Search Results
Found 26 results
510(k) Data Aggregation
(130 days)
Trade/Device Name: Medconn 8K Glycated Hemoglobin Test System
Regulation Number: 21 CFR 862.1373
Effectiveness)
This 510(k) Summary is being submitted in accordance with requirements of Title 21 CFR 862.1373
Common Name:** Hemoglobin A1c test system
Product Code: PDJ, LCP
Regulation Number: 21 CFR 862.1373
Glycated Hemoglobin Analyzer | D-100™ Hemoglobin Testing System | NA |
| Regulation number | 21 CFR 862.1373
| 21 CFR 862.1373 | SE |
| Regulatory Class | 2 | 2 | SE |
Page 8
Sponsor: Shanghai Medconn
Medconn 8K Glycated Hemoglobin Test System is intended for the quantitative determination of hemoglobin A1c (IFCC mmol/mol and NGSP %) in human whole blood or hemolysate with ion-exchange high performance liquid chromatography (HPLC) using the Medconn 8K HbA1c Assay Kit (HPLC) on the Medconn 8K Glycated Hemoglobin Analyzer, models MQ-8000 and MQ-8000PT.
Hemoglobin A1c measurements are used as an aid in diagnosis of diabetes, as an aid to identify patients who may be at risk for developing diabetes mellitus, and for the monitoring of long-term blood glucose control in individuals with diabetes mellitus.
The Medconn 8K Glycated Hemoglobin Test System is intended for the quantitative determination of hemoglobin A1c (IFCC mmol/mol and NGSP %) in human venous blood or hemolysate using ion-exchange high performance liquid chromatography (HPLC) on the Medconn 8K Glycated Hemoglobin Analyzer, models MQ-8000 and MQ-8000PT.
A high-pressure pumping system delivers a buffer solution to an analytical cartridge and detector. Whole blood samples undergo an automatic hemolysis and dilution process before being introduced into the analytical flow path. Manually hemolyzed and prediluted samples loaded in sample cups at designated location are directly introduced for analysis.
A programmed buffer gradient of increasing ionic strength delivers the sample to the analytical cartridge where the hemoglobin species are separated based upon their ionic interactions with the cartridge material and the buffer gradient. The separated hemoglobin species then pass through the flow cell where changes in the absorbance are measured at 415nm and recorded as a digital chromatogram.
The software performs an analysis of the hemoglobin peaks in the chromatogram, recording information including retention time, peak area, and relative peak area of the detected substance over the total peak area of all substances. Peaks identified as target analytes are calibrated to generate a report and a chromatogram for each sample.
Medconn 8K Glycated Hemoglobin Test System contains the following components:
- Medconn 8K Glycated Hemoglobin Analyzer
- Medconn 8K HbA1c Assay Kit (HPLC)
- Medconn Hemoglobin A1c Calibrator
- Medconn Hemoglobin A1c Control
- Medconn 8K HbA1c Column Kit (HPLC)
- Medconn HbA1c Haemolyser
N/A
Ask a specific question about this device
(266 days)
Trade/Device Name: Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01
Regulation Number: 21 CFR 862.1373
Effectiveness)
This 510(k) Summary is being submitted in accordance with requirements of Title 21 CFR 862.1373
Name:** Hemoglobin A1c test system
Product Code: PDJ, LCP
Regulation Number: 21 CFR Part 862.1373
The Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 is intended for in vitro diagnostic use for the quantitative measurement of % hemoglobin A1c (HbA1c) (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in human venous whole blood and hemolysate specimens using ion-exchange high performance liquid chromatography (HPLC).
This test is to be used as an aid in diagnosis of diabetes and identifying patients who may be at risk for developing diabetes, and for monitoring of long-term blood glucose control in individuals with diabetes mellitus.
The Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 is intended for in vitro diagnostic use for the quantitative measurement of % hemoglobin A1c (HbA1c) (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in human venous whole blood and hemolysate specimens using ion-exchange high performance liquid chromatography (HPLC). The Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 analyzer is to be used in the clinical laboratory setting. The analyzer hemolyzes and dilutes whole blood specimens and subsequently injects the diluted and hemolyzed sample into the injection valve and onto the TSKgel GR01 HbA1c Column. Manually hemolyzed and prediluted samples can be loaded in sample cups directly for analysis on the analyzer. Separation of hemoglobin fractions is achieved by using differences in ionic interactions between the cation exchange group on the column resin surface and the hemoglobin components by a step gradient elution with three elution buffers of different ionic strengths. Changes in light absorbance at 415nm caused by hemoglobin fractions eluting from the column are monitored. The GR01 software has chromatographic peak retention time windows for the six expected hemoglobin fractions, A1a, A1b, HbF, LA1c+, SA1c, and A0. In addition, the software has retention time windows for the presumptive identification of the common hemoglobin variants, D, S, C, and E. An analysis requires 50 seconds per sample to complete. A printout of the results includes the sample ID, date, percentage, and retention time of each hemoglobin fraction, the SA1c percentage along with a chromatogram of the elution pattern of the hemoglobin fractions. The analyzer is equipped with external USB ports. These can be used to store assay results, to update and backup program versions, to attach an external device, for example an external printer or an optional handheld barcode scanner, and for software installation or backup of data. The last 150,000 assay results are automatically saved to the analyzer's internal memory and can be retrieved as list data in .CSV format and chromatograms in .PDF format.
The Tosoh Automated Glycohemoglobin Analyzer HLC-723GR01 system consists of the following components.
- GR01 HbA1c Elution Buffer No. 1, No. 2, No. 3
- TSKgel® GR01 HbA1c Column
- Hemoglobin A1c Controls Levels 1 and 2
- 21 CFR 862.1660, Product Code JJX
- Hemoglobin A1c Calibrator Set
- 21 CFR 862.1150, Product Code JIT
- Hemolysis and Wash Solution (L) and (LL)
- 21 CFR 864.8540, Product Code GGK
N/A
Ask a specific question about this device
(269 days)
K242911**
Trade/Device Name: Medconn Glycated Hemoglobin Test system
Regulation Number: 21 CFR 862.1373
Test system
This 510(k) Summary is being submitted in accordance with requirements of Title 21 CFR 862.1373
Common Name:** Hemoglobin A1c test system
Product Code: PDJ, LCP
Regulation Number: 21 CFR 862.1373
Glycated Hemoglobin Analyzer | D-100™ Hemoglobin Testing System | NA |
| Regulation number | 21 CFR 862.1373
| 21 CFR 862.1373 | SE |
| Regulatory Class | 2 | 2 | SE |
| Product Code | PDJ, LCP | PDJ, LCP, JIT
Medconn Glycated Hemoglobin Test System is intended for the quantitative determination of hemoglobin A1c (IFCC mmol/mol and NGSP %) in human whole blood or hemolysate with ion-exchange high performance liquid chromatography (HPLC) using the Medconn HbA1c Assay Kit (HPLC) on the Medconn Glycated Hemoglobin Analyzer, models MQ3000 and MQ3000PT.
Hemoglobin A1c measurements are used as an aid in diagnosis of diabetes, as an aid to identify patients who may be at risk for developing diabetes mellitus, and for the monitoring of long-term blood glucose control in individuals with diabetes mellitus.
The Medconn Glycated Hemoglobin Test System is intended for the quantitative determination of hemoglobin A1c (IFCC mmol/mol and NGSP %) in human venous blood or hemolysate using ion-exchange high performance liquid chromatography (HPLC) on the Medconn Glycated Hemoglobin analyzer, models MQ-3000 and MQ-3000PT.
A high-pressure pumping system delivers a buffer solution to an analytical cartridge and detector. Whole blood samples undergo an automatic hemolysis and dilution process before being introduced into the analytical flow path. Manually hemolyzed and prediluted samples loaded in sample cups at designated location are directly introduced for analysis. A programmed buffer gradient of increasing ionic strength delivers the sample to the analytical cartridge where the hemoglobin species are separated based upon their ionic interactions with the cartridge material and the buffer gradient. The separated hemoglobin species then pass through the flow cell where changes in the absorbance are measured at 415nm and recorded as a digital chromatogram. The software performs an analysis of the hemoglobin peaks in the chromatogram, recording information including retention time, peak area, and relative peak area of the detected substance over the total peak area of all substances. Peaks identified as target analytes are calibrated to generate a report and a chromatogram for each sample.
Medconn Glycated Hemoglobin Test system contains the following components:
- Medconn Glycated Hemoglobin analyzer
- Medconn HbA1c Assay Kit (HPLC)
- Medconn Hemoglobin A1c Calibrator
- Medconn Hemoglobin A1c Control
- Medconn HbA1c Column Kit (HPLC)
- Medconn HbA1c Haemolyser
The Medconn Glycated Hemoglobin Test System is an in vitro diagnostic device intended for the quantitative determination of hemoglobin A1c (HbA1c) in human whole blood or hemolysate. It uses ion-exchange high-performance liquid chromatography (HPLC) on the Medconn Glycated Hemoglobin Analyzer (models MQ3000 and MQ3000PT). HbA1c measurements are used as an aid in the diagnosis of diabetes, to identify patients at risk for developing diabetes mellitus, and for monitoring long-term blood glucose control in individuals with diabetes mellitus.
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria for the Medconn Glycated Hemoglobin Test System's performance are embedded within the "Summary of Performance Data" section of the 510(k) submission, specifically through demonstrating acceptable precision, linearity, analytical specificity (interference), and method comparison against a standardized reference method. The performance criteria as stipulated by the Special Controls requirements for HbA1c systems that diagnose diabetes have clearly been met.
