Search Filters

Search Results

Found 22 results

510(k) Data Aggregation

    K Number
    K191364

    Validate with FDA (Live)

    Date Cleared
    2020-02-14

    (268 days)

    Product Code
    Regulation Number
    864.5700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    California 92131

    Re: K191364

    Trade/Device Name: T-TAS 01 System with PL Chip Regulation Number: 21 CFR 864.5700
    |
    | Regulation: | 21 CFR 864.5700

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The T-TAS 01 Instrument is intended for use with T-TAS reagent chips in the clinical laboratory.

    The T-TAS 01 PL chip is intended for use in the clinical laboratory for the analysis of the platelet thrombus formation process (primary hemostatic function) in patients age 21 and older with a history of conditions associated with impaired primary hemostatic function or use of antiplatelet therapy. The test uses BAPA-anticoagulated whole blood specimens to measure platelet adhesion to a thrombogenic collagen-coated surface and aggregation, which causes an increase in flow pressure inside the PL chip. The test measures primary hemostatic function as the area under the pressure-time curve (AUC), with AUC < 260 suggesting abnormal primary hemostatic function. Additional testing may be necessary to identify the cause(s) of abnormal primary hemostatic function. The test has been evaluated in patients taking antiplatelet therapy, in patients with von Willebrand disease, and in patients with Glanzmann's thrombasthenia. Other primary hemostasis disorders have not been evaluated.

    The BAPA tube for T-TAS 01 is intended to be used for the collection, transport, and storage of blood specimens for use with the T-TAS 01 system.

    Device Description

    The T-TAS 01 system is an in vitro diagnostic device that is comprised of tabletop instrument controlled by a dedicated PC and a disposable, single-use flow chamber. The PL Chip for T-TAS 01 is designed to specifically measure platelet thrombus formation (PTF) under physiological conditions on a collagen-coated analytical path consisting of 26 microcapillary channels. Platelet thrombus formation is a direct indicator of the patient's primary hemostatic function. The assay is performed under arterial flow conditions using benzylsulfonyl-D-Arg-Pro-4-amidinobenzylamide (BAPA)-anticoagulated whole blood samples. BAPA is an anticoagulant that inhibits thrombin and factor Xa, blocking the coagulation cascade and allowing the PL assay to specifically measure only the platelet thrombus formation process (primary hemostasis). During the assay, the blood sample is exposed to arterial shear stresses at 1,500 s-1 in the presence of a collagen-coated surface, which causes platelet attachment to collagen mediated by von Willebrand factor (vWF), and platelet activation. Platelet activation causes the release of endogenous factors contained within the platelets that recruit and activate other platelets and cause aggregation, and platelet thrombus formation. The growing platelet thrombus causes occlusion of the microcapillary channels, which increases the flow pressure within the assay chip. The process of platelet thrombus formation in the flow chamber is continuously monitored by a pressure sensor that tracks pressure changes in the flow path. Results are calculated automatically within 10 minutes or when the pressure a reading reaches 60 kPa above the baseline pressure, whichever occurs first. Results are displayed as AUC, which is the flow pressure curve over 10 minutes.

    AUC results less than 260 are associated with abnormal primary hemostatic function.

    AI/ML Overview

    The provided text is a 510(k) Summary for the T-TAS 01 System with PL Chip, an automated platelet aggregation system. It details the device's intended use, comparison to a predicate device, and non-clinical and clinical performance data.

    Here's an analysis to extract the requested information, noting that this document describes a diagnostic test, not an AI model. Therefore, some questions related to AI-specific studies (e.g., human-in-the-loop, AI effect size, training data ground truth establishment) are not directly applicable or answerable from this document.


    Device Name: T-TAS 01 System with PL Chip

    Device Type: Automated Platelet Aggregation System (In Vitro Diagnostic, IVD)

    Acceptance Criteria for Performance (based on clinical performance data):

    The device's performance is demonstrated through its ability to differentiate between individuals with normal primary hemostatic function and those with impaired function due to specific conditions. The key metric is the Area Under the Pressure-Time Curve (AUC), with a cutoff of < 260 AUC suggesting abnormal primary hemostatic function.

    Since this is a diagnostic device comparison to a predicate, the "acceptance criteria" are implicitly met by demonstrating substantial equivalence to the predicate device (Dade Behring PFA-100) and by providing clinical performance metrics (negative agreement and sensitivity) for various target populations. The document does not explicitly state pre-defined quantitative acceptance criteria thresholds for these metrics that were required for clearance, but rather presents the results.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given that explicit "acceptance criteria" (thresholds for success) are not stated in the document as typical for AI/ML performance metrics, we will present the key performance indicators reported and their values. The underlying "acceptance" is the FDA's determination of substantial equivalence based on these results.

    Performance MetricImplicit Acceptance Criteria (based on predicate equivalence & clinical utility)Reported Device Performance (T-TAS 01 AUC < 260 cutoff)
    PrecisionCV ≤ 15% or SD ≤ 39 (as stated in the document)Met (e.g., Total CV for High: 2.8%, Middle: 14.3%, Low: 19.6% - note: Low is slightly above 15% CV but acceptable given SD ≤ 39, as 25.7 is less than 39 for low signal values)
    Negative Agreement (Healthy Donors)High agreement expected, comparable to predicate's ability to identify normal.95.8% (95% CI: 91.1-98.0%)
    Sensitivity (Aspirin Monotherapy)Clinically meaningful detection of impairment due to aspirin.68.4% (95% CI: 55.5-79.0%)
    Sensitivity (Clopidogrel + ASA DAPT)Clinically meaningful detection of impairment.100.0% (95% CI: 81.5-100.0%)
    Sensitivity (Prasugrel + ASA DAPT)Clinically meaningful detection of impairment.100.0% (95% CI: 78.2-100.0%)
    Sensitivity (Ticagrelor + ASA DAPT)Clinically meaningful detection of impairment.100.0% (95% CI: 76.8-100.0%)
    Sensitivity (von Willebrand Disease)Clinically meaningful detection of impairment.72.0% (95% CI: 50.6-87.9%)
    Sensitivity (Glanzmann's Thrombasthenia)Clinically meaningful detection of impairment.100.0% (95% CI: 43.9-100.0%)
    InterferenceNo significant effect on AUC results at specific concentrations of tested compoundsMet (list of tested compounds provided, some known to affect platelet activity still acceptable as device measures their effect)
    Stability (PL chip & blood samples)Demonstrated stability for declared durations.Met (Closed pouch PL chip: 12 months; Open pouch PL chip: 8 hours; BAPA tube: 10 months; Blood sample: 6 hours)

    2. Sample Size Used for the Test Set and Data Provenance

    • Clinical Performance Test Set Sample Size:
      • Healthy Donors: 142 individuals
      • Aspirin Monotherapy: 57 patients
      • Clopidogrel + ASA DAPT: 18 patients
      • Prasugrel + ASA DAPT: 15 patients
      • Ticagrelor + ASA DAPT: 14 patients
      • von Willebrand Disease: 25 patients (12 Type 1, 10 Type 2, 3 Type 3)
      • Glanzmann's Thrombasthenia: 3 patients
      • Total Clinical Subjects: 142 + 57 + 18 + 15 + 14 + 25 + 3 = 274 subjects across 6 investigational sites.
    • Data Provenance: The study was conducted at 6 investigational sites. The document does not explicitly state the country of origin, but given it's an FDA submission, the primary data likely comes from the US. It is a prospective collection of data for the purpose of this clinical evaluation, as it refers to "subjects enrolled at ... investigational sites."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This is a diagnostic device for measuring a physiological parameter (platelet thrombus formation) based on direct blood sample analysis. The "ground truth" is established by the patient's diagnosed medical condition or medication use (e.g., "healthy donor," "patients taking antiplatelet therapy," "patients with von Willebrand disease," "patients with Glanzmann's thrombasthenia") and, in some cases, by other laboratory tests (e.g., PFA-100 results, vWF activity, FVIII:C levels to confirm vWD).

    Therefore, the concept of "experts establishing ground truth" in the sense of image annotation or AI model output review by multiple radiologists is not directly applicable here. The "experts" are the physicians who diagnosed the patients and the laboratory personnel who performed confirmatory tests. The document does not specify the number or qualifications of these individuals directly, as their role is standard medical practice for diagnosis.

    4. Adjudication Method for the Test Set

    Not applicable in the context of this diagnostic device study. Ground truth is based on clinical diagnosis and other objective laboratory parameters, not on interpretation requiring adjudication among human readers/experts.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This is not an AI-based device, nor does it involve human readers interacting with an AI system. It is an automated in vitro diagnostic test.

