anaerobic bacteria, fungi/yeasts, fastidious organisms or those associated with sterile pyuria.

    Device Description

    The BacterioScan 216Dx System is a semi-automated, in vitro diagnostic system (consisting of an instrument, software, and disposable Multicuvettes) that analyzes light scattering to measure any bacterial growth directly from urine sample incubated in trypticase soy broth. The BacterioScan 216Dx is for qualitative determination (presumptive positive or presumptive negative) of bacteriuria at a density of ≥5x104 CFU/mL. The BacterioScan 216Dx results are intended for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of Urinary Tract Infections (UTIs).

    The BacterioScan 216Dx System uses laser light scattering to measure turbidity (particle density) of a prepared urine-broth sample during a period of incubation. The system is designed to kinetically capture digital measurements of sample turbidity at very low particle densities. Classification software analyzes these optical signal measurements, and interprets increasing turbidity as bacterial growth. Samples with robust growth during the second half of the test cycle, and/or high turbidity, are reported as "Presumptive Positive" for bacteriuria and/or UTI. Low turbidity samples that do not exhibit growth are classified as "Presumptive Negative".

    The user logs into the 216Dx interface and is prompted to associate a patient sample ID with a Multicuvette serial number and well location. The user will then pipette the prepared liquid samples into a Multicuvette. Once the Multicuvette is filled, it is loaded into a 216Dx Instrument. The Multicuvette is then incubated for approximately three (3) hours. During the instrument run, the 216Dx appliance will analyze optical measurements to provide a qualitative determination of the presence or absence of viable bacteria in urine specimens at a cut-off of >5x10 CFU/ml. The Classification algorithm qualitatively assesses a result by analyzing periodic measurements taken throughout this incubation period.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the BacterioScan 216Dx System, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are not explicitly stated as distinct pass/fail thresholds in the document, but rather demonstrated through the reported performance data. The device's performance is presented in terms of sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) for detecting bacteriuria at densities ≥5x10⁴ CFU/mL.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance (Primary Analysis: Bacteriuria ≥5x10⁴ CFU/mL)Reported Device Performance (Secondary Analysis: UTI Associated* ≥5x10⁴ CFU/mL)
    SensitivityHigh (e.g., above 80-90%)89.1% (95% CI: 86.8%; 91.1%)97.7% (95% CI: 96.2%; 98.6%)
    SpecificityModerate to High (e.g., above 70%)74.4% (95% CI: 72.6%; 76.2%)72.0% (95% CI: 70.2%; 73.8%)
    PPVModerate (context dependent)55.3% (95% CI: 52.6%; 58.0%)46.8% (95% CI: 44.1%; 49.6%)
    NPVHigh (e.g., above 90%)95.1% (95% CI: 93.9%; 96.0%)99.2% (95% CI: 98.7%; 99.5%)
    Reproducibility≥95% for High and Moderate Densities91.7-100% across operators/sites/days for High/Moderate densities; 96.5% for negativesN/A
    Limit of Detection (LoD)Approximately equivalent to clinical cut-off (10⁴ CFU/mL)LoD for most common and rare UTI pathogens confirmed to be approximately equivalent to 10⁴ CFU/mL. Some pathogens showed higher LoD and were noted in limitations.N/A

    *Limited to bacterial pathogens encountered in the clinical studies.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set (Method Comparison Study):
      • Overall: 3,153 urine samples enrolled, with 3,052 evaluable specimens for primary analysis of bacteriuria (≥5x10⁴ CFU/mL).
      • For secondary analysis (UTI Associated*), 3,010 evaluable specimens for ≥5x10⁴ CFU/mL and 3,020 for ≥1x10⁵ CFU/mL.
    • Data Provenance: The method comparison study was conducted at three geographically diverse sites. It involved urine samples "received in the laboratory for routine urine culture," suggesting these were clinical samples. The study design implies a prospective collection of samples for the purpose of the comparison. The country of origin is not explicitly stated, but the FDA documentation implies a US context.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The ground truth for the test set was established using "Reference cultures performed simultaneously with 216Dx testing." This involved using 1- and 10μL loops, performing colony count plates in triplicate, and generating a median result. Organisms isolated were identified by MALDI.

    The document does not specify:

    • The number of experts.
    • The qualifications of those experts (e.g., microbiologists, medical technologists).

    4. Adjudication Method for the Test Set

    The ground truth was established by median colony count from triplicate plating, which acts as a form of internal reconciliation/adjudication. There is no mention of external expert adjudication (e.g., 2+1, 3+1).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a TRADITIONAL MRMC study involving human readers with and without AI assistance was not conducted.
    • Instead, this study compared the device (BacterioScan 216Dx System) directly against a reference culture method, not against human interpretation of diagnostic images or data, nor did it involve human readers using or not using the AI to interpret data.
    • Therefore, an effect size of how much human readers improve with AI vs without AI assistance is not applicable to this study.

