510(k) Data Aggregation

    K Number
    K132231

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-11-27

    (132 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K111302, K100931, K081794, K091009, K110688

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Esaote's Model 6400 is a mainframe ultrasound system diagnostic general ultrasound studies including Cardiac, Transesophageal Cardiac, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small organs, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatric and Other: Urologic. The 6400 system provides imaging for guidance of biopsy and imaging to assist in the placement of needles in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.

    Esaote's Model 7400 is a compact ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Transesophageal Cardiac, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small organs, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatric and Other: Urologic. The 7400 system provides imaging for guidance of biopsy and imaging to assist in the placement of needles in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.

    Device Description

    Model 6400 is a mainframe system equipped with wheels allowing them to move the system. Model 7400 is a portable system equipped with a handle. The system size and weight allow it to be carried using its handle. The primary modes of operation are for both models: B-Mode, M-Mode, Tissue Enhancement Imaging (TEI), XView, Multi View (MView), Trapezoidal View (TPView), Doppler, Color Flow Mapping (CFM), Amplitude Doppler (AD) and Tissue Velocity Mapping (TVM). Both 6400 and 7400 are equipped with a LCD color display where acquired images and advanced image features are shown. Both 6400 and 7400 can drive Phased, Convex, Linear array and Doppler probes. 6400 control panel is equipped with a pull-out qwerty alphanumeric keyboard that allows data entry. On 7400 model the touchscreen has an emulation of the qwerty alphanumeric keyboard that allows data entry. Both 6400 and 7400 models are equipped with wireless capability. Model 7400 was designed to be powered by battery.

    Both 6400 and 7400 have been cleared via K111302.

    6400 and 7400 Upgrade, defined herein, combine the cleared features of both 6400 and 7400 systems with other new software capabilities, listed below:

      1. Management of motorized probes (Bi-Scan probes) that allow volumetric acquisition.
      1. Management of 3D/4D mode of operation.
      1. Management of the acquisition of consecutive ultrasound images on extended surface. The final image is composed of the consecutive frames placed side (Panoramic View or VPan).

    The 6400 and 7400 Upgrades are manufactured under an ISO 9001:2000 and ISO 13485 certified quality system.

    AI/ML Overview

    This 510(k) summary for the Esaote 6400 and 7400 Ultrasound Systems (K132231) indicates that no clinical tests were performed to establish acceptance criteria or demonstrate device performance.

    Instead, the submission relies on demonstrating substantial equivalence to previously cleared predicate devices and conformity to applicable medical device safety and performance standards.

    Therefore, most of the requested information regarding acceptance criteria and performance studies, if they were based on clinical data, cannot be extracted from this document as such studies were explicitly stated as not performed.

    However, I can extract information related to the device's technological characteristics and the non-clinical tests performed.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since no clinical studies were performed, there are no specific performance metrics (e.g., sensitivity, specificity, accuracy) or quantitative acceptance criteria for diagnostic capability that can be reported from this document.
    The acceptance criteria are implicitly based on:

    • Substantial Equivalence: The new features and devices (6400 and 7400 Upgrades) are deemed substantially equivalent to previously cleared predicate devices (Esaote 6200, 7340, 6400, 7400 models). This means their performance is expected to be no worse than the predicate devices, which were already deemed safe and effective.
    • Conformity to Safety Standards: The devices meet various international and national safety standards.
    Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
    Technological EquivalenceClinical UsesEquivalent to predicate devices (K111302, K100931, K081794, K091009, K110688)
    Bi-Scan probe managementEquivalent to predicate devices
    3D/4D mode of operationEquivalent to predicate devices
    Panoramic View (VPan)Equivalent to predicate devices
    Safety StandardsIEC 60601-1Conforms to standard
    IEC 60601-1-2Conforms to standard
    IEC 60601-2-37Conforms to standard
    NEMA UD-3 (Acoustic Output Display)Conforms to standard, equivalent Ispta and MI maximal values to predicate
    NEMA UD-2 (Acoustic Output Measurement)Conforms to standard
    Non-Clinical TestsAcoustic OutputEvaluated and conforms
    BiocompatibilityEvaluated and conforms
    Cleaning & Disinfection EffectivenessEvaluated and conforms
    Thermal SafetyEvaluated and conforms
    Electromagnetic SafetyEvaluated and conforms
    Mechanical SafetyEvaluated and conforms

    2. Sample Size Used for the Test Set and Data Provenance:

    • Not Applicable. No clinical test set data was used as "No clinical tests were performed." The submission relies on comparative data with predicate devices and non-clinical engineering tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • Not Applicable. No clinical tests were performed, so no expert ground truth for a clinical test set was established as part of this submission.

