    510k Summary Text (Full-text Search) :

    [Follicle Stimulating Hormone (FSH) Test, Over the Counter]

    Regulatory Classification:

    21 CFR 862.1300
    CA 92128

    K070484 Re:

    Trade/Device Name: OuantRxTM Female Fertility Test Regulation Number: 21 CFR§862.1300

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QuantRx Female Fertility Test is intended to measure follicle-stimulating hormone (FSH) in urine as an ancillary screen of fertility for home use by women who are attempting to conceive.

    Device Description

    The QuantRx Female Fertility Test is a rapid, immunochromatographic assay for the measurement of follicle-stimulating hormone (FSH) in human urine. The test is performed on day 3 of the menstrual cycle by urinating directly on the absorbent tip or by dipping it in a cup of urine, and observing for the formation of colored lines after 10 minutes. A test line of color intensity greater than or equal to the reference line indicates an FSH level of at least 10 mIU/mL is indicative of diminished ovarian reserve, which corresponds to reduced fertility. A test line intensity of less than the reference line is indicative of normal ovarian reserve.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the QuantRx Female Fertility Test, based on the provided 510(k) summary:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Primary Study: - Concordance (trained user) between proposed device vs. laboratory method89%
    - Concordance (layperson) between proposed device vs. laboratory method88%
    - Concordance trained user vs. layperson for primary study98%
    Supplemental Study: - Concordance (trained user) between proposed device vs. predicate94%
    - Concordance (layperson) between proposed device vs. predicate92%
    - Concordance trained user vs. layperson for supplemental study98%
    The device is intended to measure FSH in urine, with a test line intensity equal to or darker than the reference line (calibrated at 10 mIU/mL) indicating FSH ≥ 10 mIU/mL, and a lighter intensity indicating FSH < 10 mIU/mL.The study results indicate concordance with laboratory and predicate methods, supporting the device's ability to discriminate FSH levels at the 10 mIU/mL threshold for both trained and lay users.
    The Control line must always appear, indicating the test performed properly; its absence results in an "invalid" test.Not explicitly quantified in the provided summary, but implied by the device description as a functional requirement for valid results.

    Study Details

    1. Sample size used for the test set and the data provenance:

      • Primary Study: 209 females (for consumer study, and the same combination of fresh/blinded control specimens for professional study). The origin of this data (e.g., country) is not specified. It is likely prospective for the consumer and professional studies, as it describes a comparison to a validated lab method using freshly collected and blinded control specimens.
      • Supplemental Consumer Study: 48 females. 40 of these provided paired serum specimens. Data provenance (e.g., country) is not specified. This also appears to be prospective as it involves women performing the test and providing specimens.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document refers to a "published, validated, quantitative, chemiluminescent method for urine follicle-stimulating hormone (FSH)" and an "FDA cleared serum FSH assay (DPC IRMA FSH; K893976)" as the ground truth.
      • The number of experts involved in the initial validation of these laboratory methods is not specified, nor are their individual qualifications (e.g., specific years of experience or board certifications). The ground truth is established by the results of these validated and cleared laboratory methods, rather than by human expert consensus on interpretations of the device's results.
    3. Adjudication method for the test set:

      • No adjudication method (e.g., 2+1, 3+1) is described for the test set. The comparison is made directly between the QuantRx device's results and the results from the established laboratory methods.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study involving AI or human reader improvement with AI assistance was conducted or mentioned. This device is a rapid diagnostic test, not an AI-powered image analysis tool. The study did compare "trained user" vs. "layperson" performance, which is a form of multi-reader evaluation, but not focused on AI assistance. The concordance between trained user and layperson for the primary and supplemental studies was 98%, suggesting very high agreement between these two user groups.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • This is not applicable to the QuantRx Female Fertility Test, as it is a visual, rapid immunochromatographic assay that requires human interpretation of color lines. It is not an algorithm-only device. The performance data explicitly includes "trained user" and "layperson" results, indicating human-in-the-loop performance.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth was established using results from a published, validated, quantitative, chemiluminescent method for urine follicle-stimulating hormone (FSH) and an FDA cleared serum FSH assay (DPC IRMA FSH; K893976). This is a laboratory-based, quantitative measurement of the analyte of interest.
    7. The sample size for the training set:

      • The document describes performance studies (primary and supplemental) which serve as validation. It does not explicitly mention a separate "training set" in the context of machine learning, as this is not an AI/ML device. The studies described are evaluation studies comparing the device to reference methods.
    8. How the ground truth for the training set was established:

      • See point 7. No specific "training set" or method for establishing its ground truth is described, as the device is a chemical immunoassay, not an AI/ML system that requires a distinct training phase.
    Ask a Question

    Ask a specific question about this device

    K Number
    K052662

    Validate with FDA (Live)

    Date Cleared
    2005-11-30

    (64 days)

    Product Code
    Regulation Number
    862.1300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    V3M 6S7 Canada

    Re: K052662 Trade/Device Name: One Step FSH Menopausal Test Regulation Number: 21 CFR 862.1300

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FSH Menopausal Test is a rapid Step The One chromatographic immunoassay for the qualitative detection of follicle stimulating hormone (FSH) in urine to aid in the detection of the onset of menopause. The test utilizes a combination of antibodies including mouse monoclonal anti-FSH antibodies and goat polyclonal anti-mouse antibodies to selectively detect elevated levels of FSH. This device is intended for both professional and lay person use.

    Device Description

    One Step FSH Menopausal Test is a rapid chromatographic immunoassay for the qualitative detection of follicle stimulating hormone (FSH) in urine. The test utilizes a combination of antibodies including mouse monoclonal anti-FSH antibodies and goat polyclonal anti-mouse antibodies to selectively detect elevated levels of FSH.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided FDA 510(k) summary for the "One Step FSH Menopausal Test":

    A. Acceptance Criteria and Reported Device Performance

    The acceptance criteria for diagnostic tests often involve measures like sensitivity, specificity, and accuracy, especially when comparing to a reference method or clinical endpoint. While the specific numerical acceptance criteria established by the manufacturer are not explicitly detailed in this FDA letter (which typically focuses on substantial equivalence), the provided data implies a performance standard for "positive agreement" and "negative agreement" in comparison to a predicate device or clinical reference.

    Acceptance Criteria (Implied)Reported Device Performance
    Professional Use
    Positive Agreement90.3%
    Negative Agreement92.2%
    Overall Agreement91.5%
    Lay Use
    Positive Agreement90.3%
    Negative Agreement92.2%
    Overall Agreement91.5%
    Accuracy (Against Clinical Reference)
    Negative Reference92.2%
    Positive Reference90.3%

    Note: The FDA 510(k) summary usually provides the specific acceptance criteria. This document indicates the device is "substantially equivalent," meaning its performance is comparable to a legally marketed predicate device, implicitly meeting accepted performance standards for its intended use.

    B. Sample Size Used for the Test Set and Data Provenance

    The provided document does not explicitly state the specific sample size used for the test set in a clinical study to validate these performance metrics. The agreement percentages (Positive Agreement, Negative Agreement, Overall Agreement) are listed, but the accompanying sample sizes for "Professional Use" and "Lay Use" are not present in this excerpt.

    • Data Provenance: Not specified in this document. It does not mention the country of origin of the data or whether the study was retrospective or prospective.

    C. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This information is not available in the provided document. For an FSH Menopausal Test, the "ground truth" for elevated FSH levels would likely be established by a laboratory assay (e.g., ELISA) for quantitative FSH levels, interpreted by a qualified medical professional (e.g., endocrinologist, gynecologist). The document doesn't detail the involvement of experts in establishing the ground truth or their qualifications.

    D. Adjudication Method for the Test Set

    This information is not available in the provided document. Adjudication methods (e.g., 2+1, 3+1) are typically used in imaging or subjective interpretation studies. For a diagnostic test like this, the ground truth is usually established by a reference method (e.g., quantitative laboratory FSH measurement), making multi-reader adjudication less relevant in the same way.