    6. Standalone Performance

    Yes, the method comparison study directly assesses the standalone performance of the BacterioScan 216Dx System (algorithm only without human-in-the-loop performance) against the reference culture method.

    7. Type of Ground Truth Used

    The ground truth used was microbiological culture with colony count and identification by MALDI, which is a form of pathology/laboratory ground truth.

    8. Sample Size for the Training Set

    The document does not specify the sample size used for the training set. It focuses on the validation studies (e.g., LoD, linearity, reproducibility, and method comparison). The "Classification software analyzes these optical signal measurements" suggests an algorithm, but details about its development (and thus training data) are not provided in this excerpt.

    9. How the Ground Truth for the Training Set Was Established

    Since the document does not mention the training set size, it also does not detail how the ground truth for any training set was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173873

    Validate with FDA (Live)

    Date Cleared
    2018-03-16

    (86 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    F Culture Vials Soybean-Casein Digest Broth with Resins in a Plastic Vial Regulation Number: 21 CFR 866.2560
    media: BD BACTEC™ Peds Plus™/F (in Plastic)

    Regulatory Information

    Regulation section: 21 CFR §866.2560

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BD BACTEC Peds Plus™F culture vials (enriched Soybean-Casein Digest broth with CO2) are for aerobic blood cultures. Principal use is with the BD BACTEC fluorescent series instruments for the qualitative culture and recovery of aerobic microorganisms (mainly bacteria and yeast) from pediatric and non-pediatric blood specimens which are generally less than 3 mL in volume.

    Device Description

    The sample to be tested is inoculated into one or more vials which are inserted into the BD BACTEC fluorescent series instrument for incubation and periodic reading. Each vial contains a chemical sensor which can detect increases in CO2 produced by the growth of microorganisms. The sensor's fluorescence measurement is detected at regular intervals to determine the change in CO2 present in the system. A positive reading indicates the presumptive presence of viable microorganisms in the vial. BD BACTEC Peds Plus™F culture vial detection is limited to microorganisms that will grow in a particular type of medium.

    AI/ML Overview

    The document describes the analytical performance criteria and the study conducted for the BD BACTEC™ Peds Plus™/F Culture Vials.

    Here's the breakdown of the information requested:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance (BD BACTEC Peds Plus/F Plastic Vial vs. Glass Vial)
    Instrument Time to Detection (TTD)Median TTD difference < 10% between test and control devices.Wilcoxon estimated median TTD difference was approximately 30 minutes, favoring the test device. This is < 10% and deemed not clinically relevant.
    Percent RecoveryNo significant difference in percent recovery between test and control devices.Both test and control devices were positive in all 492 paired sets (10-100 CFU). No difference was observed from the control.
    Microbial Detection LimitNo statistically significant difference in recovery between test and control devices.McNemar chi-square analysis showed no statistically significant difference in recovery (p=0.644) between test and control devices.
    False Negative RateNo false negatives observed.No false negatives observed in either the control or test device.
    Antimicrobial NeutralizationNo statistically significant difference in recovery between test and control devices when evaluating antimicrobial neutralization.Recovery was the same in the test and control devices, indicating no statistically significant difference.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Instrument Time to Detection: 672 paired sets evaluated. 596 paired sets were positive in both devices.
    • Percent Recovery: 492 paired sets (from the 672, specifically with 10-100 CFU) evaluated.
    • Microbial Detection Limit: 180 paired sets (from the 672, specifically with 0-1, 1-10 CFU) evaluated.
    • Antimicrobial Neutralization: Representative drug of its class was evaluated at the MIC level for selected strains. (Specific sample size for this part of the study is not explicitly stated as a number of paired sets, but focuses on the number of strains and drug class).

    The data provenance is not explicitly stated regarding country of origin or whether it's retrospective or prospective. However, such studies for regulatory submissions are typically prospective, controlled laboratory studies.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The study described is an analytical performance study comparing two medical devices (plastic vial vs. glass vial) in a laboratory setting, not a clinical study involving human patients where expert readers would establish ground truth for diagnostic interpretations. Therefore, this information is not applicable to the provided text. The "ground truth" here refers to whether microorganisms grew and were detected as per the gold standard laboratory methods.

    4. Adjudication Method for the Test Set:

    Not applicable. This was an analytical performance study, not a clinical study requiring adjudication of expert interpretations.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is not an AI-based device, nor a diagnostic imaging device typically evaluated with MRMC studies.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    The device is a culture vial used with an instrument (BD BACTEC fluorescent series instrument) for automated detection of microbial growth. The performance described is "standalone" from human interpretation in the sense that the instrument provides the detection results. There is no "human-in-the-loop" AI algorithm being evaluated.

    7. The Type of Ground Truth Used:

    The ground truth used for this analytical performance study is the actual growth and detection of microorganisms in the culture vials, confirmed by laboratory methods (i.e., whether growth occurred and was detected by the instrument, and confirmed by potentially subcultures or other microbiological techniques if discrepancies arose, though the document doesn't detail this level of confirmation). The study design compares the performance of a new plastic vial with an existing glass vial (control).