    4. Adjudication Method for the Test Set:

    • Not Applicable. No clinical tests were performed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This device is an ultrasound imaging system, not an AI-assisted diagnostic tool in the sense of software interpreting images for human readers. No MRMC comparative effectiveness study was done for AI assistance. The submission focuses on hardware and software feature upgrades to existing ultrasound systems.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is an ultrasound imaging system, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not Applicable. For the purpose of establishing clinical performance, no ground truth data from clinical trials was used, as no clinical trials were performed. The "ground truth" for substantial equivalence is based on the cleared status and established safety and effectiveness of the predicate devices and conformity to engineering standards.

    8. The sample size for the training set:

    • Not Applicable. No clinical data was used for training purposes, as no clinical tests or AI/machine learning components requiring such training are described.

    9. How the ground truth for the training set was established:

    • Not Applicable. No clinical training set was used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K111302

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-07-22

    (74 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Why did this record match?
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Esaote's Model 6400 is a mainframe ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Transesophageal Cardiac, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small organs, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatric, and Other: Urologic. The 6400 system provides imaging for guidance of biopsy and imaging to assist in the placement of needles in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.

    Esaote's Model 7400 is a compact ultrasound system ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Transesophageal Cardiac, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small organs, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatric, and Other: Urologic. The 7400 system provides imaging for guidance of biopsy and imaging to assist in the placement of needles in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.

    Device Description

    6400 is a mainframe system equipped with wheels allowing to move the system. 7400 is a portable system equipped with a handle. The system size and weight allow it to be carried using its handle. The primary modes of operation are: B-Mode, M-Mode, Tissue Enhancement Imaging (TEI), XView, Multi View (MView), Trapezoidal View (TPView), Doppler, Color Flow Mapping (CFM), Amplitude Doppler (AD) and Tissue Velocity Mapping (TVM). Both 6400 and 7400 are equipped with a LCD color display where acquired images and advanced image features are shown. Both 6400 and 7400 can drive Phased, Convex, Linear array, Doppler probes and Volumetric probes. 6400 control panel is equipped with a pull-out Qwerty alphanumeric keyboard that allows data entry. On 7400 model the touchscreen has an emulation of the Qwerty alphanumeric keyboard that allows data entry. Both 6400 and 7400 systems are equipped with wireless capability. The 6400 and 7400 systems are manufactured under an ISO 9001:2000 and ISO 13485 certified quality system.

    AI/ML Overview

    The provided text is a 510(k) summary for the Esaote 6400 and 7400 Ultrasound Systems. This type of submission focuses on demonstrating substantial equivalence to a predicate device rather than efficacy through clinical studies and acceptance criteria as would be expected for a novel device. As such, the information typically requested in your prompt (e.g., acceptance criteria, test set, experts, adjudication, MRMC studies, standalone performance, ground truth for test/training sets, and training set size) is not explicitly present in this document.

    The document explicitly states: "No clinical tests were performed." This indicates that the device's performance was not evaluated against specific acceptance criteria through a clinical study in the way your prompt describes. Instead, the submission relies on demonstrating similarity to existing, legally marketed predicate devices.

    However, I can extract information related to the device's characteristics and the non-clinical tests performed to support its safety and performance.

    Here's the summary of what can be inferred and what is explicitly stated:


    1. Table of Acceptance Criteria and Reported Device Performance:

    • Acceptance Criteria: Not explicitly stated as this is a 510(k) submission based on substantial equivalence, not a novel device requiring new efficacy criteria. The "acceptance criteria" here are effectively demonstrating equivalence to the predicate devices.
    • Reported Device Performance: The primary performance characteristic claimed is "substantial equivalence" to predicate devices, meaning it performs similarly in its intended uses. The document highlights functional equivalence and compliance with safety standards.
    CharacteristicAcceptance Criteria (Implied for Substantial Equivalence)Reported Device Performance
    Technological CharacteristicsEquivalent to predicate devices (Esaote 7350, 7340, 8100) in fundamental technology."The 6400 and 7400 systems employ the same fundamental technological characteristics as their predicate devices." (K050326, K052805, K060827 for 7350; K081794, K091009 for 7340; K101605 for 8100 Wi-Fi).
    Clinical Uses/Intended UseEquivalent to those cleared for predicate devices (Esaote 7350 and 7340)."Clinical uses for which respectively 6400 and 7400 are designed, are equivalent to those cleared for Esaote 7350 and 7340." Both devices are intended for diagnostic general ultrasound studies including Cardiac, Transesophageal Cardiac, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small organs, Musculoskeletal, Abdominal, Fetal, Transvaginal, Transrectal, Pediatric, Urologic, and imaging for guidance of biopsy and needle placement.
    Safety RequirementsConform to IEC60601-1 and IEC60601-2-37."Both 6400 and 7400 are designed to meet the IEC60601-1 and IEC60601-2-37 safety requirements." Device evaluated for acoustic output, biocompatibility, cleaning/disinfection, thermal, electromagnetic, and mechanical safety, and found to conform to these standards, as well as NEMA UD-3 and NEMA UD-2.
    Acoustic Output DisplayProvide Acoustic Output Display feature per AIUM / NEMA standards, with equivalent Ispta and MI maximal values to predicate devices."Both 6400 and 7400 provide an Acoustic Output Display feature per AIUM / NEMA standards, with equivalent Ispta and MI maximal values."
    Measurements and Analysis PackageSimilar to predicate devices (Esaote 6400, 7400, 7350, 7340 systems)."Esaote 6400, 7400, 7350 and 7340 systems provide a similar measurements and analysis package."
    Digital Storage Capabilities/ConnectivityInclude Network connectivity."Esaote 6400, 7400, 7350 and 7340 systems have digital storage capabilities, including Network connectivity." Both 6400 and 7400 are equipped with wireless capability.
    7400 Battery PoweringEquivalent to 7340 system cleared for battery powering."Esaote 7400, designed to be powered by battery, is equivalent to 7340 system, cleared for battery powering via K081794."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Not applicable. The document explicitly states "No clinical tests were performed." Therefore, there is no test set in the context of clinical performance evaluation. Non-clinical tests were performed on the devices themselves to ensure compliance with standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable. As no clinical tests were performed, there was no test set requiring ground truth established by experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. No clinical tests were performed.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. No clinical tests were performed, and this device is an ultrasound imaging system, not an AI-powered diagnostic tool that assists human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is an ultrasound imaging system, not a standalone algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable. No clinical tests were performed to evaluate an algorithm against ground truth. The "ground truth" in this submission relates to compliance with engineering and safety standards (e.g., whether acoustic output measures meet NEMA standards, or whether the device can perform the listed modes of operation as intended).

    8. The sample size for the training set:

    • Not applicable. This submission is for an ultrasound imaging system, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. This submission is for an ultrasound imaging system, not an AI/ML algorithm that requires a training set and associated ground truth.

    Conclusion from the Document:

    The Esaote 6400 and 7400 Ultrasound Systems received 510(k) clearance based on demonstrating substantial equivalence to existing predicate devices (Esaote 7350, 7340, and 8100). The submission focuses on showing that the new devices have the same technological characteristics, intended uses, and meet the same non-clinical safety and performance standards as the legally marketed predicates. No clinical studies were performed to establish performance against specific acceptance criteria for a novel functionality.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103152

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-01-07

    (73 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K051308, K052805, K060827, K081386, K091009

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Esaote's Model 6250 is a mainframe ultrasound system used to perform diagnostic general ultrasound studies including Cardiac. Transesophageal, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic. Small organ, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatric, Intraoperative (Abdominal), Laparoscopic and Other: Urologic. The 6250 system provides imaging for guidance of biopsy and imaging to assist in the placement of needles in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.

    The Virtual Navigator is a MyLab optional license that provides additional image information from a second modality like CT or MR, during a clinical ultrasound session. The second modality provides additional security in assessing the morphology of the ultrasound image.

    Virtual Navigator can be used in the following application: Abdominal, Musculo-skeletal, Urologic, and Vascular.

    The second modality image is not intended to be used as a standalone diagnostic image since it represents information of a patient that could not be congruent with the current (actual) patient position and shall therefore always been seen as an additional source of information.

    The Virtual Navigator tracking system should not be used on or around persons with a cardiac pacemaker, and should not be used around life supporting equipment.

