Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K022373

    Validate with FDA (Live)

    Date Cleared
    2002-10-07

    (77 days)

    Product Code
    Regulation Number
    886.5844
    Reference & Predicate Devices
    Predicate For
    N/A
    Why did this record match?
    Reference Devices :

    K012132, K994320

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not-aphakic persons.

    Prescribed as a colored filter to aid individuals who experience reading discomfort not related to binocular vision problems or uncorrected refractive error.

    Prescribed to enhance color discrimination in patients with protan or deutan (red-green) color vision deficiencies.

    Device Description

    The ChromaGen v3.0 Reading Aid & Color Discrimination Enhancement, Spectacle Lenses are prescription spectacles that have been precision coated with varying hue and saturation filters similar to antireflection coatings and colored prescription sunglass coatings that are already included in the prescription spectacle lens classification. ChromaGen v30 Reading Aid & Color Discrimination Enhancement are available in varying hues and saturation levels duplicate of that available in the ChromaGen v30 Reading Aid & Color Discrimination Enhancement, Soft Contact Lenses. The ChromaGen v30 Reading Aid & Color Discrimination Enhancement, Spectacle Lens has been designed for binocular use, and when used in combination, have been shown to be of use for patients experiencing visual discomfort when reading.

    The ChromaGen v3.0 Reading Aid & Color Discrimination Enhancement, Spectacle Lens coating/tinting process does not alter the optical and/or performance characteristics of the finished spectacle lens.

    AI/ML Overview

    Here's an analysis of the provided text regarding the ChromaGen v3.0 Reading Aid & Color Discrimination Enhancement Spectacle Lenses, focusing on acceptance criteria and supporting studies:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not explicitly state numerical acceptance criteria for the ChromaGen v3.0 Spectacle Lenses. Instead, the basis for clearance is substantial equivalence to previously cleared predicate devices (ChromaGen v3.0 Reading Aid Soft Contact Lenses (K012132) and ChromaGen v2.0 Color Discrimination Enhancement Soft Contact Lenses (K994320)).

    The reported device performance, in this context, is that the ChromaGen v3.0 Spectacle Lenses achieve "Substantially Equivalent" efficacy to the predicate soft contact lenses.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Explicitly Stated in Document)Reported Device Performance (as stated in Document)
    Efficacy for Reading AidNot explicitly stated as a quantifiable metric for the spectacle lenses. Implied to be equivalent to predicate."Substantially Equivalent, Clinical study (reference K994320 & K012132)"
    Efficacy for Color Discrimination EnhancementNot explicitly stated as a quantifiable metric for the spectacle lenses. Implied to be equivalent to predicate."Substantially Equivalent, Clinical study (reference K994320 & K012132)"
    Prescription Spectacle Lens ClassificationAdherence to Class I regulatory requirements.Classified as Class I.
    Manufacturing ProcessSpectacle light transmittance identical to predicate contact lenses, and no alteration of optical/performance characteristics."coated/tinted to the exact same spectacle light transmittance as the predicate device soft contact lens." "does not alter the optical and/or performance characteristics of the finished spectacle lens."

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: No new clinical studies were conducted for the ChromaGen v3.0 Spectacle Lenses directly. The submission relies on the efficacy established in the predicate contact lens studies (K994320 & K012132). Therefore, the sample size for a new test set for these spectacle lenses is 0.
    • Data Provenance: Not specified within this document for the original predicate studies. However, the applicant is based in England. The 510(k) is submitted to the US FDA. We can infer the predicate studies (K994320 & K012132) were likely either performed in the UK or internationally, and were considered acceptable by the FDA for the previous contact lens clearances. This is a retrospective reliance on previously established data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Since no new studies were conducted, there were no new experts specifically used to establish ground truth for a new test set for these spectacle lenses. The ground truth for the predicate contact lens studies is not detailed in this 510(k) summary.

    4. Adjudication Method for the Test Set

    • Not applicable, as no new test set was created or adjudicated for these spectacle lenses.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, an MRMC comparative effectiveness study was not done for the ChromaGen v3.0 Spectacle Lenses. The document explicitly states: "Clinical performance studies are not required for the spectacle version of the ChromaGen v3.0 Reading Aid & Color Discrimination Enhancement product line, as efficacy has been established in K994320 & K012132."
    • Effect size of human readers improvement with AI vs. without AI assistance: Not applicable as this is not an AI device, and no MRMC study was performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was done

    • Not applicable. This device is a physical spectacle lens with colored filters and is not an algorithm or AI.

    7. The Type of Ground Truth Used

    • The ground truth for the efficacy of the predicate contact lenses (which this device relies upon) was established via "Clinical study" as stated repeatedly in the table. The specific type of clinical outcome or assessment used to establish efficacy (e.g., improved reading speed, improved scores on color vision tests, patient-reported outcomes) is not detailed in this 510(k) summary but would have been part of the K994320 and K012132 submissions.