    8. The Sample Size for the Training Set:

    Not applicable. This device is a culture vial, not an AI algorithm that requires a training set. The study focuses on evaluating the analytical performance of the device itself.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161816

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-03-28

    (270 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K161816

    Trade/Device Name: BacT/ALERT® VIRTUO™ Microbial Detection System Regulation Number: 21 CFR 866.2560
    BacT/ALERT® VIRTUO™ Microbial DetectionSystem |
    | Classification Name: | 21 CFR 866.2560

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BacT/ALERT® VIRTUO(TM) Microbial Detection System is an automated microbial test system capable of incubating, agitating, and continuously monitoring for the detection of aerobic microorganism growth from blood and other normally sterile body fluids.

    Device Description

    The VIRTUO Instrument is the next generation of the bioMerieux BacT/ALERT Microbial Detection System. This blood culture instrument consists of an incubator, agitation mechanism, robotic apparatus for automated loading and unloading of bottles and a tactile graphical interface. The VIRTUO is used in conjunction with existing, commercialized BacT/ALERT reagent bottles for clinical use (BacT/ALERT SA, SN, FA Plus, FN Plus and PF Plus culture bottles). The VIRTUO system utilizes a colorimetric sensor and reflected light to monitor the presence and production of carbon dioxide (CO2) dissolved in the culture medium. If microorganisms are present in the inoculated sample, carbon dioxide is produced as the organisms metabolize the substrates in the culture medium. When growth of the microorganisms produces CO2, the color of the gas-permeable sensor installed in the bottom of each culture bottle changes from blue-green to yellow. The color change results in an increase of reflectance units monitored by the system. The VIRTUO optically monitors the reflectance of each bottle over time and will store and interpret readings against algorithms, which are embedded in the firmware and/or software.

    AI/ML Overview

    Here's a breakdown of the BacT/ALERT VIRTUO Bacterial Detection System's acceptance criteria and study proving its performance, based on the provided FDA 510(k) Summary.

    This device is not an AI/ML device in the modern sense (e.g., using deep learning for diagnostic imaging), but rather an automated system that uses embedded algorithms to detect microbial growth. Therefore, some of the requested information regarding AI/ML-specific concepts (like MRMC studies, training set details for complex models) might not be directly applicable or detailed in the provided document. The study performed is a comparative clinical study against a predicate device.


    Device Name: BacT/ALERT® VIRTUO™ Microbial Detection System

    Device Description: An automated microbial test system capable of incubating, agitating, and continuously monitoring for the detection of aerobic, facultative, and anaerobic microorganism growth from blood and other normally sterile body fluids. It utilizes a colorimetric sensor and reflected light to monitor CO2 production as an indicator of microbial growth. Its detection algorithms are proprietary to the system.


    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document doesn't explicitly state the "acceptance criteria" for performance as a numerical threshold (e.g., "sensitivity must be > X%"). Instead, it demonstrates substantial equivalence to the predicate device (BacT/ALERT® 3D Microbial Detection System) by comparing their performance head-to-head. The key performance metric assessed is the "Ratio of True Positives" and its 95% Confidence Interval (CI), with the implicit acceptance being that the performance of the VIRTUO is comparable to the established predicate device, and the false positive/negative rates are within acceptable limits.

