510(k) Data Aggregation

    K Number
    K200904

    Validate with FDA (Live)

    Date Cleared
    2021-08-05

    (486 days)

    Product Code
    Regulation Number
    862.1373
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K112015, K891235, K142448

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is intended for in vitro diagnostic use for the measurement of % hemoglobin A1c (HbA1c) (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in venous whole blood specimens using ion-exchange high-performance liquid chromatography (HPLC). This test is an aid in diagnosis of diabetes and identifying patients who may be at risk for developing diabetes, and for monitoring of long-term blood glucose control in individuals with diabetes mellitus.

    Device Description

    The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is an automated High-Performance Liguid Chromatography (HPLC) system that separates and reports stable hemoglobin A1c (sA1c) percentage in venous whole blood. The operational portion of the G8 is composed of a sampling unit, liquid pump, degasser, column, detector, microprocessors, sample loader, smart media card, operation panel, and a printer. The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 uses ion-exchange HPLC for rapid, accurate, and precise separation of the stable form of HbA1c (sA1c) from other hemoglobin fractions. The G8 uses a non-porous cation exchange column and separates the hemoglobin components in the blood. Separation is achieved by utilizing differences in ionic interactions between the cation and exchange group on the column resin surface and the hemoglobin components in a step gradient elution. The hemoglobin fractions (designated as A1a. A1b. F. LA1c+, SA1c, A0, and, if present, H-V0, H-V2, H-V2 and H-V3) are subsequently removed from the column by performing a step-wise elution gradient using the varied salt concentrations in the Variant Elution Buffers HSi 1, 2 and 3. The peaks, H-V0, H-V1, H-V2 and H-V3 are typically presumptive HbAD, HbAS, HbAC and HbAE respectively. The software compares the retention times of hemoglobin fractions in a sample to the expected "windows of retention" and labels each fraction that correctly elutes within a defined expected window of retention. The software designates a hemoglobin fraction as POX (where X is the order of the peak as it elutes from the column) if it does not match a defined window of retention. All automated processes in the G8 are controlled by internal microprocessors, using software downloaded via a smart media card. The result report is printed and can be stored on the instrument. The data can be transmitted to a host computer through a bi-directional interface. The result report includes the sample ID, date, percentage and retention time of each fraction of hemoglobin, sA1c percentage and total A1 percentage, along with a chromatogram of the elution pattern of the hemoglobin fractions. If a sample contains a hemoglobin variant, the column elutes the fraction depending upon its charge.

    AI/ML Overview

    The provided text describes the non-clinical performance testing of the Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 (subject device) to support its substantial equivalence to a predicate device. This document focuses on the analytical performance of a diagnostic device rather than an AI/ML powered device, so some of the specific questions regarding AI/ML study design (e.g., number of experts, adjudication methods, MRMC studies) are not applicable.

    Here's the information extracted from the document:

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by the statement "All performance testing results met their pre-determined acceptance criteria." While explicit numerical acceptance criteria for each test are not listed in a consolidated table, the discussion throughout the "Summary of Non-Clinical Performance Testing" implicitly defines them through the methodology and results. For example, for precision/repeatability, the claim of "imprecision at ≤ 2%" was a pre-established criterion. Similarly, for hemoglobin variant interference, "Non-clinically significant interference was defined as <= 6% relative difference in the results from a comparative method."

    It's important to note that the provided text is a 510(k) summary, which often summarizes detailed study reports. The full reports would contain explicit acceptance criteria thresholds.

    Table of Acceptance Criteria (Implied) and Reported Device Performance:

    Acceptance Criteria (Implied/Defined)Reported Device Performance
    Precision/Repeatability: Pre-established claim of imprecision at ≤ 2% CV (for Total CV across all factors, or similar)Total CV for combined analyzers: - 5.46% HbA1c: 0.97% - 6.38% HbA1c: 1.04% - 7.60% HbA1c: 1.28% - 11.91% HbA1c: 0.87% (All reported Total CVs are well below the implied 2% or potentially a specific lower threshold per concentration)
    Method Comparison (Bias): No explicit acceptance criterion given for bias in the tables, but implied by outcome.Passing-Bablok Bias: - 5.0% HbA1c: 0.1753 (3.5% bias) - 6.5% HbA1c: 0.2068 (3.2% bias) - 8.0% HbA1c: 0.2383 (3.0% bias) - 12.0% HbA1c: 0.3224 (2.7% bias) Deming Bias: - 5.0% HbA1c: 0.1979 (4.0% bias) - 6.5% HbA1c: 0.2172 (3.3% bias) - 8.0% HbA1c: 0.2366 (3.0% bias) - 12.0% HbA1c: 0.2884 (2.4% bias) (The device is found to be "substantially equivalent" based on these results, implying acceptance)
    Hemoglobin Variant Interference: Non-clinically significant interference defined as <= 6% relative difference in the results from a comparative method at 6.5% or 8.0% HbA1c.Reported Relative Bias from Reference Method: HbAD: -0.5% (~6.5% HbA1c), -1.7% (~8.0% HbA1c) HbAS: -2.7% (~6.5% HbA1c), -3.2% (~8.0% HbA1c) HbAC: -1.9% (~6.5% HbA1c), -1.1% (~8.0% HbA1c) HbAE: -1.3% (~6.5% HbA1c), -1.2% (~8.0% HbA1c) HbA2: -4.2% (~6.5% HbA1c), -5.1% (~8.0% HbA1c) HbF: -0.7% (~6.5% HbA1c), -1.6% (~8.0% HbA1c) (All reported relative biases are within the <= 6% redefined significant interference threshold)
    Endogenous Interfering Substances: Significant interference defined as percent recovery ± 5% of the expected 100% recovery.No Interference concluded up to specific concentrations: - Albumin: 5000 mg/dL - Ascorbic Acid: 25 mg/dL - Bilirubin C: 21 mg/dL - Bilirubin F: 18 mg/dL - Lipemia: 1000 mg/dL - Rheumatoid Factor: 550 IU/mL (Implies results were within ± 5% recovery for these substances up to the tested concentrations)
    Cross Reactivity with Hemoglobin Derivatives: Concluded as "not interfering with the assay".Not interfering concluded for: - Acetylated Hb up to 50 mg/dL - Carbamylated Hb up to 25 mg/dL - Aldehyde Hb up to 25 mg/dL - Labile HbA1c up to 1000 mg/dL
    Matrix Comparison: No clinical or statistical difference between K2-EDTA and K3-EDTA.Data supports interchangeable use of K2-EDTA and K3-EDTA blood collection tubes.
    Linearity and Detection Limit: Reportable range 4.0 to 16.9% HbA1c.Previously established under 510(k) K071132, and the range is maintained.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Precision/Repeatability Study:
      • Sample Size: Four concentrations of HbA1c% in K2EDTA venous whole blood. A total of 720 measurements per concentration were performed (three analyzers, over 20 non-consecutive days, with three reagent lots, samples run in duplicate, two times per day).
      • Data Provenance: The study was implicitly conducted retrospectively on collected samples. The country of origin is not explicitly stated, but the testing was performed at "Tosoh Bioscience, Inc." laboratories, likely in the US (given the South San Francisco address).
    • Method Comparison Study:
      • Sample Size: 220 K₂EDTA venous whole blood specimens with HbA1c concentrations spanning the measuring range (4.0-16.9%). Each specimen was tested in duplicate.
      • Data Provenance: The study was conducted at "two separate sites," one a "moderate complexity clinical laboratory" (for the candidate device) and the other an "NGSP SRL" (for the comparator device). The precise country of origin is not explicitly stated. The specimens were collected and likely used retrospectively for the comparison.
    • Endogenous Interfering Substances / Hemoglobin Variant Interference Studies:
      • Sample Size:
        • Endogenous interference: Not explicitly stated per substance, but specimens were "spiked with increasing amounts of the interfering substance."
        • Hemoglobin Variant Interference:
          • HbC: 26 samples
          • HbD: 24 samples
          • HbE: 26 samples
          • HbS: 29 samples
          • HbA2: 20 samples
          • HbF: 21 samples
      • Data Provenance: Not explicitly stated, but implies lab-tested, potentially spiked, and clinical samples from unknown origin (likely retrospective).