    6. If a Standalone (i.e. Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The T-TAS 01 System is a standalone automated diagnostic device. Its performance, as measured by AUC, is the "algorithm only" or "device only" performance. The results (AUC values) are automatically calculated by the instrument when the pressure reading reaches 60 kPa or after 10 minutes.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The ground truth was established by:

    • Clinical Diagnosis: Patients with a history of conditions associated with impaired primary hemostatic function (e.g., Von Willebrand disease, Glanzmann's thrombasthenia) or those confirmed to be taking antiplatelet therapy (aspirin, clopidogrel, prasugrel, ticagrelor).
    • Absence of Diagnosis/Conditions: For the healthy control group, individuals without a history of inherited or acquired platelet dysfunction, and without laboratory evidence of von Willebrand disease.
    • Confirmatory Laboratory Tests: For von Willebrand disease patients, comparison to PFA-100 Col/EPI and Col/ADP sensitivity, and reference to vWF activity and FVIII:C results.

    8. The Sample Size for the Training Set

    This document describes the validation of a lab diagnostic device, not an AI/ML model. Therefore, the concept of a "training set" for model development is not explicitly discussed. The device's underlying principles (measuring pressure changes due to platelet adhesion and aggregation) are based on established physiological and engineering principles, not on learned patterns from a training dataset in the AI sense.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as this is not an AI/ML device relying on a training set in the typical AI sense. The device is designed and validated based on known biological mechanisms and engineering specifications.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181777

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-03-29

    (269 days)

    Product Code
    Regulation Number
    864.5700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Houston, Texas 77099

    Re: K181777

    Trade/Device Name: AggreGuide A-100 ADP Regulation Number: 21 CFR 864.5700
    Classification Name: | System automated plateletaggregation |
    | Regulation Number: | 864.5700

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AggreGuide A-100 ADP Assay is used with the AggreGuide A-100 instrument in non-CLIA waived physician's office or clinical laboratory for the detection of platelet dysfunction in patients age 22 or older receiving P2Y 12 antiplatelet drugs, prasugrel and ticagrelor, using 3.2% sodium citrated whole blood. The AggreGuide A-100 ADP Assay is a semi-quantitative assay. The level of platelet aggregation is determined by the platelet activity index (PAI) where values < 4.7 PAI suggest that platelet dysfunction is due to the presence of P2Y 12 antiplatelet drugs, prasugrel and ticagrelor. The test results should be interpreted in conjunction with all laboratory data available to the clinician.

    Device Description

    The AggreGuide A-100 ADP Assay is an individual use, disposable assay cartridge for use with the AggreGuide A-100 instrument. The cartridge contains preloaded freeze dried agonist. The level of platelet aggregation induced by the adenosine diphosphate (ADP) agonist in a sample of whole blood is detected within the cartridge. The amount of platelet aggregation is measured by detecting and quantifying the laser light scattering caused by platelet aggregates. P2Y12 inhibitor drugs e.g. clopidogrel, and ticagrelor are known to inhibit the level of platelet aggregation, causing platelet dysfunction.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    MetricAcceptance CriteriaReported Device Performance (AggreGuide A-100 ADP Assay)
    Sensitivity
    Prasugrel (Post-Loading, 24 hours)≥ 0.801.000 (95% CI: 0.918 – 1.000)
    Prasugrel (Post-Maintenance, 7 days)≥ 0.800.907 (95% CI: 0.784 - 0.963)
    Ticagrelor (Post-Loading, 3-6 hours)≥ 0.800.906 (95% CI: 0.825 – 0.952)
    Ticagrelor (Post-Maintenance, 7 days)≥ 0.800.839 (95% CI: 0.770 - 0.890)
    SpecificityNot explicitly stated as a numerical acceptance criterion, but "0.907" is given as an overall specificity value for the device.0.907

    Note: The document only explicitly states numerical acceptance criteria for sensitivity. Specificity is provided as a reported value, implying it met an internal or expected threshold, though the exact numerical acceptance criterion for specificity isn't listed.

    2. Sample Size Used for the Test Set and Data Provenance

    The sample sizes for the sensitivity analysis (test set) are:

    • Prasugrel: 43 patients for both Post-Loading (24 hours) and Post-Maintenance (7 days).
    • Ticagrelor: 85 patients for Post-Loading (3-6 hours) and 143 patients for Post-Maintenance (7 days).

    The document does not explicitly state the country of origin of the data or whether the study was retrospective or prospective. However, the study is described as "clinical testing" and evaluated against a "clinical truth data set comprising off-therapy (baseline) versus on-therapy with P2Y12 inhibitor medications. On-therapy clinical truth utilized high potency P2Y12 medications at a time of full-effect," which suggests a prospective clinical study design.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not specify the number of experts used to establish the ground truth or their qualifications. It simply refers to a "clinical truth data set."

    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method used for establishing the ground truth of the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. The study evaluated the standalone performance of the AggreGuide A-100 ADP Assay and compared its performance (sensitivity values) against another device (VerifyNow) on the same clinical truth data set, but not in the context of human reader improvement with and without AI assistance.

    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance)

    Yes, a standalone performance study was done. The reported sensitivity and specificity values are for the AggreGuide A-100 ADP Assay as a diagnostic device, without human intervention in the interpretation of the primary PAI result. The document says, "The level of platelet aggregation is determined by the platelet activity index (PAI) where values < 4.7 PAI suggest that platelet dysfunction..." This indicates the algorithm provides the direct output, which is then used for interpretation.

    7. Type of Ground Truth Used

    The ground truth used was clinical truth data. Specifically, it was established by comparing "off-therapy (baseline) versus on-therapy with P2Y12 inhibitor medications." The "on-therapy clinical truth utilized high potency P2Y12 medications at a time of full-effect." This implies a reference standard based on the known physiological state of the patient (presence or absence of platelet dysfunction due to specific P2Y12 inhibitors).

    8. Sample Size for the Training Set

    The document does not provide information regarding the sample size for the training set. It focuses solely on the performance testing carried out.

    9. How the Ground Truth for the Training Set Was Established

    Since the document does not provide information on a training set, it also does not describe how the ground truth for any training set would have been established. The provided data relates to clinical performance testing, implying a pre-defined or externally validated cut-off point. It states, "Cut Off Point is confirmed by ROC Decision Threshold / J-statistic (Youden) analysis of the clinical truth data set," suggesting the cut-off was determined using the available clinical truth data, rather than being established pre-emptively on a separate training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160502

    Validate with FDA (Live)

    Date Cleared
    2017-04-19

    (421 days)

    Product Code
    Regulation Number
    864.5425
    Age Range
    2 - 12
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    |
    | RegulationNumber | 21 CFR 864. 5425 (K150041)21 CFR 864.5700

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TEG 6s System is intended for in vitro diagnostic use to provide semi-quantitative indications of the hemostasis state of a blood sample. The TEG 6s System records the kinetic changes in a venous sample of 3.2% citrated whole blood as the sample clots, and retracts in real time. The system output consists of a table of numerical values for parameters R, K, Angle, MA, and FLEV. The TEG 6s System provides specific blood modifiers, in the form of reagents dried-in-place within TEG 6s cartridges.

    Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests.

    The indication for use for the TEG 6s System is with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluations are commonly used to assess clinical conditions in cardiovascular surgery and cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure.

    Device Description

    The TEG6s analyzer is a four-channel in vitro diagnostic instrument with an integrated computer module, a display touch screen for operator interaction, and a slot for inserting one TEG 6s cartridge. The TEG 6s analyzer is for use in laboratories and near patient use. It consists of an assembly of controllers, sensors, and displays, all managed and sequenced by a central processor. The embedded programming in the processor provides the necessary information for the automation of hemostasis testing. The program sequences the instrument hardware through the appropriate cycles to perform the test. Using a compressor, pressure and vacuum sensors, and a series of valves, actuators and controls, blood samples are introduced into the microfluidics of a disposable cartridge.

    To perform a test, a disposable cartridge is inserted into the instrument. The instrument reads the bar code on the cartridge, identifies the type of cartridge for operator confirmation. Then, the operator adds a blood sample to the entry port on the cartridge and uses the touch screen to issue the command to the instrument to proceed with the test. The sample is then drawn into the cartridge under instrument control. The amount of the sample drawn into the cartridge is automatically determined by the volume of the reagent chambers in the cartridge.

    The TEG 6s analyzer firmware provides features for and manages all functions of the instrument, including user interface (via the touch display screen) and external communications for service and installation via Service Maintenance Software (SMS). SMS is run on a separate computer connected to one or more TEG 6s analyzer(s) via the Ethernet port and a router. Its purpose is to allow additional control of analyzer functions by authorized remote users, such as administrative and service personnel.

    The TEG Manager 2.0.0 is an optional accessory to Haemonetics TEG 6s Hemostasis System. TEG Manager 2.0.0 is an application that provides remote viewing of current and historical patient tracing and test results created by the TEG 6s analyzers, and administration of all connected TEG 6s devices. TEG Manager interfaces with the TEG analyzers to obtain clinical data and retrieves patient information from external Hospital Information System (HIS). Users cannot manipulate the data that is stored and displayed within TEG Manager or input any additional clinical data in the software.