    Performance Goal: Demonstrate substantial equivalence to the BacT/ALERT® 3D Microbial Detection System in detecting microbial growth.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance (BacT/ALERT® VIRTUO™ vs. BacT/ALERT® 3D™)
    Ratio of True Positives (VIRTUO / BTA3D) – Overall95% CI for the ratio of true positives should largely encompass or be close to 1.0, indicating comparable detection rates between the two systems for various specimen types and clinical determinations (Significant, Contaminant, Unknown).Blood Cultures (Compliant)
    For FA Plus (Blood)Ratio of True Positives should be close to 1.0, and 95% CI should indicate comparable performance.Significant: 0.970 (95% CI: 0.821, 1.119)
    Contaminant: 0.636
    Total: 0.924 (95% CI: 0.766, 1.082)
    For FN Plus (Blood)Same as above.Significant: 0.979 (95% CI: 0.823, 1.135)
    Contaminant: 1.300
    Total: 1.034 (95% CI: 0.852, 1.216)
    For PF Plus (Blood)Due to low numbers, potentially wider CIs are accepted, but still demonstrate detection.Significant: 1.000 (95% CI: -1.772, 3.772*)
    Total: 1.000 (95% CI: -1.772, 3.772*)
    For Low Fill FA Plus (Blood)Same as above.Significant: 0.966 (95% CI: 0.790, 1.142)
    Contaminant: 1.333
    Total: 1.031 (95% CI: 0.790, 1.272)
    For SA (Blood)Same as above.Significant: 1.026 (95% CI: 0.857, 1.195)
    Contaminant: 0.400
    Total: 0.920 (95% CI: 0.736, 1.104)
    For SN (Blood)Same as above.Significant: 1.067 (95% CI: 0.797, 1.337)
    Contaminant: 0.250
    Total: 1.000 (95% CI: 0.730, 1.270)
    Sterile Body Fluid Cultures
    For FA Plus (SBF)Same as above.Significant: 1.000 (95% CI: 0.852, 1.148)
    Contaminant: 1.333
    Total: 1.058 (95% CI: 0.864, 1.252)
    For FN Plus (SBF)Same as above.Significant: 1.026 (95% CI: 0.838, 1.214)
    Contaminant: 1.125
    Total: 1.000 (95% CI: 0.794, 1.206)
    For PF Plus (SBF)Same as above.Significant: 0.933 (95% CI: 0.807, 1.059)
    Contaminant: 0.000
    Total: 0.882 (95% CI: 0.665, 1.099)
    For SA (SBF)Same as above.Significant: 1.000 (95% CI: 0.837, 1.163)
    Contaminant: 1.000
    Total: 1.056 (95% CI: 0.806, 1.306)
    False Positive Rate (Overall)Should be low and comparable to the predicate device.VIRTUO: 0.09% (5/5862)
    BTA3D: 0.19% (11/5862)
    False Negative Rate (Overall)Should be low and comparable to the predicate device.VIRTUO: 0.38% (22/5862)
    BTA3D: 0.38% (22/5862)
    Positive Control PerformanceHigh percentage of positive controls detected.1487/1490 (99.8%) signaled positive.
    Negative Control PerformanceHigh percentage of negative controls detected as negative.1486/1486 (100%) signaled negative.
    Overall Control PerformanceHigh percentage of controls giving expected results.2973/2976 (99.9%) of controls gave expected results.

    Note on CI for PF Plus (Blood): The document explicitly states "Since the confidence interval contains a negative, the interval does not provide a meaningful interpretation" due to extremely low positive counts for this specific category. However, the 1:1 ratio of detected positives itself implies equivalence for the limited positive cases observed.


    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (Clinical Evaluation):

      • Total Bottle Pairs Tested (Clinical Samples): 5862 bottle pairs (one VIRTUO bottle and one BTA3D bottle per patient sample).
      • Detailed Sample Sizes per Category:
        • BacT/ALERT FA Plus (Blood): 1057 (compliant)
        • BacT/ALERT FN Plus (Blood): 912 (compliant)
        • BacT/ALERT PF Plus (Blood): 161 (compliant)
        • Low Fill BacT/ALERT FA Plus (Blood): 379 (compliant)
        • BacT/ALERT SA (Blood): 780 (compliant)
        • BacT/ALERT SN (Blood): 830 (compliant)
        • BacT/ALERT FA Plus (SBF): 374
        • BacT/ALERT FN Plus (SBF): 437
        • BacT/ALERT PF Plus (SBF): 77
        • BacT/ALERT SA (SBF): 81
        • BacT/ALERT SN (SBF): 81
      • Data Provenance: "External performance evaluations were conducted at eight external clinical sites." The document does not specify the country of origin, but given the FDA submission, it is typically either US-based or multi-national with significant US participation. The data is prospective as it involves clinical evaluation performed specifically for the 510(k) submission.
    • Control Sample Size:

      • 1490 positive controls
      • 1486 negative controls
      • Total controls: 2976

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number or qualifications of experts used. However, for a device detecting microbial growth, the "ground truth" for positive culture would typically be established by standard microbiological laboratory procedures:

    • Positive Culture: Visual detection of growth indicators (e.g., turbidity, gas production, or color change) in the culture bottle, followed by subculture onto agar plates and identification of microbial isolates using standard laboratory techniques (e.g., Gram stain, biochemical tests, mass spectrometry).
    • Negative Culture: No growth detected after a specified incubation period, confirmed by terminal subculture on rich media.
    • Clinical Determination (Significant/Contaminant/Unknown): This classification would typically be made by a qualified microbiologist or infectious disease physician, based on the identified organism, patient's clinical presentation, and other relevant factors.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    The document does not explicitly describe an adjudication method. For comparison studies of this nature, discrepancies between the investigational device and the predicate (or deviations from expected results) would generally be investigated by the clinical sites' microbiology laboratories to determine the definitive "ground truth" (e.g., re-subculture, review of patient charts).


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This type of study (assessing human reader performance with and without AI assistance on diagnostic images) is not applicable to this device. The BacT/ALERT VIRTUO is an automated microbial detection system, not an imaging AI diagnostic aid for human readers. Its primary function is to automatically detect the presence of microbial growth, replacing or assisting a manual observation process, rather than augmenting a radiologist's interpretation of an image.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, in essence, the primary performance evaluation measures the standalone performance of the BacT/ALERT VIRTUO system against the predicate BacT/ALERT 3D system. Both systems operate autonomously to detect microbial growth. The reported "True Positives," "False Positives," and "False Negatives" for VIRTUO directly reflect its algorithm-only performance in detecting growth from clinical samples, without human intervention required for the detection itself (humans are involved in loading bottles and interpreting the overall clinical significance, but not in the detection of growth by the device).