    3. Number of Experts and Qualifications for Ground Truth:

    This device is an automated in vitro diagnostic analyzer for quantitative measurement of HbA1c. The "ground truth" for such devices is typically established by:

    • Reference methods (e.g., NGSP SRL certified methods).
    • Traceability to international standards (DCCT, IFCC).
    • Direct analytical measurements and statistical comparisons, rather than expert consensus on image interpretation.

    Therefore, the concept of "number of experts used to establish the ground truth" similar to that in an AI imaging study is not applicable here. The ground truth is the chemical measurement itself, validated against recognized reference standards.

    4. Adjudication Method for the Test Set:

    Not applicable, as this is an analytical device for quantitative measurement, not an interpretative AI/ML system requiring human adjudication of results. The "adjudication" is through statistical agreement with reference methods and assessment against predetermined analytical performance criteria.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    No. An MRMC study is relevant for AI/ML systems that assist human readers in tasks like image interpretation. This device performs automated quantitative measurements, and its performance is evaluated against reference methods and analytical standards, not human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Yes, the studies described are standalone performance studies of the device (Tosoh Automated Glycohemoglobin Analyzer HLC-723G8) without human intervention in the measurement process itself. The device is an automated HPLC system.

    7. The Type of Ground Truth Used:

    The ground truth for evaluating the subject device's performance is based on:

    • Reference Methods: Specifically, the NGSP SRL (National Glycohemoglobin Standardization Program Secondary Reference Laboratory) using the Trinity Biotech Premier Hb9210™ HbA1c Analyzer (for method comparison) and Primus Model CLC 330 (for some hemoglobin variant interference testing), and the Bio-Rad VARIANT II TURBO HbA1c Kit - 2.0 (for some hemoglobinopathy interference testing).
    • Traceability and Standardization: The device's results are traceable to the Diabetes Control and Complications Trial (DCCT) reference method and IFCC, and it is certified by the NGSP.
    • Analytical Standards: For precision, linearity, and interference, the "ground truth" reflects the known concentrations of spiked samples or validated concentrations in patient samples, against which the device's measurements are compared.

    8. The Sample Size for the Training Set:

    The provided document describes non-clinical performance testing to support substantial equivalence after a software modification (v5.24). It does not provide information on the training set used for the development of the device's algorithms or software. This 510(k) summary is focused on verification and validation of the modified device.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not provided in the 510(k) summary. Given that the device is an automated HPLC system, its "training" per se would involve optimization of its operational parameters, calibration curves, and potentially peak detection/integration algorithms, likely using well-characterized samples with known HbA1c concentrations established by reference methods. However, the details of such a "training set" and its ground truth establishment are not discussed here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112015

    Validate with FDA (Live)

    Date Cleared
    2011-11-22

    (132 days)

    Product Code
    Regulation Number
    864.7470
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K891235

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Premier Hb9210™ system is intended for the quantitative measurement of hemoglobin A1c (HbA1c) in human capillary and venous whole blood. HbA1c is used for the monitoring of long-term glycemic control in individuals with diabetes mellitus. For in vitro diagnostic use only.

    Device Description

    The Premier Hb9210™ is an update of the predicate HPLC device, the Ultra2, which is the current model under K891235. The Premier Hb9210™ is a compact, integrated HPLC system and workstation with the Trinity Biotech AFFINITY Software for the quantitative determination of glycated hemoglobin using the patented boronate affinity chemistry with high performance liquid chromatography.

    AI/ML Overview

    Here's an analysis of the provided information, addressing your requested points:

    Device: Trinity Biotech Premier Hb9210™ HbA1c Analyzer

    The Premier Hb9210™ is an automated High-Performance Liquid Chromatography (HPLC) system designed for the quantitative measurement of HbA1c in human capillary and venous whole blood. This device uses boronate affinity chemistry for HbA1c measurement, building upon the technology of its predicate device, the Ultra2 (K891235).


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" for the performance metrics in a pass/fail format. Instead, it presents the results of performance tests aimed at demonstrating substantial equivalence to a predicate device. For the purpose of this response, I will interpret the reported performance from the precision and accuracy studies as the device's demonstrated performance.