Here's a table summarizing the acceptance criteria and the reported device performance, derived from the provided text:
| Performance Characteristic | Acceptance Criteria (Implied by study design and FDA clearance) | Reported Device Performance (Summary) |
|---|---|---|
| Precision/Reproducibility | Based on CLSI EP05-A3. Acceptable CV% across different levels and sources of variation. | Whole Blood Mode: Total Precision (CV%) for 4 samples (5.11% to 12.25% NGSP): 0.86% to 1.59%.Dilution Mode: Total Precision (CV%) for 4 samples (5.20% to 12.15% NGSP): 0.87% to 1.43%. All CV% values are considered acceptable and within expected ranges for this type of test. |
| Linearity | Based on CLSI EP06-A2. Regression parameters (slope, intercept, R2) demonstrating linearity across the claimed measuring range (3.0% to 15.0% NGSP / 9.3-140.5 mmol/mol IFCC). Maximum Bias (mean) between observed and theoretical values should be minimal. | Whole Blood Mode: Slope 0.9967, Intercept 0.0318, R2 0.9999 (NGSP).Dilution Mode: Slope 0.9971, Intercept 0.0352, R2 0.9999 (NGSP).Maximum Bias (mean) observed: ± 0.1 (NGSP %) and ±1 (IFCC mmol/mol). Results support the claimed measuring range of 3.0% to 15.0% HbA1c. |
| Analytical Specificity (Interference) | Based on CLSI EP07-A3. Significant interference defined as a ± 6% change in %HbA1c value from the control. No significant interference by common endogenous substances, drugs, or common hemoglobin derivatives/variants at specified concentrations. | Endogenous Interference: No significant interference observed for Unconjugated Bilirubin (21.3 mg/dL), Conjugated Bilirubin (19.2 mg/dL), Lipemia (Triglycerides) (6000 mg/dL), RF (750 IU/mL), TP (21.0 g/dL).Drug Interference: No significant interference observed for a wide range of drugs (e.g., Ascorbic acid 100 mg/dL, Acetaminophen 200 mg/dL, Ibuprofen 500 mg/dL, etc.).Cross Reactivity with Hemoglobin Derivatives: No interference from Acetylated Hb, Carbamylated Hb, or Labile A1c at tested concentrations.Hemoglobin Variant Interference: No significant interference observed for HbC (≤ 40%), HbD (≤ 43%), HbS (≤ 42%), HbE (≤ 30%), HbA2 (≤ 6.2%) and HbF (≤33%) at concentrations tested. |
| Method Comparison (Accuracy vs. NGSP Reference) | Based on CLSI EP09c 3rd Edition. Acceptable regression analysis (Deming, Passing-Bablok) and Total Error (TE) values near clinical decision points, showing good agreement with a standardized reference method. | Whole Blood Mode:Deming Regression: y-Intercept 0.0004, Slope 1.0074 (95% CI for slope includes 1).Passing-Bablok Regression: y-Intercept 0.0500, Slope 1.0000 (95% CI for slope includes 1).Total Error (TE) at key HbA1c levels for Whole Blood Mode: 2.44% to 4.15% (depending on method and level).Dilution mode data:Deming Regression: y-Intercept 0.1940, Slope 0.9894.Passing-Bablok Regression: y-Intercept 0.1000, Slope 1.0000.Total Error (TE) at key HbA1c levels for Dilution Mode: 2.27% to 5.70% (depending on method and level).Correlates well with current cleared methods and NGSP standardized testing. |
| Traceability | To international and national standardization programs (IFCC, NGSP, DCCT). | Device standardization is traceable to IFCC reference calibrators and certified via NGSP Program. Results are provided in both NGSP (%) and IFCC (mmol/mol) units using the Master Equation. |
| Measuring Interval/Reportable Range | The device should accurately measure within its claimed range. | Claimed: 3.0% to 15.0% (NGSP) and 9.3-140.5 mmol/mol (IFCC). Linearity study results support this claimed measuring range. |
2. Sample Sizes Used for the Test Set and Data Provenance
-
Precision/Reproducibility:
- Four K2-EDTA whole blood samples (at concentrations near 5%, 6.5%, 8%, and 12% HbA1c) and two-level quality control products were used.
- For each sample, there were 720 measurements (analyzed in duplicate, twice a day, with three lots of reagents, over 20 non-consecutive days, on three analyzers).
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, the study design follows laboratory testing standards (CLSI guidelines), suggesting well-controlled, likely prospective, laboratory-based experiments.
-
Linearity:
- Eleven test samples (high, low, and 9 intermediate levels) were used. These were altered patient samples collected using K2-EDTA.
- Each was run three times with 3 lots of reagents on one instrument.
- Data Provenance: Not explicitly stated. Likely laboratory-prepared and tested.
-
Analytical Specificity (Interference):
- Endogenous Interference: Two EDTA whole blood sample pools (low (~6.5% HbA1c) and high (~8.0% HbA1c)). Ten replicates of each pool prepared with test and control samples were analyzed.
- Drug Interference: Two EDTA whole blood sample pools (low (~6.5% HbA1c) and high (~8.0% HbA1c)). Ten replicates of each drug prepared with test and control samples were analyzed.
- Hemoglobin Derivatives: Low (~6.5% HbA1c) and high (~8.0% HbA1c) whole blood EDTA samples. Each sample was analyzed using ten replicates.
- Hemoglobin Variant Interference:
- Study #1: 67 variant samples (HbS=11, HbC=10, HbD=13, HbE=12, HbA2=10, HbF=11).
- Study #2: 143 variant samples (HbS=25, HbC=25, HbD=25, HbE=25, HbA2=20, HbF=23).
- Samples were whole blood EDTA patient variant samples.
- Data Provenance: Not explicitly stated for endogenous/drug/derivative interference. For hemoglobin variants, they were "patient variant samples." It is implied these were collected and tested specifically for the study, likely in a controlled laboratory setting.
-
Method Comparison:
- 124 variant-free whole blood K2-EDTA samples.
- The distribution of samples spanned around clinical decision points (e.g., 25% of samples between 6-6.5%, 24.19% between 6.5-7%).
- Data Provenance: Not explicitly stated. The samples were patient samples. Given the nature of an FDA 510(k) submission, these would typically be from a well-defined cohort, likely retrospective collections or prospectively collected samples in a controlled clinical laboratory or research setting. The comparison was made against a "NGSP Secondary Reference Laboratory," implying adherence to established quality and calibration standards.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not provided in the given text. For an in vitro diagnostic device for HbA1c, the "ground truth" for the test set is established by the reference methods themselves (e.g., NGSP Secondary Reference Laboratory using a previously cleared HPLC HbAlc assay method - BIO-RAD D-100 analyzer, and other reference methods like Trinity Biotech Premier Hb9210, Bio-Rad VARIANT II Hemoglobin Testing System, and Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 for variant studies). The accuracy of these reference methods is inherent in their standardization and certification, rather than requiring expert adjudication of individual results as one might see in imaging studies.
4. Adjudication Method for the Test Set
Since the ground truth is established by well-defined and certified reference laboratory methods (e.g., NGSP Secondary Reference Laboratory using HPLC), there is no human adjudication method described or typically required for this type of quantitative diagnostic device. The comparison is purely analytical, comparing the numerical output of the candidate device to the numerical output of the reference method.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC study was done. MRMC studies are typically performed for devices where human interpretation is a critical part of the diagnostic process, such as medical imaging AI algorithms (e.g., radiologists reading images with and without AI assistance). This device is a quantitative in vitro diagnostic test, meaning the output is a numerical value directly from the instrument, not an interpretation by a human reader. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply here.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
This entire submission describes a standalone performance study of the device. The "Medconn Glycated Hemoglobin Test System" is an automated analytical system (analyzer plus reagents) that provides a quantitative result. Its performance evaluation (precision, linearity, interference, method comparison) is based on the result produced by the device itself, without human intervention in the interpretation of the final HbA1c value.
7. The Type of Ground Truth Used
The ground truth used for performance evaluation is established through:
- Certified/Standardized Reference Laboratory Methods: For the method comparison study, the samples were compared to results obtained by a "NGSP Secondary Reference Laboratory using a previously cleared HPLC HbAlc assay method (BIO-RAD D-100 analyzer)." For hemoglobin variant interference studies, comparison was made to reference methods "demonstrated to be free from interference with the hemoglobin variant being tested (Trinity Biotech Premier Hb9210, Bio-Rad VARIANT II Hemoglobin Testing System and Tosoh Automated Glycohemoglobin Analyzer HLC-723G8)."
- A Priori Known Values/Controlled Samples: For precision, linearity, and interference studies, ground truth is established by the careful preparation of samples with known or targeted concentrations (e.g., spiked samples, diluted samples, or control materials with assigned values).
This is a quantitative analytical ground truth established by established and validated laboratory methodologies, not expert consensus (as in clinical diagnosis from imaging), pathology (as in histopathology), or long-term outcomes data.
8. The Sample Size for the Training Set
The concept of a "training set" is not explicitly applicable in the context of this 510(k) submission for a traditional in vitro diagnostic device like an HbA1c assay. This device is based on a chemical/physical separation principle (HPLC) and established analytical methodologies, not machine learning or artificial intelligence algorithms that inherently require training data sets. The performance data presented (precision, linearity, interference, method comparison) constitutes analytical validation, not the evaluation of a "trained" model.
9. How the Ground Truth for the Training Set Was Established
As noted above, there isn't a "training set" in the machine learning sense for this device. The underlying principles and calibration of the HPLC system and assay kits are established through standard chemical and metrological practices, traceable to international reference materials and methods (IFCC, NGSP, DCCT). This traceability and analytical validation inherently serve the purpose equivalent to ensuring the "accuracy" or "correctness" of the system's measurements.
Ask a specific question about this device
(892 days)
Trade/Device Name: Hipro Glycosylated Hemoglobin (HbA1c) Test System Regulation Number: 21 CFR 862.1373
| 862.1373 |
| {5}------------------------------------ |
|---|
| Regulationnumber |
| 862.1373 |
The Hipro® Glycosylated Hemoglobin (HbA1c) Test System comprised of the Hipro Glycosylated hemoglobin (HbA1c) test kit and the HP-AFS/1 automatic immunoassay analyzer is used as an aid in diagnosis of diabetes mellitus, as an aid to identify patients who may be at risk for developing diabetes mellitus, and for the monitoring of long-term blood glucose control in individuals with diabetes mellitus. It is an in vitro diagnostics reagent system intended for quantitative determination of % hemoglobin A1c (DCCT/NGSP) in venous whole blood.