    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth was established by standard microbiological culture and identification methods, combined with a clinical review to classify isolates as "significant," "contaminant," or "unknown." This is essentially a definitive laboratory diagnosis of microbial presence and identity.


    8. The Sample Size for the Training Set

    The document does not specify a separate "training set" sample size in the context of typical AI/ML model development. The device uses "proprietary algorithms unique for BacT/ALERT VIRTUO," which implies embedded logic or rules rather than a dynamically trained machine learning model in the contemporary sense. The algorithms would have been developed and validated internally by the manufacturer using a combination of laboratory studies, simulated data, and potentially a distinct internal dataset. The data presented in the 510(k) is for the clinical validation (test set performance) of the final device.


    9. How the Ground Truth for the Training Set was Established

    Given that this is not a modern AI/ML device with a distinct training phase documented for public review, the methods for establishing ground truth for any internal algorithm development ("training") would also implicitly rely on controlled laboratory studies and established microbiological techniques to generate data on microbial growth, CO2 production kinetics, and associated reflectance changes. This would involve inoculating bottles with known strains and concentrations of microorganisms and observing their growth characteristics under various conditions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161810

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-02-09

    (223 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    /10 Aerobic/F Culture Vials Soybean-Casein Digest Broth in a Plastic Vial Regulation Number: 21 CFR 866.2560

    Common Names:

    Aerobic blood culture medium

    Regulatory Information

    Classification: 21 CFR§866.2560

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BD BACTECT™ Standard/10 Aerobic/F culture vials (enriched Soybean-Casein Digest broth with CO2) are for aerobic blood cultures. Principal use is with the BD BACTEC fluorescent series instruments for the qualitative culture and recovery of aerobic microorganisms (bacteria and yeast) from blood.

    Device Description

    The sample to be tested is inoculated into one or more vials which are inserted into the BD BACTEC fluorescent series instrument for incubation and periodic reading. Each vial contains a chemical sensor which can detect increases in CO2 produced by the growth of microorganisms. The sensor is monitored by the instrument every ten minutes for an increase in its fluorescence, which is proportional to the amount of CO2 present. A positive reading indicates the presumptive presence of viable microorganisms in the vial. Detection is limited to microorganisms that will grow in a particular type of medium.

    AI/ML Overview

    The provided document describes the BD BACTEC™ Standard/10 Aerobic/F Culture Vials (plastic version). The device is a culture vial for aerobic blood cultures, intended for use with BD BACTEC fluorescent series instruments for qualitative detection of microorganisms (bacteria and yeast) in blood. The study presented aims to demonstrate that the plastic version of the culture vial is substantially equivalent to the predicate device (glass version).

    Here's an analysis of the acceptance criteria and study details based on the provided text, focusing on the comparisons between the modified (plastic) device and the predicate (glass) device:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Time to Detection (TTD): No relevant difference from current product (predicate device).The estimated median TTD difference for 682 positive paired sets was -1.084 hours, favoring the modified device. Conclusion: The modified device meets the acceptance criteria, and the effect of differences between the modified and predicate devices on TTD was minimal; the modified device performs equivalently.
    Percent Recovery: Equivalent between devices.984 paired sets evaluated. 948 were positive in both. 4 grew in predicate only, 9 grew in modified only. McNemar chi-square analysis: 0.2673, indicating no relevant difference.
    Microbial Detection Limit: No relevant difference between devices.360 paired sets tested. 207 grew in both. 37 grew in predicate only, 48 grew in modified only. McNemar chi-square analysis: 0.2781, indicating no relevant difference.
    False Positive Rate: No false positives in the modified device, or statistically insignificant difference from predicate.240 paired sets evaluated. No false positives in the modified device. One anomaly (a false positive) observed in the predicate device (positive at 16.98 hours without organism). A statistical analysis could not be performed due to insufficient failures in the test (only one failure in predicate and none in modified). Implicitly, the modified device met this better than the predicate.
    False Negative Rate: No statistically significant difference in recovery between devices.91 paired sets were instrument negative. 41 modified devices and 57 predicate devices were instrument negative. Terminal subculture of instrument negatives identified 34 false negatives total. 18 were in the predicate device, 16 in the modified device. McNemar chi-square analysis (p=0.8638) indicates no statistically significant difference in recovery between the modified and predicate devices.
    Reproducibility: No statistical difference across lots in TTD and recovery.The modified device was evaluated for reproducibility across lots. No statistical difference in time to detection and recovery was observed between lot comparisons.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. It describes analytical studies comparing the modified device to a predicate device.