    E. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study, which typically evaluates how AI assistance impacts human reader performance, is not applicable to a single-use, qualitative in-vitro diagnostic test like the One Step FSH Menopausal Test. This device is not an AI-assisted interpretation tool.

    F. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes, in essence. This device inherently operates as a "standalone" test. The "performance" figures provided (Positive Agreement, Negative Agreement, Overall Agreement) reflect the device's ability to accurately detect FSH in urine samples, without human intervention in the detection process itself. The "Lay Use" section specifically evaluates its performance when interpreted by non-professionals, which is a key aspect of its standalone functionality for over-the-counter use.

    G. The Type of Ground Truth Used

    The ground truth used for performance evaluation is indicated by "Clinical Reference" and "Negative Reference" / "Positive Reference." This strongly suggests that the device's results were compared against an established clinical reference method for FSH levels (likely a quantitative laboratory assay) to determine true positive and true negative cases.

    H. The Sample Size for the Training Set

    This information is not available in the provided document. For a traditional in-vitro diagnostic test, there isn't a "training set" in the machine learning sense. The device's components (antibodies, reagents) are developed and optimized through research and development, but the sample sizes involved in that development process are distinct from the clinical validation (test set) discussed above.

    I. How the Ground Truth for the Training Set Was Established

    Not applicable in the machine learning sense. The "ground truth" for the development of such a test would involve confirming the specificity and sensitivity of the antibodies and reagents against known concentrations of FSH, using established laboratory standards and methods. This is part of the product development and quality control process, rather than a "training set" in AI/ML terminology.

    Ask a Question

    Ask a specific question about this device

    K Number
    K043599

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-02-25

    (58 days)

    Product Code
    Regulation Number
    862.1300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    92121

    K043599 Re:

    Trade/Device Name: Moments Menopause Check (Model 9113) Regulation Number: 21 CFR 862.1300

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Moments Menopause Check is an in-vitro diagnostic screen for the detection of FSH (follicle stimulating hormone) in urine. Change in FSH levels may be associated with changes in menopause. This kit provides a preliminary screening result. This kit is intended for over-the-counter sales.

    Device Description

    Immunoassay for the qualitative detection of FSH in urine.

    AI/ML Overview

    Here's an analysis of the provided information regarding the Moments Menopause Check (Model 9113), focusing on its acceptance criteria and the supporting study:

    Acceptance Criteria and Device Performance for Moments Menopause Check (Model 9113)

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Qualitative detection of FSH in urine.The device performs qualitative detection of FSH in urine.
    Minimum detectable concentration for FSH is 25 mIU/ml.The minimum detectable concentration for this test is 25 mIU/ml.
    Substantially equivalent to predicate devices for qualitative detection of FSH."Laboratory studies, using clinical specimens, produced a >98% correlation when compared to the predicate devices."
    Excellent overall accuracy for consumer interpretation."Consumer interpretation of the FSH test showed accuracy to be greater than 98%."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: The document does not explicitly state the sample size used for the clinical sample correlation study or the blind labeled spiked study.
    • Data Provenance: The studies used "clinical specimens." The country of origin for the data is not specified, but the manufacturer is Phamatech, located in San Diego, California, USA. The studies appear to be prospective, as they were conducted to evaluate the performance of the Moments Menopause Check.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The document does not provide information on the number of experts used or their qualifications for establishing ground truth in the performance studies.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method used for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study being performed to assess how much human readers improve with AI vs. without AI assistance. This device is a rapid, qualitative immunoassay for over-the-counter use, and such a study design is generally not applicable to this type of diagnostic.

    6. Standalone Performance Study

    A standalone performance study was performed. The "clinical sample correlation study" and "blind labeled spiked study" evaluated the device's performance (qualitative detection of FSH in urine) independently. The "consumer study" also assessed the device in a standalone context, focusing on user interpretation.