The Hipro® Glycosylated hemoglobin (HbA1c) test system is intended for quantitative determination of % hemoglobin A 1c (DCCT/NGSP) in venous whole blood. It is composed of Glycosylated hemoglobin (HbA1c) test kit and automatic immunoassay analyzer. Glycosylated hemoglobin (HbA1c) test kit consists of two reagents R1 and R2, which are liquid and ready to use. Reagent R1 contains glycine buffer and latex, and reagent R2 contains glycine buffer and two types of antibodies, Goat anti-mouse IgG, mouse anti-human HbAlc monoclonal antibody.
HP-AFS/1 Automatic Immunoassay Analyzer is made up of light absorption test module, scattered light test module, fluorescence test module, press components, data transmission interface and printer. And the absorption test module is made up of absorption light path unit, and lightabsorption sensor part. The scattered light test module is made up of scattered light path unit and scattered light sensor part.
Here's a breakdown of the acceptance criteria and the study details for the Hipro® Glycosylated Hemoglobin (HbA1c) Test System:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state pre-defined acceptance criteria values for most tests. Instead, it describes performing studies and concluding whether "All acceptance criteria were met" or providing summary statistics. For the purpose of this table, I will infer relevant criteria based on common industry standards for diagnostic devices and the structure of the reported results. For Method Comparison, Total Error, and Linearity, the reported results are directly in line with typical performance metrics. For Interference and Hemoglobin Variants, "met acceptance criteria" and specific thresholds are mentioned.
| Performance Characteristic | Acceptance Criterion (Inferred/Stated) | Reported Device Performance |
|---|---|---|
| Precision (Total %CV) | Not explicitly stated. Implied to be within acceptable clinical limits. | For HbA1c concentrations ranging from 5.2% to 12.3%, Total %CV ranged from 1.5% to 2.7% across multiple analyzers and lots. For controls (6%, 11%), Total %CV was 3.0% and 1.6% respectively. |
| Method Comparison (Mean Bias) | Not explicitly stated. Implied to show good agreement. | Mean bias vs. NGSP Bio-Rad: -0.0148. 95% CI for mean bias: -0.0470 to 0.0174. |
| Bias at Specific Concentrations | Not explicitly stated. | -0.5% at 5.23% HbA1c, -0.26% at 6.34%, -0.02% at 8.03%, 0.31% at 12.53%. |
| Total Error (%TE) | Not explicitly stated. Implied to be within acceptable clinical limits. | %TE ranged from 3.3% to 5.5% for HbA1c concentrations 5.23% to 12.53%. |
| Linearity (R value) | Not explicitly stated. Implied to be high (close to 1). | R = 0.999 |
| Linearity (Range) | Not explicitly stated. | 4.3% - 14% HbA1c |
| Endogenous Interference (Deviation) | >7% deviation in % HbA1c considered significant interference. | All tested substances (Lipemia, Unconjugated Bilirubin, Conjugated Bilirubin, Glucose, Rheumatoid Factor, Total Protein) showed no significant interference at the highest tested concentrations. |
| Drug Interference (Deviation) | >7% deviation from reference value considered significant interference. | All 16 commonly used drugs tested showed no significant interference. |
| Cross-reactivity | Not explicitly stated. Implied no significant cross-reactivity. | Studies performed for Hb A0, Labile Hb A1c, Carbamylated Hb, Acetylated Hb, Glycated Albumin, Hb A1a+b. Results not detailed, but implied acceptance. |
| Hemoglobin Variants (Interference) | Not explicitly stated. Implied to be minimal or within acceptable clinical limits. Interference with HbF > 8% was observed. | Average bias for HbC: -0.17 at ~6.5% HbA1c, 2.27 at ~9.0% HbA1c. Average bias for HbD: 1.67 at ~6.5% HbA1c, 2.10 at ~9.0% HbA1c. Average bias for HbE: 1.58 at ~6.5% HbA1c, 1.38 at ~9.0% HbA1c. Interference was observed when the concentration of HbF is > 8%. Average bias for HbS: 1.69 at ~6.5% HbA1c, 1.20 at ~9.0% HbA1c. Average bias for HbA2: 1.57 at ~6.5% HbA1c, -0.51 at ~9.0% HbA1c. |
2. Sample Size Used for the Test Set and Data Provenance
- Precision: Four EDTA K2 whole blood samples with targeted HbA1c concentrations (5.2%, 6.4%, 8.0% and 12.3%) and two controls (6% and 11%). Each sample was analyzed in duplicate per run, two runs per day for 20 days on three analyzers (total of 240 measurements per sample/control).
- Method Comparison: One hundred and twenty (120) samples.
- Linearity: One dilution series consisting of 9 levels. Each level measured in triplicate.
- Endogenous Interference: Pooled whole blood samples with two HbA1c levels, spiked with 5 interferents (10 spiked samples), and one unspiked pool. A 10-level dilution series was created for each. Tested ten-fold.
- Drug Interference: Native patient samples at 2 different HbA1c levels, spiked with 16 drugs at two defined concentrations. Measured ten-fold.
- Hemoglobin Variants: 20 samples for each variant (HbC, HbD, HbE, HbF, HbS, HbA2) for a total of 120 samples. Each sample was tested twice.
Data Provenance:
- Method Comparison: The 120 samples were obtained from the NGSP reference laboratory.
- General: The document does not explicitly state the country of origin or whether samples were retrospective or prospective, but the reliance on the NGSP reference laboratory suggests a standardized, potentially retrospective, approach for the method comparison component. The use of "native patient samples" for drug interference suggests clinical sources.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
The establishment of ground truth is primarily through a traceable reference method and a standardization program.
- Method Comparison: The test system was compared against the Bio-Rad Hemoglobin Testing System which was certified by the National Glycohemoglobin Standardization Program (NGSP). The NGSP is a program that standardizes HbA1c results to align with the results of the Diabetes Control and Complications Trial (DCCT). This implies that the 'ground truth' is established by the NGSP's reference methods and processes, rather than individual experts adjudicating each case. The document specifies that the NGSP reference laboratory measured the "X axis" samples.
The document does not specify a number of "experts" in the traditional sense of clinicians or radiologists adjudicating individual cases.
4. Adjudication Method for the Test Set
Not applicable. This is not a study requiring human adjudication of imaging or clinical cases. The "ground truth" is established by a reference laboratory measurement (NGSP Bio-Rad Hemoglobin Testing System) for chemical analysis.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. This device is an in vitro diagnostic (IVD) test system for quantitative determination of HbA1c, not a diagnostic imaging or AI-assisted interpretation device that involves human readers.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, the studies described are standalone performance evaluations of the Hipro® Glycosylated Hemoglobin (HbA1c) Test System (the automatic immunoassay analyzer and test kit) without human intervention in the measurement process. The system quantifies HbA1c levels directly.
7. The Type of Ground Truth Used
- Reference Method: For method comparison, the ground truth was established by the NGSP (National Glycohemoglobin Standardization Program) Bio-Rad Hemoglobin Testing System. The device is certified with NGSP.
- Reference Materials: For precision, linearity, interference, and hemoglobin variant studies, the ground truth for HbA1c concentrations was established using characterized samples (e.g., targeted HbA1c concentrations, spiked samples, varying hemoglobin variant concentrations) which would derive their values from established reference methods or known preparations.
8. The Sample Size for the Training Set
The document does not mention a separate "training set" in the context of an AI/ML algorithm that is trained. This device is a chemical assay system, not an AI/ML-based diagnostic. The described studies are all for performance evaluation (test set) of the final device.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no specific "training set" for an AI/ML algorithm mentioned for this device.
Ask a specific question about this device
(486 days)
Trade/Device Name: Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 Regulation Number: 21 CFR 862.1373
K200904
Trade/Device Name: Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 Regulation Number: 21 CFR 862.1373
Device Trade Name: Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 Regulation Numbers: 21 CFR Part 862.1373
| |
| Regulation # | 21 CFR 862.1373
864.7470 as well as the same analyzer cleared in K131580 as a diagnostic HbA1c device under 21 CFR 862.1373
The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is intended for in vitro diagnostic use for the measurement of % hemoglobin A1c (HbA1c) (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in venous whole blood specimens using ion-exchange high-performance liquid chromatography (HPLC). This test is an aid in diagnosis of diabetes and identifying patients who may be at risk for developing diabetes, and for monitoring of long-term blood glucose control in individuals with diabetes mellitus.
The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is an automated High-Performance Liguid Chromatography (HPLC) system that separates and reports stable hemoglobin A1c (sA1c) percentage in venous whole blood. The operational portion of the G8 is composed of a sampling unit, liquid pump, degasser, column, detector, microprocessors, sample loader, smart media card, operation panel, and a printer. The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 uses ion-exchange HPLC for rapid, accurate, and precise separation of the stable form of HbA1c (sA1c) from other hemoglobin fractions. The G8 uses a non-porous cation exchange column and separates the hemoglobin components in the blood. Separation is achieved by utilizing differences in ionic interactions between the cation and exchange group on the column resin surface and the hemoglobin components in a step gradient elution. The hemoglobin fractions (designated as A1a. A1b. F. LA1c+, SA1c, A0, and, if present, H-V0, H-V2, H-V2 and H-V3) are subsequently removed from the column by performing a step-wise elution gradient using the varied salt concentrations in the Variant Elution Buffers HSi 1, 2 and 3. The peaks, H-V0, H-V1, H-V2 and H-V3 are typically presumptive HbAD, HbAS, HbAC and HbAE respectively. The software compares the retention times of hemoglobin fractions in a sample to the expected "windows of retention" and labels each fraction that correctly elutes within a defined expected window of retention. The software designates a hemoglobin fraction as POX (where X is the order of the peak as it elutes from the column) if it does not match a defined window of retention. All automated processes in the G8 are controlled by internal microprocessors, using software downloaded via a smart media card. The result report is printed and can be stored on the instrument. The data can be transmitted to a host computer through a bi-directional interface. The result report includes the sample ID, date, percentage and retention time of each fraction of hemoglobin, sA1c percentage and total A1 percentage, along with a chromatogram of the elution pattern of the hemoglobin fractions. If a sample contains a hemoglobin variant, the column elutes the fraction depending upon its charge.