    AI/ML Overview

    The provided document is a 510(k) summary for the Haemonetics TEG® 6s Hemostasis System and TEG Manager 2.0.0. The primary purpose of this submission is to demonstrate substantial equivalence to previously cleared predicate devices, specifically the TEG 6s Hemostasis System (K140893 and K150041), with the addition of TEG Manager 2.0.0 as an optional accessory.

    Here's the breakdown of the requested information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of specific numerical acceptance criteria (e.g., sensitivity, specificity, accuracy targets) for the TEG Manager 2.0.0. Instead, the acceptance criterion for the software is described in a more general sense:

    Acceptance Criteria (General)Reported Device Performance
    TEG Manager 2.0.0 Software Functionality: The software must be thoroughly verified and validated at different levels (unit, system/integration, verification, validation, beta testing)."The TEG Manager 2.0.0 software was thoroughly verified and validated via test standards. The overall approach to testing was to test the software at different levels, including unit, system/integration, verification, validation and beta testing."
    Test Case Results: Acceptance criteria for verification and validation testing were identified in individual test cases as expected results. Test cases were comprised of scenarios with multiple steps along with their respective expected results."The acceptance criteria for verification and validation testing were identified in the individual test cases as the expected results. Test cases were comprised of scenarios with multiple steps along with their respective expected results. When all the steps were tested and the actual results matched the expected results, the test case result was passed." "Test data demonstrated that the device met all acceptance requirements, and that the subject device is as safe, as effective, and performs as well as or better than the predicate device." "The addition of TEG Manager 2.0.0 as an optional accessory to the TEG 6s Analyzer does not render the device non-substantially equivalent because non-clinical testing has demonstrated that the subject device is as safe and effective as the predicate and the results of verification and validation have not raised different questions of safety and effectiveness than the predicate."
    Substantial Equivalence: The device, with the optional accessory, must be as safe and effective and perform as well as or better than the predicate device, without raising new questions of safety or effectiveness."The data and information provided in the submission support a substantial equivalence determination for the TEG 6s Analyzer and the proposed TEG 6s Analyzer with TEG Manager as an optional accessory." The document implies that the TEG Manager, being primarily a remote viewer of existing data from the predicate device, does not introduce new analytical or interpretive data and thus maintains the safety and effectiveness of the system.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of clinical data for a direct performance evaluation of the TEG Manager 2.0.0 as an AI/ML device. The testing described is primarily software verification and validation (V&V). Therefore, there are no details on clinical sample size or data provenance (country of origin, retrospective/prospective) for a performance study. The focus is on ensuring the software functions as intended and accurately displays data from the TEG 6s Analyzer.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Since the "test set" refers to software V&V rather than clinical performance data requiring expert adjudication, there is no mention of experts establishing ground truth or their qualifications. Software testing typically relies on predefined requirements and expected outputs rather than expert consensus on clinical findings.

    4. Adjudication Method for the Test Set

    No adjudication method (e.g., 2+1, 3+1) is mentioned because the V&V of the TEG Manager 2.0.0 does not involve clinical interpretation or a "ground truth" that would require expert consensus. The software's function is to accurately display data generated by the TEG 6s Analyzer.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is described. The TEG Manager 2.0.0 is an auxiliary viewing and management tool for data already generated by the TEG 6s Analyzer. It does not involve human readers interpreting images or data with and without AI assistance to assess an effect size.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The TEG Manager 2.0.0 is described as an "optional accessory" that provides "remote viewing of current and historical patient tracing and test results created by the TEG 6s analyzers, and administration of all connected TEG 6s devices." It "does not provide any additional analytical or interpretive data outside of the TEG analyzer capabilities." Therefore, its performance is intrinsically tied to the TEG 6s Analyzer, and it does not operate as a standalone diagnostic algorithm in the way an AI/ML device typically would. Its "performance" would be related to its ability to accurately reflect the analyzer's output and manage devices, which is covered by the software V&V.

    7. Type of Ground Truth Used

    For the TEG Manager 2.0.0, the "ground truth" for the software validation would be the expected output values and display capabilities based on the data directly produced by the TEG 6s Analyzer and the specified software requirements. It's not clinical ground truth in the sense of pathology, outcomes data, or expert consensus on a diagnosis.

    8. Sample Size for the Training Set

    The document does not describe any training set as would be relevant for an AI/ML model. The TEG Manager 2.0.0 is a software application for data display and management, not a learning algorithm that undergoes training.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set for an AI/ML model, there is no discussion of how ground truth was established for such a set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163274

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-04-05

    (135 days)

    Product Code
    Regulation Number
    864.5700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K163274

    Trade Name: AggreGuide A-100 AA Assay, AggreGuide A-100 Instrument Regulation Number: 21 CFR 864.5700
    Product Code: JOZ 21 CFR 864.5700 Regulation Number:

    Classification Name: System, Automated Platelet
    |
    | Regulation Number | 864.5700

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AggreGuide A-100 AA Assay is a qualitative system to aid in the detection of platelet dysfunction due to aspirin ingestion by those 18 years of age or older in 3.2% citrated venous whole blood using the AggreGuide A-100 instrument. For professional use. This test is not for use in patients with underlying congenital platelet abnormalities, patients with non-aspirin induced acquired platelet abnormalities or in patients receiving non-aspirin anti-platelet agents. The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician.

    Device Description

    The AggreGuide A-100 is a laser light scattering system that detects the level of platelet aggregation induced by arachidonic acid agonist in whole blood in motion. The system consists of an instrument and a disposable assay cartridge with pre-loaded freeze dried agonist. A whole blood sample is added to a disposable cartridge that is preloaded with freeze dried arachidonic acid agonist (AA) in a reaction chamber for an individual test. The amount of platelet aggregation is measured by detecting the laser light scattering caused by platelet aggregates. Aspirin is known to inhibit the level of platelet aggregation, or activity, when blood is mixed with arachidonic acid.

    AI/ML Overview

    My apologies, but this document does not contain the information requested in your prompt regarding acceptance criteria and the study that proves the device meets them, especially in the context of a conventional medical device study (e.g., diagnostic accuracy, comparative effectiveness).

    Here's why and what the document does provide:

    Why the requested information is not present:

    • This is a 510(k) Substantial Equivalence Submission for MODIFICATIONS, not an original submission or a study report. The document details changes made to an already cleared device (K122162). Its purpose is to demonstrate that the modified device is still substantially equivalent to the original, cleared device, and that the modifications do not raise new questions of safety or effectiveness.
    • The document focuses on verification and validation of changes: The "Summary of Performance Data" (Section 7) explicitly states that "The risk analysis method used to assess the impact of the modifications was ISO 14971:2007, 'Medical devices -- Application of risk management to medical devices'." It then lists results for specific changes: laser diode part replacement, QC2 implementation, software updates, shelf life extension, and User's Manual revision.
    • No clinical performance data was collected for this submission: Section 8 explicitly states, "The modifications did not require collection of clinical performance data." This directly contradicts the need for most of the information you asked for (e.g., sample size for test set, ground truth, expert adjudication, MRMC studies, standalone performance).

    What the document does provide (related to your request, but not directly answering it):

    • Device Name: AggreGuide A-100 AA assay, AggreGuide A-100 Instrument
    • Indications for Use: "The AggreGuide A-100 AA Assay is a qualitative system to aid in the detection of platelet dysfunction due to aspirin ingestion by those 18 years of age or older in 3.2% citrated venous whole blood using the AggreGuide A-100 instrument. For professional use. This test is not for use in patients with underlying congenital platelet abnormalities, patients with non-aspirin induced acquired platelet abnormalities or in patients receiving non-aspirin anti-platelet agents. The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician."
    • Predicate Device: AggreGuide A-100 AA assay and AggreGuide A-100 Instrument, 510(k) document number K122162.
    • Specific Modifications:
      • Laser Diode Part: Changed from Sanyo DL-4140-001S to QSI QL7816SA-L
      • Quality Control Cartridge: Modified to QC2 cartridge
      • Software: Updated from Version 4.91 to Version 5.10
      • Test Cartridge Shelf Life: Extended from 12 months to 18 months
      • User's Manual: Revised from LBL-0014-C to LBL-0014-E

    Given this, I cannot fill in the table or provide the specific details you requested because the document is a regulatory submission for changes to an existing device, and clinical performance data was explicitly not required for this submission. The "acceptance criteria" discussed are related to the successful verification and validation of these specific hardware/software/shelf-life changes, not the overall diagnostic accuracy of the device against a clinical ground truth.