    7. Type of Ground Truth Used

    • Clinical Sample Correlation Study: The ground truth for this study was likely established by the "predicate devices" to which the Moments Menopause Check was compared. This implies a comparison to existing, legally marketed FSH tests which themselves have established accuracy.
    • Blind Labeled Spiked Study: For this study, the ground truth would have been based on known, controlled concentrations of FSH that were "spiked" into samples, allowing for a direct assessment of the device's ability to detect FSH at the specified threshold (25 mIU/ml).
    • Consumer Study: The ground truth for the consumer interpretation was likely derived from the objective results of the FSH test (e.g., whether FSH was actually present at the detectable level) and then compared to the consumer's visual interpretation.

    8. Sample Size for the Training Set

    The document does not provide information about a training set since this is a chemical test (immunoassay) and not a machine learning or AI-based device. Therefore, the concept of a "training set" as understood in AI/ML is not applicable here. The device's performance is based on its inherent chemical properties and design.

    9. How the Ground Truth for the Training Set Was Established

    As mentioned above, the concept of a training set and its ground truth is not applicable to this device, which relies on immunoassay technology rather than machine learning.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041165

    Validate with FDA (Live)

    Date Cleared
    2004-06-17

    (44 days)

    Product Code
    Regulation Number
    862.1300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Beverly, MA 01915

    K041165 Re:

    Trade/Device Name: FSH Menopause Predictor Test Regulation Number: 21 CFR 862.1300

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FSH Menopause Predictor Test is a qualitative, one-step, midstream assay for the detection of Follicle Stimulating Hormone (FSH) in urine to be used as an aid in predicting menopause. The FSH Menopause Predictor Test is intended for over-the-counter use by the lay consumer.

    Device Description

    The FSH Menopause Predictor Test is a midstream test used for the qualitative measurement of FSH and the detection of Follicle Stimulating Hormone in a woman's urine as an aid in predicting menopause. The FSH predictor test is intended for use outside the body (in vitro diagnostic use) by women at home. FSH Menopause Predictor is an over-the-counter (OTC) device and may be sold under various private labels.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the ACON Laboratories Inc. FSH Menopause Predictor Test, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device PerformanceComments
    Accuracy (ACON FSH Menopause Predictor Test (trained Lab Tech) vs. InstaCheck®)>99% (99% - 99% CI)This appears to be the primary criterion for substantial equivalence to the predicate device when used by trained personnel. Due to rounding, the confidence interval is very narrow.
    Accuracy (ACON FSH Menopause Predictor Test (consumer) vs. ACON FSH Menopause Predictor Test (trained Lab Tech))94% (92.1% - 95.6% CI)This criterion assesses the performance of the device when used by the intended lay consumer compared to a trained professional using the same device. This demonstrates user-friendliness and reliability in the intended OTC environment.
    Accuracy (ACON FSH Menopause Predictor Test (Consumer) vs. InstaCheck®)94% (92.1% - 95.6% CI)This criterion directly compares the consumer-used device to the predicate device, reflecting overall consumer performance against the established standard.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 70 females.
    • Data Provenance: The study was a prospective clinical trial as participants conducted tests at home according to instructions, and urine samples were collected for testing by a study coordinator. The location of the study (e.g., country of origin) is not explicitly stated in the provided text.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • The ground truth in this study was established using a substantially equivalent product (InstaCheck®) and the ACON FSH Menopause Predictor Test used by a trained lab tech.
    • The text refers to a "trained Lab Tech," implying a professional with specific training in operating such devices. No further details on the qualifications (e.g., years of experience, specific certifications) are provided for this "trained Lab Tech."