The provided text describes the non-clinical performance testing of the Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 (subject device) to support its substantial equivalence to a predicate device. This document focuses on the analytical performance of a diagnostic device rather than an AI/ML powered device, so some of the specific questions regarding AI/ML study design (e.g., number of experts, adjudication methods, MRMC studies) are not applicable.
Here's the information extracted from the document:
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally implied by the statement "All performance testing results met their pre-determined acceptance criteria." While explicit numerical acceptance criteria for each test are not listed in a consolidated table, the discussion throughout the "Summary of Non-Clinical Performance Testing" implicitly defines them through the methodology and results. For example, for precision/repeatability, the claim of "imprecision at ≤ 2%" was a pre-established criterion. Similarly, for hemoglobin variant interference, "Non-clinically significant interference was defined as <= 6% relative difference in the results from a comparative method."
It's important to note that the provided text is a 510(k) summary, which often summarizes detailed study reports. The full reports would contain explicit acceptance criteria thresholds.
Table of Acceptance Criteria (Implied) and Reported Device Performance:
| Acceptance Criteria (Implied/Defined) | Reported Device Performance |
|---|---|
| Precision/Repeatability: Pre-established claim of imprecision at ≤ 2% CV (for Total CV across all factors, or similar) | Total CV for combined analyzers: - 5.46% HbA1c: 0.97% - 6.38% HbA1c: 1.04% - 7.60% HbA1c: 1.28% - 11.91% HbA1c: 0.87% (All reported Total CVs are well below the implied 2% or potentially a specific lower threshold per concentration) |
| Method Comparison (Bias): No explicit acceptance criterion given for bias in the tables, but implied by outcome. | Passing-Bablok Bias: - 5.0% HbA1c: 0.1753 (3.5% bias) - 6.5% HbA1c: 0.2068 (3.2% bias) - 8.0% HbA1c: 0.2383 (3.0% bias) - 12.0% HbA1c: 0.3224 (2.7% bias) Deming Bias: - 5.0% HbA1c: 0.1979 (4.0% bias) - 6.5% HbA1c: 0.2172 (3.3% bias) - 8.0% HbA1c: 0.2366 (3.0% bias) - 12.0% HbA1c: 0.2884 (2.4% bias) (The device is found to be "substantially equivalent" based on these results, implying acceptance) |
| Hemoglobin Variant Interference: Non-clinically significant interference defined as <= 6% relative difference in the results from a comparative method at 6.5% or 8.0% HbA1c. | Reported Relative Bias from Reference Method: HbAD: -0.5% (~6.5% HbA1c), -1.7% (~8.0% HbA1c) HbAS: -2.7% (~6.5% HbA1c), -3.2% (~8.0% HbA1c) HbAC: -1.9% (~6.5% HbA1c), -1.1% (~8.0% HbA1c) HbAE: -1.3% (~6.5% HbA1c), -1.2% (~8.0% HbA1c) HbA2: -4.2% (~6.5% HbA1c), -5.1% (~8.0% HbA1c) HbF: -0.7% (~6.5% HbA1c), -1.6% (~8.0% HbA1c) (All reported relative biases are within the <= 6% redefined significant interference threshold) |
| Endogenous Interfering Substances: Significant interference defined as percent recovery ± 5% of the expected 100% recovery. | No Interference concluded up to specific concentrations: - Albumin: 5000 mg/dL - Ascorbic Acid: 25 mg/dL - Bilirubin C: 21 mg/dL - Bilirubin F: 18 mg/dL - Lipemia: 1000 mg/dL - Rheumatoid Factor: 550 IU/mL (Implies results were within ± 5% recovery for these substances up to the tested concentrations) |
| Cross Reactivity with Hemoglobin Derivatives: Concluded as "not interfering with the assay". | Not interfering concluded for: - Acetylated Hb up to 50 mg/dL - Carbamylated Hb up to 25 mg/dL - Aldehyde Hb up to 25 mg/dL - Labile HbA1c up to 1000 mg/dL |
| Matrix Comparison: No clinical or statistical difference between K2-EDTA and K3-EDTA. | Data supports interchangeable use of K2-EDTA and K3-EDTA blood collection tubes. |
| Linearity and Detection Limit: Reportable range 4.0 to 16.9% HbA1c. | Previously established under 510(k) K071132, and the range is maintained. |
2. Sample Size Used for the Test Set and Data Provenance:
- Precision/Repeatability Study:
- Sample Size: Four concentrations of HbA1c% in K2EDTA venous whole blood. A total of 720 measurements per concentration were performed (three analyzers, over 20 non-consecutive days, with three reagent lots, samples run in duplicate, two times per day).
- Data Provenance: The study was implicitly conducted retrospectively on collected samples. The country of origin is not explicitly stated, but the testing was performed at "Tosoh Bioscience, Inc." laboratories, likely in the US (given the South San Francisco address).
- Method Comparison Study:
- Sample Size: 220 K₂EDTA venous whole blood specimens with HbA1c concentrations spanning the measuring range (4.0-16.9%). Each specimen was tested in duplicate.
- Data Provenance: The study was conducted at "two separate sites," one a "moderate complexity clinical laboratory" (for the candidate device) and the other an "NGSP SRL" (for the comparator device). The precise country of origin is not explicitly stated. The specimens were collected and likely used retrospectively for the comparison.
- Endogenous Interfering Substances / Hemoglobin Variant Interference Studies:
- Sample Size:
- Endogenous interference: Not explicitly stated per substance, but specimens were "spiked with increasing amounts of the interfering substance."
- Hemoglobin Variant Interference:
- HbC: 26 samples
- HbD: 24 samples
- HbE: 26 samples
- HbS: 29 samples
- HbA2: 20 samples
- HbF: 21 samples
- Data Provenance: Not explicitly stated, but implies lab-tested, potentially spiked, and clinical samples from unknown origin (likely retrospective).
- Sample Size:
3. Number of Experts and Qualifications for Ground Truth:
This device is an automated in vitro diagnostic analyzer for quantitative measurement of HbA1c. The "ground truth" for such devices is typically established by:
- Reference methods (e.g., NGSP SRL certified methods).
- Traceability to international standards (DCCT, IFCC).
- Direct analytical measurements and statistical comparisons, rather than expert consensus on image interpretation.
Therefore, the concept of "number of experts used to establish the ground truth" similar to that in an AI imaging study is not applicable here. The ground truth is the chemical measurement itself, validated against recognized reference standards.
4. Adjudication Method for the Test Set:
Not applicable, as this is an analytical device for quantitative measurement, not an interpretative AI/ML system requiring human adjudication of results. The "adjudication" is through statistical agreement with reference methods and assessment against predetermined analytical performance criteria.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:
No. An MRMC study is relevant for AI/ML systems that assist human readers in tasks like image interpretation. This device performs automated quantitative measurements, and its performance is evaluated against reference methods and analytical standards, not human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Yes, the studies described are standalone performance studies of the device (Tosoh Automated Glycohemoglobin Analyzer HLC-723G8) without human intervention in the measurement process itself. The device is an automated HPLC system.
7. The Type of Ground Truth Used:
The ground truth for evaluating the subject device's performance is based on:
- Reference Methods: Specifically, the NGSP SRL (National Glycohemoglobin Standardization Program Secondary Reference Laboratory) using the Trinity Biotech Premier Hb9210™ HbA1c Analyzer (for method comparison) and Primus Model CLC 330 (for some hemoglobin variant interference testing), and the Bio-Rad VARIANT II TURBO HbA1c Kit - 2.0 (for some hemoglobinopathy interference testing).
- Traceability and Standardization: The device's results are traceable to the Diabetes Control and Complications Trial (DCCT) reference method and IFCC, and it is certified by the NGSP.
- Analytical Standards: For precision, linearity, and interference, the "ground truth" reflects the known concentrations of spiked samples or validated concentrations in patient samples, against which the device's measurements are compared.
8. The Sample Size for the Training Set:
The provided document describes non-clinical performance testing to support substantial equivalence after a software modification (v5.24). It does not provide information on the training set used for the development of the device's algorithms or software. This 510(k) summary is focused on verification and validation of the modified device.
9. How the Ground Truth for the Training Set Was Established:
This information is not provided in the 510(k) summary. Given that the device is an automated HPLC system, its "training" per se would involve optimization of its operational parameters, calibration curves, and potentially peak detection/integration algorithms, likely using well-characterized samples with known HbA1c concentrations established by reference methods. However, the details of such a "training set" and its ground truth establishment are not discussed here.
Ask a specific question about this device
(515 days)
| Glycosylated hemoglobin assay |
| Regulation | 21 CFR 862.1373
submission and the risks associated with compliance to the FDA Special Controls set forth in 21 CFR 862.1373
The risks identified to the assay Special Controls set forth in 21 CFR 862.1373(b) and the device modification
The ADVIA® Chemistry Enzymatic Hemoglobin A1c (A1c E) assay is an in vitro diagnostic assay for the quantitative determination of mmol/mol HbA1c (IFCC) and % HbA1c (DCCT/NGSP) in human anticoagulated venous whole blood and hemolysate for use on the ADVIA® Chemistry Systems. Measurement of Hemoglobin A1c is used as an aid in the diagnosis and monitoring of long-term blood glucose control in patients with diabetes mellitus, and as an aid in the identification of patients at risk for developing diabetes mellitus.