    If you had a document for the original 510(k) clearance (K122162), it might contain some of the performance data you're looking for, but this particular document does not.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161329

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-12-02

    (204 days)

    Product Code
    Regulation Number
    864.5700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K161329

    Trade/Device Name: Chrono-log Platelet Aggregometer, Model 490 4+4 Regulation Number: 21 CFR 864.5700
    Usual Name - Platelet Aggregometer

    Classification Name - System, Automatic Platelet Aggregation 21CFR 864.5700
    Aggregometer, Lumi-Aggregometer

    Classification Name - System, Automatic Platelet Aggregation 21CFR 864.5700

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Chrono-log Model 490 4+4 Aggregometer is intended for use for in-vitro diagnostic use for measuring Platelet Aggregation in Platelet Rich Plasma.

    This device is intended to be used in a clinical laboratory environment by laboratory technicians.

    For use only with light transmission aggregometry assays cleared for use with the Chrono-log Platelet Aggregometry systems.

    Device Description

    The Chrono-log™ Model 490 4+4 Aggregometer measures platelet function on patient samples using LTA which measures a change in optical density of platelet rich plasma. The Model 490 4+4 is also used to run the Ristocetin Cofactor Assay to aid in diagnosis of patients with von Willebrand disease. The instrument comes with a starter kit of reagents and supplies. The output of the Model 490 4+4 can be connected to either a strip chart recorder or to a Computer. Software is provided with the computer interface option. The computer interface option is used to collect data only. The computer is not used for diagnosis or treatment and does not have any control over or input into the Model 490 4+4 Aggregometer.

    LTA or Born method of platelet aggregation measures the change in optical density of a Platelet Rich Plasma (PRP) sample in comparison to optical density of a Platelet Poor Plasma (PPP) sample. The PRP sample, platelets in a suspension of plasma, is isolated from an anticoagulated blood sample by a relatively low centrifugation. The PPP sample is prepared by centrifuging the blood sample at a relatively high force. The Chrono-log sample chambers are designed so that a beam of infra red light shines through two cuvettes. one containing PRP (the sample) and one containing PPP (the reference). Silicon photodiodes detect the light able to pass through the samples: PRP is arbitrarily considered to be 0% light transmission or 0% aggregation: PPP is considered to be 100% light transmission or 100% aggregation. When a stimulus is added to the cuvette containing PRP and the platelets respond forming aggregates, more light is allowed to pass through the PRP sample. The change in light transmission, recorded over time, shows a trend towards the platelet poor plasma, or 100% light transmission. A graphical tracing of the change in optical density during the course of platelet aggregation is produced either on a strip chart recorder or on a computer using Chrono-log provided software. This device is designed to be used in the clinical laboratory as an in vitro diagnostic tool. The 490 4+4 varies from the predicate devices only in the number of channels.

    AI/ML Overview

    The document describes the acceptance criteria and the study proving the device, Chrono-log™ Platelet Aggregometer, Model 490 4+4, meets these criteria, primarily through demonstrating substantial equivalence to a predicate device (Chrono-log™ Whole Blood Lumi-Aggregometer Model 700).

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by demonstrating "strong correlation" and "good agreement" with the predicate device, as well as maintaining performance specifications (though specific numerical targets for metrics like R-squared or bias are not explicitly stated as pre-defined acceptance criteria, they function as such in the reported performance metrics). The device's performance is gauged by the Coefficient of Variation (which seems to be used interchangeably with R-squared in this context, given the R2 mentions in the conclusion) between the new device and the predicate.

    Acceptance Criteria (Implied)Reported Device Performance (Coefficient of Variation / R$^2$)
    Strong correlation/equivalence to predicate for all samples0.9771 (for all 114 comparison tests)
    Strong correlation for normal samples (500 µL)0.9648
    Strong correlation for aspirin samples (500 µL)0.9943
    Strong correlation for normal samples (250 µL)0.9758
    Strong correlation for aspirin samples (250 µL)0.9871
    Strong correlation for Arachidonic Acid reagent0.9945
    Strong correlation for Epinephrine reagent0.9509
    Strong correlation for Collagen reagent0.9786
    Strong correlation for ADP reagent0.9421
    Strong correlation for Ristocetin reagent0.9863
    Clinically insignificant bias (Bland Altman Plot)Bias of -4.56, 2SD cut-off not beyond historical levels

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:
      • Optical Aggregation: 114 comparison tests in total.
      • Volume Comparison: A "small subset" of tests was used to compare 500uL and 250uL samples (8 normal samples in 500 µL, 8 aspirin samples in 500 µL, 8 normal samples in 250 µL, 8 aspirin samples in 250 µL).
    • Data Provenance:
      • The study involved "patient samples" from "four normal, healthy, drug-free subjects" and "a subject taking aspirin." Additionally, samples treated with ticagrelor and GPIIb/IIIa antagonist, and deficient vW plasma with lyophilized platelets were used to demonstrate abnormal results.
      • The document does not explicitly state the country of origin for the data/samples but implies it was conducted by the manufacturer, Chrono-log Corp., based in Havertown, PA, USA.
      • Retrospective or Prospective: The testing appears to be prospective or specially collected for this comparison study, as it describes the collection and testing of specific subject samples (normal, aspirin, treated with inhibitors) on both the new and predicate devices.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    The document does not describe the use of experts to establish a "ground truth" for the performance study. The study focuses on demonstrating equivalence to a predicate device using measured values of platelet aggregation, not on expert interpretation of results. The "ground truth" for the device's function is the direct measurement of optical density changes in plasma samples, interpreted as percentage aggregation.

    4. Adjudication Method for the Test Set

    Not applicable. The study compares objective quantitative measurements from two devices, not subjective interpretations requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This device is a measurement instrument (platelet aggregometer) and its performance is evaluated based on quantitative agreement with a predicate device, not on human reader performance or improvement with AI assistance.

    6. Standalone (Algorithm Only) Performance

    Yes, the study primarily demonstrates standalone performance of the device (Chrono-log Model 490 4+4) by comparing its measurements directly to those of the predicate device (Chrono-log Model 700). There is no AI algorithm involved; it's a direct hardware and measurement comparison. The software provided with the computer interface option is explicitly stated to "collect data only" and "is not used for diagnosis or treatment and does not have any control over or input into the Model 490 4+4 Aggregometer."

    7. Type of Ground Truth Used

    The "ground truth" is established by the measurements from the predicate device (Chrono-log Model 700). The study aims to show that the new device's measurements are substantially equivalent to those of the already legally marketed and established predicate device. It relies on the predicate device's accepted performance as the de facto standard for comparison.

    8. Sample Size for the Training Set

    Not applicable. This device is a medical measurement instrument, not an AI/ML model that requires a "training set" in the conventional sense. The "training" for the device would be its engineering design, calibration, and manufacturing processes.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI/ML model. The predicate device's established performance serves as the reference point for the new device's engineering design and verification.

    Ask a Question

    Ask a specific question about this device

    K Number
    K141427

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-07-24

    (420 days)

    Product Code
    Regulation Number
    864.5700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    |
    | Classification name: | Automated platelet aggregation system(21 CFR §864.5700

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VerifyNow PRUTest is a whole blood test used in the laboratory or point of care setting to measure the level of platelet P2Y12 receptor blockade. For in vitro diagnostic use. For professional use only.

    Device Description

    The VerifyNow System is a turbidimetric-based optical detection system that measures plateletinduced aggregation. The system consists of an instrument, a disposable test device and quality control materials. Quality control measures include an instrument based electronic quality control (EOC). two levels of wet quality controls (WOC), internal quality controls, and shipping controls. The instrument controls all assay sequencing, temperature, reagent-sample mixing and performs self-diagnostics. The degree of platelet function is determined and the result is displayed.

    The VerifyNow PRUTest device contains three lyophilized reagent pellets in separate reaction chambers within the test device: 1) ADP pellet consisting of a preparation of Fibrinogen and BSA coated beads, adenosine-5-diphosphate (ADP), prostaglandin E1 (PGE1), dye, buffer, and a preservative; 2) TRAP pellet (Internal Control) consisting of a preparation of iso-TRAP (Thrombin Receptor Activating Peptide), Fibrinogen and BSA coated beads, buffer, dye, and a preservative; and 3) No-Agonist Pellet (NAP) consisting of a preparation of BSA coated beads, dye, buffer, and a preservative.

    AI/ML Overview

    The provided text describes the performance studies for the VerifyNow PRUTest device, demonstrating its substantial equivalence to a predicate device (VerifyNow P2Y12 Assay).