    Device Description

    The 6150 is a mainframe ultrasound system, used to perform diagnostic general ultrasound studies. Its primary modes of operation are: B-Mode, XView, Multi View (MView), Trapezoidal View (TPView), Doppler, Color Flow Mapping, Amplitude Doppler (AD), Tissue Velocity Mapping (TVM) and Tissue Enhancement Imaging (TEI). The system is equipped with a LCD Color Display, a control panel and is capable of operating Linear, Convex, and Phased array probes.

    The modified 6150, with respect to the cleared version 6150 via K051308, K052805, K060827, K081386 and K091009, is due to the improvements of the system. These modifications, that do not affect the intended use or alter the fundamental scientific technology of the device, are the following:

    • a. New keyboard (control panel), where some control keys (knobs and keys) have been replaced by a touch screen.
    • b. New plastic housing of the system both to include the touch screen and to give a new style
    • c. New keyboard group PCBs lay out to interface the touch screen and to match the new organization of the panel keys.
    • Software/Firmware modification to translate the touch screen information for the d. software: main software characteristics and performances have not been changed.

    The 6250 is the model name of the modified 6150. The set of probes of the 6150 is exactly the same.

    The 6250 system is manufactured under an ISO 9001:2000 and ISO 13485 certified quality system.

    AI/ML Overview

    The provided Special 510(k) Summary for the Esaote 6250 System describes modifications to a previously cleared ultrasound system (6150). The document does not describe a new study to establish acceptance criteria or device performance, but rather attests that the modifications do not alter the fundamental scientific technology or intended use. Therefore, the device meets the previously established acceptance criteria by demonstrating substantial equivalence to the predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a Special 510(k) for device modifications, explicit "acceptance criteria" for a new study are not presented. Instead, the justification relies on the modified device (6250) maintaining the same performance and safety characteristics as the predicate devices (6150 cleared via K051308, K052805, K060827, K081386 and K091009). The "performance" assessment is based on the nature of the modifications not affecting clinical functionality.

    The document implicitly refers to the performance of the predicate device for all clinical applications. The "Reported Device Performance" for the 6250 system is therefore considered to be equivalent to the previously cleared 6150 system, as the modifications were primarily cosmetic (new keyboard, plastic housing, PCB layout) and software/firmware changes not affecting main software characteristics and performances.

    Feature/CharacteristicAcceptance Criteria (Implicit, based on predicate 6150)Reported Device Performance (6250)
    Intended UseIdentical to predicate 6150Maintained identical intended uses
    Fundamental Scientific TechnologyIdentical to predicate 6150No alteration to fundamental technology
    Clinical Modes of OperationIdentical to predicate 6150 for each transducer and application (B-Mode, XView, Multi View, Trapezoidal View, Doppler, Color Flow Mapping, Amplitude Doppler, Tissue Velocity Mapping, Tissue Enhancement Imaging, and other specified modes per application)Maintained identical modes of operation and equivalent performance for all transducers and applications.
    Safety ProfileEquivalent to predicate 6150No adverse impact on safety identified by modifications

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a new test set or a new study with a specific sample size. The modifications are justified based on their nature, not requiring new clinical performance data to demonstrate equivalence. Therefore, there is no mention of sample size or data provenance (country of origin, retrospective/prospective). This is typical for a Special 510(k) submission where changes are considered minor and do not impact safety or effectiveness.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Since no new test set or clinical study is described, there is no mention of experts establishing ground truth for a test set.

    4. Adjudication Method for the Test Set

    As no new test set or clinical study is described, no adjudication method is mentioned.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was performed or is mentioned in this Special 510(k) summary, as the modifications did not warrant a new clinical efficacy trial. The submission seeks to demonstrate substantial equivalence based on the nature of the changes.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This device is an ultrasound imaging system with optional features (Virtual Navigator). It is not an "algorithm-only" or AI-driven diagnostic device as understood in the context of standalone performance studies for AI. The modifications described pertain to hardware and user interface elements, and minor software updates that did not change the core performance. Therefore, a standalone performance study is not applicable and not described.

    7. Type of Ground Truth Used

    No new ground truth was established for this Special 510(k). The regulatory approval relies on the ground truth previously established for the predicate 6150 system, which would have been based on established clinical and technical standards for ultrasound imaging.