The ADVIA® Chemistry Enzymatic Hemoglobin A1c (A1c E) assay measures hemoglobin A1c in human anticoagulated whole blood and hemolysate. The assay consists of three reagents (R1, R2, and Pretreatment Solution), which are liquid and ready to use.
The assay offers both an automated and a manual application. The automated application (A1c_E) lyses the anticoagulated whole blood specimen on the system for the automated application (A1c E). Samples may also be lysed manually using the ADVIA® Chemistry A1c_E Pretreatment Solution to obtain hemolysate for the manual application (A1c_EM). The two applications yield the same results.
The provided document is a 510(k) Summary for a medical device called the ADVIA® Chemistry Enzymatic Hemoglobin A1c (A1c E) Assay. This document describes a submission seeking FDA clearance for a modification to an existing device, specifically to extend the low end of the analytical measuring range of total hemoglobin (tHb).
However, the document does not contain the detailed study results, acceptance criteria tables with performance data, information on sample sizes for test and training sets, expert qualifications, or adjudication methods that would be typically found in a comprehensive study report proving a device meets acceptance criteria.
The 510(k) Summary states that "Performance data were needed to evaluate the change" and "The verification study of linearity was done in accordance with the CLSI standard recognized by the FDA. This study along with other verification activities demonstrate equivalent performance to the predicate and effective risk mitigations. The studies met pre-determined acceptance criteria." and "Testing verified all acceptance criteria were met."
While these statements confirm that studies were conducted and met acceptance criteria, the specific details requested in your prompt (e.g., the actual table of acceptance criteria vs. performance, sample sizes, expert involvement, etc.) are not present in this summary document.
Therefore, I cannot provide the requested information from this document. To answer your prompt, I would need a more detailed study report or clinical trial summary that includes the actual performance data and study design specifics.
Ask a specific question about this device
(146 days)
46250
Re: K193053
Trade/Device Name: Tina-quant Hemoglobin A1cDx Gen.3 Regulation Number: 21 CFR 862.1373
|
| Product Codes,Regulation Numbers | PDJ, 862.1373
The Tina-quant Hemoglobin A1cDx Gen.3 assay is intended for use as an aid in diagnosis of diabetes and as an aid in identifying patients who may be at risk for developing diabetes. It is an in vitro diagnostics reagent system intended for quantitative determination of mmol/mol hemoglobin A1c (IFCC) and % hemoglobin A1c (DCCT/NGSP) in hemolysate or venous whole blood on the cobas c 503 clinical chemistry analyzer. HbA1c determinations are useful for monitoring of long-term blood glucose control in individuals with diabetes mellitus.
Tina-quant Hemoglobin A1cDx Gen.3 assay is an in vitro diagnostics reagent system intended for quantitative determination of mmol/mol hemoglobin A1c (IFCC) and % hemoglobin A1c (DCCT/NGSP) in hemolysate or whole blood on the cobas c 503 clinical chemistry analyzer. The assay offers separate applications that are specific to the sample types whole blood and hemolysate. The Whole Blood Application differs from the Hemolysate Application in the hemolyzing step. For the Whole Blood Application, whole blood samples are placed on the analyzer and hemolysis occurs onboard the analyzer. For the Hemolysate Application, hemolyzed samples are placed on the analyzer and hemolysis occurs manually before placing the samples onboard the analyzer. The two applications yield the same results. Hemolyzing reagent is part of the test system and is either placed on board the analyzer for the Whole Blood Application or used manually for the Hemolysate Application. Anticoagulated whole blood is hemolyzed either manually or automatically prior to determination of HbAlc by a turbidimetric inhibition immunoassay. Liberated hemoglobin (Hb) in the hemolyzed sample is converted to a derivative having a characteristic absorption spectrum and measured bichromatically. The instrument calculates the % HbAlc from the HbAlc/Hb ratio according to a user selected protocol, either IFCC or NGSP protocols.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary for the Tina-quant Hemoglobin A1cDx Gen.3:
1. A table of acceptance criteria and the reported device performance
The document doesn't explicitly state "acceptance criteria" as a single, consolidated table with pass/fail thresholds. Instead, it presents various performance studies and their results. The implicit acceptance criteria are that the device's performance metrics are acceptable for its intended use and comparable to or better than the predicate device.
Here’s a table summarizing the reported device performance for key metrics:
| Performance Metric | Acceptance Criteria (Implicit) | Reported Device Performance (Tina-quant Hemoglobin A1cDx Gen.3) |
|---|---|---|
| Precision | Acceptable levels of repeatability and intermediate precision for clinical use. | Hemolysate Application: - Reproducibility (total CV%): Most samples/controls show CV% between 1.0% and 1.9%. - Repeatability (CV%): Most samples/controls show CV% between 0.4% and 0.6%. Whole Blood Application: - Reproducibility (total CV%): Most samples/controls show CV% between 1.2% and 1.6%. - Repeatability (CV%): Most samples/controls show CV% between 0.4% and 0.8%. |
| Limit of Blank (LoB) | Detection limit sensitive enough for clinical application. | Hb LoB: 0.0530 mmol/L (0.085 g/dL) HbA1c LoB: 0.0220 mmol/L (0.035 g/dL) This corresponds to 15 mmol/mol (IFCC) and 3.5% HbA1c (DCCT/NGSP) at 13.2 g/dL Hb. |
| Limit of Detection (LoD) | Detection limit sensitive enough for clinical application. | Hb LoD: 0.119 mmol/L (0.192 g/dL) HbA1c LoD: 0.0437 mmol/L (0.07 g/dL) This corresponds to 22 mmol/mol (IFCC) and 4.2% HbA1c (DCCT/NGSP) at 13.2 g/dL Hb. |
| Linearity/Reportable Range | Linear response across the claimed measuring range. | Hemoglobin (Hb): 4 – 40 g/dL (2.48 – 24.8 mmol/L) HbA1c: 0.3 – 2.6 g/dL (0.186 – 1.61 mmol/L) This corresponds to a measuring range of 23-196 mmol/mol HbA1c (IFCC) and 4.2-20.1% HbA1c (DCCT/NGSP). Empirical First Order Regression: Pearson's r for Hb = 0.9999, HbA1c = 0.9990. |
| Endogenous Interferences | No significant interference from common endogenous substances at specified concentrations. | Demonstrated claimed maximum concentrations without interference for: Bilirubin (60 mg/dL), Ditaurobilirubin (60 mg/dL), Lipemia (400 mg/dL), Rheumatoid Factors (750 IU/mL), Total Protein (21 g/dL), Albumin (60 g/L), Immunoglobulin (IgG) (60 g/L), Glucose (1000 mg/dL), Triglycerides (1584 mg/dL). A percent deviation criteria was used (not explicitly stated but implied to be within acceptable limits). |
| Cross-Reactivity | No significant cross-reactivity with specified hemoglobin fractions and glycated albumin. | Max Whole Blood Cross-Reactant Concentration with no Interference: HbA0 (120 g/dL), HbA1(a+b) (0.96 g/dL for Level 1, 1.6 g/dL for Level 2), Carbamylated Hb (2.0 g/dL), Acetylated Hb (2.0 g/dL), Glycated Albumin (10 g/dL), Labile HbA1c (1000 mg/dL). Note: Specimens with >7% HbF may yield lower than expected HbA1c values. |
| Hemoglobin Variants | Accurate results for common hemoglobin variants (HbS, HbC, HbE, HbD, HbA2) within acceptable bias. | Relative % Bias from Reference Method at Low (around 6.5%) and High (around 9%) HbA1c: - HbS: -2.5% (Low), -4.0% (High) - HbC: -3.9% (Low), -6.0% (High) - HbE: -0.1% (Low), -1.2% (High) - HbD: -1.8% (Low), -2.6% (High) - HbA2: -1.0% (Low), 0.4% (High) Note: Specimens with >7% HbF may yield lower than expected HbA1c values. |
| Exogenous Interferences (Drugs) | No significant interference from a list of commonly used drugs at specified concentrations. | No significant interference reported for 18 listed drugs (N-Acetylcysteine, Acetylsalicylic acid, Ampicillin-Na, Ascorbic acid, Cefoxitin, Heparin, Levodopa, Methyldopa + 1.5, Metronidazole, Doxycyclin, Rifampicin, Gammagard, Cyclosporine, Phenylbutazone, Acetaminophen, Ibuprofen, Theophylline, Tolbutamide) at tested concentrations. |
| Sample Matrix Comparison | Acceptable agreement across different anticoagulants and fill levels. | Mean Difference for K2-EDTA, K3-EDTA, Na Heparin, Li Heparin, NaF/Potassium oxalate, EDTA/Fluoride at full and half-full tubes indicate close agreement (e.g., -0.026 to 0.019). Upper and Lower 95% CIs are provided (e.g., -0.181 to 0.172). |
| Method Comparison | Good agreement with the NGSP reference method (Tosoh Automated Glycohemoglobin Analyzer HLC-723G8). | Whole Blood Application: Mean bias vs. NGSP TOSOH = -0.046%. Hemolysate Application: Mean bias vs. NGSP TOSOH = 0.046%. Bias at concentrations: (e.g., 5% HbA1c: WB -2.4%, Hemolysate 0.6%; 12% HbA1c: WB 0.7%, Hemolysate 1.2%). |
| Total Error (TE) | Total Error within clinical requirements, considering both bias and precision. | Hemolysate Application: TE ranging from 3.0% to 4.4% across HbA1c levels. Whole Blood Application: TE ranging from 3.1% to 5.2% across HbA1c levels. |
2. Sample size used for the test set and the data provenance
-
Precision (Repeatability and Intermediate Precision):
- Sample Size: 10 samples (2 controls, 8 human samples) for each application (Hemolysate and Whole Blood). Measured for 21 days.