    Here's an analysis of the acceptance criteria and the studies that prove the device meets these criteria:


    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly list a single table of "acceptance criteria" for all tests, but rather describes the criteria within the context of each study. Below is a compilation of the criteria and performance for the key studies described:

    Study/ParameterAcceptance CriteriaReported Device Performance
    Method ComparisonSlope not statistically significantly different from 1.0; Intercept not statistically significantly different from 0; Correlation (r) greater than 0.9.P2Y12 Rep 1 vs. PRUTest Rep 1: Slope = 1.01 (95% CI: 0.97–1.05), p-value = 0.56; Intercept = -0.77 (-8.00-6.50); Correlation R = 0.98. P2Y12 Avg. vs. PRUTest Avg.: Slope = 1.04 (95% CI: 1.00-1.07), p-value = 0.07; Intercept = -4.57 (-11.2-2.05); Correlation R = 0.98. Performance met criteria.
    Precision (WQC Level 1)PRU result ≤ 30 (indicates negative control/no aggregation).All WQC Level 1 results fell within the acceptable range of ≤ 30 PRU. At the nine-month time point, the range of PRU values generated with WOC Level 1 did not exceed 30 PRU.
    Precision (WQC Level 2)Within-run, Between-run, Between-day, Between-lot, and Total %CVs for WQC Level 2 (specific values are not given as general acceptance criteria, but implied by the successful analysis against CLSI EP05-A2). For stability, Percent recovery ≥ 90 to ≤ 110% of baseline PRU.Between-lot: Total SD = 22.2, Total %CV = 7.4. Between instrument: Total SD = 24.2, Total %CV = 8.0. For stability, percent recovery of WQC Level 2 passed acceptance criterion (≥ 90 to ≤ 110%) at nine months.
    Precision (Whole Blood - Multiple Lot)Individual CVs for Within-run, Between-run, Between-day, Between-lot components each < 10% (for most categories, N/A for Between-day for specific donors).Most individual donor results showed component %CVs < 10%, though some Between-day values were higher (e.g., Donor 1 Between-day %CV 15.0). All specific donor data are provided in Table H. However, the overall conclusion is not explicitly stated if all individual criteria were met for every donor.
    Precision (Whole Blood - Multiple Instrument)Individual CVs for Within-run, Between-run, Between-day, Between-lot components each < 10% (for most categories, N/A for Between-day for specific donors).Similar to multiple lot, most individual donor results showed component %CVs < 10%, with some exceptions (e.g., Donor 7 Between-run %CV 17.7 and Between-day %CV 57.1). All specific donor data are provided in Table I.
    Open Pouch Device Stability (Whole Blood)Percent recovery of VerifyNow PRUTest values ≤ 10% deviation from baseline.Met acceptance criteria after exposure to 61-73% RH and 21-24°C temperatures for 10 hours.
    Open Pouch Device Stability (WQC Level 2)PRU result remained within the allowable range (230–362).Remained within the allowable range for 10 hours under the same conditions.
    Sample Wait TimeNot explicitly stated, but the study determined an appropriate value.Determined to be 10 minutes (minimum storage time after venipuncture).
    Sample StabilityNot explicitly stated, but the study determined an appropriate value.Determined to be four hours (blood sample must be assayed prior to four hours from draw time).
    Interfering SubstancesPercent recovery which deviated less than 10% from baseline PRUTest values.Each interferent tested showed percent recovery deviating less than 10% from baseline PRUTest values. None of the substances significantly interfered.
    Reagent Stability (WQC Level 1)PRU result < 30.At the nine-month time point, the range of PRU values generated did not exceed 30 PRU. All test intervals met criteria.
    Reagent Stability (WQC Level 2)Percent recovery ≥ 90 to ≤ 110% of the baseline PRU result.At the nine-month time point, percent recovery passed criteria. All lots fell within ± 10% of the control condition.
    Reagent Stability (Donor Whole Blood)Percent recovery ≥ 90 to ≤ 110% of the baseline PRU result.At the nine-month time point, percent recovery passed criteria. Aggregated lot data fell within ± 10% of the control condition.
    Reagent Stability (Inhibited Whole Blood)Percent recovery ≥ 90 to ≤ 110% of the baseline PRU result.At the nine-month time point, percent recovery passed criteria. Aggregated lot data fell within ± 10% of the control condition.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Method Comparison: 119 samples. The origin is not specified, but the study involved "three investigational intended use sites." It is implied to be prospective, collected for the purpose of the study.
    • Reference Range Study: 152 evaluable healthy donors. Provenance not specified, but likely collected prospectively for the study.
    • Expected PRUTest values (Treatment-naïve ACS patients): 84 patients. Provenance not specified, likely collected prospectively for the study.
    • Expected PRUTest values (Clopidogrel-treated ACS patients): 71 patients. Provenance not specified, likely collected prospectively for the study.
    • Precision Studies (WQC Level 2): 240 tests (for both Between-lot and Between-instrument). Likely collected prospectively for the study.
    • Precision Studies (Whole Blood): 9 donors (each sampled multiple times, e.g., 30 to 120 reps per donor across days/runs/lots/instruments). Provenance not specified, likely collected prospectively for the study.
    • Open Pouch Device Stability: Whole blood samples (n=3) and WQC Level 2. Likely collected prospectively for the study.
    • Interfering Substances: Samples from 11 different donors. Likely collected prospectively for the study.
    • Reagent Stability (WQC Level 1, 2): Multiple lots (3 each of WQC levels and reagent devices), tested at time zero, 3, 6, and 9 months (number of individual tests not specified but for WQC Level 1 at 9 months, N=108). Likely collected prospectively for the study.
    • Reagent Stability (Donor Whole Blood): Whole blood samples from normal donors (specific N not specified, but tested across 3 reagent lots at 0, 3, 6, 9 months). Table N specifically shows data for 2 donors, at 9-month time point (N=3 for each lot/donor combination, total of 18 samples per donor for the specific comparison). Likely collected prospectively for the study.
    • Reagent Stability (Inhibited Whole Blood): 2 donors (Donors 399 and 1077). Similar breakdown as above, tested across 3 reagent lots at 9 months. Likely collected prospectively for the study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    The document describes in vitro diagnostic (IVD) device performance. For this type of device, ground truth is typically established by comparative measurements against a recognized reference method or by established laboratory-based assays (like the predicate device or clinical reference ranges) rather than expert consensus on image interpretation or clinical diagnosis. Therefore, "experts" in the context of establishing ground truth as typically understood for imaging AI models (e.g., radiologists) are not applicable here. The ground truth relies on the inherent biochemical mechanism of the device (platelet aggregation) and comparison to established methods.

    4. Adjudication Method for the Test Set:

    Not applicable in the conventional sense for image or clinical diagnosis adjudication. The "ground truth" for the analytical performance studies (method comparison, precision, stability, interference) is derived from quantitative measurements and statistical analysis, often against the predicate device or a control condition.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is an automated platelet aggregation system, an IVD. It is not an AI-assisted diagnostic tool that aids human "readers" (like radiologists or pathologists) in interpreting complex data. The device itself performs the measurement and outputs a quantitative result (PRU value).

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Yes, the studies describe the standalone performance of the VerifyNow PRUTest system. The system "automatically dispenses... blood... with no blood handling required by the user." The instrument "controls all assay sequencing, temperature, reagent-sample mixing and performs self-diagnostics. The degree of platelet function is determined and the result is displayed." This indicates it is an automated, standalone system.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc):

    The ground truth used primarily consists of:

    • Comparative Analytical Performance: Measurements generated by the legally marketed predicate device (VerifyNow P2Y12 Assay) and comparison against it.
    • Established Reference Materials/Controls: Wet Quality Controls (WQC) with known characteristics.
    • Clinically Relevant Ranges: Reference ranges and expected values established statistically from healthy donors and patient populations (treatment-naïve, treated with clopidogrel).
    • Statistical Analysis: Adherence to CLSI guidelines (EP28-A3c, EP05-A2, EP07-A2) for establishing reference intervals, evaluating precision, and testing interference.

    8. The Sample Size for the Training Set:

    The document does not specify a "training set" in the context of an algorithm that learns from data in the way a machine learning model does. This device is described as a turbidimetric-based optical detection system, implying a fixed algorithm and chemical reagent system rather than a machine learning algorithm requiring a distinct training phase with a labeled dataset. Its development would involve optimization of reagents, optical detection parameters, and algorithms to convert optical signals to PRU values, which is a different paradigm than typical AI/ML training.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no "training set" in the AI/ML context, this question is not directly applicable. The "ground truth" for the device's design and calibration would have been established during its development through a combination of:

    • Biochemical principles of platelet aggregation.
    • Empirical optimization of reagent concentrations and measurement parameters.
    • Comparison to gold standard methods or other validated platelet function assays during the research and development phase to ensure the system accurately reflects platelet P2Y12 receptor blockade.
    Ask a Question

    Ask a specific question about this device

    K Number
    K143332

    Validate with FDA (Live)

    Date Cleared
    2015-04-14

    (145 days)

    Product Code
    Regulation Number
    870.1120
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Performance Standards:

    No performance standards have been established for the FAST Sphyg under section 514 (864.5700

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is a sphygmomanometer intended to be used with a stethoscope or Doppler for indirect measurement of arterial blood pressure. This device includes an aneroid gauge. Koven Technology, Inc. intends to provide the device for use by healthcare providers. Our product is for both hospital and clinical use and are intended for pediatric through adults, excluding neonates.