    • Instrument Time to Detection: 682 paired sets recovered organisms in both devices.
    • Percent Recovery: 984 paired sets.
    • Microbial Detection Limit: 360 paired sets.
    • False Positive Rate: 240 paired sets (comprised of 40 bottles from each of 3 lots).
    • False Negative Rate: 91 paired sets that were instrument negative.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide any information regarding the number or qualifications of experts used to establish ground truth. The studies described are laboratory-based comparisons of detection capabilities, not evaluations of human interpretation.

    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method involving human interpretation. The "ground truth" (or reference standard) in these studies appears to be based on whether microorganisms were successfully grown and detected, or confirmed by terminal subculture in the case of false negatives.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    No MRMC comparative effectiveness study was done. This device is a culture vial designed for automated detection by an instrument, not an AI-assisted diagnostic tool that requires human reader interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The performance described in the document is effectively "standalone" in terms of the device's ability to detect microbial growth. The BD BACTEC fluorescent series instruments are automated systems that detect increases in CO2 produced by microorganisms. The studies evaluate the performance of the modified culture vial within this automated system, essentially demonstrating the algorithm (instrument's detection mechanism) only performance as it relates to the culture media. There is no human intervention in the detection process itself once the vials are loaded into the instrument.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the analytical studies appears to be based on:

    • Presence of microbial growth: Confirmed by instrument detection and/or terminal subculture for definitive growth determination.
    • Known inoculum levels: For studies like Microbial Detection Limit and implied for Time to Detection and Percent Recovery, predefined organism concentrations are used to challenge the system.

    8. The Sample Size for the Training Set

    The document does not mention any "training set." The studies described are analytical performance evaluations of a medical device, not a machine learning model that requires a training phase.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no mention of a training set or machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161306

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-08-08

    (90 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Standard Anaerobic/F Culture Vials Soybean-Casein Digest Broth in a Plastic Vial Regulation Number: 21 CFR 866.2560

    Common Names:

    Anaerobic blood culture medium

    Regulatory Information

    Classification: 21 CFR§866.2560

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BD BACTECTM Standard Anaerobic/F culture vials (prereduced enriched Soybean-Casein Digest broth with CO2) are for anaerobic blood cultures. Principal use is with the BD BACTEC fluorescent series instruments for the qualitative culture and recovery of anaerobic microorganisms from blood.

    Device Description

    The sample to be tested is inoculated into one or more vials which are inserted into the BD BACTEC fluorescent series instrument for incubation and periodic reading. Each vial contains a chemical sensor which can detect increases in CO2 produced by the growth of microorganisms. The sensor is monitored by the instrument every ten minutes for an increase in its fluorescence, which is proportional to the amount of CO2 present. A positive reading indicates the presumptive presence of viable microorganisms in the vial. Detection is limited to microorganisms that will grow in a particular type of medium.

    AI/ML Overview

    The provided document describes the BD BACTEC™ Standard Anaerobic/F Culture Vials Soybean-Casein Digest Broth in a Plastic Vial (modified device) and its substantial equivalence to the predicate device (glass vial). The studies performed are analytical and focus on the performance of the culture medium and vial characteristics.

    Here's a breakdown of the requested information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative "acceptance criteria" for each test. Instead, it indicates that the modified device should perform "equivalently" or show "no relevant difference" compared to the predicate device. For the purpose of this table, "Acceptance Criteria" are inferred from the study conclusions.

    Performance MetricAcceptance Criteria (Inferred)Reported Device Performance
    Instrument Time to DetectionNo relevant difference from the predicate deviceWilcoxon estimated median TTD difference for 443 positive paired sets is -0.500 hours (30 minutes), favoring the modified device. Conclusion: Effect of differences on TTD was minimal; modified device performs equivalently to the predicate device.
    Percent RecoveryRecovery equivalent between modified and predicate devices528 paired sets evaluated; 504 paired sets positive in both at 10-100 CFU inoculum. McNemar p-value could not be calculated as neither device showed exclusive detection. Conclusion: Recovery was equivalent.
    Microbial Detection LimitNo statistically significant difference in recovery compared to the predicate device312 paired sets inoculated: 191 grew in both, 36 in predicate only, 39 in modified only, 46 in neither. McNemar chi-square analysis: p=0.8174. Conclusion: No statistically significant difference in recovery.
    False Positive RateNo false positive bottles for the modified device240 paired sets (40 from each of 3 lots) inoculated with fresh human blood at varying levels. All expected to be instrument-negative. Conclusion: No false positive bottles observed for the modified device with blood volumes (2, 4, 6, 8, and 10mL).
    False Negative RateNo statistically significant difference in recovery compared to the predicate device70 paired sets end of protocol negative evaluated for false negative rate. 36 modified device only instrument negatives, 39 predicate device only instrument negatives. 1 false negative observed in the predicate device. McNemar chi-square analysis: p=1.00. Conclusion: No statistically significant difference in recovery.
    Reproducibility (Lot-to-Lot)No relevant difference in time to detection across lotsNo relevant difference in time to detection observed between lot comparisons.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes analytical studies. The "test set" in this context refers to the samples used in these analytical performance studies.