- Data Provenance: Human samples and controls mentioned. No specific country of origin is stated, but implied to be laboratory-based testing. Prospective study design.
-
Analytical Sensitivity (LoB, LoD):
- Sample Size:
- LoB: One analyte-free sample. 60 measurements per lot across 3 lots.
- LoD: Five unique human samples with low-analyte concentrations. 60 measurements per lot across 3 lots.
- Data Provenance: Human samples mentioned. Laboratory-based testing. Prospective study design.
- Sample Size:
-
Linearity/Assay Reportable Range:
- Sample Size: Separate dilution series (at least eleven levels) prepared from human hemolysate sample pools.
- Data Provenance: Human hemolysate samples. Laboratory-based testing. Prospective study design.
-
Endogenous Interferences:
- Sample Size: Pooled whole blood samples at two HbA1c levels, spiked with 9 different interferents. 18 spiked samples + interferent-free pools. Each tested in ten-fold.
- Data Provenance: Pooled whole blood samples. Laboratory-based testing. Prospective study design.
-
Cross-Reactivity:
- Sample Size: Not explicitly stated as a number of unique samples, but rather a "series of experiments" with specific cross-reactants. Ten replicates of each sample were analyzed for each dilution level.
- Data Provenance: Laboratory-based testing using prepared samples to introduce cross-reactants. Prospective study design.
-
Hemoglobin Variants:
- Sample Size: 30 HbS, 30 HbC, 30 HbE, 29 HbD, 15 HbA2, 19 Elevated HbF. Total = 153 samples.
- Data Provenance: Not specified, but implied to be characterized patient samples containing the variants. Laboratory-based testing. Prospective.
-
Exogenous Interferences (Drugs):
- Sample Size: Hemolysate samples at two HbA1c levels, spiked with 18 common drugs at two concentrations. Each drug/concentration combination tested in ten-fold.
- Data Provenance: Hemolysate samples and spiked drugs. Laboratory-based testing. Prospective study design.
-
Sample Matrix Comparison:
- Sample Size: At least 40 samples of each anticoagulant type (e.g., K2-EDTA, K3-EDTA, Na Heparin, Li Heparin, NaF/Potassium oxalate, EDTA/Fluoride) at full and half-filled tubes. Samples from one donor for full/half-filled comparisons.
- Data Provenance: Human donor samples. Laboratory-based testing. Prospective study design.
-
Method Comparison:
- Sample Size: 171 whole blood samples and 173 hemolysate samples.
- Data Provenance: Samples from a "secondary NGSP reference laboratory." This suggests these are clinical samples that have been previously characterized by a gold standard (NGSP certified) method. The nature (retrospective/prospective clinical samples) is not explicitly stated but is consistent with a clinical validation using real patient samples.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document does not describe the use of human experts to establish ground truth in the way one would for image-based diagnostic AI. This device is an in vitro diagnostic (IVD) reagent system for quantifying a biomarker (HbA1c).
- Ground truth for the performance studies (e.g., precision, linearity, interferences) is established through:
- Reference materials (e.g., controls like PreciControl HbA1c norm/path).
- Dilution series from characterized sample pools.
- Spiking experiments with known concentrations of interferents or cross-reactants.
- For Hemoglobin Variants and Method Comparison:
- The "ground truth" or reference method is the NGSP Tosoh HPLC system. This system undergoes certification by the National Glycohemoglobin Standardization Program (NGSP) and is considered a highly accurate laboratory assay method, not dependent on human expert interpretation of a result. No individual human expert counts or qualifications are reported for establishing these reference values.
4. Adjudication method for the test set
Not applicable. This is an IVD device, not an AI or imaging diagnostic device that typically requires expert adjudication for ground truth establishment. The performance is assessed by comparing results to established reference methods, spiked concentrations, or statistical analysis of replicates.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The Tina-quant Hemoglobin A1cDx Gen.3 assay is an in vitro diagnostic (IVD) device for quantitative determination of HbA1c. It does not involve human readers interpreting images or data with or without AI assistance. It provides a numerical result.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the studies described are all standalone (algorithm only) performance evaluations. The device autonomously measures HbA1c and Hb concentrations, calculates the ratio, and reports the final HbA1c result. There is no human intervention in the result generation process once the sample is loaded onto the analyzer.
7. The type of ground truth used
The ground truth used in various parts of the study consists of:
- Reference Methods: The NGSP Tosoh HPLC system for method comparison, which is a recognized standardized method for HbA1c determination.
- Certified Reference Materials/Controls: PreciControl HbA1c norm and path with assigned values.
- Known Concentrations: For linearity, LoB, LoD studies, the "truth" is derived from preparing samples with known concentrations or by statistical analysis of repeat measurements of low/negative samples.
- Spiked Samples: For interference and cross-reactivity studies, the "truth" is the un-spiked sample value, and the effect is measured by deviation after adding a known amount of interferent.
- Characterized Patient Samples: For hemoglobin variant testing, samples from patients known to carry specific hemoglobin variants were used.
8. The sample size for the training set
This document does not describe a "training set" in the context of machine learning or AI. The Tina-quant Hemoglobin A1cDx Gen.3 is a reagent system for a well-established immunoassay technique. Its development involved chemical and assay optimization, not machine learning model training.
9. How the ground truth for the training set was established
Not applicable, as there is no mention or indication of a machine learning "training set" for this IVD device. The assay's analytical principles are based on turbidimetric inhibition immunoassay (TINIA) and bichromatic photometric determination, which are traditional chemical analysis methods, not AI-driven algorithms requiring training data.
Ask a specific question about this device
(114 days)
Clare Ireland V94 PP63
Re: K182651
Trade/Device Name: HbA1c Advanced Regulation Number: 21 CFR 862.1373
5.0 Recommended Reagent Classification: Name: Hemoglobin A1c Test System Requlation: 862.1373
The HbA 1c (Hemoglobin A 1c) Advanced assay on the Beckman Coulter DxC700 AU Clinical Chemistry Analyzer, is intended for the quantitative determination of mmol/mol HbA1c (DCCT/NGSP) concentration in human venous whole blood. The determination of HbA1c is used as an aid in the diagnosis of diabetes mellitus, for the monitoring of long-term glucose control in individuals with diabetes mellitus and identifying patients who may be at risk for developing diabetes mellitus. For in vitro diagnostic use only.
The HbA1c Advanced reagent kit is in a liquid format and is ready to use. It contains four reagents HbA1c R1 and HbA1c R2, Total Hemoglobin R1 and Hemolyzing reagent R1. The HbA1c calibrator is supplied with the reagent, in a liquid, ready to use format and contains 5 x 2mL calibrator levels. The sample hemolysis is automated on the DxC700 AU Clinical Chemistry analyzer. Sample handling is performed as follows: 200 uL of hemolyzing reagent is aspirated from the Hemolyzing Reagent R1and dispensed into a cuvette. Tetradecyltrimethylammonium bromide (TTAB) in the hemolyzing reagent eliminates interference from leukocytes. 2 µL of whole blood sample is then aspirated from the patient sample and added to the hemolyzing reagent in the cuvette. This hemolyzed whole blood is then added to the THb assay cuvette and HbA1c assay cuvette as per the assay parameters. The concentrations of both HbA1c and Total Hemoglobin are determined. The HbA1c/Total Hemoglobin ratio is expressed either as mmol/mol (IFCC) or %HbA1c (DCCT/NGSP). Total Hemoglobin Reagent is used to measure total hemoglobin concentration by a colorimetric method. Change in absorbance is measured at 570/660 nm. HbA1c reagent is used to measure hemoglobin A1c concentration by a turbidimetric immunoinhibition method. In the reaction, hemoglobin A1c antibodies combine with HbA1c from the sample to form soluble antigen-antibody complexes. Polyhaptens from the reagent then bind with the excess antibodies and the resulting agglutinated complex is measured turbidimetrically. Change in absorbance is measured at 340/700 nm.