    4. Adjudication Method for the Test Set

    • The text describes a comparison study rather than an adjudication process for a single ground truth. The "adjudication" was effectively done by comparing the results of the ACON FSH Menopause Predictor Test (both consumer-used and lab tech-used) against the InstaCheck® device. The data obtained from both devices were recorded as "Negative or Positive" without mention of an adjudication panel resolving discrepancies between the devices.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No, a traditional MRMC comparative effectiveness study was not done in the context of comparing human readers with and without AI assistance. This device is a diagnostic test kit, not an AI-powered diagnostic imaging tool that would typically involve radiologists or other human readers interpreting complex images.
    • However, the study does compare performance between the "trained Lab Tech" and the "consumer" using the ACON device, which could be seen as a form of multi-user comparison, albeit without an AI component. There is no "effect size of how much human readers improve with AI vs without AI assistance" as AI assistance is not part of this device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Yes, a standalone performance was done in the sense that the ACON FSH Menopause Predictor Test (when operated by a trained lab tech) was evaluated for its accuracy against the InstaCheck® product. This represents the device's inherent performance under controlled conditions. The "consumer" arm of the study also represents a standalone performance of the device in the hands of its intended user.

    7. The Type of Ground Truth Used

    • The ground truth used was comparative performance against a legally marketed predicate device (InstaCheck®) and the device itself when operated by a trained professional. While not explicitly "pathology" or "outcomes data" in the traditional sense, the InstaCheck® device serves as an established clinical reference for FSH levels in this context. The study also uses the ACON device operated by a trained professional as a reference for assessing consumer performance.

    8. The Sample Size for the Training Set

    • The provided text does not specify a training set sample size. This is likely because the FSH Menopause Predictor Test is a chemical-based in vitro diagnostic, not a machine learning or AI-driven device that requires a separate training set for algorithm development. The performance data presented are from a clinical trial (test set) designed to validate the device's performance.

    9. How the Ground Truth for the Training Set Was Established

    • As there's no mention of a training set in the context of AI or machine learning for this device, a separate ground truth establishment method for a training set is not applicable based on the provided information. The device's mechanism is based on antibody-antigen reactions, not on learned patterns from a large dataset.
    Ask a Question

    Ask a specific question about this device

    K Number
    K040575

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-05-20

    (77 days)

    Product Code
    Regulation Number
    862.1300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    K040575

    Trade/Device Name: Moments Menopause Check (Models 9111 and 9112) Regulation Number: 21 CFR 862.1300

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Moments Menopause Check is an in-vitro diagnostic screen for the detection of FSH (follicle stimulating hormone) in urine. Change in FSH levels may be associated with stages in menopause. This kit provides a preliminary result for the detection/presence of FSH in urine. It is intended for over-the-counter sales.

    Device Description

    Immunoassay for the qualitative detection FSH (follicle stimulating hormone) in urine. The Moments Menopause Check, like many commercially available FSH screening test kits, qualitatively measures the presence of FSH by visual color sandwich one step immunoassay technology.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study details for the Moments Menopause Check device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device PerformanceStudy Type
    Correlation to Predicate Devices98% correlationClinical sample correlation study
    Overall Accuracy (Lay Users)97% accuracyConsumer study

    2. Sample Sizes Used for the Test Set and Data Provenance

    The exact sample sizes for the clinical sample correlation study and the consumer study are not explicitly stated in the provided text.

    • Data Provenance: The text indicates "Laboratory studies, using clinical specimens" and "A consumer study was also performed." This implies that the data was collected from human participants, likely in a prospective manner for the consumer study, and potentially retrospective or prospective for the clinical sample correlation study, though this is not specified. The country of origin is not mentioned, but the manufacturer is based in San Diego, California, USA, suggesting the studies likely originated there.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Clinical Sample Correlation Study: The "predicate devices" were used as the comparison for establishing ground truth. The expertise involved in generating results from the predicate devices is not specified, but it's implied that these are established, commercially available tests.
    • Consumer Study: The ground truth for the consumer study's accuracy (97%) is not explicitly described in terms of "experts." It refers to the device's accuracy in the hands of lay users, suggesting the "truth" against which their results were compared was likely derived from the established laboratory methods (predicate devices) on the same samples, although this is inferred.

    4. Adjudication Method for the Test Set

    The adjudication method is not explicitly stated in the provided text for either study.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned. The studies focus on comparing the device to predicate devices and assessing accuracy with lay users, not on how human readers' performance improves with or without AI assistance. The device itself is a qualitative immunoassay, not an AI-assisted diagnostic.