    Note: The document focuses on demonstrating substantial equivalence to a predicate device (Ultra2, K891235), not necessarily meeting specific pre-defined acceptance criteria for novel device approval.

    Performance MetricReported Device Performance (Premier Hb9210™)
    Intra-run Precision (Hemolysate - All Sites)
    %HbA1c Concentration5.76%
    Repeatability SD (Sr)0.07
    Repeatability %CV1.26
    Within-device Precision SD (ST)0.09
    Within-device Precision %CV1.62
    Accuracy (vs IFCC HbA1c Standards)r = 0.9998
    Accuracy (vs Predicate Device Ultra2 - 51 samples)r = 0.9976
    Linearity3.7% - 18.5% HbA1c

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Precision Test Set:
      • Three hemolysate samples, each representing normal, decision point, and abnormal levels of %HbA1c.
      • Each of these three samples was analyzed 80 times per site (2 analyses per run, 2 runs per day, over 20 non-consecutive days) across 3 sites.
      • Total replicates: 240 replicates per HbA1c level (3 HbA1c levels * 80 replicates per site * 3 sites) if referring to individual measurements contributing to the overall analysis. The document states "a total of 80 replicates of each level per site, 240 replicates overall".
      • Total observations for precision study: 1200 total observations (implicitly across the 3 HbA1c levels and 3 sites).
    • Sample Size for Accuracy Test Set (vs. predicate): 51 samples.
    • Data Provenance: Not explicitly stated, but the study was conducted at one internal and two external sites. The phrase "aliquotted and frozen whole blood" suggests controlled sample handling, but the origin country of the patient samples or whether they were retrospective or prospective samples is not specified. Given the context of a 510(k) submission, it's typically based on controlled laboratory studies rather than a full clinical trial with diverse patient populations and retrospective/prospective designations.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This device is an in vitro diagnostic (IVD) for quantitative measurement, not an AI/ML-driven diagnostic requiring human expert interpretation for ground truth. Therefore, the concept of "experts establishing ground truth" in the way it applies to image interpretation or clinical diagnosis by human readers does not apply in this context.

    • Ground truth for the HbA1c levels in the precision study samples would have been established by a highly accurate reference method or certified reference materials, not by human expert consensus or adjudication.
    • For the accuracy study comparing to the predicate device, the predicate device itself served as the reference for comparison of the 51 samples.
    • For the accuracy study against IFCC HbA1c Standards, the standards themselves provide the ground truth.

    4. Adjudication Method for the Test Set

    Not applicable. As described in point 3, this is a quantitative measurement device, not one where human interpretation requires adjudication.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an automated standalone in vitro diagnostic analyzer, not an AI-assisted diagnostic tool that aids human readers. Therefore, an MRMC study is not relevant.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the performance studies described (precision, accuracy, linearity) are inherently standalone performance studies for the Premier Hb9210™ analyzer. The device performs the measurement and outputs the HbA1c value without human interpretative input in the measurement process itself.


    7. The Type of Ground Truth Used

    • Precision Study: Hemolysate samples representing different HbA1c levels. The "actual" concentrations were likely assigned using a reference method or certified values.
    • Accuracy Study (IFCC HbA1c Standards): Certified IFCC HbA1c Standards. These are internationally recognized reference materials providing an agreed-upon "true" value for comparison.
    • Accuracy Study (against predicate): The predicate device (Ultra2) results were used as the comparator for the 51 samples. While the predicate device itself is not "ground truth" in the strictest sense of a gold standard, it serves as the benchmark against which substantial equivalence is claimed.

    8. The Sample Size for the Training Set

    Not applicable. This document describes an in vitro diagnostic device using a well-established chemical and physical principle (boronate affinity HPLC). It is not an AI/ML device that requires a "training set" of data in the common sense of machine learning. The device's operational parameters are determined through engineering and chemical optimization, not through data-driven training of an algorithm.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this type of IVD device in the context of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062658

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-11-26

    (446 days)

    Product Code
    Regulation Number
    864.7470
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K891235

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Primus A1care Assay A1c test, for use with the TRI•stat™ Instrument, is a rapid in vitro diagnostic test for measurement of the percent of glycated hemoglobin (%HbA1c) level in human blood from finger stick or venous samples for clinical laboratory and point-of-care use. Measurement of percent HbA1c is used to monitor long-term glucose control in individuals with diabetes mellitus.