Here's an analysis of the acceptance criteria and study details for the HbA1c Advanced device, based on the provided document:
Acceptance Criteria and Device Performance
| Parameter | Acceptance Criteria | Reported Device Performance (NGSP Units) | Pass/Fail |
|---|---|---|---|
| Precision | |||
| Repeatability (Within-run) | CV ≤1.5% or SD ≤0.1% HbA1c | Human Whole Blood 1 (5.06%): 0.88% CV, 0.04 SD Human Whole Blood 2 (6.72%): 1.01% CV, 0.07 SD Human Whole Blood 3 (8.06%): 0.77% CV, 0.06 SD Human Whole Blood 4 (11.70%): 0.79% CV, 0.09 SD Spiked Human Whole Blood (14.02%): 0.74% CV, 0.10 SD Whole Blood Control 1 (5.32%): 1.19% CV, 0.06 SD Whole Blood Control 2 (9.88%): 0.77% CV, 0.08 SD | Pass |
| Total Precision | CV ≤2% or SD ≤0.13% HbA1c | Human Whole Blood 1 (5.06%): 1.63% CV, 0.08 SD Human Whole Blood 2 (6.72%): 1.64% CV, 0.11 SD Human Whole Blood 3 (8.06%): 1.57% CV, 0.13 SD Human Whole Blood 4 (11.70%): 1.26% CV, 0.15 SD Spiked Human Whole Blood (14.02%): 1.19% CV, 0.17 SD Whole Blood Control 1 (5.32%): 2.08% CV, 0.11 SD Whole Blood Control 2 (9.88%): 1.54% CV, 0.15 SD | Pass |
| Linearity (NGSP) | |||
| Linear Range | 4-15% HbA1c | 3.94% HbA1c to 15.37% HbA1c | Pass |
| Regression Parameters | Slope: 1.0 ± 0.05; Intercept: ≤ ± 0.5 % HbA1c; R: ≥ 0.990; N: ≥ 9 | Slope: 1.0 ± 0.05; Intercept: ≤ ± 0.5 % HbA1c; R: ≥ 0.990; N: ≥ 9 (All met) | Pass |
| Method Comparison (NGSP) | |||
| Slope (Weighted Deming) | 1.0 ± 0.05 | 0.990 (0.978; 1.002) | Pass |
| Intercept (Weighted Deming) | ≤ ± 0.5% HbA1c | 0.010 %HbA1c (-0.070; 0.089) %HbA1c | Pass |
| R (Weighted Deming) | ≥ 0.975 | 0.998 | Pass |
| Slope (Passing-Bablok) | 1.0 ± 0.5 | 0.980 (0.964; 0.992) | Pass |
| Intercept (Passing-Bablok) | ≤ ± 0.5% HbA1c | 0.090 %HbA1c (-0.006; 0.187) % HbA1c | Pass |
| R (Passing-Bablok) | ≥ 0.975 | 0.998 | Pass |
| Total Error | ≤6% | 5.0% HbA1c: 4.3% 6.5% HbA1c: 4.2% 8.0% HbA1c: 4.3% 12.0% HbA1c: 3.3% | Pass |
| Analytical Specificity | No Significant Interference (recovery within 7% of initial value) | Endogenous Interference: No significant interference up to stated concentrations for Conjugated Bilirubin (60 mg/dL), Unconjugated Bilirubin (60 mg/dL), Lipemia (500 mg/dL), Ascorbic Acid (300 mg/dL), RF (1000 IU/ml), Total Protein (21 g/dL), Glucose (2000 mg/dL). Drug Interference: No significant interference up to stated concentrations for numerous drugs (e.g., Glyburide 0.12 mg/dL, Salicylic Acid 4.76 mg/dL, Acetaminophen 26 mg/dL, etc.). Hemoglobin Derivative and Cross Reactants: No significant interference up to stated concentrations for Labile Hemoglobin (2000 mg/dL), Acetylated Hemoglobin (0.5 mg/mL), Carbamylated Hemoglobin (1.5 mg/mL), Glycated Albumin (5mg/mL), HbA0 (12 mg/mL), HbA1a + 1b (0.16 mg/mL). | Pass |
| Hemoglobin Variants (Bias) | No Significant Interference (recovery within 7% of reference value) | HbC: -2.57% bias (range -4.30% to -1.80%) at ~6.5% HbA1c; -3.19% bias (range -6.48% to 0.41%) at ~9.0% HbA1c. HbD: -0.77% bias (range -4.81% to 2.99%) at ~6.5% HbA1c; -1.22% bias (range -6.30% to -0.22%) at ~9.0% HbA1c. HbE: -1.12% bias (range -9.16% to 2.48%) at ~6.5% HbA1c; 0.47% bias (range -1.76% to 4.21%) at ~9.0% HbA1c. HbS: -1.18% bias (range -2.17% to 3.04%) at ~6.5% HbA1c; -1.04% bias (range -3.33% to 4.41%) at ~9.0% HbA1c. HbA2: 0.48% bias (range -1.92% to 5.60%) at ~6.5% HbA1c; 2.49% bias (range -0.98% to 3.60%) at ~9.0% HbA1c. HbF: Specimens containing >7% HbF may yield lower than expected HbA1c values (disclaimer). | Pass (with disclaimer for HbF) |
Study Details
-
Sample size used for the test set and the data provenance:
- Precision (EP05-A3): Four levels of HbA1c K2 EDTA human venous whole blood patient samples (approx. 5.0%, 6.5%, 8.0%, 12%, and 14%) and two whole blood controls. Samples were analyzed in duplicate, twice daily, over 20 working days on 3 different instrument lots. (n=2 for each sample/control, 2x daily, 20 days: 80 measurements per sample/control per instrument, total for all samples/controls/instruments: 8073 = 1680 individual measurements for precision)
- Linearity (EP06-A): High and low pools of human whole blood were used to create a linearity series to span the analytical range. The exact number of samples in the linearity series is implied to be at least 9 (N: ≥ 9 for regression parameters).
- Method Comparison (EP09-A3): 138 venous human frozen whole blood specimens (K2 EDTA anticoagulant type). These were patient samples. The provenance is not explicitly stated as country of origin, but they are referred to as "venous human frozen whole blood specimens." The study design is prospective in the sense that the samples were collected and then tested for the study; it is not explicitly called out as retrospective/prospective.
- Analytical Specificity (Interference - EP07): Two % HbA1c concentrations (approx. 6.5% and 8.0% HbA1c) using low and high pools prepared from human whole blood. Interfering substances were tested at a minimum of 5 levels each, with 10 replicates per level. The interference assessment was done on human venous whole blood K2 EDTA samples.
- Hemoglobin Variants: A minimum of 20 samples were tested for each variant (HbC, HbD, HbE, HbF, HbS, HbA2), totaling at least 120 samples. These are patient samples with identified hemoglobin variants.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- For the Method Comparison study, the ground truth was established by an "NGSP Secondary Reference Laboratory (SRL) using a test system (method X - HA8180V HPLC)." The specific number and qualifications of experts at the SRL are not detailed in this document.
- For Hemoglobin Variants, the reference methods used to establish the ground truth were "Trinity Biotech Hb9210 and Ultra2, Menarini HA8181V and TOSOH G8," which are demonstrated to be free from hemoglobin interference. Again, the number and qualifications of the operators of these reference methods are not specified.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- The document does not describe an adjudication method involving multiple human readers for establishing ground truth. The ground truth for method comparison and hemoglobin variant studies relies on reference methods/laboratories.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study involving human readers or AI assistance was conducted or described, as this device is an in-vitro diagnostic (IVD) test system (HbA1c assay) and not an imaging AI device that would typically involve human reader interpretation.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, the performance characteristics described are for the "HbA1c Advanced reagent on the DxC 700 AU Clinical Chemistry Analyzer" as a standalone device (algorithm only performance, in a laboratory setting). The entire submission details the analytical performance of the device itself.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Precision: Internal validation against the device's own specifications.
- Linearity: Internal validation for demonstrating accurate measurement across specific concentration ranges.
- Method Comparison: Comparison to an NGSP Secondary Reference Laboratory (SRL) standardized method (HA8180V HPLC).
- Analytical Specificity (Interference): Comparison to baseline measurements of samples without interferents, assessed against a 7% recovery criteria.
- Hemoglobin Variants: Comparison to reference methods demonstrated to be free from hemoglobin interference (Trinity Biotech Hb9210 and Ultra2, Menarini HA8181V and TOSOH G8).
-
The sample size for the training set:
- This document describes the analytical validation of a biochemical assay on an automated analyzer. There is no "training set" in the context of machine learning, as the device is a reagent and instrument system, not an AI/ML algorithm that is trained on data. The device's performance is rigorously tested as described above.
-
How the ground truth for the training set was established:
- Not applicable, as this is not an AI/ML device with a training set. The assay's analytical parameters are established through chemical and immunochemical principles and optimized during development.
Ask a specific question about this device
(31 days)
0884 Norway
Re: K182988
Trade/Device Name: Afinion HbA1c Dx on Afinion 2 Regulation Number: 21 CFR 862.1373
-------------|--------------------|----------------------|
| PDJ | Class II | 21 CFR 862.1373
Afinion™ HbA1c Dx is an in vitro diagnostic test for quantitative determination of glycated hemoglobin (% homoglobin Alc, HbAlc) in human venous and capillary whole blood.
This test is to be used as an aid in the diagnosis of diabetes and as an aid in identifying patients who may be at risk for developing diabetes.
The measurement of % HbA1c is recommended as a marker of long-term metabolic control in persons with diabetes mellitus.
The Afinion™ 2 System consisting of the Afinion™ 2 Analyser and the Afinion™ Test Cartridges is for in vitro diagnostic use only. The Afinion™ 2 Analyzer is a compact multi-of-care testing and is designed to analyze the Afinion™ Test Cartridges.
The Afinion™ HbA1c Dx test system is a CLIA moderate complexity test for diagnosing diabetes and identifying patients who may be at risk for developing diabetes, as a marker of long-term metabolic control in persons with diabetes mellitus.
The Afinion 2 is a multi-assay analyzer for point-of-care testing, designed for use with Afinion assay test cartridges and Afinion controls. It has the same functionality as the Afinion AS100 analyzer performing identical assay processing.
The provided text describes the 510(k) premarket notification for the Afinion™ HbA1c Dx on Afinion™ 2 device. It focuses on demonstrating substantial equivalence to a predicate device (Afinion™ HbA1c Dx on Alere Afinion™ AS100 Analyzer) rather than explicitly outlining a standalone clinical study for new acceptance criteria.
However, the document states: "Verification and validation studies were performed as required by risk analysis and all acceptance criteria were met." It implies that acceptance criteria were established and subsequently met through these studies. The core of the submission is to show that the modified Afinion 2 analyzer performs equivalently to the previous AS100 analyzer when running the same Afinion HbA1c Dx test.
Based on the provided text, here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly list specific numerical acceptance criteria (e.g., specific accuracy, precision targets) and their corresponding reported device performance values in a table format. Instead, it broadly states that "all acceptance criteria were met" and that the analytical performance was not adversely affected.
However, by comparing the candidate device (Afinion™ HbA1c Dx with Afinion™ 2 analyzer) to the predicate device, we can infer that the acceptance criterion for the new analyzer is to maintain the same performance characteristics as the predicate. The "reported device performance" is the demonstration that its performance is equivalent.
| Performance Characteristic | Acceptance Criteria (Inferred from Predicate) | Reported Device Performance |
|---|---|---|
| Analytical Performance | Not adversely affected compared to predicate | All acceptance criteria met; analytical performance not adversely affected |
| Assay Sequence Timing | Maintained as in the predicate (AS100) | Software modifications ensured timing is maintained |
| Functionality | Same as the predicate (AS100) | Same functionality as predicate |
| User Interface | Same as the predicate (AS100) | Same user interface as predicate |
| Error Rates | Equivalent to predicate | Risk analysis showed no adverse effect on risk of erroneous results |
2. Sample Size Used for the Test Set and Data Provenance:
The document mentions "in-house analytical performance verification studies" but does not specify the sample size for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It refers to these as part of "design control activities" to address risk analysis.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
This information is not provided in the document. The nature of the device (an HbA1c test system) suggests that "ground truth" would likely be based on established reference methods or certified values, rather than expert consensus on images or interpretations.