    • Instrument Time to Detection: 443 paired sets
    • Percent Recovery: 528 paired sets
    • Microbial Detection Limit: 312 paired sets
    • False Positive Rate: 240 paired sets (40 from each of 3 lots)
    • False Negative Rate: 70 paired sets

    Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. Given the nature of these analytical studies (e.g., inoculation with specific CFU levels), they are inherently prospective and controlled laboratory experiments. The origin of the human blood used for the false positive rate study is not specified beyond "fresh human blood."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The studies are analytical performance comparisons of a microbial growth medium, not diagnostic accuracy studies that rely on expert interpretation of results. The "ground truth" for these studies is typically defined by microbial culture techniques (e.g., presence/absence of growth, colony counts) rather than expert consensus on images or clinical classifications.

    4. Adjudication Method for the Test Set

    This information is not applicable/provided. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies where multiple human readers interpret data, and discrepancies need to be resolved. These analytical studies focus on the objective measurement of microbial growth and detection by the instrument.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This information is not applicable. The device (BD BACTEC™ Standard Anaerobic/F Culture Vials) is a culture medium used with an automated instrument (BD BACTEC fluorescent series instruments) for detecting microbial growth in blood. It is not an AI-powered diagnostic device or an imaging device that involves human readers interpreting results with or without AI assistance. Therefore, an MRMC comparative effectiveness study involving human readers and AI is outside the scope of this device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This information is not applicable in the context of an "algorithm only" device. The device itself is a culture vial (medium) that is used with an instrument. The instrument's algorithms detect growth. The studies evaluate the performance of the vial and medium in conjunction with the instrument, not an isolated algorithm. The "standalone" performance is essentially what the analytical studies describe – the ability of the combined system to detect growth.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for these analytical studies is based on microbiological culture and laboratory methods.

    • Presence/absence of microbial growth: Confirmed by standard microbiological techniques (e.g., subculture to agar plates, colony identification).
    • Time to detection (TTD): Objectively measured by the BACTEC instrument.
    • Colony-forming units (CFU): Used to prepare inoculum levels for certain studies.

    8. The Sample Size for the Training Set

    This information is not provided and is not applicable in the context presented. The document describes an analytical performance comparison study for a modified device against a predicate device. It does not mention any machine learning or AI components that would require a distinct "training set" in the conventional sense. The instrument's algorithms are assumed to be pre-existing and validated.

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided and is not applicable for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151866

    Validate with FDA (Live)

    Date Cleared
    2016-02-25

    (231 days)

    Product Code
    Regulation Number
    866.2560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K151866

    Trade/Device Name: BD BACTEC Peds Plus/F Culture Vials (plastic) Regulation Number: 21 CFR 866.2560
    | Aerobic blood culture medium |
    | DEVICE CLASSIFICATION: | 21 CFR§866.2560

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    BD BACTEC Peds Plus™/F culture vials (enriched Soybean-Casein Digest broth with CO2) are for aerobic blood cultures. Principal use is with the BACTEC fluorescent series instruments for the qualitative culture and recovery of aerobic microorganisms (mainly bacteria and yeast) from pediatric and non-pediatric blood specimens which are generally less than 5 mL in volume.

    Device Description

    The sample to be tested is inoculated into one or more vials which are inserted into the BD BACTEC fluorescent series instrument for incubation and periodic reading. Each vial contains a chemical sensor which can detect increases in CO2 produced by the growth of microorganisms. The sensor is monitored by the instrument every ten minutes for an increase in its fluorescence, which is proportional to the amount of CO2 present. A positive reading indicates the presumptive presence of viable microorganisms in the vial. Detection is limited to microorganisms that will grow in a particular type of medium.

    AI/ML Overview

    The document describes the BD BACTEC Peds Plus™/F Culture Vials (plastic), which is an aerobic blood culture medium. It compares this modified device to its predicate, the BD BACTEC Peds Plus/F medium (glass). The purpose of the studies is to demonstrate that the modified device performs equivalently to the predicate device.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state formal "acceptance criteria" in a quantitative format for each performance metric. Instead, it aims to demonstrate equivalence or non-inferiority to the predicate device. The reported performance is generally framed as "no statistically significant difference" or "performs equivalently."