    6. Standalone (Algorithm Only) Performance Study

    • Yes, a standalone performance was done. The "clinical sample correlation study" and the "consumer study" directly assess the performance of the Moments Menopause Check device itself (the "algorithm" in this context refers to the device's inherent detection mechanism) without explicit human-in-the-loop assistance influencing the detection mechanism itself, beyond the user performing the test.

    7. Type of Ground Truth Used

    • Clinical Sample Correlation Study: The ground truth was established by comparison to predicate devices. This implies the results from established, commercially available FSH screening test kits were considered the "truth."
    • Consumer Study: The ground truth for the accuracy assessment against the lay user's results was likely established by laboratory methods (predicate devices) on the same samples, though this is not explicitly stated.

    8. Sample Size for the Training Set

    The provided text does not mention a training set or its sample size. This type of immunoassay device typically does not involve a "training set" in the machine learning sense. Performance is established through validation studies.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned for this type of device, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032002

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-10-30

    (125 days)

    Product Code
    Regulation Number
    862.1300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    510k Summary Text (Full-text Search) :

    . § 862.1300 |

    Legally Marketed Predicate Devices:

    Instacheck Menopause Predictor Test (K023408)
    D.C. 20004

    K032002 Re:

    Trade/Device Name: Fertell Female Fertility Test Regulation Number: 21 CFR 862.1300

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fertell Female Fertility Test is intended to measure Follicle Stimulating Hormone ("FSH") in urine as an adjunctive screen of fertility for home use by women who are attempting to conceive but have been unsuccessful.

    Device Description

    The Fertell Female Fertility Test is a dipstick, lateral flow rapid test that measures FSH in an instream urine sample. The test is performed two days after the onset of menses (day 3 of the menstrual cycle) and is similar in design and use to a pregnancy test or an ovulation prediction test. It uses a monoclonal antibody to one antigenic determinant that is conjugated with colloidal gold and a second monoclonal antibody to a different antigenic determinant, the latter antibody being immobilized as a line on a nitrocellulose strip. FSH present in the urine reacts with the conjugated antibody and immobilized antibody, forming a "sandwich" immunocomplex at the site of the immobilized antibody. Unreacted conjugate is washed from the strip by the flow of excess sample. The appearance of a clear red line (test result) indicates FSH in the urine sample, with the intensity of the line proportional to the FSH concentration.

    The nitrocellulose strip also has a comparator line of fixed intensity that corresponds to an FSH concentration of 10 IU/L. A test line of color intensity greater than or equal to the comparator line indicates a concentration of FSH of 10 IU/L or greater. This level is indicative of diminished ovarian reserve.

    AI/ML Overview

    Here's a summary of the acceptance criteria and the study details for the Fertell Female Fertility Test, based on the provided 510(k) summary:

    1. Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Correct performance by general public>97%
    Lay user accuracy (general public vs. professional)>90%
    Accuracy (device vs. quantitative serum FSH)>95%

    2. Sample Size and Data Provenance for the Test Set

    The document does not explicitly state the sample size used for the test set in the consumer study. It also does not specify the country of origin of the data or whether the study was retrospective or prospective.

    3. Number of Experts and Qualifications for Ground Truth

    The document does not provide information on the number of experts used to establish ground truth for the test set or their qualifications. The professional comparisons are mentioned, but no details on their role in defining ground truth.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method used for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study is mentioned in the provided text. The study focuses on lay user performance and comparison to professional or quantitative methods rather than evaluating human reader improvement with AI assistance.

    6. Standalone (Algorithm Only) Performance

    A standalone performance evaluation was conducted for the device. The "lay user also demonstrated >90% accuracy when compared to a professional. and when compared to a quantitative serum FSH tests, the Fertell Female Fertility test demonstrated >95% accuracy." This indicates the device's performance independent of real-time human interpretation, comparing its output directly to reference standards.

    7. Type of Ground Truth Used

    The ground truth used for the performance evaluation includes:

    • Professional observation/interpretation: For assessing lay user accuracy.
    • Quantitative serum FSH tests: For assessing the accuracy of the device's FSH measurement.