4. Adjudication Method for the Test Set:
This information is not provided in the document. Given the type of diagnostic test (quantitative determination of glycated hemoglobin), clinical adjudication by experts is unlikely to be the primary method for determining ground truth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
A MRMC study is typically for evaluating human reader performance, often in image-based diagnostics. This device is an automated in vitro diagnostic test system. Therefore, an MRMC comparative effectiveness study involving human readers is not applicable and was not performed.
6. Standalone (Algorithm Only) Performance:
The entire submission focuses on the performance of the "Afinion™ HbA1c Dx on Afinion™ 2" as a complete system, which is an automated diagnostic test. Therefore, the "standalone" performance is the algorithm's performance integrated into the device, using the specified test cartridges. There is no human-in-the-loop component for the measurement itself. The study details, though not fully disclosed, would pertain to the device operating in this standalone manner.
7. Type of Ground Truth Used:
The document states that the Afinion HbA1c Dx is "traceable to the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Reference Method for Measurement of HbA1c" and "certified by NGSP (National Glycohemoglobin Standardization Program)." This indicates that the ground truth for performance evaluation (e.g., accuracy, bias) is established by reference methods and standardization programs rather than expert consensus, pathology, or direct outcomes data.
8. Sample Size for the Training Set:
The document does not provide details on a "training set" or its sample size. This type of device is likely developed and validated using a more traditional analytical validation approach based on measurement science, rather than a machine learning paradigm that typically involves distinct training and test sets. It implies a process of design, calibration, and verification/validation.
9. How Ground Truth for the Training Set Was Established:
As there is no mention of a traditional "training set" in the context of machine learning, this information is not applicable and a method for establishing its ground truth is not provided. The development and calibration of the test would rely on rigorous analytical methods and established reference materials, as indicated by its traceability to IFCC and NGSP certification.
Ask a specific question about this device
(203 days)
DE 19702
Re: K173909
Trade/Device Name: Dimension Hemoglobin A1c Assay Regulation Number: 21 CFR 862.1373
System |
| Classification Name | Hemoglobin A1c Test System |
| Regulation | 21 CFR 862.1373
The Dimension® Hemoglobin A1c assay is an in vitro diagnostic assay for the quantitative determination of %HbA 1c (DCCT/NGSP) and mmol/mol HbA1c (IFCC) in human anticoagulated venous whole blood for use on the Dimension® clinical chemistry system. Measurement of Hemoglobin A1c is used as an aid in diagnosis and monitoring of long-term blood glucose control in patients with diabetes mellitus and as an aid in the identification of patients at risk for developing diabetes mellitus.
The Dimension® Hemoglobin A1C assay is an in vitro diagnostic device intended to measure the concentration of hemoglobin A1c in venous human anticoagulated whole blood. The assay consists of three reagents packaged in Dimension® Flex® cartridges. The reagents are liquid and ready to use.
The provided text describes the Siemens Healthcare Diagnostics Inc.'s Dimension® Hemoglobin A1c Assay, a Class II medical device for quantitative determination of %HbA1c and mmol/mol HbA1c in human anticoagulated venous whole blood. The document details studies undertaken to demonstrate the device's performance, primarily for a 510(k) premarket notification to show substantial equivalence to a predicate device.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" with numerical targets for each performance characteristic. Instead, it presents study results and, for some criteria, qualitative conclusions (e.g., "No significant interference," "No deviations from linearity"). The assessment of "substantial equivalence" to the predicate device implies that the reported performance meets acceptable clinical standards for HbA1c measurement.
| Performance Metric | Acceptance Criteria (Implied / Qualitative) | Reported Device Performance |
|---|---|---|
| Method Comparison (against NGSP Reference Method) | Agreement with NGSP reference method; low bias. | Passing-Bablok: Slope [0.966 to 1.001], Y-int [-0.095 to 0.144] (%HbA1c); Bias at decision levels: -1.10% to -1.45% (%HbA1c) |
| Deming: Slope [0.957 to 1.000], Y-int [-0.094 to 0.198] (%HbA1c); Bias at decision levels: -1.16% to -1.77% (%HbA1c) | ||
| Precision | Low variability (Repeatability, Between-Run, Between-Day, Between-Instrument, Between-Lot, Total CV). | Total CV: 1.7% to 2.6% for %HbA1c (QC samples), 1.6% to 2.4% for %HbA1c (patient pools) |
| Total Error | Acceptable total error at decision levels. | Ranged from 4.4% to 5.8% (%TE) for %HbA1c at various decision levels. |
| Endogenous & Exogenous Interference | No significant interference (e.g., < ± 5.0% bias) from common interferents. | No significant interference observed for interferents listed in Table J. |
| Hemoglobin Variant Interference | No significant interference from common Hb variants (e.g., < ± 5.0% bias). | No significant interference observed for HbC, HbD, HbE, HbA2. Significant interference observed for HbF (-23.2% to -24.7% bias). |
| Hemoglobin Derivatives | Inaccuracies/biases ≤ 5% due to acetylated, carbamylated, or labile Hb. | Inaccuracies (biases) due to these substances are less than or equal to 5%. |
| Linearity | Linearity across the assay range. | No deviations from linearity observed for results from 3.6 to 15.9% HbA1c. |
| Limit of Blank (LoB) | Defined LoB value. | LoB: 3.6% for %HbA1c, 0.0 g/dL for tHb, 0.21 g/dL for HbA1c. |
| Limit of Detection (LoD) | Defined LoD value. | LoD: 3.7% for %HbA1c, 0.5 g/dL for tHb, 0.23 g/dL for HbA1c. |
| Anticoagulant Comparison | Equivalence between different anticoagulants. | Slopes [0.982 to 1.023] and y-intercepts close to zero in comparison to K2-EDTA. |
2. Sample size used for the test set and the data provenance
- Method Comparison: 147 human whole blood samples. The document does not specify the country of origin or if the data was retrospective or prospective.
- Precision: Two (2) commercial quality controls and four (4) whole blood patient pools (with specific target values). The number of individual patient samples in these pools is not specified, but the testing was performed over 20 days by 2 operators on 3 instruments.
- Endogenous & Exogenous Interference: Not explicitly stated as a "test set" of patient samples, but each interferent was evaluated using samples at two HbA1c levels (6.5% ± 1.0% and 8.0% ± 1.0%).
- Hemoglobin Variant Interference:
- HbC: 37 samples
- HbD: 20 samples
- HbE: 22 samples
- HbS: 22 samples
- HbA2: 23 samples
- HbF: 20 samples
- These were "anticoagulated human blood samples with known concentrations of hemoglobin variant and HbA1c." The document does not specify the country of origin or if the data was retrospective or prospective.
- Linearity: Nine (9) samples with HbA1c levels across the assay range.
- Limit of Blank (LoB) and Limit of Detection (LoD):
- LoB: 60 determinations with 5 blank samples each for tHb, and 75 determinations with 5 blank samples each for % HbA1c/HbA1c.
- LoD: 60 determinations with 5 low-level samples each for tHb, and 75 determinations with 5 low samples each for % HbA1c/HbA1c.
- Anticoagulant Comparison: 79 samples. The document does not specify the country of origin or if the data was retrospective or prospective.
Data Provenance: For all studies, the document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. However, based on the nature of lab-based validation studies for a diagnostic assay, they are typically prospective, controlled experimental setups.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
For the method comparison study, the ground truth was established by "NGSP reference method results from testing performed at an NGSP reference laboratory." The document does not specify the number or qualifications of experts at this reference laboratory, but an NGSP (National Glycohemoglobin Standardization Program) reference laboratory adheres to strict standardization protocols.
For other studies (e.g., interference, linearity), the ground truth was based on preparing samples with known concentrations or characteristics, rather than expert consensus on diagnostic images or conditions.
4. Adjudication method for the test set
Not applicable in the context of this device. The studies directly measure quantitative values and compare them against reference methods or known concentrations, rather than relying on multiple expert interpretations that would require an adjudication process.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an in vitro diagnostic assay for quantitative laboratory measurement, not an AI-assisted diagnostic imaging or interpretation device. There are no "human readers" in the traditional sense involved in interpreting results that would be improved with AI assistance.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, the studies described are for the standalone performance of the Dimension® Hemoglobin A1c Assay. The device is a clinical chemistry system designed for automated quantitative determination, implying an "algorithm only" or automated performance. The studies evaluate the accuracy, precision, and interference characteristics of this automated system.
7. The type of ground truth used
- Method Comparison: NGSP reference method results. This is a standardized, high-accuracy laboratory method.
- Interference (Endogenous, Exogenous, Hemoglobin Variant, Derivatives): Samples with known concentrations of interferents or variants, and known HbA1c values.
- Precision: Commercial quality controls and meticulously prepared patient pools with target values.
- Linearity: Prepared samples with a range of known HbA1c levels.
- LoB/LoD: Blank samples and low-level samples with known very low concentrations.
- Anticoagulant Comparison: K2-EDTA samples served as the comparator (reference) for other anticoagulants.
In essence, the ground truth is established through highly controlled laboratory preparations and comparisons to recognized reference methods or established comparator methods, rather than expert consensus or pathology in the context of imaging.
8. The sample size for the training set
Not applicable. This is an in vitro diagnostic assay, not a machine learning or AI-based device that typically requires a separate training set. The performance evaluation is based on standard analytical validation studies.
9. How the ground truth for the training set was established
Not applicable, as there is no training set in the context of this traditional in vitro diagnostic assay.
Ask a specific question about this device