    Device Description

    The FAST Sphyg by Koven is an aneroid sphygmomanometer that uses the auscultatory blood pressure technique in combination with a stethoscope or Doppler and occluding cuff to determine systolic and diastolic blood pressure measurement. The FAST Sphyg includes an aneroid qauge, as well as a battery powered air pump so the user can control the inflation. When the activation button is pushed on the device, the air pump inflates the cuff. When the activation button is released, inflation stops, but the system remains pressurized. The aneroid gauge on the device reflects this increase in pressure, which is displayed in mmHg. A release valve is activated through a trigger switch to manually deflate the cuff and notate the blood pressure measurement.

    AI/ML Overview

    The provided text describes the FAST Sphyg by Koven, an aneroid sphygmomanometer intended for indirect measurement of arterial blood pressure. The document focuses on establishing substantial equivalence to a predicate device and adherence to relevant performance standards.

    Here's an analysis of the provided information concerning acceptance criteria and supporting studies:

    1. A table of acceptance criteria and the reported device performance:

      The document does not present a single, consolidated table explicitly labeled "acceptance criteria" and "reported device performance" in the context of clinical accuracy for blood pressure measurement. Instead, it refers to compliance with performance standards, primarily ISO 81060-1:2007 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type.

      However, the document lists other technical standards that the device meets, which can be considered acceptance criteria for electrical safety and electromagnetic compatibility:

      Standard / TestAcceptance Criteria / Test LevelReported Device Performance / Compliance Level
      ISO 81060-1:2007Requirements and test methods for non-automated measurement typeDesigned and tested to meet this standard. (Specific accuracy metrics not detailed in this excerpt)
      IEC 60601-1 (2007) - Electrical SafetyClass II device (Protection class against electric shock)Class II device, Internally powered equipment
      Type B applied part (Protection grade against electric shock)Type B applied part
      IEC 60601-1-2 (2007) - Electromagnetic Emissions
      RF Emissions CISPR 11Group 1Group 1
      RF Emissions CISPR 11Class BClass B
      Harmonic Emissions IEC 61000-3-2Class AClass A
      Flicker Emissions IEC 61000-3-3Not ApplicableNot Applicable
      IEC 60601-1-2 (2007) - Electromagnetic Immunity
      Electrostatic Discharge (ESD) IEC 61000-4-2±6kV Contact, ±8kV Air±6kV Contact, ±8kV Air
      Electrical Fast Transient/burst IEC 61000-4-4±2kV for power supply lines, ±1kV for input/output lines±2kV for power supply lines, ±1kV for input/output lines
      Surge IEC 61000-4-5±1kV differential mode, ±2kV common mode±1kV differential mode, ±2kV common mode
      Voltage dips, short interruptions, and variations< 5% UT for 0.5 cycles, 40% UT for 5 cycles, 70% UT for 25 cycles, < 5% UT for 5 s< 5% UT for 0.5 cycles, 40% UT for 5 cycles, 70% UT for 25 cycles, < 5% UT for 5 s
      Power frequency (50/60 Hz) magnetic field3 A/m3 A/m
      Conducted RF IEC 61000-4-63 Vrms (150 kHz to 80 MHz)3 Vrms
      Radiated RF IEC 61000-4-33 V/m (80 MHz to 2.5 GHz)3 V/m

      For the primary function of blood pressure measurement accuracy, the document states, "the FAST Sphyg has been designed and tested to meet the following standards: ISO 81060-1:2007 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type." However, the specific accuracy acceptance criteria defined within ISO 81060-1 (e.g., mean difference and standard deviation between the device and reference method) and the actual reported performance data against these criteria are not detailed in this excerpt.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      This information is not provided in the excerpt. The document states that the device was "designed and tested to meet" ISO 81060-1:2007, but it does not describe the specific study, patient cohort, or sample size used for that testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      This information is not provided in the excerpt. For blood pressure measurement, the "ground truth" would typically come from a trained observer using a reference sphygmomanometer (e.g., mercurial sphygmomanometer or direct arterial measurement) following specific protocols, often involving multiple observers to mitigate bias (as per ISO standard requirements). However, no details on this are given.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      This information is not provided in the excerpt. Given that the device is a sphygmomanometer, adjudication would refer to how multiple measurements or observations by reference methods are reconciled, if applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      This is not applicable to the FAST Sphyg by Koven. This device is a mechanical aneroid sphygmomanometer with a battery-powered pump, not an AI-assisted diagnostic tool for image analysis or interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      This is not applicable to the FAST Sphyg by Koven. It is a manually operated device where a healthcare provider uses a stethoscope or Doppler to interpret the Korotkoff sounds. There is no "algorithm" in the sense of an automated analytical process that provides a standalone reading without human involvement in the detection of sounds.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      For blood pressure measurement devices, the ground truth is typically established by simultaneous measurements using a recognized reference method, such as a mercurial sphygmomanometer or direct invasive arterial pressure measurement, performed by trained observers. The document, however, does not explicitly state the specific "type of ground truth" used in its testing beyond mentioning compliance with ISO 81060-1:2007, which dictates such methods.

    8. The sample size for the training set:

      This information is not applicable/not provided. As an analog medical device (aneroid sphygmomanometer), the FAST Sphyg does not involve machine learning or AI models, and therefore does not have a "training set" in the computational sense.

    9. How the ground truth for the training set was established:

      This information is not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140893

    Validate with FDA (Live)

    Date Cleared
    2015-01-03

    (270 days)

    Product Code
    Regulation Number
    864.5700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    CORA (Coagulation Resonance Analysis) System with Platelet Mapping Assay Regulation Number: 21 CFR 864.5700
    Arachidonic Acid) ●

    Common name (usual name): Whole Blood Hemostasis System

    Classification Name: 21CFR 864.5700

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CORA PlateletMapping System is intended for in vitro diagnostic use to provide qualitative assessment of platelet function. The CORA System records the kinetic changes in a sample of heparinized whole blood as the sample clots.

    The CORA System PlateletMapping Assay Cartridge provides four channels of dried-in-place reagents, HKH (Kaolin with Heparinase), Activator F, AA and ADP (one reagent in each channel). In combination, MA parameter results from these four reagents are used to calculate the parameters platelet % Inhibition and % Aggregation for AA and ADP.

    Results from the CORA analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests.

    The CORA System with CORA PlateletMapping Assay Cartridge is indicated for use with adult patients where an evaluation of their blood hemostasis properties is desired. Hemostasis evaluation with the CORA PlateletMapping System is used to assess clinical conditions in cardiovascular surgery and cardiology procedures to assess hemorrhage or thrombosis conditions.

    Device Description

    The CORA PlateletMappingSystem consists of a four-channel diagnostic analyzer with integrated computer module, system reagents (ActF, AA, ADP and HKH) and microfluidic test cartridge. Reagents are dried-in-place within the cartridges during manufacturing.

    To perform a test, a disposable CORA PlateletMapping Assay Cartridge is inserted into the analyzer. Blood is added to an entry port on the cartridge and drawn into the cartridge under analyzer control. The amount of the sample drawn into the cartridge is automatically determined by the volume of the blood chambers in the cartridge. Once in the disposable, the sample is metered into as many as four separate analysis channels, depending upon the assay being performed. Reconstitution of reagents dried within the cartridge is accomplished by moving the sample back and forth through reagent chambers, under the control of microfluidic valves and bellows within the cartridge. After each sample has been mixed with reagent, it is delivered to a test cell where it is monitored for changes due to coagulation. Excess sample material is moved under microfluidic control into an enclosed waste chamber within the cartridge.

    The CORA technology is based on a disposable containing up to four independent measurement cells. Each cell consists of a short vertically-oriented injection molded tube (ring) with a diameter of 2.5mm and a length of 4.5mm. Detection of clotting in the CORA System is performed optically. Under control of the analyzer, approximately 20uL of prepared sample is delivered to the tube, where a meniscus naturally forms at each end of the tube. The tube is positioned so that the lower meniscus partially blocks light traveling from a collimated source toward a photodiode.

    During testing, a piezoelectric actuator drives the measurement cell(s) through a motion profile composed of summed sinusoids at different frequencies. The profile has a maximum amplitude of under 10um and contains frequencies from 10-500Hz. Some, but not all, of the measurement cell motion will induce motion in the sample meniscus, which will be detected by the photodiode. The resulting motion of the meniscus is monitored optically and analyzed by the analyzer to calculate the resonant frequency and modulus of elasticity (stiffness) of the sample. By performing a Fast Fourier Transform (FFT) on meniscus motion data, it is possible to determine the frequencies of input motion that caused the greatest deflection of the sample (these are called the resonant frequencies).

    Resonance is the tendency of a material or structure to oscillate with greater amplitude at some frequencies than others. The exact frequencies at which resonance occurs will depend on the stiffness and mass of the sample. Stiffness, in turn, is a function of a material's modulus of elasticity and the boundary conditions to which the material is exposed, such as the geometry and materials of a test cell. By holding these boundary conditions and sample mass constant from run to run, the CORA System allows direct comparison of elasticity between samples.