    8. Sample Size for the Training Set

    The document does not provide information about a specific training set or its sample size. This is a rapid diagnostic test, and typically, such devices are developed and validated rather than "trained" in the machine learning sense. The testing described likely covers validation rather than a separate training phase.

    9. How the Ground Truth for the Training Set Was Established

    As no specific training set is mentioned in the provided text, information on how its ground truth was established is not available.

    Ask a Question

    Ask a specific question about this device

    K Number
    K030058

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-07-30

    (204 days)

    Product Code
    Regulation Number
    862.1300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    Orange, CA 92865

    Re: K030058

    Trade/Device Name: EarlyDETECT® MENOPAUSE TEST Regulation Number: 21 CFR 862.1300

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EarlyDETECT MENOPAUSE TEST is a qualitative rapid membrane immunoassay for the in vitro diagnostic detection of follicle stimulating hormone (FSH) in human urine as a confirmation of hormone changes related to the symptoms associated with the stages of menopause. This device is intended for professional and Over the Counter (OTC) use.

    Device Description

    Rapid Membrane ImmunoAssay

    AI/ML Overview

    The provided document is a 510(k) premarket notification letter from the FDA for the EarlyDETECT® MENOPAUSE TEST. This document does not contain the information requested regarding acceptance criteria, study details, sample sizes, ground truth establishment, or expert qualifications.

    The letter confirms that the device is substantially equivalent to legally marketed predicate devices and can, therefore, be marketed subject to general controls provisions. It specifies the device name, regulation number, regulation name, and product code, along with its intended use as a qualitative rapid membrane immunoassay for in vitro diagnostic detection of follicle-stimulating hormone (FSH) in human urine for confirming hormone changes related to menopause, intended for professional and Over-The-Counter (OTC) use.

    Therefore, I cannot provide the requested information based on the text provided. A separate submission, often referred to as the 510(k) summary or the full 510(k) submission, would typically contain such detailed study information.

    Ask a Question

    Ask a specific question about this device

    K Number
    K972720

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1997-09-02

    (43 days)

    Product Code
    Regulation Number
    862.1300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    510k Summary Text (Full-text Search) :

    classification name is - follicle-stimulating hormone test system product code CGJ (per 21 CRF section 862.1300

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The quantitative determination of follicle stimulating hormone (FSH) concentration in human serum and plasma by a microplate enzymeimmunoassay.Measurements of follicle-stimulating hormone are used in the diagnosis of pituitary gland and gonadotropin disorders.

    Device Description

    Monobind Inc., registration number 2020726, plans to introduce Into commercial distribution an enzymeimmunoassay (ELISA) kit for the determination of folliclestimulating hormone (FSH) in human serum and plasma.

    The Monobind ELISA method is based on two-site immunoassay (sandwich) technology utilizing the streptavidin-blotin reaction to effect separation. Upon mixing monoclonal blotinylated anti-FSH antibody, the enzyme-labeled anti-FSH antibody and a serum containing the native antigen (FSH), reaction results between the native antigen (FSH) and the antibodies, without competition or steric hindrance, to form a soluble sandwich complex. Simultaneously, the complex is deposited to the well through the high affinity reaction of streptavidin and biotinylated antibody. After incubation is complete, decantation or aspiration separates the bound fraction. The enzyme activity on the well Is directly proportional to the native antigen (FSH) concentration. By utilizing several different serum references of known antigen values, a dose response curve can be generated from which the antigen concentration of an unknown can be ascertained.

    AI/ML Overview

    The provided document describes the Monobind Inc. Follicle-stimulating hormone (FSH) ELISA kit and its submission for 510(k) clearance. The focus of the acceptance criteria and study is on demonstrating substantial equivalence to a predicate device.