    Device Description

    The Primus TRI•stat™ Instrument is a small (10"Wx11"Lx4"H), in vitro diagnostic instrument used with the Primus A1care Assay test to quantitate HbA1c using a patented two-phase optical method. The TRI•stat™ is capable of analyzing a total of 3 samples simultaneously.

    AI/ML Overview

    The provided document is a 510(k) summary for the Primus TRI•stat™ Instrument and A1Care Assay, indicating substantial equivalence to a predicate device (K891235, Primus Boronate Affinity HPLC Method). However, it does not contain detailed information about specific acceptance criteria, study methodologies, sample sizes, ground truth establishment, or expert qualifications that would typically be found in a comprehensive study report.

    The document primarily focuses on establishing equivalent intended use, technology, and performance to the predicate device. It states: "The Primus TRI•stat™ Instrument and Arcare Assay A1c Test was evaluated for nonclinical and clinical performance characteristics in comprehensive studies. These studies demonstrate that the instrument and test substantially equivalent to the predicate device and are safe and effective for their intended use."

    Without the actual study reports, the following information cannot be extracted:


    Here's a table based on the information not available in the provided document, indicating what would typically be requested for such a description:

    Acceptance Criteria and Device Performance (Not Provided in Document)

    MetricAcceptance CriteriaReported Device Performance
    Accuracy / BiasNot specifiedNot specified
    Precision (Within-run, Between-run, Total)Not specifiedNot specified
    Linearity / Measuring RangeNot specifiedNot specified
    InterferenceNot specifiedNot specified
    Method Comparison (vs. Predicate/Reference Method)Not specifiedNot specified
    Correlation Coefficient (R)Not specifiedNot specified
    Regression Equation (Slope, Intercept)Not specifiedNot specified

    Here's a breakdown of the other requested information, indicating what is available and what is not:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Sample Size (Test Set): Not specified in the provided document. The document only mentions "comprehensive studies."
      • Data Provenance: Not specified.
      • Retrospective or Prospective: Not specified.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not applicable as the device measures a quantitative biomarker (HbA1c). Ground truth for such devices is typically established through reference methods or certified reference materials, not expert consensus in the diagnostic imaging sense. The document implies comparison to the predicate device (Primus Boronate Affinity HPLC Method) and "known standards."
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not applicable for a quantitative diagnostic assay. Performance is assessed by statistical comparison to a reference method or certified values.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not applicable. This is a standalone in-vitro diagnostic instrument for measuring a biomarker, not an AI-assisted diagnostic imaging device that involves human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • The document describes the TRI•stat™ Instrument and A1Care Assay as standalone in-vitro diagnostic device. It quantifies HbA1c using a "patented two-phase optical method." While the instrument is "Partially Automated" and "Requires Sample Insertion and Tube Insertion," the measurement and result generation appear to be fully automated once the sample is loaded. Therefore, performance evaluation would be standalone. The document states "The Primus TRI•stat™ Instrument and Arcare Assay A1c Test was evaluated for nonclinical and clinical performance characteristics in comprehensive studies."
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • The document implies that performance is benchmarked against "known standards" and the predicate device (Primus Boronate Affinity HPLC Method). For HbA1c measurements, ground truth would typically be established by:
        • Reference methods: Such as HPLC methods (like the predicate device) or methods traceable to IFCC reference systems.
        • Certified reference materials: Samples with known, validated HbA1c concentrations.
    7. The sample size for the training set

      • Not specified. The document does not describe the development or training of any machine learning model, only the evaluation of the full device. It mentions "factory calibrated" for optical measurements, which implies internal calibration data but not a distinct "training set" in the context of machine learning.
    8. How the ground truth for the training set was established

      • Not applicable as no specific "training set" for a machine learning model is mentioned. Calibration and performance are likely based on reference materials and comparisons to established methods.