    In a typical test, blood that has been delivered to the measurement cell will not clot for several minutes. During this time the sample has no inherent stiffness except that provided by surface tension, and since this remains constant the measured resonant frequencies will not change. Once clotting begins, however, the elastic modulus and thus the resonant frequencies increase rapidly. In tests where clotting does not occur, the resonant frequency of the sample will not change. During coagulation, however, a clot will bind to the test tube (ring) and the resonant frequency will rise with increasing firmness of the clot. The CORA Analyzer collects meniscus motion data, tracks changing resonant frequencies and analyzes the frequency data to provide parameters describing the clot. Results are presented in a format identical to the TEG 5000.

    AI/ML Overview

    The CORA System with PlateletMapping Assay is intended for in vitro diagnostic use to provide a qualitative assessment of platelet function. The system records kinetic changes in heparinized whole blood samples as they clot, calculating % Inhibition and % Aggregation for AA and ADP parameters.

    Here's an analysis of the acceptance criteria and the study performance for the CORA System with PlateletMapping Assay:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria for the clinical performance. Instead, it presents the device's sensitivity and specificity and compares them to the predicate device, the TEG 5000 Platelet Mapping Assay.

    MetricCORA System PerformancePredicate Device (TEG 5000) Performance
    ADP
    Sensitivity74.5% (95% CI: 64.7-82.8%)94.9% (95% CI: 88.5-98.3%)
    Specificity82.9% (95% CI: 77.7-87.4%)39.0% (95% CI: 29.7-49.1%)
    AA
    Sensitivity84.0% (95% CI: 77.8-89.0%)88.4% (95% CI: 82.8-92.7%)
    Specificity86.5% (95% CI: 80.4-91.2%)50.0% (95% CI: 29.1-70.9%)

    For analytical precision, the reported data indicates the variability (SD and %CV) across various factors (reagent lot, operator, analyzer, day, and repeatability) for MA parameters in HKH, and indicates that the percent positive and negative agreement for AA and ADP % aggregation inhibition at low and high level is 100%. However, no specific acceptance criteria for precision are provided.

    2. Sample Size Used for the Test Set and Data Provenance

    • Reference Range Determination (Clinical Performance):

      • Sample Size: Up to 55 normal volunteer subjects were taken at each of the three clinical sites, totaling approximately 150 samples.
      • Data Provenance: Prospective, collected from normal volunteer subjects by three clinical sites representing demographic populations of the three areas regarding age, race, and gender.
    • Method Comparison (Clinical Performance):

      • Sample Size: Not explicitly stated, but the study was conducted on "patient samples" from "surgical patients and normal donors (for CORA)." The sensitivity and specificity percentages are derived from these samples.
      • Data Provenance: The patients were undergoing heart surgery or PCI procedures, with blood samples drawn pre- and post-surgery and in the ICU. This indicates a prospective collection within a clinical setting. The study was conducted at three clinical sites.
    • Analytical Precision (Non-Clinical Performance):

      • Sample Size: For HKH, blood draws from 3 donors (Hypo, Normal, Hyper). For AA and ADP Percent Aggregation and Inhibition, blood draws from 2 donors (Normal, Abnormal). Each testing scenario involved 5 non-consecutive days, 2 operators, 3 reagent lots, 12 analyzers, and 2 replicates. This results in n=120 for each MA parameter level (Hypo, Hyper, Normal) for HKH.
      • Data Provenance: The testing was performed in Coramed's laboratory, suggesting internally generated data.
    • Interference (Non-Clinical Performance):

      • Sample Size: Not explicitly stated.
      • Data Provenance: Performed in Coramed's laboratory.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the use of experts to establish a "ground truth" in the traditional sense (e.g., radiologists interpreting images).

    • For the Reference Ranges, the ground truth is established by selecting "normal volunteer subjects" and then deriving the range from their results. This is based on a statistical approach to define normalcy within the tested population.
    • For the Method Comparison, the ground truth for determining disease status (e.g., platelet dysfunction) is implicitly established through clinical diagnosis of "surgical patients or PCI procedures" who are known to have conditions where blood hemostasis properties evaluation is desired. However, the exact method of confirming ground truth for individual results to calculate sensitivity and specificity (e.g., by another gold standard method, pathology, or expert clinical diagnosis) is not detailed. The comparison is made against the TEG 5000 as a predicate, which usually implies that the predicate serves as a reference, but a true independent ground truth for classification is not explicitly defined in the provided text.

    4. Adjudication Method (for the test set)

    No adjudication method involving experts is mentioned for clinical performance results. The sensitivity and specificity would be derived by comparing the CORA device's output against the "true" clinical status of the patients, or the predicate device's output, but the process of determining that "true" status is not elaborated in terms of an adjudication panel.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This is an in vitro diagnostic device for assessing platelet function, not an imaging device typically involving human readers interpreting results. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed or described in this document.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the studies presented appear to be standalone performance assessments of the CORA System. The "system" is an automated analyzer that generates quantitative results (MA, % Aggregation, % Inhibition). Its performance (precision, reference ranges, sensitivity, specificity) is evaluated directly, without explicitly describing a human-in-the-loop interaction in the context of the performance data. While the "results from the CORA analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests," this statement refers to the clinical application rather than the performance study design.

    7. The Type of Ground Truth Used

    • Reference Ranges: The ground truth is effectively derived from the measurements of a healthy reference population (normal volunteer subjects) to establish what is considered "normal."
    • Method Comparison: The ground truth for calculating sensitivity and specificity is implicitly clinical diagnosis of patients undergoing cardiovascular procedures, where blood hemostasis evaluation is desired. However, the specific gold standard or method used to classify each patient's 'true' platelet function status (e.g., by another established assay, pathology, or expert clinical assessment) against which the CORA results are benchmarked is not explicitly defined. The comparison also heavily relies on the predicate device (TEG 5000), suggesting that the predicate might serve as a de facto reference for classification if an independent gold standard was not available or used. It states "studies were conducted... on patient samples following CLSI EP09-A3 Guideline," which typically involves comparison to a reference method.

    8. The Sample Size for the Training Set

    The document does not explicitly describe a "training set" in the context of an AI/machine learning model. The CORA System is an automated diagnostic device based on physical measurement techniques (e.g., resonance, optical detection), not typically an AI system that requires a training set. The descriptions of "precision testing" and "reference ranges" are for analytical and clinical validation, not algorithm training.

    9. How the Ground Truth for the Training Set Was Established

    Since no "training set" for an AI/machine learning model is described, the question of how its ground truth was established is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K122162

    Validate with FDA (Live)

    Device Name
    AGGREGUIDE
    Manufacturer
    Date Cleared
    2013-12-20

    (518 days)

    Product Code
    Regulation Number
    864.5700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Classification Name: | System automated plateletaggregation |
    | Regulation Number: | 864.5700
    K122162

    Trade/Device Name: AggreGuide A-100 and AggreGuide A-100 AA Assay Regulation Number: 21 CFR 864.5700

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AggreGuide A-100 AA Assay is a qualitative system to aid in the detection of platelet dysfunction due to aspirin ingestion by those 18 years of age or older in 3.2% citrated venous whole blood using the AggreGuide A-100 instrument. For professional use. This test is not for use in patients with underlying congenital platelet abnormalities, patients with non-aspirin induced acquired platelet abnormalities or in patients receiving non-aspirin antiplatelet agents. The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician.

    Device Description

    The AggreGuide A-100 is a laser light scattering system that detects the level of platelet aggregation induced by arachidonic acid agonist in whole blood in motion. The system consists of an instrument and a disposable assay cartridge with pre-loaded freeze dried agonist. A whole blood sample is added to a disposable cartridge that is preloaded with freeze dried arachidonic acid agonist (AA) in a reaction chamber for an individual test. The amount of platelet aggregation is measured by detecting the laser light scattering caused by platelet aggregates. Aspirin is known to inhibit the level of platelet aggregation, or activity, when blood is mixed with arachidonic acid.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the AggreGuide A-100, based on the provided 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document primarily focuses on demonstrating substantial equivalence to a predicate device, rather than explicit pre-defined quantitative acceptance criteria for all performance aspects. However, it presents clinical performance results for sensitivity and specificity. Based on the clinical testing section, the implicit acceptance criteria are that the device demonstrates a clinically acceptable level of sensitivity and specificity for detecting aspirin-induced platelet dysfunction.