    Here's an analysis based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific CriteriaReported Device Performance and Outcome
    Substantial Equivalence (Clinical Comparison)Correlation Coefficient: The new device's results should have a strong correlation with the predicate device's results. This implies a high R-value (typically close to 1). Linear Regression: The relationship between the new device and the predicate device should be linear, allowing for predictable conversion or direct comparison, expressed as y = mx + b. Mean Values: The mean values of the new device should be comparable to the predicate device, indicating a similar central tendency in measurements.Correlation Coefficient: 0.994, indicating very good method agreement. Linear Regression Equation: y = 0.93(x) - 1.5, showing a strong linear relationship. Mean Values: Predicate device (ICMA) mean = 21.0 mIU/ml; New device (ELISA) mean = 18.0 mIU/ml.
    Analytical RecoveryAverage Recovery Rate: When exogenous FSH is added to human serum specimens, the device should accurately recover the added amount, reflecting minimal interference or loss.Average recovery of 99.9% when exogenous added FSH was introduced into human serum specimens.
    Linearity (Dilution Studies)Average Linearity: When specimens with varying FSH concentrations are diluted and compared to the dose response curve, the device should maintain accurate and proportional readings, demonstrating linearity across its measurement range.Average linearity of 93.8% when specimens were diluted and compared to the dose response curve.

    Study Proving Device Meets Acceptance Criteria:

    The study proving the device meets the acceptance criteria is a clinical comparison study (sometimes referred to as a method comparison study) specifically designed to demonstrate substantial equivalence to the predicate device, the Ciba Corning ACS 180 chemiluminescence (ICMA) test.

    2. Sample Size and Data Provenance:

    • Sample Size for Test Set: 128 specimens.
    • Data Provenance: The document does not explicitly state the country of origin. However, the manufacturer is Monobind Inc., located in Costa Mesa, CA (USA), and the submission is to the FDA. It is highly probable the data is from the USA. The study used specimens "from low to elevated populations," indicating a range of clinical samples.
    • Retrospective or Prospective: Not explicitly stated, but clinical comparison studies for 510(k) submissions are typically prospective or involve retrospectively collected samples that are then prospectively tested with both devices. The phrase "clinical comparison... using 128 specimens" suggests these were real-world samples tested with both methods.

    3. Number of Experts and Qualifications:

    Not applicable for this type of device and study. The ground truth itself is established by the predicate device's measurement, not by expert consensus on visual assessment or interpretation. For an in vitro diagnostic (IVD) device like an ELISA kit, the "ground truth" for the comparison study is the result from the established, legally marketed predicate device.

    4. Adjudication Method:

    Not applicable. This is a quantitative measurement device comparison, not an interpretation task requiring adjudication. The "ground truth" is the numerical result from the predicate device.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

    Not applicable. This is an in vitro diagnostic (IVD) device that provides a quantitative measurement of a biomarker in a sample, not an imaging or diagnostic interpretation device where human readers are involved. Therefore, there is no "human reader" component to improve in this context.

    6. Standalone Performance:

    Yes, a standalone performance study was done in the sense that the Monobind ELISA kit generated its own quantitative results independent of human interpretation. The study then compared these standalone results to those of the predicate device. The performance characteristics reported (mean values, recovery, linearity) are standalone performance metrics of the Monobind ELISA.

    7. Type of Ground Truth Used:

    The ground truth for the comparison was the measurement obtained from the legally marketed predicate device, the Ciba Corning ACS 180 chemiluminescence (ICMA) test. This is an accepted method for demonstrating substantial equivalence for IVDs.

    8. Sample Size for the Training Set:

    The document does not explicitly mention a "training set" in the context of an algorithm or machine learning. For an ELISA kit, "training" might refer to assay development and optimization rather than a separate dataset for an algorithm. The reported 128 specimens were used for the clinical comparison/validation study. There is no indication of a separate training set for a machine learning model.

    9. How the Ground Truth for the Training Set was Established:

    As there is no explicit mention of a "training set" for an algorithm, this question is not directly applicable. If one considers the process of developing and optimizing the ELISA kit itself (which could be loosely termed "training" for a traditional assay), the "ground truth" would be established by various biochemical and analytical methods to ensure the assay accurately detected and quantified FSH in reference materials and spiked samples. This would precede the formal clinical comparison study.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1