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Instrument Time to Detection (TTD)Equivalence to predicate device.The Wilcoxon estimated median TTD difference for 473 positive sets is -0.250 hours (15 minutes), favoring the modified device. The data indicate that the effect of differences between the modified and predicate devices on TTD under these test conditions was minimal and that the modified device performs equivalently to the predicate device.
    Percent RecoveryNo statistically significant difference in recovery compared to the predicate device.A total of 1344 paired sets were evaluated. 953 paired sets were positive in both devices. The McNemar p-value for this data set equals 0.2673, indicating no statistically significant difference in recovery. The document notes a "significant difference in recovery...favoring the modified device contained in plastic vials," which seemingly contradicts the p-value unless it refers to a non-statistical observation or a different comparison. Assuming the p-value is the statistical conclusion, it suggests no significant difference.
    Microbial Detection LimitNo statistically significant difference in recovery compared to the predicate device.A total of 360 inoculated cultures. 196 grew and detected in both. 42 in predicate only, 57 in modified only. McNemar chi-square analysis (p=0.1594) indicates no statistically significant difference in recovery between the modified and predicate devices. Of 164 sets, 62.2% had plate counts of 0 CFU per vial.
    False Positive RateNo false positives within specified blood volumes (0.5 to 1 mL).A total of 288 paired sets (96 bottles/lot across 3 lots) inoculated with fresh human blood. There were no false positive bottles of the modified device observed within blood volumes (0.5 to 1 mL).
    False Negative RateNo statistically significant difference in false negative rates compared to the predicate device.83 paired sets were end of protocol negative in both. 129 sets were predicate-only detected, 146 sets were modified-only detected. Terminal subculture of non-detected vials showed 45 false negative bottles (25 in predicate, 20 in modified). A Chi-square analysis (p=0.456) indicates no significant difference favoring the modified in the plastic bottle. Retesting of Haemophilus influenzae strains initially leading to high false negatives resulted in detection in both vials.
    Antimicrobial Neutralization CapabilityNo statistically significant difference in recovery when antimicrobials are present, compared to the predicate device.Eleven drugs evaluated at MIC level. A total of 51 paired sets tested. Predicate detected 48 times (94%), modified detected 49 times (96%). McNemar's test p-value = 1.000, indicating no statistically significant difference in recovery between the modified and predicate devices.
    ReproducibilityNo statistical difference across lots in Time to Detection and Percent Recovery.Tested across three lots for TTD and Percent Recovery. There was no statistical difference observed across lots comparison.

    2. Sample Size Used for the Test Set and Data Provenance

    • Instrument Time to Detection: 473 paired sets (positive in both devices).
    • Percent Recovery: 1344 paired sets.
    • Microbial Detection Limit: 360 inoculated cultures (196 grew and detected in both).
    • False Positive Rate: 288 paired sets (96 bottles from each of 3 lots).
    • False Negative Rate: 83 paired sets (end-of-protocol negative in both), and other sets where only one device detected (129 predicate-only, 146 modified-only).
    • Antimicrobial Neutralization Capability: 51 paired sets.
    • Reproducibility: Tested across three lots (implied using subsets of the TTD and Percent Recovery data).

    Data Provenance: The document does not specify the country of origin. The studies appear to be prospective analytical studies conducted in a laboratory setting, involving the inoculation of culture media with specific microorganisms and human blood.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. The studies described are analytical performance studies of a culture medium, not diagnostic studies involving interpretation by human experts. The "ground truth" for these tests (e.g., presence/absence of microbial growth, CFU counts) would have been established through standard microbiology laboratory techniques (e.g., subculture, plate counts) rather than expert consensus on diagnostic images or clinical data.

    4. Adjudication Method for the Test Set

    This information is not applicable/provided. As mentioned above, these are analytical performance studies where the outcome is objectively measured (e.g., instrument detection of CO2, colony counts from subcultures), not subjective interpretations requiring adjudication by human experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. The device is a culture medium, and the studies assess its analytical performance (e.g., time to detection, recovery rates) directly from instrument readings and laboratory methods, not human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies reflect standalone performance of the device in conjunction with the BACTEC fluorescent series instruments. The instrument automatically detects growth based on CO2 production. Human intervention would be for initial sample loading, result interpretation (positive/negative), and subsequent microbiological workup, but the "detection" itself is algorithmic. The document states: "The sensor is monitored by the instrument every ten minutes for an increase in its fluorescence, which is proportional to the amount of CO2 present. A positive reading indicates the presumptive presence of viable microorganisms in the vial." This describes the standalone algorithmic detection by the instrument.

    7. The Type of Ground Truth Used

    The ground truth used for these analytical studies includes:

    • Microbiological Culture and Growth: The presence or absence of viable microorganisms, confirmed by growth in culture or subculture, and typically quantified by plate counts (e.g., CFU per vial shown in the Microbial Detection Limit section).
    • Instrument Readings: The BACTEC fluorescent series instruments provide quantitative data on CO2 production and time to detection.
    • Terminal Subculture: Used to confirm true positive/negative status where the instrument did not detect growth, particularly for false negative assessments.

    8. The Sample Size for the Training Set

    This information is not provided as this is not an AI/ML device that requires a distinct "training set" in the conventional sense. The device is a culture medium where performance is evaluated through a series of analytical experiments with known microbial inocula and clinical samples (human blood). There is no "training phase" for an algorithm in the context of this product.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable as there is no training set for an AI/ML algorithm involved with this device. The product is a physical culture medium, and its performance is evaluated against established microbiological methods and instrument readings as ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 3