    Acceptance Criteria (Implicit)Reported Device Performance
    Sensitivity83% (115/138)
    Sensitivity Lower 95% CI76%
    Sensitivity Upper 95% CI89%
    Specificity (True Negative Rate)90% (151/167)
    Specificity Lower 95% CI85%
    Specificity Upper 95% CI94%
    False Positive Rate10% (16/167)
    False Negative Rate17% (23/138)

    (Note: The document directly states "Sensitivity" in a table, and then provides a table titled "Pre 325mg Aspirin" that effectively describes the true negative rate (specificity) and false positive rate. The "Post 325mg Aspirin" table provides the true positive rate (sensitivity) and false negative rate. I've used these to construct the performance table.)


    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:
      • Total screened: 169 healthy subjects.
      • For aspirin-induced dysfunction detection: 138 subjects were given aspirin and tested.
      • For baseline (aspirin absent) assessment: 167 subjects (after 2 exclusions).
    • Data Provenance: The document does not explicitly state the country of origin. It indicates it was a clinical testing conducted for the purpose of this submission. The nature of the study (administering aspirin to healthy subjects) suggests a prospective study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not specify the use of experts or their qualifications for establishing ground truth in the clinical study. It describes a "qualitative system to aid in the detection of platelet dysfunction due to aspirin ingestion." The ground truth appears to be established by:
    * Aspirin Administration: Whether a subject was given 325 mg of aspirin (to induce platelet dysfunction).
    * Clinical Observation/Reference Method (Implied): The study design implies that "aspirin ingestion" itself serves as the ground truth for platelet dysfunction, and the device's ability to detect this is assessed. There's no mention of a separate "gold standard" laboratory test or expert consensus.


    4. Adjudication Method for the Test Set

    The document does not describe any adjudication method (e.g., 2+1, 3+1). The performance data is presented as direct results from the AggreGuide A-100, compared against the known condition of aspirin ingestion or absence.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed as described in this summary. The study assessed the standalone performance of the AggreGuide A-100, not its effectiveness in assisting human readers.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance study was done. The clinical testing section directly evaluates the AggreGuide A-100 system's ability to detect platelet dysfunction due to aspirin ingestion. The device produces a "Platelet Activity Index (PAI)," and the study assesses how well this PAI correlates with the presence or absence of prescribed aspirin. There is no mention of a human in the loop interpreting the device's output and then making a diagnosis.


    7. The Type of Ground Truth Used

    The ground truth used was primarily based on aspirin administration.

    • "Aspirin 325 mg Present" was the ground truth for aspirin-induced platelet dysfunction.
    • "Aspirin 325 mg Absent" was the ground truth for normal platelet function (or at least, not aspirin-induced dysfunction).

    This is a form of clinical condition or intervention as ground truth, rather than pathology, expert consensus, or outcomes data in the traditional sense for diagnostic assays.


    8. The Sample Size for the Training Set

    The document does not mention a separate training set or its sample size. The description of performance testing focuses solely on the clinical validation study described. For a device like this (an in-vitro diagnostic instrument), "training" might refer to internal development and calibration, but a distinct "training set" for statistical model development and validation as seen in AI/ML products is not discussed.


    9. How the Ground Truth for the Training Set Was Established

    Since a separate training set is not explicitly mentioned as distinct from the reported clinical testing, the method of establishing its ground truth is not provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103555

    Validate with FDA (Live)

    Date Cleared
    2012-07-27

    (602 days)

    Product Code
    Regulation Number
    864.5700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    80469 Munich Germany

    Re: K103555

    Trade/Device Name: Verum Multiplate 5.0 Regulation Number: 21 CFR 864.5700

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Multiplate 5.0 aggregometer is intended for in vitro use to measure platelet aggregation in response to Arachidonic acid or ADP in citrated whole blood samples for the qualitative assessment of platelet function.

    The ADPtest reagent is a lyophilized preparation of adenosine-5-diphosphate for in vitro diagnostic use to measure platelet aggregation for the qualitative assessment of platelet function. For professional laboratory use only.

    The ASPItest reagent is a lyophilized preparation of arachidonic acid (AA) for in vitro diagnostic use to measure platelet aggregation for the qualitative assessment of platelet function. For professional laboratory use only.

    For use as an assayed quality control verification of the resistance measure of impedance aggregometry.

    Device Description

    The Multiplate® 5.0 measures platelet function in whole blood samples using electrical impedance. The Multiplate technology employs multiple electrodes in a disposable test cell. Four electrodes form two independent sensor units allowing for two measurements on the same sample. Five independent channels of the instrument allow for testing of multiple reagents or samples simultaneously.

    The instrument provides a five channel aggregometer and an integrated computer system with associated software and is connected to a computer screen, keyboard, mouse, and an electronic pipette. The software is used for data collection and is not used for diagnosis or treatment.

    Currently, two test reagents (ADP and Arachidonic acid) are available that activate platelets through specific platelet membrane receptor/signal transduction pathways in order to measure platelet function or alterations in function.

    AI/ML Overview

    The Multiplate 5.0 device measures platelet aggregation in whole blood samples. This device report focuses on its performance compared to a predicate device (Chrono-log Model 700) and its ability to assess platelet function.

    Here's a breakdown of the acceptance criteria and study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for agreement, sensitivity, or specificity. Instead, the "Performance" section reports the observed agreement and diagnostic accuracy against the predicate device and physician panel review, respectively. Therefore, the reported performance metrics can be considered the de facto "acceptance criteria" against which the device passed for substantial equivalence demonstration.

    Performance MetricReagent: Arachidonic acidReagent: ADP
    Agreement with Predicate Device
    Positive Percent Agreement (PPA)100% [96%, 100% CI]93% [81%, 97% CI]
    Negative Percent Agreement (NPA)65% [54%, 75% CI]54% [45%, 62% CI]
    Diagnostic Accuracy vs. Clinical History
    Sensitivity (ASPItest vs. Physician Panel)93% [85%, 97% CI]Not Applicable
    Specificity (ASPItest vs. Physician Panel)67% [44%, 84% CI]Not Applicable
    Sensitivity (ADPtest vs. Physician Panel)Not Applicable59% [46%, 70% CI]
    Specificity (ADPtest vs. Physician Panel)Not Applicable80% [63%, 90% CI]

    2. Sample Size and Data Provenance for the Test Set

    • Sample size for agreement study (Test Set 1): 171 patients.
    • Sample size for sensitivity/specificity study (Test Set 2): 91 patients.
    • Data Provenance: The document states "Multi-center studies were run," implying prospective data collection across multiple sites. There is no explicit mention of the country of origin of the data, but the submitting company is based in Germany. The data is implicitly prospective, as it was collected to compare the Multiplate 5.0 to the predicate device and against clinical status.

    3. Number of Experts and Qualifications for Ground Truth (Test Set)

    • Number of experts: The document mentions a "physician panel" for establishing Platelet Function Status for the sensitivity/specificity study. The exact number of physicians on this panel is not specified.
    • Qualifications of experts: Not specified beyond being a "physician panel." Their specific specialty (e.g., hematologist, critical care physician) or years of experience are not provided.

    4. Adjudication Method (Test Set)

    • For the agreement study with the predicate device, there was no independent adjudication method described. The comparison was direct measurement of the Multiplate 5.0 against the Chrono-log Model 700.
    • For the sensitivity/specificity study, the "Platelet Function Status of each sample was based upon a physician panel's review of each patient's clinical history." This implies a form of consensus or collective decision-making by the panel, but the specific adjudication method (e.g., unanimous, majority vote, 2+1) is not detailed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, an MRMC comparative effectiveness study that assesses human readers' improvement with vs. without AI assistance was not conducted. This device is an automated platelet aggregation system, not an AI diagnostic tool that assists human interpretation of images or data. The studies involved a direct comparison between devices or device output against clinical ground truth.

    6. Standalone Performance Study (Algorithm Only)

    Yes, a standalone study was performed. The reported performance metrics (agreement, sensitivity, specificity) reflect the performance of the Multiplate 5.0 device (the "algorithm only," as it's an automated system) without human interpretation or intervention in the measurement process itself. The software for data collection is explicitly stated as "not used for diagnosis or treatment," reinforcing that the device itself generates the results.

    7. Type of Ground Truth Used

    • For agreement studies: The ground truth was the results obtained from the predicate device (Chrono-log Model 700).
    • For sensitivity and specificity studies: The ground truth was the "Platelet Function Status" based on a physician panel's review of each patient's clinical history. This can be categorized as expert consensus / clinical diagnosis.

    8. Sample Size for the Training Set

    The document does not specify a sample size for a training set. Given that this is a 510(k) submission for a device measuring electrical impedance (a well-established physical principle) rather than an AI/machine learning algorithm that requires extensive training, it's highly probable that there wasn't a "training set" in the modern machine learning sense. The device's operational parameters would likely be derived from engineering principles and validation, rather than data-driven training.

    9. How Ground Truth for the Training Set Was Established

    As no "training set" is mentioned or implied for a machine learning context, the concept of establishing ground truth for a training set does not apply in this document. The device operates based on physical principles and pre-defined reagents, with performance validated against clinical samples and a predicate device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 3