    8. The Sample Size for the Training Set

    • Not applicable. This device does not involve a "training set" in the context of machine learning or AI algorithms. It is a physical medical device.

    9. How the Ground Truth for the Training Set was Established

    • Not applicable. As stated above, this device does not utilize a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K012132

    Validate with FDA (Live)

    Date Cleared
    2002-05-09

    (304 days)

    Product Code
    Regulation Number
    886.5925
    Reference & Predicate Devices
    Predicate For
    Why did this record match?
    Reference Devices :

    K990757, K994320

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ChromaGen v3.0 Reading Aid Soft Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not-aphakic persons with non-diseased eyes.

    Theses lenses may also be prescribed as a colored filter to aid individuals who experience reading discomfort not related to binocular vision problems or uncorrected refractive error. The lenses are disinfected using a hydrogen peroxide lens care system only and are available in a frequent replacement program.

    Device Description

    The ChromaGen v3.0 Reading Aid, Soft Contact Lenses are a range of soft lenses with precision tinted I ho on onlying hue and saturation which, when used in combination, have been shown to be of use for patients experiencing visual discomfort when reading. The ChromaGen v3.0 Reading Aid, Soft Contact Lenses are tinted with FDA "listed" color additives. The color additives are used in amounts not to exceed the minimum reasonably required to accomplish the intended coloring effect. As part of the encoufacturing process, the lens containing the color additives are thoroughly washed to remove unbound reactive color additives. The manufacturing process alters and/or changes the specifications to the clear version of a contact lens by affixing a listed reactive color additive on that portion of the anterior (front) surface of the lens that corresponds to the iris.

    The ChromaGen color additive effect is formed by reacting one or more of the reactive color additives listed in this paragraph with (poly hydroxyethyl methacrylate). The reactive color additives that may be used in and paragraph with (pation are: reactive black 5, reactive blue 21, reactive blue 19, reactive blue 4, reactive blue 163, reactive red 11, reactive red 180, reactive yellow 15, reactive yellow 86, or reactive orange 78. The color additives used are not removed by lens handling or approved cleaning/disinfecting procedures. The ChromaGen v3.0 Reading Aid, Soft Contact Lenses tinting process does not alter the optical and/or performance characteristics of the finished tinted soft contact lens.

    AI/ML Overview

    Acceptance Criteria and Study for ChromaGen v3.0 Reading Aid Soft Contact Lens

    This response details the acceptance criteria and the study conducted to demonstrate the effectiveness of the ChromaGen v3.0 Reading Aid Soft Contact Lens, based on the provided 510(k) summary.


    1. Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device PerformanceMetric
    Subjective preference for ChromaGen lenses over placebo lenses72.5% of subjects preferred ChromaGen lenses to placebo lenses (29 out of 40 subjects who expressed an opinion)Percentage
    Improved "ease of reading" with ChromaGen lenses compared to placebo lensesMedian "ease of reading" grading was 7.0 for ChromaGen lenses vs. 4.5 for placebo lenses (on a scale from -10 to +10)Median Score
    Statistical significance of improved "ease of reading"p-value < 0.001 (Wilcoxon analysis)p-value

    2. Sample Size and Data Provenance

    • Sample Size for Test Set: 47 subjects completed the study. (53 subjects initially entered, but 6 were excluded from the analysis for not providing an opinion on lens preference)
    • Data Provenance: The document does not explicitly state the country of origin. The study was conducted by Cantor & Nissel Limited, an English company, implying the data likely originated from the UK. The study was prospective, as it involved recruiting subjects and conducting new tests.

    3. Number and Qualifications of Experts for Ground Truth

    • The study did not use experts to establish ground truth in the traditional sense of diagnostic accuracy.
    • The ground truth was based on the subjective experience of the participants, who were asked to grade their own "ease of reading" and express their preference between lenses.

    4. Adjudication Method

    • None. There was no adjudication method described. The outcome was based on individual subjective reports from the participants.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This study focused on the subjective experience of individual users with and without the device.

    6. Standalone (Algorithm Only) Performance Study

    • Not applicable. The ChromaGen v3.0 Reading Aid Soft Contact Lens is a physical device (tinted contact lens), not an algorithm or AI system. Therefore, a standalone algorithm performance study was not performed.

    7. Type of Ground Truth Used

    • Subjective Patient Self-Report / Patient Reported Outcomes (PROs): The ground truth was based on the participants' subjective grading of "ease of reading" and their expressed preference between the ChromaGen lenses and placebo lenses.

    8. Sample Size for Training Set

    • Not applicable. This device is a physical product (tinted contact lens) and does not involve a "training set" in the context of machine learning or algorithms. The clinical study evaluated the device's performance in a test set of patients.

    9. How Ground Truth for Training Set Was Established

    • Not applicable, as there was no training set for this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1