    8. Sample Size for the Training Set

    Since no new algorithms or machine learning components are introduced that require a training set, the document does not mention any sample size for a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no new training set or machine learning algorithms are described in this Special 510(k) submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K100931

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-06-25

    (81 days)

    Product Code
    Regulation Number
    892.1550
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K091009

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Esaote's Model 6200 is a mainframe ultrasound system used to perform diagnostic general ultrasound studies including Cardiac, Transesophageal, Peripheral Vascular, Neonatal Cephalic, Adult Cephalic, Small organ, Musculoskeletal (Conventional and Superficial), Abdominal, Fetal, Transvaginal, Transrectal, Pediatric, Intraoperative (Abdominal), Laparoscopic and Other: Urologic. The 6200 system provides imaging for guidance of biopsy and imaging to assist in the placement of needles in vascular or other anatomical structures as well as peripheral nerve blocks in Musculoskeletal applications.

    The Virtual Navigator is a MyLab optional license that provides additional image information from a second modality like CT or MR, during a clinical ultrasound session. The second modality provides additional security in assessing the morphology of the ultrasound image.

    Virtual Navigator can be used in the' following application: Abdominal. Musculo-skeletal. Urologic, and Vascular,

    The second modality image is not intended to be used as a standalone diagnostic image since it represents information of a patient that could not be congruent with the current (actual) patient position and shall therefore always been seen as an additional source of information.

    The Virtual Navigator tracking system should not be used on or around persons with a cardiac pacemaker, and should not be used around life supporting equipment.

    Device Description

    The 6100 is a mainframe ultrasound system, used to perform diagnostic general ultrasound studies. Its primary modes of operation are: B-Mode, XView, Multi View (MView), Trapezoidal View (TPView), Doppler, Color Flow Mapping, Amplitude Doppler (AD), Tissue Velocity Mapping (TVM) and Tissue Enhancement Imaging (TEI). The system is equipped with a LCD Color Display, a control panel and is capable of operating Linear, Convex, and Phased array probes.

    The 6100 system has been cleared by FDA via K051837, K060827 and K081386.

    The modified 6100, with respect to the cleared version 6100 via K051837, K060827 and K081386, is due to the improvements of the system. These modifications, that do not affect the intended use or alter the fundamental scientific technology of the device, are the following:

    • a. New keyboard (control panel), where some control keys (knobs and keys) have been replaced by a touch screen.
    • b. New plastic housing of the system both to include the touch screen and to give a new style
    • c. New keyboard group PCBs lay out to interface the touch screen and to match the new organization of the panel keys.
    • d. Software/Firmware modification to translate the touch screen information for the software: main software characteristics and performances have not been changed.

    The 6200 is the model name of the modified 6100. The set of probes of the 6100 is exactly the same.

    The 6200 system is manufactured under an ISO 9001:2000 and ISO 13485 certified quality system.

    AI/ML Overview

    The provided text is a Special 510(k) Summary for the Esaote 6200 Ultrasound Imaging System. It describes modifications to an existing device (6100) and confirms that these modifications do not alter the intended use or fundamental scientific technology. The document focuses on demonstrating substantial equivalence to predicate devices and lists various clinical applications and operation modes for different transducers.

    However, the provided text does not contain any information about acceptance criteria, device performance studies, sample sizes for test or training sets, data provenance, expert ground truth, adjudication methods, multi-reader multi-case (MRMC) studies, or standalone algorithm performance.

    The tables within the document primarily indicate:

    • "P" (Previously cleared by FDA): This means the indication and mode of operation for a specific transducer were already approved based on prior submissions (K051837, K060827, K081386, or K091009). For these, new performance studies are generally not required to re-demonstrate safety and effectiveness unless the modification changes the fundamental scientific technology or intended use.
    • "N" (New indication): This indicates an indication not previously cleared for a specific transducer. For these, the applicant would typically provide data to support the new indication, but this data is not included in the provided text.
    • "E" (Added under Appendix): This also refers to expanded or new indications. Similar to "N," supporting data would be submitted, but it's not present here.

    Therefore, I cannot fulfill your request for an acceptance criteria table and the details of a study proving the device meets those criteria, as this information is not present in the provided document. The 510(k) summary focuses on justifying substantial equivalence through a comparison of technological characteristics and intended uses rather than presenting new performance data for device modifications that are deemed not to significantly affect safety or effectiveness.